(112 days)
No
The device description and performance studies focus on mechanical properties and MRI compatibility of bone plates and screws, with no mention of AI or ML.
Yes
The device is indicated for treating fractures, non-unions, deformities, and malunions, which are conditions requiring therapeutic intervention.
No
Explanation: The device is described as an internal fixation device (plates, screws, washers) used for treating fractures and other conditions of the pelvic ring and acetabulum. Its intended use focuses on surgical repair and stabilization rather than diagnosis. The mention of "Image Artifacts per ASTM F 2119" refers to testing compatibility with MRI imaging, not that the device itself performs image processing for diagnostic purposes.
No
The device description explicitly states it consists of plates, screws, and washers, which are hardware components for internal fixation.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use/Indications for Use: The intended use clearly describes the device as being used for the internal fixation of fractures, non-unions, deformities, and malunions of the pelvic ring and acetabulum, as well as sacroiliac joint dislocations. These are all surgical procedures performed on the body, not tests performed on samples taken from the body.
- Device Description: The device is described as an internal fixation device consisting of plates and screws. This is consistent with surgical implants, not diagnostic tests.
- Performance Studies: The performance studies mentioned are related to mechanical testing (bending, fatigue) and compatibility with MRI, which are relevant for surgical implants. There are no mentions of studies related to analyzing biological samples or diagnosing conditions based on such analysis.
- Lack of IVD Characteristics: There is no mention of the device being used to examine specimens derived from the human body (blood, tissue, urine, etc.) to provide information for the diagnosis, prevention, monitoring, treatment, or alleviation of disease.
In summary, the device is a surgical implant used for internal fixation of the pelvis, which falls under the category of medical devices used in vivo (within the body), not in vitro (in glass/outside the body).
N/A
Intended Use / Indications for Use
The PRO Plating System is indicated for:
· Fractures, non-unions, deformities and malunions of the pelvic ring and acetabulum
· Sacroiliac joint dislocations
The Stryker Trauma Pelvic Set (Matta) is indicated for:
- · Fractures of the acetabulum, sacrum, illium, and entire pelvic ring;
- · Revision surgery of pseudarthrosis, non-unions and mal-unions;
- · Osteonomies:
- · Arthrodeses;
- · Sacroilic joint dislocations;
- Symphysis pubis disruptions
Product codes
HRS, HWC
Device Description
This Traditional 510(k) submission is being supplied to the U.S. FDA to gain clearance to market the new devices of the PRO Plating System, to align indication of the existing PRO Pelvis System and the new devices and add MRI labeling to the existing PRO Pelvis System and the Trauma Pelvic Set System (Matta). This submission encompasses multiple systems (PRO Pelvis System and Trauma Pelvic Set System (Matta)) that have similar intended use and/or will be used together during the surgical procedure.
The PRO Plating System (Pelvis II Implant System) is an internal fixation device that consists of different types of plates used with compatible screws to fit different types of fractures in the pelvis. All devices in the system are provided sterile and non-sterile.
The Stryker Trauma Pelvic Set System (Matta) consists of plate, screw, and washer components of various sizes. All are manufactured from stainless steel (ASTM F-138/139). All devices in the system are provided sterile and non-sterile.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Pelvic ring, acetabulum, sacroiliac joint, sacrum, illium, symphysis pubis
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-Clinical Performance and Conclusions:
The following non-clinical laboratory testing, and performance assessments were made in support of substantial equivalence:
• Single Cycle Bend and Bending Fatigue Testing per ASTM F382
Tests performed to establish compatibility with a magnetic resonance environment:
• Magnetically Induced Displacement per ASTM F2052
• Magnetically Induced Torque per ASTM F2213
• RF Heating per ASTM F2182
• Image Artifacts per ASTM F 2119
Clinical Performance and Conclusions:
Clinical data and conclusions were not needed for this device.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.
(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.
0
April 7, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Stryker GMbH Danielle Madureira Associate Manager Regulatory Affairs 325 Corporate Drive Mahwah. New Jersev 07430
Re: K223772
Trade/Device Name: PRO Plating System, Stryker Trauma Pelvic Set (Matta) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: December 16, 2022 Received: December 16, 2022
Dear Danielle Madureira:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
1
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K223772
Device Name PRO Plating System
Indications for Use (Describe)
The PRO Plating System is indicated for:
· Fractures, non-unions, deformities and malunions of the pelvic ring and acetabulum
· Sacroiliac joint dislocations
Type of Use (Select one or both, as applicable) | |
---|---|
☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Indications for Use
510(k) Number (if known) K223772
Device Name Stryker Trauma Pelvic Set (Matta)
Indications for Use (Describe)
The Stryker Trauma Pelvic Set (Matta) is indicated for:
- · Fractures of the acetabulum, sacrum, illium, and entire pelvic ring;
- · Revision surgery of pseudarthrosis, non-unions and mal-unions;
- · Osteonomies:
- · Arthrodeses;
- · Sacroilic joint dislocations;
- Symphysis pubis disruptions
Type of Use (Select one or both, as applicable) |
---|
------------------------------------------------- |
|X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4
510(k) Summary
| Proprietary Name: | PRO Plating System
Trauma Pelvic Set System (Matta) |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Plate, Fixation, Bone and Screw, Fixation, Bone |
| Regulation Description: | 21 CFR 888.3030: Single/Multiple component metallic bone fixation
appliances and accessories
21 CFR 888.3040: Smooth or threaded metallic bone fastener |
| Regulation Number: | 21 CFR 888.3040 (primary), 21 CFR 888.3030 |
| Classification Product Code: | HRS, HWC |
| Device Class: | II |
| Sponsor: | Stryker GMBH
Bohnackerweg 1
2545 Selzach, Switzerland |
| Contact Person: | Danielle Jannuzzi Madureira, PhD
Associate Manager Regulatory Affairs
Dr. Homer Stryker Strasse, 1
CH-2545 Selzach, Switzerland
Phone: +41 79 890 02 89
Fax: +41 32 641 66 60 |
| Date: | December 05, 2022 |
| Primary Predicate: | PRO Plating System K131132 |
| Additional Predicate: | Trauma Pelvic Set System (Matta) K001614 |
| Reference Device: | aap Bone Plates and Screw Implants K072411 |
| Device Description: | This Traditional 510(k) submission is being supplied to the U.S. FDA to
gain clearance to market the new devices of the PRO Plating System, to
align indication of the existing PRO Pelvis System and the new devices and
add MRI labeling to the existing PRO Pelvis System and the Trauma Pelvic
Set System (Matta). This submission encompasses multiple systems (PRO
Pelvis System and Trauma Pelvic Set System (Matta)) that have similar
intended use and/or will be used together during the surgical procedure.
The PRO Plating System (Pelvis II Implant System) is an internal fixation
device that consists of different types of plates used with compatible screws
to fit different types of fractures in the pelvis. All devices in the system are
provided sterile and non-sterile.
The Stryker Trauma Pelvic Set System (Matta) consists of plate, screw, and
washer components of various sizes. All are manufactured from stainless
steel (ASTM F-138/139). All devices in the system are provided sterile and
non-sterile. |
| Indications for Use: | The PRO Plating System is indicated for:
Fractures, non-unions, deformities and malunions of the pelvic ring
● |
| | and acetabulum
• Sacroiliac joint dislocations
The Stryker Trauma Pelvic Set (Matta) is indicated for:
• Fractures of the acetabulum, sacrum, illium, and entire pelvic ring;
• Revision surgery of pseudarthrosis, non-unions and mal-unions;
• Osteonomies;
• Arthrodeses;
• Sacroilic joint dislocations;
• Symphysis pubis disruptions |
| Comparison to Predicate
Device: | A comparison of the system demonstrated that the subject PRO Plating
System is substantially equivalent to the PRO Plating System, Stryker
Trauma Pelvic Set (Matta) and aap Bone Plates regarding intended use,
material, design, and operational principles.
A comparison of the system demonstrated that the subject Stryker Trauma
Pelvic Set (Matta) is substantially equivalent to the Stryker Trauma Pelvic
Set (Matta) regarding intended use, material, design, and operational
principles. |
| Performance Data (Nonclinical): | Non-Clinical Performance and Conclusions:
The following non-clinical laboratory testing, and performance assessments
were made in support of substantial equivalence:
• Single Cycle Bend and Bending Fatigue Testing per ASTM F382
Tests performed to establish compatibility with a magnetic resonance
environment:
• Magnetically Induced Displacement per ASTM F2052
• Magnetically Induced Torque per ASTM F2213
• RF Heating per ASTM F2182
• Image Artifacts per ASTM F 2119
Clinical Performance and Conclusions:
Clinical data and conclusions were not needed for this device. |
| Conclusion: | Except for the modifications described in this submission the subject
devices are identical to the predicate device, and the performance data and
analyses demonstrate that any differences do not raise new questions of
safety and effectiveness as established with performance testing and
therefore, the subject devices PRO Plating System and Stryker Trauma
Pelvic Set (Matta) is substantially equivalent to the previously cleared
predicate device PRO Plating System and Stryker Trauma Pelvic Set
(Matta) |
5