K Number
K223726
Date Cleared
2023-03-07

(84 days)

Product Code
Regulation Number
892.1750
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is indicated to acquire and display cross-sectional volumes of the whole the head. The Aquilion Precision has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.

FIRST is an iterative reconstruction algorithm intended to reduce exposure dose and improve high contrast spatial resolution for abdomen, pelvis, chest, cardiac, extremities, and head applications.

AiCE is a noise reduction algorithm that improves image quality and reduces image noise by employing Deep Convolutional Network methods for abdomen, pelvis, lung, cardiac, extremities, head, and inner ear applications.

Device Description

Aquilion Precision (TSX-304A/4) V10.14 with AiCE is an ultra-high resolution whole body multislice helical CT scanner, consisting of a gantry, couch and a console used for data processing and display. Aquilion Precision incorporates a 160-row, 0.25 mm detector, a 5.7- MHU large-capacity tube, and 0.35 s scanning, enabling wide-range scanning with short scan times to capture cross sectional volume data sets used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician. In addition, the subject device incorporates the latest reconstruction technology, FIRST, intended to reduce exposure dose while maintaining and/or improving image quality as well as, AiCE (Advanced intelligent Clear-IQ Engine), intended to reduce image noise and improve image quality by utilizing Deep Convolutional Neural Network methods to 1024x1024 HR/SHR images. These methods can more fully explore the statistical properties of the signal and noise. By learning to differentiate structure from noise, the algorithm produces fast, high quality CT reconstruction.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for the Aquilion Precision (TSX-304A/4) V10.14 with AiCE:

Context: The submission is for a modification to an existing device, where the AiCE (Advanced intelligent Clear-IQ Engine) noise reduction algorithm is expanded to new anatomical regions: extremities, head (brain CTA), and inner ear. Therefore, the focus of the performance testing is to demonstrate sufficient image quality and potential improvements in spatial resolution in these new applications.


1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly list "acceptance criteria" as pass/fail thresholds with specific numerical values for each image quality metric. Instead, the performance testing aims to demonstrate equivalence to the predicate device and highlight quantitative improvements where applicable. The stated performance demonstrates that "AiCE is substantially equivalent to the predicate device as demonstrated by the results of the above testing" and that "the reconstructed images using the subject device were of diagnostic quality."

Implicit Acceptance Criteria & Reported Performance:

Image Quality Metric / CapabilityImplicit Acceptance Criteria (Demonstrated)Reported Device Performance
General Image Quality (across all applications)- Diagnostic quality of reconstructed images. - Equivalence to predicate device in previously cleared applications."reconstructed images using the subject device were of diagnostic quality." "AiCE is substantially equivalent to the predicate device" (based on Phantom Test Results)
Contrast-to-Noise Ratios (CNR)Maintained or improved compared to predicate/baseline.Assessed (specific values not provided, but deemed "substantially equivalent").
CT Number AccuracyMaintained accuracy.Assessed (specific values not provided, but deemed "substantially equivalent").
UniformityMaintained uniformity.Assessed (specific values not provided, but deemed "substantially equivalent").
Slice Sensitivity Profile (SSPz)Maintained or improved profile.Assessed (specific values not provided, but deemed "substantially equivalent").
Modulation Transfer Function (MTF)-WireMaintained or improved spatial resolution.Assessed (specific values not provided, but deemed "substantially equivalent").
Modulation Transfer Function (MTF)-EdgeMaintained or improved spatial resolution.Assessed (specific values not provided, but deemed "substantially equivalent").
Standard Deviation of Noise (SD)Maintained or reduced noise.Assessed (specific values not provided, but deemed "substantially equivalent").
Noise Power Spectra (NPS)Acceptable noise characteristics.Assessed (specific values not provided, but deemed "substantially equivalent").
Low Contrast Detectability (LCD)Maintained or improved detectability.Assessed (specific values not provided, but deemed "substantially equivalent").
Pediatric conditionsAcceptable performance for pediatric imaging.Assessed (specific values not provided, but deemed "substantially equivalent").
High Contrast Spatial Resolution (Quantitative Improvement)Demonstrable improvement in specific applications for newly added AiCE regions.Bone in HR mode: 5.37 lp/cm improvement (at 10% MTF and same dose). Inner Ear in HR mode: 7.50 lp/cm improvement (at 10% MTF and same dose). Brain CTA in HR mode: 11.31 lp/cm improvement (at 10% MTF and same dose).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document refers to "Representative Abdomen Bone, Brain CTA, and Inner Ear images." This implies a qualitative selection rather than a large, statistically defined sample size of clinical cases. The quantitative spatial resolution improvement was likely assessed using phantoms, not clinical images, which is common for such technical measurements.
  • Data Provenance: Not explicitly stated regarding country of origin. The images were "obtained using the subject device," implying prospective acquisition for the purpose of this evaluation, but it's not explicitly stated if they were newly acquired for this submission or if they were existing images. The evaluation states "reviewed by an American Board Certified Radiologist," which might suggest clinical images from the US, but this is not definitive.
  • Retrospective or Prospective: Unspecified, but the phrasing "obtained using the subject device" suggests they could be prospectively acquired.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: One.
  • Qualifications of Experts: "an American Board Certified Radiologist." No specific years of experience are provided.

4. Adjudication Method for the Test Set

No explicit adjudication method is described. The single radiologist reviewed the images and "it was confirmed that the reconstructed images using the subject device were of diagnostic quality." This implies a sole reader assessment for the qualitative clinical image evaluation.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not reported in this summary. The clinical evaluation was a single radiologist's qualitative assessment of diagnostic quality.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance assessment was conducted for objective image quality metrics using phantoms. This includes:

  • Contrast-to-Noise Ratios (CNR)
  • CT Number Accuracy
  • Uniformity
  • Slice Sensitivity Profile (SSPz)
  • Modulation Transfer Function (MTF)-Wire
  • Modulation Transfer Function (MTF)-Edge
  • Standard Deviation of Noise (SD)
  • Noise Power Spectra (NPS)
  • Low Contrast Detectability (LCD)
  • Pediatric conditions

The quantitative spatial resolution improvement claims (5.37 lp/cm, 7.50 lp/cm, 11.31 lp/cm) are also results of standalone (algorithm only) performance tests, likely performed on phantoms.


7. The Type of Ground Truth Used

  • For objective image quality metrics (phantoms): The ground truth is the known physical properties and measurements derived from the phantoms themselves (e.g., known dimensions for MTF, known densities for CT number accuracy).
  • For clinical image quality: The ground truth for "diagnostic quality" was established by expert consensus / opinion from a single "American Board Certified Radiologist." There's no mention of pathology or outcomes data being used as ground truth for this aspect.

8. The Sample Size for the Training Set

The document does not provide any information about the sample size (or any details) for the training set used for the AiCE Deep Convolutional Network. This information is typically not included in FDA 510(k) summaries, as the focus is on validation and verification rather than the internal development process of the AI model.


9. How the Ground Truth for the Training Set was Established

As no information is provided about the training set, there is also no information on how its ground truth was established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 7, 2023

Canon Medical Systems Corporation % Orlando Tadeo Sr. Manager, Regulatory Affairs Canon Medical Systems USA 2441 Michelle Drive TUSTIN, CA 92780

Re: K223726

Trade/Device Name: Aquilion Precision (TSX-304A/4) V10.14 with AiCE Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: February 14, 2023 Received: February 15, 2023

Dear Orlando Tadeo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K223726

Device Name

Aquilion Precision (TSX-304A/4) V10.14 with AiCE

Indications for Use (Describe)

This device is indicated to acquire and display cross-sectional volumes of the whole the head. The Aquilion Precision has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.

FIRST is an iterative reconstruction algorithm intended to reduce exposure dose and improve high contrast spatial resolution for abdomen, pelvis, chest, cardiac, extremities, and head applications.

AiCE is a noise reduction algorithm that improves image quality and reduces image noise by employing Deep Convolutional Network methods for abdomen, pelvis, lung, cardiac, extremities, head, and inner ear applications.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

  1. SUBMITTER'S NAME: Fumiaki Teshima Senior Manager, Quality Assurance Department Canon Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
    1. ESTABLISHMENT REGISTRATION: 9614698

3. OFFICIAL CORRESPONDENT/CONTACT PERSON:

Orlando Tadeo, Jr. Sr. Manager, Regulatory Affairs Canon Medical Systems USA, Inc 2441 Michelle Drive Tustin, CA 92780 (714) 669-7459

    1. DATE PREPARED: November 29, 2022
  • TRADE NAME(S): Aquilion Precision (TSX-304A/4) V10.14 with AiCE

COMMON NAME: 6.

Computed Tomography X-ray System

7. DEVICE CLASSIFICATION:

a) Classification Name: Computed Tomography X-ray system b) Regulation Number: 21 CFR §892.1750 c) Regulatory Class: Class II

8. PRODUCT CODE:

JAK

9. PERFORMANCE STANDARD:

This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR, Subchapter J, Part 1020]

PHONE: 800-421-1968 2441 Michelle Drive, Tustin, CA 92780

{4}------------------------------------------------

10. PREDICATE DEVICE:

ProductMarketed byRegulation NumberRegulation NameProduct Code510(k) NumberClearance Date
Aquilion Precision(TSX-304A/4) V10.10with AiCECanonMedicalSystems, USA21 CFR§892.1750ComputedTomographyX-ray SystemJAK:System, X-ray,Tomography,ComputedK220986September12, 2022

11. REASON FOR SUBMISSION:

Modification of existing medical device

12. DEVICE DESCRIPTION:

Aquilion Precision (TSX-304A/4) V10.14 with AiCE is an ultra-high resolution whole body multislice helical CT scanner, consisting of a gantry, couch and a console used for data processing and display. Aquilion Precision incorporates a 160-row, 0.25 mm detector, a 5.7- MHU large-capacity tube, and 0.35 s scanning, enabling wide-range scanning with short scan times to capture cross sectional volume data sets used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician. In addition, the subject device incorporates the latest reconstruction technology, FIRST, intended to reduce exposure dose while maintaining and/or improving image quality as well as, AiCE (Advanced intelligent Clear-IQ Engine), intended to reduce image noise and improve image quality by utilizing Deep Convolutional Neural Network methods to 1024x1024 HR/SHR images. These methods can more fully explore the statistical properties of the signal and noise. By learning to differentiate structure from noise, the algorithm produces fast, high quality CT reconstruction.

13. INDICATIONS FOR USE:

This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head.

The Aquilion Precision has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software, by a trained and qualified physician.

FIRST is an iterative reconstruction algorithm intended to reduce exposure dose and improve high contrast spatial resolution for abdomen, pelvis, chest, cardiac, extremities, and head applications.

AiCE is a noise reduction algorithm that improves image quality and reduces image noise by employing Deep Convolutional Network methods for abdomen, pelvis, lung, cardiac, extremities, head, and inner ear applications.

14. SUBSTANTIAL EQUIVALENCE:

The Aquilion Precision (TSX-304A/4) V10.14 with AiCE is substantially equivalent to Aquillion Precision (TSX-304A/4) V10.10 with AiCE, which received premarket clearance under K220986, and is marketed by Canon Medical Systems USA. The intended use of the Aquilion Precision is the same as that of the predicate device. The Aquilion Precision (TSX-304A/4) V10.14 with AiCE includes changes made to the predicate device that expand the use of AiCE to extremities, head, and inner ear applications. A comparison of the technological characteristics between the subject and the predicate device is included below.

{5}------------------------------------------------

Subject DevicePredicate DeviceComment
Device Name,Model NumberAquilion Precision (TSX-304A/4) V10.14 with AiCEAquilion Precision (TSX-304A/4) V10.10 with AiCE
510(k) NumberThis submissionK220986
AiCE Modifications
• Scan RegionsAbdomen and Pelvis/Chest/Cardiac/Extremities*/Brain*/Inner Ear*Abdomen and Pelvis/Chest/Cardiac*Added
• Magnifiedreconstruction -AvailablemagnificationSize (D-FOV)BODY / BODY SHARP:Min. 100 mmLUNG:Min. 100 mmCARDIAC:Min. 70 mmBONE / INNER EAR:Min. 50 mmBRAIN CTA:Min. 100 mmMax. 500 mmBODY / BODY SHARP:Min. 100 mmLUNG:Min. 100 mmCARDIAC:Min. 70 mmMax. 500 mmThe minimum FOVdiffers dependingon the anatomicalregion.
• Reconstructionprocessingsystem [AiCE](CSAL-002A)Option[Reconstruction ProcessingUnit required]N/A[AiCE implemented withReconstruction ProcessingUnit (CCRS-003B)]
• Applicableanatomicalregions• BODY and BODY SHARP(For Abdomen andPelvis)• LUNG (For Chest)• CARDIAC (For Cardiac)• BONE (For extremities)• BRAIN CTA (For brain)• INNER EAR (For innerear)*• BODY and BODY SHARP(For Abdomen andPelvis)• LUNG (For Chest)• CARDIAC (For Cardiac)*Added

{6}------------------------------------------------

15. SAFETY:

The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the following standards IEC60601-1, IEC60601-1-3, IEC60601-1-6, IEC60601-1-6, IEC60601-1-9, IEC60601-2-28, IEC60601-2-44, IEC60825-1, IEC62304, IEC62366, NEMA XR-26 and NEMA XR-29. Additionally, this device complies with all applicable requirements of the radiation safety performance standards, as outlined in 21 CFR §1010 and §1020.

This device conforms to applicable Performance Standards for lonizing Radiation Emitting Products [21 CFR, Subchapter J, Part 1020]

16. TESTING

Risk analysis and verification/validation activities conducted through bench testing demonstrate that the established specifications for the device have been met.

Performance Testing - Bench

Image Quality Evaluation

CT image quality metrics were performed, utilizing phantoms, to assess Contrast-to-Noise Ratios (CNR), CT Number Accuracy, Uniformity, Slice Sensitivity Profile (SSPz), Modulation Transfer Function (MTF)-Wire, Modulation Transfer Function (MTF)-Edge, Standard Deviation of Noise (SD), Noise Power Spectra (NPS), Low Contrast Detectability (LCD) and Pediatric conditions. AiCE is substantially equivalent to the predicate device as demonstrated by the results of the above testing.

Quantitative Spatial Resolution Improvement

A spatial resolution comparison study was conducted to support a high contrast spatial resolution improvement claim of 5.37 lp/cm improvement for Bone in HR mode, 7.50 lp/cm improvement for Inner Ear in HR mode, and 11.31 lp/cm improvement for Brain CTA in HR mode, all compared to hybrid iterative reconstruction at 10% MTF and at the same dose.

Performance Testing – Clinical Images

Representative Abdomen Bone, Brain CTA, and Inner Ear images, reviewed by an American Board Certified Radiologist, were obtained using the subject device and it was confirmed that the reconstructed images using the subject device were of diagnostic quality.

A summary of the risk analysis and verification/validation testing conducted through bench and clinical testing is included in this submission which demonstrates that the requirements for the system have been met.

Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is included in this documentation includes justification for the Moderate Level of Concern determination as well as testing which demonstrates that the verification and validation requirements for the modifications described above have been met.

Cybersecurity documentation, per the FDA cybersecurity premarket guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" issued on October 2, 2014, is also included as part of this submission.

{7}------------------------------------------------

17. CONCLUSION

The Aquilion Precision (TSX-304A/4) V10.14 with AiCE performs in a manner similar to and is intended for the same use as the predicate device, as indicated in product labeling. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this submission it is concluded that the subject device has demonstrated substantial equivalence to the predicate device and is safe and effective for its intended use.

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.