(26 days)
The SmoothCool HR System is indicated for hair removal (permanent hair reduction).
The SmoothCool HR System is an Intense Pulsed Light (IPL) system used hair removal (Permanent hair reduction) in the area of dermatology. The system consists of a console containing the power unit and control electronics with control and display panel including software. Applicators/hand pieces are connected to the system in order to generate light energy for treatment in the waveband 420-950 nm. Four different hand pieces are attached with the system.
The provided text describes a 510(k) premarket notification for the "SmoothCool HR System," an Intense Pulsed Light (IPL) system for hair removal. However, it does not contain any information regarding acceptance criteria, device performance metrics, or any study involving a test set, expert adjudication, or AI assistance. The document focuses on establishing substantial equivalence to a predicate device based on technical characteristics and modifications (adding more filters).
Therefore, I cannot provide the requested information about acceptance criteria and the study proving the device meets them from the given text. The text only mentions "risk analysis and verification/validation tests" and lists "Energy Output Test" and "Button Operation Test" as performed, but does not provide specific criteria, results, or study details.
Here's what I can extract from the provided text, highlighting the absence of the requested information:
-
A table of acceptance criteria and the reported device performance:
- Information not provided. The document states that "risk analysis and verification/validation tests per modification" were performed, and "the test results support that the modifications do not raise a question in safety and performance." However, no specific acceptance criteria or performance metrics are detailed.
-
Sample size used for the test set and the data provenance:
- Information not provided. There is no mention of a test set, sample sizes, or data provenance (e.g., country of origin, retrospective/prospective study design).
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Information not provided. No experts or ground truth establishment for a test set are mentioned.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Information not provided. No adjudication method is mentioned as a test set or clinical study is not described.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Information not provided. This device is an IPL system for hair removal; it is not an AI-assisted diagnostic or imaging device, so an MRMC study with human readers and AI assistance would not be relevant.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Information not provided. This device is a physical IPL system, not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Information not provided. Ground truth is not discussed as no clinical study or diagnostic performance evaluation is described.
-
The sample size for the training set:
- Information not provided. Training sets are not relevant for this type of device and the information provided.
-
How the ground truth for the training set was established:
- Information not provided. Ground truth for a training set is not relevant here.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 4, 2023
Jeisys Medical Inc % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 18881 Von Karman Ave. STE 160 Irvine, California 92612
Re: K223685
Trade/Device Name: SmoothCool HR System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: December 7, 2022 Received: December 9, 2022
Dear Priscilla Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jianting Wang -S
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K223685
Device Name SmoothCool HR System
| Indications for Use (Describe) |
|---|
| --------------------------------------- |
The SmoothCool HR System is indicated for hair removal (permanent hair reduction).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) SUMMARY
(K223685)
Dec 30, 2022
1. Submitted by:
Jeisys Medical Inc 307, 308, 401, Daeryung Techno Town 8th, 96, Gamasan-ro, Geumcheon-Gu, Seoul, 08501, Republic of Korea
2. US Agent/ Official Correspondent:
Priscilla Chung
LK Consulting Group USA, Inc. 18881 Von Karman Ave STE 160 Irvine, CA 92612 Tel: 714.202.5789 Fax: 714.409.3357 Email: juhee.c@LKconsultingGroup.com
3. Device Name:
| - Trade Name | : SmoothCool HR System |
|---|---|
| - Classification | : Class II |
| - Classification Name | : Powered Laser Surgical Instrument |
| - Product Code | : GEX |
| - Regulation Number | : 21 CFR 878.4810 |
| - Review Panel | : General Hospital |
4. Predicate Device:
- SmoothCool HR System (K213484) by Jeisys Medical Inc.
5 Device Description:
The SmoothCool HR System is an Intense Pulsed Light (IPL) system used hair removal (Permanent hair reduction) in the area of dermatology. The system consists of a console containing the power unit and control electronics with control and display panel including
{4}------------------------------------------------
software. Applicators/hand pieces are connected to the system in order to generate light energy for treatment in the waveband 420-950 nm. Four different hand pieces are attached with the system.
6 Indications for Use Statement
The SmoothCool HR System is indicated for hair removal (permanent hair reduction).
7 Substantial Equivalence Discussion:
| Proprietary | Subject Device | Primary Predicate Device |
|---|---|---|
| Manufacturer | Jeisys Medical Inc. | Jeisys Medical Inc. |
| Device Name | SmoothCool HR System | SmoothCool HR System |
| 510(k)Number | K223685 | K213484 |
| DeviceClassificationName | Laser surgical instrument for use in general andplastic surgery and in dermatology | Laser surgical instrument for use in general andplastic surgery and in dermatology |
| Product Code | GEX | GEX |
| RegulationNumber | 878.4810 | 878.4810 |
| Intended Use | The SmoothCool HR System is indicated forhair removal (permanent hair reduction). | The SmoothCool HR System is indicated forhair removal (permanent hair reduction). |
| Components | • Main Console• Hand Pieces:420nm(S) Handpiece560nm(S) Handpiece700nm(S) Handpiece700nm(L) Handpiece• Filters800, 700, 640, 590, 560, 545(V), 530, 530(D),530(s), 500, 420(s)nm• Eyes Protector• Power Cord | • Main Console• Hand Pieces:420nm(S) Handpiece560nm(S) Handpiece700nm(S) Handpiece700nm(L) Handpiece• Filters800, 700, 560, 530, 530(s), 420(s)nm• Eyes Protector• Power Cord |
| Power Output | Up to 65 J/cm2 | Up to 65 J/cm2 |
| Sterile | No | No |
7.1.Comparison Chart
{5}------------------------------------------------
7.2. Substantial Equivalence Discussion
The indications for use and the technological characteristics of the subject device are the same as the predicate device, the unmodified device (K213484). The modification is adding more filters. The power output and also the range of the filter remained the same. We also performed risk analysis and verification/validation tests per modification, and the test results support that the modifications do not raise a question in safety and performance.
8. Technological Characteristics:
The SmoothCool HR System and the predicate devices in the market have the substantially equivalent technological characteristics. Risks associated with the changes were identified and appropriate design controls implemented to mitigate the risks.
To validate the control of the risks, the following tests were performed.
| No | Test |
|---|---|
| 1 | Energy Output Test |
| 2 | Button Operation Test |
9. Conclusion:
Based on the information provided in this special 510(k), the SmoothCool HR System is substantially equivalent to the predicate devices.
§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.
(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.