K Number
K213484

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-04-20

(173 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SmoothCool HR System is indicated for hair removal (permanent hair reduction).

Device Description

SmoothCool HR System is an Intense Pulsed Light (IPL) system used for hair removal (permanent hair reduction) in the area of dermatology. The system consists of a console containing the power unit and control electronics with control and display panel including software. Applicators/hand pieces are connected to the system in order to generate light energy for treatment. Four different hand pieces are offered with the system.

AI/ML Overview

The provided document is a 510(k) summary for the SmoothCool HR System, which is an Intense Pulsed Light (IPL) system for hair removal. It aims to demonstrate substantial equivalence to predicate devices, not to describe the acceptance criteria and study that proves the device meets those criteria in the way typically associated with AI/ML-based medical devices or diagnostic devices involving performance metrics like sensitivity, specificity, or AUC.

Therefore, the specific information requested in the prompt (e.g., acceptance criteria table, sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, ground truth type, training set sample size, how ground truth was established for training set) is not present in this document.

This document focuses on regulatory equivalence for a hardware device (IPL system) based on updated components and technical characteristics rather than a software algorithm's performance against clinical endpoints or expert consensus. The "tests" mentioned are for energy output and electrical safety/EMC, which are standard for such physical devices, not performance evaluation against clinical outcomes or a ground truth as would be relevant for devices that interpret data or diagnose conditions.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.

April 20, 2022

Jeisys Medical Inc % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 18881 Von Karman Ave. STE 160 Irvine, California 92612

Re: K213484

Trade/Device Name: SmoothCool HR System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: October 18, 2021 Received: October 29, 2021

Dear Priscilla Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K213484

Device Name SmoothCool HR System

Indications for Use (Describe)

The SmoothCool HR System is indicated for hair removal (permanent hair reduction).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

(K213484)

Apr 20, 2022

1. Submitted by:

Jeisys Medical Inc 307, 308, 401, Daeryung Techno Town 8th, 96, Gamasan-ro, Geumcheon-Gu, Seoul, 08501, Republic of Korea

2. US Agent/ Official Correspondent:

Priscilla Chung

LK Consulting Group USA, Inc. 18881 Von Karman Ave STE 160 Irvine, CA 92612 Fax: 714.409.3357 Tel: 714.202.5789 Email: juhee.c@LKconsultingGroup.com

3. Device Name:

  • Trade Name :: SmoothCool HR System
  • Classification Class II :
  • Classification Name Powered Laser Surgical Instrument :
  • Product Code GEX ::
  • Regulation Number 21 CFR 878.4810 :
  • Review Panel : General Hospital

4. Predicate Device:

  • SmoothCool (K082911) by Jeisys Medical Inc.
  • Stellar M22 for Intense Pulsed Light (IPL) and Laser system (K193500) by Lumenis Ltd.

{4}------------------------------------------------

Device Description: ഗ

SmoothCool HR System is an Intense Pulsed Light (IPL) system used for hair removal (permanent hair reduction) in the area of dermatology. The system consists of a console containing the power unit and control electronics with control and display panel including software. Applicators/hand pieces are connected to the system in order to generate light energy for treatment. Four different hand pieces are offered with the system.

Indications for Use Statement e

The SmoothCool HR System is indicated for hair removal (permanent hair reduction).

ProprietarySubject DevicePrimary Predicate DeviceReference Device• Main Console• Main Console• Main Console
ManufacturerJeisys Medical Inc.Jeisys Medical Inc.Lumenis Ltd.Components• Hand Pieces:420nm(S) Handpiece560nm(S) Handpiece700nm(S) Handpiece700nm(L) Handpiece• Filters800, 700, 560, 530, 420nm• Eyes Protector• Power Cord• Hand Pieces:560nm(S) Handpiece700nm(S) Handpiece700nm(L) Handpiece• Filters700, 560nm• Eyes Protector• Power Cord• IPL Handpiece400-1200 nm• ResurFX Handpiece1565 nm• Multi-Spot Nd:YAG Handpiece1064 nm• Q-Switched Nd:YAG Handpiece1064 nm
Device NameSmoothCoolSmoothCoolStellar M22 for Intense PulsedLight (IPL) and Laser systemPower OutputUp to 65 J/cm2Up to 65 J/cm2• IPL HandpieceUp to 56 J/cm2• Multi-Spot Nd:YAG Handpiece10225 J/cm2• Q-Switched Nd:YAG Handpiece0.914 J/cm2
510(k)Number-K082911K193500SoftwareSoftware ValidationSoftware ValidationSoftware Validation
DeviceClassificationNameLaser surgical instrument foruse in general and plasticsurgery and in dermatologyLaser surgical instrument for usein general and plastic surgery andin dermatologyLaser surgical instrument for use ingeneral and plastic surgery and indermatologySterileNoNoNo
Product CodeGEXGEXGEX
RegulationNumber878.4810878.4810878.4810
Intended UseThe SmoothCool HR System isindicated for hair removal(permanent hair reduction).The SmoothCool HR System isindicated for hair removal(permanent hair reduction).The subject Stellar M22 hasconnection capability with thefollowing available treatmenthandpieces, for multi-applicationtreatment options. All handpiecesare designed for aesthetic anddermatologicalskin procedure applications, asfollows:□ The Intense Pulsed Light (IPL)handpiece with a spectrum of400-1200 nm (with 9different filters) is indicated for:o Benign epidermal lesions,including dyschromia,hyperpigmentation, melasma, andephelides (freckles)o Cutaneous lesions, includingwarts, scars and striaeo Benign cutaneous vascular
lesions, including port wine stains,hemoangiomas, facial,truncal andleg telangiectasias, erythema ofrosacea, angiomas and spiderangiomas, poikiloderma of Civatte,leg veins and venousmalformationso Removal of unwanted hair and toeffect stable long term, orpermanent* hair reduction in skintypes I-V through selectivetargeting of melanin in hair follicleso Mild to moderate inflammatoryAcne (Acne vulgaris)□ The Nd:YAG Laser handpiecewith a wavelength of 1064 nm(Multi-Spot Nd:YAG) isindicated for:o The coagulation and hemostasisof vascular lesions and soft tissue,including the treatment andclearance of superficial and deeptelangiectasias (venulectasias) andreticular veins (0.1-4.0 mm.diameter) of the lego The removal of unwanted hairand to effect table long term, orpermanent* hair reduction in skintypes I-V through selectivetargeting of melanin in hair follicleso The non-ablative treatment offacial wrinkles□ ResurFX module andhandpiece, with wavelength of1565 nm, is indicated for:o Use in dermatological proceduresrequiring fractional skin resurfacingand coagulation of soft tissue□ The Q-Switched Nd:YAGLaser Handpiece with awavelength of 1064 nm isindicated for:o Removal of dark tattooso Treatment of pigmented lesions*NotePermanent hair reduction is definedas long-term, stable reduction in thenumber of hairs regrowing whenmeasured at 6, 9, and 12 monthsafter completion of treatmentregime.

Substantial Equivalence Discussion: 7

{5}------------------------------------------------

.

{6}------------------------------------------------

The indications for use and the technological characteristics of the subject device is the same as the predicate device, the unmodified device (K082911). Modifications are the addition of handpieces and filters. We identified a reference device which has the same indications for use (hair removal) and encompasses the wavelength range of the subject device.

We performed risk analysis and verification/validation tests per modifications, and the test results support that the modifications do not raise a question in safety and performance.

8. Technological Characteristics:

The SmoothCool HR System and the predicate devices in the market have the substantially equivalent technological characteristics. Risks associated with the changes were identified and appropriate design controls implemented to mitigate the risks.

To validate the control of the risks, the following tests were performed.

NoTestStandard
1Energy Output Test-
2Electrical Safety and EMCEN 60601-1EN 60601-1-2EN 60601-2-22

{7}------------------------------------------------

9. Conclusion:

Based on the information provided in this special 510(k), the SmoothCool HR System is substantially equivalent to the predicate devices.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.