(200 days)
No
The document describes a dental implant system and its components, focusing on materials, dimensions, surface treatments, and mechanical/biocompatibility testing. There is no mention of AI or ML technology in the intended use, device description, or performance studies.
No
The device, a dental implant system, aims to restore chewing function and support prosthetic devices, which are restorative rather than therapeutic functions.
No
Explanation: The provided text describes dental implants and their associated prosthetic components, used for restoring chewing function by providing support for artificial teeth. There is no mention of the device being used to diagnose a medical condition or disease.
No
The device description and intended use clearly describe physical implants and abutments made of titanium and titanium alloy, intended for surgical placement. There is no mention of software as the primary or sole component of the device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostic devices are used to examine specimens derived from the human body (like blood, urine, or tissue) to provide information for diagnosis, monitoring, or screening.
- Device Description and Intended Use: The descriptions and intended uses clearly state that these devices are surgically implanted into the jawbone to support prosthetic teeth and restore chewing function. They are physical implants and components used in vivo (within the body), not for testing samples in vitro (outside the body).
- Lack of IVD Characteristics: There is no mention of analyzing biological samples, providing diagnostic information, or any of the typical characteristics of an IVD.
Therefore, the Neodent Implant System and its components described are medical devices, but they do not fit the definition of an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
Indications for Use for Helix Short Implant:
The Neodent Implant System is recommended for surgical procedures on maxilla or mandible bones. It provides support for prosthetic components such as artificial teeth, thus restoring the chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Indications for Use for Mini Straight, Angled Abutment and screws:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Indications for Use for Helix Short Attachment:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Indications for Use for Helix Short Healing Abutment:
This product is used for the maintenance of soft tissue, during the bone integration stage of Neodent implants to be rehabilitated with the late loading technique. The Healing Abutment may be used in the implant installation surgery or the reopening surgery (second surgical stage).
Indications for Use for Helix Short Cover Screw:
This product is used for the maintenance of soft tissue, during the bone integration stage of Neodent implants to be rehabilitated with the late loading technique. The Cover Screw must remain intragingival, preventing the growth of tissue over the implant platform. The Cover Screw takes place in the conical region of the implant interface.
Indications for Use for Temporary Abutment for Helix Short Implant
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Temporary Abutments are installed over Helix Short Implants and to provide support for prosthetic structures for up to 6 months.
Indications for Use for Helix Short Titanium Base:
The Helix Short Implant Titanium base that is placed over Neodent dental implants to provide support for customized prosthetic restorations, such as copings and crowns. It is indicated for single- and multiple-structure restorations, screw- or cement-retained on implants installed in the maxilla or mandible. All digitally-designed copings and/or crowns to be used with the Neodent Titanium Base System must be sent to Straumann for manufacture at a validated milling center.
Product codes
DZE, NHA
Device Description
- . All the subject devices are intended for single use;
- The subject implants are provided sterile via gamma irradiation and the subject abutments are provided sterile via Ethylene Oxide to an SAL of 1x10-9;
- . The subject implants are manufactured of Commercially Pure Titanium (Grade 4), the subject abutments, prosthetic screws and abutment screws are manufactured of Titanium alloy Ti-6Al-4V ELI (ASTM F136) and the restorations placed onto Helix Short Titanium Bases are made of various top half materials.
- . The subject implants have a wide cone prosthetic interface with an internal hexagon for short implants lengths available from 4.0 to 7.0; cylindrical shape with double threads and conical apex with three helical flutes for diameters from 3.75 to 7.0mm.
- . The implants are provided in two different surfaces: Neoporos: Sand blasted and acid etched, and Acqua: Hydrophilic surface aggregated to the rough surface obtained by double treatment: abrasive blasting and acid subtraction.
- The Helix Short Implant Titanium Base abutments is used as two-piece abutment, where the base is premanufactured from titanium alloy and the top half is created via burn out coping or a digital workflow with CAD/CAM at a validated milling center. The final top half can be created from Ticon, Coron, Zerion LT, IPS e.max, or PMMA. The top half and base pieces are cemented together to form the final abutment.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
maxilla or mandible bones, upper or lower jaw
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Mechanical testing: Dynamic Fatigue Test per ISO 14801 was performed to determine the fatigue strength for the dental implant system, according to FDA Guidance. The tested subject devices exhibit a level of performance equivalent to that reviewed for the predicate devices. Torsion Test was performed to evaluate Helix Short Implant under static torsional loading. The results met the acceptance criteria. Insertion test was performed to evaluate the insertion torque of the Helix Short System when it is inserted in bones type I, II, IIII and IV. The Implant Surface Area simulation and Pull Out Test were also made to evaluate and represent the clinical use of the subject implants. Results demonstrated that the subject devices, despite having a slightly smaller surface area than the reference devices, presented a higher BIC rate and pull out resistance upon placement. Thus, the subject devices exhibit a level of benchtop performance equivalent to that reviewed for the reference device.
MR Compatibility testing: The MR compatibility was performed to access the risk of exposing patients who have implantable medical devices. An assessment was made to demonstrate that the subject devices do not configure a new worst case and can be represented by the previously conducted studies reviewed for reference devices, since both have the same raw material and similar dimensions. The subject devices are therefore MR conditional devices and a patient treated with the subject devices can be safely scanned observing the parameters previously established per reference devices.
Surface treatment: The surface treatments applied to subject devices are identical to these applied and previously evaluated for primary predicate devices. Although some reference devices have a surface with different nomenclature, the chemical processes applied to the surface are similar to create the roughness surface. Acqua implants are submitted to an additional step in order to increase their hydrophilicity, as already cleared for primary predicate device.
Sterilization validation: Sterilization of the subject implants via gamma irradiation according to ISO 11137-1 Sterilization of health care products -Radiation -Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices and ISO 11137-2 Sterilization of health care products -Radiation -Part 2: Establishing the sterilization dose. A minimum Sterility Assurance Level (SAL) of 1 x 10-6 has been validated.
Sterilization of the subject abutments via ethylene oxide gas using the overkill method has been performed according to the requirements of ISO 11135-1 Sterilization of Health Care Products - Ethylene Oxide - Part 1: Requirements For Development, Validation and Routine Control of a Sterilization Process For Medical Devices. A minimum Sterility Assurance Level (SAL) of 1 x 10-6 has been validated.
Endotoxin testing (LAL): The LAL Endotoxin Analysis is the method used to determine that the device meets pyrogen limit specifications, based in the referred FDA guidance (510(k) Submission for Devices Labeled as Sterile, issued on 21 January 2016). Routine endotoxin testing is performed with representative devices as an assessment of environmentally derived pyrogenicity. Subject devices meet the acceptance criteria for endotoxin levels as well as the primary predicate and reference devices. The subject devices are not represented to be "pyrogen free".
Shelf Life validation: The expiration date of the products was determined considering the integrity of the product and the packaging tests after shelf life testing. The packaging of the subject Helix Short Implant System is identical to the packaging of the primary predicate and reference devices. The shelf life for devices provided sterile is 5 years, except Helix Short Acqua Implants that have a shelf life of 4 years.
Biocompatibility: Representative samples of each of the subject devices was subjected to the following: Cytotoxicity testing was performed per ISO 10993-5. Chemical characterization was performed per ISO 10993-18. Biocompatibility sample preparation was performed per ISO 10993-12. Biological Safety Assessment guided by ISO 10993-1. EO Residuals evaluation was performed in accordance with 10993-7. No new issues of biocompatibility are raised for the subject devices when compared to primary predicate and reference devices. Therefore, no additional biocompatibility testing was required.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
K200586, K123022, K202942, K203309, K220251, K182620
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 872.3640 Endosseous dental implant.
(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
JJGC Industria e Comercio de Materiais Dentarios S.A. % Jennifer Jackson Sr. Director, Regulatory Affairs and Quality Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810
Re: K223638
Trade/Device Name: Neodent Implant System - Helix Short Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: May 17, 2023 Received: May 25, 2023
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K223638
Device Name
Neodent Implant System - Helix Short Implant System
Indications for Use (Describe)
Indications for Use for Helix Short Implant:
The Neodent Implant System is recommended for surgical procedures on maxilla or mandible bones. It provides support for prosthetic components such as artificial teeth, thus restoring the chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Indications for Use for Mini Straight, Angled Abutment and screws:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Indications for Use for Helix Short Attachment:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Indications for Use for Helix Short Healing Abutment:
This product is used for the maintenance of soft tissue, during the bone integration stage of Neodent implants to be rehabilitated with the late loading technique. The Healing Abutment may be used in the implant installation surgery or the reopening surgery (second surgical stage).
Indications for Use for Helix Short Cover Screw:
This product is used for the maintenance of soft tissue, during the bone integration stage of Neodent implants to be rehabilitated with the late loading technique. The Cover Screw must remain intragingival, preventing the growth of tissue over the implant platform. The Cover Screw takes place in the conical region of the implant interface.
Indications for Use for Temporary Abutment for Helix Short Implant
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Temporary Abutments are installed over Helix Short Implants and to provide support for prosthetic structures for up to 6 months.
Indications for Use for Helix Short Titanium Base:
The Helix Short Implant Titanium base that is placed over Neodent dental implants to provide support for customized prosthetic restorations, such as copings and crowns. It is indicated for single- and multiple-structure restorations, screw- or cement-retained on implants installed in the maxilla or mandible. All digitally-designed copings and/or crowns to be used with the Neodent Titanium Base System must be sent to Straumann for manufacture at a validated milling center.
3
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4
510(k) Summary | |
---|---|
ADMINISTRATIVE INFORMATION | |
Sponsor | JJGC Indústria e Comércio de Materiais Dentários SA |
(dba Neodent) | |
Av. Juscelino Kubitschek de Oliveira, 3291 | |
Curitiba, Paraná, Brazil 81270-200 | |
Registration No.: 3008261720 | |
Owner/Operator No.: 10031702 | |
Contact Person | Jennifer M. Jackson, MS |
Director of Regulatory Affairs, | |
Straumann USA | |
E-mail: jennifer.jackson@straumann.com | |
Telephone (978) 747-2509 | |
Date Prepared | 22/June/2023 |
Preparer / Alternate Contact | Bárbara Uzae |
Regulatory Affairs Analyst | |
E-mail: barbara.uzae@neodent.com | |
DEVICE NAME AND CLASSIFICATION | |
Trade/Proprietary Name | |
Common Name | Neodent Implant System – Helix Short Implant System |
Endosseous dental implant | |
Classification Name | Endosseous dental implant |
Classification Regulations | |
Product Code | 21 CFR 872.3640, Class II / 21 CFR 872.3630, Class II |
DZE/NHA | |
Classification Panel | |
Reviewing Branch | Dental Products Panel |
Dental Devices Branch | |
PREDICATE DEVICE INFORMATION | |
Primary Predicate Device | K163194, Neodent Implant System - GM Line, JJGC |
Indústria e Comércio de Materiais Dentários S.A | |
Reference Devices | K200586 - Straumann TLX Implant System, Institut |
Straumann AG | |
K123022 - Neodent Implant System, JJGC Indústria e | |
Comércio de Materiais Dentários S.A |
510(k) Summary
5
K202942 - Straumann® 4 mm Short Implants, Straumann USA, LLC
K203309 – NUVO CF Implant System, JJGC Indústria e Comércio de Materiais Dentários S.A
K220251 - Neodent Implant System - Narrow Implant System, JJGC Indústria e Comércio de Materiais Dentários S.A
K182620 - MRI Compatibility For Existing Neodent Implant System, JJGC Indústria e Comércio de Materiais Dentários S.A
INDICATIONS FOR USE
Indications for Use for Helix Short Implant:
The Neodent Implant System is recommended for surgical procedures on maxilla or mandible bones. It provides support for prosthetic components such as artificial teeth, thus restoring the chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Indications for Use for Mini Straight, Angled Abutment and screws:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Indications for Use for Helix Short Attachment:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Indications for Use for Helix Short Healing Abutment:
This product is used for the maintenance of soft tissue, during the bone integration stage of Neodent implants to be rehabilitated with the late loading technique. The Healing Abutment may be used in the implant installation surgery or the reopening surgery (second surgical stage).
6
Indications for Use for Helix Short Cover Screw:
This product is used for the maintenance of soft tissue, during the bone integration stage of Neodent implants to be rehabilitated with the late loading technique. The Cover Screw must remain intragingival, preventing the growth of tissue over the implant platform. The seating of the Cover Screw takes place in the conical region of the implant interface.
Indications for Use for Temporary Abutment for Helix Short Implant
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Temporary Abutments are indicated to be installed over Helix Short Implants and to provide support for prosthetic structures for up to 6 months.
Indications for Use for Helix Short Titanium Base:
The Helix Short Implant Titanium Base is a titanium base that is placed over Neodent dental implants to provide support for customized prosthetic restorations, such as copings and crowns. It is indicated for single- and multiple-structure restorations, screw- or cementretained on implants installed in the maxilla or mandible. All digitally-designed copings and/or crowns to be used with the Neodent Titanium Base System must be sent to Straumann for manufacture at a validated milling center.
SUBJECT DEVICE DESCRIPTIONS
- . All the subject devices are intended for single use;
- The subject implants are provided sterile via gamma irradiation and the subject abutments are provided sterile via Ethylene Oxide to an SAL of 1x10-9;
- . The subject implants are manufactured of Commercially Pure Titanium (Grade 4), the subject abutments, prosthetic screws and abutment screws are manufactured of Titanium alloy Ti-6Al-4V ELI (ASTM F136) and the restorations placed onto Helix Short Titanium Bases are made of various top half materials.
- . The subject implants have a wide cone prosthetic interface with an internal hexagon for short implants lengths available from 4.0 to 7.0; cylindrical shape with double threads and conical apex with three helical flutes for diameters from 3.75 to 7.0mm.
- . The implants are provided in two different surfaces: Neoporos: Sand blasted and acid etched, and Acqua: Hydrophilic surface aggregated to the rough surface obtained by double treatment: abrasive blasting and acid subtraction.
- The Helix Short Implant Titanium Base abutments is used as two-piece abutment, where the base is premanufactured from titanium alloy and the top half is created via burn out coping or a digital workflow with CAD/CAM at a validated milling center. The final top half can be created from Ticon, Coron, Zerion LT, IPS e.max, or PMMA. The top half and base pieces are cemented together to form the final abutment.
7
The various materials that can be used to create top halves for titanium bases and cements were already evaluated and cleared by FDA. Please find below a table with all clearance information for these materials:
Material | Raw material | 510(k) of the material |
---|---|---|
Ticon | Titanium | Exempt according to CFR |
872.3710 (subpart E): Dental Base | ||
Metal Alloys. | ||
Coron | CoCr alloy | |
Zerion LT | Zirconia | K061804 |
IPS e.max | Lithium Dissilicate | K132209 |
Polycon ae | PMMA | K071548 |
Panavia-Kuraray cement | Chemically-active resin | K183537 |
Ivoclar Multilink cement | Lithium Dissilicate | K130436 |
8
TECHNOLOGICAL CHARACTERISTIC COMPARISON TABLE
Table 1 - Technological Characteristic Comparison Table – Helix Short Implants
SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | REFERENCE DEVICE | REFERENCE DEVICE | ||
---|---|---|---|---|---|---|
Neodent Implant System - Helix | ||||||
Short Implant System | ||||||
JJGC Indústria e Comércio de | ||||||
Materiais Dentários S.A. | K163194 | |||||
Neodent Implant System - GM | ||||||
Line | ||||||
JJGC Indústria e Comércio de | ||||||
Materiais Dentários S.A. | K200586 | |||||
Straumann TLX Implant System | ||||||
Institut Straumann AG | K123022 | |||||
Neodent Implant System | ||||||
JJGC Indústria e Comércio de | ||||||
Materiais Dentários S.A. | K202942 | |||||
Straumann® 4 mm Short | ||||||
Implants | ||||||
Straumann USA, LLC | Equivalence | |||||
Discussion | ||||||
Indications for | ||||||
Use | The Neodent Implant | |||||
System is recommended | ||||||
for surgical procedures on | ||||||
maxilla or mandible bones. | ||||||
It provides support for | ||||||
prosthetic components | ||||||
such as artificial teeth, thus | ||||||
restoring the chewing | ||||||
function. It may be used | ||||||
with single-stage or two- | ||||||
stage procedures, for | ||||||
single or multiple unit | ||||||
restorations, and may be | ||||||
loaded immediately when | ||||||
good primary stability is | ||||||
achieved and with | ||||||
appropriate occlusal | ||||||
loading. | The Neodent Implant System | |||||
is intended to be surgically | ||||||
placed in the bone of the | ||||||
upper or lower jaw to | ||||||
provide support for | ||||||
prosthetic devices such as | ||||||
artificial teeth, to restore | ||||||
chewing function. It may be | ||||||
used with single-stage or | ||||||
two-stage procedures, for | ||||||
single or multiple unit | ||||||
restorations, and may be | ||||||
loaded immediately when | ||||||
good primary stability is | ||||||
achieved and with | ||||||
appropriate occlusal loading. | Straumann TLX Implants are | |||||
suitable for endosteal | ||||||
implantation in the upper and | ||||||
lower jaws and for the | ||||||
functional and esthetic oral | ||||||
rehabilitation of edentulous | ||||||
and partially edentulous | ||||||
patients. TLX Implants can be | ||||||
placed with immediate | ||||||
function on single-tooth and | ||||||
multiunit restorations when | ||||||
good primary stability is | ||||||
achieved and with appropriate | ||||||
occlusal loading to restore | ||||||
chewing function. The | ||||||
prosthetic restorations are | ||||||
connected to the implants | ||||||
through the corresponding | ||||||
abutment components. | Equivalent | |||||
Although the | ||||||
language is slightly | ||||||
different, the | ||||||
indications for use | ||||||
of the subject | ||||||
devices is the same | ||||||
as the primary | ||||||
predicate device | ||||||
(K163194) and | ||||||
equivalent to the | ||||||
indications for use | ||||||
of reference | ||||||
predicate device | ||||||
(K200586). The | ||||||
subject and | ||||||
predicate devices | ||||||
are intended to be | ||||||
surgically placed in | ||||||
the bone of the | ||||||
upper or lower jaw | ||||||
to provide support | ||||||
for prosthetic | ||||||
devices such as | ||||||
artificial teeth, to | ||||||
restore chewing | ||||||
function. | ||||||
SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | REFERENCE DEVICE | REFERENCE DEVICE | ||
Neodent Implant System - Helix | ||||||
Short Implant System | ||||||
JJGC Indústria e Comércio de | ||||||
Materiais Dentários S.A. | K163194 | |||||
Neodent Implant System – GM | ||||||
Line | ||||||
JJGC Indústria e Comércio de | ||||||
Materiais Dentários S.A. | K200586 | |||||
Straumann TLX Implant System | ||||||
Institut Straumann AG | K123022 | |||||
Neodent Implant System | ||||||
JJGC Indústria e Comércio de | ||||||
Materiais Dentários S.A. | K202942 | |||||
Straumann® 4 mm Short | ||||||
Implants | ||||||
Straumann USA, LLC | Equivalence | |||||
Discussion | ||||||
Implant- | ||||||
Abutment | ||||||
interface | Internal Hex | Morse taper | TorcFit (with conical fitting) | Morse taper | Internal octagon | Equivalent |
The subject and | ||||||
predicate devices | ||||||
have internal | ||||||
implant to | ||||||
abutment | ||||||
connection. | ||||||
Design | Cylindrical body | |||||
Tissue level | Conical body (Helix) | |||||
Bone level | Tapered body | |||||
Tissue level | Cylindrical body | |||||
Bone level | Cylindrical body | |||||
Tissue level | Identical | |||||
Cylindrical shape and | ||||||
tissue level as the | ||||||
reference devices. | ||||||
Reusable | No | No | No | No | No | Identical |
The subject | ||||||
devices and the | ||||||
predicate devices | ||||||
are indicated for | ||||||
single use. | ||||||
Length (mm) | 4.0 / 5.5 / 7.0 / 8.5 | 8 / 10 / 11.5 / 13 / 16 / 18 | ||||
(GM Helix) | 6.0 / 8.0 / 10 / 12 / 14 / 16 / 18 | - 5.0 / 6.0 (WS) | 4.0 | Equivalent | ||
Subject implant | ||||||
lengths are within | ||||||
the range of the | ||||||
predicate devices. | ||||||
Collar length | ||||||
(mm) | 1.8 | N/A | 1.8 | N/A | 1.8 | Identical |
Subject implant and | ||||||
reference devices with | ||||||
tissue level have a | ||||||
collar neck of 1.8mm. | ||||||
SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | REFERENCE DEVICE | REFERENCE DEVICE | ||
Neodent Implant System - Helix | ||||||
Short Implant System | ||||||
JJGC Indústria e Comércio de | ||||||
Materiais Dentários S.A. | K163194 | |||||
Neodent Implant System – GM | ||||||
Line | ||||||
JJGC Indústria e Comércio de | ||||||
Materiais Dentários S.A. | K200586 | |||||
Straumann TLX Implant System | ||||||
Institut Straumann AG | K123022 | |||||
Neodent Implant System | ||||||
JJGC Indústria e Comércio de | ||||||
Materiais Dentários S.A. | K202942 | |||||
Straumann® 4 mm Short | ||||||
Implants | ||||||
Straumann USA, LLC | Equivalence | |||||
Discussion | ||||||
Diameter (Ø) | ||||||
(mm) | 3.75 / 4.0 / 5.0 / 6.0 / 7.0 | 3.5 / 3.75 / 4.0 / 4.3 / 5.0 | ||||
(GM Helix) | 3.75 / 4.0 / 4.5 / 5.0 / 5.5 / 6.5 | 4.0 / 5.0 / 6.0 (WS) | 4.1 / 4.8 | Equivalent | ||
Subject device | ||||||
diameters are | ||||||
within the range of | ||||||
diameters of the | ||||||
predicates devices | ||||||
or larger (Ø 7.0). | ||||||
Larger diameters | ||||||
do not represent a | ||||||
worst case in | ||||||
terms of | ||||||
performance. | ||||||
Material | Commercially pure | |||||
Titanium grade 4 (ASTM | ||||||
F67) | Commercially pure Titanium | |||||
grade 4 (ASTM F67) | Titanium-13 Zirconium alloy | Commercially pure titanium | ||||
grade 4 (ASTM F67) | Titanium-13 Zirconium | |||||
alloy | Identical | |||||
Identical to the | ||||||
primary predicate | ||||||
device and the | ||||||
Neodent reference | ||||||
devices (K123022) | ||||||
Sterilization | ||||||
Method | Gamma irradiation to an | |||||
SAL of 1x10-6 | Gamma irradiation to an SAL of | |||||
1x10-6 | Gamma irradiation to an SAL of | |||||
1×10-6 | Gamma irradiation to an SAL of | |||||
1×10-6 | Gamma irradiation to an SAL | |||||
of 1x10-6 | Identical | |||||
Subject | ||||||
devices | ||||||
and | ||||||
predicate | ||||||
devices | ||||||
are | ||||||
provided sterile by | ||||||
the | ||||||
same | ||||||
sterilization | ||||||
method. | ||||||
Surface | ||||||
treatment | NeoPoros | |||||
Acqua | Neoporos | |||||
Acqua | SLActive® | NeoPoros | SLActive® | Identical | ||
Identical | ||||||
to the | ||||||
primary | ||||||
predicate | ||||||
device. |
9
10
11
SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | ||
---|---|---|---|---|
Neodent Implant System - Helix Short | ||||
Implants System | ||||
JJGC Indústria e Comércio de Materiais | ||||
Dentários S.A. | K163194 | |||
Neodent Implant System - GM Line | ||||
JJGC Indústria e Comércio de | ||||
Materiais Dentários S.A | K203309 | |||
NUVO CF Implant System | ||||
JJGC Indústria e Comércio de Materiais | ||||
Dentários S.A. | Equivalence | |||
Discussion | ||||
Indications for Use | Mini Straight and Angled Abutment: | |||
The Neodent Implant System is | ||||
recommended for surgical procedures | ||||
on maxilla or mandible bones. It | ||||
provides support for | ||||
prosthetic | ||||
components such as artificial teeth, | ||||
thus restoring the chewing function. It | ||||
may be used in one- or two-stage | ||||
procedures, multiple restorations, as | ||||
well as immediate loading when there | ||||
is good primary stability and adequate | ||||
occlusal load. Multiple rehabilitations | ||||
may be splinted rigidly. | GM Mini Conical and GM Exact Mini | |||
Conical Abutments: | ||||
The Neodent Implant System is | ||||
intended to be surgically placed in the | ||||
bone of the upper or lower jaw to | ||||
provide support for prosthetic devices | ||||
such as artificial teeth, to restore | ||||
chewing function. It may be used with | ||||
single stage or two-stage procedures | ||||
for single or multiple unit restorations | ||||
and may be loaded immediately when | ||||
good primary stability is achieved and | ||||
with appropriate occlusal loading. | Multi-Unit Screw Retained Abutment | |||
CF (Straight/Angled): | ||||
The Implant System is intended to be | ||||
surgically placed in the maxilla or | ||||
mandible to provide support for | ||||
prosthetic devices such as artificial | ||||
teeth in order to restore chewing | ||||
function. It may be used with single- | ||||
stage or two-stage procedures, for | ||||
single or multiple unit restorations, | ||||
and may be loaded immediately when | ||||
good primary stability is achieved and | ||||
with appropriate occlusal loading. | ||||
Multiple tooth applications may be | ||||
rigidly splinted. | Equivalent | |||
Although the | ||||
language is slightly | ||||
different, the indications for use of the | ||||
subject devices is included into the | ||||
indications of the predicate devices. The | ||||
subject and predicate device intended to | ||||
be surgically placed in the bone of the | ||||
upper or lower jaw to provide support | ||||
for prosthetic devices such as artificial | ||||
teeth, to restore chewing function. It | ||||
may be used with single stage or two- | ||||
stage procedures, for single or multiple | ||||
unit restorations, and may be loaded | ||||
immediately when good primary stability | ||||
is achieved and with appropriate occlusal | ||||
loading. | ||||
Abutment Design | Surface: | |||
Machined | ||||
with | ||||
pink | ||||
electrolysis | ||||
Mini Straight Abutment: | ||||
Gingival height: 0.2, 1.5, 2.5, 3.5, 4.5 | ||||
Mini angled abutment: | ||||
Gingival height: 0.6, 1.5, 2.5, 3.5 | ||||
Angulation: 17° | Surface: Machined | |||
GM Mini Conical Abutments: | ||||
Gingival height: 0.8; 1.5, 2.5, 3.5, 4.5, | ||||
5.5 | ||||
GM Exact Mini Conical Abutments: | ||||
Gingival height: 1.5, 2.5, 3.5 | ||||
Angulation: 17° and 30° | Surface: Machined and electrolysis | |||
Straight Abutment: | ||||
Gingival height: 1.0, 2.0, 3.0, 4.0 and 5.0 | ||||
mm; | ||||
Angled Abutment: | ||||
Gingival height: 2.0 and 3.0 mm | ||||
Angulation: 17° and 30° | Equivalent | |||
The subject device have the same surface of | ||||
reference predicate device and similar | ||||
geometry and range of dimensions and | ||||
angulation of primary predicate device. | ||||
Single Use | Yes | Yes | Yes | Identical |
The subject devices and predicates | ||||
devices are indicated for single use. |
Table 2 - Technological Characteristic Comparison Table – Mini Abutments
JJGC Indústria e Comércio de Materiais Dentários S.A.
12
K223638 - Neodent Implant System – Helix Short Implant System
SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | ||
---|---|---|---|---|
Neodent Implant System - Helix Short | ||||
Implants System | ||||
JJGC Indústria e Comércio de Materiais | ||||
Dentários S.A. | K163194 | |||
Neodent Implant System - GM Line | ||||
JJGC Indústria e Comércio de | ||||
Materiais Dentários S.A | K203309 | |||
NUVO CF Implant System | ||||
JJGC Indústria e Comércio de Materiais | ||||
Dentários S.A. | Equivalence | |||
Discussion | ||||
Raw material | Titanium alloy ASTM F136. | Titanium alloy ASTM F136. | Titanium alloy ASTM F136. | Identical |
The subject devices and predicates | ||||
devices are manufactured of the same | ||||
raw material. | ||||
Sterilization | ||||
Method | Provided sterile via Ethylene Oxide | |||
to an SAL of 10-6 | Provided sterile via Ethylene Oxide to | |||
an SAL of 10-6 | Provided sterile via Ethylene Oxide to an | |||
SAL of 10-6 | Identical | |||
The subject devices and the primary | ||||
predicate devices are provided in sterile | ||||
condition to an SAL of 1 x 10-6 by the same | ||||
sterilization method. |
Table 3 - Technological Characteristic Comparison Table – Attachments
SUBJECT DEVICE | REFERENCE DEVICE | ||
---|---|---|---|
Neodent Implant System - Helix Short Implants | |||
System | |||
JJGC Indústria e Comércio de Materiais | |||
Dentários S.A. | K220251 | ||
Neodent Implant System - Narrow Implant System | |||
JJGC Indústria e Comércio de Materiais Dentários | |||
S.A | Equivalence | ||
Discussion | |||
Indications for Use | Helix Short Attachment: | ||
The Neodent Implant System is | |||
recommended for surgical procedures on | |||
maxilla or mandible bones. It provides | |||
support for prosthetic components such as | |||
artificial teeth, thus restoring the chewing | |||
function. It may be used in one- or two-stage | |||
procedures, multiple unit restorations, as | |||
well as immediate loading when there is | |||
primary stability and adequate occlusal load. | NGM Attachments | ||
The Neodent Implant System is intended to be | |||
surgically placed in the bone of the upper or lower | |||
jaw to provide support for prosthetic devices, such | |||
as artificial teeth, to restore chewing function. It may | |||
be used with single stage or two-stage procedures | |||
for single or multiple unit restorations, and may be | |||
loaded immediately when good primary stability is | |||
achieved and with appropriate occlusal loading. | Equivalent | ||
Although the language is slightly different, the indications | |||
for use of the subject devices is the same of the indications | |||
of the predicate devices. The subject and predicate device | |||
are intended to be surgically placed in the bone of the | |||
upper or lower jaw to provide support for prosthetic | |||
devices such as artificial teeth, to restore chewing | |||
function. It may be used with single stage or two-stage | |||
procedures, for single or multiple unit restorations, and | |||
may be loaded immediately when good primary stability is | |||
achieved and with appropriate occlusal loading. |
JJGC Indústria e Comércio de Materiais Dentários S.A.
13
SUBJECT DEVICE | REFERENCE DEVICE | ||
---|---|---|---|
Neodent Implant System - Helix Short Implants | |||
System | |||
JJGC Indústria e Comércio de Materiais | |||
Dentários S.A. | K220251 | ||
Neodent Implant System - Narrow Implant System | |||
JJGC Indústria e Comércio de Materiais Dentários | |||
S.A | Equivalence | ||
Discussion | |||
Abutment Design | HS Attachments: | ||
Gingival height: 0.2, 1.5, 2.5, 3.5 | |||
Surface Treatment: TiN Coating | NGM Attachments: | ||
Gingival height: 0.8; 1.5, 2.5, 3.5 and 4.5 | |||
Surface Treatment: TiN Coating | Equivalent | ||
The subject devices and predicates devices have the same | |||
surface treatment and similar geometry and range of | |||
dimensions. | |||
Single Use | Yes | Yes | Identical |
The subject devices and predicates devices are indicated for | |||
single use. | |||
Raw material | Titanium alloy ASTM F136. | Titanium alloy ASTM F136. | Identical |
The subject devices and predicates devices are | |||
manufactured of the same raw material. | |||
Sterilization Method | Provided sterile via Ethylene Oxide to an | ||
SAL of 10-6. | Provided sterile via Ethylene Oxide to an SAL of | ||
10-6 | Identical | ||
The subject devices and the primary predicate devices are | |||
provided in sterile condition to an SAL of 1 x 10 -6 by the | |||
same sterilization method. |
14
SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | ||
---|---|---|---|---|
Neodent Implant System - Helix | ||||
Short Implants System | ||||
JJGC Indústria e Comércio de | ||||
Materiais Dentários S.A. | K163194 | |||
Neodent Implant System - GM Line | ||||
JJGC Indústria e Comércio de | ||||
Materiais Dentários S.A | K203309 | |||
NUVO CF Implant System | ||||
JJGC Indústria e Comércio de | ||||
Materiais Dentários S.A. | Equivalence | |||
Discussion | ||||
Indications for Use | Helix Short Healing Abutment | |||
and Cover Screw: | ||||
This product is used for the | ||||
maintenance of soft tissue, | ||||
during the bone integration stage | ||||
of Neodent implants to be | ||||
rehabilitated with the late | ||||
loading technique. |
The Healing Abutment may be
used in the implant installation
surgery or the reopening surgery
(second surgical stage).
The Cover Screw must remain
intragingival, preventing the
growth of tissue over the implant
platform. The seating of the
Cover Screw takes place in the
conical region of the implant
interface. | GM Healing abutment and Cover
Screw:
The Neodent Implant System is
intended to be surgically placed in
the bone of the upper or lower
jaw to provide support for
prosthetic devices, such as
artificial teeth, to restore chewing
function. It may be used with
single stage or two-stage
procedures, for single or multiple
unit restorations, and may be
loaded immediately when good
primary stability is achieved and
with appropriate occlusal loading. | NUVO CF Healing Abutments and
Cover Screws:
The Implant System is intended to
be surgically placed in the maxilla
or mandible to provide support
for prosthetic devices such as
artificial teeth in order to restore
chewing function. It may be used
with single-stage or two-stage
procedures, for single or multiple
unit restorations, and may be
loaded immediately when good
primary stability is achieved and
with appropriate occlusal loading.
Multiple tooth applications may
be rigidly splinted. | Equivalent
Although the language is
different, the indications for use
of the subject devices are the
same of predicate devices. The
subject and the predicate devices
are used to the maintenance of
soft tissue, during the bone
integration stage of implants to
be rehabilitated with the late
loading technique. The
indications for use are written
differently because those for the
predicate devices are talking
about the implant system as a
whole and not the healing
abutment and cover screws
specifically. |
Table 4 - Technological Characteristic Comparison Table – Healing Abutment and Cover Screw
15
SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | ||
---|---|---|---|---|
Neodent Implant System - Helix | ||||
Short Implants System | ||||
JJGC Indústria e Comércio de | ||||
Materiais Dentários S.A. | K163194 | |||
Neodent Implant System - GM Line | ||||
JJGC Indústria e Comércio de | ||||
Materiais Dentários S.A | K203309 | |||
NUVO CF Implant System | ||||
JJGC Indústria e Comércio de | ||||
Materiais Dentários S.A. | Equivalence | |||
Discussion | ||||
Abutment Design | Surface: Machined with pink | |||
electrolysis. | ||||
HS Healing Abutment: | ||||
Gingival height: 1.5, 2.5, 3.5, 4.5, | ||||
5.5 | ||||
Cover Screw: | ||||
0.35 mm height above the implant | ||||
platform level | Surface: Machined | |||
GM Healing abutment: | ||||
Gingival height: 0.8, 1.5, 2.5, 3.5, 4.5, | ||||
5.5 | ||||
Cover Screw: | ||||
Gingival height: at the bone level, | ||||
or with a height of 2 mm | Surface: Machined with pink | |||
electrolysis. | ||||
NUVO CF Healing Abutments and | ||||
Cover Screws: | ||||
Gingival height: 1.0, 2.0, 3.0, 4.0 and | ||||
5.0 | ||||
Cover Screw: 0.3, 2. 0 mm | Equivalent | |||
The subject devices and the | ||||
reference predicate devices have | ||||
the same surface treatment. The | ||||
subject and predicate devices have | ||||
similar geometry and dimensions. | ||||
Single Use | Yes | Yes | Yes | Identical |
The | ||||
subject | ||||
devices | ||||
and | ||||
predicates devices are indicated | ||||
for single use. | ||||
Raw material | Titanium alloy ASTM F136. | Titanium alloy ASTM F136. | Titanium alloy ASTM F136. | Identical |
The subject devices and predicate | ||||
devices are manufactured of the | ||||
same raw material. | ||||
Sterilization | ||||
Method | Provided sterile via Ethylene | |||
Oxide to an SAL of 10-6. | Provided sterile via Ethylene | |||
Oxide to an SAL of 10-6. | Provided sterile via Ethylene | |||
Oxide to an SAL of 10-6. | Identical | |||
The subject devices and the | ||||
predicate devices are provided in | ||||
sterile condition to an SAL of 1 x 10-6 | ||||
by the same sterilization | ||||
method. |
16
| | SUBJECT DEVICE | REFERENCE DEVICE | REFERENCE DEVICE | Equivalence
Discussion |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Neodent Implant System - Helix Short Implants System
JJGC Indústria e Comércio de Materiais Dentários S.A. | K203309
NUVO CF Implant System
JJGC Indústria e Comércio de Materiais Dentários
S.A. | K220251
Neodent Implant System - Narrow Implant System
JJGC Indústria e Comércio de Materiais Dentários
S.A | |
| Indications for
Use | Temporary Abutment:
The Implant System is indicated for surgical
procedures on maxilla or mandible bones, providing
support for dental prostheses with the purpose of
restoring the chewing function as well as aesthetics.
It may be used in one- or two-stage procedures, single
or multiple restorations, and immediate loading
when there is good primary stability and adequate
occlusal load. Temporary Abutments are indicated to
be installed over Helix Short Implants and to provide
support for prosthetic structures for up to 6 months. | Temporary Abutments:
The Implant System is intended to be surgically
placed in the maxilla or mandible to provide
support for prosthetic devices such as artificial
teeth in order to restore chewing function. It
may be used with single-stage or two-stage
procedures, for single or multiple unit
restorations, and may be loaded immediately
when good primary stability is achieved and
with appropriate occlusal loading. Multiple
tooth applications may be rigidly splinted. | Temporary Abutments:
Prosthetic component to be installed in the
Narrow GM Implant, providing support for
prosthetic structures. Indicated for temporary
restorations with unitary structures, screwed,
on implants installed in the maxilla or mandible.
Temporary Abutments Narrow GM have a
maximum duration of usage of 180 days. | Equivalent
Although the language is
slightly different, the
indications for use of the
subject devices is included into
the indications of the
predicate devices. The subject
and predicate devices are
intended to be surgically
placed in the maxilla or
mandible to provide support
for prosthetic devices such as
artificial teeth in order to
restore chewing function. It
may be used in one- or two-
stage procedures, single or
multiple restorations, and
immediate loading when there
is good primary stability and
adequate occlusal load. All are
indicated to be used for up to
6 months. |
| Abutment
Design | Temporary Abutments:
Gingival height: 0.2; 1.5; 2.5; 3.5
Surface: Machined with pink electrolysis. | Temporary Abutments:
Gingival height: 0.5, 1.0 and 3.0 mm
Surface: Machined with pink electrolysis. | Temporary Abutments:
Gingival height: 0,8; 1.5, 2.5, 3.5 and 4.5
Surface: Machined | Equivalent
The subject devices and the
primary predicate devices have
the same surface treatment. The
subject and predicate devices
have similar geometry and
dimensions. |
Table 5 - Technological Characteristic Comparison Table – Temporary Abutments
JJGC Indústria e Comércio de Materiais Dentários S.A.
17
| | SUBJECT DEVICE | REFERENCE DEVICE | REFERENCE DEVICE | Equivalence
Discussion |
|-------------------------|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Neodent Implant System - Helix Short Implants System
JJGC Indústria e Comércio de Materiais Dentários S.A. | K203309
NUVO CF Implant System
JJGC Indústria e Comércio de Materiais Dentários S.A. | K220251
Neodent Implant System - Narrow Implant System
JJGC Indústria e Comércio de Materiais Dentários S.A | |
| Single Use | Yes | Yes | Yes | Identical
The subject devices and
predicates devices are
indicated for single use. |
| Raw material | Titanium alloy ASTM F136. | Titanium alloy ASTM F136. | Titanium alloy ASTM F136. | Identical
The subject devices and
predicates devices are
manufactured of the same raw
material. |
| Sterilization
Method | Provided sterile via Ethylene Oxide to an SAL of 10-6 | Provided sterile via Ethylene Oxide to an SAL
of 10-6 | Provided sterile via Ethylene Oxide to an SAL
of 10-6 | Identical
The subject devices and the
primary predicate devices are
provided in sterile condition to
an SAL of 1 x 10 -6 by the same
sterilization method. |
18
SUBJECT DEVICES | PRIMARY PREDICATE | REFERENCE DEVICE | ||
---|---|---|---|---|
COMPARISON | Subject | |||
Neodent Implant System - Titanium Bases for | ||||
Helix Short Implant Line | ||||
JJGC Indústria e Comércio de Materiais | ||||
Dentários S.A. | K163194 | |||
Neodent Implant System - GM Line | ||||
JJGC Indústria e Comércio de Materiais Dentários | ||||
S.A. | K203309 | |||
NUVO CF Implant System | ||||
JJGC Indústria e Comércio de Materiais Dentários | ||||
S.A. | EQUIVALENCE DISCUSSION | |||
Indications for Use | The Helix Short Implant Titanium Base is a | |||
titanium base that is placed over Neodent | ||||
dental implants to provide support for | ||||
customized prosthetic restorations, such as | ||||
copings and crowns. It is indicated for | ||||
single- and multiple-structure restorations, | ||||
screw- or cement-retained on implants | ||||
installed in the maxilla or mandible. All | ||||
digitally-designed copings and/or crowns to | ||||
be used with the Neodent Titanium Base | ||||
System must be sent to Straumann for | ||||
manufacture at a validated milling center. | Titanium Base Abutment is a titanium base | |||
placed onto Neodent dental implants to | ||||
provide support for customized prosthetic | ||||
restorations. It is used with a coping and | ||||
crown, or crown alone, and is indicated for | ||||
cement- or screw-retained single or multi-unit | ||||
restorations single. All digitally designed | ||||
copings and/or crowns to be used with the | ||||
Neodent Titanium Base Abutment System are | ||||
intended to be sent to Straumann for | ||||
manufacture at a validated milling center. The | ||||
GM Titanium Base for Bridge is indicated for | ||||
cement or screw-retained | ||||
multi-unit | ||||
restorations. | Titanium Base for Bridge: | |||
CF Titanium Base for Bridge is a titanium | ||||
abutment placed onto dental implants to | ||||
provide support for customized prosthetic | ||||
restorations. The CF Titanium Base for Bridge | ||||
is indicated for cement or screw-retained | ||||
multi-unit restorations. All digitally designed | ||||
copings and/or crowns to be used with the | ||||
Titanium Base Abutment System are intended | ||||
to be sent to Straumann for manufacture at a | ||||
validated milling center. | ||||
Titanium Base for Crown: | ||||
CF Titanium Base for Crown is a titanium | ||||
abutment placed onto dental implants to | ||||
provide support for customized prosthetic | ||||
restorations, as copings or crowns. It is | ||||
indicated for single-unit restorations, cement- | ||||
retained or screw-retained in aesthetic areas | ||||
on implants installed in the maxilla or | ||||
mandible. All digitally designed copings and/or | ||||
crowns to be used with the Titanium Base | ||||
Abutment System are intended to be sent to | ||||
Straumann for manufacture at a validated | ||||
milling center. | Equivalent | |||
Although the language is | ||||
slightly different, the | ||||
indications for use of the | ||||
subject and predicate devices | ||||
are the same. The subject and | ||||
predicate devices are | ||||
intended to place onto dental | ||||
implants to provide support | ||||
for customized prosthetic | ||||
restorations, such as copings | ||||
and crowns. |
Table 6 - Technological Characteristic Comparison Table – Helix Short Titanium Bases
19
SUBJECT DEVICES | PRIMARY PREDICATE | REFERENCE DEVICE | ||||||||
---|---|---|---|---|---|---|---|---|---|---|
COMPARISON | Subject | |||||||||
Neodent Implant System – Titanium Bases for | ||||||||||
Helix Short Implant Line | ||||||||||
JJGC Indústria e Comércio de Materiais | ||||||||||
Dentários S.A. | K163194 | |||||||||
Neodent Implant System - GM Line | ||||||||||
JJGC Indústria e Comércio de Materiais Dentários | ||||||||||
S.A. | K203309 | |||||||||
NUVO CF Implant System | ||||||||||
JJGC Indústria e Comércio de Materiais Dentários | ||||||||||
S.A. | EQUIVALENCE DISCUSSION | |||||||||
Abutment Design | Cylindrical format provided in different | |||||||||
gingival heights. | ||||||||||
Angulation: 0° | ||||||||||
Gingival Heights: 0.2; 1.5, 2.5 and 3.5 mm | ||||||||||
Surface: Machined and Electrolysis | ||||||||||
Cementable portion height: 6.0 and 4.5 | ||||||||||
mm | Cylindrical format provided in different | |||||||||
gingival heights. | ||||||||||
Angulation: 0° | ||||||||||
Gingival Heights: 0.8; 1.5; 2.5; 3.5; 4.5 mm | ||||||||||
Surface: Machined | ||||||||||
Cementable portion height: 6.0 and 4.0 | Conical format provided in different gingival | |||||||||
heights. | ||||||||||
Angulation: 0° | ||||||||||
Gingival Heights: 0.5; 1.0; 3 mm | ||||||||||
Surface: Machined and Electrolysis | ||||||||||
Cementable portion height: 6.0 and 4.5 mm | Equivalent | |||||||||
The subject devices and the | ||||||||||
reference predicate devices | ||||||||||
share the same surface finish, | ||||||||||
similar geometry and range | ||||||||||
of dimensions. | ||||||||||
Single Use | Yes | Yes | Yes | Identical | ||||||
Subject and predicate devices | ||||||||||
are not reusable. | ||||||||||
Raw material | Titanium Alloy (Ti-6Al-4V) – according to | |||||||||
ASTM F136. | Titanium Alloy (Ti-6Al-4V) – according to | |||||||||
ASTM F136. | Titanium Alloy (Ti-6Al-4V) – according to ASTM | |||||||||
F136. | Identical | |||||||||
The subject and predicate | ||||||||||
devices are manufactured of | ||||||||||
the same raw material. | ||||||||||
Top Half Materials | Material | Minimum | ||||||||
thickness | Maximum | |||||||||
allowable | ||||||||||
angulation | Material | Minimum | ||||||||
thickness | Maximum | |||||||||
allowable | ||||||||||
angulation | Material | Minimum | ||||||||
thickness | Maximum | |||||||||
allowable | ||||||||||
angulation | Equivalent | |||||||||
The top half materials of the | ||||||||||
subject devices are contained | ||||||||||
within the scope of the top | ||||||||||
half materials of the | ||||||||||
predicate device. The | ||||||||||
maximum allowable | ||||||||||
angulation of subject device | ||||||||||
is smaller than the maximum | ||||||||||
angulation of the predicate | ||||||||||
devices, therefore it is less | ||||||||||
critical. | ||||||||||
Coron | ||||||||||
(CoCr) | 0.3 | Coron (CoCr) | 0.3 | Coron (CoCr) | 0.3 | |||||
Ticon | ||||||||||
(Titanium) | 0.4 | 16° | Ticon (Titanium) | 0.4 | ||||||
IPS e.max | 0.9 | Zerion LT | 0.5 | 30° | Zerion LT | 0.5 | ||||
Zerion LT | 0.5 | IPS e.max CAD | 0.9 | IPS e.max CAD | 0.9 | 30° | ||||
Polycon | ||||||||||
ae*(PMMA) | 1.0 | 3M ESPE Lava | ||||||||
Plus Zirconia | 0.7 | Polycon | ||||||||
ae*(PMMA) | 1.0 | |||||||||
* Polycon ae material is only recommended | ||||||||||
for temporary/provisional prosthetic | ||||||||||
restorations. | Polycon | |||||||||
ae*(PMMA) | 1.0 | *Polycon ae is indicated to remain in the mouth | ||||||||
only for up to 180 days. |
20
SUBJECT DEVICES | PRIMARY PREDICATE | REFERENCE DEVICE | EQUIVALENCE DISCUSSION | |
---|---|---|---|---|
COMPARISON | Subject | |||
Neodent Implant System – Titanium Bases for | ||||
Helix Short Implant Line | ||||
JJGC Indústria e Comércio de Materiais | ||||
Dentários S.A. | K163194 | |||
Neodent Implant System – GM Line | ||||
JJGC Indústria e Comércio de Materiais Dentários | ||||
S.A. | K203309 | |||
NUVO CF Implant System | ||||
JJGC Indústria e Comércio de Materiais Dentários | ||||
S.A. | ||||
Sterilization Method | • Provided sterile via Ethylene Oxide | |||
to an SAL of 1x10-6 |
The final prosthesis must be sterilized by the
end used before installation in the mouth
via moist heat (steam), gravity displacement
or dynamic air removal in 132°C (270°F) for
3 minute exposure. | • Provided sterile via Ethylene Oxide to
an SAL of 1x10-6
To be sterilized by user after superstructure
cementation, before placed in patient mouth
(moist steam sterilization). | • Provided sterile via Ethylene Oxide to
an SAL of 1x10-6
If customized on the chairside, must be sterilized
before the installation via moist heat (steam),
Gravity displacement or dynamic air removal in
132°C (270°F) for 3 minute exposure. | Identical
The subject devices and
predicate devices share the
same sterilization method. |
21
The subject implants have equivalent indications for use and equivalent range of lengths as the primary predicate devices, being contemplated within the range of lengths of the predicate devices. Subject and predicate implants have the similar implant-to-abutment interface. Both present the same surfaces, are manufactured of the same materials and are sterilized using the same sterilization method. Subject and reference predicate devices present the same range of sizes and similar overall design.
The subject conventional abutments have equivalent indications for use and overall design as the predicate devices. They also present the same sterilization method and raw materials.
The subject Base abutments have the equivalent indications for use and same sterilization method as the primary predicate devices. Both, subject and primary predicate devices, are made of the same material. The top half materials of the subject devices are contained within the scope of the top half materials of the predicate devices and all digitally designed copings and/or crowns for use with the subject Titanium Bases are intended to be sent to Straumann for manufacture at a validated milling center as the predicate devices. The subject devices and the reference predicate devices share the same surface finish, similar geometry and range of dimensions.
Overall, the subject devices are substantially equivalent to the predicate devices as follows:
- · same intended use,
- · same operating principle,
- · incorporate the equivalent basic design,
- incorporate the same materials, and
- are sterilized using the same materials and processes
PERFORMANCE DATA
Mechanical testing
Dynamic Fatigue Test per ISO 14801 was performed to determine the fatigue strength for the dental implant system, according to FDA Guidance. The tested subject devices exhibit a level of performance equivalent to that reviewed for the predicate devices.
Torsion Test was performed to evaluate Helix Short Implant under static torsional loading. The results met the acceptance criteria. Insertion test was performed to evaluate the insertion torque of the Helix Short System when it is inserted in bones type I, II, IIII and IV.
The Implant Surface Area simulation and Pull Out Test were also made to evaluate and represent the clinical use of the subject implants. Results demonstrated that the subject devices, despite having a slightly smaller surface area than the reference devices, presented a higher BIC rate and pull out resistance upon placement. Thus, the subject devices exhibit a level of benchtop performance equivalent to that reviewed for the reference device.
22
MR Compatibility testing
The MR compatibility was performed to access the risk of exposing patients who have implantable medical devices. An assessment was made to demonstrate that the subject devices do not configure a new worst case and can be represented by the previously conducted studies reviewed for reference devices, since both have the same raw material and similar dimensions. The subject devices are therefore MR conditional devices and a patient treated with the subject devices can be safely scanned observing the parameters previously established per reference devices.
Surface treatment
The surface treatments applied to subject devices are identical to these applied and previously evaluated for primary predicate devices. Although some reference devices have a surface with different nomenclature, the chemical processes applied to the surface are similar to create the roughness surface. Acqua implants are submitted to an additional step in order to increase their hydrophilicity, as already cleared for primary predicate device.
Sterilization validation
Sterilization of the subject implants via gamma irradiation according to ISO 11137-1 Sterilization of health care products -Radiation -Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices and ISO 11137-2 Sterilization of health care products -Radiation -Part 2: Establishing the sterilization dose. A minimum Sterility Assurance Level (SAL) of 1 x 10-6 has been validated.
Sterilization of the subject abutments via ethylene oxide gas using the overkill method has been performed according to the requirements of ISO 11135-1 Sterilization of Health Care Products - Ethylene Oxide - Part 1: Requirements For Development, Validation and Routine Control of a Sterilization Process For Medical Devices. A minimum Sterility Assurance Level (SAL) of 1 x 10-6 has been validated.
Endotoxin testing (LAL)
The LAL Endotoxin Analysis is the method used to determine that the device meets pyrogen limit specifications, based in the referred FDA guidance (510(k) Submission for Devices Labeled as Sterile, issued on 21 January 2016). Routine endotoxin testing is performed with representative devices as an assessment of environmentally derived pyrogenicity. Subject devices meet the acceptance criteria for endotoxin levels as well as the primary predicate and reference devices. The subject devices are not represented to be "pyrogen free".
Shelf Life validation
The expiration date of the products was determined considering the integrity of the product and the packaging tests after shelf life testing. The packaging of the subject Helix Short Implant System is identical to the packaging of the primary predicate and reference devices. The shelf life for devices provided sterile is 5 years, except Helix Short Acqua Implants that have a shelf life of 4 years.
Biocompatibility
Representative samples of each of the subject devices was subjected to the following:
23
Cytotoxicity testing was performed per ISO 10993-5. Chemical characterization was performed per ISO 10993-18.
Biocompatibility sample preparation was performed per ISO 10993-12. Biological Safety Assessment guided by ISO 10993-1.
EO Residuals evaluation was performed in accordance with 10993-7.
No new issues of biocompatibility are raised for the subject devices when compared to primary predicate and reference devices. Therefore, no additional biocompatibility testing was required.
CONCLUSION
The data included in this submission demonstrate that the subject system is substantially equivalent to the primary predicate and reference devices in terms of intended use, design, technological characteristics, mechanical properties, shelf life, sterilization method and material.