K Number
K223520
Date Cleared
2023-04-27

(155 days)

Product Code
Regulation Number
884.1690
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The UNIDRIVE SIII System is intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue.

The HOPKINS Telescope when used with the obturator is intended to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.

Device Description

The UNIDRIVE® SIII system is a motorized surgical device intended for intrauterine use to hysteroscopically resect and remove tissue. The UNIDRIVE SIII System includes the following components:
UNIDRIVE SIII Control Unit (20701020-1) Footswitch (20016230) DrillCut-X II Shaver handpiece GYN (26702050) Shaver blades (26208SA, 26208SB) HOPKINS Telescope (26208AMA) Obturator (26208OC)

The UNIDRIVE S III system has a motor control unit and can be used with the DrillCut-X II Shaver handpiece GYN which houses the shaver blades. The shaver blade upon being connected to the DrillCut-X II Shaver handpiece GYN is inserted into the working channel of the HOPKINS Telescopes. The shaver blades are activated via a one-pedal footswitch that is connected the UNIDRIVE SIII control unit. The shaver blades consist of two sheaths that fit into each other. The inner sheath oscillates within the outer sheath to enable tissue resection.

AI/ML Overview

The provided text is a 510(k) summary for the UNIDRIVE SIII System. It primarily focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing rather than presenting a study with strict acceptance criteria of the type typically used for a novel AI/software performance claim.

Therefore, the requested information specifically on acceptance criteria, reported performance, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, ground truth types, and training set details as they pertain to such a study is not available in the provided document. The document describes a comparison to predicate devices, stating that non-clinical testing was adequate to establish substantial equivalence without requiring clinical testing.

However, I can extract the types of non-clinical tests performed, which could be considered as demonstrating the device meets established safety and performance standards relevant to its design.

Regarding your specific questions, based on the provided text:

  1. A table of acceptance criteria and the reported device performance:
    The document does not provide a table of acceptance criteria with specific numerical targets and reported performance values in the context of a clinical or analytical performance study of the type you're asking for. Instead, it refers to successful demonstration of performance by comparison to predicate devices and adherence to recognized standards.

    • "The cutting performance of the subject device, UNIDRIVE SIII system has been successfully demonstrated by comparing to the primary predicate device TRUCLEAR Morcellator System and TRUCLEAR Morcellators cleared in K132015." This implies the UNIDRIVE SIII System performed comparably to the predicate for cutting, but specific metrics are not given.
    • "Additional comparative bench testing between the subject and secondary predicate device was conducted to demonstrate the optical performance of the HOPKINS Telescope that is part of the subject UNIDRIVE SIII system to meet its intended use." This indicates the optical performance was deemed sufficient but lacks specific criteria or quantitative results.
  2. Sample size used for the test set and the data provenance: Not applicable, as no clinical or analytical performance study (as implied by these questions) is detailed. The testing was non-clinical (bench testing).

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth as typically defined for AI/software evaluation (e.g., expert consensus on images) is not relevant here as it was a non-clinical bench testing and comparative analysis.

  4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. No MRMC study was conducted or mentioned. The device is not an AI-assisted diagnostic tool.

  6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This device is not an algorithm.

  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. For the non-clinical testing, "ground truth" would be the expected performance or measurements from the predicate devices and relevant engineering specifications.

  8. The sample size for the training set: Not applicable. Not an AI/ML device.

  9. How the ground truth for the training set was established: Not applicable.


Summary of Non-Clinical Performance Testing (as described in the document):

The device underwent various non-clinical tests to demonstrate substantial equivalence and adherence to recognized standards. These standards implicitly contain "acceptance criteria" for the device's functionality, safety, and performance.

Non-Clinical Performance Data / Tests Performed:

  • Hysteroscopes and Gynecologic Laparoscopes Guidance: Adherence to FDA guidance for these types of devices.
  • ISO Endoscopic Standards:
    • ISO 8600-1:2015
    • ISO 8600-3:2019
    • ISO 8600-5:2020
  • Biocompatibility Summary (ISO 10993 series):
    • Cytotoxicity (ISO 10993-5)
    • Acute Systemic Toxicity (ISO 10993-11)
    • Intracutaneous Irritation (ISO 10993-10)
    • Maximization Sensitization (ISO 10993-10)
  • Electrical Safety and EMC (IEC 60601 series):
    • IEC 60601-1:2012 (3rd Edition)
    • IEC 60601-2-18:2009 (3RD Edition)
    • IEC 60601-1-2:2014 (4th Edition)
  • Reprocessing (Cleaning and Sterilization) (AAMI and ISO standards):
    • AAMI TIR12: 2010
    • AAMI TIR30: 2011
    • AAMI TIR39: 2009
    • ANSI/AAMI ST8: 2013
    • ANSI/AAMI ST77:2013
    • ANSI/AAMI ST79:2017
    • ANSI/AAMI ST81:2004/(R)2010
    • AAMI/ISO 14937:2009
    • ANSI/AAMI/ISO 17655-1:2006/2013
    • ANSI/AAMI/ISO 17655-2:2009
    • Reprocessing Medical Device in Health Care Settings: Validation Methods and Labeling
  • Software Verification and Validation Testing:
    • Guidance for the Content of Premarket Submissions for Software Contained in Medical Device
  • Cutting Performance: Successfully demonstrated by comparing to the primary predicate device.
  • Optical Performance: Demonstrated through comparative bench testing with the secondary predicate device for the HOPKINS Telescope.

Conclusion stated: The nonclinical tests demonstrate that the subject device is as safe and effective as the primary and secondary predicate devices to support a substantial equivalence determination.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue.

April 27, 2023

Karl Storz-Endoscopy-America, Inc. Alita McElroy Senior Regulatory Affairs Specialist 2151 E. Grand Avenue El Segundo, CA 90245

Re: K223520 Trade/Device Name: UNIDRIVE SIII System Regulation Number: 21 CFR§ 884.1690 Regulation Name: Hysteroscope and Accessories Regulatory Class: II Product Code: HIH Dated: March 27, 2023 Received: March 28, 2023

Dear Alita McElroy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jason Roberts -S

Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223520

Device Name UNIDRIVE SIII System

Indications for Use (Describe)

The UNIDRIVE SIII System is intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue.

The HOPKINS Telescope when used with the obturator is intended to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and consists of the word "STORZ" in large, bold letters, with the "O" in the word being a circle. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller, but still bold, letters.

510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.

Submitter:KARL STORZ Endoscopy-America, Inc.2151 E. Grand AvenueEI Segundo, CA 90245
Contact:Alita McElroySenior Regulatory Affairs SpecialistPhone: (424) 218-8376Email: Alita.McElroy@karlstorz.com
Date ofPreparation:April 26, 2023
Type of 510(k)Submission:Traditional
DeviceIdentification:Trade Name: UNIDRIVE SIII SystemClassification Name: Hysteroscope and accessories (21 CFR 884.1690)
Regulatory Class:II
Product Code:HIH
ClassificationPanel:Obstetrics/Gynecology
PredicateDevice(s):Primary Predicate: TRUCLEAR Morcellator System and TRUCLEARMorcellators (K132015)The primary predicate has been subject to a design related recallSecondary Predicate: KARL STORZ Magnifying Hysteroscopes (K935716)The secondary predicate has not been subject to any recall
DeviceDescription:The UNIDRIVE® SIII system is a motorized surgical device intended forintrauterine use to hysteroscopically resect and remove tissue. TheUNIDRIVE SIII System includes the following components:UNIDRIVE SIII Control Unit (20701020-1) Footswitch (20016230) DrillCut-X II Shaver handpiece GYN (26702050) Shaver blades (26208SA, 26208SB) HOPKINS Telescope (26208AMA) Obturator (26208OC)
The UNIDRIVE S III system has a motor control unit and can be used with the DrillCut-X II Shaver handpiece GYN which houses the shaver blades. The shaver blade upon being connected to the DrillCut-X II Shaver handpiece GYN is inserted into the working channel of the HOPKINS Telescopes. The shaver blades are activated via a one-pedal footswitch that is connected the UNIDRIVE SIII control unit. The shaver blades consist of two sheaths that fit into each other. The inner sheath oscillates within the outer sheath to enable tissue resection.
Indications forUse:The UNIDRIVE S III System is intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue.The HOPKINS Telescope when used with the obturator is intended to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.
TechnologicalCharacteristics.The subject device, UNIDRIVE SIII System is substantially equivalent to the TRUCLEAR Morcellator System and TRUCLEAR Morcellators, (510(k) Number: K132015), marketed by Medtronic as well as to secondary predicate KARL STORZ Magnifying Hysteroscopes (510(k) Number: K935716).The subject and primary predicate devices are intended for intrauterine use for resection and removal of tissue. The main differences between the subject and primary predicate devices include the following:
The subject device includes a control unit, footswitch shaver handpiece, shaver blade, hysteroscope and an obturator while the primary predicate does not include a hysteroscope and obturator but includes other common components as the subject device. However the primary predicate device requires a hysteroscope to achieve its intended use similar to the subject device.
The subject device includes oscillating shaver blades and has a default rotational speed of 2100rpm and maximum rotational speed of 5000rpm.The primary predicate device includes oscillating shaver blade and reciprocating shaver blade. When used with the oscillating shaver blade, primary predicate device has a default rotational speed of 800rpm and maximum speed of 1500rpm. When used with the reciprocating blade, the primary predicate device has a default rotational speed of 1100rpm and maximum speed of 2500rpm.
The subject device shaver handpiece and shaver blades are provided non-sterile and are reusable while the primary predicate include sterile single-use shaver handpiece and shaver blades.
The secondary predicate is chosen to support the diagnostic and surgical hysteroscopic indications of the subject device. Both the subject device and secondary predicate device include reusable rigid rod-lens scopes that are used with external light sources a light cable, a camera head, and/or a Camera Control Unit (CCU). The subject device endoscope includes a working length of 21cm while the secondary predicate endoscopes have working length in the range of 18.5cm-30cm. The main differences between the subject and secondary predicate devices include the following:
The outer diameter of the subject device is 7.3mm and the secondary predicate includes endoscopes with outer diameters that ranges from 2.4mm to 4mm.
The subject device includes an endoscope with 6° direction of views while the secondary predicate device includes endoscopes with direction of view of 0°, 30°, 45°, 70° & 120°.
The subject device endoscope has a field of view of 105° ± 3° whereas the secondary predicate device includes endoscopes with a field of view of 67°- 92° ± 4°.
The subject device endoscope has a working channel diameter of 4.3mm while the secondary predicate endoscopes do not have a working channel.
Non-ClinicalPerformanceData:There are no performance standards or special controls developed under Section 514 of the FD&C Act for the UNIDRIVE SIII System. However, the UNIDRIVE SIII System follows the FDA recognized consensus standards and is tested according to the following standards and FDA Guidance:
Hysteroscopes and Gynecologic Laparoscopes Guidance Hysteroscopes and Gynecologic Laparoscopes - Submission Guidance for a 510(k)
ISO Endoscopic Standards ISO 8600-1:2015 ISO 8600-3:2019 ISO 8600-5:2020

K223520 Page 1 of 5

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller, but still bold, letters.

K223520
Page 2 of 5

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller letters.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller letters.

Biocompatibility Summary

  • Cytotoxicity (ISO 10993-5)
  • . Acute Systemic Toxicity (ISO 10993-11)
  • Intracutaneous Irritation (ISO 10993-10) ●
  • Maximization Sensitization (ISO 10993-10) ●

Electrical Safety and EMC

  • IEC 60601-1:2012 (3rd Edition) .
  • IEC 60601-2-18:2009 (3RD Edition) ●
  • IEC 60601-1-2:2014 (4th Edition)

Reprocessing (Cleaning and Sterilization)

  • AAMI TIR12: 2010 ●
  • AAMI TIR30: 2011 ●
  • AAMI TIR39: 2009 ●
  • ANSI/AAMI ST8: 2013 ●
  • ANSI/AAMI ST77:2013 ●
  • ANSI/AAMI ST79:2017
  • ANSI/AAMI ST81:2004/(R)2010
  • AAMI/ISO 14937:2009 ●
  • ANSI/AAMI/ISO 17655-1:2006/2013 ●
  • ANSI/AAMI/ISO 17655-2:2009 ●
  • Reprocessing Medical Device in Health Care Settings: Validation ● Methods and Labeling

Software Verification and Validation Testing

  • Guidance for the Content of Premarket Submissions for Software ● Contained in Medical Device
    The cutting performance of the subject device, UNIDRIVE SIII system has been successfully demonstrated by comparing to the primary predicate device TRUCLEAR Morcellator System and TRUCLEAR Morcellators cleared in K132015.

Additional comparative bench testing between the subject and secondary predicate device was conducted to demonstrate the optical performance of the HOPKINS Telescope that is part of the subject UNIDRIVE SIII system to meet its intended use.

Clinical Clinical Testing was not required to demonstrate the substantial equivalence Performance to the predicate devices. Non-clinical testing was adequate to establish Data: substantial equivalence.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller letters.

Conclusion:The conclusions drawn from the nonclinical tests demonstrate that thesubject device is as safe and effective as the primary and secondary predicatedevices to support a substantial equivalence determination.
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

§ 884.1690 Hysteroscope and accessories.

(a)
Identification. A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.(b)
Classification. (1) Class II (performance standards).(2) Class I for hysteroscope accessories that are not part of a specialized instrument or device delivery system; do not have adapters, connectors, channels, or do not have portals for electrosurgical, laser, or other power sources. Such hysteroscope accessory instruments include: lens cleaning brush, cannula (without trocar or valves), clamp/hemostat/grasper, curette, instrument guide, forceps, dissector, mechanical (noninflatable), and scissors. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.