K Number
K223424
Device Name
Spine Auto Views
Date Cleared
2023-07-13

(241 days)

Product Code
Regulation Number
892.1750
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Spine Auto Views is a non-invasive image analysis software package which may be used in conjunction with CT images to aid in the automatic generation of anatomically focused multi-planar reformats and automatically export results to predetermined DICOM destinations.

Spine Auto Views assists clinicians by providing anatomically focused reformats of the spine, with the ability to apply anatomical labels of the vertebral bodies and intervertebral disc spaces.

Spine Auto Views may be used for multiple care areas and is not specific to any disease state. It can be utilized for the review of various types of exams including trauma, oncology, and routine body.

Device Description

Spine Auto Views is a software analysis package designed to generate batch reformats and apply labels to the spine. It is intended to streamline the process of generating clinically relevant batch reformat outputs that are requested for many CT exam types.

Spine Auto Views can generate, automatically, patient specific, anatomically focused spine reformats. Spine Auto Views brings a state-of-the-art deep learning algorithm that generates oblique axial reformats, appropriately angled through each disc space without the need for a user interface and human interaction. 3D curved coronal and curved sagittal views of the spine as well as traditional reformat planes can all be generated with Spine Auto Views, no manual interaction required. Vertebral bodies and disc spaces can be labeled, and all series networked to desired DICOM destination(s), ready to read. The automated reformats may help in providing a consistent output of anatomically orientated images, labeled, and presented to the interpreting physician ready to read.

AI/ML Overview

Here's an analysis of the acceptance criteria and study proving the device meets them, based on the provided text:


Acceptance Criteria and Device Performance Study for Spine Auto Views

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria DefinitionReported Device Performance (Spine Auto Views)
Algorithm Capability to Automatically Detect Intervertebral Discs: Successful detection of position and orientation of intervertebral discs.Passed. The algorithm successfully passed the defined acceptance criteria for automatically detecting the position and orientation of intervertebral discs using a database of retrospective CT exams.
User Acceptance of Oblique Axial Reformats: Reader acceptance of automatically generated oblique axial reformats.Greater than 95% of the time for all readers. The reader evaluation concluded that Spine Auto Views oblique axial reformats generated user-acceptable results over 95% of the time for all readers.
User Acceptance of Curved Coronal and Curved Sagittal Reformats: Reader assessment of automatically generated curved coronal and curved sagittal batch reformats compared to standard views.Assessed by readers. The batch reformats of curved coronal and curved sagittal views were assessed by readers by comparing them with corresponding standard coronal and sagittal views. (Specific success rate not quantified in the provided text, but implied as satisfactory given the overall conclusion).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: A database of retrospective CT exams for algorithm validation and a sample of clinical CT images for the reader study. The exact number of exams/images for each is not specified in the document.
  • Data Provenance: The document states that the database of exams for algorithm validation was "representative of the clinical scenarios where Spine Auto Views is intended to be used, with consideration of acquisition parameters and patient characteristics." No specific country of origin is mentioned, but the mention of "retrospective CT exams" confirms the nature of the data.

3. Number of Experts and Qualifications for Ground Truth

The document does not explicitly state the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") used to establish the ground truth for the test set.

4. Adjudication Method for the Test Set

The document does not specify an adjudication method (e.g., 2+1, 3+1). For the reader study, it indicates that readers assessed the reformats, but it doesn't detail how discrepancies or consensus were handled if multiple readers were involved in rating the same case.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? A reader study was performed to assess user acceptance, which is a form of MRMC study in that multiple readers evaluated cases. However, it was not explicitly a "comparative effectiveness study" with and without AI assistance as described in the prompt.
  • Effect Size (AI vs. No AI): The study focused on the acceptance of the AI-generated reformats, implying a comparison against traditional manual generation (which would be "without AI assistance" for reformats). The text states that "Spine Auto Views oblique axial reformats generates user acceptable results greater than 95% of the time for all readers." This indicates a high level of acceptance for the AI-generated images, suggesting a positive effect compared to the traditional manual process, which the AI aims to streamline and automate. However, a quantitative effect size in terms of clinical improvement or efficiency gain compared to a "without AI" baseline is not provided. The device's primary benefit is automation ("no manual interaction required"), implying an improvement in workflow efficiency and consistency over manual methods.

6. Standalone (Algorithm Only) Performance

  • Was standalone performance done? Yes, an engineering validation of the Spine Auto Views algorithm's capability to automatically detect the position and orientation of intervertebral discs was performed as a standalone assessment. This evaluated the algorithm's performance independent of human interaction or interpretation in a clinical setting.

7. Type of Ground Truth Used

  • Algorithm Validation: The ground truth for the algorithm validation (disc detection) seems to have been established through a reference standard derived from the "database of retrospective CT exams." While the exact method (e.g., manual expert annotation, pathology correlation) is not explicitly stated, it implies a reliable, established truth against which the algorithm's detections were compared.
  • Reader Study: For the reader study, the "user acceptable results" and comparison with "corresponding standard coronal and standard sagittal views" suggest a ground truth based on expert consensus/clinical acceptability by the evaluating readers.

8. Sample Size for the Training Set

The document does not specify the sample size for the training set used for the deep learning algorithm. It only mentions a "database of retrospective CT exams" for validation.

9. How Ground Truth for Training Set Was Established

The document does not provide details on how the ground truth for the training set was established. It only refers to the validation set and its ground truth.

{0}------------------------------------------------

July 13, 2023

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

GE Medical Systems SCS % Yonghui Han Senior Regulatory Affairs Leader 283, rue de la Miniere BUC, 78530 FRANCE

Re: K223424

Trade/Device Name: Spine Auto Views Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK, QIH Dated: June 9, 2023 Received: June 12, 2023

Dear Yonghui Han:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223424

Device Name Spine Auto Views

Indications for Use (Describe)

Spine Auto Views is a non-invasive image analysis software package which may be used in conjunction with CT images to aid in the automatic generation of anatomically focused multi-planar reformats and automatically export results to predetermined DICOM destinations.

Spine Auto Views assists clinicians by providing anatomically focused reformats of the spine, with the ability to apply anatomical labels of the vertebral bodies and intervertebral disc spaces.

Spine Auto Views may be used for multiple care areas and is not specific to any disease state. It can be utilized for the review of various types of exams including trauma, oncology, and routine body.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K223424

Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. The circle has a swirling design around the letters, giving the impression of motion or energy. The logo is simple, recognizable, and associated with a well-known multinational corporation.

GE HealthCare 510(k) Premarket Notification Submission -Spine Auto Views

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date:July 7, 2023
Submitter:GE Medical Systems SCS
Establishment Registration Number - 9611343
283 rue de la Miniere
78530 Buc, France
Primary Contact Person:Yonghui Han
Senior Regulatory Affairs Leader
GE HealthCare
(+1) (262)225-1914
Yonghui.Han@ge.com
Secondary Contact Person:Elizabeth Mathew
Senior Regulatory Affairs Manager
GE HealthCare
Tel: (+1) (262)424-7774
Email: Elizabeth.Mathew@ge.com
Device Trade Name:Spine Auto Views
Common/Usual Name:Spine Auto Views
Primary Regulation Number:Computed Tomography X-Ray System (21 CFR 892.1750)
Primary Product Code:JAK
Secondary Product Code:QIH
Classification:Class II

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are white swirls around the letters, giving the impression of movement or energy. The logo is simple and recognizable, and it is often used to represent the company's products and services.

Predicate Device
Device name:Bone VCAR
Manufacturer:GE Medical Systems SCS
510(k) number:K183204
Regulation Number:21 CFR 892.1750 Computed tomography X-Ray System
Primary Product Code:JAK
Secondary Product Code:LLZ
Classification:Class II

Device Description:

Spine Auto Views is a software analysis package designed to generate batch reformats and apply labels to the spine. It is intended to streamline the process of generating clinically relevant batch reformat outputs that are requested for many CT exam types.

Spine Auto Views can generate, automatically, patient specific, anatomically focused spine reformats. Spine Auto Views brings a state-of-the-art deep learning algorithm that generates oblique axial reformats, appropriately angled through each disc space without the need for a user interface and human interaction. 3D curved coronal and curved sagittal views of the spine as well as traditional reformat planes can all be generated with Spine Auto Views, no manual interaction required. Vertebral bodies and disc spaces can be labeled, and all series networked to desired DICOM destination(s), ready to read. The automated reformats may help in providing a consistent output of anatomically orientated images, labeled, and presented to the interpreting physician ready to read.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. There are decorative swirls around the circle, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, representing the brand identity of General Electric.

Intended Use:

Spine Auto Views is a non-invasive image analysis software package which may be used in conjunction with CT images to aid in the automatic generation of anatomically focused multiplanar reformats.

Indication for Use:

Spine Auto Views is a non-invasive image analysis software package which may be used in conjunction with CT images to aid in the automatic generation of anatomically focused multiplanar reformats and automatically export results to pre-determined DICOM destinations.

Spine Auto Views assists clinicians by providing anatomically focused reformats of the spine, with the ability to apply anatomical labels of the vertebral bodies and intervertebral disc spaces.

Spine Auto Views may be used for multiple care areas and is not specific to any disease state. It can be utilized for the review of various types of exams including trauma, oncology, and routine body.

Technology:

The proposed device Spine Auto Views employs the same fundamental scientific technology as its predicate device.

Comparison:

The table below summarizes the key feature/technological differences and similarities between the predicate device and the proposed device:

Table 5.1: Comparison

SpecificationPredicate Device:Bone VCAR (K183204)Subject Device:Spine Auto ViewsComparison
CT Spine Labelingalgorithm(VertebraeDetection andVertebrae Labels)YesYesIdentical
SpecificationPredicate Device:Bone VCAR (K183204)Subject Device:Spine Auto ViewsComparison
Disc Detectionalgorithm(Disc Position, DiscOrientation & DiscLabels)Not availableYesSubstantiallyEquivalent.Spine Auto Viewsintroduces a new deeplearning Disc detectionalgorithm which candetect discs positions,planes/orientationsand generate disclabels.
Image DisplayFormatsTraditional reformatorientations (axial, coronal,sagittal)Traditional reformat orientations(axial, coronal, sagittal)Identical
Curved coronal and curvedsagittal views of the spineas well as oblique axialviews perpendicular to the3D curved trace.Curved coronal and curved sagittalviews of the spine as well as obliqueaxial views through theintervertebral spaces.SubstantiallyEquivalentBoth the subject deviceand predicate devicecan output curvedcoronal and curvedsagittal views andoblique axial view.However, themethodology in whichthese reformat seriesare generated aredifferent between thesubject device andpredicate device.For the curvedreformats, Bone VCARuses the centerlinetraced through thevertebrae centers;Spine Auto Views usethe centerline traced
SpecificationPredicate Device:Bone VCAR (K183204)Subject Device:Spine Auto ViewsComparison
edge of the disccenters.For oblique axial viewsgeneration, the obliqueaxial of Bone VCAR isperpendicular to itscenterline. Spine AutoViews uses thecomputed position andorientation of the discs.
Field of ViewBone VCAR uses inputvolume field of view andmanually adjusted by theuser.Spine Auto Views uses input volumefield of view and additionally has anoption to automatically set the fieldof view to focus on the spineSubstantiallyEquivalentThe subject device hastwo configurations:input series field ofview, or automatedfield of view focusedon the spine.
ExportManualAutomatedSubstantiallyEquivalentIn the subject device,all generated series canbe automaticallyexported to DICOMdestinations, such asPACS, ready for reviewand interpretation.
Measurement ToolAccess to all standardVolume Viewer tools formeasuring distances, areas,Hounsfield unit values andannotating within theimagesNot availableSubject device doesn'thave a user interfaceand doesn't includemeasurement tools.Subject device providesDICOM reformatimages for export.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. There are decorative swirls or droplets surrounding the circle, giving it a dynamic and recognizable appearance.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The frame is adorned with three droplet-like shapes evenly spaced around the circle. The logo is presented in a blue color.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script inside. There are three water droplet shapes around the outside of the circle.

Determination of Substantial Equivalence:

Summary of Non-Clinical, Design Control Testing

Spine Auto Views has successfully completed the design control testing per GE's quality system. It was designed and will be manufactured under the Quality System Regulations of 21CFR 820 and ISO 13485. No additional hazards were identified, and no unexpected test results were observed. The proposed device complies with NEMA PS 3.1 - 3.20 (2016) Digital Imaging and Communications in Medicine (DICOM) Set (Radiology) standard.

The following quality assurance measures were applied to the development of the device:

  • Requirements Definition
  • Risk Analysis
  • Technical Design Reviews ●
  • Formal Design Reviews
  • Software Development Lifecycle ●
  • Performance testing (Verification, Validation)
  • Safety Testing (Verification)

The proposed Spine Auto Views has been successfully verified on the EHL platform. All the testing and results did not raise new or different questions of safety and effectiveness other than those already associated with predicate device. Software documentation is for a MODERATE level of concern.

In addition, Engineering has validated Spine Auto Views algorithm capability to automatically detect the position and orientation of intervertebral discs using a database of retrospective CT exams. This database of exams is representative of the clinical scenarios where Spine Auto Views is intended to be used, with consideration of acquisition parameters and patient

characteristics. The result of the algorithm validation showed that the algorithm successfully passed the defined acceptance criteria.

Summary of Clinical Testing

A reader study evaluation was performed with a sample of clinical CT images which were processed with the Spine Auto Views software. The purpose of this study was to assess the acceptance of spine automated reformats generated in the oblique axial, curved coronal and curved sagittal planes as these orientations traditionally require users to manually generate each of them. The reader evaluation concluded that Spine Auto Views oblique axial reformats generates user acceptable results greater than 95% of the time for all readers. Additionally, the batch reformats of curved coronal and curved sagittal views were also assessed by the readers by comparing the automatically generated curved batch reformats with the corresponding standard coronal and standard sagittal views.

Conclusion:

Spine Auto Views has the same intended use as its predicate device but differs in technological characteristics. Specifically Spine Auto Views automatically generates spine reformat DICOM

5 б

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are white swirls around the edge of the circle. The logo is simple and recognizable.

images by detecting the vertebras and discs in the scan while the predicate requires more manual interactions.

GE HealthCare's quality system's design, verification, and risk management processes did not identify any new questions of safety or effectiveness, hazards, unexpected results, or adverse effects stemming from the changes to the predicate.

Based on development under GE HealthCare's quality system, successful design verification, software documentation for a "Moderate" level of concern, along with the engineering bench testing and reader study, GE HealthCare determines that the proposed device is as safe and effective as its predicate device.

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.