K Number
K223373

Validate with FDA (Live)

Device Name
Enivo™
Date Cleared
2023-04-07

(154 days)

Product Code
Regulation Number
878.4780
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Enivo™ is indicated for use to remove surgical and bodily fluids from a closed wound for hematoma and seroma prophylaxis following plastic surgery or other general surgeries where large flaps are formed.

Device Description

Enivo™ is a surgical drainage system intended for the removal of surgical and bodily fluids from a closed wound. The system can be used in both home and healthcare environments. The system includes two primary components: - A vacuum device unit that includes a single-use, portable, and battery-powered vacuum device that provides continuous operation for up to 30 days and a disposable exudate reservoir. - a removable silicone drainage catheter

AI/ML Overview

The provided document is a 510(k) summary for the Enivo™ device, which is a powered suction pump. It outlines the device's intended use and compares it to a predicate device (SOMAVAC Device, K222856) to establish substantial equivalence.

Here's an analysis of the acceptance criteria and supporting studies based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not provide a table with explicit acceptance criteria (e.g., minimum vacuum pressure, maximum flow rate) and corresponding measured device performance values. Instead, it lists various types of non-clinical tests that were conducted to demonstrate that "the device performs as intended" and "no new types of risk to the patent have been introduced by these differences." Without specific criteria, it's impossible to create such a table.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document lists only non-clinical (bench) testing. Therefore, there is no "test set" in the context of patient data (e.g., diagnostic images, clinical samples), nor is there data provenance related to patient demographics or study design (retrospective/prospective). The tests listed are primarily engineering and performance evaluations.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This question is not applicable. The studies conducted are non-clinical engineering and performance tests, not studies requiring expert interpretation of patient data to establish ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This question is not applicable. There is no "test set" in the context of patient data that would require an adjudication method.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. The Enivo™ device is a powered suction pump, not an AI-assisted diagnostic or therapeutic device. No MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable. The Enivo™ device is a hardware device; it does not involve algorithms or AI for standalone performance evaluation in this context.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical tests conducted, the "ground truth" would be established by:

  • Voluntary Standards: Compliance with recognized standards like IEC 60601-1, ISO 10993-1, ISO 20697, ASTM F1929, etc.
  • Engineering Specifications: Internal design specifications for vacuum pressure, flow rate, battery life, mechanical strength, alarm thresholds, etc.
  • Test Method Requirements: The specifications and procedures defined in the various ASTM and ISO standards for package integrity, sterilization, biocompatibility, etc.

8. The sample size for the training set

This question is not applicable. The Enivo™ device is not an AI/machine learning device, so there is no concept of a "training set" in this submission.

9. How the ground truth for the training set was established

This question is not applicable as there is no training set mentioned or implied for this device.


Summary of Acceptance Criteria and Supporting Studies (Based on available information):

Due to the nature of the device (a powered suction pump) and the 510(k) submission type, the acceptance criteria are based on compliance with recognized consensus standards and internal engineering performance specifications rather than clinical performance metrics typically seen for diagnostic or AI-powered devices.

The document lists a comprehensive set of non-clinical tests to demonstrate substantial equivalence by showing that the device meets safety and performance requirements and does not introduce new risks compared to the predicate device. However, specific numerical acceptance criteria and reported values for each test are not detailed in this summary.

In essence, the "acceptance criteria" can be inferred as "compliance with the listed standards and satisfactory performance per internal specifications," and the "study that proves the device meets the acceptance criteria" is the entirety of the non-clinical testing performed.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

April 7, 2023

Aroa Biosurgery Ltd. Tina O'brien Director, Regulatory Affairs 2 Kingsford Smith Place Airport Oaks, Auckland 2022 New Zealand

Re: K223373

Trade/Device Name: Enivo™ Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: BTA, Dated: March 10, 2023 Received: March 10, 2023

Dear Tina O'brien:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2023.04.07 13:28:45 -04'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

K223373

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Expiration Date: 06/30/2023

See PRA Statement below.

Indications for UseSee PRA Statement below.
Submission Number (if known)K223373
Device NameEnivo™
Indications for Use (Describe)Enivo™ is indicated for use to remove surgical and bodily fluids from a closed wound for hematoma and seroma prophylaxis following plastic surgery or other general surgeries where large flaps are formed.
Type of Use (Select one or both, as applicable)☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K223373
510(k) #: K223373510(k) Summary
Page 1 of 2
Prepared on: 2023-04-07
Contact Details21 CFR 807.92(a)(1)
Applicant NameAroa Biosurgery Ltd.
Applicant Address2 Kingsford Smith Place Airport Oaks AKL 2022 New Zealand
Applicant Contact Telephone+64 098693035
Applicant ContactMs. Tina O'Brien
Applicant Contact Emailtina.obrien@aroabio.com
Device Name21 CFR 807.92(a)(2)
Device Trade NameEnivo™
Common NamePowered suction pump
Classification NamePump, Portable, Aspiration (Manual Or Powered)
Regulation Number878.4780
Product CodeBTA

Legally Marketed Predicate Devices

21 CFR 807.92(a)(3)
Predicate #Predicate Trade Name (Primary Predicate is listed first)
K222856SOMAVAC Device
Product Code
BTA
Device Description Summary21 CFR 807.92(a)(4)
-------------------------------------------------

Enivo™ is a surgical drainage system intended for the removal of surgical and bodily fluids from a closed wound. The system can be used
in both home and healthcare environments.

The system includes two primary components:

  • A vacuum device unit that includes a single-use, portable, and battery-powered vacuum device that provides continuous operation for
    up to 30 days and a disposable exudate reservoir.
  • a removable silicone drainage catheter
Intended Use/Indications for Use21 CFR 807.92(a)(5)
-------------------------------------------------------

Enivo™ is indicated for use to remove surgical and bodily fluids from a closed wound for hematoma and seroma prophylaxis following
plastic surgery or other general surgeries where large flaps are formed.

Indications for Use Comparison21 CFR 807.92(a)(5)
-----------------------------------------------------

Enivo™ shares the same intended use and indications as the predicate.

Technological Comparison21 CFR 807.92(a)(6)
-----------------------------------------------

The predicate device uses the same fundamental technology and use configuration as Enivo™ in that it is required to a Class I drainage catheter. The differences between Enivo™ and the predicate do not introduce a new intended use and do not raise new

{4}------------------------------------------------

issues of safety and effectiveness. Bench testing has demonstrated that no new types of risk to the patent have been introduced by these differences and that the device performs as intended.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The Enivo™ system complies with voluntary standards for electromagnetic compatibility, and sterile drainage catheters. The following data was provided in support of the substantial equivalence determination:

  • Software verification and validation testing

  • Electrical safety and electromagnetic compatibility testing were conducted per IEC 60601 1:2005+AMD1:2012+AMD2:2020 with US deviations per AAM ES60601-1:2005+AMD1:2012+AMD2:2021, IEC 60601-1-2:2014+AMD1:2020, IEC 60601-1-8:2006+AMD1:2012 +AMD2:2020, IEC 60601-1-11:2015+AMD1:2020, and AIM 7351731:2017

  • Biocompatibility evaluated as per ISO 10993-1:2018

  • Vacuum pressure and flow rate performance testing

  • Battery life testing

  • Endurance runtime via simulated use testing

  • Performance and safety assessed according to the applicable parts of ISO 20697:2018

  • Package integrity assessed as per ASTM F1929:2015, ASTM F2096:2019, and ISO 11607-1:2019

  • Package seal strength assessed as per ASTM F88/F88M:2015

  • Bacterial filtration efficiency assessed as per ASTM F2101:2019

  • Sterilization confirmed by assessment in accordance with ISO 11135:2018

  • Bacterial endotoxins assessed as per ANSI/AAMI ST72:2019

  • Bioburden (pre-EO) assessed as per ISO 11737-1:2021

  • Connection leak testing

  • Quantification of connection forces via uniaxial tensile and compression testing

  • Mechanical strength assessed using uniaxial tensile and compression testing

  • Alarm system assessed to perform as indicated in the instructions

  • System performance assessed following simulated transportation testing as per ASTM D4169:2022

Clinical testing was not required to demonstrate the substantial equivalence of Enivo™ to the predicate device.

The intended use and indications for use of the substantially equivalent to the predicate device. The differences between Enivo™ and the predicate device do not intended use and do not raise new issues of safety and effectiveness. Testing demonstrated that no adverse effects have been introduced by the technological differences and that Enivo™ performs as intended.

From the results of nonclinical testing described, Aroa Biosurgery concludes that Enivo™ is substantially equivalent to the legally marketed predicate device.

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.