K Number
K223352

Validate with FDA (Live)

Device Name
Tenderfoot
Date Cleared
2023-03-02

(120 days)

Product Code
Regulation Number
878.4850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Tenderfoot is a sterile incision device intended to initiate capillary blood flow via a heel stick with infants and toddlers. For medical device use. For healthcare professional use, prescription only.

Device Description

Tenderfoot is an automated skin incision device used to collect capillary blood from infants and toddlers by heel stick. The device produces an arc-like incision at a controlled depth to provide blood flow for obtaining a blood sample. The incision is made by a surgical blade that is completely enclosed in a plastic housing and deployed by pressing a trigger. Following deployment, the blade permanently retracts within the housing, rendering the device inoperable for further use. Tenderfoot is provided as a sterile, single-use disposable device intended for use by healthcare professionals only. Sterilization is achieved by gamma irradiation. Tenderfoot is offered in four models covering a range of incision depths sized for different infant populations: Micro-Preemie, Preemie, Newborn, and Toddler.

AI/ML Overview

The provided text is a 510(k) summary for a medical device called "Tenderfoot". This document focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets strict acceptance criteria through a clinical or algorithmic performance study in the way one might for a novel AI/ML device.

Therefore, the information requested in the prompt, particularly regarding MRMC studies, standalone performance, ground truth establishment for test/training sets, and expert adjudication, is not present in this document. This typically applies to AI/ML software as a medical device (SaMD) where performance is measured against reference standards. The Tenderfoot device is a physical, mechanical blood lancet.

However, I can extract the information that is available about the "performance data" provided.


Understanding the Context

The Tenderfoot device is a physical medical device (a heel incision device/lancet), not a software or AI/ML product. The "performance data" referred to in Section VII of the 510(k) summary are non-clinical bench testing and biocompatibility/sterilization/packaging validation, relevant for mechanical devices and their safety, not for evaluating algorithmic performance or diagnostic accuracy.

The FDA's review for this type of device (a Class II blood lancet, product code FMK) primarily focuses on demonstrating that the device is substantially equivalent to a legally marketed predicate device, meaning it is as safe and effective as the predicate. This often involves comparing technological characteristics and validating that the device meets relevant performance standards (e.g., incision depth/length accuracy, sharps injury prevention features, sterility, biocompatibility).


What Can Be Extracted from the Provided Text:

Given the nature of the device and the document, the acceptance criteria and supporting study details will pertain to physical performance and safety, not diagnostic accuracy or AI/ML performance.

1. A table of acceptance criteria and the reported device performance:

The document states: "Nonclinical bench testing was performed for cut depth, cut length, trigger activation, and blade retraction before and after stress conditions, such as simulated distribution and aging. The testing met all acceptance criteria."

While the specific numerical acceptance criteria and reported numerical performance values are not detailed in this summary, the statement confirms that the device did meet predefined standards for these physical aspects.

Acceptance Criterion (Category)Reported Device Performance
Nonclinical Bench Testing
Cut DepthMet all acceptance criteria
Cut LengthMet all acceptance criteria
Trigger ActivationMet all acceptance criteria
Blade RetractionMet all acceptance criteria
Performance after Stress Conditions(Simulated distribution & aging)Met all acceptance criteria
Biocompatibility
CytotoxicitySuccessfully completed
SensitizationSuccessfully completed
IrritationSuccessfully completed
Acute systemic toxicitySuccessfully completed
Material-mediated pyrogen testingSuccessfully completed
HemocompatibilitySuccessfully completed
Sterilization
Sterility Assurance Level (SAL)10-6 (validated per ISO 11137-1)
Packaging Performance
Visual inspection of seal widthSuccessfully completed (per ISO 11607-1)
Seal strength (ASTM F88/F88M)Successfully completed
Gross leaks (ASTM F2096)Successfully completed
Microbial barrier (ASTM F1608)Successfully completed
Shelf Life
Packaging shelf life4 years (established through accelerated aging per ASTM F1980)

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not explicitly stated for each test. The summary only mentions "Nonclinical bench testing was performed..." without specifying the number of units tested.
  • Data Provenance: Not applicable in terms of "country of origin of the data" or "retrospective/prospective" for this type of physical device testing. The tests are bench tests performed on the manufactured devices.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This is not applicable as the device is a physical lancet, not an AI/ML diagnostic or image interpretation device that requires human expert-established ground truth. The "ground truth" for physical properties like cut depth/length would be measured directly by instruments.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable for this type of device and testing. Adjudication methods are typically used for establishing ground truth in clinical image or data analysis.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC study was not done. This type of study is specifically for evaluating the performance of diagnostic devices, often AI-assisted, where human readers interpret medical images or data. The Tenderfoot device is a physical instrument for blood collection.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • No, a "standalone" study in the context of an algorithm was not done because the Tenderfoot is a physical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • For physical attributes like incision depth and length, the "ground truth" would be direct physical measurement using appropriate metrology tools, not expert consensus or pathology. For biocompatibility and sterilization, the ground truth is established by validated laboratory assays and standards (e.g., ISO 10993, ISO 11137).

8. The sample size for the training set:

  • Not applicable. This device does not involve a "training set" in the context of machine learning or AI.

9. How the ground truth for the training set was established:

  • Not applicable. As there is no training set (for AI/ML), there is no ground truth establishment for it.

Summary of missing information and reasons:

The prompt's questions are largely tailored to the evaluation of AI/ML-based medical devices or diagnostic software. The provided document is a 510(k) summary for a physical, mechanical blood lancet. Therefore, many of the requested details (especially those regarding AI performance, human reader studies, and expert ground truth establishment for data sets) are fundamentally not applicable to this type of device or the regulatory pathway it followed. The performance validation documented here pertains to the physical and safety characteristics of the device, demonstrated through bench testing and compliance with recognized standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.

March 2, 2023

Accriva Diagnostics, Inc. Wenni Haley Senior Principal Specialist, Regulatory Affairs 6260 Sequence Drive San Diego, California 92121

Re: K223352

Trade/Device Name: Tenderfoot Regulation Number: 21 CFR 878.4850 Regulation Name: Blood lancets Regulatory Class: Class II Product Code: FMK Dated: November 1, 2022 Received: November 2, 2022

Dear Wenni Haley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Carr -S

for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Accriva Diaqnostics, Inc. Tenderfoot - Special 510(k) K223352/S001

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known) K223352

Device Name

Tenderfoot Newborn, Tenderfoot Micro-Preemie, Tenderfoot Preemie, Tenderfoot Toddler

Indications for Use (Describe)

Tenderfoot is a sterile incision device intended to initiate capillary via a heel stick with infants and toddlers.

For medical device use. For healthcare professional use, prescription only.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

I. SUBMITTER

Applicant:Accriva Diagnostics, Inc.6260 Sequence DriveSan Diego, CA 92121USAEstablishment Registration Number: 3002721930
ApplicationCorrespondent:Wenni HaleySenior Principal Specialist, Regulatory AffairsPhone: 858-263-2322Email: whaley@werfen.com
Date Prepared:March 1, 2023

II. DEVICE

Trade Name:Tenderfoot
Common Name:Heel Incision Device
Classification Name:Single use only blood lancet with an integral sharps injury prevention feature
Regulation Number:21 CFR 878.4850
Regulatory Class:Class II
Product Code:FMK
Review Panel:General and Plastic Surgery (79)

III. PREDICATE DEVICE

Primary Predicate:Tenderfoot (K883968), Product Code: FMK
Additional Predicate:Tenderfoot (K911997), Product Code: FMK

IV. DEVICE DESCRIPTION

Tenderfoot is an automated skin incision device used to collect capillary blood from infants and toddlers by heel stick. The device produces an arc-like incision at a controlled depth to provide blood flow for obtaining a blood sample. The incision is made by a surgical blade that is completely enclosed in a plastic housing and deployed by pressing a trigger. Following deployment, the blade permanently retracts within the housing, rendering the device inoperable for further use.

{4}------------------------------------------------

Tenderfoot is provided as a sterile, single-use disposable device intended for use by healthcare professionals only. Sterilization is achieved by gamma irradiation.

Tenderfoot is offered in four models covering a range of incision depths sized for different infant populations: Micro-Preemie, Preemie, Newborn, and Toddler (see table below).

Tenderfoot ModelIncision DepthIncision LengthColor
Micro-Preemie0.65 mm1.40 mmBlue
Preemie0.85 mm1.75 mmWhite
Newborn1.00 mm2.50 mmBlue/Pink
Toddler2.00 mm3.00 mmPink

Tenderfoot is not marketed with accessories or as part of a system.

V. INDICATIONS FOR USE

Tenderfoot is a sterile incision device intended to initiate capillary blood flow via a heel stick with infants and toddlers.

For medical device use. For healthcare professional use, prescription only.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

A detailed comparison between the predicate, the currently marketed Tenderfoot device (K883968 and K911997), and the subject Tenderfoot device is provided in the table below to demonstrate substantial equivalence.

CharacteristicPredicateTenderfoot (K883968, K911997)Subject DeviceTenderfoot (this submission)
Similarities
Indications for Use /Intended UseTenderfoot is a sterile incision deviceintended to initiate capillary bloodflow via a heel stick with infants andtoddlers for use in diagnostic testing.For medical device use. For healthcareprofessional use, prescription only.Tenderfoot is a sterile incision deviceintended to initiate capillary bloodflow via a heel stick with infants andtoddlers.For medical device use. For healthcareprofessional use, prescription only.
CharacteristicPredicateSubject Device
Tenderfoot (K883968, K911997)Tenderfoot (this submission)
Principle ofOperationIncision made by a spring-loaded,cam-driven surgical steel blademoving in a single sweeping motion,actuated by pressing a trigger.Incision made by a spring-loaded,cam-driven surgical steel blademoving in a single sweeping motion,actuated by pressing a trigger.
Sharps InjuryPrevention Features1. Blade permanently retractsfollowing deployment.2. Housing completely encloses theblade and prevents finger access.3. Detachable trigger guard preventsaccidental/premature actuation.4. Indicator arrow indicates exit pathof the blade.5. Device can be activated with asingle hand.6. Trigger remains visibly depressedto show device has already beenactivated.1. Blade permanently retractsfollowing deployment.2. Housing completely encloses theblade and prevents finger access.3. Detachable trigger guard preventsaccidental/premature actuation.4. Indicator arrow indicates exit pathof the blade.5. Device can be activated with asingle hand.6. Trigger remains visibly depressedto show device has already beenactivated.
Incision Profiles(Depth/Length)Micro-Preemie: 0.65 mm / 1.40 mmPreemie: 0.85 mm / 1.75 mmNewborn: 1.00 mm / 2.50 mmToddler: 2.00 mm / 3.00 mmMicro-Preemie: 0.65 mm / 1.40 mmPreemie: 0.85 mm / 1.75 mmNewborn: 1.00 mm / 2.50 mmToddler: 2.00 mm / 3.00 mm
Housing ColorsMicro-Preemie: BluePreemie: WhiteNewborn: Blue/PinkToddler: PinkMicro-Preemie: BluePreemie: WhiteNewborn: Blue/PinkToddler: Pink
HousingDimensionsWidth: 1.3 inHeight: 1.2 in (excluding triggerguard)Depth: 0.5 inWidth: 1.3 inHeight: 1.2 in (excluding triggerguard)Depth: 0.5 in
Incision BladeDesignDouble-honed surgical bladeDouble-honed surgical blade
MaterialsPatient-Contacting1. Housing: Polystyrene2. Surgical Blade: Stainless steelPatient-Contacting1. Housing: Polystyrene2. Surgical Blade: Stainless steel
Non-Patient Contacting3. Cam follower:Polycarbonate/acrylic alloy4. Torsion Spring: Stainless steel5. Trigger: PolystyreneNon-Patient Contacting3. Cam follower:Polycarbonate/acrylic alloy4. Torsion Spring: Stainless steel5. Trigger: Polystyrene
6. Trigger Guard: Polystyrene6. Trigger Guard: Polystyrene
CharacteristicPredicateTenderfoot (K883968, K911997)Subject DeviceTenderfoot (this submission)
Packaging SystemPrimary (sterile barrier)1. Blister Pack Tray: Polyethyleneterephthalate (PETG)2. Lid: TyvekSecondary3. Box: Clay-coated news backPrimary (sterile barrier)1. Blister Pack Tray: Polyethyleneterephthalate (PETG)2. Lid: TyvekSecondary3. Box: Clay-coated news back
Packaging (Boxed)ConfigurationsMicro-Preemie: 50 packPreemie: 50 pack, 200 pack, 1000packNewborn: 50 pack, 200 pack, 1000packToddler: 50 packMicro-Preemie: 50 packPreemie: 50 pack, 200 pack, 1000packNewborn: 50 pack, 200 pack, 1000packToddler: 50 pack
BiocompatibilityConforms to ISO 10993-1Conforms to ISO 10993-1
SterilizationGamma irradiationSterility Assurance Level (SAL) of10-6Gamma irradiationSterility Assurance Level (SAL) of10-6
Number of UsesSingle use onlySingle use only
Shelf Life4 years4 years
Differences
LabelingDoes not contain all labelinginformation required by the specialcontrols under 21 CFR878.4850(a)(2).Contains all labeling informationrequired by the special controls under21 CFR 878.4850(a)(2).

{5}------------------------------------------------

Accriva Diagnostics, Inc.

Tenderfoot - Special 510(k)

{6}------------------------------------------------

VII. PERFORMANCE DATA

The following nonclinical performance data were provided in support of the substantial equivalence determination and compliance with the special controls described in 21 CFR 878.4850(a)(2).

Performance Testing - Bench

Nonclinical bench testing was performed for cut depth, cut length, trigger activation, and blade retraction before and after stress conditions, such as simulated distribution and aging. The testing met all acceptance criteria.

{7}------------------------------------------------

Biocompatibility Testing

Biocompatibility evaluation for Tenderfoot was conducted in accordance with the FDA Guidance "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" and ISO 10993-1 "Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. Testing was successfully completed for the following endpoints:

  • . Cytotoxicity
  • Sensitization ●
  • Irritation ●
  • Acute systemic toxicity
  • Material-mediated pyrogen testing ●
  • Hemocompatibility ●

Sterilization and Shelf Life

Sterilization by gamma irradiation was validated per ISO 11137-1 and demonstrated a Sterility Assurance Level (SAL) of 10-6.

Sterile barrier packaging testing was conducted for packaging performance and stability in accordance with ISO 11607-1. Package integrity tests included:

  • Visual inspection of seal width
  • Seal strength (ASTM F88/F88M)
  • Gross leaks (ASTM F2096) ●
  • . Microbial barrier (ASTM F1608)

A packaging shelf life of 4 years was established through accelerated aging per ASTM F1980.

VIII. CONCLUSIONS

Based on the information provided in this submission, Accriva Diagnostics concludes that the subject device, Tenderfoot, is substantially equivalent to the predicate device (K883968 and K911997). The labeling differences between the subject and predicate do not raise new concerns of safety and effectiveness.

§ 878.4850 Blood lancets.

(a)
Single use only blood lancet with an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base (including an integral sharps injury prevention feature) that is used to puncture the skin to obtain a drop of blood for diagnostic purposes. The integral sharps injury prevention feature allows the device to be used once and then renders it inoperable and incapable of further use.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
Single use only blood lancet without an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
Multiple use blood lancet for single patient use only —(1)Identification. A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
Multiple use blood lancet for multiple patient use —(1)Identification. A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class III (premarket approval).(3)
Date PMA or notice of completion of a PDP is required: A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.