K Number
K223256
Device Name
Optimos™ Cystotome
Date Cleared
2023-04-05

(166 days)

Product Code
Regulation Number
876.4300
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Optimos™ Cystotome is intended for use as an electrosurgical accessory to cannulate the transgastric or transduodenal wall and into a pancreatic pseudocyst that is visibly bulging into the gastrointestinal tract. This device is supplied sterile and is intended for single use only.
Device Description
Optimos™ Cystotome is an electrode which uses heat generated from the application of high frequency (HF) energy through the monopolar electrode to puncture a hole in tissue. Following a fine-needle aspiration (FNA) procedure, it is positioned over a guidewire, so the applied part is located at the area of interest. The Optimos™ Cystotome is for use to drain pancreatic pseudocysts. The Optimos™ Cystotome is a single use device provided sterile to the user. No reprocessing is performed by the user. The device is provided in three sizes, which have different diameters and tip lengths. The device itself consists of 3 parts: handle, sheath, and tip.
More Information

No
The description focuses on the electrosurgical mechanism and physical characteristics of the device, with no mention of AI or ML.

Yes

The device is intended to cannulate the transgastric or transduodenal wall to drain pancreatic pseudocysts, which is a therapeutic intervention aimed at treating a medical condition.

No

The device is an electrosurgical accessory used to puncture and drain pancreatic pseudocysts, which is a therapeutic function, not a diagnostic one.

No

The device description clearly states it is an electrode and consists of physical parts (handle, sheath, and tip), indicating it is a hardware device. The performance studies also involve bench testing on tissue, further confirming it is not software-only.

Based on the provided information, the Optimos™ Cystotome is not an IVD (In Vitro Diagnostic) device.

Here's why:

  • IVD Definition: In vitro diagnostic devices are used to examine specimens taken from the human body (like blood, urine, or tissue samples) to provide information about a person's health. This testing is done outside of the body ("in vitro").
  • Optimos™ Cystotome Function: The Optimos™ Cystotome is a surgical tool used directly on tissue within the body ("in vivo") to puncture and drain a pancreatic pseudocyst. It uses electrosurgical energy to create a hole.
  • Lack of Specimen Analysis: The device does not analyze any bodily fluids or tissues in vitro to diagnose a condition or monitor health. Its purpose is a therapeutic/interventional procedure.

The description clearly indicates it's an electrosurgical accessory used for a surgical procedure, not for laboratory testing of specimens.

N/A

Intended Use / Indications for Use

The Optimos™ Cystotome is intended for use as an electrosurgical accessory to cannulate the transgastric or transduodenal wall and into a pancreatic pseudocyst that is visibly bulging into the gastrointestinal tract. This device is supplied sterile and is intended for single use only.

Product codes

KNS

Device Description

Optimos™ Cystotome is an electrode which uses heat generated from the application of high frequency (HF) energy through the monopolar electrode to puncture a hole in tissue. Following a fine-needle aspiration (FNA) procedure, it is positioned over a guidewire, so the applied part is located at the area of interest. The Optimos™ Cystotome is for use to drain pancreatic pseudocysts.

The Optimos™ Cystotome is a single use device provided sterile to the user. No reprocessing is performed by the user. The device is provided in three sizes, which have different diameters and tip lengths. The device itself consists of 3 parts: handle, sheath, and tip.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

transgastric or transduodenal wall, pancreatic pseudocyst

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Biocompatibility Testing:

  • Cytotoxicity: Reactivity grade - test sample: 0, reagent control: 0, negative control: 0, positive control: 4 (P)
  • Intracutaneous reactivity: No abnormal signs, Normal body weight increase, Mean score was 0.00 in sterile saline extracts and 0.00 in cottonseed oil extracts (P)
  • Sensitization (GPMT): No abnormal signs, Normal body weight increase, The sensitization score was observed 0.0 (P)
  • Acute systemic toxicity: No abnormal signs, Normal body weight increase, No abnormal signs (P)
  • Material mediated pyrogenicity test: No abnormal signs, Normal body weight increase, No animal showed body temperature increase of 0.5 °C or above (P)

Electrical safety and electromagnetic compatibility (EMC):

  • EN 60601-1 (2006) + A1:2013 General: P
  • IEC 60601-1:2005/AMD1:2012 Assessment of Basic Safety and Essential Performance of Medical Devices: P
  • IEC 60601-1-2 Collateral standard: Electromagnetic compatibility: P
  • IC 60601-2-2:2017 Particular standards: HF surgical equipment: P
  • IEC 60601-2-18 Particular equipment: endoscopic equipment: P

Bench Testing:

  • Unit tests included: visual test, dimensional test, continuity test, endoscopic compatibility test, device operability test, high frequency dielectric strength test, detachable connector retention test, injection & leakage test, and tensile strength test. All tests passed the set acceptance criteria.
  • System tests to examine the thermal effects on tissue was conducted with fine needle aspiration, guidewire, endoscope, ESU as a system. The test results showed that all test samples meet the required performance, puncture tissues, and are compatible with all devices/equipment working together.
  • Thermal effects were evaluated on both the subject device (Optimos™ Cystotome, Model: OCT1906, Sample size: 18) and the predicate device (Cystotome™ Cystoenterostomy Needle Knife, Manufacturer: Wilson-Cook, Model: CST-10, Sample size: 18) at 18, 100, and 120W on porcine stomach tissue and duodenum in the mucosa, submucosa, and muscularis propria.
  • Acceptance criteria: 1. The ability to puncture 2 type tissues (Porcine stomach and Duodenum). 2. Substantial equivalence of thermal damage: There should be no significant difference in measurement of size (length, width, and depth) of the thermal damage zone between subject device and predicate device.
  • Generator power setting: Mono-polar Type, Auto Cut Mode (400 – 500Vp); Minimum: 80W, Default: 100W, Maximum: 120W.
  • Thermal damage zone measurement: Measured under microscope and depth analyzed using histological strain, NADH.
  • Temperature-time history: Digital Thermometer used, highest temperature of stomach and duodenum tissue applied by subject device/predicate device were 69.7/60.6°C and 2.5/66.2°C respectively. Damage zone measured after tissue samples cooled to 37°C.
  • Results: The performance of the Optimos™ Cystotome was comparable to the Wilson-Cook CST-10 Cystotome Cystoenterostomy Needle Knife. The test results met the acceptance criteria.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K022595, Wilson-Cook Cystotome

Reference Device(s)

K150692, AXIOS Stent with Electrocautery Enhanced Delivery System

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 876.4300 Endoscopic electrosurgical unit and accessories.

(a)
Identification. An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

April 5, 2023

Taewoong Medical Co., Ltd. % Christy Foreman Senior Consultant Biologics Consulting Group Inc. 100 Daingerfield Rd., Suite 400 Alexandria, VA 22314

Re: K223256

Trade/Device Name: Optimos™ Cystotome Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: Class II Product Code: KNS Dated: March 7, 2023 Received: March 7, 2023

Dear Christy Foreman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any

1

Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/5 description: The image shows the text "Je An-S" in a large, bold font. The letters are black and stand out against a white background. There is a faint, light blue watermark in the background that is partially obscured by the text. The text is evenly spaced and centered in the image.

Je Hi An, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K223256

Device Name Optimos™ Cystotome

Indications for Use (Describe)

The Optimos™ Cystotome is intended for use as an electrosurgical accessory to cannulate the transgastric or transduodenal wall and into a pancreatic pseudocyst that is visibly bulging into the gastrointestinal tract. This device is supplied sterile and is intended for single use only.

Type of Use (Select one or both, as applicable)

× Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Optimos™ Cystotome is provided below.

1. SUBMITTER

| Applicant: | Taewoong Medical Co., Ltd.
14 Gojeong-ro, Wolgot-myeon, Gimpo-si,
Gyeonggi-do, Korea, 10022
Phone: +82-31-996-0641
Fax: +82-31-996-0645 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Yonjin Jeff Kim
Regulatory Affairs / Quality Management Dept
Phone: +82-31-993-0641-4 (Ext.# 125)
Fax: 82-31-996-0645-6
E-mail: jinjeff@stent.net |
| Submission Correspondent: | Christy Foreman, M.B.E.
Senior Consultant, Medical Devices
Phone: (301) 325-4245
Fax: (703) 548-7457
Email: cforeman@biologicsconsulting.com |
| Date Prepared: | March 2, 2023 |

2. DEVICE

Device Trade Name:Optimos™ Cystotome
Device Common Name:Cystotome
Classification Name876.4300, Endoscopic electrosurgical unit and accessories
Regulatory Class:II
Product Code:KNS

PREDICATE DEVICE AND REFERENCE DEVICES 3.

Predicate Device: K022595, Wilson-Cook Cystotome

Reference Device: K150692, AXIOS Stent with Electrocautery Enhanced Delivery System

The AXIOS Stent is provided as a reference device for the sole electrode (electrode tip).

4

DEVICE DESCRIPTION 4.

Optimos™ Cystotome is an electrode which uses heat generated from the application of high frequency (HF) energy through the monopolar electrode to puncture a hole in tissue. Following a fine-needle aspiration (FNA) procedure, it is positioned over a guidewire, so the applied part is located at the area of interest. The Optimos™ Cystotome is for use to drain pancreatic pseudocysts.

The Optimos™ Cystotome is a single use device provided sterile to the user. No reprocessing is performed by the user. The device is provided in three sizes, which have different diameters and tip lengths. The device itself consists of 3 parts: handle, sheath, and tip.

INTENDED USE/INDICATIONS FOR USE ട്.

The Optimos™ Cystotome is intended for use as an electrosurgical accessory to cannulate the transgastric or transduodenal wall and into a pancreatic pseudocyst that is visibly bulging into the gastrointestinal tract. This device is supplied sterile and is intended for single use only.

SUBSTANTIAL EQUIVALENCE 6.

Comparison of Indications

The Wilson-Cook Cystotome is intended for use as an electrosurgical accessory to cannulate the transgastric or transduodenal wall and into a pancreatic pseudocyst that is visibly bulging into the gastrointestinal tract. This device is supplied sterile and is intended for single use only.

The indications for use statement of the subject device are identical to that of the predicate device.

Technological Comparisons

The table below compares the key technological feature of the subject devices to the predicate device, Wilson-Cook Cystotome (K022595).

Proposed DevicePredicate Device
510(k) NumberTDBK022595
ApplicantTaewoong MedicalWilson-Cook
Device NameOptimosTM CystotomeCystotome
Classification Regulation876.4300
Endoscopic electrosurgical
unit and accessories876.4300
Endoscopic electrosurgical
unit and accessories
Product CodeKNSKNS
Proposed DevicePredicate Device
Indications for UseThe Optimos™ Cystotome is
intended for use as an
electrosurgical accessory to
cannulate the transgastric or
transduodenal wall and into a
pancreatic pseudocyst that is
visibly bulging into the
gastrointestinal tract. This
device is supplied sterile and
is intended for single use only.The Wilson-Cook Cystotome
is intended for use as an
electrosurgical accessory to
cannulate the transgastric or
transduodenal wall and into a
pancreatic pseudocyst that is
visibly bulging into the
gastrointestinal tract. This
device is supplied sterile and
is intended for single use only.
UseEndoscopicEndoscopic
Number of Electrodes12
Size of ElectrodesNeedle Tip – Not Applicable
Electrode Tip – 1.4 - 1.7 mm
(Diameter)
OCT1906: 1.4 mm
OCT1907: 1.7 mm
OCT1908: 1.7 mmNeedle tip - 0.097 mm
Electrode Tip – 2.15 mm
(Diameter)
CST-10: 2.15 mm
Size of SheathCatheter attached to Needle -
Not ApplicableInner catheter attached to
Needle Diameter X length:
CST-10:1.7 mm(5Fr) X
1900 mm
Catheter attached to Electrode
tip (Diameter X Length):
OCT1906: 2.0 mm X 1850mm
OCT1907: 2.4 mm X 1850mm
OCT1908: 2.8 mm X 1850mmOuter catheter attached to
Electrode tip
Diameter X length:
CST-10: 3.3mm (10Fr) X
1650 mm
Energy Delivered1. Energy: Monopolar
  1. Maximum rated input:
    1kV peak-to-peak
  2. Recommended generator
    settings:
    Pure cut mode, 80-120 Watts
    (400-500Vp) | 1. Energy: Monopolar
  3. Maximum rated input:
    2kV peak-to-peak
  4. Recommended generator
    settings:
    Pure cut mode, 80-120 Watts
    (400-500Vp) |
    | | Proposed Device | Predicate Device |
    | Compatible
    devices/equipment | 0.025 ~ 0.035" Guide wire
    Endoscope with 3.2mm
    diameter or larger working
    channel | 0.035" Guide wire
    Endoscope with 3.7mm
    diameter or larger working
    channel |
    | Materials | Electrode: SUS 304
    Sheath: PTFE | Unknown |
    | Single Use | Yes | Yes |
    | Sterile | EO Sterilization | EO Sterilization |

Table 1: Technological Comparison

5

6

7. PERFORMANCE DATA

Biocompatibility Testing

Biocompatibility Characterization Table 2:

CategorizationOptimos™ Cystotome
Contact partDirect contact to patientSheath (PTFE), Electrode Tip (STS 304)
Indirect contact to patientSaline port of Connector (PP), Sheath Tip (STS 304),
Shaft wire (STS 304)
Categorization
by ISO 10993-1Nature of body contactExternal communicating device
Contact with Tissue
Duration of contactLimited exposure (A) – medical devices whose
cumulative sum of single, multiple or repeated
duration of contact is
up to 24 h.
Sample preparationMedical Device in Final Finished FormTable

Table 3: Biocompatibility Test Results

Test ItemsResultP/F
CytotoxicityReactivity grade
  • test sample: 0
  • reagent control: 0
  • negative control: 0
  • positive control: 4 | P |

7

Test ItemsResultP/F
Intracutaneous reactivity1. No abnormal signs
  1. Normal body weight increase
  2. Mean score was 0.00 in sterile saline extracts
    and
    0.00 in cottonseed oil extracts | P |
    | Sensitization
    (GPMT) | 1. No abnormal signs
  3. Normal body weight increase
  4. The sensitization score was observed 0.0 | P |
    | Acute systemic toxicity | 1. No abnormal signs
  5. Normal body weight increase
  6. No abnormal signs | P |
    | Material mediated
    pyrogenicity test | 1. No abnormal signs
  7. Normal body weight increase
  8. No animal showed body temperature increase
    of 0.5 °C or above | P |

Electrical safety and electromagnetic compatibility (EMC)

Table 4:Electrical Safety and Electromagnetic Compatibility Summary Results
-----------------------------------------------------------------------------------
StandardResults
EN 60601-1 (2006) + A1:2013GeneralP
IEC 60601-1:2005/AMD1:2012Assessment of Basic Safety and Essential Performance of Medical DevicesP
IEC 60601-1-2Collateral standard: Electromagnetic compatibilityP
IC 60601-2-2:2017Particular standards: HF surgical equipmentP
IEC 60601-2-18Particular equipment: endoscopic equipmentP

Bench Testing

Unit tests of the Optimos™ Cystotome included a visual test, dimensional test, continuity test, endoscopic compatibility test, device operability test, high frequency dielectric strength test, detachable connector retention test, injection & leakage test, and tensile strength test. All tests passed the set acceptance criteria.

8

Page 6 of 7

System tests to examine the thermal effects on tissue was conducted with fine needle aspiration, guidewire, endoscope, ESU as a system. The test results showed that all test samples meet the required performance, puncture tissues, and are compatible with all devices/equipment working together.

Thermal effects were evaluated on both the subject device and the equivalent device to validate the performance of the device. The performance of the Optimos™ Cystotome was comparable to the Wilson-Cook CST-10 Cystotome Cystoenterostomy Needle Knife at 18, 100, and 120W on porcine stomach tissue, and duodenum in the mucosa, submucosa, and muscularis propria. The test results met the acceptance criteria.

| Test
sample | Device name | Optimos™ Cystotome | Cystotome™
Cystoenterostomy Needle
Knife |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|------------------------------------------------|
| | Manufacturer | Taewoong Medical Co., Ltd | Wilson-Cook |
| | Model
number | OCT1906 | CST-10 |
| | Sample size | 18 | 18 |
| Acceptance criteria | 1. The ability to puncture 2 type tissues (Porcine stomach and
Duodenum) | | |
| | 2. Substantial equivalence of thermal damage: There should be
no significant | | |
| | difference in measurement of size (length, width, and depth) of
the thermal | | |
| | damage zone between subject device and predicate device. | | |
| Generator power setting | Mono-polar Type, Auto Cut Mode (400 – 500Vp) | | |
| | Minimum: 80W, Default: 100W, Maximum: 120W | | |
| Damage zone
measurement | The thermal damage zone is measured under microscope and the
depth is analyzed using histological strain, NADH
((nicotinamide adenine dinucleotide). | | |
| Temperature-time
history | Using Digital Thermometer, the temperature of the tissue surface
was measured during the test. When the generator was activated,
69.7/60.6°C and 2.5/66.2°C were the highest temperature of
stomach and duodenum tissue applied by the subject
device/predicate device, respectively. The damage zone was
measured after confirmation that the tissue samples were cool
back to 37°C, baseline temperature. | | |

Table 5:Test Summary
------------------------

Animal Testing

Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.

9

Clinical Data

Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device.

8. CONCLUSION

Performance testing was conducted on all key performance attributes of the device. All samples met their acceptance criteria, demonstrating that when manufactured to specification, the device functions as intended and can be found substantially equivalent to the predicate device.