K Number
K223170
Date Cleared
2023-12-05

(420 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This product is intended to measure systolic and diastolic blood pressure and pulse on upper arm of population over 12 years old in household or medical facilities. (Not suitable for neonate, pregnancy or pre-eclampsia)

Device Description

Electronic blood pressure monitor, model: HTD6602US is a Noninvasive Blood Pressure Measurement System that is intended to measuring blood pressure through oscillation mensuration. The proposed device will automatically start to take measurements after the inflation of the cuff is finished, the results will show the systolic pressure and diastolic pressure with pulse rate. The blood pressure monitor will store the measurements automatically; The record maybe revisited. It measures blood pressure and pulse rate through inflating cuff which rounding the upper arm of patients. The HTD6602US is small, portable and used in home or medical facilities environment. The Electronic blood pressure monitor consists of two parts: main unit and cuffs. The HTD6602US is composed of PCBA, crystal oscillator, pump, valve, enclose, and LCD. Cuffs including cuff of size 22cm32cm and cuff of size 22cm42cm.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the Electronic Blood Pressure Monitor, model: HTD6602US, based on the provided document:

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Accuracy (Blood Pressure)
±3 mmHg (±0.4kPa)±3 mmHg (±0.4kPa)
Accuracy (Pulse Rate)
±5% of reading value±5% of reading value

Note: The primary acceptance criteria for accuracy are based on compliance with ISO 81060-2 Third Edition 2018-11, as indicated in the "Performance" row of the comparison table and the discussion of clinical accuracy testing. The specific numerical accuracy values are explicitly stated in the comparison table.

Study Details

The study that proves the device meets the acceptance criteria is a clinical accuracy test.

2. Sample Size and Data Provenance

  • Sample Size for Test Set: 85 subjects
  • Data Provenance: Not explicitly stated, but since the submission is by Hetaida Technology Co., Ltd. from China and it's a 510(k) submission to the US FDA, it likely refers to a prospective clinical trial conducted to meet ISO standards. The document doesn't specify if the subjects were from a particular country.

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: Not explicitly stated. The document refers to "reference equipment" (Sphygmomanometer - Stethoscope health box) and "Measuring Method: Auscultatory," which implies measurements were taken by trained personnel (e.g., medical professionals, technicians) using a standard, validated auscultatory method simultaneously with the device being tested.
  • Qualifications of Experts: Not explicitly stated. However, for auscultatory measurements to establish ground truth in blood pressure studies, the personnel typically need to be trained clinicians or technicians adhering to standardized measurement protocols.

4. Adjudication Method for Test Set

  • Adjudication Method: Not explicitly stated, but the method used for "ground truth" is "Auscultatory" measurements. In clinical blood pressure validation studies following ISO 81060-2, this usually involves at least two trained observers taking simultaneous readings (often blinded to each other's readings) and resolving discrepancies. The document mentions "Reference equipment used for measurements," implying a direct comparison to established, validated manual methods.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? No, this type of study is not applicable to a non-invasive blood pressure monitor that directly measures physiological parameters. MRMC studies are typically used for diagnostic imaging devices where human readers interpret images.

6. Standalone Performance Study

  • Was a standalone (algorithm only without human-in-the-loop performance) study done? Yes, the entire clinical accuracy study evaluates the device's (algorithm's) performance in measuring blood pressure and pulse rate autonomously against a gold standard (auscultatory measurements). The device operates without human interpretation of its internal measurements; it directly provides readings.

7. Type of Ground Truth Used

  • Type of Ground Truth: Expert consensus via Auscultatory Measurement Method using a reference Sphygmomanometer - Stethoscope health box. This is the gold standard for clinical validation of automated blood pressure devices.

8. Sample Size for Training Set

  • Sample Size for Training Set: Not applicable or not provided. This device is a hardware-based measurement system with an embedded algorithm, not a machine learning model that typically has a distinct training phase with labeled data in the same way. The algorithm's development (training) phase would occur during the device's engineering and calibration, not necessarily with a separate "training set" as understood in modern AI/ML contexts.

9. How Ground Truth for Training Set was Established

  • How Ground Truth for Training Set was Established: Not applicable. As noted above, this isn't a typical AI/ML model where a training set with established ground truth is used for iterative learning. The accuracy of the device's algorithm would be refined through engineering and calibration processes based on physiological principles and potentially internal testing, not a "training set" of patient data in the same sense. The ISO 81060-2 clinical trial serves as the validation of the finalized algorithm's performance against clinical ground truth.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 5, 2023

HeTaiDa Technology Co., Ltd. % You Yijie Manager Qimmiq Medical Consulting Service Co., Ltd. RM.406, Building C, Run Science Park, No.18 Shenzhou Road, Huangpu, Guangzhou, Guangdong 510663 China

Re: K223170

Trade/Device Name: Electronic Blood Pressure Monitor, model: HTD6602US Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: November 14, 2023 Received: November 14, 2023

Dear You Yijie:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

3

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K223170

Device Name

Electronic Blood Pressure Monitor, model: HTD6602US

Indications for Use (Describe)

This product is intended to measure systolic and diastolic blood pressure and pulse on upper arm of population over 12 years old in household or medical facilities. (Not suitable for neonate, pregnancy or pre-eclampsia)

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Section 5: 510(k) Summary

510(k) Number: K223170

1. Submitter's Information

Establishment Registration Information

Name: Hetaida Technology Co., Ltd.

Address: Room 801,802,803,804,901, 2# Building Scientific Research Center, Songhu Intelligent Valley, No.6 Minfu Road, Liaobu Town, Dongquan City, Guangdong Province Dongguan Guangdong, CN 523423.

Establishment registration number: 3017493034

Contact Person of applicant:

Name: Tom Chen

Address: Room 801,802,803,804,901, 2# Building Scientific Research Center, Songhu Intelligent Valley, No.6 Minfu Road, Liaobu Town, Dongguan City, Guangdong Province Dongguan Guangdong, CN 523423

TEL: +86-769-87036256

Email: tomchen@hetaida.com.cn

Correspondent (510(k) preparer) Information:

Name: You Yijie

Address: RM.406, Building C, Run Science Park, No.18 Shenzhou Road, Huangpu, Guangzhou, Guangdong 510663 P.R.China.

TEL: (+86)020-82245821

FAX: (+86)020-82245821

Email: jet.you@gimmiq-med.com

Date prepared: 11/15/2022

2. Device Information

Device Common Name: Noninvasive Blood Pressure Measurement System

Trade Name: Electronic blood pressure monitor

Model: HTD6602US

Classification name: Noninvasive Blood Pressure Measurement System

Review panel: Cardiovascular

Product code: DXN

{5}------------------------------------------------

Regulation Class: II Requlation Number: 870.1130

3. Predicate Device Information

510(k) submitter/holder:Jiangsu Yuyue Medical Equipment & Supply Co.,
Ltd.
510(K) Number:K200939
Trade Name:Electronic blood pressure monitor
Model:YE680B
Classification name:Noninvasive Blood Pressure Measurement
System
Review panel:Cardiovascular
Product code:DXN
Regulation Class:II
Regulation Number:870.1130

4. Device description

Electronic blood pressure monitor, model: HTD6602US is a Noninvasive Blood Pressure Measurement System that is intended to measuring blood pressure through oscillation mensuration. The proposed device will automatically start to take measurements after the inflation of the cuff is finished, the results will show the systolic pressure and diastolic pressure with pulse rate. The blood pressure monitor will store the measurements automatically; The record maybe revisited.

It measures blood pressure and pulse rate through inflating cuff which rounding the upper arm of patients. The HTD6602US is small, portable and used in home or medical facilities environment. The Electronic blood pressure monitor consists of two parts: main unit and cuffs. The HTD6602US is composed of PCBA, crystal oscillator, pump, valve, enclose, and LCD. Cuffs including cuff of size 22cm32cm and cuff of size 22cm42cm.

Principle of operation:

This product uses the Oscillometric Measuring method to detect blood pressure. When the cuff is fully inflated to reach a pressure above systolic pressure, no blood flow occurs through the artery. As the cuff is deflated below the systolic pressure, the reducing pressure exerted on the artery allows blood to flow through it and sets up a detectable vibration in the arterial wall. When the cuff pressure falls below the patient's diastolic pressure, blood flows smoothly through the artery in the usual pulses, without any vibration being set up in the wall. Vibrations occur at any point where the cuff pressure is sufficiently high that the blood has to push the arterial wall open in order to flow through the artery. The vibrations are transferred from the arterial wall, through the air inside the cuff, into a transducer in the monitor that converts the measurements into electrical signals. Hence when it starts inflating the arm cuff, meanwhile, the unit detects pressure oscillations

{6}------------------------------------------------

generated by beat-to-beat pulsatile, which is used to determine the systolic and diastolic pressure, and pulse rate.

5. Indications for Use

This product is intended to measure systolic and diastolic blood pressure and pulse on upper arm of population over 12 years old in household or medical facilities. (Not suitable for neonate, pregnancy or pre-eclampsia).

6. Summary of technological characteristics of device compared to the

SEComparisonsSubject device(Electronic blood pressure monitor,model: HTD6602US)Predicate device(Electronic blood pressure monitor,Model: YE680B)Discussion ofdifference
510K NumberK223170K200939/
ModelHTD6602USYE680B/
Classification21CFR 870.113021CFR 870.1130Same
Product CodeDXNDXNSame
FDA ClassIIIISame
Indications forUseThis product is intended to measuresystolic and diastolic blood pressureand pulse on upper arm.This product is intended to measuresystolic and diastolic blood pressure andpulse on upper armSame
PatientPopulationpopulation over 12 years oldAdultDifferent(Discussion isindicated in D1)
Environment ofUseHome or medical facilitiesHome or medical facilitiesSame
Designtable typetable typeSame
Design MethodOscillometricOscillometricSame
MeasurementSiteUpper ArmUpper ArmSame
CuffCircumference22cm32cm;22cm42cm.22cm32cm;2245cm.Different(Discussion isindicated in D2)
Inflation MethodInflation MethodInflation MethodSame
Deflation MethodAutomatic Pressure Release ValveAutomatic Pressure Release ValveSame
DisplayBacklight LCDDigital DisplayBacklight LCDDigital DisplaySame
Memory Size180 groupsUp to 99x2 sets of dataDifferent(Discussion isindicated in D3)
Blood PressureIndicationRangeDIA: 30 mmHg195mmHgSYS: 60 mmHg255mmHgDiastolic:20210 mmHgSystolic:40260 mmHgDifferent(Discussion isindicated in D4)
MeasurementPressure Range0299mmHg (039.9kPa)0 ~ 300 mmHg (0 kPa ~ 40 kPa)Same
RangeAccuracy±3 mmHg (±0.4kPa)±3 mmHg (±0.4kPa)Same
MeasurementPulse Range40~180 beats/min40 ~ 200 beats/minDifferent(Discussion isindicated in D5)
Pulse Accuracy±5% of reading value±5% of reading valueSame
PressurizationSourceAutomatic Internal PumpAutomatic Internal PumpSame
PressureSensorSemiconductor Pressure SensorSemiconductor Pressure SensorSame
OperatingEnvironmentAmbient temperature: +5°C ~ +40°CRelative humidity (RH): ≤85%Atmospheric pressure: 800 hPa~1050hPaTemperature: +5℃ ~ +40 ℃Humidity: 15% RH ~ 90% RH (nocondensation)Different(Discussion isindicated in D6)
StorageEnvironmentAmbient temperature: -20 °C ~+55 °CRelative humidity (RH): 10%93%Atmospheric pressure: 800 hPa1050hPaTemperature: -20 °C ~ +55 °CHumidity: 15% RH ~ 90% RH (nocondensation)Different(Discussion isindicated in D7)
Energy SourceInternal power supply DC 6V (4X1.5V(AA) Alkaline battery/ IEC Type LR06);External power supply: USB interfaceDC 5V 1A4 AA batteries or 6V/600mA AC adapterDifferent(Discussion isindicated in D8)
Display Content1, User Group2,Over pressure Alarm3, Low battery symbol4, Unit mmHg5, WHO (icon)6, Pulse Rate (Measuring mode)Memory group (Memory mode)7, Unit kPa8, Memory9, Average of last 3 measurements10, Pulse signal11, Body movement12, Cuff inflation abnormal13, Pulse Rate (Memory mode)14, Diastolic pressure15, Systolic pressureCuff Pressure,Pulse, Date,Time,Systolic/DiastolicPressure, errormessage,measurementsresults inmemory,Irregular HeartBeat Feature,Body movementdetection, CuffWrappingDetection, Dualuser switchingDifferent(Discussion isindicated in D9)
ControlsSet buttonStart/Stop buttonMemory buttonMemory Button,START/PULSE Button,Member ButtonDifferent(Discussion isindicated in D10)
PerformanceANSI/IAAMI/ISO81060-2:2013ANSI/IAAMI/ISO81060-2:2013Same
PerformanceIEC80601-2-30IEC80601-2-30Same
BiocompatibilityISO 10993-1, FDA Guidance, Testsincluded Cytotoxicity, Sensitization andIntracutaneous ReactivityISO 10993-1, FDA Guidance, Testsincluded Cytotoxicity, Sensitization andIntracutaneous ReactivitySame
ElectricalSafetyIEC60601-1IEC60601-1Same
EMCIEC60601-1-2IEC60601-1-2Same
UsabilityIEC 60601-1-6IEC 60601-1-6Same
Home UseIEC 60601-1-11IEC 60601-1-11Same
Material ofPatient contactcomponentsTop Cover: ABS757Bottom Cover: ABS757Battery door: ABS757Window cover: PMMAFemale connector: ABS757Start/Stop button: ABS757Set button: ABS757Memory button: ABS757Decoration ring(button): ABS757Cuff connector: ABS757Cuff: 1. cuff: Two layers of 420DPolyester;2.air bag: PVC.3.Velcro: nylon.4.Ring: metal.5.air tube: PVC.Not publicDifferent(Discussion isindicated in D11)
PatientInterfaceCuff, buttonCuff, buttonSame
Dimensions141mm×106mm×70mm(Length×Width×Height)127x93x74 (mm)Different(Discussion isindicated in D12)
WeightApproximate 323g. (Without battery)Approx.330gDifferent(Discussion isindicated in D13)
Principle ofoperationOscillometric Measuring methodOscillometric Measuring methodSame
User InterfaceCuff, buttonCuff, buttonSame
SoftwareEmbeddedEmbeddedSame

predicate devices (K200939)

{7}------------------------------------------------

{8}------------------------------------------------

The discussion of differences exist between the subject and predicate devices is listed in following:

  • D1: The difference of Patient Population between subject device and predicate device is that the intended age range of subject device is over 12 years old and predicate device is intended for adults, this difference is addressed through clinical trial conducted with subject device and contains population over 12 years old according to ISO 81060-2 Third Edition 2018-11, the results of clinical trial meet the requirements of ISO 81060-2 Third Edition 2018-11, therefore, the difference of subject device with predicate device YE680B (K200939) will not affect the safety and effectiveness.

{9}------------------------------------------------

  • D2: The difference of Cuff Circumference between subject device and predicate device is that the intended Cuff Circumference range of subject device is 2242cm and predicate device is 2245cm, the Cuff Circumference range of subject device is in the range of predicate device, and this difference is also addressed through clinical trial conducted with subject device according to ISO 81060-2 Third Edition 2018-11, the results of clinical trial meet the requirements of ISO 81060-2 Third Edition 2018-11, therefore, the difference of subject device with predicate device YE680B (K200939) will not affect the safety and effectiveness.
  • D3: The Memory Size of subject device is different with predicate device predicate device YE680B (K200939) will not affect the safety and effectiveness.
  • D4: The difference of Blood Pressure Indication Range between subject device and predicate device is that the Blood Pressure Indication Range of subject device is DIA: 30 mmHq195mmHq & SYS: 60 mmHg255mmHg and predicate device is Diastolic:20210 mmHg & Systolic:40260 mmHg, the Blood Pressure Indication Range of subject device is in the range of predicate device and meets the IEC 80601-2-30: Edition 2.0 2018-03, and this difference is also addressed through requirements of clinical trial conducted with subject device according to ISO 81060-2 Third Edition 2018-11, the results of clinical trial meet the requirements of ISO 81060-2 Third Edition 2018-11, therefore, the difference of subject device with predicate device YE680B (K200939) will not affect the safety and effectiveness.
  • D5: The Measurement Pulse Range of subject device is in the range of predicate device, and this difference is also addressed through clinical trial conducted with subject device according to ISO 81060-2 Third Edition 2018-11, therefore, the difference of subject device with predicate device YE680B (K200939) will not affect the safety and effectiveness.
  • D6: The operating Environment of subject device are different with predicate device YE680B (K20033), the difference introduces risks mitigated by testing in accordance with IEC 60601-1-11 and ANSI AAMI ES60601-1 provided in this submission, therefore the difference does not raise new questions of safety and effectiveness.
  • D7: The Storage Environment of subject device is different with predicate device YE680B (K200939), the difference introduces risks mitigated by testing in accordance with IEC 60601-1-11 and ANSI AAMI ES60601-1 provided in this submission, therefore the difference does not raise new questions of safety and effectiveness.
  • D8: The Energy Source of subject device are different with predicate devices, the difference introduces risks mitigated by the electromagnetic compatibility and electrical safety testing in accordance with IEC 60601-1-2 and ANSI AAMI ES60601-1 provided in this submission, therefore the difference does not raise new questions of safety and effectiveness.
  • D9: The Display Content between subject device and predicate device is that excluding the same parts, the Display Content of subject device additionally includes "User Group and Average of last 3 measurements" which are not included by predicate device, and the predicate device additionally includes "Cuff Pressure, Date, Time, Irreqular Heart, Beat Feature, Dual user switching" which are not included by predicate device, since the subject device meets the requirements of IEC 80601-2-30: Edition 2.0 2018-03 and ISO 81060-2 Third Edition 2018-11, therefore, the difference of subject device with predicate device YE680B (K200939) will not affect the safety and effectiveness.
  • D10: The Controls of subject device are different with predicate device, the difference does not raise new questions of safety and effectiveness.

{10}------------------------------------------------

  • D11: The Material of Patient contact components of subject device has been validated for Cytotoxicity though testing against ISO 10993- 5, for Sensitization though testing against ISO 10993-10 tested and all test results are positive, the difference of subject device with predicate device YE680B (K200939) do not raise new questions of safety and effectiveness.
  • D12: The Dimensions of subject device is different with predicate device predicate device YE680B (K200939) will not affect the safety and effectiveness.
  • D13: The Weight of subject device is different with predicate device YE680B (K200939) will not affect the safety and effectiveness.

7. Discussion of Non-Clinical Tests Performed are as follows

The recognized consensus standards for safety of medical electrical equipment: ANSI AAM ES 60601- 1: 2005 / (R) 2012 and A1: 2012, IEC 60601-1-11 Edition 2.0 2015-01. IEC 60601-1-8 Edition 2.1 2012-11 for safety, IEC 60601-1-2: 2014 for electromagnetic compatibility, IEC 80601-2-30: Edition 2.0 2018-03, ISO 81060-2 Third Edition 2018-11 for performance, IEC 62304 Edition 1.1 2015-06 for software verification, IEC 62366-1 Edition 1.0 2015-02, IEC 60601-1-6 Edition 3.1 usability and 2013-10 for for ട്രി 10993-10:2010 for Sensitization and Irritation endpoints are complied. See below table for details:

StandardsStandards Name
ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012Medical Electrical Equipment - Part 1: General Requirements ForBasic Safety And Essential Performance
IEC 60601-1-2: 2014Medical Electrical Equipment - Part 1-2: General Requirements ForBasic Safety And Essential Performance -- Collateral Standard:Electromagnetic Disturbances - Requirements And Tests
IEC 60601-1-11 Edition 2.02015-01Medical Electrical Equipment - Part 1-2: General Requirements ForBasic Safety And Essential Performance - Collateral Standard:Electromagnetic Disturbances - Requirements And Tests
IEC 60601-1-8 Edition 2.12012-11Medical electrical equipment - Part 1-8: General requirements forbasic safety and essential performance - Collateral Standard:General requirements, tests and guidance for alarm systems inmedical electrical equipment and medical electrical systems
IEC 80601-2-30: Edition 2.02018-03Medical electrical equipment - Part 2-30: Particular requirements forthe basic safety and essential performance of automated non-invasive sphygmomanometers
ISO 81060-2 Third Edition2018-11Non-invasive sphygmomanometers - Part 2: Clinical investigation ofintermittent automated measurement type
IEC 62366-1 Edition 1.02015-02Medical devices - Part 1: Application of usability engineering tomedical devices
IEC 60601-1-6 Edition 3.12013-10Medical electrical equipment - Part 1-6: General requirements forbasic safety and essential performance - Collateral standard:Usability

{11}------------------------------------------------

ISO 10993-5:2009Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
IEC 62304 Edition 1.1 2015-06Medical device software - Software life cycle processes

● Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the subject device HTD6602US. The system complies with the AAMI ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11, IEC 80601-2-30: Edition 2.0 2018-03, ISO 81060-2 Third Edition 2018-11, standards for electrical safety and the IEC 60601-1-2 standard for EMC.

● Software Verification and Validation Testing

Software verification and validation was performed for the subject device in accordance with Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff, May 2005.

Software Description:

The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software could result in Minor Injury, either to a patient or to a user of the device. The software of the system, on the whole, is accountable for the system scheduler of the device, including Pressure and pulse signal acquisition, calculation and display, the button presses of end user, display the result of the measurement data storage and average calculation, memory data query, prompt of error message, low battery voltage detection and prompt, measurement Unit conversion, user Account Switching.

  • V Programming language C Programming Language.
    Hardware Requirements A Microprocessors: HME062

8. Discussion of Clinical Accuracy Testing Performed

The clinical accuracy test report and data analysis followed the requirements of the ISO 81060-2 Third Edition 2018-11.

Subjects requirementNumber specified in ISO81060-2 Third edition 2018-11 Amendment 1 2020-01Actual number
TotalA minimum of 8585
MaleAt least 2638
FemaleAt least 2647
Age > 12100%100%

The clinical accuracy testing evaluated 85 of subjects, division of all subjects:

{12}------------------------------------------------

arm circumference in the range of 22-27cm;≥20%31.76%
arm circumference in the range of 27-32cm;≥20%23.53%
arm circumference in the range of 32-37cm;≥20%23.53%
arm circumference in the range of 37-42cm;≥20%21.18%
arm circumference in the range of 22-24.5cm;≥10%21.18%
arm circumference in the range of 39.5-42cm≥10%11.76%
subjects shall be tested through the cuffwith specified (22-32cm)At least 25%32.94%
subjects shall be tested through the cuffwith specified (22-42cm)At least 50%67.06%
Systolic BP ≤ 100mmHg≥5%17.65%
Systolic BP ≥ 160mmHg≥5%12.94%
Systolic BP ≥140mmHg≥20%30.59%
Diastolic BP ≤ 60mmHg≥5%14.12%
Diastolic BP ≥ 100mmHg≥5%10.59%
Diastolic BP ≥ 85mmHg≥20%27.06%

The test data showed the clinical accuracy of the subject device complied with the requirements of ISO 81060-2 Third Edition 2018-11.

NameSphygmomanometer - Stethoscope health box
ModelA Type
ManufacturerJiangsu Yuyue Medical Equipment & SupplyCo., Ltd.
Measuring MethodAuscultatory

Reference equipment used for measurements:

9. Conclusions

The Electronic blood pressure monitor, model: HTD6602US, have the same intended use and similar characteristics as the cleared predicate device Electronic Blood Pressure Monitor, Model: YE680B. Moreover, bench testing contained in this submission supplied demonstrate that the differences existed between HTD6602US and YE680B do not raise any new questions of safety or effectiveness.

The non-clinical tests support the safety of the device and the hardware and software verification and validation demonstrate that the Electronic blood pressure monitor, model: HTD6602US performs as intended in the specified use conditions are same with predicate device. The clinical performance tests demonstrate that the Electronic blood pressure monitor, model: HTD6602US performs comparably to the predicate device that is currently marketed for the same intended use. Thus, the Electronic blood pressure monitor, model: HTD6602US is Substantially Equivalent (SE) to the predicate device.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).