K Number
K223162
Date Cleared
2023-03-17

(161 days)

Product Code
Regulation Number
862.3870
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Healgen® Accurate Oral Fluid Drug Test COT is a lateral flow chromatographic immunoassay for the qualitative detection of COT in oral fluid at the cut-off concentration 30 ng/mL.

This assay provides only a preliminary result. An alternative laboratory test must be used to confirm the results provided by this drug test. Gas chromatography/mass spectrometry (GC/MS) is the preferred method confirmation test.

The Healgen® Accurate Oral Fluid Drug Test is a competitive binding lateral flow immunochromatographic assay for the qualitative and simultaneous detection of Marijuana (THC) and Cotinine in human oral fluid at the cutoff concentrations listed below and their metabolites.

TestCalibratorCut-off (ng/mL)
Cotinine (COT)(-) Cotinine30
Marijuana (THC)Delta-9-Tetrahydrocannabinol40

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography mass spectrometry (GC/MS) and liquid chromatography mass spectrometry (LC/MS) are the preferred confirmatory methods.

Device Description

The Healgen® Accurate Oral Fluid Drug Test COT is immunochromatographic assay that uses a lateral flow system for the qualitative detection of cotinine in human oral fluid. The Healeen® Accurate Oral Fluid Drug Test immunochromatographic assay for the qualitative and simultaneous detection of Marijuana (THC) and Cotinine in human oral fluid. The tests are the first step in a two-step process. The second step is to send the sample for laboratory testing if preliminary positive results are obtained.

AI/ML Overview

This document describes the performance characteristics and studies for the Healgen® Accurate Oral Fluid Drug Test and Healgen® Accurate Oral Fluid Drug Test COT, which are qualitative immunoassays for detecting Cotinine (COT) and Marijuana (THC) in oral fluid.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as numerical targets in the document. However, the "Precision-Reproducibility-Cut-Off" studies implicitly define the expected performance around the cut-off values. The "Comparison Studies" and "Layuser Studies" then demonstrate the agreement of the device with a gold standard (LC/MS/MS).

Performance MetricGeneral Acceptance Goal (Implicit)Reported Device Performance (COT)Reported Device Performance (THC)
Precision-Reproducibility-Cut-Off (Analytical)High accuracy, especially near cut-off; consistent performance across lots.COT: -100% to -50% cut off: 50-/0+ (100% negative) -25% cut off: Lot 1: 50-/0+, Lot 2: 48-/2+, Lot 3: 49-/1+ (High negative agreement) Cut off: Lot 1: 24-/26+, Lot 2: 23-/27+, Lot 3: 24-/26+ (Balanced +/- at cut-off, as expected) +25% to +100% cut off: Near 100% positiveTHC: -100% to -50% cut off: 50-/0+ (100% negative) -25% cut off: Lot 1: 47-/3+, Lot 2: 48-/2+, Lot 3: 49-/1+ (High negative agreement) Cut off: Lot 1: 24-/26+, Lot 2: 28-/22+, Lot 3: 24-/26+ (Balanced +/- at cut-off, as expected) +25% to +100% cut off: Near 100% positive
Method Comparison Study (Expert Operator)High percentage of correct results across all concentration ranges, especially for Drug-Free and High Positive samples.COT:Drug-Free: 100% negative (112/112)Less than Half Cutoff: 100% negative (93/93)Near Cutoff Negative: 88% correct (35/40)Near Cutoff Positive: 88% correct (28/32)High Positive: 100% positive (150/150)THC:Drug-Free: 100% negative (35/35)Less than Half Cutoff: 100% negative (13/13)Near Cutoff Negative: 71.4% correct (5/7)Near Cutoff Positive: 60% correct (6/10)High Positive: 100% positive (61/61)
Layuser Study (OTC Use)High percentage of correct results (similar to expert), ease of use.COT (Layuser):Drug-Free: 100% negative (95/95)Less than Half Cutoff: 100% negative (39/39)Near Cutoff Negative: 89% correct (31/35)Near Cutoff Positive: 87% correct (34/39)High Positive: 100% positive (154/154)All layusers found instructions easy to understand and follow (Reading Grade Level 7).Not applicable (Layuser study only reported for COT, which is for OTC use as per the Indications for Use).

2. Sample Sizes Used for the Test Set and Data Provenance

  • Precision-Reproducibility-Cut-Off (Analytical Performance):

    • For each of the 9 concentration levels and 3 device lots: 50 samples (2 runs per day for 25 days).
    • Total samples = 9 concentrations * 50 samples/concentration * 3 lots = 1350 samples for COT, and 1350 samples for THC.
    • Data Provenance: Samples were "spiked" with cotinine or marijuana in negative oral fluid samples, indicating a controlled laboratory setting. The origin (country/retrospective/prospective) is not specified, but the nature of spiking implies prospective sample preparation for the study.
  • Method Comparison Studies (Expert Operator):

    • COT: 427 samples
    • THC: 126 samples
    • Data Provenance: Oral fluid samples. The data provenance (e.g., country of origin, retrospective or prospective collection) is not specified.
  • Layuser Studies:

    • COT: 362 samples
    • Data Provenance: Oral fluid samples. The data provenance is not specified.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Ground Truth Establishment: The ground truth for all performance studies (Precision, Method Comparison, Layuser) was established by LC/MS/MS (Liquid Chromatography/Mass Spectrometry/Mass Spectrometry), which is stated as the "preferred method confirmation test."
  • Qualifications of Experts: This implies that the ground truth was not established by human experts interpreting images or subjective data, but by highly accurate analytical laboratory equipment and the qualified personnel operating it. The document does not specify the number or qualifications of the lab personnel who performed the LC/MS/MS analysis, as GC/MS and LC/MS/MS are considered an objective gold standard in drug testing.

4. Adjudication Method for the Test Set

  • LC/MS/MS serves as the objective ground truth, providing quantitative results against which the device's qualitative results are compared. Therefore, an "adjudication method" in the sense of reconciling disagreements between human readers is not applicable here. Any discrepancies between the device result and LC/MS/MS are reported as "discordant results."

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

  • This is not an AI-assisted diagnostic device, but a rapid diagnostic test (lateral flow immunoassay). Therefore, an MRMC comparative effectiveness study involving human readers with and without AI assistance is not applicable.
  • The "Layuser Studies" evaluate the performance of the device in the hands of untrained users, essentially assessing human performance with the device instructions. It shows that layusers can achieve comparable results to expert operators for COT detection.

6. If a Standalone (i.e., algorithm-only without human-in-the-loop performance) Was Done

  • This question is not applicable as the device is a lateral flow immunoassay kit, not an algorithm. Its "standalone" performance is inherently demonstrated by the results it produces directly from the oral fluid sample, without further human-in-the-loop interpretation beyond reading the visual lines on the strip. The "Precision-Reproducibility-Cut-Off" and "Method Comparison Studies" characterize this inherent device performance.

7. The Type of Ground Truth Used

  • The ground truth used was analytical confirmation by LC/MS/MS. This is an objective, quantitative laboratory method considered the gold standard for drug detection and quantification in biological samples.

8. The Sample Size for the Training Set

  • The document describes performance studies, but it does not mention a "training set" in the context of machine learning. This is a traditional immunoassay device, which does not employ machine learning algorithms that require separate training and test sets. The studies described are for analytical and clinical validation of the device's performance.

9. How the Ground Truth for the Training Set Was Established

  • As there is no mention of a "training set" for a machine learning model, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 17, 2023

Healgen Scientific LLC % Joe Shia Director LSI International Inc. 504 East Diamond Ave. Suite I Gaithersburg, MD 20877

Re: K223162

Trade/Device Name: Healgen® Accurate Oral Fluid Drug Test, Healgen® Accurate Oral Fluid Drug Test COT Regulation Number: 21 CFR 862.3870 Regulation Name: Cannabinoid Test System Regulatory Class: Class II Product Code: LDJ, MKU Dated: February 12, 2023 Received: February 13, 2023

Dear Joe Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Paula
Caposino -S

Digitally signed by Paula
Caposino -S
Date: 2023.03.17 13:12:26
-04'00'

Paula Caposino, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K223162

Device Name Healgen® Accurate Oral Fluid Drug Test COT

Indications for Use (Describe)

The Healgen® Accurate Oral Fluid Drug Test COT is a lateral flow chromatographic immunoassay for the qualitative detection of COT in oral fluid at the cut-off concentration 30 ng/mL.

This assay provides only a preliminary result. An alternative laboratory test must be used to confirm the results provided by this drug test. Gas chromatography/mass spectrometry (GC/MS) is the preferred method confirmation test.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known)

K223162

Device Name

Healgen® Accurate Oral Fluid Drug Test

Indications for Use (Describe)

The Healgen® Accurate Oral Fluid Drug Test is a competitive binding lateral flow immunochromatographic assay for the qualitative and simultaneous detection of Marijuana (THC) and Cotinine in human oral fluid at the cutoff concentrations listed below and their metabolites.

TestCalibratorCut-off (ng/mL)
Cotinine (COT)(-) Cotinine30
Marijuana (THC)Delta-9-Tetrahydrocannabinol40

This assay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas chromatography mass spectrometry (GC/MS) and liquid chromatography mass spectrometry (LC/MS) are the preferred confirmatory methods.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{5}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{6}------------------------------------------------

510(k) SUMMARY K223162

    1. Date: March 16, 2023 2. Submitter: HEALGEN SCIENTIFIC LLC 5213 Maple Street Bellaire, TX 77401
  • Joe Shia 3. Contact person: LSI International Inc. 504 East Diamond Ave., Suite I Gaithersburg, MD 20877 Telephone: 240-505-7880 Fax: 301-916-6213 Email:shiajl(@yahoo.com
  • Healgen® Accurate Oral Fluid Drug Test 4. Device Names: Healgen® Accurate Oral Fluid Drug Test COT

Classification:

Product CodeCFR #Panel
LDJ21 CFR, 862.3870 Cannabinoids Test SystemToxicology
MKU21 CFR, 862.3220 Cotinine Test SystemToxicology
    1. Predicate Devices:
      LZI Oral Fluid Cannabinoids Enzyme Immunoassay, LZI Oral Fluid Cannabinoids Calibrators and LZI Oral Fluid Cannabinoids Controls, K141320 LabOne Micro-Plate Cotinine EIA (Oral Fluid), K033601
    1. Intended Use
      The Healgen® Accurate Oral Fluid Drug Test COT is a lateral flow chromatographic immunoassay for the qualitative detection of COT in oral fluid at the cut-off concentration 30 ng/mL.

This assay provides only a preliminary result. An alternative laboratory test must be used to confirm the results provided by this drug test. Gas chromatography/mass spectrometry (GC/MS) is the preferred method confirmation test.

The Healgen® Accurate Oral Fluid Drug Test is a competitive binding lateral flow immunochromatographic assay for the qualitative and simultaneous detection of Marijuana (THC) and Cotinine in human oral fluid at the cutoff concentrations listed below and their metabolites.

{7}------------------------------------------------

TestCalibratorCut-off (ng/mL)
Cotinine (COT)(-) Cotinine30
Marijuana (THC)Delta-9-Tetrahydrocannabinol40

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography mass spectrometry (GC/MS) and liquid chromatography mass spectrometry (LC/MS) are the preferred confirmatory methods.

7. Device Description

The Healgen® Accurate Oral Fluid Drug Test COT is immunochromatographic assay that uses a lateral flow system for the qualitative detection of cotinine in human oral fluid. The Healeen® Accurate Oral Fluid Drug Test immunochromatographic assay for the qualitative and simultaneous detection of Marijuana (THC) and Cotinine in human oral fluid. The tests are the first step in a two-step process. The second step is to send the sample for laboratory testing if preliminary positive results are obtained.

    1. Substantial Equivalence Information
      A summary comparison of features of the Healgen® Accurate Oral Fluid Drug Test COT and Healgen® Accurate Oral Fluid Drug Test and the predicate devices are provided in following tables.
ItemDevicePredicate - K033601
Indication(s)for UseFor the qualitative determination of cotinine inhuman oral fluid.For the qualitative and semi-quantitative analysis ofcotinine in human oral fluid
CalibratorsCotinineSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assaysA solid phase competitiveenzyme-linked immunoassay
Type of TestQualitativequalitative and semi-quantitative
Specimen TypeHuman Oral fluidSame
Intended UseFor Over-The-Counter usesFor prescription uses

Table 1: Features Comparison of Healgen® Accurate Oral Fluid Drug Test COT and the Predicate Device

{8}------------------------------------------------

Table 2: Features Comparison of Healgen® Accurate Oral Fluid Drug Test and the Predicate Device

ItemDevicePredicate - K141320
Indication(s)for UseSamePreliminary Drug screening test for thequalitative detection of drug analytes inoral fluid (human saliva)For In Vitro Diagnostic Use
MethodologySameCompetitive binding,lateral flowimmunochromatographicassays
Type of TestSameQualitative
Specimen TypeSameHuman Oral Fluid

9. Test Principle

The Healgen® oral fluid drug test device is an immunoassay based on the principle of competitive binding. Drugs that may be present in the oral fluid specimen compete against their respective drug conjugates for binding sites on their specific antibody.

During testing, a portion of the oral fluid specimen migrates upward by capillary action. A drug, if present in the oral fluid specimen below its cut-off concentration, will not saturate the binding sites of its specific antibody. The antibody will then react with the drug-protein conjugate and a visible colored line will show up in the test line region of the specific drug strip. The presence of drug above the cut-off concentration in the oral fluid specimen will saturate all the binding sites of the antibody. Therefore, the colored line will not form in the test line region.

A drug-positive oral fluid specimen will not generate a colored line in the specific test line region of the strip because of drug competition, while a drug-negative oral fluid specimen will generate a line in the test line region because of the absence of drug competition.

To serve as a procedural control, a colored line will always appear at the control line region, indicating that proper volume of specimen has been added and membrane wicking has occurred.

10. Performance Characteristics

    1. Analytical Performance
    • Precision-Reproducibility-Cut-Off a.

Precision-Reproducibility-Cut-Off studies were carried out for samples with concentrations of -100% cut off, -75% cut off, -50% cut off, -25% cut off, cut off, +25% cut off, +50% cut off, +75% cut off and +100% cut off. These samples were prepared by spiking cotinine or marijuana in negative oral fluid samples. Each cotinine or marijuana concentration was confirmed by LC/MS/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days per device lot in a randomized order. The following are summaries.

{9}------------------------------------------------

Results-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Device Lot
LC-MS (ng/mL)07.415.422.630.737.945.952.960.6
Lot 150-/0+50-/0+50-/0+50-/0+24-/26+48+/2-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+48-/2+23-/27+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+49-/1+24-/26+49+/1-50+/0-50+/0-50+/0-
THC
Result-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Device LotLC-MS (ng/mL)0.009.520.730.240.851.459.268.381.5
Lot 1:50-/0+50-/0+50-/0+47-/3+24-/26+49+/1-50+/0-50+/0-50+/0-
Lot 2:50-/0+50-/0+50-/0+48-/2+28-/22+48+/2-50+/0-50+/0-50+/0-
Lot 3:50-/0+50-/0+50-/0+49-/1+24-/26+47+/3-50+/0-50+/0-50+/0-

The following cut-off values for the candidate device has been verified.

CalibratorCut-off (ng/mL)
cotinine30
marijuana40

b. Linearity

Not applicable.

c. Stability

The devices are stable at 4-30 ℃ for 24 months based on the real time stability study at 4°C and 30°C.

d. Interference

Potential interfering substances were added to drug-free oral fluid and target drugs oral fluid with concentrations at 50% below and 50% above Cut-Off levels. These oral fluid samples were tested using three batches of the Healgen device. Compounds that showed no interference for cotinine and THC at a concentration of 100μg/mL are summarized in the following tables.

{10}------------------------------------------------

AcetaminophenDiacetylmorphineNaltrexone
(IR,2S)-(-)Ephedrine HydrochlorideDiazepamNaproxen
AcetylcodeineDigoxinNiacinamide
Acetylsalicylic AcidDihydrocodeineNicotinamide
AllobarbitalDiltiazem HClNiordiazepan
AlprazolamDiphenhydramine HClNoscapine
AmobarbitalDL-MethadoneOmeprazole
AmpicillinDL-PropranololPapaverine
ApomorphineDoxylaminePenicillin
ascorbic acidEcgonine HydrochloridePentazocine
AtenololEcgonine methylesterPentobarbital
AtropineEstradiolPhencyclidine (PCP)
BaclofenEstronePhenobarbital
BenzocaineFenoprofenPhentermine
BenzoylecgonineFluconazolePhenylephrine
beta-PhenethylamineFurosemidePhenylpropanolamine
bilirubinGemfibrozilPhenytoin
ButabarbitalGentisic AcidPioglitazone HCl
ButalbitalHeroin hydrochloridePrednisolone
CaffeineHexobarbitalPrednisone
CarbamazepineHydrochlorothiazideProcainamide HCl
ChlordiazepoxideHydrocodoneProcaine HCL
ChlorpromazineHydromorphonePromethazine
CimetidineIbuprofenPseudoephedrine
Citalopram HBrImipramineQuinidine
ClobazamL-EphedrineQuinine HCl
ClomipramineL-MethamphetamineR,R(-)-Pseudoephedrine
ClonazepamL-ThyroxineSalicylic Acid
ClonidineLamotrigineSertraline HCL
Clopidogrel bisulfateLevetiracetamSimvastin
ClorazepateLidocaineTemazepam
CocaethyleneLormetazepamTheophylline
CocaineMeperidineTheophylline
CodeineMetformin HClThiamine
CortisolMethylphenidate HClTopiramate
D-AmphetamineMetoprololValproic Acid
D-MethamphetamineMetronidazoleVerapamil
d,l-SalbutamolMontelukast sodium saltZomepirac
DeoxycorticosteroneMorphine SulfateZonisamide
Delta-9-THC (except for THC test)Nalorphine
DextromethorphanNaloxone

{11}------------------------------------------------

Food items such as methanol cough drops, cough syrup, cola, mouthwash, coffee, tea, milk, sugar, chewing gum, alcohol, baking soda, salt, cranberry juice, orange juice, food coloring (red, blue, green), toothpaste, tomatoes and MSG were added in either drug-free oral fluid or oral fluid containing the target drugs with concentrations of 50% below and 50% above cutoff levels to a concentration of 5%. None of the substances showed interference. Hemoglobin showed no interference at 10mg/dL.

e. Specificity

To test specificity, drug metabolites and other components that are likely to interfere in oral fluid samples were tested using three batches of the Healgen device. The following are summaries.

COT
(-) Cotinine(Cut-off=30 ng/mL)ResultPositive at (ng/mL)% Cross-Reactivity
(-) Cotinine30100%
S(-)-Nicotine30001%
Trans-3-hydroxycotinine10300%
(+)-anabasine1,350,0000.002%
(+/-)-nornicotine150,0000.02%
Niacinamide>100,000< 0.03%

тно

THCMarijuana(Cut-off=40 ng/mL)ResultPositive at (ng/mL)% Cross-Reactivity
△9-Tetrahydrocannabinol40100%
△8-Tetrahydrocannabinol8050%
11-nor-△9-THC-9 COOH41000%
11-hydroxy-△9-THC4589%
Cannabinol20020%
Cannabidiol (CBD)2,2001.8%
11-Nor-△9-THC-carboxy-glucuronide6066.7%
(+)-11-nor-9-carboxy-△9-THC5080%
11-nor-A8-THC-9-COOH20200%
8-beta-11-dihydroxy-△9-THC20020%
8-beta-hydroxy-△9-THC20020%
Exo-THC7553.3%
1-11-Nor-△9-THC-9-CarboxylicAcyl-Glucuronide15266.7%
△8-THC Carboxylic Acid20200%
△9-THC Carboxylic Acid41000%

f. Effect of Oral fluid pH

To investigate the effect of oral fluid pH, oral fluid samples with pH 4 to 9 were spiked with target drugs at 50% below and 50% above Cut-Off levels. These samples were tested using three lots of the device. Results were all positive for samples at and above +50% Cut-Off and all negative for samples at and below -50% Cut-Off.

{12}------------------------------------------------

g. Drug Recovery Study

Negative oral fluid samples in glass bottles were spiked with target drugs to concentrations of -50% and +50% of the cutoffs. The samples were transferred to Healgen devices and store at room temperature, at -20°C and at 40°C. Over 90% recoveries were observed for both cotinine and marijuana in the Healgen devices. Oral fluid samples can be stored in the device at -20°C for at least 90 days. Oral fluid samples can be shipped overnight in the device for LC-MS confirmation.

2. Comparison Studies

Method comparison studies for the Healgen Device were performed at three testing sites for each target drug with one operator at each site. Operators tested total 427 samples for COT and 126 samples for THC. The obtained test results are compared to LC/MS/MS results. The results are presented in the tables below.

Concentration Range(by LC-MS/MS)NumberofsamplesTest ResultsThe percentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
Drug-Free1120112100%
Less than Half theCutoff Concentration93093100%
Near Cutoff Negative4053588%
Near Cutoff Positive3228488%
High Positive1501500100%

COT

Discordant Results

SitesSample NumberLC/MS ResultTest Results
Site 1CM-09823.43Positive
Site 1CM-00327.70Positive
Site 2LH-14024.62Positive
Site 2LH-16128.88Positive
Site 3LS-00729.32Positive
Site 1CM-09630.38Negative
Site 2LH-06932.64Negative
Site 3LS-08230.75Negative
Site 3LS-00231.84Negative

THC

Concentration Range(by LC-MS/MS)NumberofsamplesTest ResultsThe percentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
Drug-Free35035100
Less than Half theCutoff Concentration13013100

{13}------------------------------------------------

Near Cutoff Negative72571.4
Near Cutoff Positive106460
High Positive61610100

Discordant Results

SitesSample NumberLC/MS ResultTest Results
Site 2S066512338.3Positive
Site 2S066513938.7Positive
Site 3S058336740.6Negative
Site 3S058336342.2Negative
Site 3S058336846.0Negative
Site 3S058337147.0Negative

3. Layuser Studies

Layuser studies for the Healgen® Accurate Oral Fluid Drug Test COT were performed at three testing sites. Total 362 layusers with diverse educational and professional backgrounds performed tests based on the product insert. The layuser results are compared to LC/MS/MS results. The results are presented in the tables below.

Concentration Range(by LC-MS/MS)NumberofsamplesTest ResultsThe percentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
Drug-Free95095100%
Less than Half theCutoff Concentration39039100%
Near Cutoff Negative3543189%
Near Cutoff Positive3934587%
High Positive1541540100%

Discordant Results

SitesSample NumberLC/MS ResultTest Results
Site ANX-03227.97Positive
Site BPH-01223.54Positive
Site BPH-02829.77Positive
Site CAJ-09629.89Positive
Site ANX-01030.53Negative
Site ANX-01038.64Negative
Site BPH-08335.80Negative
Site CAJ-01930.26Negative
Site CAJ-01632.54Negative

Layusers were given surveys on the ease of understanding the instruction for use. All layusers indicated that the device instruction is easy to understand and follow. A Flesch-Kincaid reading analysis was performed on the package insert and the scores revealed a reading Grade Level of 7.

{14}------------------------------------------------

4. Clinical Studies

Not applicable.

11. Conclusion

Based on the test principle and acceptable performance characteristics including precision, interference, specificity, method comparison and layuser studies of the device, it's concluded that the Healgen® Accurate Oral Fluid Drug Test and Healgen® Accurate Oral Fluid Drug Test COT are substantially equivalent to the predicates.

§ 862.3870 Cannabinoid test system.

(a)
Identification. A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds includedelta -9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.(b)
Classification. Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).