K Number
K223037

Validate with FDA (Live)

Date Cleared
2023-07-21

(295 days)

Product Code
Regulation Number
880.2910
Age Range
22 - 120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Infrared Thermometer (Model: IR8807) is intended for the intermittent measurement of human body temperature from the forehead. It is a non-contact clinical thermometer and intended for use on people of all ages except for neonates/newborns.

Device Description

The Infrared Thermometer (Model: IR8807) is an electronic thermometer using an infrared sensor to measure infrared energy radiated from the forehead. This energy is collected through the lens and converted to a temperature value. The Infrared Thermometer (Model: IR8807) has a forehead measurement mode and an object measurement mode.

The Infrared Thermometer (Model: IR8807), consists of the following parts:
a) Thermopile Sensor
b) Application-Specific Integrated Circuitry
c) Erasable Programmable Read-Only Memory Integrated Circuit
d) Capacitance-touch Integrated Circuit
e) LCD display
f) 2 buttons (Scan button, Setting button,)
q) 2×1.5V AAA dry batteries

The operation principle is based on infrared sensor technology. The sensor in the thermometer detects the infrared energy, and the device processes the data to body temperature result with its internal microcontroller unit. It operates in an adjusted mode.

The main functions of Infrared Thermometer (Model: IR8807) are as followings:
Forehead temperature measure functions
Wide range of temperature readings: from 32.0ºCto 42.9ºC (89.6ºF ~ 109.2ºF)
The measurement results can be displayed in either °Cor °F units
The LCD display showing clear measurement result.
Memory data reading/removal function: The device can store a total of 32 groups of memorable values. All memory values can be removed after the battery is removed
Low battery indicator
Buzzer reminder function
Users can set temperature alarm value
Environment or object temperature measurement function
The reference body site of the output temperature is Oral.

AI/ML Overview

The provided document describes the Infrared Thermometer, Model: IR8807, and its compliance with relevant standards and performance criteria. Here's a breakdown of the requested information based on the text:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (from ASTM E 1965-98)Reported Device Performance (Model: IR8807)
Measurement Range32°C ~ 42.9°C (89.6°F ~ 109.2°F)
Accuracy (Forehead mode)±0.2°C for 35.0°C ~ 42.0°C
±0.3°C outside this temperature range
Measurement Time≤3 seconds
Measurement Distance2-5 cm

Note: The document states that the device's accuracy requirements meet ASTM E1965-98. The specific ASTM E1965-98 requirements for measurement range, accuracy, measurement time, and measurement distance are not explicitly detailed in the provided text beyond what's shown for the device's performance. The table above uses the device's reported performance as the "reported device performance" and implies these meet the ASTM E 1965-98 criteria because the document explicitly states compliance.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample size for clinical test set: 169 subjects.
  • Data Provenance: The document does not specify the country of origin. The test was a "clinical test" which implies a prospective study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the document. The document only mentions "clinical tests were conducted" and that the "clinical performance of the subject device complied with the requirement of ASTM E1965-98 (2016)." It does not describe the method or personnel used to establish the ground truth (reference temperature).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided in the document.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is an Infrared Thermometer. The document describes clinical tests for device accuracy, not a comparative effectiveness study involving human readers or AI assistance. Therefore, no MRMC comparative effectiveness study was done in the context of human readers improving with or without AI.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Yes, a standalone performance study was done for the device. The "clinical tests" described in section 8 were conducted on the Infrared Thermometer (Model: IR8807) itself, evaluating its direct measurement performance against a ground truth (implied by compliance with ASTM E1965-98 for clinical accuracy). This assesses the algorithm/device's performance without human interpretation or intervention in the measurement process beyond standard use.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The document does not explicitly state the type of ground truth used. However, given that it's a clinical thermometer study and states compliance with ASTM E1965-98, the ground truth would most likely be a reference clinical thermometer (e.g., a highly accurate rectal, oral, or tympanic thermometer, depending on the standard's protocol for correlating to forehead temperature) used simultaneously or sequentially with the test device.

8. The sample size for the training set

This device is a physical thermometer, not an AI/ML model that typically requires a "training set." The concept of a training set is not applicable to this device.

9. How the ground truth for the training set was established

As the concept of a training set is not applicable to this device, this question is not relevant.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in a larger, bolder font, and "ADMINISTRATION" in a smaller font below.

July 21, 2023

Shen Zhen Rong Feng Technology Co., Ltd Iris Fung Regulation Manager 3/F R Building ShaSi Industrial Park, Shajing Town BaoAn district Shenzhen, Guangdong China

Re: K223037

Trade/Device Name: Infrared Thermometer, Model: IR8807 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: June 18, 2023 Received: June 20, 2023

Dear Iris Fung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Porsche Bennett

Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223037

Device Name Infrared Thermometer, Model: IR8807

Indications for Use (Describe)

The Infrared Thermometer (Model: IR8807) is intended for the intermittent measurement of human body temperature from the forehead. It is a non-contact clinical thermometer and intended for use on people of all ages except for neonates/newborns.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary-K223037

This summary of 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR 807.92.

1. Submitter Information

Sponsor Company Name: SHEN ZHEN RONG FENG TECHNOLOGY CO., LTD

  • Address: 3/F R building ShaSi Industrial Park, ShajingTown, BaoAndistrict, ShenZhen � city, GuangDong Province, China
  • Phone: +86-13715151187 � Fax: +86-020-82591255
  • Contact Person (including title): Hong Xiao Lan (Engineer) �
  • E-mail: 844352259@qq.com

Application Correspondent: SHEN ZHEN RONG FENG TECHNOLOGY CO., LTD

Address: 3/F R building ShaSi Industrial Park, Shajing Town, BaoAn district, ShenZhen city,

GuangDong Province, China

  • � Contact Person: Iris Fung
  • Tile: Requlation Manager �
  • Tel: +86-13211147965 �
  • � Email: mdc-fs@foxmail.com; med-jl@foxmail.com
  • Summary Prepared Date � July 21, 2023
  • Subject Device Information: 2.
  • � Type of 510(k) submission: Traditional
  • � Common Name: Infrared Thermometer
  • Trade Name: Infrared Thermometer, Model: IR8807 �
  • Classification Name: Clinical Electronic Thermometer �
  • Review Panel: General Hospital �
  • � Product Code: FLL
  • Regulation Number: 21 CFR 880.2910 �
  • Regulation Class: 2 �

Predicate Device Information: 3.

  • 510(k) number: K173048
  • � Sponsor: Dongguan SIMZO Electronic Technology Co., Ltd.

{4}------------------------------------------------

  • Trade Name: Non-contact Forehead Thermometer, Model: HW-2/HW-3/HW-3/HW-4 /HW-� 4S/HW-302/HW-303
  • Common Name: Infrared Thermometer �
  • Review Panel: General Hospital �
  • � Product Code: FLL
  • Regulation Number: 21 CFR 880.2910 �
  • Requlation Class: 2 �

Device Description 4.

The Infrared Thermometer (Model: IR8807) is an electronic thermometer using an infrared sensor to measure infrared energy radiated from the forehead. This energy is collected through the lens and converted to a temperature value. The Infrared Thermometer (Model: IR8807) has a forehead measurement mode and an object measurement mode.

The Infrared Thermometer (Model: IR8807), consists of the following parts:

  • a) Thermopile Sensor
  • b) Application-Specific Integrated Circuitry
  • c) Erasable Programmable Read-Only Memory Integrated Circuit
  • d) Capacitance-touch Integrated Circuit
  • e) LCD display
  • f) 2 buttons (Scan button, Setting button,)
  • q) 2×1.5V AAA dry batteries

The operation principle is based on infrared sensor technology. The sensor in the thermometer detects the infrared energy, and the device processes the data to body temperature result with its internal microcontroller unit. It operates in an adjusted mode.

The main functions of Infrared Thermometer (Model: IR8807) are as followings:

  • Forehead temperature measure functions �
  • Wide range of temperature readings: from 32.0ºCto 42.9ºC (89.6ºF ~ 109.2ºF) �
  • The measurement results can be displayed in either °Cor °F units �
  • The LCD display showing clear measurement result. �
  • Memory data reading/removal function: The device can store a total of 32 groups of � memorable values. All memory values can be removed after the battery is removed
  • Low battery indicator �
  • Buzzer reminder function �
  • Users can set temperature alarm value �
  • Environment or object temperature measurement function � The reference body site of the output temperature is Oral.

{5}------------------------------------------------

5. Indications for use

The Infrared Thermometer (Model: IR8807) is intended for the intermittent measurement of human body temperature from the forehead. It is a non-contact clinical thermometer and intended for use on people of all ages except for neonates/newborns.

Elements ofComparisonSubject DevicePredicate DeviceComparison
ManufacturerSHEN ZHEN RONG FENGTECHNOLOGY CO., LTD.Dongguan SIMZO ElectronicTechnology Co.Ltd.
510 (k) NumberK223037K173048
Product NameInfrared ThermometerNon-contact ForeheadThermometer
ModelsModel: IR8807HW-2/HW-2S/HW-3/HW-4/HW-4S/HW-302/HW-303Note 1
Intended UseThe Infrared Thermometer (Model:IR8807) is intended for theintermittent measurement of humanbody temperature from the forehead.It is a non-contact clinicalthermometer and intended for useon people of all ages except forneonates/newborns.The Non-contact ForeheadThermometer is an infraredthermometer intended for theintermittent measurement ofhuman body temperature inpeople of all ages.Note 1
SensorInfrared SensorInfrared SensorSame
MeasurementModeForehead measurement ModeForehead measurement ModeSame
Measurementcontact typeNon-contact typeNon-contact typeSame
MeasurementtypeAdjustedAdjustedSame
MeasurementmethodInfrared radiation detectionInfrared radiation detectionSame
Measurementtime≤3S1SNote 2
MeasurementDistance2-5cmHW-2/HW-2S/HW-3/HW-302/HW-303:5-8cm HW-4/HW-4S:1-2cmNote 3
MeasuringRange32°C ~ 42.9°CForehead mode: 35.5°C~42.9°C (95.9°F~109.2°F)Note 4
AccuracyForehead mode:±0.2°C , forrange 35.0°C ~ 42.0°C±0.3°C, outside this temperaturerangeForehead mode:±0.2°C (0.4°F) within 35.5°C42.9°C(95.9°F109.2°F),±0.3°C(0.5°F) other rangeNote 5
DisplayResolution0.1°C/0.1°F0.1°C/0.1°FSame
Sensor typeThermopileThermopileSame
Elements ofComparisonSubject DevicePredicate DeviceComparison
HightemperatureindicationYesYesSame
BuzzerYesYesSame
Display screenLCDLCDSame
Auto power-offwhile nooperationYesYesSame
°C/°FswitchableYesYesSame
Memory32 setsHW-2/2S:32sets HW-3/4/4S: 1 setHW-302/303:64 SetsNote 6
Power Supply2 x AAA batteries2 x AAASame
OperatingConditionsTemperature: 10°C ~ 40°C Relativehumidity: ≤85% RHTemperature: 10~40°CRelative humidity: ≤85%RHSame
StorageConditionsTemperature: -20°C ~55°C Relativehumidity: ≤93%RH, non-condensingTemperature: -25°C- +55°CRelative humidity: ≤90%Note 7
Dimension182(L) x 54(W) x 47(H) mm9315341mmNote 8
WeightApproximately 125 g90-125 gNote 8
ConformancestandardEN60601-1(Safety),IEC60601-1-2(EMC)ASTME1965-98(performance)EN60601-1(Safety),IEC60601-1-2(EMC)ASTME1965-98(performance)Same
BiocompatibilityISO 10993-10 Skin Irritation test ISO10993-10Skin Sensitization testISO 10993-5 Vitro Cytotoxicity testISO 10993-10 Skin Irritationtest ISO 10993-10SkinSensitization testISO 10993-5 Vitro CytotoxicitytestSame
Sensor andpatient contactmaterialsInfrared Sensor ABSplastic shellInfrared SensorABS plastic shellSame

Comparison to Predicate Device 6.

{6}------------------------------------------------

Note 1

The Intended Use of subject device and predicate device is non-contact measurement of human body temperature from the forehead. The target population of the subject device is a subset of the predicate. This difference does not raise new or different questions of safety and effectiveness.

Note 2

Measurement time of the subject device is ≤3 seconds, the predicate device is 1 second. Both the subject and predict devices comply with ASTM E 1965-98. It is demonstrated that Page 4 of 7

{7}------------------------------------------------

the measure time does not affect the measurement accuracy. This difference does not raise new or different questions of safety and effectiveness.

Note 3

Measurement distance of the subject devices is 2-5cm, the predicate device measurement distance is 5- 8cm (models HW-2/HW-2S/HW-3/HW-302/ HW-303) or 1-2cm (models HW-4/HW-4S). Performance testing of the of the subject device demonstrates the accuracy meets the requirements within the distance range. The difference does not raise new or different questions of safety and effectiveness.

Note 4

Both devices have different measurement ranges; however, the measurement range of the subject device meets the requirements of ASTME1965-98. Therefore, the difference does not raise new or different questions of safety and effectiveness.

Note 5

Both devices have different accuracy requirements ; however, the measurement accuracy of the subject device and the predicate device meet the requirements of ASTME1965-98. Performance testing was conducted on the subject device according to ASTM E1965-98. Therefore, the difference does not raise new or different questions of safety and effectiveness.

Note 6

The memory capacity of the subject device is the same as the predicate device model: HW-2/2S. The memory capacity is only an auxiliary function. Software verification and validation testing conducted on the subject device demonstrates that the difference does not raise any new or different safety and effectiveness questions.

Note 7

There is a difference between the storage conditions of the predicate device and the subject device; however, both devices comply with IEC60601-1-11 and ASTM E1965-98. Therefore, the difference does not raise new or different questions of safety and effectiveness.

Note 8

There is a difference between the dimensions and weight of the predicate device and the subject device, however, the differences do not affect the safety and effectiveness as the subject device demonstrated adequate performance to requirements.

{8}------------------------------------------------

7. Test Summary - Non-Clinical Test

The subject device conforms to applicable standards that include:

  • � ASTM E 1965-98:2016 Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
  • � IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
  • � IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility
  • � IEC 62304:2006+AMD1:2015 Medical Device Software - Software Life Cycle Processes
  • � IEC60601-1-11:2015 medical electrical equipment - part 1-11: general requirements for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
  • � ISO 10993-5:2009, Biological evaluation of medical devices --Part 5: Testsfor In Vitro cytotoxicity,
  • � ISO 10993-10:2010, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization

Summary of Clinical Test 8.

Clinical tests were conducted on Model: IR8807. The study excluded subjects with medical conditions such as inflammation at the measuring sites and subjects using medications known to affect body temperature. The clinical tests evaluated 169 subjects which were divided into three age groups: infant (less than 1 year), children (1 t o 5 years old) and adult (greater than 5 years old). No less than 50 subjects in each group and 30% of them had temperature equaling or exceeding 37.5 °C.

Based on the result, the clinical performance of the subject device complied with the requirement of ASTM E1965-98 (2016).

9. Summary of software verification and validation Test

According to Software Safety Classification of IEC 62304, the infrared forehead thermometer IR8807 software shall be Class B. The Infrared Thermometer (model: IR8807) display data about the patient's body temperature supported by the software. According to the software unitlevel test, integration test and system level test, we hereby certify that all the functions those support the infrared forehead thermometer system specifications are proper. The Infrared Thermometer (model: IR8807) can be used safely and effectively to measure body temperature.

{9}------------------------------------------------

10. Conclusion

After analyzing the intended use, technological characteristics, verification and validation testing, it can be concluded that the differences between the subject and predicate device do not raise any new or different questions of safety and effectiveness. Therefore, subject device, Infrared Thermometer, Model: IR8807 is substantially equivalent to the predicate device, Noncontact Forehead Thermometer, Model: HW-2/HW-2S/HW-3/HW-4 /HW-4S/HW-302/HW-303, cleared under K173048.

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.