(9 days)
Not Found
No
The device is a simple examination glove and the summary contains no mention of AI or ML.
No
The device is a disposable examination glove used to prevent contamination, not to treat or diagnose a medical condition.
No
Explanation: A diagnostic device is used to detect, diagnose, or monitor a medical condition. This device is a glove, intended to prevent contamination between patient and examiner, not to diagnose.
No
The device is a physical examination glove, which is a hardware medical device, not software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states the device is "worn on the examiner's hands to prevent contamination between patient and examiner." This is a barrier function, not a diagnostic test performed on samples taken from the body.
- Device Description: The description is of a physical barrier (a glove) and its specifications (size, material, sterility). There is no mention of reagents, assays, or any components used to analyze biological samples.
- Lack of IVD Characteristics: The document does not mention any of the typical characteristics of an IVD, such as:
- Analyzing biological samples (blood, urine, tissue, etc.)
- Detecting specific substances or markers
- Providing diagnostic information about a patient's health status
- Using reagents or analytical methods
The device is a medical device, specifically a Class I medical device (based on the predicate device K180861, which is a Class I device). Its purpose is to provide a physical barrier for infection control, not to perform an in vitro diagnostic test.
N/A
Intended Use / Indications for Use
The Disposable Vinyl Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
Product codes
LYZ
Device Description
The subject device is powder free non-colored vinyl examination glove. It can be available in four specifications: S, M, L and XL. The subject device is non-sterile.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
hands
Indicated Patient Age Range
Not Found
Intended User / Care Setting
examiner's hands.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
ISO 10993-5 Third edition 2009-06-01, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves
ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D5250-19, Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.
Key Results:
Physical Dimensions Test (ASTM D5250): Length: >230/Pass; Width: S: 84-87 / Pass, M: 92-96 / Pass, L: 103-106/ Pass, XL: 112-117/ Pass; Thickness: Finger:0.13-0.15/ Pass, Palm:0.09-0.12/Pass.
Watertightness Test for Detection of Holes (ASTM D5151): 0/125 / Pass.
Powder Content (ASTM D6124): 0.070.15mg / Pass.24MPa / Pass, Ultimate Elongation: 347
Physical properties (ASTM D5250): Before Aging Tensile Strength: 13579% / Pass; After Aging Tensile Strength: 1119MPa / Pass, Ultimate Elongation: 340~529 % / Pass.
Cytotoxicity (ISO 10993-5): Non-Cytotoxicity / Pass.
Irritation (ISO 10993-10): Non-irritating / Pass.
Sensitization (ISO 10993-10): Non-sensitizing / Pass.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 8, 2022
Jiangxi Handspro Products Solutions CO., LTD. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801,No.161 Lujiazui East Rd.,Pudong Shanghai, Shanghai 200120 China
Re: K223008
Trade/Device Name: Disposable Vinyl Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: September 29, 2022 Received: September 29, 2022
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
1
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K223008
Device Name Disposable Vinyl Examination Gloves
Indications for Use (Describe)
The Disposable Vinyl Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
K223008
510(k) Summary
This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.
1.0 Submitter's Information
JIANGXI HANDSPRO PRODUCTS SOLUTIONS CO., LTD. Name: Address: Chuangye Avenue, Fenglin Industrial Zone of Dean County, Jiujiang City, Jiangxi, China Phone Number: +86-18616636918 Contact: Jun Yin Date of Preparation: Sep.29, 2022
Designated Submission Correspondent
Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn
2.0 Device Information
Trade name: | Disposable Vinyl Examination Gloves |
---|---|
Common name: | Vinyl Patient Examination Glove |
Classification name: | Non-powdered patient examination glove |
Model(s): | S, M, L, XL |
3.0 Classification
Production code: LYZ Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital
4.0 Predicate Device Information
Manufacturer: | ZHICHENG TRADING CO., LTD. |
---|---|
Device: | Vinyl Examination Glove (Clear, Non-Colored) |
510(k) number: | K180861 |
5.0 Indication for Use
4
The Disposable Vinyl Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
6.0 Device Description
The subject device is powder free non-colored vinyl examination glove. It can be available in four specifications: S, M, L and XL. The subject device is non-sterile.
7.0 Technological Characteristic Comparison Table
| Item | Subject Device
(K223008) | Predicated Device
(K180861) | Remark |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Product Code | LYZ | LYZ | Same |
| Regulation No. | 21CFR880.6250 | 21CFR880.6250 | Same |
| Class | I | I | Same |
| Intended Use | The DISPOSABLE
VINYL EXAMINATION
GLOVES is a disposable
device intended for
medical purposes that is
worn on the examiner's
hands to prevent
contamination between
patient and examiner. | The Vinyl Examination
Glove (Clear,
Non-Colored) is a
disposable device
intended for medical
purposes that is worn on
the examiner's hands to
prevent contamination
between patient and
examiner. | Same |
| Material | Vinyl | Vinyl | Same |
| Powdered or
Powered free | Powdered free | Powdered free | Same |
| Design Feature | Ambidextrous | Ambidextrous | Same |
| Colorant | Clear | Clear | Same |
| Labeling
Information | Single-use indication,
powder free, device
color, device name,
glove size and quantity,
Non-Sterile | Single-use indication,
powder free, device
color, device name,
glove size and quantity,
Non-Sterile | Same |
| Dimensions(mm) | Length:
S/M/L/XL: ≥230;
Width:
S: 85±5;
M: 95±5;
L: 105±5; | Length:
S≥230;
M≥235;
L≥245;
XL≥245;
Width: | Similar
Analysis 1 |
Table1-General Comparison
5
XL: 115±5 | S: 85±5; | |||||
---|---|---|---|---|---|---|
M: 95±5; | ||||||
L: 105±5; | ||||||
XL: 115±5 | ||||||
Thickness(mm) | Finger: ≥0.08; | |||||
Palm: ≥0.08 | Finger: ≥0.05; | |||||
Palm: ≥0.08 | Similar | |||||
Analysis 1 | ||||||
Tensile | 11MPa, | |||||
min | Tensile | |||||
Befor | ||||||
e | Strength | Strength | ||||
Aging | Ultimate | |||||
Elongation | 300% min | Ultimate | ||||
Elongation | 380% min | Similar | ||||
Analysis 2 | ||||||
Physical | ||||||
Properti | ||||||
es | ||||||
After | ||||||
Aging | Tensile | |||||
Strength | 11MPa, | |||||
min | Tensile | |||||
Strength | 15MPa, | |||||
min | Similar | |||||
Analysis 2 | ||||||
Ultimate | ||||||
Elongation | 300%min | Ultimate | ||||
Elongation | 380%min | Similar | ||||
Analysis 2 | ||||||
Freedom from | ||||||
Holes | Be free from holes when | |||||
tested in accordance | ||||||
with | ||||||
ASTMD5151 | ||||||
AQL=2.5 | Be free from holes when | |||||
tested in accordance | ||||||
with | ||||||
ASTMD5151 | ||||||
AQL=1.5 | Similar | |||||
Analysis 3 | ||||||
Powder Content | Meet the requirements of | |||||
ASTM D6124 | Meet the requirements of | |||||
ASTM D6124 | Same | |||||
Biocompatibility | ISO 10993-10; | |||||
Under the conditions of | ||||||
the study, not an irritant | ||||||
or a sensitizer | Comply with | |||||
ISO10993-10 | Same | |||||
Under conditions of the | ||||||
study, did not show | ||||||
potential toxicity to L-929 | ||||||
cells. Complies with ISO | ||||||
10993-5 | Under conditions of the | |||||
study, did not show | ||||||
potential toxicity to L-929 | ||||||
cells. Complies with ISO | ||||||
10993-5 | Same |
Analysis 1:
The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D5250.
Analysis 2:
The tensile strength and ultimate elongation are different with that of the predicate, but they all meet the requirements of ASTM D5250.
Analysis 3:
Freedom from holes of subject device is similar with that of the predicate, but they all meet the requirements of ASTMD5151.
6
8.0 Summarv of Non-clinical Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
ISO 10993-5 Third edition 2009-06-01, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves
ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D5250-19, Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.
| Test
Method | Purpose | Acceptance Criteria | Results |
---|---|---|---|
ASTM | |||
D5250 | Physical | ||
Dimensions | |||
Test | Length(mm): | ||
S/M/L/XL: ≥230; | |||
Width(mm): | |||
S: 85±5; | |||
M: 95±5; | |||
L: 105±5; | |||
XL: 115±5 |
Thickness (mm):
Finger: ≥0.08
Palm: ≥0.08 | Length(mm):
230/Pass;
Width(mm):
S: 84-87 / Pass
M: 92-96 / Pass
L: 103-106/ Pass
XL: 112-117/ Pass
Thickness (mm):
Finger:0.13-0.15/ Pass
Palm:0.09-0.12/Pass |
| ASTM
D5151 | Watertightness
Test for
Detection of
Holes | Meet the requirements of ASTM
D5151 AQL 2.5 | 0/125 / Pass |
| ASTM
D6124 | Powder
Content | Meet the requirements of ASTM
D6124