AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. It is not intended for use in bone.

Semi-Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. It is not intended for use in bone.

Coaxial Biopsy Needle is intended for use as a guiding needle in obtaining core biopsy samples from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. It is not intended for use in bone.

Device Description

Automatic Disposable Biopsy Needle, Semi-automatic Disposable Biopsy Needle and Coaxial Biopsy Needle are hand-operated, non-electronic, surgical instruments.

Automatic Disposable Biopsy Needle is designed for the automatic extraction of a specimen from soft tissues, while causing minimal surrounding tissue damage, for tissue pathological examination/ testing. Automatic Disposable Biopsy Needle is first loaded and then inserted into the edge of the target tissue. Then, the inner needle rod is threaded into the target lesion (automatic firing), then, the outer needle tube is fired to push forward, and the tissue sample is cut through the relative movement of the outer needle tube and the inner needle rod of the biopsy needle, later, the tissue sample is cut off and stored in the sampling groove. Finally, specimen was removed after withdrawing the biopsy needle.

Semi-automatic Disposable Biopsy Needle is designed for the extraction of a specimen from soft tissues, while causing minimal surrounding tissue damage, for tissue pathological examination/testing. The semi-automated biopsy needle requires manual advancement of the inner needle to expose the specimen notch. With pressure on its plunger, a spring action rapidly advances the outer needle (cutting cannula) over the specimen notch of the inner needle.

Coaxial Biopsy Needle is used with biopsy needles to quide the insertion of biopsy needle into the soft tissue under imaging control (ultrasound, X-ray, CT, etc.). It is supplied with trocar tip stylet with or without blunt tip needle.

All of these devices are sterile with a Sterility Assurance Level (SAL) of 10-6, nonpyrogenic and single-use devices.

AI/ML Overview

This submission is a 510(k) Pre-Market Notification for a traditional medical device (a biopsy needle), not an AI/ML-enabled device. As such, the information provided does not include the details typically found in submissions for AI/ML devices regarding acceptance criteria, training/test sets, expert adjudication, or MRMC studies.

The document focuses on demonstrating substantial equivalence to predicate devices through technical characteristics, non-clinical bench testing (biocompatibility, packaging, shelf life), and the assertion that "Bench testing is sufficient to demonstrate performance of the device. No preclinical testing of the subject device is necessary" and "Performance Test is sufficient to demonstrate safety and effectiveness of the subject devices with the predicate devices."

Therefore, I cannot extract the requested information about acceptance criteria for an AI/ML device, training/test set details, expert ground truth establishment, or clinical study methodologies (like MRMC) from this document.

The document discusses the following types of tests and their conformity:

1. Acceptance Criteria and Device Performance (Summary of Non-Clinical Testing):

The acceptance criteria are not explicitly numerical thresholds like sensitivity/specificity for an AI model. Instead, they are met by demonstrating compliance with recognized consensus standards and by comparing performance to predicate devices. The "reported device performance" is essentially that the device passed these tests and met the design requirements.

Acceptance Criterion TypeStandard/Test ConductedReported Device Performance (Met Acceptance Criteria)
BiocompatibilityISO 10993 seriesCytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogen Test, Hemolysis (All passed)
Package ValidationISO 11607-1, ISO 11607-2Integrity performance (Passed)
Transport-Packaging not damaged after compression, vibration, shock (Passed)
Shelf LifeAccelerated agingPhysical/chemical performance, package integrity tests on aged samples complied with pre-determined acceptance criteria (Passed)
Comparative Performance-Comparison tests (scale mark identification, puncture force, biopsy sample testing, stiffness, resistance to breakage, resistance to corrosion, joint strength, total heavy metal content) showed no significant risks compared to predicate devices (Passed)
EO/ECH ResidualISO 10993-7Residuals comply with allowable limits (Passed)
Standard ComplianceISO 9626, ISO 10993 series, ISO 11607-1, ISO 11607-2, ASTM F 1980Results confirmed design requirements (Passed)

2. Sample size used for the test set and the data provenance:

  • This information is not applicable and not provided in the context of this traditional medical device submission. The "test set" here refers to physical devices undergoing bench testing, not a dataset for an AI model.
  • The data provenance would be that the tests were conducted by the manufacturer according to specified standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable as this is not an AI/ML device requiring expert-labeled ground truth for a test set. Ground truth for the device's performance is established through objective physical and chemical testing against established standards and comparisons to predicate devices.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable. This concept is relevant for AI/ML model training/testing, not for physical device bench testing.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is not an AI/ML device used for reading or interpretation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is a physical biopsy needle, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Ground truth for this device's performance is based on pre-defined engineering specifications, material properties, sterility assurance levels, biocompatibility standards, and functional requirements (e.g., puncture force, resistance to breakage, sample integrity). It is not derived from clinical outcomes or expert consensus on image interpretation.

8. The sample size for the training set:

  • Not applicable. This device does not involve a "training set" in the context of machine learning.

9. How the ground truth for the training set was established:

  • Not applicable. No machine learning training set is involved.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.

April 3, 2023

Canyon Medical Inc. Huiqi Jiang RA Building 3, Phase 2 Accelerator, No. 11 Yaogu Avenue Jiangbei New Area Nanjing, Jiangsu Province 210032 China

Re: K222865

Trade/Device Name: M Biopsy /SureCore Automatic Disposable Biopsy Needle, M Biopsy /SureCore Semi-Automatic Disposable Biopsy Needle, M Biopsy /SureAim Coaxial Biopsy Needle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: March 3, 2023 Received: March 3, 2023

Dear Huiqi Jiang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-

542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

MarkDigitally signed by
TrumboreMark Trumbore -S
-SDate: 2023.04.0311:20:56-04'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K222865

Device Name

Automatic Disposable Biopsy Needle Semi-automatic Disposable Biopsy Needle Coaxial Biopsy Needle

Indications for Use (Describe)

Automatic Disposable Biopsy Needle:

Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. It is not intended for use in bone.

Semi-automatic Disposable Biopsy Needle:

Semi-Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. It is not intended for use in bone.

Coaxial Biopsy Needle:

Coaxial Biopsy Needle is intended for use as a guiding needle in obtaining core biopsy samples from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. It is not intended for use in bone.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

{4}------------------------------------------------

1. 510(k) Submitter Information:

Submitter Name: Canyon Medical Inc.

Address: Building 3, Phase 2 Accelerator, No. 11 Yaogu Avenue, Jiangbei New Area, 210032, Nanjing, Jiangsu Province, People's Republic of China

Submission Correspondent: Ms. Huiqi Jiang, RA

E-mail: jhq@canyonmedical.com.cn

Telephone: +86-13241196637

2. Date of Preparation: January 24, 2023

Common Name:Automatic Disposable Biopsy NeedleSemi-automatic Disposable Biopsy NeedleCoaxial Biopsy Needle
Trade Name:M·Biopsy /SureCore Automatic Disposable Biopsy NeedleM·Biopsy /SureCore Semi-automatic Disposable Biopsy NeedleM·Biopsy /SureAim Coaxial Biopsy Needle
Regulation Number:21 CFR 876.1075
Classification Name:Instrument, Biopsy
Device Class:II
Product Code:KNW
Review Panel:Gastroenterology/Urology

3. Subject Device(s) Information:

4. Predicate Device(s) Information:

Subject devicePredicate device
Automatic DisposableBiopsy NeedleK133948 BARD® MAX-CORE® Disposable Core BiopsyInstrument
Semi-automatic DisposableBiopsy NeedleK171953 BARD® MISSION® Disposable Core BiopsyInstrument
Coaxial Biopsy NeedleK171953 BARD® TRUGUIDE® Disposable Coaxial BiopsyNeedle

5. Device Description

Automatic Disposable Biopsy Needle, Semi-automatic Disposable Biopsy Needle and Coaxial Biopsy Needle are hand-operated, non-electronic, surgical instruments.

Automatic Disposable Biopsy Needle is designed for the automatic extraction of

{5}------------------------------------------------

510(K) Summarv

a specimen from soft tissues, while causing minimal surrounding tissue damage, for tissue pathological examination/ testing. Automatic Disposable Biopsy Needle is first loaded and then inserted into the edge of the target tissue. Then, the inner needle rod is threaded into the target lesion (automatic firing), then, the outer needle tube is fired to push forward, and the tissue sample is cut through the relative movement of the outer needle tube and the inner needle rod of the biopsy needle, later, the tissue sample is cut off and stored in the sampling groove. Finally, specimen was removed after withdrawing the biopsy needle.

Semi-automatic Disposable Biopsy Needle is designed for the extraction of a specimen from soft tissues, while causing minimal surrounding tissue damage, for tissue pathological examination/testing. The semi-automated biopsy needle requires manual advancement of the inner needle to expose the specimen notch. With pressure on its plunger, a spring action rapidly advances the outer needle (cutting cannula) over the specimen notch of the inner needle.

Coaxial Biopsy Needle is used with biopsy needles to quide the insertion of biopsy needle into the soft tissue under imaging control (ultrasound, X-ray, CT, etc.). It is supplied with trocar tip stylet with or without blunt tip needle.

All of these devices are sterile with a Sterility Assurance Level (SAL) of 10-6, nonpyrogenic and single-use devices. The ultrasound, X-ray, CT and other equipments are used to guide the puncture and sampling. These devices can't be used under MRI.

6. Indications for Use

Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. lt is not intended for use in bone.

Semi-Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. It is not intended for use in bone.

Coaxial Biopsy Needle is intended for use as a guiding needle in obtaining core biopsy samples from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. It is not intended for use in bone.

7. Technical Characteristics

The subject devices are substantially equivalent to the predicate devices in terms of intended use and technological characteristics. The differences between the subject

{6}------------------------------------------------

devices and predicate devices do not affect the basic design principle, usage, effectiveness and safety of the subject device. Equivalence has been identified as

follows:

Subject devicePredicate device
Automatic DisposableBiopsy NeedleK133948 BARD® MAX-CORE® Disposable Core BiopsyInstrument
Semi-automatic DisposableBiopsy NeedleK171953 BARD® MISSION® Disposable Core BiopsyInstrument
Coaxial Biopsy NeedleK171953 BARD® TRUGUIDE® Disposable Coaxial BiopsyNeedle

{7}------------------------------------------------

ItemSubject Device-CanyonPredicate Device-Bard
DeviceAutomatic Disposable BiopsyNeedleK133948 BARD® MAX-CORE®Disposable Core BiopsyInstrument
ClassificationClass IISame as subject device
Product CodeKNWSame as subject device
Regulation Number21 CFR 876.1075Same as subject device
Regulation NameGastroenterology-Urology Biopsy InstrumentSame as subject device
Indications for UseIntended for use in obtainingbiopsies from soft tissues such asliver, kidney, prostate, spleen,lung, thyroid and lymph nodes. It isnot intended for use in bone.Same as subject device
Target PopulationIndividuals requiringbiopsy for sampling ofsoft tissue abnormalitiesSame as subject device
Operation MechanicsSingle hand automatic activationSame as subject device
Gauge12G, 14G, 16G, 18G, 20G14G, 16G, 18G, 20G
Length (mm)100, 130, 160, 200, 250100, 160, 200, 250
PenetrationDepth(mm)11, 22Same as subject device
Slot Size (mm)1818, 19
SterilizationEO SterilizationSame as subject device
Single UseYesYes
Patient-contactingInfomarionExternally communicating device,in contact with the patient for alimited duration (no more than 24hours)Same as subject device
BiocompatibilityBiocompatible according to ISO10993 applicable partsBiocompatible according to ISO10993 applicable parts
ItemSubject Device-CanyonPredicate Device-Bard
DeviceSemi-automatic Disposable BiopsyNeedleK171953 BARD® MISSION®Disposable Core BiopsyInstrument
ClassificationClass IISame as subject device
Product CodeKNWSame as subject device
RegulationNumber21 CFR 876.1075Same as subject device
Regulation NameGastroenterology-Urology Biopsy InstrumentSame as subject device
Indications forUseIntended for use in obtaining biopsiesfrom soft tissues such as liver, kidney,prostate, spleen, lung, thyroid and lymphnodes. It is not intended for use in bone.Same as subject device
Target PopulationIndividuals requiringbiopsy for sampling ofsoft tissue abnormalitiesSame as subject device
OperationMechanicsSingle hand semi-automatic activationSame as subject device
Gauge14G, 16G, 18G, 20G14G, 16G, 18G, 20G
Length (mm)60, 100, 130, 160, 200, 250, 30060, 100, 160, 200, 250
Penetration Depth(mm)10, 20Same as subject device
SterilizationEO SterilizationSame as subject device
Single UseYesYes
Patient-contactingInformationExternally communicating device, incontact with the patient for a limitedduration (no more than 24 hours)Same as subject device
BiocompatibilityBiocompatible according to ISO 10993applicable partsBiocompatible according toISO 10993 applicable parts

Table 1. Comparison of the Automatic Disposable Biopsy Needle to the predicate device.

{8}------------------------------------------------

Table 2. Comparison of the Semi-automatic Disposable Biopsy Needle to the predicate device.

{9}------------------------------------------------

ItemSubject Device-CanyonPredicate Device-Bard
DeviceCoaxial Biopsy NeedleK171953 BARD® TRUGUIDE®Disposable Coaxial Biopsy Needle
ClassificationClass IISame as subject device
Product CodeKNWSame as subject device
Regulation Number21 CFR 876.1075Same as subject device
Regulation NameGastroenterology-Urology Biopsy InstrumentSame as subject device
Indications for UseIntended for use as a guiding needlein obtaining core biopsy samples fromsoft tissues such as liver, kidney,prostate, spleen, lung, thyroid andlymph nodes. It is not intended for usein boneSame as subject device
Operation MechanicsSingle hand automatic activationSame as subject device
Target PopulationIndividuals requiringbiopsy for sampling ofsoft tissue abnormalitiesSame as subject device
Gauge13G, 15G, 17G, 19G11G, 13G, 15G, 17G, 19G
SterilizationEO SterilizationSame as subject device
Single useYesYes
Patient contactingExternally communicating device, incontact with the patient for a limitedduration (no more than 24 hours)Same as subject device
BiocompatibilityBiocompatible according to ISO10993 applicable partsBiocompatible according to ISO10993 applicable parts

Table 3. Comparison of the Coaxial Biopsy Needle to the predicate device.

The technological characteristics of the subject device are identical to those of predicate device. The subject device has the same basic design as the predicate device. The comparison between the subject and predicate devices is based on the following:

  • · Same indications for use
  • Same material types that meet ISO 10993 biocompatibility requirements
  • Same sterilization methods
  • Same fundamental technology

There is no significant risk raised by the difference.

{10}------------------------------------------------

8. Summary of Non-Clinical Testing

Summary of non-clinical and performance bench testing was performed to evaluate the performance and functionality of the subject device aqainst requirement specification. The subject device has been subjected to compliance testing according to, by FDA, recognized consensus standards ISO 9626, ISO 10993 series standards, ISO 11607-1, ISO 11607-2, ASTM F 1980 etc. Results from testing performed confirm the design requirement.

    1. Biocompatibility
      The subject devices are externally communicating medical devices with tissue less than 24h which means limited duration. Therefore, as per "Table A.1: Biocompatibility Evaluation Endpoints" of FDA quidance, following items shall be carried out.
No.Test ItemStandard
1CytotoxicityISO10993-5:2009
2Maximization SensitizationISO10993-10:2010
3Intracutaneous ReactivityISO10993-10:2010
4Acute Systemic ToxicityISO10993-11:2017
5Pyrogen TestISO10993-11:2017/ USP General Chapter <151>
6HemolysisISO 10993-4:2017
    1. Package Validation
      The primary package of the proposed device is intended to maintain the sterility of the product during its claimed shelf life. The integrity performance as per ISO 11607-1:2019 and ISO 11607-2:2019 is carried out.
    1. Transport
      Subject devices are processed by compression, first vibration, shock and second vibration, and then check. The result indicates that packaging is not damaged.
    1. Shelf Life
      Accelerated aging was used to simulate the storage of 5 years, then the physical performance tests, chemical performance tests as well as the package integrity tests were performed on the accelerated aged samples. The test results demonstrate that the aged samples complied with the pre-determined acceptance criteria.
    1. Comparative Claims/Performance Comparison with predicate devices
      The test was conducted to the predicate device and subject device to compare their performance, including scale mark identification, puncture force, biopsy sample testing, stiffness, resistance to breakage, resistance to corrosion, joint strength, total heavy metal content.

{11}------------------------------------------------

6) EO/ECH Resudial

Examination of the EO and ECH residuals have been conducted in accordance with ISO 10993-7:2008+Amd.1:2019 to evaluate whether the sterilized proposed device comply with the above selected allowable limits, and the results meet the requirements.

9. Preclinical (Animal) Studies

Bench testing is sufficient to demonstrate performance of the device. No preclinical testing of the subject device is necessary.

10. Clinical Test

No. Use of biopsy needle is proven technology and is well accepted by the medical community. Performance Test is sufficient to demonstrate safety and effectiveness of the subject devices with the predicate devices.

11. Conclusion

The differences between both devices are insignificant in terms of safety and effectiveness. The conclusions drawn from information above demonstrate that the proposed devices are as safe, as effective, and performs as well as the legally marketed predicate devices and raises no new risks of safety or effectiveness.

§ 876.1075 Gastroenterology-urology biopsy instrument.

(a)
Identification. A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification. (1) Class II (performance standards).(2) Class I for the biopsy forceps cover and the non-electric biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.