(123 days)
The Sentimag Magnetic Localization System when used with the Magseed® marker is indicated to assist in localizing soft tissue lesions.
The Endomag Sentimag System is the only non-imaging guidance system intended for use with the Magsed® magnetic marker.
The Sentimag Gen 3 System is the 3d generation of Sentimag detector, intended for the nonimaging detection and localization of Magseed® magnetic marker that has been implanted in a lumpectomy site intended for surgical removal. It consists of a base unit and a probe.
This generation of the device has two modes:
- a) Counts Mode: This mode is the exact equivalent of the current generation of the Sentimag detector (Sentimag Gen 2 - described in K153044). In this mode Magseed (K173587) and is detectable based on a counts signal displayed on the base unit of Sentimag device.
- b) Measure Mode: Please note that this mode has a different graphic display and is currently dormant in Sentimag Gen 3
The Sentimag Gen 3 System comprises of a mains-powered base unit, a detachable hand-held probe available in 2 sizes (Standard Probe and Thinner Probe) that is connected to the base unit with a flexible cable, and a detachable electronic footswitch that is connected to the base unit with a flexible cable. A single use sterile sheath is required to be placed over the probe prior to the procedure, this is to be supplied by the hospital.
The hand-held probe which emits an alternating magnetic field that detects the magnetic response of the marker. When the marker is exposed to a magnetic field it becomes magnetic, and this magnetism can then be detected using the probe which gives an audible and visual signal of the level of magnetism. In count mode this is in the format of a numerical reading.
Acceptance Criteria and Device Performance Study for Sentimag System (K222832)
Based on the provided document, the Sentimag System (Gen 3) is a magnetic localization system intended to assist in localizing soft tissue lesions when used with the Magseed® marker. The submission describes the device's substantial equivalence to previous generations, particularly regarding its "Counts Mode." The "Measure Mode" is currently dormant and subject to specific risk control measures to prevent its use.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a formal table of "acceptance criteria" with numerical targets for clinical performance metrics (e.g., sensitivity, specificity, accuracy) for the Sentimag System (Gen 3) itself. Instead, the performance studies primarily focus on demonstrating substantial equivalence to its predicate device (Sentimag Gen 2) in its "Counts Mode" and verifying basic functional performance.
The key acceptance criterion, though implicitly, is that the Sentimag Gen 3 in Counts mode performs equivalently to the Sentimag Gen 2 in detecting the Magseed® marker.
| Acceptance Criteria (Implicit) | Reported Device Performance (Sentimag Gen 3 - Counts Mode) |
|---|---|
| Equivalent Sensitivity to Sentimag Gen 2 in detecting Magseed® | Sentimag (Gen 3) had equivalent sensitivity to Sentimag (Gen 2) in counts mode. |
| Equivalent Functional Performance with Standard and Thin Probes | Functional performance of both standard and thin probes is equivalent in both Gen 3 modes (and in Counts mode, equivalent to Sentimag Gen 2 predicate). |
| Detectable signal with Magseed® in various simulated tissues | Simluated Use of Sentimag (Gen 3) with Magseed (K173587) in Counts mode showed detection in Liver, Thyroid (CIRS Model 074), Breast (CIRS Model 073), Abdominal Tissue Plate (SynDaver 141720, 2N Std, 5mm Fat, 5mm muscle), and Lung Model (SynDaver). |
| Visual and audible signal generation | Provides a visual and audible signal of the level of magnetism, similar to the predicate device. |
| Risk control measures for dormant Measure Mode are effective | Specific risk control measures (disabling the mode, distinct display, training, IFU directions) are in place to prevent wrong use. |
2. Sample Size and Data Provenance for Test Set
The reported studies are pre-clinical/laboratory testing, not human clinical trials with specific test sets of patients. Therefore, information on sample size for a "test set" in the traditional clinical sense, country of origin, or retrospective/prospective nature of patient data is not applicable or provided.
The "test set" in this context refers to the simulated tissues and the devices themselves. The document mentions:
- Usability testing: of Sentimag (Gen 3) - Standard and thin probe. (Number of users/tests not specified).
- Sensitivity testing: of Sentimag (Gen 3) standard and thin probes against Magseed (K173587). (Number of tests not specified).
- Comparison of Sentimag (Gen 3) and Sentimag (Gen 2) sensitivity: in counts mode. (Number of comparisons not specified).
- Simulated Use: Using various simulated tissues (Liver, Thyroid, Breast, Abdominal Tissue Plate, Lung Model). The number of instances or "samples" for each tissue type is not specified.
The data provenance is from laboratory testing, implying it was conducted by Endomagnetics Ltd. or associated testing facilities.
3. Number of Experts and Qualifications for Ground Truth
Since the studies are pre-clinical/laboratory tests focusing on device functionality and equivalence, there were no human experts used to establish a ground truth for clinical diagnostic performance. The "ground truth" here is the physical presence and detectability of the Magseed® marker within controlled laboratory settings and simulated tissues.
4. Adjudication Method for the Test Set
As there were no human experts or clinical cases requiring interpretation, an adjudication method (like 2+1, 3+1) is not applicable or described. The "ground truth" was established by the experimental setup and the known presence/absence of the Magseed® marker.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No multi-reader multi-case (MRMC) comparative effectiveness study was done. The submission focuses on device functionality and substantial equivalence for detection, not on the improvement of human reader performance with AI assistance. The Sentimag System is a non-imaging guidance system and does not involve AI for interpretation by human readers.
6. Standalone Performance Study (Algorithm Only)
The Sentimag System inherently operates in a standalone manner in its "Counts Mode" to detect the Magseed® marker. The function is the algorithm only in terms of detecting the magnetic signal and converting it into visual and audible output. The studies described (sensitivity testing, simulated use) represent this standalone performance, implicitly demonstrating that the algorithm (the detection mechanism) functions as intended. It's not an "AI algorithm" in the conventional sense that processes images or makes diagnostic recommendations, but rather a direct physical detection system.
7. Type of Ground Truth Used
The ground truth used in these pre-clinical/laboratory studies was:
- Known physical presence and location of the Magseed® marker: within the controlled laboratory environment and specific simulated tissue models.
- Controlled experimental conditions: allowing for direct measurement of the device's response to the marker.
8. Sample Size for the Training Set
The document does not describe a "training set" in the context of machine learning or AI. The Sentimag System is a physical detection device, not an AI-based system that requires training data. The device's operational parameters are designed and validated, not "trained" on a dataset.
9. How Ground Truth for Training Set Was Established
As there is no "training set" for an AI algorithm, the concept of establishing ground truth for it is not applicable. The device's detection capabilities are based on established principles of electromagnetism and magnetometry, developed through engineering design and validated through testing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
1/21/2023
Endomagnetics Ltd Mehryar Behizad Regulatory Director 330 Cambridge Science Park, Milton Road Cambridge, Cambridgeshire CB4 0WN United Kingdom
Re: K222832
Trade/Device Name: Sentimag System Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: December 19, 2022 Received: December 20, 2022
Dear Mehryar Behizad:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Deborah A. Fellhauer -S
Deborah Fellhauer RN. BSN Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K222832
Device Name Sentimag System
Indications for Use (Describe)
The Sentimag Magnetic Localization System when used with the Magseed® marker is indicated to assist in localizing soft tissue lesions.
The Endomag Sentimag System is the only non-imaging guidance system intended for use with the Magsed® magnetic marker.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY
SUBMITTER INFORMATION
| Submitter's Name:Address: | Endomagnetics Ltd.330 Cambridge Science Park, Milton RoadCambridgeCB4 0WN |
|---|---|
| Contact Person: | Mehryar Behizad PhD |
| Regulatory Director | |
| Tel: | +44 1223 652540 |
| Email: | mbehizad@endomag.com |
| Date summary prepared: | 25th August 2022 |
DEVICE INFORMATION
| Trade name: | Sentimag System |
|---|---|
| Common name: | Gen 3 |
| Classification name: | Marker, Radiographic Implantable |
| Regulation: | 21 CFR 878.4300 |
| Device Classification: | Class II |
| Product Code: | NEU |
PREDICATE DEVICE
-
Principal Predicate K153044 Sentimark Magnetic Marker System
-
Secondary Predicate K173587 Magseed® Magnetic Marker System
-Tertiary Predicate is Special 510(k) K163541 Magseed® Magnetic Marker System
The rationale for the ranking of the predicates is as follows:
The Principal Predicate is K153044. This Traditional 510(k) submission describes the SentiMark magnetic marker in conjunction with Sentimag. This system is used to mark and locate soft tissue for surgical removal. It describes the indication and the magnetic mechanism of detection and that the Sentimag detector is the unit that induces and detects the magnetic response in the Magseed magnetic marker. Magseed marker in combination with Sentimag (Gen 2) has been legally marketed in the United States since its 510(k) clearance on 2nd March 2016.
{4}------------------------------------------------
The Secondary Predicate is K173587. This has been identified as the second most important predicate because this is a derivative predicate that describes the use of the same system (Sentimag/Magseed). However, in this predicate, data for implantation of Magseed for more than 30 days was presented, reviewed and cleared by the FDA (16 February 2018).
The target tissue, magnetic mode of detection and materials of construction are exactly the same as the primary predicate. The overall indication (marking, detection and removal of soft tissue) remained the same as the Principal Predicate, but the extended duration to more than 30 days for the Magseed marker implantation was approved. This has no bearing on the Sentimag Gen 2 detector.
The Tertiary Predicate is Special 510(k) K163541. This Premarket Submission is solely concerned with the change of name for the marker from SentiMark to Magseed. There is no revision to the implanted marker and it has no bearing on the Sentimag Gen 2 detector. This was cleared on 13th January 2017.
DESCRIPTION OF DEVICE
The Sentimag Gen 3 System is the 3d generation of Sentimag detector, intended for the nonimaging detection and localization of Magseed® magnetic marker that has been implanted in a lumpectomy site intended for surgical removal. It consists of a base unit and a probe.
This generation of the device has two modes:
- a) Counts Mode: This mode is the exact equivalent of the current generation of the Sentimag detector (Sentimag Gen 2 - described in K153044). In this mode Magseed (K173587) and is detectable based on a counts signal displayed on the base unit of Sentimag device.
b) Measure Mode: Please note that this mode has a different graphic display and is currently dormant in Sentimag Gen 3
Risk control measures in relation to users wrondly using Sentimag Gen 3 in Measure mode are outlined in Risk Management File: Use risk Analysis Appendix 1, line U 6.4.
Briefly these are:
- a) Magseed does not provide a clinically usable signal with Measure mode,
- b) The displays from the two modes are visually clearly distinguishable. Counts mode looks very similar to the Gen2 display with yellow graphics, while Measure mode is graphically very different and in a grey color (initially) as shown from the screen shots of the initial 'ready to use' display below:
Image /page/4/Figure/13 description: The image shows three different modes of a device. The first mode is "Gen 2" and displays a value of "0" along with a speaker icon and a probe icon. The second mode is "Gen3 Counts mode" and also displays a value of "0" along with a graph. The third mode is "Gen3 Measure mode" and displays a value of ">40mm" along with a visual representation of distance.
Initial 'Ready to use' screenshots from the Gen2 and Gen3 Sentimag displays
{5}------------------------------------------------
- c) The Gen 3 system remembers which mode was previously selected when it is next used, so once set in Counts mode, users would not have any need to change modes.
- d) Specific training will be provided to ensure end users use Sentimag Gen 3 system in the correct mode.
- e) The IFU will provide explicit direction in relation to not using the Measure mode in addition to the above.
Note: Other components of risk management file are enclosed as part of this submission.
The probability of end users utilising Sentimag Gen 3 in the inappropriate mode has been minimised through a series of risk control measures outlined in Use risk analysis, Appendix 1, lines U 6.4.
Whilst we recognize that there is the possibility that end users might accidentally switch to measure mode, we believe the visual difference, lack of signal and suitable training, would ensure the use of the appropriate mode. End users are trained to seek a signal before commencing surgery, so as to make sure the system is fully operational. Existing end users are familiar with counts mode through their use with Gen 2 and new users would be trained to use only this mode for the Magseed, and provided with instruction via the IFU.
The predicate for the Sentimaq Gen 3 device is the previous generation of the detector and was introduced in previous submissions to the FDA as Sentimaq but in fact was the second generation in this family of devices. So. throughout this submission in order to distinguish between the predicate detector and the detector which is the subject of this submission the predicate will be referred to as "Sentimag Gen 2" and the detector in this submission will be referred to as "Sentimag Gen 3".
Sentimag Gen 3 also supports a standard diameter and a thin probe. The functional performance of the two different probes is equivalent in the two different Gen 3 modes. In Counts mode both the probes are functionally equivalent to Sentimag (Gen 2) predicate.
The Sentimag System is designed for use in an operating room environment by suitably trained physicians who are experienced in diagnosis and treatment of breast lesions. The Sentimag System aids the surgeon to detect surgically invasively magnetic marker material that has been placed for the purpose of detecting a non-palpable lesion, and to locate target excision sites.
Prior to a lumpectomy procedure, the Magseed® Magnetic Marker which already has clearance (K173587) is placed percutaneously into the breast, using imaging guidance such as ultrasound or radiography, to temporarily mark a site intended for surgical removal. During a surgical procedure, the hand-held Sentimag probe emits an alternating magnetic field that detects the magnetic response of the Magseed® magnetic marker, this signal is converted by the base unit into a visual and audible response that is similar in use to the predicate device.
The Sentimag Gen 3 device is the successor to the Sentimag Gen 2 device (Principal Predicate K153044. Secondary Predicate K173587 and Tertiary Predicate 510(k) K163541). In common with the principal predicate, Endomag Sentimag Gen 3 device is able to detect the Magseed Magnetic Marker using proximity detection based on susceptometry.
{6}------------------------------------------------
INTENDED USE
The Sentimag system is intended for intra-operative use, to help surgeons detect and locate magnetic marker in the body. The system is specifically intended to detect Magseed® implantable magnetic marker used to mark soft tissue lesions.
The Endomag Magseed® Magnetic Marker (K173587) is placed within the target soft tissue prior to planned surgical removal. The currently cleared marker (K173587), when used in conjunction with the Sentimag Gen 3 System in Counts mode, can be used as a guide for the surgeon to follow in the excision of tissue.
SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
As the predicate Sentimag Gen 2 detector (K153044), the Sentimag Gen 3 System utilizes nonimaging technologies that are comprised of a console that incorporates electronics and a simple user interface, plus a probe handpiece. Like the predicate (Sentimag Gen 2), Sentimag Gen 3 is a proximity detector. The Sentimag Gen3 System comprises of a mains-powered base unit, a detachable hand-held probe available in 2 sizes (Standard Probe and Thinner Probe) that is connected to the base unit with a flexible cable, and a detachable electronic footswitch that is connected to the base unit with a flexible cable. A single use sterile sheath is required to be placed over the probe prior to the procedure, this is to be supplied by the hospital.
The hand-held probe which emits an alternating magnetic field that detects the magnetic response of the marker. When the marker is exposed to a magnetic field it becomes magnetic, and this magnetism can then be detected using the probe which gives an audible and visual signal of the level of magnetism. In count mode this is in the format of a numerical reading. In this mode the Sentimag Gen 3 device can measure Magseed® in exactly the same way as the predicate (Sentimag Gen 2).
A location marker is placed percutaneously in situ at the clinical target site by a delivery system using 5 cm to 12 cm delivery systems described in predicates K173587 and K163541 and then the detector handpiece is used for the intraoperative detection and localization of the implanted marker. The handpiece is connected by a flexible cable to a console unit that provides the user with a visual indication of the presence and proximity of the marker.
In Counts mode Sentimag Gen 3 can detect Magseed® (K173587) in exactly the same way and with exactly the same sensitivity as the predicate (Sentimag Gen 2 - described in K153044). In common with the Sentimag Gen 2 predicate, in Sentimag Gen 3 in Count mode a visible numerical representation of the detected signal level is simultaneously displayed on the base unit's liquid crystal display.
Additionally, Sentimag Gen 3 supports a Measure Mode. Please note that this mode is currently dormant in Sentimag Gen 3.
In both Count mode and Measure mode the sensing of the magnetic signal is indicated by a change in pitch (frequency) of an audio output from the Sentimag base unit.
The Gen 3 Sentimaq also provides an option for a thinner (13.5 mm diameter) probe. This probe functions in exactly the same way as a normal diameter (18.5 mm) probe in each mode described above.
{7}------------------------------------------------
DISCUSSION OF NON-CLINICAL TESTS SUBMITTED
Performance testing was conducted to evaluate and characterize the performance of the Sentimag System (Gen 3) with the previously approved Magseed®(K173587) Preclinical/Laboratory testing included:
- Usability testing of Sentimag (Gen 3)- Standard and thin probe .
- Sensitivity testing of Sentimag (Gen 3) standard and thin probes against Magseed . (K173587).
- . Comparison of Sentimag (Gen 3) and Sentimag (Gen 2) sensitivity in counts mode
- Simulated Use of Sentimag (Gen 3) with Magseed (K173587) in Counts mode using the . following simulated tissue:
-
- Liver
-
- Thyroid (CIRS Model 074)
-
- Breast (CIRS Model 073)
-
- Abdominal Tissue Plate (SynDaver 141720, 2N Std, 5mm Fat, 5mm muscle)
-
- Lung Model (SynDaver)
-
CONCLUSION
Endomagnetics believes that Sentimag System (K153044) is the predicate device because the Sentimag (Gen 3) is the same as the predicate Sentimag in Counts mode. The Measure mode in Sentimag Gen 3 is an additional function. The Sentimag Gen3 Measure mode is currently dormant.
Specific risk control measures have been adopted to prevent the end user from using this mode. These risk control measures are outlined in section 5.4 above. Sentimag Gen 3 has the same Intended Use and very similar technological characteristics to the predicate device.
The different technological characteristics do not raise any new questions of safety or effectiveness. The test, verification and validation data presented in this submission demonstrate substantial equivalence of the Sentimag Gen 3 to the predicate (Sentimag Gen 2).
§ 878.4300 Implantable clip.
(a)
Identification. An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.