K Number
K222829
Date Cleared
2023-04-17

(210 days)

Product Code
Regulation Number
866.3255
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Curian Shiga Toxin assay, for use with the Curian Analyzer, is a rapid, qualitative, fluorescent immunoassay for the simultaneous detection and differentiation of Shiga toxin 1 (Stx1) and Shiga toxin 2 (Stx2) in a single test device. It is intended for use with cultures derived from human stool specimens to aid in the diagnosis of disease caused by Shiga toxin producing Escherichia coli (STEC) infections. Test results are to be used in conjunction with the patient's clinical symptoms and history.

Device Description

The Curian® Shiga Toxin assay is a qualitative in vitro diagnostic test for the detection of Shiga toxin 1 (Stx1) and Shiga toxin 2 (Stx2) in cultures derived from human stool specimens. The Curian® Shiga Toxin assay utilizes fluorescence technology with the cleared Curian® Analyzer (K192817) to detect Stx1 and Stx2 in cultures derived from human stool.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria, but it presents the performance of the Curian Shiga Toxin assay against a reference method (Vero cell Cytotoxin Assay). We can infer the implicit acceptance criteria from the reported performance, which demonstrates high sensitivity and specificity.

Performance MetricAcceptance Criteria (Implied by High Performance)Reported Device Performance (Prospective Specimens)Reported Device Performance (Archived Specimens)
Stx1 SensitivityHigh (e.g., >90%)100.0% (95% CI: 56.6% - 100.0%)100.0% (95% CI: 92.3% - 100.0%)
Stx1 SpecificityHigh (e.g., >90%)99.4% (95% CI: 98.9% - 99.7%)97.8% (95% CI: 92.2% - 99.4%)
Stx2 SensitivityHigh (e.g., >90%)100.0% (95% CI: 51.0% - 100.0%)97.0% (95% CI: 84.7% - 99.5%)
Stx2 SpecificityHigh (e.g., >90%)99.5% (95% CI: 99.1% - 99.8%)98.0% (95% CI: 93.1% - 99.5%)

Note: The low end of the 95% CI for sensitivity in prospective specimens (e.g., 56.6% for Stx1) is due to the very small number of positive cases in that cohort (5 for Stx1 and 4 for Stx2). The 100% point estimate, while positive, has a wide confidence interval reflecting this small sample size. The archived cohort provides more robust sensitivity estimates.

2. Sample Size Used for the Test Set and Data Provenance

  • Prospective Test Set: 1,627 stool specimens collected from patients suspected of having a Shiga toxin-producing Escherichia coli (STEC) infection.
    • Evaluable Prospective Specimens: 1,538
    • Data Provenance: Prospective collection from five clinical study sites representing geographically distinct regions throughout the United States.
  • Archived Test Set: 140 archived stool samples.
    • Evaluable Archived Specimens: 135 (5 excluded due to inconclusive reference results).
    • Data Provenance: Retrospective testing of archived samples at all five clinical study sites.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the number of experts or their qualifications for establishing the ground truth. It mentions that the reference method was the "Vero cell Cytotoxin Assay (with neutralization) performed on the broth culture obtained from the stool specimen." This is a laboratory-based assay, and its interpretation would typically be performed by trained laboratory personnel rather than a panel of clinical experts (e.g., radiologists).

4. Adjudication Method for the Test Set

The document does not describe any adjudication method for the test set. The ground truth was established by the "Vero cell Cytotoxin Assay (with neutralization)."

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not conducted. This device is an in vitro diagnostic (IVD) assay interpreted by an analyzer, not a medical imaging or diagnostic aid that multiple human readers would interpret with and without AI assistance. The Curian Analyzer automates the interpretation of results ("Results interpretation automated by Curian® Analyzer").

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, the clinical performance study (both prospective and archived) represents the standalone performance of the device (Curian Shiga Toxin assay and Curian Analyzer). The results are "produced by the Curian Analyzer," indicating an automated, algorithm-only interpretation without a human-in-the-loop performance evaluation in the context of these clinical studies.

7. The Type of Ground Truth Used

The type of ground truth used was a laboratory reference method: the Vero cell Cytotoxin Assay (with neutralization) performed on broth cultures obtained from the stool specimens.

8. The Sample Size for the Training Set

The document does not provide information about the sample size for a training set. This is common for IVD submissions, where the focus is on the analytical and clinical performance of the device itself rather than the development of the underlying algorithms through a specific training set. The device is a "lateral flow fluorescent immunoassay," which is a biochemical detection method, not a machine learning algorithm that typically requires a large training dataset in the same way.

9. How the Ground Truth for the Training Set Was Established

Since no training set information is provided, there is no description of how ground truth for a training set was established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 17, 2023

Meridian Bioscience Inc. Heather Planck Regulatory Affairs Specialist 3471 River Hills Drive Cincinnati, Ohio 45244

Re: K222829

Trade/Device Name: Curian® Shiga Toxin Regulation Number: 21 CFR 866.3255 Regulation Name: Escherichia Coli Serological Reagents Regulatory Class: Class I Product Code: GMZ Dated: September 19, 2022 Received: September 19, 2022

Dear Heather Planck:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ribhi Shawar -S

Ribhi Shawar, Ph.D. Chief General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222829

Device Name Curian Shiga Toxin

Indications for Use (Describe)

The Curian Shiga Toxin assay, for use with the Curian Analyzer, is a rapid, qualitative, fluorescent immunoassay for the simultaneous detection and differentiation of Shiga toxin 2 (Stx2) in a single test device. It is intended for use with cultures derived from human stool specimens to aid in the diagnosis of disease caused by Shiga toxin producing Escherichia coli (STEC) infections. Test results are to be used in conjunction with the patient's clinical symptoms and history.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

510(k) number: K222829

Date of Preparation: April 17, 2023

A. 510(k) Number:

K222829

B. Purpose for Submission:

To determine substantial equivalence for the Curian® Shiga Toxin assay used to detect Shiga toxin 1 (Stx1) and Shiga toxin 2 (Stx2) in cultures derived from stool samples.

C. Measurand:

Shiga toxin 1 (Stx1) and Shiqa toxin 2 (Stx2)

D. Type of Test:

Lateral flow fluorescent immunoassay

E. Applicant:

Meridian Bioscience, Inc.

F. Proprietary and Established Names:

Curian® Shiga Toxin

G. Regulatory Information:

Product CodeClassificationRegulation SectionPanel
GMZI21 CFR 866.3255 – Escherichia coliserological reagentsMI – Microbiology (83)

H. Intended Use:

    1. Intended Use:
      The Curian Shiga Toxin assay, for use with the Curian Analyzer, is a rapid, qualitative, fluorescent immunoassay for the simultaneous detection and differentiation of Shiga toxin 1 (Stx1) and Shiga toxin 2 (Stx2) in a single test device. It is intended for use with cultures derived from human stool specimens to aid in the diagnosis of disease caused by Shiga toxin producing Escherichia coli (STEC) infections. Test results are to be used in conjunction with the patient's clinical symptoms and history.
    1. Indications for use:
      Same as Intended Use.
  • Special Conditions for use Statement(s): റ്റ്
    For prescription use only.

    1. Special Instrument Requirements:
      Curian Analyzer

{4}------------------------------------------------

l. Device Description:

The Curian® Shiga Toxin assay is a qualitative in vitro diagnostic test for the detection of Shiga toxin 1 (Stx1) and Shiga toxin 2 (Stx2) in cultures derived from human stool specimens. The Curian® Shiga Toxin assay utilizes fluorescence technology with the cleared Curian® Analyzer (K192817) to detect Stx1 and Stx2 in cultures derived from human stool.

Reagents and Test Components:

The Curian® Shiga Toxin assay kit contains the following components:

  • Curian® Shiga Toxin Test Card: A test strip enclosed in a plastic frame which is in a foil pouch with . a desiccant. The desiccant is pink; if the packaging is compromised, the desiccant will be blue. Supplied ready to use.
  • . Curian® Shiga Toxin Aioprep™ Sample Preparation Device/ Negative Control: A buffered aqueous protein solution containing blue dye and 0.094% sodium azide. The Aioprep™ device is fitted with a dropper tip. Supplied ready to use.
  • . Curian® Shiga Toxin Positive Control: Inactivated Shiga toxins 1 and 2 in an aqueous phosphate buffered solution containing 0.094% sodium azide. Supplied ready to use.
  • . Pipette I: Transfer pipettes graduated to 50 µL and 175 µL.
  • Pipette II: Transfer pipettes graduated up to 1.0 mL. .

J. Substantial Equivalence Information:

    1. Predicate device name(s): SHIGA TOXIN QUIK CHEK
    1. Predicate 510(k) number: K121364
    1. Comparison with predicate:
NEW DEVICECurian® Shiga ToxinPREDICATE DEVICESHIGA TOXIN QUIK CHEKK121364
Similarities
Intended Use/Indications for UseThe Curian Shiga Toxin assay, for use withthe Curian Analyzer, is a rapid, qualitative,fluorescent immunoassay for thesimultaneous detection and differentiation ofShiga toxin 1 (Stx1) and Shiga toxin 2 (Stx2)in a single test device. It is intended for usewith cultures derived from human stoolspecimens to aid in the diagnosis of diseasecaused by Shiga toxin producingEscherichia coli (STEC) infections. Testresults are to be used in conjunction with thepatient's clinical symptoms and history.The SHIGA TOXIN QUIK CHEK test is arapid membrane enzyme immunoassay forthe simultaneous qualitative detection anddifferentiation of Shiga toxin 1 (Stx1) andShiga toxin 2 (Stx2) in a single test device.It is intended for use with human fecalsamples from patients with gastrointestinalsymptoms to aid in the diagnosis ofdisease caused by Shiga toxin producingEscherichia coli (STEC). It may be usedwith fecal specimens, or broth or platecultures derived from fecal specimens. Thetest results should be considered inconjunction with the patient history.
ClassificationClass ISame
Product CodeGMZSame
Regulation21 CFR 866.3255Same

{5}------------------------------------------------

NEW DEVICECurian® Shiga ToxinPREDICATE DEVICESHIGA TOXIN QUIK CHEKK121364
Measured AnalyteShiga toxin 1 (Stx1) and Shiga toxin 2 (Stx2)Same
Antibody FormatMonoclonal/PolyclonalSame
Target PopulationPersons suspected of having diseasecaused by Shiga toxin producing E. coli(STEC) infectionsSame
Type of TestQualitativeSame
Specimen TypeHuman Stool SpecimenSame
Sample MatrixBroth and Plate Cultures Derived fromHuman Stool SpecimensSame
ControlsPositive and negative control included in thekit. Internal Control line.Same
Read Result Time< 30 minutesSame
FormatSingle Use CassetteSame
StorageRefrigerated (2-8 °C)Same
Differences
Sample MatrixNot for use with Direct Human StoolSpecimenDirect Human Stool Specimen
TechnologyLateral flow fluorescent immunoassayRapid membrane enzyme immunoassay
Interpretation ofResultsResults interpretation automated by Curian®AnalyzerVisual Reading

K. Test Principle:

The Curian Shiga Toxin assay is a lateral flow-based fluorescent immunoassay for the direct detection of Shiga toxin 1 (Stx1) and Shiga toxin 2 (Stx2) in cultures derived from human stool specimens. The Curian Shiga Toxin assay is based on standard immunoassay sandwich technology utilizing two detector antibodies (E. coli monoclonal antibodies specific to Stx1 or Stx2, which are conjugated to Europium as a fluorescent tag), and three capture antibodies: (a) anti-E. coli monoclonal antibodies specific to Stx1 or Stx2, which are affixed at the "Stx1 Test Line" or "Stx2 Test Line" positions of the test strip, and (b) a polyclonal goat anti-mouse IgG antibody, which is affixed at the "Control Line" position of the test strip.

The Aioprep sampling device (the "Aioprep") is pre-filled with blue tinted Sample Diluent / Negative Control and contains a filter and a dropper tip. A sample of the patient's culture enriched stool specimen is transferred from the enrichment media to the Aioprep, using either the included transfer pipettes (Pipette II) or a swab (not provided with the kit) to add the sample directly into the Sample Diluent / Negative Control. The diluted sample is mixed directly in the Aioprep barrel and then dispensed dropwise into the sample port of the Curian Shiga Toxin test card.

If Stx1 and/or Stx2 antigens are present in the sample, they bind to the respective monoclonal detector antibodies conjugated to fluorescent particles, forming a complex moves through the test strip, the anti-Shiga toxin 1 and/or 2 monoclonal capture antibodies, each bound to the assay membrane at their respective Stx1 and Stx2 test positions of the strip, bind the complexes to yield the respective test lines. When antigen is not present, a complex is not formed, and therefore, test lines do not form. As the sample continues to move further up the test strip, the polyclonal capture antibody, bound to the assay membrane at the control position of the strip, binds the conjugated antibodies and

{6}------------------------------------------------

yields the control line. A line at the control position of the test strip should be present each time a sample or external control is tested. If the control line is not generated, adequate sample flow has not occurred, and the Curian Analyzer will consider the test invalid. The output is a reported test result produced by the Curian Analyzer, indicating if Stx1 and/or Stx2 has been detected. The assay detects and differentiates between Stx1 and Stx2 in a single test device.

L. Performance Characteristics

1. Analytical Performance

Precision/Reproducibility a.

The reproducibility of the Curian Shiga Toxin assay was determined by testing contrived cultured broth samples across three independent laboratories. Samples were created with Stx1 and/or Stx2 spiked into pooled negative cultured broth matrix at high negative, low positive, and moderate positive concentrations, along with a true negative cultured broth sample. Ten panels consisting of 30 blinded samples comprising of various combinations of Stx1 and Stx2 concentrations were provided to each of the three laboratories for a total of 900 samples. Testing was conducted at each laboratory over 5 different days. Each day two separate operators tested each one a separate panel while alternating between kit lots. Testing included three different kit lots (2 lots per site). In addition, positive and negative controls were run daily.

For Stx1. the overall agreement between the Curian Shiga Toxin assay result and the expected assay result was 99.6% (95% Cl: 98.9% - 99.8%) with 100% for all sample types except for combined Stx1 and Stx2 high negative samples, which showed an agreement of 95.6% (95% Cl: 89.1% - 98.3%). For Stx2, the overall agreement between the Curian Shiga Toxin assay result and the expected assay result was 99.7% (95% C1: 99.0% - 99.9%) with 100% for all sample types except for Stx2 high negative samples, which showed an agreement of 97.8% (95% -99.4%), and combined Stx1 and Stx2 high negative samples, which showed an agreement of 98.9% (95% Cl: 94.0% - 99.8%). Results are summarized in the table below.

Panel Sample% Agreement (n)
Stx1 ResultsStx2 Results
Shiga Toxin 1 Sample Types
True Negative100% (90/90)100% (90/90)
High Negative100% (90/90)100% (90/90)
Low Positive100% (90/90)100% (90/90)
Moderate Positive100% (90/90)100% (90/90)
Shiga Toxin 2 Sample Types
True Negative100% (90/90)100% (90/90)
High Negative100% (90/90)97.8% (88/90)
Low Positive100% (90/90)100% (90/90)
Moderate Positive100% (90/90)100% (90/90)
Shiga Toxin 1 and 2 Sample Types
True Negative100% (90/90)100% (90/90)
High Negative95.6% (86/90)98.9% (89/90)
Low Positive100% (90/90)100% (90/90)
Moderate Positive100% (90/90)100% (90/90)

b. Linearity/reportable range:

Not applicable as this is a qualitative test.

Traceability, Stability, Expected values (controls, calibrators, or methods) C.

{7}------------------------------------------------

Specimen Stability:

The specimen stability claims include storage of up to 7 days at refrigerated temperature (2-8 C) or for up to 21 days frozen (-20°C and/or -80°C) for clinical broth culture enriched stool specimens prior to testing with the Curian® Shiga Toxin assay.

Freeze/Thaw:

The specimen freeze/thaw stability claims for clinical broth culture enriched stool specimens include freezing and thawing specimens for up to two (2) freeze/thaw cycles when specimens are stored frozen (≤ - 20 C) prior to testing with the Curian® Shiga Toxin assay.

Reagent Stability:

Accelerated Stability testing produced acceptable results to support up to eighteen (18) months expiry date at 2-8 C for the Curian® Shiga Toxin assay and kit components (i.e., test cards, Aioprep and Positive Control).

d. Analytical Sensitivity

Analytical sensitivity studies were performed to determine the analytical limit of detection (LoD) of quantified Stx1 and Stx2 in broth culture matrix derived from human stool for the Curian Shiga Toxin assay. Three lots of the Curian Shiqa Toxin assay were evaluated. For each kit lot, an LoD was established and confirmed in separate studies for each target toxin (Stx1 and Stx2). The LoD was defined as the lowest concentration of the target toxin that produced positive results ≥ 95% of the time.

The LoD values determined for the Curian Shiga Toxin assay for each toxin detected by the assay in culture matrix derived from human stool are 0.185 ng/mL for Stx1 and 0.125 ng/mL for Stx2.

e. Prozone / Hook Effect

A study was performed to determine the potential for a high dose prozone/hook effect with the Curian Shiga Toxin assay. Dilutions of quantified Stx1 and Stx2 were prepared in negative broth culture matrix to create contrived positive samples containing known concentrations of the target toxin. Individual reactions were prepared such that the concentration in each replicate was that of a high positive specimen, approximately 5xLoD - 377xLoD for Stx1 samples (ranging from 0.932 to 69.78 ng/mL) and 7xLoD - 558xLoD for Stx2 samples (ranging from 0.932 to 69.78 ng/mL). A total of n=7 dilutions were prepared for each toxin (Stx1 and Stx2), with an additional dilution prepared for a contrived combined Stx2 and Stx2 positive sample representing the highest concentration of Stx1 and Stx2 tested individually (69.78 ng/mL, approximately 377X LoD for Stx1 and 558X LoD for Stx2). Therefore, n=15 total dilutions were evaluated for this study. Each sample dilution was tested in replicates of five to determine whether a hook/prozone effect was observed with the Curian Shiga Toxinassay. A prozonel hook effect was not observed with the Curian Shiga Toxin assay when testing samples containing high concentrations of Stx1 and/or Stx2.

f. Cross-Reactivity:

A cross-reactivity and microbial interference study was performed to determine if potential cocontaminants of broth culture derived from human stool specimens would non-specifically react with the Curian Shiga Toxin assay, or interfere with detection of Stx1 or Stx2 when present at high concentrations. The specificity of Curian Shiga Toxin was evaluated by testing bacteria. fungi, and viral strains. Each organism was tested with a true negative sample and two contrived low positive samples (3xLoD; one for each of Stx1 and Stx2) at a minimum concentration of 1.2x107 CFU/mL (for bacterialfungi, with the exception of Shigella dysenteriae) or 1.0x105 TCIDsolmL (for viruses).

No cross-reactivity or microbial interference with the Curian Shiga Toxin assay was observed except for S. dysenteriae (strain ATCC 9361), which was found to be Stx1 positive at concentrations greater than 1.25 x 10° CFU/mL in the Curian Shiga Toxin assay. The organisms evaluated for cross-reactivity are listed below.

{8}------------------------------------------------

MicroorganismStrain IDMicroorganismStrain ID
Aeromonas hydrophilaATCC 35654Helicobacter pyloriCCUG 39500
Bacillus subtilisATCC 6051Klebsiella pneumoniaeATCC BAA-1900
Bacteroides fragilisATCC 23745Lactobacillus acidophilusATCC 4356
Campylobacter coliATCC 43482Proteus vulgarisATCC 6380
Campylobacter concisusATCC 33237Providencia stuartiiATCC 33672
Campylobacter fetusATCC 25936Pseudomonas aeruginosaATCC 10145
Campylobacter hyointestinalisATCC 35217Pseudomonas fluorescensATCC 13525
Campylobacter jejuniATCC 33292Salmonella enterica susb. entericaserovar HilversumATCC 15784
Candida albicansATCC 18804Salmonella enterica susb. entericaserovar TyphimuriumATCC 13311
Citrobacter freundiiATCC 43864Salmonella minnesotaATCC 9700
Clostridium difficileATCC 43255Serratia liquefaciensATCC 27592
Clostridium perfringensATCC 12915Serratia marcescensATCC 43862
Enterobacter cloacaeATCC 15337Shigella boydiiATCC 9207
Enterococcus faecalisATCC 49532Shigella dysenteriaeATCC 9361
Escherichia coli (non-toxigenic)ATCC BAA-2190Shigella flexneriATCC 12022
Escherichia coli EIECATCC 43893Shigella sonneiATCC 25931
Escherichia coli EPECATCC 33780Staphylococcus aureusATCC 51153
Escherichia coli ETECATCC 35401Staphylococcus aureus (Cowan's)ATCC 12598
Escherichia coli O157:H7 (non-toxigenic)ATCC 700728Staphylococcus epidermidisATCC 49134
Escherichia fergusoniiATCC 35469Streptococcus equisimilis subsp.dysgalactiaeATCC 12388
Escherichia hermaniiATCC 33650Yersinia enterocoliticaATCC 23715
Gardnerella vaginalisATCC14019Human Coxsackievirus A9VR-1311
Human Adenovirus 2Type 2; Species CHuman Coxsackievirus B1VR-687
Human Adenovirus 14VR-15Human Enterovirus 69Tolouca-1
Human Adenovirus 40DuganHuman RotavirusRV4
Human Adenovirus 41TakFeline calicivirusVR-782

g. Interfering Substances:

Interference testing was performed in the presence of chemical and biological substances introduced directly into contrived Stx1 and Stx2 low positive and negative cultured broth samples. Substances and their respective test concentrations evaluated are listed below. Interference was not observed with the Curian Shiga Toxin assay for any of the substances evaluated at their respective test concentrations.

Substance(active ingredient(s))Test Concentration
Barium Sulfate5% w/v
Ciprofloxacin0.25% w/v (2.500 mg/mL)
Hog gastric mucin3.5% w/v (35 mg/mL)
Human blood (whole)40% v/v
Human hemoglobin10.0% w/v (100 mg/mL)
Human urine5% v/v
Imodium AD(Loperamide HCl, 1 mg/7.5 mL)5% v/v
Kaopectate(Bismuth subsalicylate 262 mg/15 mL)5% v/v
Leukocytes0.05% v/v

{9}------------------------------------------------

Mylanta (per 10 mL: Aluminum hydroxide 800 mg, Magnesiumhydroxide 800 mg, Simethicone 80 mg)5% v/v (8.400 mg/mL)
Palmitic acid (fecal fat)40% w/v (400 mg/mL)
Pepto-bismol(Bismuth subsalicylate 525 mg/30 mL)5% v/v
Prilosec OTC (Omeprazole 20 mg/tablet)5 µg/mL
Stearic acid (fecal fat)40% w/v (400 mg/mL)
Tagamet (Cimetidine 200 mg/tablet)5 µg/mL
TUMS50 µg/mL
Naproxen sodium0.5% w/v (5 mg/mL)
Metronidazole0.25% w/v (2.500 mg/mL)
Vancomycin0.25% w/v (2.500 mg/mL)

Assav Reactivity/ Inclusivity: h.

Various Shiga toxin-producing Escherichia coli strains (STEC) were evaluated in the Curian Shiga Toxin assay by the Sorbitol MacConkey Agar (SMAC) plate, Gram Negative (GN) broth, and MacConkey (MAC) broth culture enrichment methods. Each strain was a clinical isolate tested by an FDA-cleared commercial assay and a cytotoxin assay to confirm the presence of the Shiga toxin gene(s). All strains representing various serotypes and toxin combinations tested showed reactivity with the Curian Shiga Toxin assay, detailed below:

STX1 Type Strains: 026:H11 (6), 0111:H8 (3), 045:H2 (3), O103:H25 (2), O145 (2), O103:H11 (2), O157:H7, O103:H2, O111

STX2 Type Strains: 0121:H19 (4), O145 (3), O104:H21 (2), O113:H21 (2), O157:H7, O157:NM, 0145:H25, O145:H28, O91:H21

STX1+2 Type Strains: 0111:H8 (3), 0157:H7 (2), 0111 (2), 0145, 0113:H21, 026:H11

  • i. Assay Cut-Off:
    Not Applicable

2. Comparison Studies

  • a. Method Comparison with Predicate Device:
    See Clinical Studies Section below.

  • Matrix Comparison: b. Not applicable.

3. Clinical Performance

  • Clinical Sensitivity: a.

Prospective:

The Curian Shiga Toxin assay was evaluated from September 2021 to June 2022 at five clinical study sites representing geographically distinct regions throughout the United States. There were 1,627 stool specimens from patients suspected of having a Shiga toxin-producing Escherichia coli (STEC) infection for whom a diagnostic E. coli test had been ordered by a practicing physician. Specimens were prospectively collected and tested with the Curian Shiga Toxin assay using stool specimens inoculated into appropriate broth. Of those 1,627, evaluable reference data was available for 1,538, all of which were evaluable prospective specimens. All specimens were tested at the study sites with the Curian Shiga Toxin assay and in a central laboratory with the reference method, the Vero cell Cytotoxin Assay (with neutralization) performed on the broth culture obtained from the stool specimen. Clinical performance (sensitivity and specificity) for prospective specimens against the reference method (Vero cell Cytotoxin Assay) for both Stx2 are

{10}------------------------------------------------

presented in the following tables. There were no observable differences in performance of the Curian Shiga Toxin assay with respect to study site, broth type, kit lot, or patient gender or age.

Reference Method:Vero cell Cytotoxin Assay
Stx1PositiveStx1 NegativeTotalParameterEstimate95% CI
Curian Shiga Toxin AssayStx1Positive5914Sensitivity100.0%[56.6% - 100.0%]
Stx1Negative015241524Specificity99.4%[98.9% - 99.7%]
Total515331538

Curian Shiga Toxin Overall Performance for Prospective Specimens versus Vero Cell Cytotoxin Assay

Reference Method:Vero cell Cytotoxin Assay
Stx2PositiveStx2NegativeTotalParameterEstimate95% CI
Curian Shiga Toxin AssayStx2Positive4711Sensitivity100.0%[51.0% - 100.0%]
Stx2Negative015271527Specificity99.5%[99.1% - 99.8%]
Total415341538

Archived:

To further estimate sensitivity and specificity of the Curian Shiga Toxin assay, 140 archived stool samples were retrospectively tested for Stx1 and Stx2 using the Curian Shiga Toxin assay at all five study sites. The clinical performance (sensitivity and specificity) for archived samples against the reference method (Vero cell Cytotoxin Assay) are presented in the table below. Of the 140 eligible samples enrolled, 5 were excluded due to inconclusive Vero cell Cytotoxin results, leaving a total of 135 evaluable archived specimens. There were no observable differences in performance of the Curian Shiga Toxin assay with respect to study site, broth type, kit lot, or patient gender or age.

Curian Shiga Toxin Overall Performance for Archived Samples Vero Cell Cytotoxin Assay
Reference Method:Vero cell Cytotoxin Assay
Stx1PositiveStx1NegativeTotalParameterEstimate95% CI
Curian Shiga Toxin AssayStx1Positive46248Sensitivity100.0%[92.3% - 100.0%]
Stx1Negative08787Specificity97.8%[92.2% - 99.4%]
Total4689135
Reference Method:Vero cell Cytotoxin Assay
Stx2PositiveStx2NegativeTotalParameterEstimate95% Cl
Curian Shiga Toxin AssayStx2Positive32234Sensitivity97.0%[84.7% - 99.5%]
Stx2Negative1100101Specificity98.0%[93.1% - 99.5%]
Total33102135

b. Clinical Specificity:

See section L.3.a. above.

{11}------------------------------------------------

  • c. Other clinical supportive data (when a. and b. are not applicable): Not applicable.

4. Clinical Cut-off:

Not applicable.

5. Expected values/Reference Range:

Not applicable.

M. Proposed Labeling:

The labeling is sufficient, and it satisfies the requirements of 21 CFR Part 809.10.

N. Conclusion:

The Curian® Shiga Toxin assay, as supported by the information submitted in this premarket submission, is substantially equivalent to the predicate device (SHIGA TOXIN QUIK CHEK; K121364).

§ 866.3255

Escherichia coli serological reagents.(a)
Identification. Escherichia coli serological reagents are devices that consist of antigens and antisera used in serological tests to identifyEscherichia coli from cultured isolates derived from clinical specimens. Additionally, some of these reagents consist ofEscherichia coli antisera conjugated with a fluorescent dye used to identifyEscherichia coli directly from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium belonging to the genusEscherichia, and provides epidemiological information on diseases caused by this microorganism. AlthoughEscherichia coli constitutes the greater part of the microorganisms found in the intestinal tract in humans and is usually nonpathogenic, those strains which are pathogenic may cause urinary tract infections or epidemic diarrheal disease, especially in children.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9.