K Number
K222773
Device Name
Feeding Tube
Date Cleared
2023-08-17

(337 days)

Product Code
Regulation Number
876.5980
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This product is intended for use in neonatal and pediatric patients to provide nutrition via nasal or orogastric placement. The Polyurethane Feeding Tube is not intended for use beyond 30 days. The PVC Feeding Tube is not intended for use beyond 24 hours.

Device Description

The Feeding Tube consists of the following main components: a feeding tube single lumen catheter and an enteral only connector hub with integral tethered connection closure plug. The catheter tubing is made of Polyurethane or PVC. The catheter tubing has an orange or purple radiopaque stripe of barium sulfate embedded in the tubing wall which can be visualized on x-ray, for exact placement of the tip. The single lumen catheter tubing has side holes for better flow and to provide multiple openings for aspiration. The proposed Feeding Tube is sterilized by Ethylene Oxide Gas to achieve a SAL of 10-6 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of three years.

AI/ML Overview

This document is a 510(k) premarket notification decision letter from the FDA to Anhui Tiankang Medical Technology Co., Ltd. regarding their "Feeding Tube" device. It primarily focuses on the device's substantial equivalence to a predicate device and outlines the non-clinical performance data provided to support this claim.

Based on the provided text, the device is a medical feeding tube, and the "acceptance criteria" and "proof" relate to demonstrating its safety and effectiveness through non-clinical performance testing and biocompatibility testing, rather than a study involving human subjects or an AI algorithm. Thus, many of the typical elements of AI/ML device acceptance criteria (e.g., sample size for test set, expert adjudication, MRMC studies, standalone algorithm performance, training set details) are not applicable here as this is a traditional medical device submission.

Here's an analysis of the acceptance criteria and the study (non-clinical testing) that proves the device meets them, based only on the information provided:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are primarily defined by adherence to recognized international standards and successful completion of specific physical and biological tests. The "reported device performance" indicates that the subject device met these standards and test requirements.

Acceptance Criteria (Test/Standard)Reported Device Performance (as stated in the document)
Biocompatibility Testing
In Vitro Cytotoxicity Test (ISO 10993-5:2009)Performed, result not explicitly stated but implied to be acceptable as "does not show any adverse effect" for the material difference.
Intracutaneous Reactivity Test (ISO 10993-10:2010)Performed, result not explicitly stated but implied to be acceptable as "does not show any adverse effect" for the material difference.
Skin Sensitization Test (ISO 10993-10:2010)Performed, result not explicitly stated but implied to be acceptable as "does not show any adverse effect" for the material difference.
Acute Systemic Toxicity (ISO 10993-11:2017)Performed, result not explicitly stated but implied to be acceptable as "does not show any adverse effect" for the material difference.
Subacute Toxicity (ISO 10993-11:2017)Performed, result not explicitly stated but implied to be acceptable as "does not show any adverse effect" for the material difference.
Implantation Effects (ISO 10993-6:2017)Performed, result not explicitly stated but implied to be acceptable as "does not show any adverse effect" for the material difference.
Sterilization & Shelf Life Testing
Sterilization Validation (ISO 11135)Validated to achieve a SAL of 10-6.
EO Residue (ISO 10993-7:2008)Performed, result not explicitly stated but compliant.
ECH Residue (ISO 10993-7:2008)Performed, result not explicitly stated but compliant.
Ageing Test (ASTM F1980-16 for 3-year shelf life)Shelf life of three years determined based on stability studies.
Package Integrity (ASTM F88/F88M-15, ASTM F1929-2015, ASTM F1886/F1886M-16)All packaging deemed acceptable for protection of product and sterility maintenance after testing (Seal strength, Blue Dye Penetration, Visual Inspection).
Performance Testing (Mechanical/Physical)
Radiopacity Verification (ISO 20695:2020)Compliant with ISO 20695:2020.
Tube Markings (ISO 20695:2020)Compliant with ISO 20695:2020.
Liquid Leakage Testing (ISO 20695:2020, 50-60 kPa internal pressure)"Feeding tube set shall not show signs of leakage sufficient to form a falling drop of water" - stated as tested and met the updated standard.
Fluid Leakage - Connector (ISO 80369-3 and ISO 80369-20, 300-330 kPa)"The connector shall not leak" at applied pressure - stated as tested and met the standards.
Stress Cracking - Connector (ISO 80369-3 and ISO 80369-20)"The connector shall have adequate resistance to stress cracking" - stated as tested and met the standards.
Resistance to Separation from Axial Load - Connector (ISO 80369-3 and ISO 80369-20)"The connector shall not separate from the reference connector" - stated as tested and met the standards.
Resistance to Separation from Unscrewing - Connector (ISO 80369-3 and ISO 80369-20, <0.26 N.m)"The connector shall separate from the reference connector applied unscrewing torque less than 0.26 N.m" - stated as tested and met the standards.
Resistance to Overriding - Connector (ISO 80369-3 and ISO 80369-20, 0.15-0.17 N.m)"The connector shall not override the threads or lugs of the reference connector" - stated as tested and met the standards.
Connector Compatibility (ISO 80369-3 reference connectors)"The connector shall be compatible with ISO 80369-3 reference connectors" - stated as evaluated per ISO 80369-3 for ENFit dimensional verification. "Non-ENFit Connector" also tested and results showed it "does not present the risks contemplated by ISO 80369-1".
Tensile Testing (ISO 20695:2020)Tested and met the updated standard.

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes (number of units tested) for each non-clinical performance and biocompatibility test. It simply states that the tests were "performed" according to the relevant standards.
Data provenance is not directly addressed in terms of "country of origin for data" or "retrospective/prospective" as these are non-clinical, in-vitro/bench tests, not clinical studies. The testing would have been conducted by the manufacturer or a contracted lab to demonstrate compliance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. Ground truth, in the context of device performance, is established by adherence to codified international standards and test methods (e.g., ISO, ASTM). These tests are objective measurements, not subjective evaluations requiring expert consensus from human readers or assessors.

4. Adjudication method for the test set

Not applicable. As described above, these are objective physical and biological tests, not human assessments requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a physical medical device (feeding tube), not an AI/ML diagnostic or assistive device for human readers. No MRMC study was performed or is relevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device does not involve an algorithm.

7. The type of ground truth used

The "ground truth" for this device's acceptance is established by international standards and validated test methodologies. For example, a successful biocompatibility test result (per ISO 10993) is the "ground truth" that the material is biocompatible. Similarly, meeting specified force or leakage requirements (per ISO 80369-3, ISO 20695) constitutes the "ground truth" for mechanical performance.

8. The sample size for the training set

Not applicable. This device is a physical medical device and does not involve AI/ML requiring a training set.

9. How the ground truth for the training set was established

Not applicable. As above, no training set is relevant for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.

August 17, 2023

Anhui Tiankang Medical Technology Co., Ltd. Zhang Yong Management No. 228, Weiyi Road, Economic Development Zone Tianchang, Anhui 239300 China

Re: K222773

Trade/Device Name: Feeding Tube Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal Tube And Accessories Regulatory Class: Class II Product Code: PIF Dated: December 5, 2022 Received: December 6, 2022

Dear Zhang Yong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Sivakami Venkatachalam -S

  • for
    Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222773

Device Name Feeding Tube

Indications for Use (Describe)

This product is intended for use in neonatal and pediatric patients to provide nutrition via nasal or orogastric placement. The Polyurethane Feeding Tube is not intended for use beyond 30 days. The PVC Feeding Tube is not intended for use beyond 24h

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) summary

l Submitter

Device submitter: Anhui Tiankang Medical Technology Co., Ltd. No.228, Weiyi Road, Economic Development Zone, Tianchang City, Anhui, China.

Contact person: Name: Zhang yong Title: Management representative Phone: +86-13705505106 Fax: +86-550-7309158 E-mail: zy@tkmedical.com

II Device

Trade Name of Device: Feeding Tube Common Name: Gastrointestinal tubes with enteral specific connectors Regulation Name: Gastrointestinal tube and accessories Regulatory Class: Class II Product code: PIF Review Panel: Gastroenterology/Urology

III Predicate Device 1

Trade name:Pediatric Nasogastric Feeding Tubes - Single ENFit Port
Regulation name:Gastrointestinal tube and accessories
Classification:Class II
Product Code:PIF
Premarket Notification:K210598
Manufacturer:Cair Lgl

IV Device description

The Feeding Tube consists of the following main components: a feeding tube single lumen catheter and an enteral only connector hub with integral tethered connection closure plug. The catheter tubing is made of Polyurethane or PVC. The catheter tubing has an orange or purple radiopaque stripe of barium sulfate embedded in the tubing wall which can be visualized on x-ray, for exact placement of the tip. The single lumen catheter tubing has side holes for better flow and to provide multiple openings for aspiration.

The proposed Feeding Tube is sterilized by Ethylene Oxide Gas to achieve a SAL of 10-6 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of three years.

{4}------------------------------------------------

V Indications for use

This product is intended for use in neonatal and pediatric patients to provide nutrition via nasal or orogastric placement.The Polyurethane Feeding Tube is not intended for use beyond 30 days. The PVC Feeding Tube is not intended for use beyond 24 hours.

VI Comparison of technological characteristics with the predicate devices

The Feeding Tube are substantially equivalent to the currently marketed predicate Pediatric Nasogastric Feeding Tubes - Single ENFit Port. Table 6-1 is a detailed comparison of the Feeding Tube to the predicate devices regarding substantial equivalence.

Device featureSubject DevicePredicate Device K210598Comment
ProductFeeding TubePediatric Nasogastric FeedingTubes - Single ENFit Port
Product codePIFPIFSame
RegulationNumber21CFR 876.598021CFR 876.5980Same
IndicationsforuseThis product is intended foruse in neonatal and pediatricpatients to provide nutritionvia nasal or orogastricplacement. The PolyurethaneFeeding Tube is not intendedfor use beyond 30 days. ThePVC Feeding Tube is notintended for use beyond 24hours.The Pediatric NasogastricFeeding Tubes - Single ENFitPort are intended for enteralfeeding to deliver nutrition,fluids, and medications to thepatient from an enteral feedingsyringe or feeding set designedwith ENFit connectors forenteral applications.This product is single use for nolonger than 29 days.DifferenceComment 1
Environment ofUseUnspecified -PrescriptionOnlyUnspecified -Prescription OnlySame
Intended UsersTrained personnel wellversed in anatomicallandmarks, safe technique,and potential complicationsmust perform the procedure.Neonatal and PediatricpatientsPhysicians, nurses, andtrained clinicians (by facilitypolicy)Same
PatientPopulationNeonatal and Pediatric patients
Single UseSingle UseSingle UseSame
Device featureSubject DevicePredicate Device K210598Comment
ConnectorCompliant with ISO 80369-3Compliant with ISO 80369-3Same
RadiopacityVerificationCompliant with ISO20695:2020Compliant with ISO 20695:2020Same
Tube MarkingsCompliant with ISO20695:2020Compliant with ISO 20695:2020Same
French Sizes4,5,6.5,8,10,124, 5, 6, 8, 10, 12DifferenceComment 2
Lengths40cm, 60cm, 90cm, 105cm,109cm40cm, 60cm, 90cmDifferenceComment 3
Liquid LeakageTestingFeeding tube set shall notshow signs of leakagesufficient to form a fallingdrop of water while beingsubjected to the internallyapplied pressure of 50 kPa to60 kPa.Tested and met updatedstandard ISO 20695:2020Enteral Feeding Systems -Design and TestingSame
Component andMaterialHead Melted Tube: TPU/PVCHub: TPU/ABSTethered Bolt: PVCTube: PolyurethaneCap: PolypropyleneConnector: ABSTether: thermoplasticpolyurethaneDifferenceComment 4
Tensile TestingTested and met updatedstandard ISO 20695:2020Enteral Feeding Systems -Design and TestingTested and met updatedstandard ISO 20695:2020Enteral Feeding Systems -Design and TestingSame
Fluid Leakage:ConnectorThe connector shall not leakapplied pressure of between300 kPa and 330 kPa perISO 80369-3 and ISO80369-20.Tested per ISO 80369-20 andmet the standards of 80369-3for fluid leakage.Same
StressCracking:ConnectorThe connector shall haveadequate resistance to stresscracking per ISO 80369-380369-20.Tested per ISO 80369-20 andmet the standards of ISO80369-3 for stress cracking.Same
Resistance toseparation fromaxial load:connectorThe connector shall notseparate from the referenceconnector applied axial forceper ISO 80369-3 and ISO80369-20.Tested per ISO 80369- 20 andmet the standards of ISO80369-3 for resistance toseparation from axial load.Same
Device featureSubject DevicePredicate Device K210598Comment
Resistance toseparation fromunscrewing:connectorThe connector shall separatefrom the reference connectorapplied unscrewing torqueless than 0.26 N.m per ISO80369-3 and ISO 80369-20.Tested per ISO 80369-20 andmet the standards of ISO80369-3 for separation fromunscrewing.Same
Resistance tooverriding:connectorThe connector shall notoverride the threads or lugsof the reference connectorapplied torque of between0.15 N.m to 0.17 N.m perISO 80369-3 and ISO80369-20.Tested per ISO 80369-20 andmet the standards of ISO80369-3 for resistance tooverriding.Same
ConnectorThe connector shall becompatible with ISO 80369-3reference connectors.Evaluated per ISO 80369-3 forENFit dimensional verificationSame
ConnectorNon- ENFit ConnectorNoneDifferenceComment 5
BiocompatibilityCompliant with Use ofInternational Standard ISO10993-1, “Biologicalevaluation of medical devices– Part 1: Evaluation andtesting within a riskmanagement process”Compliant with Use ofInternational Standard ISO10993-1, “Biological evaluationof medical devices – Part 1:Evaluation and testing within arisk management process”Same
SterilityConditionSterileSterileSame
Sterile MethodEO SterilizedEO SterilizedSame

Table 6-1 Substantial equivalence discussion for Feeding Tube

{5}------------------------------------------------

{6}------------------------------------------------

Difference comment 1

The difference between subject device and predicate device is only in textual expressions, and the actual indications for use are not different, which will not affect the safety and performance of the device.

Difference comment 2

The french sizes of predicate device are covered by subject device, so this difference will not affect the safety and performance of the device

Difference comment 3

{7}------------------------------------------------

The lengths of 105cm and 109cm of the predicate device are not covered by the predicate device. The operator will select the corresponding length of the device according to the clinical needs. At the same time, the device has passed the performance test, so the difference will not affect the safety and performance of the device.

Difference comment 4

subject device is different from predicate device in material, the biocompatibility test has been performed on subject device according to ISO 10993, the result does not show any adverse effect which can demonstrate the safety of proposed device. Therefore, this difference does not raise new safety and effectiveness issues for the device.

Difference Comment 5

The predicate device does not contain a Non-ENFit Connector, and since the subject device has been tested by the connector and the results show that it does not present the risks contemplated by ISO 80369-1, this does not affect the safety and effectiveness of the product device

VII Performance data

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

Biocompatibility of the Feeding Tube was evaluated in accordance with ISO 10993-1:2018 for the body contact category of "Surface medical device-mucosal membrane" with a contact duration of "prolonged ( > 24h to 30d)". The following tests were performed, as recommended:

In Vitro Cytotoxicity TestISO 10993-5: 2009
Intracutaneous Reactivity TestISO 10993-10: 2010
Skin Sensitization TestISO 10993-10: 2010
Acute systemic toxicityISO 10993-11:2017
Subacute toxicityISO 10993-11:2017
Implantation effectsISO 10993-6:2017

Sterilization and shelf life testing

The sterilization method has been validated to ISO11135, which has thereby determined the routine control and monitoring parameters.

The testing is performed according to the following standards:

EO residueISO 10993-7:2008
ECH residueISO 10993-7:2008

The shelf life of three years is determined based on stability studies which include ageing test according to FDA recognized standard ASTM F1980-16.

Package integrity testing was conducted on the final, packaged, and sterile devices affer

{8}------------------------------------------------

environmental conditioning and simulated transportation. All packaging deemed acceptable for protection of product and sterility maintenance.

The testing is performed according to the following standards:

Seal strengthASTM F88/F88M-15
Blue Dye PenetrationASTM F 1929-2015
Seal Integrity (Visual Inspection)ASTM F 1886/ F 1886M-16

Performance testing

Performance testing is performed according to the following standards:

  • ISO 80369-3:2016: Small-bore connectors for liquids and gases in healthcare △ applications - Part 3: Connectors for enteral applications
  • ISO 80369-20:2015: Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods

  • A ISO 20695 Enteral feeding systems - Design and testing

XI Clinical Test Conclusion

No clinical study is included in this submission.

X Conclusion

The Feeding Tube is substantially equivalent to its predicate device. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.