K Number
K222556

Validate with FDA (Live)

Device Name
BTL-785X
Date Cleared
2023-05-31

(281 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BTL-785X device has the following indications for use:

The BTL-785X with BTL-785-1 applicator is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The BTL-785-1-4 massage device is intended to provide a temporary reduction in the appearance of cellulite.

The BTL-785X with BTL-785-2 applicator is indicated to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

The BTL-785X with BTL-785-3 applicator is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

The BTL-785-4 with BTL-785-4 applicator used with tips BTL-785-4-2, BTL-785-4-5 and BTL-785-4-6 is indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis. The Applicator BTL-785-4 of BTL-785X device used with tips BTL-785-4-4, BTL-785-4-4, BTL-785-4-7 and BTL-785-4-8 is intended for dermatological procedures requiring fractional treatment of the skin. At higher energy levels greater than 62 mJ/pin, use of the BTL-785-4 applicator is limited to Skin Types I-IV.

The BTL-785X with BTL-785-7 hands-free applicator used with forehead BTL-785-7-1 and cheek BTL-785-7-2 singleuse electrodes is intended to provide:

  • heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

  • non-invasive temporary reduction of facial wrinkles

Device Description

The BTL-785X is a state-of-the-art radiofrequency device that enables the application of therapy by a high-frequency field.

The control unit of the system is equipped with a large color touch screen that significantly facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen of the device. During the therapy the device displays information about the applied therapy type, remaining therapy time and main therapy parameters on the screen. For easier control, the hand-pieces are equipped with buttons, enabling operation of the device during therapy. The energy flow's quality is indicated by the illuminated treatment tip. The BTL-785X device comes with five different types of applicators.

The BTL-785X device consists of the following main components:

  • microprocessor-driven control unit
  • radiofrequency generator
  • user interface with 15.6" color touch screen
  • applicators for an application of radiofrequency
  • exchangeable applicator tips
AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets those criteria, based on the provided document:

Acceptance Criteria and Device Performance

The core acceptance criterion for the new indication (non-invasive temporary reduction of facial wrinkles) of the BTL-785X with the BTL-785-7 hands-free applicator was based on a statistically significant reduction in wrinkle severity.

Acceptance Criterion (Primary Endpoint)Reported Device Performance (Primary Endpoint Result)
A statistically significant reduction of at least 1.0 score in the average score of Fitzpatrick Wrinkle and Elastosis Scale (FWES) score according to at least 2 out of 3 blinded dermatologists at 3-months follow-up compared to baseline.The average FWES score of the subjects showed an overall improvement of 1.44 points (p<0.001) at the 3-month follow-up, with three of the three blinded evaluators finding a statistically significant reduction of at least 1.0 score.

Secondary Acceptance Criteria and Device Performance:

Acceptance Criterion (Secondary Endpoints)Reported Device Performance (Secondary Endpoints Result)
Evaluation of the safety of the BTL-785 device with BTL-785-7 applicator for non-invasive reduction of wrinkles.No adverse events were recorded by the investigator during this clinical investigation, neither during treatments nor in the follow-up period.
The majority of the treated subjects to report satisfaction (level satisfied and higher) with the therapy.All forty-two participants (100%) were either satisfied (N=14, 33.3%) or very satisfied (N=28, 66.7%) at the 3-month follow-up. Furthermore, 100% of the subjects found the procedure comfortable.

Study Details

2. Sample size used for the test set and the data provenance:

  • Sample Size: N = 42 subjects completed all study treatments and follow-up visits.
  • Data Provenance: The study was conducted at a "single site." The document does not specify the country of origin, but given the applicant and sponsor are based in Marlborough, Massachusetts, USA, it is likely a US-based study. The study design is classified as "interventional," indicating prospective data collection.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Number of Experts: 3 blinded dermatologists.
  • Qualifications: The document identifies them as "blinded dermatologists." Specific years of experience or other additional qualifications are not provided.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • The primary efficacy outcome measure required at least 2 out of 3 blinded dermatologists to agree on a statistically significant reduction of at least 1.0 score in the FWES scale. This indicates an adjudication method, where a simple majority (2 out of 3) determined the positive outcome.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC comparative effectiveness study involving AI assistance for human readers was not conducted. This study focused on the standalone performance of the device in reducing wrinkles, with dermatologists assessing the outcome.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • This question is not applicable in the context of the provided document. The device is a radiofrequency system with applicators for therapeutic use, not an AI-driven algorithm for image analysis or diagnosis. The performance related to "non-invasive temporary reduction of facial wrinkles" was evaluated in human subjects.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Expert Consensus: The primary ground truth for wrinkle reduction was established by expert consensus, specifically the assessment of 3 blinded dermatologists using the Fitzpatrick Wrinkle and Elastosis Scale (FWES).
  • Outcomes Data: Patient satisfaction and adverse events were also collected as outcomes data.

8. The sample size for the training set:

  • The document describes a clinical study to demonstrate the performance, efficacy, and safety of the device for a new indication. This is a validation study (test set), not a training set for an algorithm. Therefore, there is no information about a separate training set.

9. How the ground truth for the training set was established:

  • As mentioned above, there is no description of a training set for an algorithm in this document, so this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

May 31, 2023

BTL Industries Inc. David Chmel VP of Operations 362 Elm Street Marlborough, Massachusetts 01752

Re: K222556

Trade/Device Name: Btl-785x Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: PBX, GEI Dated: April 12, 2023 Received: May 3, 2023

Dear David Chmel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Figure/6 description: The image shows a digital signature. The signature is for Mark Trumbore -S. The date of the signature is 2023.05.31, and the time is 13:44:10 -04'00'.

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222556

Device Name BTL-785X

Indications for Use (Describe)

The BTL-785X device has the following indications for use:

The BTL-785X with BTL-785-1 applicator is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The BTL-785-1-4 massage device is intended to provide a temporary reduction in the appearance of cellulite.

The BTL-785X with BTL-785-2 applicator is indicated to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

The BTL-785X with BTL-785-3 applicator is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

The BTL-785-4 with BTL-785-4 applicator used with tips BTL-785-4-2, BTL-785-4-5 and BTL-785-4-6 is indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis. The Applicator BTL-785-4 of BTL-785X device used with tips BTL-785-4-4, BTL-785-4-4, BTL-785-4-7 and BTL-785-4-8 is intended for dermatological procedures requiring fractional treatment of the skin. At higher energy levels greater than 62 mJ/pin, use of the BTL-785-4 applicator is limited to Skin Types I-IV.

The BTL-785X with BTL-785-7 hands-free applicator used with forehead BTL-785-7-1 and cheek BTL-785-7-2 singleuse electrodes is intended to provide:

  • heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

  • non-invasive temporary reduction of facial wrinkles

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three nested squares on the left side, with the letters "BTL" inside the innermost square. To the right of the squares is the text "BTL Industries" in a simple, sans-serif font. The logo is clean and modern, with a focus on geometric shapes and clear typography.

510(k) Summary

K222556

General Information

Classification Product Code:

Sponsor:BTL Industries, Inc.362 Elm StreetMarlborough, MA 01752Tel: +1-866-285-1656Fax: +1-888-499-2502
Applicant:BTL Industries, Inc.362 Elm StreetMarlborough, MA 01752Tel: +1-866-285-1656Fax: +1-888-499-2502
Contact Person:David ChmelBTL Industries, Inc.chmel@btlnet.com
Summary PreparationDate:May 29, 2023
Device
Trade/Proprietary Name:BTL-785X
Primary Classification Name:Electrosurgical cutting and coagulation device and accessories
Classification Regulation:21 CFR 878.4400, Class II

GEI, PBX

{5}------------------------------------------------

Legally Marketed Predicate Device

The BTL-785X is a state-of-the-art radiofrequency device with accessories, and is substantially equivalent to the following products that are already cleared for distribution in the USA under the following 510(k) Premarket Notification numbers:

Product Description

The BTL-785X is a state-of-the-art radiofrequency device that enables the application of therapy by a high-frequency field.

The control unit of the system is equipped with a large color touch screen that significantly facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen of the device. During the therapy the device displays information about the applied therapy type, remaining therapy time and main therapy parameters on the screen. For easier control, the hand-pieces are equipped with buttons, enabling operation of the device during therapy. The energy flow's quality is indicated by the illuminated treatment tip. The BTL-785X device comes with five different types of applicators.

The BTL-785X device consists of the following main components:

  • · microprocessor-driven control unit
  • · radiofrequency generator
  • user interface with 15.6" color touch screen
  • applicators for an application of radiofrequency
  • · exchangeable applicator tips

Technological characteristics

The BTL-785X device has identical technological characteristics compared to its predicate device. The BTL-785X device and the predicate are comprised of a system console and applicators.

The system console consists of the RF generator, computer, and a touch-screen control panel. The device is accompanied by the following applicators:

  • . BTL-785-1 applicator providing treatment by integration of radiofrequency, ultrasound and active cooling. Suitable for the treatment of large body areas.
  • BTL-785-2 applicator providing treatment by integration of radiofrequency and ultrasound. . Suitable for the treatment of small areas.

{6}------------------------------------------------

  • BTL-785-3 applicator providing radiofrequency treatment. The therapy is provided with single . use tips only.
  • BTL-785-4 applicator delivering radiofrequency via an array of microneedles and/or superficial pins. Therapy is provided with single use tips only.
  • . BTL-785-7 hands-free applicator providing treatment by integration of radiofrequency heating and muscle stimulation resulting in induced muscle workout naturally increases local blood circulation. Suitable for the treatment of small and sensitive areas. The therapy is provided with single use electrodes only.

The compared devices' applicators use the RF signal of the same monopolar mode of operation, waveform and frequency. The devices have the same properties regarding their RF tips, including material, size, biocompatibility and sterilization method where applicable.

Indications for Use

The BTL-785X device has the following indications for use:

The BTL-785X with BTL-785-1 applicator is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The BTL-785-1-4 massage device is intended to provide a temporary reduction in the appearance of cellulite.

The BTL-785X with BTL-785-2 applicator is indicated to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

The BTL-785X with BTL-785-3 applicator is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

The BTL-785X with BTL-785-4 applicator used with tips BTL-785-4-1, BTL-785-4-2, BTL-785-4-5 and BTL-785-4-6 is indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis.

The Applicator BTL-785-4 of BTL-785X device used with tips BTL-785-4-3, BTL-785-4-4, BTL-785-4-7 and BTL-785-4-8 is intended for dermatological procedures requiring fractional treatment of the skin.

At higher energy levels greater than 62 mJ/pin, use of the BTL-785-4 applicator is limited to Skin Types I-IV.

{7}------------------------------------------------

The BTL-785X with BTL-785-7 hands-free applicator used with BTL-785-7-1 and BTL-785-7-2, singleuse electrodes is intended to provide:

  • heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.
  • . non-invasive temporary reduction of facial wrinkles.

Performance Data

The BTL-785X device has been thoroughly evaluated for electrical safety. The device has been found to comply with applicable medical device safety standards:

IEC 60601-1Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests
IEC 60601-2-2Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgery equipment and high frequency surgical accessories
IEC 60601-2-5Medical electrical equipment - Part 2-5: Particular requirements for the basic safety and essential performance of ultrasonic physiotherapy equipment
IEC 60601-1-6Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability
IEC 62304Medical device software - Software life cycle processes
ISO 14971Medical devices - Application of risk management to medical devices
ISO 10993-1Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
ISO 10993-5Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
ISO 10993-7Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
ISO 10993-10Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
ISO 10993-11Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity
ISO 11135Sterilization of health-care products -- Ethylene oxide -- Requirements for the development, validation and routine control of a sterilization process for medical devices

{8}------------------------------------------------

ISO 11607-1Packaging for terminally sterilized medical devices - Part 1: Requirements for
materials, sterile barrier systems and packaging systems
  • Packaging for terminally sterilized medical devices Part 2: Validation ISO 11607-2 requirements for forming, sealing and assembly processes

BTL-785-7 applicator clinical performance data

A clinical study was conducted to demonstrate the performance, clinical efficacy and safety of the BTL-785-7 hands-free applicator for non-invasive temporary reduction of wrinkles. This study has a single-center single-arm, open-label, interventional design.

The study was conducted at a single site. All the enrolled participants were treated with the BTL-785-7 hands-free applicator.

In total, N=42 subjects respected the study schedule and completed all the required study treatment visits and follow-up visits.

No adverse events occurred throughout the whole clinical investigation.

The study's primary efficacy endpoint was determined at the 3-month follow-up with three of the three blinded evaluators finding a statistically significant reduction of at least 1.0 score in the average score of FWES score at the 3-months follow-up compared to baseline. The average FWES score of the subjects showed an overall improvement of 1.44 points (p<0.001).

The overall satisfaction with the study treatment outcome was high with all forty-two participants (100%) either satisfied (N=14, 33.3%) or very satisfied (N=28, 66.7%) at the 3-month follow-up. None of the patients was dissatisfied. Furthermore, 100% of the subjects found the procedure comfortable.

Based on 3-month follow-ups, the treatment with the BTL-785X device equipped with applicator BTL-785-7 has shown to be both safe and effective for non-invasive temporary treatment of facial wrinkles with high satisfaction levels.

{9}------------------------------------------------

The below table is a summary of the study design and results:

Study designSingle-center single-arm, open-label, interventional design
Sample size42 patients completed all study treatments and follow-up visits.
Number of treatmentsand follow-up visits4 treatments and 2 follow-ups at 1 and 3 months
Primary endpointsTo gather clinical evidence that BTL-785 system equipped with the BTL-785-7 applicator is able to provide reduction of wrinkle severity according to theFitzpatrick Wrinkle and Elastosis Scale.The primary efficacy outcome measure is a statistically significant reductionof at least 1.0 score in the average score of Fitzpatrick Wrinkle and ElastosisScale score according to at least 2 out of 3 blinded dermatologists at 3-months follow-up compared to baseline.
Secondary endpointsEvaluation of the safety of the BTL-785 device with BTL-785-7 applicator fornon-invasive reduction of wrinklesThe majority of the treated subjects to report satisfaction (level satisfied andhigher) with the therapy.
Primary endpoint resultThe average FWES score of the subjects showed an overall improvementof 1.44 points (p<0.001).
Secondary endpointresultsNo adverse events were recorded by the investigator during this clinicalinvestigation neither during treatments, nor in the follow-up period.All forty-two participants (100%) were either satisfied (N=14, 33.3%) or verysatisfied (N=28, 66.7%) at the 3-month follow-up.

{10}------------------------------------------------

Comparison with the Predicate Device

510(k) numberK222556K211639K092191
Device nameBTL-785XBTL-785WExilis 5000
CompanynameBTL Industries, Inc.BTL Industries, Inc.BTL Industries, Inc.
TypeSubject devicePrimary predicateSecondary predicate
Product Codeand RegulationGeneral & Plastic Surgery21 CFR 878.4400GEI - Electrosurgical,Cutting & Coagulation &AccessoriesPBX - Massager,Vacuum, RadiofrequencyInduced HeatGeneral & Plastic Surgery21 CFR 878.4400GEI - Electrosurgical,Cutting & Coagulation &AccessoriesPBX - Massager,Vacuum, RadiofrequencyInduced HeatGeneral & Plastic Surgery21 CFR 878.4400GEI - Electrosurgical,Cutting & Coagulation &AccessoriesPBX - Massager,Vacuum, RadiofrequencyInduced Heat
Indications forUseThe BTL-785X device hasthe following indicationsfor use:The BTL-785X with BTL-785-1 applicator isintended to provideheating for the purpose ofelevating tissuetemperature for selectedmedical conditions suchas temporary relief ofpain, muscle spasms, andincrease in localcirculation. The BTL-785-1-4 massage device isintended to provide aThe BTL-785W devicehas the followingindications for use:The BTL-785W with BTL-785-1 applicator isintended to provideheating for the purpose ofelevating tissuetemperature for selectedmedical conditions suchas temporary relief ofpain, muscle spasms, andincrease in localcirculation. The BTL-785-1-4 massage device isintended to provide aThe EXILIS device isindicated for the primarytreatment of dermatologicand general surgicalprocedures for non-invasive treatment ofwrinkles and rhytids.

{11}------------------------------------------------

temporary reduction in theappearance of cellulite.temporary reduction in theappearance of cellulite.
The BTL-785X with BTL-785-2 applicator isindicated to provideheating for the purpose ofelevating tissuetemperature for selectedmedical conditions suchas temporary relief ofpain, muscle spasms, andincrease in localcirculation.The BTL-785W with BTL-785-2 applicator isindicated to provideheating for the purpose ofelevating tissuetemperature for selectedmedical conditions suchas temporary relief ofpain, muscle spasms, andincrease in localcirculation.
The BTL-785X with BTL-785-3 applicator isintended to provideheating for the purpose ofelevating tissuetemperature for selectedmedical conditions suchas temporary relief ofpain, muscle spasms, andincrease in localcirculation.The BTL-785W with BTL-785-3 applicator isintended to provideheating for the purpose ofelevating tissuetemperature for selectedmedical conditions suchas temporary relief ofpain, muscle spasms, andincrease in localcirculation.
The Applicator BTL-785-4of BTL-785X device usedwith tips BTL-785-4-1,BTL-785-4-2, BTL-785-4-5 and BTL-785-4-6 isindicated for use indermatological andgeneral surgicalprocedures forelectrocoagulation andhemostasis.The Applicator BTL-785-4of BTL-785W device usedwith tips BTL-785-4-1,BTL-785-4-2, BTL-785-4-5 and BTL-785-4-6 isindicated for use indermatological andgeneral surgicalprocedures forelectrocoagulation andhemostasis.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three interconnected blue squares with the letters "BTL" inside them. To the right of the squares is the text "BTL Industries" in a dark gray sans-serif font.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image contains the logo for BTL Industries. The logo consists of a geometric design on the left and the text "BTL Industries" on the right. The geometric design is made up of three squares that are nested inside each other and slightly offset.

Application of the heat tothe tissue via RF energy.Application of the heat tothe tissue via RF energy.Application of the heat tothe tissue via RF energy.
Principle ofActionMassaging of body partswith massage attachment.(BTL-785-1 applicatoronly)Radiofrequencyaccompanied byelectromagneticstimulation (BTL-785-7applicator only)Massaging of body partswith massage attachment.(785-1 applicator only)Radiofrequencyaccompanied byelectromagneticstimulation (785-7applicator only)Application of the heat tothe tissue via RF energy.
Clinical UsePrescription usePrescription usePrescription use
Energy Source100 - 120 V AC,50/60 Hz200 – 240 V AC,50/60 Hz100 - 120 V AC,50/60 Hz200 – 240 V AC,50/60 Hz100-120 V AC, 5 A,50/60 Hz208-240 V AC, 2.5 A,50 Hz
Type of EnergyAppliedElectromagnetic Energy -RadiofrequencyElectromagnetic Energy -RadiofrequencyElectromagneticEnergy -Radiofrequency
Frequency3.2 MHz ± 5%(BTL-785-1, BTL-785-2,and BTL-785-3, BTL-785-7)1 MHz ± 5%(BTL-785-4)3.2 MHz ± 5%(BTL-785-1, BTL-785-2,and BTL-785-3, BTL-785-7)1 MHz ± 5%(BTL-785-4)3.25 Mhz ± 50 kHzN/A
Mode ofOperationMonopolarMonopolarMonopolar
User InterfaceColor Touch-screenColor Touch-screenColor Touch-screen

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of a blue geometric design on the left and the text "BTL Industries" on the right. The geometric design is made up of three squares, each with the letters "BTL" inside.

140 W (BTL-785-1-1)140 W (BTL-785-1-1)125 W (BTL-785-1-2)170W
62 W (BTL-785-2-1)62 W (BTL-785-2-1)53 W (BTL-785-2-2)62 W (BTL-785-2-3)
MaximumOutput Power48 W (BTL-785-3-1)48 W (BTL-785-3-1)
30 W (BTL-785-4-1, 2, 5, 6)30 W (BTL-785-4-1, 2, 5, 6)
25 W (BTL-785-4-4, 8)25 W (BTL-785-4-4, 8)
20 W (BTL-785-4-3, 7)20 W (BTL-785-4-3, 7)
39.8 W (BTL-785-7)120 W (BTL-785-7)
EffectiveTreatmentTemperature40 - 45°C(104 - 113°F)40 - 45°C(104 - 113°F)39 - 42°C(102 - 108°F)
SkinTemperatureMonitoringIntegrated thermometer +patient's feedback(BTL-785-1, 2, 3)Integrated thermometer +patient's feedback(BTL-785-1, 2, 3)Integrated thermometer +Skin temperaturemeasurement
Ultrasonic TipPre-heatingFunctionYes (BTL-785-1, 2)Yes (BTL-785-1, 2)No

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three blue squares nested inside each other, with the letters "BTL" in white inside the innermost square. To the right of the logo is the text "BTL Industries" in a dark gray sans-serif font. The logo is simple and modern, and the colors are clean and professional.

MassageAttachmentYes (BTL-785-1)Yes (BTL-785-1)Yes
Number ofMicroneedles6 x 66 x 6N/A
HandsfreeapplicatorYesYesN/A
Depth ofMicroneedleElectrodes0.5 — 4 mm0.5 — 4 mmN/A
Number of Pinsof SuperficialTips32643264N/A
SterilizationMethodEthylene oxideEthylene oxideN/A
NeutralElectrode Area169 cm²169 cm²118 cm²
System Weight65 kg(143 lb)60 kg(132 lb)15.3 kg(33.7 lb)
SystemDimension(W×H×D)1370 mm x 670 x 670(53.94" x 26.38" x 26.38")1370 mm x 670 x 670(53.94" x 26.38" x 26.38")230 x 390 x 260 mm(9.06" x 15.35" x 10.24")

New Indication for Use for Applicator BTL-785-7

The BTL-785X and its hands-free applicator BTL-785-7 are newly intended for non-invasive temporary reduction of facial wrinkles. This indication is supported by comparison with the primary and secondary predicate device and data from the clinical investigation. The new indication is proposed based on the

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three interlocking squares with the letters "BTL" inside the leftmost square. To the right of the squares is the text "BTL Industries" in a sans-serif font. The logo and text are both in a dark blue color.

clinical trial's positive feedback from the participants, as well as the results achieved in the study. Additionally, the safety of the device has been evaluated during the clinical investigation and no new risks have been identified.

We believe the difference does not raise any new questions of safety or effectiveness.

Substantial Equivalence

The BTL-785X device has the same technological characteristics and similar intended use compared to the primary predicate device and secondary predicate device. Any differences between the predicate devices and BTL-785X device have no significant influence on safety or effectiveness of the BTL-785X device.

Therefore, the BTL-785X device is substantially equivalent to the predicate devices.

Conclusion

Based upon the intended use, comparison of technical characteristics and performance testing provided in this premarket notification, the BTL-785X device has been shown to be substantially equivalent to the currently cleared predicate device and secondary predicate device for requested intended use.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.