(93 days)
The Nitrile Patient Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
The subject device is powder free nitrile examination gloves. The subject device is blue. The subject device is non-sterile.
The provided text describes the acceptance criteria and performance of a Nitrile Patient Examination Glove, not an AI-powered medical device. Therefore, the questions related to AI device performance metrics, such as ground truth, expert consensus, MRMC studies, and training/test sets, are not applicable in this context.
This document is a 510(k) Premarket Notification for a Class I medical device (nitrile examination glove), which establishes substantial equivalence to a predicate device based on physical and performance characteristics, and biocompatibility.
Here's a breakdown of the relevant information provided:
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria are primarily based on established ASTM and ISO standards for medical gloves. The reported device performance is compared against these standards and the predicate device.
| Test / Characteristic | Acceptance Criteria (Standard Reference) | Reported Device Performance |
|---|---|---|
| Biocompatibility | ||
| Irritation (ISO 10993-10:2010) | Primary Irritation Index: 0-0.4 (Negligible) | Primary irritation index is 0. Response categorized as negligible. All animals survived with no abnormal signs. |
| Sensitization (ISO 10993-10:2010) | Grades less than 1 (no sensitization) | All grades are 0. Not a sensitizer under study conditions. |
| Cytotoxicity (ISO 10993-5:2009) | Viability (viab.%) of 100% extract ≥ 70% of blank (no cytotoxic potential) | Viab.% of 100% test article extract is 70.9%. No potential toxicity to L-929 in MTT method. |
| Physical Properties | ||
| Residual Powder (ASTM D6124-06) | <0.25 mg/glove (Table 4) or 2.0 mg (Table 5) | <0.25 mg/glove |
| Freedom from Holes (ASTM D5151-19) | AQL=2.5, criterion ≤ 7 gloves for water leakage (Sample size: 125 gloves) | No glove water leakage found. |
| Dimensions (ASTM D6319-19) | Min Length: S (220mm), M, L, XL (230mm) Width: S (80±10mm), M (95±10mm), L (110±10mm), XL (120±10mm) Min Thickness (Finger/Palm): 0.05mm | Length (mm): S: 297-302, M: 296-302, L: 296-302, XL: 297-302 (All meet min length) Width (mm): S: 84-87, M: 94-97, L: 104-108, XL: 113-117 (All meet ±10mm tolerance) Thickness (mm): Finger: 0.152-0.184, Palm: 0.112-0.131 (All meet min 0.05mm) |
| Physical Properties | ||
| Before Aging: Tensile Strength | ≥ 14MPa | 14.2-29.0 MPa |
| Before Aging: Ultimate Elongation | ≥ 500% | 537.700% - 552.733% |
| After Accelerated Aging: Tensile Strength | ≥ 14MPa | 14.1-26.8 MPa |
| After Accelerated Aging: Ultimate Elongation | ≥ 400% | 421.422% - 552.633% |
2. Sample size used for the test set and the data provenance
- Freedom from Holes (ASTM D5151-19): 125 gloves were tested.
- Other tests (Biocompatibility, Dimensions, Physical Properties, Powder Content): Specific sample sizes for each test are not explicitly stated in the provided text, but testing was conducted to meet the requirements of the referenced ASTM and ISO standards, which define appropriate sample sizes and methodologies.
- Data Provenance: Not specified in terms of country of origin for the test data. The manufacturer is Jiangxi Zhonghong Pulin Medical Products Co.,Ltd. in China. The data would be prospective, as it's generated specifically for this regulatory submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is a medical device (glove), not an AI diagnostic tool that requires expert interpretation of images or other data for ground truth establishment. The "ground truth" here is the physical and chemical properties of the glove measured against established engineering and biological standards.
4. Adjudication method for the test set
Not applicable for a medical glove. Testing relies on standardized methods and laboratory measurements, not human adjudication of results in the way AI model outputs are typically adjudicated.
5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done
Not applicable. MRMC studies are used for diagnostic imaging devices/algorithms to assess how human reader performance is affected by AI assistance. This device is a physical examination glove.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. There is no algorithm to assess.
7. The type of ground truth used
The "ground truth" for this device is based on:
- Standardized Test Methods: Adherence to established international and national standards (e.g., ISO 10993 for biocompatibility, ASTM D6319 for physical properties of gloves, ASTM D5151 for freedom from holes).
- Reference Materials/Controls: For biocompatibility tests, positive and negative controls would be used as part of the standard methodology (ISO 10993).
- Objective Measurements: Laboratory measurements of physical dimensions, tensile strength, elongation, powder content, and visual inspection for holes.
8. The sample size for the training set
Not applicable. This is not an AI/machine learning device that requires a training set.
9. How the ground truth for the training set was established
Not applicable for the same reason as above.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 23, 2022
Jiangxi Zhonghong Pulin Medical Products Co.,Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608,No.738,Shangcheng Rd.,Pudong Shanghai, Shanghai 200120 China
Re: K222534
Trade/Device Name: Nitrile Patient Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: October 25, 2022 Received: October 25, 2022
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Allan Guan -S
For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K222534
Device Name Nitrile Patient Examination Glove
Indications for Use (Describe)
The Nitrile Patient Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary (K222534)
This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.
1.0 Submitter's Information
Name: Jiangxi Zhonghong Pulin Medical Products Co.,Ltd. Address: Yinshawan Park, High-tech Industrial Park, Hukou County, Jiujiang City, Jiangxi Province, China Phone Number: +86-15247135174 Contact: Zhou Ziyu Date of Preparation: 2022.11.21
Designated Submission Correspondent
Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn
2.0 Device Information
Trade name: Nitrile Patient Examination Glove Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): S, M, L, XL
3.0 Classification
Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital
4.0 Predicate Device Information
Manufacturer: Yingxiang Glove Products Co., Ltd. Device: Nitrile Patient Examination Gloves 510(k) number: K211914
5.0 Indication for Use
{4}------------------------------------------------
The Nitrile Patient Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
6.0 Device Description
The subject device is powder free nitrile examination gloves. The subject device is blue. The subject device is non-sterile.
7.0 Technological Characteristic Comparison Table
| Item | Subject Device(K222534) | Predicate Device(K211914) | Remark |
|---|---|---|---|
| Product Code | LZA | LZA | Same |
| Regulation No. | 21CFR880.6250 | 21CFR880.6250 | Same |
| Class | I | I | Same |
| Intended Use | The Nitrile Patient Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner. | The Nitrile Patient Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. | Same |
| Powdered orPowered free | Powdered free | Powdered free | Same |
| Design Feature | Ambidextrous | Ambidextrous | Same |
| LabelingInformation | Single-use indication, powder free, device color, device name, glove size and quantity, Nitrile Glove Powder Free Blue, Non-Sterile | Single-use indication, powder free, device color, device name, glove size and quantity, Disposable Powder Free Nitrile Examination Glove, Non-Sterile | Same |
Table1-General Comparison
Table2 Device Dimensions Comparison
{5}------------------------------------------------
| Designation | Size | Tolerance | |||||
|---|---|---|---|---|---|---|---|
| S | M | L | XL | ||||
| PredicateDevice(K211914) | 9-inch | Length, mm | 220 | 230 | 230 | 230 | min |
| Width, mm | 80 | 95 | 110 | 120 | ±10 | ||
| 12-inch | Length, mm | 220 | 230 | 230 | 230 | min | |
| Width, mm | 80 | 95 | 110 | 120 | ±10 | ||
| Thickness, mm: | |||||||
| 9-inch/12-inch | Finger | 0.05 | min | ||||
| Palm | 0.05 | min | |||||
| SubjectDevice | 12 inch | Length, mm | 220 | 230 | 230 | 230 | min |
| Width, mm | 80 | 95 | 110 | 120 | ±10 | ||
| Thickness, mm: | |||||||
| FingerPalm | 0.05 | min | |||||
| Remark | SIMILAR |
Analysis: The physical dimensions of subject device are same with the 12inch ones of the predicate device, and they all meet the requirements of ASTM D6319-19.
| Item | Subject device(K222534) | Predicate device(K211914) | Remark | |
|---|---|---|---|---|
| Colorant | Blue | Blue | Same | |
| PhysicalProperties | BeforeAgingTensileStrength | 14MPa, min | 14MPa, min | Same |
| BeforeAgingUltimateElongation | 500% min | 500% min | Same | |
| AfterAgingTensileStrength | 14MPa, min | 14MPa, min | Same | |
| AfterAgingUltimateElongation | 400%min | 400%min | Same | |
| Comply with ASTM D6319 | Comply with ASTM D6319 | Same | ||
| Freedom from Holes | Be free from holes when tested in accordance with ASTMD5151 AQL=2.5 | Be free from holes when tested in accordance with ASTMD5151 AQL=2.5 | Same |
Table3 Performance Comparison
{6}------------------------------------------------
| Powder Content | <0.25 mg/glove | Meet the requirements of ASTM D6124 | Similar |
|---|
Table4 Safety Comparison
| Item | Subject device(K222534) | Predicatedevice(K211914) | Remark | |
|---|---|---|---|---|
| Material | Nitrile | Nitrile | Same | |
| Biocompatibility | Irritation (ISO10993-10:2010BiologicalEvaluation ofMedical Devices -Part 10: Tests ForIrritation And SkinSensitization) | Underconditions ofthe study, notan irritant | Same | |
| Sensitization (ISO10993-10:2010BiologicalEvaluation ofMedical Devices -Part 10: Tests ForIrritation And SkinSensitization) | Underconditions ofthe study, not asensitizer. | Comply withISO10993-10 | ||
| Cytotoxicity (ISO10993-5:2009BiologicalEvaluation ofMedical Devices -Part 5: Tests For InVitro Cytotoxicity) | Underconditions ofthe study,device extractis not cytotoxic | ComplywithISO10993-5 | Similar | |
| Label and Labeling | Meet FDA'sRequirement | Meet FDA'sRequirement | SAME |
8.0 Discussion of Non-clinical and Performance Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
{7}------------------------------------------------
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves
ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
| No. | Name of the TestMethodology / Standard | Purpose | Acceptance Criteria | Results |
|---|---|---|---|---|
| 1 | ISO 10993-10:2010Biological Evaluation OfMedical Devices - Part10: Tests For IrritationAnd Skin Sensitization. | This part of ISO 10993 assesses possible contact hazards from chemicals released from medical devices, which may produce skin and mucosal irritation, eye irritation or skin sensitization. | Skin Sensitization Test:provided grades less than 1, otherwise sensitization.Skin Irritation Test:If the primary irritation index is 0-0,4, the response category is Negligible.0,5-1,9 means slight2-4,9 means moderate5-8 means severe | All grades are 0.All animals were survived and no abnormal signs were observed during the study.The primary irritation index is 0.The response of the proposed device was categorized as negligible under the test condition |
| 3 | ISO 10993-5:2009Biological Evaluation OfMedical Devices - Part5: Tests For In VitroCytotoxicity | This part of ISO 10993 describes test methods to assess the in vitro cytotoxicity of medical devices. | The viab.% of the 100% extract of the test article is the final result, and if viability is reduced to <70% of the blank, it has cytotoxic potential. | Viab.% of 100% test article extract is 70.9%It means the proposed device have no potential toxicity to L-929 in the MTT method |
| 4 | ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves | This standard is designed to determine the amount ofresidual powder (or filter-retained mass) found on medical gloves | powder residue limit of 2.0 mg | <0.25 mg |
| 5 | ASTM D5151-06(Reapproved2 015), Standard Test Method for Detection of Holes in Medical Gloves. | This test method covers the detection of holes in medical gloves. | Samples number: 125 glovesAQL: 2.5 (ISO 2859)Criterion ≤ 7 gloves for water leakage | no glove water leakage found |
| 6 | ASTM D6319-10(Reapproved 2015),Standard Specification For Nitrile Examination Gloves For Medical Application. | This specification covers certain requirements for nitrile rubber gloves used in conducting medical examinations and diagnostic and therapeutic procedures. | Sterility: no needFreedom from holes: pl. Refer to No. 5 in table 5Dimensions:S: width 80±10mmLength ≥220 mmM: width 95±10mmLength ≥230 mmL: width 110±10mmLength ≥230 mmXL: width 120±10mmLength ≥230 mmThickness:Finger ≥0.05 mmPalm ≥0.05 mmPhysical properties:Before agingTensile strength ≥ 14MPaUltimate Elongation ≥ 500%After Accelerated AgingTensile strength ≥ 14MPaUltimate Elongation ≥ 400% | N.A.Dimensions:S: width: 84-87 mmLength 297-302 mmM: width 94-97 mmLength 296-302 mmL: width 104-108mmLength 296-302 mmXL: width 113-117 mmLength 297-302 mmThickness:Finger 0.152-0.184 mmPalm 0.112-0.131 mmPhysical properties:Before agingTensile strength 14.2-29.0 MPaUltimate Elongation 537.700% - 552.733%After Accelerated AgingTensile strength 14.1-26.8 MPaUltimate Elongation 421.422% - 552.633%Powder-free Residue: pl. Refer to No. 4 in table 5 |
| Powder-free Residue:pl. Refer to No. 4 in table5 |
Table 5 Summary of Non-Clinical Performance Testing
{8}------------------------------------------------
{9}------------------------------------------------
9.0 Discussion of Clinical and Performance Testing
Clinical testing is not needed for this device.
10.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.