K Number
K222439
Date Cleared
2023-08-08

(361 days)

Product Code
Regulation Number
862.3950
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Atellica® CH Phencyclidine (Pcp) assay is for in the qualitative or semiguantitative analyses of phencyclidine in human urine using the Atellica® CI Analyzer, using a cutoff of 25 ng/mL. The Pop assay provides only a preliminary analytical test result. A more specific alternative chemical must be used to obtain a confirmed analytical result. Gas chromatography-mass spectrometry (GCMS) is the preferred confirmatory method. The semiquantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas chromatography/mass spectrometry (GC-MS) or liquid chromatography/tandem mass spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

The Atellica® CH Vancomycin (Vanc) assay is for in vitro diagnostic use in the quantitative measurement of vancomycin in human serum and plasma (lithium heparin) using the Atellica® CI Analyzer. Vanc test results may be used in the diagnosis and treatment of vancomycin overdose and in monitoring levels of vancomycin to ensure appropriate therapy.

Device Description

The Atellica CH Pcp assay is a homogenous enzyme immunoassay based on competition between drug in the specimen and drug labeled with glucose-6-phosphate dehydrogenase (G6PDH) for antibody binding sites. G6PDH activity decreases upon binding to the antibody, so the drug concentration in the specimen can be measured in terms of enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD+) to NADH in the presence of glucose-6-phosphate (G6P), resulting in an absorbance change that is measured spectrophotometrically at 340/410 nm. Endogenous G6PDH does not interfere because the coenzyme NAD+ functions only with the bacterial (Leuconostoc mesenteroides) enzyme employed in the assay.

The Atellica CH Vanc assay is based on a homogeneous particle enhanced turbidimetric inhibition immunoassay (PETINIA) technique which uses a synthetic particle-vancomycin conjugate (PR) and monoclonal vancomycin specific antibody (Ab). Vancomycin present in the sample competes with vancomycin on the particles for available antibody, thereby decreasing the rate of aggregation. Hence, the rate of aggregation is inversely proportional to the concentration of vancomycin in the sample. The rate of aggregation is measured using bichromatic turbidimetric readings at 545 and 694 nm.

AI/ML Overview

This document describes the analytical performance of two in vitro diagnostic (IVD) assays, the Atellica® CH Phencyclidine (Pcp) assay and the Atellica® CH Vancomycin (Vanc) assay, and does not contain information related to AI/ML clinical studies or multi-reader multi-case (MRMC) comparative effectiveness studies. Therefore, it is not possible to address acceptance criteria and study designs typically found in AI/ML performance evaluations (e.g., sample size for training/test sets, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, etc.).

However, I can extract and present the analytical performance criteria and reported device performance based on the provided text, which are relevant for IVD devices.


Analytical Acceptance Criteria and Reported Device Performance

Atellica® CH Phencyclidine (Pcp) Assay

This assay is for qualitative or semi-quantitative analysis of phencyclidine in human urine. The primary "acceptance" is demonstrated through agreement with a confirmatory method (GC/MS) and robust precision and recovery.

Table 1: Atellica® CH Phencyclidine (Pcp) Assay - Analytical Performance

Acceptance Criteria / Performance CharacteristicReported Device Performance (Pcp Assay)
Qualitative and Semiquantitative Accuracy (vs. GC/MS)Agreement Summary:- Overall POS agreement: 95% (19 true positives + 81 high positives / 108 total positives by Atellica CH)- Overall NEG agreement: 94% (42 low negatives + 7 negatives / 54 total negatives by Atellica CH)Discordant Results (Atellica CI Pcp vs GC/MS cut-off 25 ng/mL):- Atellica POS, GC/MS NEG: Samples 47 (27 ng/mL vs 18.0 ng/mL), 51 (30 ng/mL vs 24.2 ng/mL), 52 (26 ng/mL vs 24.8 ng/mL) - 3 false positives relative to GC/MS cut-off- Atellica NEG, GC/MS POS: Samples 53 (21 ng/mL vs 25.8 ng/mL), 54 (20 ng/mL vs 26.7 ng/mL), 56 (22 ng/mL vs 27.6 ng/mL), 57 (24 ng/mL vs 27.8 ng/mL), 58 (24 ng/mL vs 28.5 ng/mL) - 5 false negatives relative to GC/MS cut-off
Precision (Repeatability)Overall Low CV% (e.g., 2.2% at 18.75 ng/mL, 2.4% at 25 ng/mL, 2.8% at 31.25 ng/mL)- 0 ng/mL: SD 0.1, N/A CV- 6.25 ng/mL: SD 0.4, 6.7% CV- 12.5 ng/mL: SD 0.4, 3.3% CV- 18.75 ng/mL: SD 0.4, 2.2% CV- 25 ng/mL (Cutoff): SD 0.6, 2.4% CV- 31.25 ng/mL: SD 0.9, 2.8% CV- 37.5 ng/mL: SD 0.9, 2.3% CV- 43.75 ng/mL: SD 1.1, 2.6% CV- 50 ng/mL: SD 1.7, 3.3% CV
Precision (Within-Lab)Overall Low CV% (e.g., 4.4% at 18.75 ng/mL and 25 ng/mL, 5.3% at 31.25 ng/mL)- 0 ng/mL: SD 0.20, N/A CV- 6.25 ng/mL: SD 0.6, 10.0% CV- 12.5 ng/mL: SD 0.6, 5.0% CV- 18.75 ng/mL: SD 0.8, 4.4% CV- 25 ng/mL (Cutoff): SD 1.1, 4.4% CV- 31.25 ng/mL: SD 1.7, 5.3% CV- 37.5 ng/mL: SD 2.3, 5.9% CV- 43.75 ng/mL: SD 2.5, 5.8% CV- 50 ng/mL: SD 3.7, 7.1% CV
Reproducibility (Total)Overall Low CV% (e.g., 6.1% for Urine QC 1, 5.8% for Urine QC 2, 6.5% for Urine QC 3)- Urine QC 1 (18 ng/mL): SD 1.1, 6.1% CV- Urine QC 2 (24 ng/mL): SD 1.4, 5.8% CV- Urine QC 3 (34 ng/mL): SD 2.2, 6.5% CV
RecoveryMean Recovery ranging from 90% to 107% across various concentrations. - 0 ng/mL: 0 ng/mL (N/A %)- 4 ng/mL: 4 ng/mL (101 %)- 5 ng/mL: 5 ng/mL (100 %)- 10 ng/mL: 9 ng/mL (90 %)- 15 ng/mL: 15 ng/mL (100 %)- 20 ng/mL: 19 ng/mL (95 %)- 25 ng/mL: 24 ng/mL (96 %)- 30 ng/mL: 30 ng/mL (100 %)- 40 ng/mL: 43 ng/mL (107 %)- 60 ng/mL: 64 ng/mL (107 %)- 80 ng/mL: 82 ng/mL (103 %)
Endogenous Substances InterferenceNo false response relative to the 25 ng/mL cutoff for tested substances (Acetone, Ascorbic Acid, Conjugated bilirubin, Creatinine, Ethanol, Gamma Globulin, Galactose, Glucose, Hemoglobin, Human Serum Albumin, Oxalic Acid, Riboflavin, Sodium Azide, Sodium Chloride, Sodium Fluoride, Urea) at specified concentrations when spiked into control pools (19 ng/mL and 31 ng/mL).
Specificity (Structurally Unrelated Compounds)No false response relative to the 25 ng/mL cutoff for listed structurally unrelated compounds (e.g., Acetaminophen, Amitriptyline, Caffeine, Ibuprofen, etc.) at specified concentrations when spiked into control pools (19 ng/mL and 31 ng/mL).
Specificity (Structurally Related Compounds - Cross-Reactivity)Values range from 0.0% to 184.4% for structurally related compounds, indicating varying levels of cross-reactivity. Notably, 1-(1-Phenylcyclohexyl)pyrrolidine (PCPy) showed 154.4% and trans-4-phenyl-4-Piperidinocyclohexanol showed 184.4% cross-reactivity. This typically means these compounds may cause a positive result even if PCP itself is not present, emphasizing the need for confirmatory testing.
Specific Gravity and pH InterferenceNo interference observed for negative urine pools with specific gravity 1.000–1.030 and pH 3–10, when tested at ±25% of the cutoff concentration.
Standardization TraceabilityTraceable to Emit Calibrators/Controls, which are referenced to gravimetrically prepared standards qualified by GC/MS from an independent laboratory (within ±10% of nominal).

Atellica® CH Vancomycin (Vanc) Assay

This assay is for quantitative measurement of vancomycin in human serum and plasma.

Table 2: Atellica® CH Vancomycin (Vanc) Assay - Analytical Performance

Acceptance Criteria / Performance CharacteristicReported Device Performance (Vanc Assay)
Limit of Detection (LoD)LoD ≤ 1.0 µg/mL. Reported LoD is 1.0 µg/mL (0.7 µmol/L). The Limit of Blank (LoB) is 0.6 µg/mL (0.4 µmol/L).
Limit of Quantitation (LoQ)LoQ ≤ 3.0 µg/mL. Reported LoQ is 3.0 µg/mL (2.1 µmol/L) (defined by total allowable error ≤ 20%).
Precision (Repeatability)Overall Low CV% (e.g., 0.8% - 2.3%).- Serum QC 1 (6.1 µg/mL): SD 0.14, 2.3% CV- Serum 1 (13.4 µg/mL): SD 0.13, 1.0% CV- Serum QC 2 (19.5 µg/mL): SD 0.15, 0.8% CV- Serum QC 3 (32.6 µg/mL): SD 0.34, 1.0% CV- Serum 2 (46.1 µg/mL): SD 0.54, 1.2% CV
Precision (Within-Laboratory)Overall Low CV% (e.g., 1.5% - 2.8%).- Serum QC 1 (6.1 µg/mL): SD 0.17, 2.8% CV- Serum 1 (13.4 µg/mL): SD 0.20, 1.5% CV- Serum QC 2 (19.5 µg/mL): SD 0.33, 1.7% CV- Serum QC 3 (32.6 µg/mL): SD 0.61, 1.9% CV- Serum 2 (46.1 µg/mL): SD 0.89, 1.9% CV
Reproducibility (Total)Overall Low CV% (e.g., 1.8% - 4.0%).- Serum QC 1 (6.0 µg/mL): SD 0.24, 4.0% CV- Serum 1 (13.4 µg/mL): SD 0.27, 2.0% CV- Serum QC 2 (19.7 µg/mL): SD 0.38, 1.9% CV- Serum QC 3 (32.9 µg/mL): SD 0.62, 1.9% CV- Serum 2 (45.9 µg/mL): SD 0.81, 1.8% CV
Assay Comparison (Correlation vs. Predicate)Correlation coefficient ≥ 0.980 and slope 1.00 ± 0.10.- Regression Equation: y = 0.97x + 0.3 µg/mL (y = 0.97x + 0.2 µmol/L)- Correlation coefficient (r): 0.999 (for 107 serum samples in range 4.1–45.9 µg/mL). Meets criteria.
Specimen Equivalency (Serum vs. Plasma (Lithium Heparin))Demonstrated equivalency between plasma and serum.- Regression Equation: y = 1.00x - 0.1 µg/mL (y = 1.00x - 0.7 µmol/L)- Correlation coefficient (r): 0.996 (for 50 samples in range 4.5–43.9 µg/mL).
Interferences (Hemolysis, Icterus, Lipemia - HIL)≤ 10% bias.- Hemoglobin (1000 mg/dL): 2% and 6% bias at two analyte levels.- Bilirubin, conjugated (30 mg/dL): 0% and 1% bias.- Bilirubin, unconjugated (30 mg/dL): -2% and -1% bias.- Lipemia (Intralipid® 2000 mg/dL): 8% and 6% bias.- Lipemia (from trig fraction 2000 mg/dL): 6% and 8% bias. All results meet the <10% bias criteria.
Interferences (Non-Interfering Substances)≤ 10% bias for listed substances (e.g., Acetaminophen, Amikacin, Caffeine, Gentamicin, Ibuprofen, etc.) at specified concentrations when tested at 10.0 and 40.0 µg/mL analyte concentrations. All results meet the ≤ 10% bias criteria.
Standardization TraceabilityTraceable to United States Pharmacopeia (USP) material.

Regarding the other points relevant to AI/ML studies:

Since the provided text describes the analytical validation of in vitro diagnostic assays (chemical laboratory tests) and not an AI/ML-driven medical device, the following points are not applicable and thus cannot be addressed from the given information:

  1. Sample sizes used for the test set and the data provenance: Not an AI/ML device. The "test set" for the Pcp assay was 157 samples compared against GC/MS. For Vanc, various sample numbers were used for different analytical performance tests (e.g., 107 for assay comparison, 50 for specimen equivalency, 80 for precision studies). Data provenance is not specified (e.g., country of origin, retrospective/prospective).
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable; ground truth is established by a reference analytical method (GC/MS for Pcp) or by defined reference materials and methods for quantitative assays (USP standards for Vanc).
  3. Adjudication method for the test set: Not applicable.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. These are lab assays, not AI algorithms.
  6. The type of ground truth used:
    • For Atellica® CH Phencyclidine (Pcp) assay: The ground truth for accuracy was established using Gas Chromatography/Mass Spectrometry (GC/MS) as the reference method.
    • For Atellica® CH Vancomycin (Vanc) assay: The ground truth/standardization is traceable to United States Pharmacopeia (USP) standards.
  7. The sample size for the training set: Not applicable. These assays are laboratory tests and do not involve "training sets" in the AI/ML sense. Validation is performed using established analytical testing protocols with various spiked and clinical samples.
  8. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Siemens Healthcare Diagnostics Inc. Joy Anoop Clinical Regulatory Affairs Specialist 511 Benedict Avenue Tarrytown, NY 10591

Re: K222439

Trade/Device Name: Atellica® CH Vancomycin (Vanc), Atellica® CH Phencyclidine (Pcp) Regulation Number: 21 CFR 862.3950 Regulation Name: Vancomycin Test System Regulatory Class: Class II Product Code: LEH, LCM Dated: January 16, 2023 Received: January 17, 2023

Dear Joy Anoop:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Joseph A. Digitally signed by Kotarek -S Date: 2023.08.08 Joseph Kotarek, Ph.D. Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222439

Device Name Atellica® CH Phencyclidine (Pcp) Atellica® CH Vancomycin (Vanc)

Indications for Use (Describe)

The Atellica® CH Phencyclidine (Pcp) assay is for in the qualitative or semiguantitative analyses of phencyclidine in human urine using the Atellica® CI Analyzer, using a cutoff of 25 ng/mL. The Pop assay provides only a preliminary analytical test result. A more specific alternative chemical must be used to obtain a confirmed analytical result. Gas chromatographymass spectrometry (GCMS) is the preferred confirmatory method. The semiquantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas chromatography/mass spectrometry (GC-MS) or liquid chromatography/tandem mass spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

The Atellica® CH Vancomycin (Vanc) assay is for in vitro diagnostic use in the quantitative measurement of vancomycin in human serum and plasma (lithium heparin) using the Atellica® CI Analyzer. Vanc test results may be used in the diagnosis and treatment of vancomycin overdose and in monitoring levels of vancomycin to ensure appropriate therapy.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) Number: K222439

1. APPLICANT

Siemens Healthcare Diagnostics Inc. 511 Benedict Avenue, Tarrytown, NY 10591 USA

Anoop Jov Contact: Clinical Regulatory Affairs Specialist Phone: (516) 232-3307 E-mail: anoop.joy@siemens-healthineers.com

Date Prepared: March 28, 2023

2. Regulatory Information

Assay: Atellica CH Phencyclidine (Pcp)

Classification Name: Enzyme Immunoassay, Phencyclidine Regulation Section: unclassified Trade Name: Atellica® CH Phencyclidine (Pcp) Classification: Unclassified, 510(k) required Product Code: LCM Panel: Toxicology

Assay: Atellica CH Vancomycin (Vanc)

Classification Name: Vancomycin test system Regulation Section: 21 CFR 862.3950 Trade Name: Atellica® CH Vancomycin (Vanc) Classification: Class II Product Code: LEH Panel: Toxicology

{4}------------------------------------------------

3. PREDICATE DEVICE INFORMATION

Predicate DeviceCandidate Device510(k) #ClassCode
Atellica CH Phencyclidine (Pcp)Atellica CH Phencyclidine (Pcp)K163220unclassifiedLCM
Trinidad CH Vancomycin (Vanc)Atellica CH Vancomycin (Vanc)K160202Class IILEH

4. DEVICE DESCRIPTION

4.1. Atellica CH Pcp

The Atellica CH Pcp assay is a homogenous enzyme immunoassay based on competition between drug in the specimen and drug labeled with glucose-6-phosphate dehydrogenase (G6PDH) for antibody binding sites. G6PDH activity decreases upon binding to the antibody, so the drug concentration in the specimen can be measured in terms of enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD+) to NADH in the presence of glucose-6-phosphate (G6P), resulting in an absorbance change that is measured spectrophotometrically at 340/410 nm. Endogenous G6PDH does not interfere because the coenzyme NAD+ functions only with the bacterial (Leuconostoc mesenteroides) enzyme employed in the assay.

4.2. Atellica CH Vanc

The Atellica CH Vanc assay is based on a homogeneous particle enhanced turbidimetric inhibition immunoassay (PETINIA) technique which uses a synthetic particle-vancomycin conjugate (PR) and monoclonal vancomycin specific antibody (Ab). Vancomycin present in the sample competes with vancomycin on the particles for available antibody, thereby decreasing the rate of aggregation. Hence, the rate of aggregation is inversely proportional to the concentration of vancomycin in the sample. The rate of aggregation is measured using bichromatic turbidimetric readings at 545 and 694 nm.

5. INTENDED USE

5.1 Atellica CH Pcp

The Atellica® CH Phencyclidine (Pcp) assay is for in vitro diagnostic use in the qualitative or semiquantitative analyses of phencyclidine in human urine using the Atellica® CI Analyzer, using a cutoff of 25 ng/mL. The Pcp assay provides only a preliminary analytical test result. A more specific alternative chemical must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas chromatography/mass spectrometry (GC-MS) or liquid chromatography/tandem mass spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

{5}------------------------------------------------

Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

5.2Atellica CH Vanc

The Atellica® CH Vancomycin (Vanc) assay is for in vitro diagnostic use in the quantitative measurement of vancomycin in human serum and plasma (lithium heparin) using the Atellica® CI Analyzer. Vanc test results may be used in the diagnosis and treatment of vancomycin overdose and in monitoring levels of vancomycin to ensure appropriate therapy.

6. INDICATIONS FOR USE

Same as Intended use

7. COMPARISION OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The following table provides a comparison between the predicate and candidate device.

7.1. Atellica CH Pcp

Below is a features comparison for the Atellica CH Pcp assay on the Atellica CI Analyzer and the predicate device Atellica IM Analyzer

FeaturePredicate Device:Atellica CH Phencyclidine(Pcp) on Atellica CHAnalyzerNewDevice:Atellica CH Phencyclidine(Pcp) on Atellica Cl Analyzer
Intended Use :The Atellica CHPhencyclidine (Pcp) assayis for in vitro diagnostic usein the qualitative orsemiquantitative analysesof phencyclidine in humanurine using the Atellica CHAnalyzer, using a cutoff of25 ng/mL. The Pcp assayprovides only a preliminaryanalytical test result. Amore specific alternativechemical method must beused to obtain a confirmedanalytical result. Gaschromatography/massspectrometry (GC/MS) isthe preferred confirmatorymethod. The semi-The Atellica CHPhencyclidine (Pcp) assayis for in vitro diagnostic usein the qualitative orsemiquantitative analysesof phencyclidine in humanurine using the Atellica ClAnalyzer, using a cutoff of25 ng/mL. The Pcp assayprovides only a preliminaryanalytical test result. Amore specific alternativechemical method must beused to obtain a confirmedanalytical result. Gaschromatography/massspectrometry (GC/MS) isthe preferred confirmatorymethod. The semi-
quantitative mode is forpurposes of enablinglaboratories to determinean appropriate dilution ofthe specimen forconfirmation by aconfirmatory method suchas gaschromatography/mass-spectrometry (GC-MS) orliquidchromatography/tandemmass spectrometry (LC-MS/MS) or permittinglaboratories to establishquality control procedures.Clinical consideration andprofessional judgmentshould be applied to anydrug-of- abuse test result,particularly whenpreliminary positive resultsare used.quantitative mode is forpurposes of enablinglaboratories to determinean appropriate dilution ofthe specimen forconfirmation by aconfirmatory method suchas gaschromatography/massspectrometry (GC-MS) orliquidchromatography/tandemmass spectrometry (LC-MS/MS) or permittinglaboratories to establishquality control procedures.Clinical consideration andprofessional judgmentshould be applied to anydrug-of-abuse test result,particularly whenpreliminary positive resultsare used
Type of Product:Analytical ReagentsSame
Measured Analyte:PcpSame
Test Matrix:UrineSame
Device Technology:Enzyme ImmunoassaySame
Materials:Matched lots of polyclonalantibody reactive tophencyclidine andphencyclidine labeled withglucose-6-phosphatedehydrogenase are used inthis Syva® Emit® II Plusmethodology.Same
Cutoff Levels:25 ng/mL PCPSame
ConfirmatoryMethod:Gas Chromatography/massspectrometrySame
CalibrationFrequency:60 days130 days

{6}------------------------------------------------

{7}------------------------------------------------

7.2. Atellica CH Vanc

Below is a features comparison for the Atellica CH Vanc assay on the Atellica CI Analyzer and the predicate device Atellica IM Analyzer

FeaturePredicate Device:Trinidad CH Vancomycin(Vanc) on Trinidad CH SystemNew Device:Atellica CH Vancomycin(Vanc) on Atellica CIAnalyzer
Intended Use :The Trinidad CH Vancomycin(Vanc) assay is for in vitrodiagnostic use in thequantitative measurement ofvancomycin in human serumand plasma (lithium heparin)using the Trinidad CHSystem.The Atellica CHVancomycin (Vanc) assayis for in vitro diagnosticuse in the quantitativemeasurement ofvancomycin in humanserum and plasma (lithiumheparin) using the AtellicaCI Analyzer.
Indications for Use:Vanc test results may beused in the diagnosis andtreatment of vancomycinoverdose and in monitoringlevels of vancomycin toensure appropriate therapy.Same
Device Technology:Homogeneous particleenhanced turbidimetricinhibition immunoassay(PETINIA) techniqueSame
Sample Type:Serum/ Lithium HeparinplasmaSame
Therapeutic Interval:Peak Intervals: Samples fromadult volunteers drawn twohours after the completion ofa 60-minute infusion ofvancomycin ranged from 18 –26 µg/mL.Samples drawn one hourafter the completion of a 60minute vancomycin infusionranged from 25 – 40 µg/mL.Samples drawn 30 minutesafter the completion of a 60minute infusion of vancomycinranged from 30 – 40 µg/mL.Trough Intervals: Samplesshould be drawn just before thenext dose. A trough interval of 5– 10 µg/mL is generallyconsidered to be effective.Same
Standardization:Traceable to United StatesPharmacopeia (USP) standards. SameSame
Calibration Frequency:30 daysSame
Analytical MeasuringInterval:3.0 – 50.0 µg/mLSame
Interferences:Bilirubin (Conjugated &Unconjugated) – 20 mg/dLLipemia (Intralipid®) – 1000mg/dLHemoglobin – 600 mg/dLBilirubin (Conjugated &Unconjugated) – 30 mg/dLLipemia (Intralipid®) – 2000mg/dLHemoglobin – 1000 mg/dL
Calibrators:Atellica CH Drug CAL IISame

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

8. PERFORMANCE CHARACTERISTICS DATA

8.1. Atellica CH Pcp

Assay Comparison

Qualitative and Semiquantitative Results

A total of one-hundred fifty-seven (157) phencyclidine samples were analyzed using the Atellica CH Pcp assay and the reference method GC/MS. All assays used a 25 ng/mL cutoff. Thirty-four (34) samples were within ± 50% of the cutoff by GC/MS. The agreement of the assay may vary depending on the study design, comparative assay, and sample population. Qualitative and Semiquantitative Accuracy Summary of Atellica CH Pcp Assay versus GC/MS

LOW NEG< 50% below thecutoff(< 13 ng/mL)NEGWithin 50% belowthe cutoff(13-24 ng/mL)POSWithin 50% abovethe cutoff(25-38 ng/mL)HIGH POS> 50% above thecutoff(> 38 ng/mL)% Agree-ment
Qualitative Summary
Atellica CHPOS03198195
NEG4275094
Semiquantitative Summary
Atellica CHPOS03198195
NEG4275094

Discordant Result Summary between Atellica CH Pcp Assay and GC/MS

Sample #Atellica CI Pcp(ng/mL)GC/MS Phencyclidine(ng/mL)Atellica CI Pcp vs GC/MS(POS/NEG)
472718.0+/-
513024.2+/-
522624.8+/-
532125.8-/+
542026.7-/+
562227.6-/+
572427.8-/+
582428.5-/+

{11}------------------------------------------------

Precision

Precision was determined in accordance with CLSI Document EP05-A3. Repeatability and within-lab precision were determined by assaying negative urine pools spiked with phencyclidine at nine different levels. The assay is designed to classify the levels as positive or negative relative to the cutoff for the spiked sample pools. Samples were assayed on an Atellica CI Analyzer in duplicate in 2 runs per day for 20 days (n = 80 for each sample). The results in the qualitative mode and semi-quantitative mode are identical. The results are summarized below.

Precision Qualitative and Semi-Quantitative Analysis
Urine Pool(ng/mL)% ofCutoff# ofResultsMean(ng/mL)RepeatabilityWithin-Lab
0-1008000.1N/A0.20N/A
6.25-758060.46.70.610.0
12.5-5080120.43.30.65.0
18.75-2580180.42.20.84.4
25Cutoff80250.62.41.14.4
31.252580320.92.81.75.3
37.55080390.92.32.35.9
43.757580431.12.62.55.8
5010080521.73.33.77.1

Reproducibility

Reproducibility was determined in accordance with CLSI Document EP05-A3. Samples were assayed n=5 in 1 run for 5 days using 3 instruments and 3 reagent lots. The data were analyzed to calculate the following components of precision: repeatability, between-day, between-lot, between-instrument, and reproducibility (total). The following results were obtained:

RepeatabilityBetween-DayBetween-InstrumentBetween-LotTotal Reproducibility
SampleNaMean(ng/mL)SD(ng/mL)CV(%)SD(ng/mL)CV(%)SD(ng/mL)CV(%)SD(ng/mL)CV(%)SD(ng/mL)CV(%)
Urine QC 1225180.52.80.63.30.31.70.73.91.16.1
Urine QC 2225240.52.10.93.80.72.90.62.51.45.8
Urine QC 3225340.82.41.54.41.23.50.82.42.26.5

a Number of results.

b Standard deviation.

Coefficient of variation.

{12}------------------------------------------------

Recovery

Recovery of Pcp samples were prepared by spiking known amounts of phencyclidine into negative urine pools. Each spiked sample was analyzed using the Atellica CH Pcp assay. Results of recovery are shown below.

Target Concentration(ng/mL)Mean Measured Concentration(ng/mL)Mean Recovery%
00N/A
44101
55100
10990
1515100
201995
252496
3030100
4043107
6064107
8082103

Endogenous Substances

The protocol used follows the CLSI Document, EP07. The endogenous substances were evaluated qualitatively at the concentrations listed below. The substances were spiked into two levels of controls at ±25% (19 ng/mL and 31 ng/mL) of the cutoff concentration. At the stated concentration, the sample did not give a false response relative to the 25 ng/mL cutoff.

CompoundConcentrationTested-25% Cutoff PoolResult(19 ng/mL)+25% Cutoff PoolResult(31 ng/mL)
Acetone1.0 g/dLNegativePositive
Ascorbic Acid0.75 g/dLNegativePositive
Conjugated bilirubin0.25 mg/dLNegativePositive
Creatinine0.5 g/dLNegativePositive
Ethanol1.0 g/dLNegativePositive
Gamma Globulin0.5 g/dLNegativePositive
Galactose0.01 g/dLNegativePositive
Glucose2.0 g/dLNegativePositive
Hemoglobin115 mg/dLNegativePositive
Human Serum Albumin0.5 g/dLNegativePositive
Oxalic Acid0.1 g/dLNegativePositive
Riboflavin7.5 mg/dLNegativePositive

{13}------------------------------------------------

Sodium Azide1% (w/v)NegativePositive
Sodium Chloride1.5 g/dLNegativePositive
Sodium Fluoride1% (w/v)NegativePositive
Urea6.0 g/dLNegativePositive

Specificity

Interference was determined in accordance with CLSI Document EP07. The interference of structurally unrelated compounds and common over the counter drugs was evaluated qualitatively at the concentrations listed below. The compounds were spiked into two levels of controls at ±25% (19 ng/mL and 31 ng/mL) of the cutoff concentration. At the stated concentration, the sample did not give a false response relative to the 25 ng/mL cutoff.

Structurally Unrelated Compounds

CompoundConcentrationTested-25% Cutoff PoolResult(19 ng/mL)+25% Cutoff PoolResult(31 ng/mL)
Acetaminophen500,000 ng/mLNegativePositive
I-a-AcetyImethadol (LAAM)25,000 ng/mLNegativePositive
N-Acetyl procainamide (NAPA)100,000 ng/mLNegativePositive
AcetyIsalicylic Acid500,000 ng/mLNegativePositive
Amitriptyline8,750 ng/mLNegativePositive
S-(+)-Amphetamine100,000 ng/mLNegativePositive
Benzoylecgonine100,000 ng/mLNegativePositive
Boric Acid1% (w/v)NegativePositive
Buprenorphine100,000 ng/mLNegativePositive
Caffeine500,000 ng/mLNegativePositive
Cannabinol100,000 ng/mLNegativePositive
Carbamazepine100,000 ng/mLNegativePositive
Chlordiazepoxide100,000 ng/mLNegativePositive
Cimetidine100,000 ng/mLNegativePositive
Clonidine100,000 ng/mLNegativePositive
Codeine25,000 ng/mLNegativePositive
Cotinine100,000 ng/mLNegativePositive
Desipramine75,000 ng/mLNegativePositive
Dextrorphan781 ng/mLNegativePositive
Diazepam100,000 ng/mLNegativePositive
Digoxin100,000 ng/mLNegativePositive
2-Ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)12,500 ng/mLNegativePositive
EMDP100,000 ng/mLNegativePositive
1R,2S-Ephedrine100,000 ng/mLNegativePositive
CompoundConcentrationTested-25% Cutoff PoolResult(19 ng/mL)+25% Cutoff PoolResult(31 ng/mL)
1S,2R-Ephedrine100,000 ng/mLNegativePositive
Fluoxetine75,000 ng/mLNegativePositive
Flurazepam50,000 ng/mLNegativePositive
Glutethimide100,000 ng/mLNegativePositive
Haloperidol100,000 ng/mLNegativePositive
Heroin25,000 ng/mLNegativePositive
Hydrocodone25,000 ng/mLNegativePositive
lbuprofen500,000 ng/mLNegativePositive
Ketamine75,000 ng/mLNegativePositive
Ketorolac Tromethamine100,000 ng/mLNegativePositive
Lidocaine100,000 ng/mLNegativePositive
Lorazepam100,000 ng/mLNegativePositive
Lormetazepam100,000 ng/mLNegativePositive
LSD100,000 ng/mLNegativePositive
MDMA100,000 ng/mLNegativePositive
Meperidine1,563 ng/mLNegativePositive
Methadone50,000 ng/mLNegativePositive
S(+) - Methamphetamine100,000 ng/mLNegativePositive
Methaqualone100,000 ng/mLNegativePositive
Morphine75,000 ng/mLNegativePositive
Naproxen100,000 ng/mLNegativePositive
Nordiazepam100,000 ng/mLNegativePositive
Nortriptyline75,000 ng/mLNegativePositive
Oxazepam100,000 ng/mLNegativePositive
Oxycodone100,000 ng/mLNegativePositive
Phenobarbital100,000 ng/mLNegativePositive
Phenylephrine100,000 ng/mLNegativePositive
Phenytoin100,000 ng/mLNegativePositive
Promethazine3,125 ng/mLNegativePositive
Propoxyphene100,000 ng/mLNegativePositive
Propranolol100,000 ng/mLNegativePositive
Protriptyline75,000 ng/mLNegativePositive
R,R - Pseudoephedrine100,000 ng/mLNegativePositive
S,S - Pseudoephedrine100,000 ng/mLNegativePositive
Ranitidine100,000 ng/mLNegativePositive
Ritalinic Acid100,000 ng/mLNegativePositive
Salicylic Acid100,000 ng/mLNegativePositive
CompoundConcentrationTested-25% Cutoff PoolResult(19 ng/mL)+25% Cutoff PoolResult(31 ng/mL)
Scopolamine100,000 ng/mLNegativePositive
Secobarbital100,000 ng/mLNegativePositive
Tapentadol50,000 ng/mLNegativePositive
11-nor-Δ9-THC-9-COOH100,000 ng/mLNegativePositive
Tramadol50,000 ng/mLNegativePositive
Trazodone100,000 ng/mLNegativePositive
Tyramine100,000 ng/mLNegativePositive
Verapamil60,000 ng/mLNegativePositive
Zidovudine (AZT)100,000 ng/mLNegativePositive
Zolpidem100,000 ng/mLNegativePositive

{14}------------------------------------------------

{15}------------------------------------------------

1

Structurally Related Compounds

CompoundConcentrationTested(ng/mL)MeanObserved PcpResponse(ng/mL)Cross-Reactivity%
Chloropromazine100,00024.00.0
Clomipramine100,00020.80.0
Cyclobenzaprine25,0007.00.0
Dextromethorphan80,00022.60.0
Diphenhydramine100,00010.80.0
Doxepin90,00013.20.0
Imipramine100,00016.20.0
Methoxetamine36,00014.00.0
4-Methoxyphencyclidine70059.48.5
Thioridazine100,00048.40.0
Venlafaxine100,0007.20.0
РСР2524.296.8
1-(4-Hydroxypiperidino)phenylcyclohexane41926.06.2
1-(1-Phenylcyclohexyl)pyrrolidine (PCPy)(Rolicyclidine)5483.4154.4
1-[1-(2-Thienyl)-cyclohexyl]piperidine (TCP)(Tenocyclidine)377.018.9
trans-4-phenyl-4-Piperidinocyclohexanol3259.0184.4

{16}------------------------------------------------

Specific Gravity and pH

Negative urine pools with specific gravity values ranging from 1.000–1.030 and pH values ranging from 3–10 were tested in the presence of two levels of controls at ±25% of the cutoff concentration (19 ng/mL and 31 ng/mL). No interference was observed.

Standardization

The Atellica CH Pcp assay is traceable to the Emit Calibrators/Controls which are referenced to gravimetrically prepared standards. These standards are qualified by GC/MS from an independent laboratory and must quantitate within ±10% of nominal.

{17}------------------------------------------------

8.2. Atellica CH Vanc

Detection Capability

Detection capability was determined in accordance with CLSI Document EP17-A2. The assay is designed to have a limit of blank (LoB) < LoD. a limit of detection (LoD) ≤ 1.0 ‍μg/mL, and a limit of quantitation (LoQ) ≤ 3.0 µg/mL . Assay results obtained at individual laboratories may vary from the data presented.

The LoD corresponds to the lowest concentration of vancomycin that can be detected with a probability of 95%. The LoD for the Atellica CH Vanc assay is 1.0 µg/mL (0.7 µmol/L), and was determined using 450 determinations, with 225 blank and 225 low level replicates, and a LoB of 0.6 µg/mL (0.4 µmol/L). Assay results obtained at individual laboratories may vary from the data presented.

The LoQ corresponds to the lowest amount of analyte in a sample that can be accurately quantitated with a total allowable error ≤ 20%. The LoQ of the Vanc assay is 3.0 µg/mL (2.1 umol/L), and was determined using n=5 replicates per sample that were assayed using 3 reagent lots, over a period of 3 days, using total analytical error definition of bias + 2SD.

Precision

Precision was determined in accordance with CLSI Document EP05-A3. Samples were assayed on an Atellica CI Analyzer in duplicate in 2 runs per day for at least 20 days (N ≥ 80 for each sample). The following results were obtained:

Sample TypeNMeanµg/mL (µmol/L)RepeatabilityWithin-Laboratory Precision
SDaµg/mL (µmol/L)CVb(%)SDaµg/mL (µmol/L)CVb(%)
Serum QC 1806.1 (4.2)0.14 (0.10)2.30.17 (0.12)2.8
Serum 18013.4 (9.2)0.13 (0.09)1.00.20 (0.14)1.5
Serum QC 28019.5 (13.5)0.15 (0.10)0.80.33 (0.23)1.7
Serum QC 38032.6 (22.5)0.34 (0.23)1.00.61 (0.42)1.9
Serum 28046.1 (31.8)0.54 (0.37)1.20.89 (0.61)1.9

Standard deviation. a

ﻣ Coefficient of variation.

Reproducibility

Reproducibility was determined in accordance with CLSI Document EP05-A3. Samples were assayed n=5 in 1 run for 5 days using 3 instruments and 3 reagent lots. The data were analyzed to calculate the following components of precision: repeatability, between-day, between-lot, between-instrument, and reproducibility (total). The following results were obtained:

{18}------------------------------------------------

RepeatabilityBetween-DayBetween-Instru-mentBetween-LotTotal Repro-ducibility
SampleNaMeanµg/mL(µmol/L)SDbµg/mL(µmol/L)CVc(%)SDµg/mL(µmol/L)CV(%)SDµg/mL(µmol/L)CV(%)SDµg/mL(µmol/L)CV(%)SDµg/mL(µmol/L)CV(%)
Serum QC 12256.0(4.1)0.11(0.1)1.80.18(0.1)3.00.09(0.1)1.50.07(0.0)1.20.24(0.2)4.0
Serum 122513.4(9.2)0.12(0.1)0.90.14(0.1)1.00.03(0.0)0.20.19(0.1)1.40.27(0.2)2.0
Serum QC 222519.7(13.6)0.16(0.1)0.80.29(0.2)1.50.10(0.1)0.50.15(0.1)0.80.38(0.3)1.9
Serum QC 322532.9(22.7)0.22(0.2)0.70.49(0.3)1.50.29(0.2)0.90.09(0.1)0.30.62(0.4)1.9
Serum 222545.9(31.7)0.36(0.2)0.80.50(0.3)1.10.48(0.3)1.00.25(0.2)0.50.81(0.6)1.8

a Number of results.

b Standard deviation.

Coefficient of variation.

Assay Comparison

The Atellica CH Vanc assay is designed to have a correlation coefficient of ≥ 0.980 and a slope of 1.00 ± 0.10 compared to Atellica CH Vanc on Atellica CH Analyzer. Assay comparison was determined using the Deming regression model in accordance with CLSI Document EP09c. The following results were obtained:

SpecimenComparative Assay (x)Regression EquationSample IntervalNarb
SerumAtellica CH Vanc onAtellica CH Analyzer$y = 0.97x + 0.3 \mu g/mL$( $y = 0.97x + 0.2 \mu mol/L$ )4.1–45.9 µg/mL(2.8–38.6 µmol/L)1070.999

a Number of samples tested.

b Correlation coefficient.

Specimen Equivalency

Specimen equivalency was determined using the Deming regression model in accordance with CLSI Document EP09c. The following results were obtained:

Specimen (y)Reference Specimen (x)Regression EquationSample IntervalNarb
Plasma (Lithium Heparin)Serum$y = 1.00x - 0.1 μg/mL$$(y = 1.00x - 0.7 μmol/L)$4.5–43.9 μg/mL(3.1–30.3 μmol/L)500.996

a Number of samples tested.

b Correlation coefficient.

{19}------------------------------------------------

Interferences

Hemolysis, Icterus, and Lipemia (HIL)

The Atellica CH Vanc assay is designed to have ≤ 10% interference from hemoglobin, bilirubin, and lipemia. Interfering substances at the levels indicated in the table below were tested in accordance with CLSI Document EP07 using the Atellica CH Vanc assay. Bias is the difference in the results between the control sample (does not contain the interferent) and the test sample (contains the interferent) expressed in percent. Bias > 10% is considered interference. Analyte results should not be corrected based on this bias.

SubstanceSubstance Test ConcentrationCommon Units (SI Units)Analyte Concentrationµg/mL (μmol/L)Percent Biasa
Hemoglobin1000 mg/dL (10.0 g/L)9.4 (6.5)2
Hemoglobin1000 mg/dL (10.0 g/L)38.1 (26.3)6
Bilirubin, conjugated30 mg/dL (513 μmol/L)9.2 (6.3)0
Bilirubin, conjugated30 mg/dL (513 μmol/L)37.1 (25.6)1
Bilirubin, unconjugated30 mg/dL (513 μmol/L)9.5 (6.6)-2
Bilirubin, unconjugated30 mg/dL (513 μmol/L)39.1 (27.0)-1
Lipemia (Intralipid®)2000 mg/dL (20.0 g/L)9.7 (6.7)8
Lipemia (Intralipid®)2000 mg/dL (20.0 g/L)37.1 (25.6)6
Lipemia (from trig fraction)2000 mg/dL (20.0 g/L)9.9 (6.8)6
Lipemia (from trig fraction)2000 mg/dL (20.0 g/L)38.4 (26.5)8

a Analyte results should not be corrected based on this bias.

Non-Interfering Substances

The following substances do not interfere with the Atellica CH Vanc assay when present in serum and lithium plasma at the concentrations indicated in the table below. Bias due to these substances is ≤ 10% at an analyte concentration of 10.0 and 40.0 µg/mL (6.9 and 27.6 µmol/L).

SubstanceSubstance Test ConcentrationCommon Units (SI Units)Percent Bias
Acetaminophen20 mg/dL (1323 $ \mu $ mol/L)≤ 10%
Acetylsalicylic Acid50 mg/dL (2778 $ \mu $ mol/L)≤ 10%
Amikacin100 $ \mu $ g/mL (171 $ \mu $ mol/L)≤ 10%
SubstanceSubstance Test ConcentrationCommon Units (SI Units)Percent Bias
Amobarbital10 mg/dL (442 µmol/L)≤ 10%
Ampicillin5 mg/dL (143 µmol/L)≤ 10%
Ascorbic Acid3 mg/dL (170.3 µmol/L)≤ 10%
Caffeine10 mg/dL (515 µmol/L)≤ 10%
Carbamazepine12 mg/dL (508 µmol/L)≤ 10%
Cefazolin500 µg/mL (1100 µmol/L)≤ 10%
Cefotaxime1000 µg/mL (2195 µmol/L)≤ 10%
Chloramphenicol100 µg/mL (309 µmol/L)≤ 10%
Chlordiazepoxide2 mg/dL (67 µmol/L)≤ 10%
Chlorpromazine5 mg/dL (157 µmol/L)≤ 10%
Cimetidine10 mg/dL (396 µmol/L)≤ 10%
Clindamycin300 µg/dL (675 µmol/L)≤ 10%
Codeine10 mg/dL (334 µmol/L)≤ 10%
Creatinine30 mg/dL (2652 µmol/L)≤ 10%
Dextran 406000 mg/dL (1500 µmol/L)≤ 10%
Dextran 702500 mg/dL (357 µmol/L)≤ 10%
Diazepam4 mg/dL (140 µmol/L)≤ 10%
Digoxin5 ng/dL (6.4 nmol/L)≤ 10%
Erythromycin20 mg/dL (273 µmol/L)≤ 10%
Ethanol350 mg/dL (76 mmol/L)≤ 10%
Ethosuximide30 mg/dL (2125 µmol/L)≤ 10%
Furosemide2 mg/dL (61 µmol/L)≤ 10%
Fusidic Acid500 µg/mL (968 µmol/L)≤ 10%
Gentamicin12 mg/dL (25 µmol/L)≤ 10%
Heparin (Porcine)8000 U/L (8000 U/L)≤ 10%
Ibuprofen40 mg/dL (1939 µmol/L)≤ 10%
SubstanceSubstance Test ConcentrationCommon Units (SI Units)Percent Bias
Lidocaine6 mg/dL (256 µmol/L)≤ 10%
Lithium3.5 mg/dL (5.04 mmol/L)≤ 10%
Methicillin500 µg/mL (1318 µmol/L)≤ 10%
Netilmicin500 µg/mL (1050 µmol/L)≤ 10%
Nicotine2 mg/dL (123 µmol/L)≤ 10%
Penicillin V80 mg/dL (2247 µmol/L)≤ 10%
Pentobarbital10 mg/dL (442 µmol/L)≤ 10%
Phenobarbital15 mg/dL (646 µmol/L)≤ 10%
Phenytoin10 mg/dL (396 µmol/L)≤ 10%
Primidone10 mg/dL (458 µmol/L)≤ 10%
Propoxyphene0.4 mg/dL (12 µmol/L)≤ 10%
Protein - Albumin12 g/dL (120 g/L)≤ 10%
Protein - IgG5 g/dL (50 g/L)≤ 10%
Protein - Total12 g/dL (120 g/L)≤ 10%
Rheumatoid Factor1465 IU/L (1465 IU/L)≤ 10%
Rifampin50 µg/mL (61 µmol/L)≤ 10%
Salicylic Acid50 mg/dL (3.62 mmol/L)≤ 10%
Secobarbital5 mg/dL (209.8 µmol/L)≤ 10%
Sodium Fluoride1 mg/dL (0.24 mmol/L)≤ 10%
Sulfamethoxazole25 µg/mL (99 µmol/L)≤ 10%
Theophylline25 mg/dL (1388 µmol/L)≤ 10%
Tobramycin100 µg/mL (214 µmol/L)≤ 10%
Trimethoprim25 µg/mL (86 µmol/L)≤ 10%
Urea500 mg/dL (83.3 mmol/L)≤ 10%
Uric Acid20 mg/dL (1.2 mmol/L)≤ 10%
Valproic Acid50 mg/dL (3467 µmol/L)≤ 10%

{20}------------------------------------------------

{21}------------------------------------------------

Standardization

The Atellica CH Vanc assay is traceable to United States Pharmacopeia (USP) material. Assigned values for calibrators are traceable to this standardization

{22}------------------------------------------------

9. CONCLUSION

The candidate devices are substantially equivalent to the Predicate devices and yields substantially equivalent Performance Characteristics. The performance data demonstrates that the devices provide consistent, reproducible, and accurate results and raises no concerns about Safety and Effectiveness.

§ 862.3950 Vancomycin test system.

(a)
Identification. A vancomycin test system is a device intended to measure vancomycin, an antibiotic drug, in serum. Measurements obtained by this device are used in the diagnosis and treatment of vancomycin overdose and in monitoring the level of vancomycin to ensure appropriate therapy.(b)
Classification. Class II.