(66 days)
The NeoCoil 16ch Breast Coils are Magnetic Resonance Imaging (MRI) RF Receive-Only Coils intended for use by trained medical professionals, in combination with and controlled by compatible 1.5T or 3T/3.0T MRI system software.
The NeoCoil 16ch Breast Coils can be used with compatible ancillary components, accessories, and/or devices to provide access to breast anatomy for diagnostic or interventional planning/procedures.
When used as intended, the 16ch Breast Coil provides information used by the MRI system to produce diagnostic and/or interventional planning images of the breast, axilla, and chest wall. The images produced are interpreted by medical professionals as part of clinical decision making.
The NeoCoil 3.0T 16ch Breast Coil is a phased array coil for imaging structures of the breast, axilla and chest wall. The 3.0T 16ch Breast Coil is a three part receive-only coil designed to provide high resolution imaging. The.3.0T 16ch Breast Coil includes a coil support structure, patient support structure, biopsy components, accessories and comfort pads.
The NeoCoil 3.0T 16ch Breast Coil is tuned to receive RF frequency corresponding to the proton precession in a 3 tesla magnetic field, which is governed by the Larmor equation.
The 16ch Breast Coils are intended for use in a manner that is identical to the predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018. The 16ch Breast Coil consists of the following arrays available in 1.5T, 3.0T and 3T field strengths:
- Medial Array, Breast Coil
- . Lateral Array Left, Breast Coil
- Lateral Array Right, Breast Coil ●
- . Baseplate Assembly, Breast Coil
- Biopsy Array Left, Breast Coil ●
- Biopsy Array Right, Breast Coil ●
The coils receive magnetic resonance signals generated in hydrogen nuclei (protons) in the Head. Neck and Brachial Plexus anatomy while blocking the high-frequency magnetic field applied by the MRI scanner at specified timings. The received signals are amplified before being transferred to the MRI scanner through the coil's system cable. The amplified signals are processed into tomographic images of the breast, axilla and chest wall anatomy by the MRI scanner. Images are typically generated as axial, sagittal, coronal oblique slices. Accessories associated with the 16ch Breast Coils include biopsy grids and biopsy drapes.
The 1.5T Breast Coil, the 3T 16ch Breast Coil, and the 3.0T 16ch Breast Coil are tuned to receive RF frequency corresponding to the proton precession in a 1.5 tesla and 3.0 tesla magnetic field (respectively), which is governed by the Larmor equation.
The 16ch Breast Coil is intended for use in a manner that is identical to the predicate device described in this submission.
Proposed labeling is documented in the Instructions for Use manual for the 16ch Breast Coil (NC149IFU-en).
The provided text describes the regulatory clearance of NeoCoil's 16ch Breast Coils (K222407) and outlines the testing performed to demonstrate its substantial equivalence to predicate devices. However, the document does not contain information related to an AI/ML powered medical device or a study involving human readers or expert consensus for ground truth establishment. Instead, it focuses on the performance and safety of a physical medical device (an MRI coil).
Therefore, I cannot extract the information required to answer your questions about AI device acceptance criteria, human reader studies, or ground truth establishment based on the provided text. The questions you've asked are typically relevant to the evaluation of AI/ML-driven diagnostic or interpretative tools, which is not what this document addresses.
The document primarily covers non-clinical bench testing for a physical MRI coil, ensuring it meets safety and performance standards comparable to an existing device.
Here's what the document does cover regarding "acceptance criteria" and "study" for the physical device:
1. A table of acceptance criteria and the reported device performance
The document provides a table of "Test Performed" (which serves as acceptance criteria) and the "Results Summary" (reported performance).
| Test Performed | Objective(s) of the Test | Test Method Description | Pre-defined pass/fail criteria | Results Summary | Discussions / Conclusions |
|---|---|---|---|---|---|
| Biocompatibility Assessment | Assess potential biological risks | Evaluation of data; historical use, biologic testing, where warranted | Acceptable level of risk | Pass | No identified significant risks. |
| Electrical Safety (60601-1) | Basic electrical safety/essential performance | Test Lab | Pre-defined performance standards | Pass | Applicable requirements for basic electrical safety and essential performance met. |
| Electrical Safety (60601-2-33) | Particular electrical requirements: MR equipment | Test Lab | Pre-defined performance standards | Pass | Applicable requirements of the particular standard were met. |
| Electrical Safety (60601-1-2) | Collateral electrical safety/essential performance | Test Lab / Bench Testing | Pre-defined performance standards | Pass | Applicable requirements of the collateral standard were met. |
| Usability Assessment | Devices meet customer, end user and patient needs | Actual, simulated or retrospective evaluation of the device and/or data | Pre-defined requirements | Pass | The devices met the needs of the customer, end user and patient. |
| Entrapment, Trapping Zone and Cable Looping (assessment w/ scanner) | Assess the device for pinch points, entrapment, cable looping - interfacing with MRI scanner | Evaluation of coil-to-scanner entrapment, trapping and cable looping not covered by test lab assessments. | Requirements based on pre-defined requirements in 60601-1 and customer requirements | Pass | Requirements were met. |
| Surface Temperature | Surface temperatures do not exceed limits | MRI scanner test | Pre-defined performance standards | Pass | Surface temperatures were within IEC limits. |
| Unplugged Surface Temperature | Devices remain safe in first fault condition | MRI scanner test | Acceptable level of risk | Pass | Surface temperatures were within IEC limits when the coil is left unplugged in the MRI scanner. |
| Blocking Network Analysis | Ensures devices are designed with adequate active and passive transmit decoupling | Theoretical calculations | Adequate transmit decoupling | Pass | Blocking network demonstrates adequate active and passive transmit decoupling. |
| B1 Field Distortion | Measure amount of distortion produced due to presence of an RF coil in the scanner | MRI scanner test | Pre-defined performance standards | Pass | B1 field inhomogeneity meets performance requirements and demonstrates adequate active and passive transmit decoupling. |
| B0 Field Distortion | Measure amount of distortion produced due to presence of an RF coil in the scanner | MRI scanner test | Pre-defined performance standards | Pass | B0 field inhomogeneity meets performance requirements and demonstrates adequate active and passive transmit decoupling. |
| NEMA MS 6-2008 | Evaluate single-channel non-volume special purpose radiofrequency (RF) coils for use with magnetic resonance (MR) imaging (MRI) systems | MRI scanner test | Pre-defined performance standards | Pass | SNR and Image Uniformity are consistent with the requirements for indications for use. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document mentions "Bench Testing" and "MRI scanner test" but does not specify sample sizes for these tests (e.g., number of coils tested, number of scans performed). It also does not provide information on data provenance like country of origin or whether the data was retrospective or prospective, as these are typically considerations for clinical data, which was not gathered for this 510(k).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable as the study involves the performance of a physical device against engineering and safety standards, not an AI model requiring human ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This concept is not relevant to the described device. The tests performed are on the physical MRI coil itself.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the physical device, the "ground truth" for the bench tests are the established engineering standards, pre-defined performance criteria (e.g., for electrical safety, temperature limits, B0/B1 field homogeneity), and consensus standards like NEMA MS 6-2008. There is no biological or diagnostic "ground truth" in the way an AI model would require.
8. The sample size for the training set
Not applicable. This is a physical device, not an AI model that requires a training set.
9. How the ground truth for the training set was established
Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
October 14, 2022
NeoCoil, LLC % Katie Gonzalez Ouality Systems & Regulatory Manager N27 W23910A Paul Road PEWAUKEE WI 53072
Re: K222407
Trade/Device Name: 16ch Breast Coil / 16ch Breast Coils, 1.5T 16ch Breast Coil, 3.0T 16ch Breast Coil, 1.5T 16ch Breast Coil, 3T 16ch Breast Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: September 9, 2022 Received: September 15, 2022
Dear Katie Gonzalez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name
16ch Breast Coils: 1.5T 16ch Breast Coil / 3.0T 16ch Breast Coil, 1.5T 16ch Breast Coil / 3T 16ch Breast Coil
Indications for Use (Describe)
The NeoCoil 16ch Breast Coils are Magnetic Resonance Imaging (MRI) RF Receive-Only Coils intended for use by trained medical professionals, in combination with and controlled by compatible 1.5T or 373.0T MRI system software.
The NeoCoil 16ch Breast Coils can be used with compatible ancillary components, accessories, and or devices to provide access to breast anatomy for diagnostic or interventional planning/procedures.
When used as intended, the 16ch Breast Coil provides information used by the MRI system to produce diagnostic and or interventional planning images of the breast, axilla, and chest wall. The images produced are interpreted by medical professionals as part of clinical decision making.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
N27 W23910A Paul Rd Pewaukee, WI 53072 USA Direct: (262) 347-1250 (262) 347-1251 Fax:
Image /page/3/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black circular shape on the left, with a white curved line inside that forms an arrow pointing to the right. To the right of the shape is the word "NeoCoil" in blue. The font is sans-serif and the letters are bold.
Special 510(k) Summary
Applicant
NeoCoil, LLC N27 W23910A Paul Rd Pewaukee, WI 53072 USA
Contact
Katie Gonzalez Quality Systems and Regulatory Manager 262-522-6124 (office) 262-347-1251 (fax) Katie.Gonzalez@neocoil.com
Preparation Date
August 3, 2022
Name of Device
- . Trade/Proprietary name(s):
16ch Breast Coils
| Submission | Device Trade/Proprietary Name |
|---|---|
| Existing model | 3.0T 16ch Breast Coil |
| Existing model | 1.5T 16ch Breast Coil |
| New model | 3T 16ch Breast Coil |
| New model | 1.5T 16ch Breast Coil |
- Magnetic Resonance Specialty Coil ● Common name:
- . Classification name: 21 CFR 892.1000, Magnetic resonance diagnostic device, Product Code MOS
Predicate Device
3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017, manufactured by NeoCoil, LLC
Reference Device
1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018, manufactured by NeoCoil, LLC
Device Description
The NeoCoil 3.0T 16ch Breast Coil is a phased array coil for imaging structures of the breast, axilla and chest wall. The 3.0T 16ch Breast Coil is a three part receive-only coil designed to provide high resolution imaging. The.3.0T 16ch Breast Coil includes a coil support structure, patient support structure, biopsy components, accessories and comfort pads.
{4}------------------------------------------------
N27 W23910A Paul Rd Pewaukee, WI 53072 USA Direct: (262) 347-1250 (262) 347-1251 Fax:
Image /page/4/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black circular shape with an arrow pointing to the right, followed by the word "NeoCoil" in blue. The "N" and "e" are in a lighter shade of blue, while the "o" is in a darker shade of blue. The "Coil" part of the word is in black.
The NeoCoil 3.0T 16ch Breast Coil is tuned to receive RF frequency corresponding to the proton precession in a 3 tesla magnetic field, which is governed by the Larmor equation.
The 16ch Breast Coils are intended for use in a manner that is identical to the predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018. The 16ch Breast Coil consists of the following arrays available in 1.5T, 3.0T and 3T field strengths:
- Medial Array, Breast Coil
- . Lateral Array Left, Breast Coil
- Lateral Array Right, Breast Coil ●
- . Baseplate Assembly, Breast Coil
- Biopsy Array Left, Breast Coil ●
- Biopsy Array Right, Breast Coil ●
The coils receive magnetic resonance signals generated in hydrogen nuclei (protons) in the Head. Neck and Brachial Plexus anatomy while blocking the high-frequency magnetic field applied by the MRI scanner at specified timings. The received signals are amplified before being transferred to the MRI scanner through the coil's system cable. The amplified signals are processed into tomographic images of the breast, axilla and chest wall anatomy by the MRI scanner. Images are typically generated as axial, sagittal, coronal oblique slices. Accessories associated with the 16ch Breast Coils include biopsy grids and biopsy drapes.
The 1.5T Breast Coil, the 3T 16ch Breast Coil, and the 3.0T 16ch Breast Coil are tuned to receive RF frequency corresponding to the proton precession in a 1.5 tesla and 3.0 tesla magnetic field (respectively), which is governed by the Larmor equation.
The 16ch Breast Coil is intended for use in a manner that is identical to the predicate device described in this submission.
Proposed labeling is documented in the Instructions for Use manual for the 16ch Breast Coil (NC149IFU-en).
Intended Use, including indications
The NeoCoil 16ch Breast Coils are Magnetic Resonance Imaging (MRI) RF Receive-Only Coils intended for use by trained medical professionals, in combination with and controlled by compatible 1.5T or 3T/3.0T MRI system software.
The NeoCoil 16ch Breast Coils can be used with compatible ancillary components, accessories, and/or devices to provide access to breast anatomy for diagnostic or interventional planning/procedures.
When used as intended, the 16ch Breast Coil provides information used by the MRI system to produce diagnostic and/or interventional planning images of the breast, axilla, and chest wall. The images produced are interpreted by medical professionals as part of clinical decision making.
Technological Characteristics
16ch Breast coils are similar in design, material, chemical composition and energy source to the legally marketed predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018.
{5}------------------------------------------------
N27 W23910A Paul Rd Pewaukee, WI 53072 USA Direct: (262) 347-1250 (262) 347-1251 Fax:
Image /page/5/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black circular shape with a white crescent moon shape inside, and a black arrow pointing to the right. To the right of the logo is the word "NeoCoil" in blue and black font. The word "Neo" is in blue, and the word "Coil" is in black.
At a high level, the 16ch Breast Coils included as part of this submission, and the predicate device, 3.0T 16ch Breast Coil (K173377) are based on the following same technological elements:
- Prescription use;
- . Coil designs are receive-only phased array coils;
- . Preamplification methodology
- Decoupling methodology
- Patient contacting materials and chemical composition are known materials that have ● been assessed for compliance with recognized biocompatibility standards;
- Energy source for the coils is the MRI scanner; ●
- Energy is not intended to be supplied by the coils; ●
- . Mechanical designs are contoured for the breast anatomy;
- Facilitates imaging for diagnostic and interventional planning of the breast, axilla and chest wall anatomy:
- Channel count
- Field strength when compared to the predicate device and reference device ●
The following technological differences exist between the subject and predicate and reference devices:
- MRI specific system cabling
- MRI specific interface circuitry .
- MRI specific RF loop and preamplifier tuning ●
- Addition of fuses on antenna elements ●
Testing performed per defined standards demonstrates that the safety and/or effectiveness of the 16ch Breast Coils compared to the predicate device is not adversely affected as a result of the differences.
Testing
Through NeoCoil's design controls process, a risk assessment leveraging FDA's quidance document decision trees for when to submit a new 510(k) for changes to an existing device, a summary of non-clinical performance data using well-established methodologies is included, referenced, or relied on to demonstrate that the 16ch Breast Coils are safe and effective and performs in a manner that demonstrates substantial equivalence to the predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018.
Performance testing - Bench:
A Test Report Summary for non-clinical Bench Testing performed, including testing to FDArecognized consensus standards identified as relevant in FDA guidance document Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices, issued November 18, 2016, is outlined below:
| TestPerformed | Objective(s) of theTest | Test MethodDescription | Pre-definedpass/failcriteria | ResultsSummary | Discussions /Conclusions |
|---|---|---|---|---|---|
| BiocompatibilityAssessment | Assess potentialbiological risks | Evaluation ofdata; historicaluse, biologictesting, wherewarranted | Acceptablelevel of risk | Pass | No identified significantrisks. |
| TestPerformed | Objective(s) of theTest | Test MethodDescription | Pre-definedpass/failcriteria | ResultsSummary | Discussions /Conclusions |
| ElectricalSafety | Basic electricalsafety/essentialperformance, 60601-1 | Test Lab | Pre-definedperformancestandards | Pass | Applicablerequirements for basicelectrical safety andessential performancemet. |
| ElectricalSafety | Particular electricalrequirements: MRequipment, 60601-2-33 | Test Lab | Pre-definedperformancestandards | Pass | Applicablerequirements of theparticular standardwere met. |
| ElectricalSafety | Collateral electricalsafety/essentialperformance, 60601-1-2 | Test Lab /Bench Testing | Pre-definedperformancestandards | Pass | Applicablerequirements of thecollateral standard weremet. |
| UsabilityAssessment | Devices meetcustomer, end userand patient needs | Actual.simulated orretrospectiveevaluation ofthe deviceand/or data | Pre-definedrequirements | Pass | The devices met theneeds of the customer,end user and patient. |
| Entrapment,Trapping Zoneand CableLooping(assessment w/scanner) | Assess the device forpinch points,entrapment, cablelooping - interfacingwith MRI scanner | Evaluation ofcoil-to-scannerentrapment,trapping andcable loopingnot covered bytest labassessments. | Requirementsbased on pre-definedrequirementsin 60601-1and customerrequirements | Pass | Requirements weremet. |
| SurfaceTemperature | Surface temperaturesdo not exceed limits | MRI scannertest | Pre-definedperformancestandards | Pass | Surface temperatureswere within IEC limits. |
| UnpluggedSurfaceTemperature | Devices remain safe infirst fault condition | MRI scannertest | Acceptablelevel of risk | Pass | Surface temperatureswere within IEC limitswhen the coil is leftunplugged in the MRIscanner. |
| BlockingNetworkAnalysis | Ensures devices aredesigned withadequate active andpassive transmitdecoupling | Theoreticalcalculations | Adequatetransmitdecoupling | Pass | Blocking networkdemonstrates adequateactive and passivetransmit decoupling. |
| B1 FieldDistortion | Measure amount ofdistortion produceddue to presence of anRF coil in the scanner | MRI scannertest | Pre-definedperformancestandards | Pass | B1 field inhomogeneitymeets performancerequirements anddemonstrates adequateactive and passivetransmit decoupling. |
| TestPerformed | Objective(s) of theTest | Test MethodDescription | Pre-definedpass/failcriteria | ResultsSummary | Discussions /Conclusions |
| B0 FiledDistortion | Measure amount ofdistortion produceddue to presence of anRF coil in the scanner | MRI scannertest | Pre-definedperformancestandards | Pass | B0 field inhomogeneitymeets performancerequirements anddemonstrates adequateactive and passivetransmit decoupling. |
| NEMA MS 6-2008 | Evaluate single-channel non-volumespecial purposeradiofrequency (RF)coils for use withmagnetic resonance(MR) imaging (MRI)systems | MRI scannertest | Pre-definedperformancestandards | Pass | SNR and ImageUniformity areconsistent with therequirements forindications for use. |
{6}------------------------------------------------
N27 W23910A Paul Rd Pewaukee, WI 53072 USA Direct: (262) 347-1250 Fax: (262) 347-1251
Image /page/6/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black circular shape with an arrow pointing to the right, followed by the word "NeoCoil" in blue. The word "Neo" is in a slightly larger font than the word "Coil".
{7}------------------------------------------------
N27 W23910A Paul Rd Pewaukee, WI 53072 USA Direct: (262) 347-1250 (262) 347-1251 Fax:
Image /page/7/Picture/1 description: The image contains the logo for NeoCoil. To the left of the company name is a black circular logo with an arrow pointing to the right. The company name is to the right of the logo, with "Neo" in blue and "Coil" in black.
Performance testing - Clinical:
Due to the substantial equivalency of the devices and the evaluation of performance testing against defined standards, no additional clinical performance testing has been performed above that present for the predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018.
No adverse events have been reported against the predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018.
Conclusion
This submission demonstrates by means of nonclinical testing that the 16ch Breast Coils are substantially equivalent to and perform as well as the predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018.
§ 892.1000 Magnetic resonance diagnostic device.
(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.