(269 days)
The swiftPro™ System is a surface based device intended for the coagulation of soft tissue during non-invasive procedures.
The swiftProTM System is not indicated for use in cardiac procedures.
The swiftPro® system is a hand-held microwave generator intended to provide the functionality of the Swift® microwave treatment system in a compact and portable package. The system comprises a hand-held microwave generator and an applicator tip assembly. The generator produces microwave energy at a frequency of 8 GHz with a maximum output power of 6 Watts, the output power is restricted depending on the applicator tip attached, which delivers the microwave energy to the tissue to effect thermal heating, The swiftPro® System interface is intuitive with a quick setup for setting treatment; output power level & treatment duration (seconds). The applicator tip and the handheld generator have been designed to be only connected in a certain way with a single, simple connection mechanism. This prevents the possibility of system malfunction due to incorrect setup or confusion. The SwiftPro® applicator tips have been designed to prevent re-use to mitigate the risk of cross-contamination.
The SwiftPro® system compromises of the following components:
- SwiftPro® Handheld Generator
- SwiftPro® Cradle
- SwiftPro® URT Applicator tip (non-sterile, single use)
- SwiftPro® LRT Applicator tip (non-sterile, single use)
The provided text describes a medical device called the swiftPro™ System and its comparison to a predicate device, the Swift® System, to demonstrate substantial equivalence for FDA clearance. However, the document does not contain information about acceptance criteria, reported device performance in the context of clinical or diagnostic accuracy, sample sizes for test sets, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, or training set details as typically apply to AI/ML powered devices.
The document focuses on the engineering and safety aspects of an electrosurgical cutting and coagulation device. The "performance data" section refers to:
- Biocompatibility testing
- Electrical safety and electromagnetic compatibility (EMC)
- Software verification and validation testing
- Cybersecurity
- Human Factors and Usability Engineering
- Bench Testing (thermal effects analysis in ex-vivo porcine skin, bovine liver, porcine kidney, and porcine muscle)
These are technical performance characteristics of the device itself and not related to diagnostic accuracy or clinical effectiveness in a way that would require the typical AI/ML study details requested in the prompt.
Therefore, I cannot populate the requested table and information based on the provided text, as the nature of the device and its evaluation for FDA clearance (as described here) do not involve those types of criteria or studies. The device is for "coagulation of soft tissue," which is a direct physical action, not a diagnostic or AI-assisted interpretation task.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 4, 2023
Emblation Ltd. % Angela Blackwell Senior Consultant Blackwell Device Consulting P.O. Box 718 Gresham, Oregon 97030-0172
Re: K222388
Trade/Device Name: swiftPro System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: NEY Dated: April 3, 2023 Received: April 4, 2023
Dear Angela Blackwell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2023.05.04 08:54:38 -04'00'
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K222388
Device Name swiftPro™ System
Indications for Use (Describe)
The swiftPro™ System is a surface based device intended for the coagulation of soft tissue during non-invasive procedures.
The swiftProTM System is not indicated for use in cardiac procedures.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!"
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "emblation". The logo consists of a red oval shape with a white and black wave-like design inside it. To the right of the oval is the word "emblation" in black letters, with the "em" in red.
510(k) Summary
The 510(k) summary of safety and effectiveness information is submitted as part of the Premarket Notification in compliance with requirements of 21 CFR Part 807.92.
I. SUBMITTER
Emblation Limited Forrester Lodge. Inalewood Alloa, FK10 2HU, UK Phone: +44(0) 1259 236 132 Fax: +44(0) 1259 210 430
Contact Person: Mairi MacFayden Chief Requlatory Officer
Date Prepared: May 2, 2023
II. DEVICE
| Name of Device: | swiftPro® System |
|---|---|
| Common Name: | Microwave ablation system and accessories |
| Classification Name: | Electrosurgical Cutting & Coagulation Device and Accessories, 21 CFR 878.4400 |
| Regulatory Class | II |
| Product Code: | NEY |
III. PREDICATE DEVICE
| Trade Name: | Swift® System |
|---|---|
| Common Name: | Microwave ablation system and accessories |
| 510(k) Number: | K181941 |
| Manufacturer: | Emblation Limited |
| Regulatory Class: | II |
| Product Code: | NEY |
This predicate has not been subject to a design-related recall.
IV. DEVICE DESCRIPTION
The swiftPro® system is the subject of this application and is substantially equivalent to the device described and cleared in K181941, the subject device is an all-encompassing version of the Swift® System.
The swiftPro® system is a hand-held microwave generator intended to provide the functionality of the Swift® microwave treatment system in a compact and portable package. The system comprises a hand-held microwave generator and an applicator tip assembly. The generator produces microwave energy at a frequency of 8 GHz with a maximum output power of 6 Watts, the output power is restricted depending on the applicator tip attached, which delivers the microwave energy to the tissue to effect thermal heating, The swiftPro® System interface is intuitive with a quick setup for setting treatment; output power level & treatment duration (seconds). The applicator tip and the handheld generator have been designed to be only connected in a certain way with a single, simple connection mechanism. This prevents the possibility of system malfunction due to incorrect setup or confusion. The SwiftPro® applicator tips have been designed to prevent re-use to mitigate the risk of cross-contamination.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "emblation". The logo consists of a red oval shape on the left side, with a stylized "e" inside. The "e" is made up of a white and black line. To the right of the oval is the word "emblation" in black font.
The SwiftPro® system compromises of the following components:
- SwiftPro® Handheld Generator
- · SwiftPro® Cradle
- · SwiftPro® URT Applicator tip (non-sterile, single use)
- · SwiftPro® LRT Applicator tip (non-sterile, single use)
V. INDICATIONS FOR USE
The swiftPro® System is a surface-based device intended for the coagulation of soft tissue during noninvasive procedures.
The swiftPro® System is not indicated for use in cardiac procedures.
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Microwave coagulation is the technological principle for both the swiftPro® System and predicate device. It is based on the use of microwave energy for the coagulation of soft tissue, where the user has the ability to select the desired power and time limits based on size of the target area.
The technological characteristics are the same regarding frequency and magnitude of microwave energy delivered to a target tissue through the existing applicator/antenna design as used in the existing Swift® System (predicate) in a controlled manner for the purpose of coagulation of soft tissue. The main differences relate to the addition of FPGA firmware and a software user interface which replicates the core experience (treatment power/time settings and a single button to enable/disable energy deposition) of using the existing Swift device.
The operation of the swiftPro® user interface is controlled via device-embedded software, responsible for handling user inputs and providing system status indications. Microwave control and functionality and system safety is provided by firmware. Both the application software and firmware are installed on devices located on the System Controller Module (microcontroller IC and FPGA respectively). However, both the subject device and predicate have hardware or firmware (FPGA) based safety controls, the addition of software controls relates purely to improve user experience.
The predicate device operation is controlled by a digital logic state machine which, by its nature, requires a high number of components to implement for this approach to be capable of providing the level of functionality required. This component count requires a number of large PCBAs - and interconnecting components - to support, and therefore results in a larqe system footprint and high cost.
The swiftPro® system replicates this functionality through use of embedded software, significantly reducing the system real estate consumed by PCBAs and their peripherals and allows for a high degree of compact integration in the system architecture. The inclusion of software within the design also permits the system functionality to be enhanced, and results in further streamlining of the system by permitting a graphical user interface (GUI) to be implemented, removing the need for discrete LED indicators as present in the Swift® system. The GUI has been designed to replicate the experience of using the existing swift® device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "emblation". The logo consists of a red oval shape with a white and black wave-like design inside, followed by the word "emblation" in red and black letters. The "em" part of the word is in red, while the "blation" part is in black.
SwiftPro® Handheld Generator Treatment System
The swiftPro® handheld generator is constructed from control PC-ABSs which are the same biocompatible materials as the applicator tip with an integrated user interface. Includes functional indicators, display auditory feedback, and programmable system control capability. The handheld generates microwave power up to 6W using hardware controlled internally requlated microwave source. The output signal from the SwiftPro™ generator is provided by the integrated user interface. The output power can be varied from 1W increments to 6W using pulse width modulation (PWM) techniques. This provides regulated microwave energy at a central frequency of 8GHz and provides real-time reflected power monitoring which ensures that the treatment is progressing as normal and that the handheld generator is performing as expected.
The swiftPro® System does not require a handpiece or cable to operate as it is an all-encompassing device and this is the main reason for the device weighing less and smaller than the predicate device. The design is different from the predicate device but the differences do not affect the fundamental principles of technology.
SwiftPro® Applicator Tip
The swiftPro® Applicator tip is identical to the predicate in design and materials used. There are no technological differences. The swiftPro® Applicator Tips are mechanically, physically, and electricaly identical to the Swift® SWF-AT01 Applicator Tip, differing only in the value of resistor fitted to their (internal) PCBA which denotes the Tip DUID.
| DeviceFeature | swiftPro® System | Swift® System(Predicate) | Comments |
|---|---|---|---|
| Classification Regulation21 CFR 878.4400 | 21 CFR 878.4400 | 21 CFR 878.4400 | Same |
| Product Code | NEY | NEY | Same |
| Indicationsfor Use | The swiftPro® System is a surface-based device intended for thecoagulation of soft tissue during non-invasive procedures.The swiftPro™ System is notindicated for use in cardiacprocedures. | The Swift® System is a surface-baseddevice intended for the coagulation ofsoft tissue during non-invasiveprocedures.The Swift® System is not indicatedfor use in cardiac procedures. | Same |
| DimensionsWidth x Height x Depth | 245x75x135mm(10x3x5.3 inch) | 300x90x300mm(11.8x3.5x11.8 inch) | Different, the subjectdevice is smaller |
| Weight | Less than 0.75kg (~1.65lbs) | Less than 4.5kg (~10 lbs) | Different, the subjectdevice is lighter |
| Material | Enclosure Material:Handpiece - ABS/PC CYCOLOY™Resin HC1204HFTip material: ABS/PC CYCOLOY™Resin HC1204HFBiobarrier material – SILPURAN®6000/70 Liquid Silicone Rubber | Enclosure Material:Handpiece - ABS/PC CYCOLOY™Resin HC1204HFGenerator - AluminiumInterconnect Cable- Silicone RubberTEKBOND TB1723Tip material - ABS/PC CYCOLOY™Resin HC1204HFBiobarrier material - SILPURAN®6000/70 Liquid Silicone Rubber | Different but as detailedwhere there is analogybetween the devicesthey are constructedfrom the same material |
| Power Supply | 24Vdc | 90-645Vac | Different, the subjectdevice requires lowerpower and can use alow voltage dc adaptor |
| Generator | Yes | Yes | Same |
| Energy | Microwave | Microwave | Same |
| Microwave OutputFrequency | 8 GHz | 8 GHz | Same |
| Maximum MicrowavePower | 0-10 in 1power level increments =treatment,6W = GeneratorTip ATO2 restricts delivered energy to | 0-10W in 1W increments =treatment,20W = GeneratorTip ATO1 restricts delivered energy to | Different, but asdetailed swiftPro®performs within theequivalent AT01(predicate) output |
| level setting on generator of 10) | generator setting) | range. | |
| Time settings | 0-10 secs | 0-10 secs | Same |
| Coagulation cycle | Maximum 15 minutes but treatmentdependant & limited by applicatortips (AT02 and AT03). | Maximum 15 mins but treatmentdependant & limited by applicator(AT01). | Different, two tip (AT02and AT03) options tosegregate energydelivery. |
| Shut Offs/Alarms | Alerts:MSM overtemp, Expired tip, Noapplicator detected, Reflection,Manual Switch Status System error,SD Card Error, Service error | Alerts:MSM overtemp, Expired tip, Noapplicator detected, Reflection,Continuous wave/SweptManual Switch Status System error | Existing alerts &descriptive Icons havebeen reused. Additionalalerts have beenincorporated intoswiftPro® System forexpanded functionality. |
| Software/Firmware Platform | Yes | No | Addition of FPGAfirmware and asoftware user interface |
| Hardware control | Yes | Yes | Same, implemented inFPGA firmware |
| Monitored parameters | Power, Time, Reuse status,Applicator connectedReflected power | Power, Time, Reuse status,Applicator connectedReflected power | Expanded parametersbased on additionalsystem functions. |
| Display parameters | Set Power (1-10 units)Set Time/Treatment Time (secs),Settings MenuAlertsDevice InformationScreen Brightness SelectionAuto/Manual Mode SelectionSet Time and DateRecall Treatment Menu | Set Power (1 -10 Watts)Set Time/Treatment Time(mins/secs),Actual power delivered (% of set),Alerts | Different due to powersetting unit change buthaving equivalentoutput.As with the predicatedevice, the systemmonitors its ownapplied and reflectedpower and will displayan alert in the event ofa trip. |
| Manual or automaticsetting | Manual orAutomatic can be selected. | Manual orAutomatic, via (footswitch) can beselected. | Same, excludesfootswitch as was neverutilised in the market |
| Operation Mode | PWM & fixed frequency | PWM/& Swept | Different due to fixedfrequency |
| Surface Applicator/Antenna | SurfaceApplicator | SurfaceApplicator | Same |
| Patient contact partsupplied sterile: | Non-steriledisposable applicator tip | Non-steriledisposable applicator tip | Same |
| Applicator/Antenna Design | Ceramic waveguideDisposable silicone barrier | Ceramic waveguideDisposable silicone barrier | Same |
| Applicator/Antenna head Size | Applicator Tip approximately 4 cm x 3cmSilicone/Ceramic contact face6.7 mm diameter | Applicator Tip approximately 4 cm x 3cmSilicone/Ceramic contact face6.7 mm diameterCable constructed of coaxial cableHandpiece approximately 12 cm x 3cm | Same Applicator/Antenna head size butsystem no longerincorporates a separatehandpiece andinterconnect cable |
| Max applicator surfaceTemp (non-activeareas) | 42°C | 42°C | Same |
| Operation Procedure | Surface use | Surface use | Same |
| Average energy densityof surface application(Area J/mm2) | 10 J/mm2Set power level 10Set time 1 Seconds6J delivered | 10 J/mm2Set power 10WSet time 1 Seconds10J set, 6J delivered | Same (in terms ofenergy delivered) |
Comparison Table
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the word "emblation" with the "em" in red and the "blation" in black. To the left of the word is a red circle with a black and white design inside. The design is a curved line that starts at the top of the circle and ends at the bottom.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for "emblation". The logo consists of a red oval shape on the left side, with a white and black curved line running through it. To the right of the oval is the word "emblation" in red and black lettering. The "embla" part of the word is in red, while the "tion" part is in black.
VII. PERFORMANCE DATA
The following performance data were provided in support of substantial equivalence determination, using the either the same or similar methods and information used for the predicate device.
Biocompatibility testing
The applicator tips were subject to biocompatibility testing, the swiftPro® System uses the same type of applicator tip as the predicate and the following Biocompatibility tests were completed in accordance with ISO10993: Biological Evaluation of Medical Devices Part 1: Evaluation and testing within a risk management process:
- · Cytotoxicity
- · Sensitisation
- · Irritation
- · Systemic toxicity
- · Pvrogenicitv
Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted and the swiftPro® system complies with all the applicable Medical electrical equipment standards for safety and essential performance.
- IEC 60601-1:2005 + AMD1:2012+AMD2:2020 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-11:2015 Medical electrical equipment – Part1-11: General requirements for basic safety and essential performance – Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- IEC 60601-2-6: 2012 + AMD1:2016 Medical electric equipment Part 2: Particular reguirements for basic safety and essential performance of microwave therapy equipment.
- IEC 60601-1-2:2014 +AMD1:2020 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances-Requirements and tests.
Software Verification and Validation Testing
The swiftPro® System includes embedded software which has been developed for use with an STM32 device, which incorporates a 32-bit ARM Cortex-M4 microcontroller. The embedded software produced for the swiftPro™ System includes source code, object files, executable load images, and all software documentation required for compliance with IEC 62304:2006 (AMD1: 2015) and FDA quidance for software contained in medical devices.
Cybersecurity
The swiftPro® system is not intended to be used as a networked medical device, nor does it contain any offthe-shelf (OTS) software to support connection to a private or public internet. The device does not incorporate an OTS operating system or OTS software/drivers which supported hard-wired or wireless network connections. Throughout the lifecycle of the product, an assessment to mitigate cybersecurity vulnerabilities has been completed. The following have been documented:
- Software Bill of Materials ●
- Threat Modelling ●
- . Total Product Lifecycle
- Security Architecture
- . Cybersecurity Testing
- . Labelling
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image is a logo for a company called "emblation". The logo consists of a red oval shape with a stylized "e" inside of it. The "e" is made up of two curved lines, one black and one white. The word "emblation" is written in black letters to the right of the oval shape. The font is sans-serif and the letters are all lowercase.
Human Factors and Usability Engineering
The swiftPro® system complies with all the applicable reguirements of the following standards:
- ANSI/ AAMI HE75:2009/R2013 American National Standard / Association for the Advancement of Medical Instrumentation. Human Factors Engineering - Design of Medical Devices
- · IEC 60601-1-6:2010+AMD1:2013 +AMD2:2020, Medical electrical equipment General requirements for basic safety and essential performance. Collateral standard: Usability
- · IEC 62366-1:2015+AMD1:2020 Medical Devices. Application of Usability Engineering to Medical Devices
- · FDA-2011-D-0469 Applying Human Factors and Usability Engineering to Medical Devices.
Bench Testing
Equivalence bench testing, including ex-vivo, within a common subset of settings has been performed to assess the common performance equivalence of the swiftPro® handheld system in comparison to the predicate device (K181941). The bench testing proves the swiftPro® System performs substantially equivalent to the predicate device.
- Thermal effects analysis in ex-vivo porcine skin .
- Thermal effects analysis in ex-vivo bovine liver, porcine kidney and porcine muscle ●
VIII. CONCLUSIONS
The difference between the subject device (swiftPro® System) and the predicate device (Swift® System) technological characteristics do not raise any new or different questions regarding safety and effectiveness. The performance testing demonstrates the subject device is as safe and effective as the predicate device as indicated for use.
§ 878.4400 Electrosurgical cutting and coagulation device and accessories.
(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.