K Number
K222377
Date Cleared
2022-11-03

(90 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

LED MASK PLATINUM MD is an over the counter device that is indicated for the treatment of full face wrinkles.

Device Description

This product irradiates the skin with red light in the wavelength range of 630 to 690 nm and infrared light in the wavelength range of 820 to 880 nm. Treatment of wrinkles all over the face. This product is designed as a portable personal medical device with a built-in battery, not a wired power supply. It is a mask-type product that is worn on the face. The operating mode of the product can be easily changed by hand. After connecting the charging adapter to the cradle, attach the main unit to the cradle to wirelessly charge the main unit with only the cradle. In addition, eye shield (for eye protection, opening/closing covers, vacuum cleaner towels, power cables, etc. are provided for convenient service use.

AI/ML Overview

This document is a 510(k) summary for the LED MASK PLATINUM MD, a device indicated for the treatment of full face wrinkles. The submission aims to demonstrate substantial equivalence to legally marketed predicate devices.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the comparison to predicate devices, particularly K180856, and by conforming to established safety standards for LED-based devices. The "reported device performance" section focuses on technical specifications and safety compliance rather than direct clinical efficacy metrics.

Acceptance Criteria (Implied)Reported Device Performance (LED MASK PLATINUM MD)
Intended Use: Treatment of full face wrinklesIndicated for the treatment of full face wrinkles.
Type of Use: Over-the-Counter (OTC)OTC
Device Configuration: Mask typeMask type
Wavelengths: Within acceptable range for wrinkle treatment630-690 nm (red light), 820-880 nm (infrared light)
Power (Irradiance): Within safe and effective range. Specifically, within 1.32 mW/cm2 to 62 mW/cm2 based on cleared OTC devices.2 mW/cm2
Treatment Time: Comparable to predicate devices.8 minutes
Energy Dose: Comparable to predicate devices.0.96 J/cm2
Safety: Compliance with relevant electrical, radiation, and risk management standards.ES 60601-1 (Basic safety & essential performance), IEC 60601-1-2 (EMC), IEC 62471 (Photobiological safety), IEC 60601-2-57 (Safety of laser product), ISO 14971 (Risk management), IEC 60601-1-6 (Usability), IEC 60601-1-11 (Home healthcare environment)
Effectiveness: Substantial equivalence to predicate device with similar intended use.Claims substantial equivalence to predicate devices (e.g., K180856).

2. Sample Size Used for the Test Set and Data Provenance

The document does not describe a clinical study with a "test set" in the traditional sense of evaluating device efficacy on patients. The performance data presented relates to non-clinical tests (measurement of wavelength, output power, total irradiance, and safety compliance). Therefore, no sample size for a test set or data provenance from patient data is applicable or reported for this type of submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. As no clinical "test set" data is presented for efficacy, there is no mention of experts establishing ground truth for such a set. The expertise involved would be in the engineering and safety testing domains, not clinical adjudication.

4. Adjudication Method for the Test Set

Not applicable. No clinical test set requiring adjudication is described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an LED mask for wrinkle treatment, not an AI-assisted diagnostic tool. An MRMC study is not relevant to this device's regulatory pathway.

6. If a Standalone (i.e. algorithm only, without human-in-the-loop performance) was done

Not applicable. This is not an algorithmic device. The device's performance is standalone in the sense that it directly emits light for treatment, not that it's an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance is established through physical measurements and compliance with recognized consensus standards.

  • For technical specifications (wavelength, power, energy dose): The ground truth is the direct measurement of these physical properties using calibrated equipment.
  • For safety: The ground truth is established by demonstrating compliance with the referenced IEC and ISO standards (e.g., IEC 60601-1 for basic safety, IEC 62471 for photobiological safety, ISO 14971 for risk management).
  • For substantial equivalence: The ground truth is the regulatory determination that the device's technical characteristics, intended use, and performance are sufficiently similar to a legally marketed predicate device to be considered equally safe and effective.

8. The Sample Size for the Training Set

Not applicable. This device does not involve a "training set" as it is not an AI/machine learning device.

9. How the Ground Truth for the Training Set Was Established

Not applicable. No training set is described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

November 3, 2022

CellReturn Co., LTD. % Do-Hyun Kim CEO BT Solutions, Inc. Unit 904, Eonju-ro 86-gil 5, Gangnam-gu Seoul, Seoul 06210 Korea. South

Re: K222377

Trade/Device Name: Led Mask Platinum Md Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: August 1, 2022 Received: August 5, 2022

Dear Do-Hyun Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jianting Wang, PhD Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222377

Device Name

LED MASK PLATINUM MD

Indications for Use (Describe)

LED MASK PLATINUM MD is an over the counter device that is indicated for the treatment of full face wrinkles.

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

LED MASK PLATINUM MD 510(k) Summary

5. 510(k) Summary

5.1 General Information

Applicant/Submitter:Cellreturn Co.,Ltd.
Address:237, Namdongseo-ro, Namdong-gu,Incheon, Republic of Korea, 21634
Contact Person:Do Hyun Kim, BT Solutions, Inc.
Address:Unit 904, Eonju-ro 86gil 5,Gangnam-gu, Seoul 06210, Korea.Tel: +82-2-538-9140Email: ceo@btsolutions.co.kr
Preparation Date:August 4, 2022

5.2 Device Name and Code

Device Trade Name:LED MASK PLATINUM MD
Model Name:MQ-M21M4
Classification Name:Light Based Over The Counter Wrinkle Reduction
Product Code:OHS
Regulation Number:21 CFR 878.4810
Classification:Class II
Review Panel:General & Plastic Surgery

5.3 Technical Characteristics in Comparison to Predicate Devices

The LED MASK PLATINUM MD, is substantially equivalent to the following legally marketed predicate devices

Proposed DeviceK180856
CompanyLED Mask Platinum MDNeutrogena Light TherapyAging Mask+
Product nameCellreturn Co.,Ltd.Johnson & Johnson Consumer Inc.
Product codeOHSOHS
Regulationnumber21 CFR 878.481021 CFR 878.4810
ClassificationClass IIClass II
Intended UseLED MASK PLATINUM MD isan over the counter device that isindicated for the treatment of fullface wrinkles.The Neutrogena Light TherapyAging Mask+ is an over thecounter device that is indicated

{4}------------------------------------------------

510(k) Summary

for the treatment of full facewrinkles.
Type of useOTCOTC
Technologicalcharacteristics
Shape of deviceMask typeMask type
Wavelength630-690 nm820-880 nm620-640 nm820-880 nm
Power(mW/cm2)21.32
Treatment time(min)810
Energy dose(J/cm2)0.960.79

Irradiance of already cleared OTC device that is intended for the treatment of full face wrinkles ranges from 1.32 mW/cm2 (K180856) to 62 mW/cm2 (K133896), which means irradiance that lies between these two values can be regarded as "safe" to use for the intended purpose.

5.4 Device Description

This product irradiates the skin with red light in the wavelength range of 630 to 690 nm and infrared light in the wavelength range of 820 to 880 nm. Treatment of wrinkles all over the face.

This product is designed as a portable personal medical device with a built-in battery, not a wired power supply. It is a mask-type product that is worn on the face. The operating mode of the product can be easily changed by hand. After connecting the charging adapter to the cradle, attach the main unit to the cradle to wirelessly charge the main unit with only the cradle. In addition, eye shield (for eye protection, opening/closing covers, vacuum cleaner towels, power cables, etc. are provided for convenient service use.

5.5 Indications / Intended Use

LED MASK PLATINUM MD is an over the counter device that is indicated for the treatment of full face wrinkles.

6.6 Performance Data

tests: Measurement of wavelength, average output power, Non-clinical and density, in in units of of J/cm2) of treatment total irradiance (power LEDs performance, such as safety of laser product, were performed. Other electromagnetic compatibility and electrical safety, etc, were tested using following consensus the

Basic safety and essential performance of the LED MASK PLATINUM MD is tested and evaluated according to the FDA-recognized consensus standard, ES 60601-1. Effect to the device by electromagnetic disturbances were tested and evaluated according to the FDA-recognized consensus standard IEC 60601-1-2.

{5}------------------------------------------------

510(k) Summary

  • -Safety of laser product is evaluated in accordance with IEC 62471, and IEC 60601-2-57.
  • Risk management was recorded by referring to ISO 14971. -
  • -Usability was documented by referring to IEC 60601-1-6.
  • -Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment were tested with IEC 60601-1-11.

5.7 Substantial Equivalence

The proposed device uses similar or identical technology as the predicate devices and has same intended uses. Based upon the predicted overall performance characteristics for LED MASK PLATINUM MD , Cellreturn Co.,Ltd.. believes that no significant differences in usage of its underlying technological principles between LED MASK PLATINUM MD and the predicate device.

5.8 Conclusions

On the basis of the information provided in this summary, Cellreturn Co.,Ltd. believes that LED MASK PLATINUM MD is substantially equivalent to legally commercialized predicate devices for the purposes of this 510 (k) submission.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.