K Number
K133896

Validate with FDA (Live)

Device Name
REJUVALITE MD
Manufacturer
Date Cleared
2014-11-13

(328 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Rejuvalite MD is an Over the Counter device that is intended for the treatment of full face wrinkles.

Device Description

The Rejuvalite MD, is a tabletop device that contains 120 LEDs split evenly into 4 different wavelengths. The total power density for combined wavelengths is approximately 62 mW/cm2 at a distance of 4" from the illuminating surface of the device. The labeling requires Users to maintain the distance between the illuminating surface and the treatment site at 4"during the treatment. There is a 4" distance limiting arm attached to the device to help users on determining and maintaining the 4" distance between the device's illuminating surface and the treatment site. The Rejuvalite MD system components include the tabletop unit consisting of the base, arm, and swivel head containing the LED module , 4" distance limiting arm, timer, internal UL approved power adapter, and goggles. The device incorporates a power adapter mounted inside of the lamp base that converts the AC to DC voltage. The power switch is operated to turn the device on or off. The unit is designed to be placed on a table for easy exposure to the face to ensure consistent administration of light during each treatment. The device does not contain any user serviceable components. The device is sold as Over the Counter (OTC).

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Rejuvalite MD device. Please note that the document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a de novo clinical study with strict acceptance criteria for a novel device. Therefore, some of the requested information, particularly regarding detailed clinical acceptance metrics and their achievement, may not be explicitly stated in the context of this type of submission.

1. Table of Acceptance Criteria and Reported Device Performance

Based on the document, the "acceptance criteria" are implied by the demonstration of substantial equivalence to the predicate device, particularly regarding safety, effectiveness, and key technical specifications. There are no explicit quantitative clinical performance acceptance criteria listed.

Acceptance Criterion (Implied by Substantial Equivalence)Reported Device Performance (Rejuvalite MD)
Indications for UseIntended for the treatment of full face wrinkles (matches predicate).
Treatment Regime3 minutes daily, 5 days per week for 8 weeks (identical to predicate).
Power Density per WavelengthPower density per wavelength (mW/cm²) at 4" distance is identical to the predicate device.
Total Power Density of Combined Wavelengths62 mW/cm² at a distance of 4" from the LED head (identical to predicate's 62 mW/cm² in contact).
Wavelengths600, 622, 660, 860nm (similar to predicate's 605, 630, 660, 855nm).
Energy Source (LEDs)120 LEDs, over 90cm² (predicate: 72 LEDs, over 40cm²). While the number of LEDs and surface area differ, the submission claims the "amount of energy (irradiance per area) delivered to the skin measured at 4" are identical to the predicate device."
Safety - Electrical & OpticalConforms to IEC 60601-1 (general requirements for basic safety and essential performance) and IEC 60601-1-2 (EMC). Performance testing on optical and electrical characteristics demonstrated it is "almost identical" to the predicate.
Safety - BiocompatibilityDevice body made of ABS plastic (same as predicate), considered safe for skin contact. Conclusion: "does not raise any new safety issues."
Usability/Label Comprehension97% of participants comprehended the labeling. 100% of participants used the device successfully. The device is designed for layman understanding and use (OTC).
Adverse EventsNo adverse effects found for the predicate device in MAUDE since clearance in 2005 (implies a similarly low risk profile for the applicant device given substantial equivalence).
Operational MethodOn/Off switch, no software (similar to predicate). Tabletop device used at 4" distance (differs from predicate's handheld, in-contact use, but power density equivalent is maintained by design).

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Description: The primary "test set" described is for a Usability/Label Comprehension Study.
  • Sample Size: 36 participants.
  • Data Provenance: The document does not explicitly state the country of origin or whether the study was retrospective or prospective. Given it's a usability study for an OTC device being submitted for FDA clearance, it would almost certainly be a prospective study conducted within a population relevant to the U.S. market.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This type of information is generally not applicable to a usability/label comprehension study. The "ground truth" for this study is essentially the participants' ability to understand instructions and use the device. There's no clinical "ground truth" (e.g., presence of a disease) being established by experts in this context.

4. Adjudication Method for the Test Set

Not applicable to a usability/label comprehension study as described. The outcomes (comprehension percentage, successful use percentage) are direct observations of participant interaction.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

  • No, an MRMC comparative effectiveness study was not done.
  • This device is a light therapy device for wrinkles, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and "human readers improve with AI vs without AI assistance" are not relevant to this submission.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Again, this question is not applicable to
the Rejuvalite MD device. It is a physical light therapy device, not an algorithm.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

For the usability study, the "ground truth" was directly observed user performance and self-reported comprehension. For the broader claim of effectiveness for wrinkle treatment, the device relies on substantial equivalence to a predicate device (LightStim for Wrinkles, K120775) which was presumably established through its own clinical data or literature at the time of its clearance. The Rejuvalite MD itself did not perform independent clinical outcomes data to establish de novo effectiveness.

8. The Sample Size for the Training Set

Not applicable. The Rejuvalite MD is a physical device, not an AI/algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of three faces in profile, stacked on top of each other. The faces are stylized and appear to be made of flowing lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 13, 2014

Trophy Skin Incorporated % Ms. Susan R. Anthoney-DeWet AEGIS Regulatory Incorporated 2424 Dempster Drive Coralville, Iowa 52241

Re: K133896 Trade/Device Name: Rejuvalite MD Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHS Dated: October 6, 2014 Received: October 7, 2014

Dear Ms. Anthoney-DeWet:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

for

Sincerely yours,

David Krause -S

  • Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K133896

Device Name Rejuvalite MD

Indications for Use (Describe)

The Rejuvalite MD is an Over the Counter device that is intended for the treatment of full face wrinkles.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K133896

This summary of 510(k) summary information is being submitted in accordance with the requirements of 21 CFR § 878.4810.

Submission Date: December 11, 2013

1. Submitter Information:AEGIS Regulatory, Inc. - Robert T. Wagner2424 Dempster DriveCoralville, IA 52241Tel.: 865-982-5552Email: bob@fdalistingconsultants.com
For Sponsor:Trophy Skin, Inc.Attn: Imran Karim4372 Kenmare TrailFrisco, TX 75034Tel.: 469-698-6270Fax: 972-692-7194Email: trophyskin@gmail.com

2. General Information

  • 2.1 Classification Name: Light Based Over-The-Counter Wrinkle Reduction Device
  • 2.2 Common/Usual Name: Rejuvalite MD
  • 2.3 Proprietary Names: Rejuvalite MD
  • 2.4 Classification: Class II
  • 2.5 Classification Number: 878.4810
  • 2.6 Product Code: OHS
  • 2.7 Review Panel: General & Plastic Surgery

{4}------------------------------------------------

3. Device Description:

The Rejuvalite MD, is a tabletop device that contains 120 LEDs split evenly into 4 different wavelengths. The total power density for combined wavelengths is approximately 62 mW/cm2 at a distance of 4" from the illuminating surface of the device.

The labeling requires Users to maintain the distance between the illuminating surface and the treatment site at 4"during the treatment. There is a 4" distance limiting arm attached to the device to help users on determining and maintaining the 4" distance between the device's illuminating surface and the treatment site.

The Rejuvalite MD system components include the tabletop unit consisting of the base, arm, and swivel head containing the LED module , 4" distance limiting arm, timer, internal UL approved power adapter, and goggles.

The device incorporates a power adapter mounted inside of the lamp base that converts the AC to DC voltage. The power switch is operated to turn the device on or off.

The unit is designed to be placed on a table for easy exposure to the face to ensure consistent administration of light during each treatment. The device does not contain any user serviceable components. The device is sold as Over the Counter (OTC).

4. Indications / Intended Use:

The RejuvaliteMD is an Over -the- Counter device that is intended for the use in the

treatment of full face wrinkles.

Rx or OTC:

The Rejuvalite MD device is an Over the Counter (OTC) device. The labeling, instructions, and User Operations (21 CFR § 801.60 and 61), are designed for layman understanding and use. The predicate device is OTC.

5. Predicate Device (s):

This device is substantially equivalent to the following predicates, which are currently in safe

and effective commerce:

1. K120775 - LightStim for Wrinkles (LED Intellectual)

{5}------------------------------------------------

Predicate Comparison Chart

DeviceRejuvalite MDLightStim forWrinkles
Trophy SkinLED Intellectual
K133896K120775
This SubmissionA Predicate Device
IndicationsThe RejuvaliteMDis an Over theCounter device thatis intended for theuse in the treatmentof full face wrinkles.The Lightstim forWrinkles is an OTChand-held deviceintended for the usein the treatment offull-face wrinkles
HandheldTable topYes
Wavelengths600,622,660,860nm605,630,660,855nm
ModesOn/OffOn/Off
Energy Source120 LEDs. Over90cm272 LEDs. Over40cm2
Total PowerDensity62 mW/cm2 at adistance of 4" fromthe LED head62 mW/cm2 incontact with LEDhead
Power SupplyAC to DC9-volt DC powertransformer
TreatmentRegime3 minutes daily, 5days per week for 8weeks3 minutes daily, 5days per week for 8weeks
TargetPopulationIndividuals withwrinkles on the faceIndividuals withwrinkles on the face
Location forUseOTCOTC

Summary of the technological characteristics of the device compared to predicate device:

The predicate device K120775 (LightStim for Wrinkles) has a smaller surface area with fewer LEDs than the proposed Rejuvalite MD device, but number of wavelengths, treatment duration, treatment regime and amount of energy (irradiance per area) delivered to the skin are the same.

{6}------------------------------------------------

This device also differs from the predicate device K120775 (LightStim for Wrinkles) in that it is not a handheld device in contact with the skin. The Rejuvalite MD is used at a 4" distance from the illuminated surface, but the power density per wavelength (mW/cm2), total power density of combined wavelengths (62 mW/cm2), amount of energy (irradiance per area) delivered to the skin measured at 4" are identical to the predicate device . The method of operation of the Rejuvalite MD device is similar to the predicate device K120775 (LightStim for Wrinkles) in that both devices have no software and use an On/Off switch.

No adverse effects on any of the predicate devices were found in MAUDE since clearance in 2005.

After analysis of safety, indications, intended uses, power density per wavelength (mW/cm2) , total power density of combined wavelengths (62 mW/cm2), performance, features, design materials, , technological properties, treatment areas, treatment reqimes and methods of operation, the manufacturer asserts that no significant differences exist between the applicant device and predicates listed in the predicate chart, and no new issues arise for safety and effectiveness. Therefore, substantial equivalency is hereby requested.

6. Biocompatibility:

The only patient contact material in the Reiuvalite MD is the body of the device.

The body of the device is constructed of ABS plastic, the same material used in the predicate device. The biocompatibility of this material is well known and considered safe when in contact with healthy skin.

The conclusion is that the Rejuvalite MD does not raise any new safety issues.

7. Performance Testing and Standards:

This device has been tested under and are in compliance with performance standards that have

been established for such devices under Section 878 of the Federal Food, Drug, and Cosmetics Act.

Safety and functionality testing demonstrates that the Rejuvalite MD conforms to various international consensus standards:

IEC 60601-1: (2005) +Corr. 1 (2006) + Corr. + (2007): Medical Electrical Equipment Part 1:

General Requirements for Basic Safety and Essential Performance.

IEC 60601-1-2: (2007): Medical Electrical Equipment Part 1: General Requirements for Basic

Safety and Essential Performance: Collateral Standard: Electromagnetic Compatibility

{7}------------------------------------------------

Performance Testing on optical and electrical characteristics demonstrated that the Rejuvalite MD device is almost identical to that of the predicate devices.

A Usability/Label Comprehension Study was conducted with 36 participants.

The results of the study found that: 97% of the participants were able to comprehend the labeling. 100% of the participants were able to use the device successfully.

The conclusions drawn from nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed devices.

8. Statement of Safety and Effectiveness:

The information in this 510(k) submission was used to support the safety and effectiveness of this device with respect to its cited predicates.

9. Substantial Equivalence Discussion

After analysis of safety, indications, intended uses, power density (mW/cm2) per wavelength, total power density of combined wavelengths (62 mW/cm2), performance, features, design materials, technological properties, treatment areas, treatment regimes and methods of operation, the manufacturer asserts that no significant differences exist between the applicant device and predicates listed in the predicate chart, and no new issues arise for safety and effectiveness. Therefore, substantial equivalency is hereby requested.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.