K Number
K222350
Date Cleared
2023-04-13

(253 days)

Product Code
Regulation Number
878.4460
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A Sterile Nitrile Surgical Gloves Powder Free is a device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

Device Description

The proposed device, Sterile Nitrile Surgical Gloves Powder Free is a sterile and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination. The proposed device is made of synthetic rubber latex, as per standard ASTM D3577-2019, Standard Specification for Rubber Surgical Gloves. The classification is: Type II - gloves compounded from synthetic rubber latex. The proposed device is Sterile Nitrile Surgical Gloves Powder Free, and is produced in sizes 6, 6 ½, 7, 7 ½, 8, 8 ½, and 9. All sizes share the same White color. The proposed device is sterilized using Gamma Radiation method to achieve the Sterility Assurance Level (SAL) of 10-6 while packaged to maintain sterility. The shelf life is 3 years.

AI/ML Overview

The provided text describes the acceptance criteria and the results from non-clinical performance testing for a medical device, "Sterile Nitrile Surgical Gloves Powder Free." It does not contain information about studies involving human readers or AI assistance. Therefore, I will focus on the non-clinical performance testing detailed.

Here's the information extracted and organized as requested, with "N/A" for sections not applicable to this type of submission:

1. A table of acceptance criteria and the reported device performance

Test MethodPurposeAcceptance Criteria (AC)Reported Device Performance (Result)Met AC
ASTM D5151-2019Freedom from holes (Before aging)AQL 1.5 (3 non-consecutive lots)PassYes
ASTM D5151-2019Freedom from holes (After accelerated aging: 70 ± 2°C for 166 ± 2 h)AQL 1.5 (3 non-consecutive lots)PassYes
ASTM D5151-2019Freedom from holes (After accelerated aging: 50 ± 2°C for 90 ± 1 days)AQL 1.5 (3 non-consecutive lots)PassYes
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesTensile strength (Before aging)Minimum 17MPa for all sizes (AQL 4%, 3 non-consecutive lots)PassYes
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesTensile strength (After accelerated aging: 70 ± 2°C for 166 ± 2 h)Minimum 12MPa for all sizes (AQL 4%, 3 non-consecutive lots)PassYes
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesTensile strength (After accelerated aging: 50 ± 2°C for 90 ± 1 days)Minimum 12MPa for all sizes (AQL 4%, 3 non-consecutive lots)PassYes
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesUltimate Elongation (Before aging)Minimum 650% for all sizes (AQL 4%, 3 non-consecutive lots)PassYes
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesUltimate elongation (After accelerated aging: 70 ± 2°C for 166 ± 2 h)Minimum 490% for all sizes (AQL 4%, 3 non-consecutive lots)PassYes
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesUltimate elongation (After accelerated aging: 50 ± 2°C for 90 ± 1 days)Minimum 490% for all sizes (AQL 4%, 3 non-consecutive lots)PassYes
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesStress at 500% elongation (Before accelerated aging)Maximum 7.0MPa for all sizes (AQL 4%, 3 non-consecutive lots)PassYes
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesStress at 500% elongation (After accelerated aging: 70 ± 2°C for 166 ± 2 h)Maximum 7.0MPa for all sizes (AQL 4%, 3 non-consecutive lots)PassYes
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesStress at 500% elongation (After accelerated aging: 50 ± 2°C for 90 ± 1 days)Maximum 7.0MPa for all sizes (AQL 4%, 3 non-consecutive lots)PassYes
ASTM F1929, Test Method for detecting seal leaks in porous medical packaging by dye penetrationPackage integrity (After accelerated aging: 70 ± 2°C for 166 ± 2 h)No penetration (AQL 0.65%, 3 non-consecutive lots)PassYes
ASTM F1929, Test Method for detecting seal leaks in porous medical packaging by dye penetrationPackage integrity (After accelerated aging: 50 ± 2°C for 90 ± 1 days)No penetration (AQL 0.65%, 3 non-consecutive lots)PassYes
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesTo determine the length of the glovesMin 265mm for all sizesPass for all sizes listedYes
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesTo determine the width of the glovesSpecific width ranges for each size (e.g., 76±6mm for Size 6)Pass for all sizes listedYes
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesTo determine finger thicknessMinimum 0.10mm for all sizes (AQL 4%)Pass for all sizes listedYes
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesTo determine palm thicknessMinimum 0.10mm for all sizes (AQL 4%)Pass for all sizes listedYes
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesTo determine cuff thicknessMinimum 0.10mm for all sizes (AQL 4%)Pass for all sizes listedYes
ASTM D6124-06R17, Standard Test Method for Residual Powder on Medical GlovesLess than 2mg/gloveMaximum 2mg/glove for all sizesPassYes
ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10- Tests for irritation and skin sensitizationIrritationUnder the conditions of the testing, not an irritantPassYes
ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10- Tests for irritation and skin sensitizationSensitizationUnder the conditions of the testing, not a sensitizerPassYes
ISO 10993-5:2009(E) Biological Evaluation of Medical Devices - Part 5-Tests for in-vitro CytotoxicityCytotoxicityUnder the conditions of the testing, non-cytotoxicPassYes
ISO 10993-11:2017(E) Biological Evaluation of Medical Devices - Part 11- Tests for Systemic Toxicity and Biological TestsAcute systemic toxicityUnder the conditions of the testing, no acute systemic toxicityPassYes
USP 41 <151>Pyrogen TestMaterial mediated pyrogenicityUnder the conditions of the testing, non-pyrogenicPassYes
USP 42 <85>Bacterial Endotoxin TestBacterial endotoxin limitsLess than 20EU/glovePassYes

2. Sample size used for the test set and the data provenance
The document refers to "3 non-consecutive lots" for AQL (Acceptable Quality Limit) testing in several instances. However, the specific number of gloves sampled from each lot for each test is not explicitly stated.
Data provenance: Not specified (origin of the gloves/materials for testing). This is a non-clinical device performance test, not a study on patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
N/A - This submission concerns non-clinical performance testing of a physical medical device (gloves), not a diagnostic algorithm requiring expert "ground truth." The "ground truth" here is established by standardized test methods and their defined pass/fail criteria.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
N/A - See point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
N/A - This is not an AI/algorithm-assisted diagnostic device. No MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
N/A - This is not an AI/algorithm device. Standalone performance is demonstrated by the physical and biological testing results.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" or reference standard for these tests is defined by the technical specifications and methodologies outlined in the referenced ASTM and ISO standards (e.g., ASTM D3577-19 for surgical gloves, ISO 10993 for biological evaluation). These standards prescribe the testing procedures and define acceptable limits for various physical, chemical, and biological properties.

8. The sample size for the training set
N/A - This is not a machine learning or AI device that requires a training set.

9. How the ground truth for the training set was established
N/A - See point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.

April 13, 2023

JR Engineering & Medical Technologies (M) SDN. BHD. % Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Dr. STE 114 Aurora, Illinois 60504

Re: K222350

Trade/Device Name: Sterile Nitrile Surgical Gloves Powder Free Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I, reserved Product Code: KGO Dated: March 13, 2023 Received: March 13, 2023

Dear Manoj Zacharias:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K22350

Device Name

Sterile Nitrile Surgical Gloves Powder Free

Indications for Use (Describe)

A Sterile Nitrile Surgical Gloves Powder Free is a device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510K SUMMARY

K22350

Date of Preparation: April 10, 2023 As required by: 21CFR § 807.92

A. APPLICANT INFORMATION

Submitter NameJR Engineering & Medical Technologies (M) SDN. BHD
AddressLot 8 &10, Jalan Zurah 3 & Lot 1&3, Jalan Zurah 3A/1, PusatPerindustrian 2, 44200 Rasa, HuluSelangor, Selangor Darul Ehsan, Malaysia.
Phone+603-60572081
Fax+603-60572181
E-mailganeshjrmt@gmail.com
Contact PersonMr. Ganesan Subramaniam
DesignationManaging Director
Contact Number+6012 224 6677
Contact Emailganeshjrmt@gmail.com
US AgentManoj Zacharias
US Agent AddressLiberty Management Group Ltd.75 Executive Dr. Ste 114 Aurora, IL 60504 USA
US Agent Telephone1 (630) 270-2921
US Agent Fax1 (815) 986-2632
US Agent Emailmanoj@libertymanagement.us
B.Device Identification
Common NameSurgical Gloves
Device NameSterile Nitrile Surgical Gloves Powder Free
Product Proprietary orTrade NameJR MEDIC
Classification NameSurgeon's Gloves

1 KGO 21 CFR 878.4460

C. Predicate Device

Product Code

Device Classification

Regulation Number

510k Number Common Name Device Name Classification name Device Classification Product Code Regulation Number

K170515 Surgical Gloves Sterile Nitrile Surgical Gloves, Powder Free Surgeon's Gloves 1 KGO 21 CFR 878.4460

D. Description of the Device

The proposed device, Sterile Nitrile Surgical Gloves Powder Free is a sterile and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination.

{4}------------------------------------------------

The proposed device is made of synthetic rubber latex, as per standard ASTM D3577-2019, Standard Specification for Rubber Surgical Gloves.

The classification is: Type II - gloves compounded from synthetic rubber latex.

The proposed device is Sterile Nitrile Surgical Gloves Powder Free, and is produced in sizes 6, 6 ½, 7, 7 ½, 8, 8 ½, and 9. All sizes share the same White color.

The proposed device is sterilized using Gamma Radiation method to achieve the Sterility Assurance Level (SAL) of 10 6 while packaged to maintain sterility. The shelf life is 3 years.

E. Technological Characteristics

CharacteristicSubject device K222350Predicate device K170515Remarks
Product CodeKGOKGOSame
Regulation No.21 CFR 878.446021 CFR 878.4460Same
Class11Same
Intended UseA Sterile Nitrile Surgical GlovesPowder Free is a device made ofsynthetic rubber intended to beworn by operating roompersonnel to protect a surgicalwound from contamination.The powder-free nitrile surgicalGloves, is a sterile and single usedevise made of synthetic rubberintended to be worn by operationroom personnel to protect a surgicalwound from contamination. Thegloves do not contain lubricating ordusting powderSimilar
Powdered or Powder freePowder freePowder freeSame
MaterialSynthetic RubberSynthetic RubberSame
ColorWhiteWhiteSame
Classification as perASTMD3577-19,Standard Specificationfor Rubber SurgicalGlovesType II - gloves compoundedfrom Synthetic rubber latexType II - gloves compounded fromSynthetic rubber latexSame
SterilizationRadiation, SAL-10-6ETO/as well as Radiation, SAL- 10-6Similar
Label and LabelingMeet FDA's label RequirementsMeet FDA's label RequirementsSame
Type of useOver the counter useOver the counter useSame
Dimensions Length:-Min 265mm307mm265mm minSimilar
Size: Width
6.0 (76±6mm)78mm76±6mm
6.5 (83±6mm)85mm83±6mmSimilar
7.0 (89±6mm)88mm89±6mm
7.5 (95±6mm)97mm95±6mm
8.0 (102±6mm)103mm102±6mm
8.5 (108±6mm)110mm108±6mm
9.0 (114±6mm)116mm114±6mm
Cuff, Palm, Finger TipThicknessMin 0.10mmCuff- 0.12mmPalm- 0.18mmFinger Tip- 0.21mmCuff, Palm & Finger has a min0.10mm thicknessSimilar
Tensile StrengthUnaged17MPa minimum24.20MPa17MPa minSimilar
Ultimate ElongationUnaged650% minimum871%650% minSimilar
Stress at 500%
Unaged7.0 MPa Max6.1MPa7.0MPa maxSimilar
Tensile Strength
Aged12MPa minimum18.53MPa12 MPa minSimilar
Ultimate Elongation
Aged490% minimum653%490% minSimilar
Freedom from HolesAQL 1.5AQL 2.5Similar
Powder residue forpowder free glovePowder content< 2 mg/Glove0.40mg/Glove< 2mg/GloveSimilar
Skin IrritationUnder conditions of the testing, not an irritantUnder conditions of the testing, not an irritantSimilar
Skin SensitizationUnder the conditions of the testing, not a sensitizerUnder the conditions of the testing, not a sensitizerSimilar
In vitro cytotoxicityUnder the conditions of the testing, not cytotoxicUnder the conditions of the testing, not cytotoxicSimilar
Material MediatedpyrogenicityUnder the conditions of the testing, non-pyrogenicNo data availableDifferent
Bacterial Endotoxin Test< 20 EU/gloveNo data availableDifferent
Systemic ToxicityUnder the conditions of the testing, no acute systemictoxicityNo data availableDifferent
Sterilization ModalityRadiationRadiation/EtOSimilar
Sterility Assurance Level10-610-6Same
Shelf Life3 yearsUnknownDifferent

{5}------------------------------------------------

F. Summary of Non-Clinical Performance Testing

Test MethodPurposeAcceptance CriteriaResult
ASTM D5151-2019Freedom from holesBefore agingAQL 1.53 non-consecutive lotsPass
ASTM D5151-2019Freedom from holesAfter accelerated aging70 $\pm$ 2°C for 166 $\pm$ 2 hAQL 1.53 non-consecutive lotsPass
ASTM D5151-2019Freedom from holesAfter accelerated aging50 $\pm$ 2°C for 90 $\pm$ 1 daysAQL 1.53 non-consecutive lotsPass
ASTM D3577-19,StandardSpecification forRubber SurgicalGlovesTensile strengthBefore agingMinimum 17MPa for allsizesAQL 4%3 non-consecutive lotsPass
ASTM D3577-19,StandardSpecification forRubber SurgicalTensile strengthAfter accelerated aging70 $\pm$ 2°C for 166 $\pm$ 2 hMinimum 12MPa for allsizesAQL 4%3 non-consecutive lotsPass
Gloves
ASTM D3577-19,StandardSpecification forRubber SurgicalGlovesTensile strengthAfter accelerated aging$50\pm2$ °C for $90\pm1$ daysMinimum 12MPa for allsizesAQL 4%3 non-consecutive lotsPass
ASTM D3577-19,StandardSpecification forRubber SurgicalGlovesUltimate ElongationBefore agingMinimum 650% for allsizesAQL 4%3 non-consecutive lotsPass
ASTM D3577-19,StandardSpecification forRubber SurgicalGlovesUltimate elongationAfter accelerated aging$70\pm2$ °C for $166\pm2$ hMinimum 490% for allsizesAQL 4%3 non-consecutive lotsPass
ASTM D3577-19,StandardSpecification forRubber SurgicalGlovesUltimate elongationAfter accelerated aging$50\pm2$ °C for $90\pm1$ daysMinimum 490% for allsizesAQL 4%3 non-consecutive lotsPass
ASTM D3577-19,StandardSpecification forRubber SurgicalGlovesStress at 500% elongationBefore accelerated agingMaximum 7.0MPa for allsizesAQL 4%3 non-consecutive lotsPass
ASTM D3577-19,StandardSpecification forRubber SurgicalGlovesStress at 500% elongationAfter accelerated aging$70\pm2$ °C for $166\pm2$ hMaximum 7.0MPa for allsizesAQL 4%3 non-consecutive lotsPass
ASTM D3577-19,StandardSpecification forRubber SurgicalGlovesStress at 500% elongationAfter accelerated aging$50\pm2$ °C for $90\pm1$ daysMaximum 7.0MPa for allsizesAQL 4%3 non-consecutive lotsPass
ASTM F1929,Test Method fordetecting seal leaks inporous medicalpackaging by dyepenetrationPackage integrityAfter accelerated aging$70\pm2$ °C for $166\pm2$ hNo penetrationAQL 0.65%3 non-consecutive lotsPass
ASTM F1929,Test Method fordetecting seal leaks inporous medicalpackaging by dyepenetrationPackage integrityAfter accelerated aging$50\pm2$ °C for $90\pm1$ daysNo penetrationAQL 0.65%3 non-consecutive lotsPass
ASTM D3577-19,StandardSpecification forRubber SurgicalTo determine the length ofthe glovesMin 265mm for all sizesSize 6Pass
Size 6 ½Pass
Size 7Pass
GlovesSize 7 ½Pass
Size 8Pass
Size 8½Pass
Size 9Pass
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesTo determine the width of the glovesSize 676±6mmSize 6Pass
Size 6 ½83±6mmSize 6 ½Pass
Size 789±6mmSize 7Pass
Size 7 ½95±6mmSize 7 ½Pass
Size 8102±6mmSize 8Pass
Size 8 ½108±6mmSize 8 ½Pass
Size 9114±6mmSize 9Pass
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesTo determine finger thicknessMinimum 0.10mm for all sizesAQL 4%Size 6Pass
Size 6 ½Pass
Size 7Pass
Size 7 ½Pass
Size 8Pass
Size 8 ½Pass
Size 9Pass
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesTo determine palm thicknessMinimum 0.10mm for all sizesAQL 4%Size 6Pass
Size 6 ½Pass
Size 7Pass
Size 7 ½Pass
Size 8Pass
Size 8 ½Pass
Size 9Pass
ASTM D3577-19, Standard Specification for Rubber Surgical GlovesTo determine cuff thicknessMinimum 0.10mm for all sizesAQL 4%Size 6Pass
Size 6 ½Pass
Size 7Pass
Size 7 ½Pass
Size 8Pass
Size 8 ½Pass
Size 9Pass
ASTM D6124-06R17, Standard Test Method for Residual Powder on Medical GlovesLess than 2mg/gloveMaximum 2mg/glove for all sizesPass
ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10- Tests for irritation and skin sensitizationIrritationUnder the conditions of the testing, not an irritantPass
ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10- Tests for irritation and skin sensitizationSensitizationUnder the conditions of the testing, not a sensitizerPass
ISO 10993-5:2009(E) CytotoxicityUnder the conditions ofPass

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Biological Evaluationof Medical Devices -Part 5-Tests forin-vitro Cytotoxicitythe testing, non-cytotoxic
ISO 10993-11:2017(E)Biological Evaluationof Medical Devices -Part 11- Tests forSystemicToxicity andBiological TestsAcute systemic toxicityUnder the conditions ofthe testing, no acutesystemic toxicityPass
USP 41<151>Pyrogen TestMaterial mediatedpyrogenicityUnder the conditions ofthe testing, non-pyrogenicPass
USP 42 <85>Bacterial EndotoxinTestBacterial endotoxin limitsLess than 20EU/glovePass

G. Summary of Clinical Performance Testing

No clinical performance testing was performed in support of this submission.

H. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device Sterile Nitrile Surgical Gloves Powder Free is as safe, as effective and performs as well or better than the legally marketed predicate device "PRIMUS NITRILE GLOVES" Sterile Nitrile Surgical Gloves Powder Free (K170515).

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).