(295 days)
The Babylog intensive care Ventilator (Babylog VN600) is intended for the ventilation of neonates with a body weight of 0.4 kg (0.88 lb) to 10 kg (22 lb) and for pediatric patients with an IBW of 5.2 kg (11.5 lb) to 20.1 kg (44.3 lb).
This device provides mandatory pressure-controlled ventilation modes for supporting spontaneous breathing, and ventilation monitoring.
The intensive care ventilators Babylog VN800 and Babylog VN600 were developed and are manufactured by Dräger in Lübeck, Germany.
Babylog VN800 and Babylog VN600 are specified for the ventilation of pediatric patients and neonates. These devices provide mandatory ventilation modes and ventilation modes for supporting spontaneous breathing as well as ventilation monitoring.
Babylog VN800 and Babylog VN600 are available in different device variants and can additionally be upgraded by software and hardware options as well as attachable accessories.
Babylog VN800 and Babylog VN600 are available with a basic device configuration that comprises the following:
- Trolley with four castors and brakes
- Display unit (graphical user interface)
- Ventilation unit
The changes have been made in the device design including functional integration of the Infinity series C cockpit into the device by changing "PC based Cockpit technology" to an "Embedded Control Display (ECD) technology" including a visually updated (color) in the graphical user interface. The hardware complexity has been reduced. In addition, a Gas supply Unit has been added as well as the Power supply Unit as an option. The device software and firmware have been redesigned. The functionality and features of Babylog VN800 / VN600 are identical to the predicate device.
It appears the provided text describes a 510(k) premarket notification for a medical device, the Babylog VN800/VN600 ventilator. However, the document focuses on demonstrating substantial equivalence to a predicate device based on technical specifications, intended use, and adherence to consensus standards. It does not present a clinical study to prove the device meets acceptance criteria related to its performance in a clinical setting with human subjects or a clinical outcome.
Specifically, the document lists "Non-clinical Testing" including:
- Sterilization
- Biocompatibility
- Software, including cybersecurity
- Electrical safety
- Electromagnetic compatibility (EMC)
- Alarm Systems in medical electrical equipment
- Respiratory gas monitors
- Waveforms, including comparisons to the predicate device and performance
- Technical System Requirements
- Accessories compatibility
- Human factors engineering, usability
The "Conclusion" states that "The conclusions drawn from non-clinical tests and the comparison of intended use and technological characteristics with its predicate demonstrate that the new product Babylog VN800 / Babylog VN600 is substantially equivalent to the predicate device... and does not raise new questions of safety or effectiveness."
Therefore, based solely on the provided text, I cannot fulfill your request for:
- A table of acceptance criteria and the reported device performance: The document does not describe specific performance metrics (e.g., accuracy, sensitivity, specificity) for a clinical application. The "performance" mentioned refers to technical performance related to waveforms, not clinical outcomes.
- Sample size used for the test set and data provenance: There is no mention of a human test set. The testing described is non-clinical.
- Number of experts used to establish the ground truth: Not applicable as there is no clinical test set requiring ground truth established by experts.
- Adjudication method: Not applicable.
- Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable.
- Standalone (algorithm only) performance: Not applicable, as this is a physical ventilator device, not an AI algorithm for diagnosis/detection.
- Type of ground truth used: Not applicable.
- Sample size for the training set: Not applicable, as this is not an AI/ML-based device requiring a training set in the typical sense.
- How ground truth for the training set was established: Not applicable.
The document demonstrates that the device functionally meets standards and is similar to a previously cleared device, but it does not detail a study proving clinical performance against specific acceptance criteria in the way you've outlined for an AI-powered diagnostic or similar product.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
May 16, 2023
Draegerwerk AG & CO KGaA Nataliia Semenova Regulatory Affairs Manager 53/55 Moislinger Allee Luebeck, Schlewsig-Holstein 23542 Germany
Re: K222207
Trade/Device Name: Babylog VN800, Babylog VN600 Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: CBK Dated: April 17, 2023 Received: April 17, 2023
Dear Nataliia Semenova:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ting Song -S
For James J. Lee, Ph.D. Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K22207
Device Name Babylog VN800 / Babylog VN600
Indications for Use (Describe)
The Babylog intensive care Ventilator (Babylog VN600) is intended for the ventilation of neonates with a body weight of0.4 kg (0.88 lb) to 10 kg (22 lb) and for pediatric patients with an IBW of 5.2 kg (11.5 lb) to 20.1 kg (44.3 lb).
This device provides mandatory pressure-controlled ventilation modes for supporting spontaneous breathing, and ventilation monitoring.
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
--- Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue. The two dots above the "a" are also blue.
510(k) Premarket Notification Summary
| Submitter: | Drägerwerk AG & Co. KGaAMoislinger Allee 53-5523542 Lübeck, GermanyEstablishment's registration number: 9611500 |
|---|---|
| Official Correspondent: | Dr. Bettina MoebiusHead of Regulatory Affairs CentralE-Mail: bettina.moebius@draeger.comTelephone: +49 451 882 4100 |
| US Agent: | Darlene ThibodeauHead of Quality Assurance and ComplianceE-mail: darlene.thibodeau@draeger.comTelephone: (978) 773-1060 |
| Date prepared: | 16 May 2023 |
| Device Name: | Common name: VentilatorTrade Name: Babylog VN800, Babylog VN600Classification Name: Ventilator, Continuous, Facility UseRegulation Number: 21 CFR §868.5895Product Code: CBKClass: II |
Predicate Device: Infinity Acute Care System Workstation Neonatal Care, K093632
Drägerwerk AG & Co. KGaA is submitting a traditional 510(k) premarket notification for a new device, Babylog VN800 / Babylog VN600.
Device Description
The intensive care ventilators Babylog VN800 and Babylog VN600 were developed and are manufactured by Dräger in Lübeck, Germany.
Babylog VN800 and Babylog VN600 are specified for the ventilation of pediatric patients and neonates. These devices provide mandatory ventilation modes and ventilation modes for supporting spontaneous breathing as well as ventilation monitoring.
Babylog VN800 and Babylog VN600 are available in different device variants and can additionally be upgraded by software and hardware options as well as attachable accessories.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "Dräger" in blue font. The font is bold and sans-serif. The two dots above the "a" are also blue. The word is the logo for the Dräger company, which is a German company that makes medical and safety technology.
Babylog VN800 and Babylog VN600 are available with a basic device configuration that comprises the following:
- Trolley with four castors and brakes
- Display unit (graphical user interface)
- Ventilation unit
The changes have been made in the device design including functional integration of the Infinity series C cockpit into the device by changing "PC based Cockpit technology" to an "Embedded Control Display (ECD) technology" including a visually updated (color) in the graphical user interface. The hardware complexity has been reduced. In addition, a Gas supply Unit has been added as well as the Power supply Unit as an option. The device software and firmware have been redesigned. The functionality and features of Babylog VN800 / VN600 are identical to the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is written in blue. The two dots above the "a" are also blue. The word is horizontally oriented and centered.
Intended Use / Indications for Use
The Babylog intensive care ventilator (Babylog VN800 / Babylog VN600) is intended for the ventilation of neonates with a body weight of 0.4 kg (0.88 lb) to 10 kg (22 lb) and for pediatric patients with an IBW of 5.2 kg (11.5 lb) to 20.1 kg (44.3 lb).
This device provides mandatory pressure-controlled ventilation modes, ventilation modes for supporting spontaneous breathing, and ventilation monitoring.
List of Consensus Standards
| Standard Number and Version | Title |
|---|---|
| ANSI AAMI ES60601-1:2005/(R)2012and A1:2012, C1:2009/(R)2012 andA2:2010/(R)2012 | Medical electrical equipment - Part 1:General requirements for basic safety andessential performance |
| IEC 60601-1-2 Edition 4.1 2020-09CONSOLIDATED VERSION | Medical electrical equipment - Part 1-2:General requirements for basic safety andessential performance - CollateralStandard: Electromagnetic disturbances -Requirements and tests |
| IEC 60601-1-6 Edition 3.1 2013-10 | Medical electrical equipment - Part 1-6:General requirements for basic safety andessential performance - Collateralstandard: Usability |
| IEC 60601-1-8 Edition 2.1 2012-11 | Medical electrical equipment - Part 1-8:General requirements for basic safety andessential performance - CollateralStandard: General requirements, testsand guidance for alarm systems inmedical electrical equipment and medicalelectrical systems |
| ISO 80601-2-12 First edition 2011-04-15 | Medical electrical equipment - Part 2-12:Particular requirements for the safety oflung ventilators - Critical care ventilators[Including: Technical Corrigendum 1(2011)] |
| ISO 80601-2-55 Second edition 2018-02 | Medical electrical equipment - Part 2-55:Particular requirements for the basicsafety and essential performance ofrespiratory gas monitors |
| ISO 14971 Third Edition 2019-12 | Medical devices - Application of riskmanagement to medical devices |
| IEC 62304 Edition 1.1 2015-06CONSOLIDATED VERSION | Medical device software - Software lifecycle processes |
| IEC 62366-1 Edition 1.0 2015-02 | Medical devices - Part 1: Application ofusability engineering to medical devices |
| ISO 10993-1 Fifth edition 2018-08 | Biological evaluation of medical devices -Part 1: Evaluation and testing within a riskmanagement process |
| ISO 18562-1 First edition 2017-03 | Biocompatibility evaluation of breathinggas pathways in healthcare applications -Part 1: Evaluation and testing within a riskmanagement process |
| IEC 62133-2 Edition1.0 2017-02 | Secondary cells and batteries containingalkaline or other non-acid electrolytes -Safety requirements for portable sealedsecondary cells, and for batteries madefrom them, for use in portable applications- Part 2: Lithium systems |
| IEC 62133-1 Edition1.0 2017-02 | Secondary cells and batteries containingalkaline or other non-acid electrolytes -Safety requirements for portable sealedsecondary cells, and for batteries madefrom them, for use in portable applications- Part 1: Nickel systems |
| /TR 60601-4-2 Edition 1.0 2016-05 | Medical electrical equipment - Part 4-2:Guidance and interpretation -Electromagnetic immunity: performanceof medical electrical equipment andmedical electrical systems |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue. The "a" in Dräger has an umlaut.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue. The two dots above the "a" are also blue.
510(k) Summary
Comparison to Predicate
| Topic | Proposed deviceBABYLOG VN800, BABYLOG VN600 | Predicate deviceInfinity Acute Care System WorkstationNeonatal Care | Comments |
|---|---|---|---|
| Device Trade Name | Babylog VN800Babylog VN600 | Infinity Acute Care System Workstation NeonatalCare | – |
| Manufacturer | Drägerwerk AG & Co. KGaA | Drägerwerk AG & Co. KGaA | Same |
| 510(k) number | K222207 | K093632 | – |
| Regulation number -Classification description | 868.5895 - Continuous Ventilator | 868.5895 - Continuous Ventilator | Same |
| Regulatory class | Class II | Class II | Same |
| Product code | CBK | CBK | Same |
| Patient population | • Pediatric• Neonate | • Pediatric• Neonate | Same |
| Topic | Proposed deviceBABYLOG VN800, BABYLOG VN600 | Predicate deviceInfinity Acute Care System WorkstationNeonatal Care | Comments |
| Intended Use /Indications for Use | The Babylog intensive care ventilator (BabylogVN800 / Babylog VN600) is intended for theventilation of neonates with a body weight of 0.4 kg(0.88 lb) to 10 kg (22 lb) and for pediatric patientswith an IBW of 5.2 kg (11.5 lb) to 20.1 kg (44.3 lb).This device provides mandatory pressure-controlledventilation modes, ventilation modes for supportingspontaneous breathing, and ventilation monitoring. | The Infinity Acute Care System WorkstationsNeonatal Care consists of monitoring and controldisplays and additional therapy. They are intendedto be used as integrated, networked, andconfigurable Workstations to provide specifictherapy in neonatal intensive care. The InfinityAcute Care System Workstations Neonatal Careare intended to be used by qualified and trainedmedical personnel.The Infinity CSeries Medical Cockpits, consisting ofthe C500 and and the C700, are monitoring andcontrol displays for the Infinity Acute Care System(IACS). Medical Cockpits are intended to be used tomonitor waveforms, parameter information, andalarms as well as to control settings. The InfinityCSeries Medical Cockpits are intended to be usedin environments where patient care is provided bytrained healthcare Professionals.The Babylog VN500 Ventilation unit of theInfinity Care System is intended for theventilation of neonatal patients from 0.4 kg (0.88lbs) up to 10 kg (22 lbs) and pediatric patientsfrom 5 kg (11 lbs) up to 20 kg (44 lbs)bodyweight. Babylog VN500 offers mandatoryventilation modes for spontaneous breathingsupport and airway monitoring. The BabylogVN500 ventilation unit is intended for use indifferent medical care areas.Babylog VN500 is intended for stationary use inhospitals and medical rooms or for patienttransportation within the hospital. | Similar |
| Topic | Proposed deviceBABYLOG VN800, BABYLOG VN600 | Predicate deviceInfinity Acute Care System WorkstationNeonatal Care | Comments |
| Intended user | The operating organization must ensure the following:Every user group has the required qualifications (e.g., has undergone specialist training or acquired specialist knowledge through experience). Every user group has been trained to perform the task. Clinical usersThis user group operates the product in accordance with the intended use.Users have medical specialist knowledge in the field of ventilation. Users have knowledge of device monitoring and ventilation care. | The Infinity Acute Care System Workstations Neonatal Care are intended to be used by qualified and trained medical personnel.The Infinity CSeries Medical Cockpits are intended to be used in environments where patient care is provided by trained healthcare Professionals. | Similar |
| Environment of use | The device is intended for stationary use in hospitals and medical rooms or for intrahospital patient transport.Do not use the device in the following environments of use:- Hyperbaric chambers- Magnetic resonance imaging- Together with flammable gases or flammable solutions that can mix with air, oxygen, or nitrous oxide- Areas with danger of explosion- Areas with combustible and highly flammable substances- Rooms with insufficient ventilationDo not operate the device with helium or helium mixtures. | Babylog VN500 is intended for stationary use in hospitals and medical rooms or for patient transportation within the hospital.Do not use the device in the following environments:- In hyperbaric chambers- For magnetic resonance imaging (MRI, NMR, NMI)- In conjunction with flammable gases or flammable solutions that can mix with air, oxygen or nitrous oxide- In areas of explosion hazard- In areas with combustible or explosive substances- In rooms without sufficient ventilationDo not operate the device with helium or helium mixtures. | Similar |
| Topic | Proposed deviceBABYLOG VN800, BABYLOG VN600 | Predicate deviceInfinity Acute Care System WorkstationNeonatal Care | Comments |
| Gas supply | Central gas supply | Central gas supply | Same |
| Gas cylinders• Transport supply unit (optional)• Gas cylinder holder (optional) | Gas cylinders• Transport supply unit (optional)• Gas cylinder holder (optional) | Same | |
| Gas supply unit GS500 (optional) | - | Different | |
| Gas dosage, mixing anddelivery | Dosage of Air/O2 in mixing chamber | Dosage of Air/O2 in mixing chamber | Same |
| Inspiratory valve | Inspiratory valve | Same | |
| Expiratory valve | Expiratory valve | Same | |
| Pneumatic nebulizer (optional, part of "Nebulizer"option) | Pneumatic nebulizer (optional, part of "Medicationnebulization" option) | Same | |
| Pressure and volumemonitoring | Inspiratory and expiratory pressure | Inspiratory and expiratory pressure | Same |
| Inspiratory flow/volume | Inspiratory flow/volume | Same | |
| Proximal flow/volume | Proximal flow/volume | Same | |
| 2 ambient pressure sensors | 1 ambient pressure sensor and user setting altitude | Similar | |
| Oxygen monitoring | Paramagnetic side-stream sensor | Paramagnetic side-stream sensor | Same |
| Flow balancing of Air and O2 | Flow balancing of Air and O2 | Same | |
| Carbon dioxide monitoring | Main-stream sensor usinginfrared absorptionspectroscopy (optional, part of “CO2 monitoring”option)• part no.6871950 MCable - Mainstream CO2• part no. 6873570 CO2 mainstream sensor | Main-stream sensor usinginfrared absorptionspectroscopy (optional, part of “CO2 monitoring”option)• part no.6871950 MCable -Mainstream CO2 | Differentoptional component6873570 (K221118)was introduced. |
| Batteries | Internal battery (NiMH) | Internal battery (NiMH) | Same |
| Topic | Proposed deviceBABYLOG VN800, BABYLOG VN600 | Predicate deviceInfinity Acute Care System WorkstationNeonatal Care | Comments |
| Power supply unit PS500 (optional)part no. 8418950 lead-acid part no. 8422900 lithium iron phosphate (LFP) | - | Different | |
| User interface | TFT LCD capacitive touchscreen display Babylog VN800: 18.5 in Babylog VN600: 15.6 in | TFT LCD resistive touchscreen display IACS Medical cockpit C500: 17 in | Similar |
| Rotary knob for selecting, adjusting and confirming | Rotary knob for selecting, adjusting and confirming | Same | |
| Power supply indicators | Power supply indicators | Same | |
| On/off key | On/off key | Same | |
| Operation display | Operation display | Same | |
| Bed coupling (optional) | - | Different | |
| User interface, screendisplays | Waveforms Graphical trends Tabular trends Loops Alarm logbook Logbook Numeric parameters Lists of measured values and set values User-specific list for measured values and set values Smart Pulmonary View (optional) | Curves Graphical trends Numeric trends Loops Alarm history Logbook Numeric parameters Preconfigured lists of measured values and set values Customized lists for measured values and set values Smart Pulmonary View (optional) | Same |
| General monitoring, alarmprinciples | Optical and acoustical alarm indication | Optical and acoustical alarm indication | Same |
| Alarm silence key | Alarm silence key | Same | |
| High, medium and low alarm priorities | High, medium and low alarm priorities | Same | |
| Topic | Proposed deviceBABYLOG VN800, BABYLOG VN600 | Predicate deviceInfinity Acute Care System WorkstationNeonatal Care | Comments |
| Device monitoring | Downgrading and resetting of defined alarms | Downgrading and resetting of defined alarms | Same |
| Alarm logbook | Alarm history | Same | |
| Nurse call (optional) | Nurse call (optional) | Same | |
| User-initiated system test | User-initiated system test | Same | |
| Gas supply | Gas supply | Same | |
| Power supply | Power supply | Same | |
| Device temperature | Device temperature | Same | |
| Mutual Processor Monitoring | Mutual Processor Monitoring | Same | |
| Connection to display unit | Connection to display unit | Same | |
| Communication protocols | MEDIBUS | MEDIBUS | Same |
| MEDIBUS.X | - | Different | |
| MED.X.Comp | - | Different | |
| Therapy types | Invasive ventilation (Tube) | Invasive ventilation (Tube) | Same |
| Non-invasive ventilation (NIV) (optional) | Non-invasive ventilation (NIV) (optional) | Same | |
| O2 Therapy | O2 Therapy | Same | |
| Type of patient interface | Invasive ventilation (Tube):Endotracheal tube (for pediatric patients and neonates)Tracheostomy cannula (for pediatric patients) | Invasive ventilation (Tube):Endotracheal tube (for pediatric patients and neonates)Tracheostomy cannula (for pediatric patients) | Same |
| Non-invasive ventilation (NIV) (optional):NIV mask (for pediatric patients and neonates)Prongs (for pediatric patients and neonates) | Non-invasive ventilation (NIV) (optional):NIV mask (for pediatric patients and neonates)Prongs (for pediatric patients and neonates) | Same | |
| Traditional 510(k) | 510(k) Summary | Section 005 | |
| Topic | Proposed deviceBABYLOG VN800, BABYLOG VN600 | Predicate deviceInfinity Acute Care System WorkstationNeonatal Care | Comments |
| O2 Therapy:– Nasal cannula (for pediatric patients andneonates) | O2 Therapy:– Nasal cannula (for pediatric patients andneonates)– Oxygen mask (for pediatric patients andneonates) | Same | |
| Ventilation modes | Pressure Control - Synchronized IntermittentMandatory Ventilation,PC-SIMV | Pressure Control - Synchronized IntermittentMandatory Ventilation,PC-SIMV | Same |
| Pressure Control - Assist Control,PC-AC | Pressure Control - Assist Control,PC-AC | Same | |
| Pressure Control - Controlled MandatoryVentilation,PC-CMV | Pressure Control - Continuous MandatoryVentilation,PC-CMV | Same | |
| Pressure Control - Airway Pressure ReleaseVentilation,PC-APRV (optional) | Pressure Control - Airway Pressure ReleaseVentilation,PC-APRV (optional) | Same | |
| Pressure Control - Pressure Support Ventilation,PC-PSV | Pressure Control - Pressure Support Ventilation,PC-PSV | Same | |
| Pressure Control - Mandatory Minute Ventilation,PC-MMV (optional, part of "Volume ventilation"option) | Pressure Control - Mandatory Minute VolumeVentilation, PC-MMV (optional, part of "Volumeventilation" option) | Same | |
| Spontaneous - Continuous Positive AirwayPressure, Pressure Support,SPN-CPAP/PS | Spontaneous - Continuous Positive AirwayPressure, Pressure Support,SPN-CPAP/PS | Same | |
| Spontaneous - Continuous Positive AirwayPressure, Volume Support, SPN-CPAP/VS(optional, part of "Volume ventilation" option) | Spontaneous - Continuous Positive AirwayPressure, Volume Support, SPN-CPAP/VS(optional, part of "Volume ventilation" option) | Same | |
| Topic | Proposed deviceBABYLOG VN800, BABYLOG VN600 | Predicate deviceInfinity Acute Care System WorkstationNeonatal Care | Comments |
| Spontaneous - Continuous Positive AirwayPressure, SPN-CPAP (optional, part of "Non-invasive ventilation" option) | Spontaneous - Continuous Positive AirwayPressure, SPN-CPAP (optional, part of "Non-invasive ventilation" option) | Same | |
| Spontaneous - Proportional Pressure Support,SPN-PPS (optional) | Spontaneous - Proportional Pressure Support,SPN-PPS (optional) | Same | |
| Additional settings forventilation | Apnea ventilation | Apnea ventilation | Same |
| Trigger | Flow trigger | Same | |
| Sigh | Sigh | Same | |
| Volume guarantee (optional, part of "Volumeventilation" option) | Volume Guarantee (optional, part of "Volumeventilation" option) | Same | |
| ATC (optional) | ATC (optional) | Same | |
| AutoRelease | AutoRelease | Same | |
| Volume ventilation | Software option which contains PC-MMV and SPN-CPAP/VS and Volume guarantee | Software option which contains PC-MMV and SPN-CPAP/VS and Volume guarantee | Same |
| Anti-air shower | Reduced flow after detected disconnection untildetected reconnection | Reduced flow after detected disconnection untildetected reconnection | Same |
| Maneuvers | ● Manual insp./inspiration hold● 02/suctioning● Nebulization (optional) | ● Manual inspiration/hold● Suction maneuver● Medication nebulization (optional) | Same |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "Dräger" in blue font. The "ä" in the word has two blue dots above it. The font is bold and sans-serif.
Traditional 510(k)
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue. The two dots above the "a" are also blue.
Traditional 510(k)
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the word "Dräger" in a bold, blue font. The "ä" in Dräger has two dots above it. The logo is simple and modern.
Traditional 510(k)
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue. The two dots above the "a" are also blue.
Traditional 510(k)
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the word "Dräger" in blue font. The word is written in a bold, sans-serif font. The two dots above the "a" are also blue. The background is white.
Traditional 510(k)
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image shows the word "Dräger" in a bold, blue font. The two dots above the "a" are also blue. The word is centered in the image and takes up most of the space.
{14}------------------------------------------------
Image /page/14/Picture/1 description: The image shows the word "Dräger" in a bold, blue font. The two dots above the "a" are also blue. The word is the logo for the Dräger company, which specializes in medical and safety technology.
Traditional 510(k)
{15}------------------------------------------------
Image /page/15/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is in blue color. The two dots above the "a" are also blue. The logo is simple and modern.
510(k) Summary
Discussion of Non-clinical Testing
The Babyloq VN800 / Babyloq VN600 ventilator is a new device and has undergone extensive testing to qualify it with e.g., national and international consensus standards, technical system requirements and other requirements. The following verification and validation activities were deemed necessary to establish substantial equivalence to the predicate device and were carried out under well-established methods, their results summarized in Test Summary tables and the evidence included in this submission.
- . Sterilization
- Biocompatibility
- Software, including cybersecurity ●
- Electrical safety ●
- . Electromagnetic compatibility (EMC)
- . Alarm Systems in medical electrical equipment
- . Respiratory gas monitors
- Waveforms, including comparisons to the predicate device and performance
- Technical System Requirements, covering:
- o Risk control measures
- o Technical data
- Essential safety and performance o
- . Accessories compatibility
- Human factors engineering, usability .
Conclusion
The conclusions drawn from non-clinical tests and the comparison of intended use and technological characteristics with its predicate demonstrate that the new product Babylog VN800 / Babylog VN600 is substantially equivalent to the predicate device Infinity Acute Care System Workstation Neonatal Care K093632 and does not raise new questions of safety or effectiveness.
- END -
§ 868.5895 Continuous ventilator.
(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).