K Number
K222205
Date Cleared
2022-10-07

(74 days)

Product Code
Regulation Number
878.4860
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cold Sore Device is indicated for shortening the time to healing of herpes simplex labialis lesions on or around the lips with time to healing defined as the time to patient described re-epithelialization.

Device Description

The Cold Sore Device is a solid state opto-electronic device that emits a controlled quantity of 1072mm +/-12nm peak wavelength near infrared light for a period of approximately 3 minutes. The maximum peak light intensity across the treatment surface is 20mW/cm². The light output and duration are monitored by a microprocessor. The tip of the device that contacts the patient is made out of Acrylonitrile Butadiene Styrene (ABS) + PC. Treatment with the device is commenced at the first symptoms of a cold sore 3 times a day with 4 hours in between each treatment for 2 consecutive days. The treatment area is approximately 3 cm². There are 2 light emitting diodes (LEDs) in the treatment area. The light within the device is activated by opening the cover of the device and automatically shut down after the preprogrammed treatment time (3 minutes). The device is designed for external, limited duration skin contact in an environment free from fluids and is provided non-sterile. The LEDs do not come in direct contact with the patient based upon the design of the device. The device is for OTC use and single patient use as described in the patient and box labeling.

AI/ML Overview

The provided text is a 510(k) summary for the Cold Sore Device (Model: QPZ-01). It outlines the device's characteristics, its comparison to a predicate device, and the testing conducted to support its substantial equivalence. However, the document explicitly states that "no clinical test on our device" was performed. This means the information requested about acceptance criteria and studies proving the device meets those criteria through clinical trials (e.g., sample size for test set/training set, expert involvement, adjudication, MRMC studies, standalone performance, ground truth establishment) cannot be extracted from this document because such studies were not conducted for this submission.

The FDA cleared this device based on its substantial equivalence to a predicate device (ViruLite Cold Sore Machine) and extensive non-clinical testing, including electrical safety, electromagnetic compatibility, photobiological safety, usability, and biocompatibility.

Here's a breakdown of what can be extracted and what cannot:

What can be extracted/inferred:

  • Acceptance Criteria (Implied): The acceptance criteria are implicitly met by demonstrating substantial equivalence to the predicate device and compliance with relevant safety and performance standards. The key performance measure for the intended use is "shortening the time to healing of herpes simplex labialis lesions on or around the lips with time to healing defined as the time to patient described re-epithelialization." The submission argues that since the new device has the same wavelength, energy density, and treatment schedule as the predicate, it is expected to achieve similar efficacy.
  • Device Performance (Reported): The device is reported to emit a controlled quantity of 1072nm +/-12nm peak wavelength near infrared light with a maximum peak light intensity of 20mW/cm² for approximately 3 minutes.
  • Sample size used for the test set and the data provenance: No clinical test set was used to evaluate efficacy directly. Usability testing was conducted with 45 initial subjects (3 withdrew for not being suitable), and then 42 subjects participated in a usability study and label comprehension. The provenance of this usability data is not specified beyond "simulated user's home environment".
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable for efficacy data, as no clinical study was performed. For usability, "researchers" made judgments, but their qualifications are not specified beyond their role in the study.
  • Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable for efficacy data. For usability, "investigator will record the operation of the subject" and "investigators will record and analyze responses," implying a direct observation/analysis by the researchers. No formal adjudication process with multiple independent reviewers is described.
  • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No MRMC study was done, as this is not an AI-assisted diagnostic device.
  • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is a physical light therapy device, not an algorithm.
  • The type of ground truth used: Not applicable for efficacy data. For usability, the ground truth was the ability of subjects to correctly self-select, perform device operations, and comprehend labeling based on researcher observation/evaluation.
  • The sample size for the training set: Not applicable, as no algorithm/AI training was performed.
  • How the ground truth for the training set was established: Not applicable, as no algorithm/AI training was performed.

Summary Table of Applicable Information (based on the provided document):

FeatureDetails (as per K222205)
1. Acceptance CriteriaImplied for Efficacy: Demonstrate substantial equivalence to predicate device (ViruLite Cold Sore Machine) by having comparable design principles, intended use, indications for use, functions, wavelength (1072nm +/- 12nm), energy density (20mW/cm2), treatment schedule (3X/day, 4 hours between treatments, for 2 consecutive days), auto-off feature, treatment time (3 min).
Reported Device PerformanceEmits 1072nm +/-12nm peak wavelength near infrared light. Max peak light intensity: 20mW/cm². Treatment duration: approx. 3 minutes. Treatment area: 3 cm². Power supply: Adapter (Input: 100-240Va.c., 50/60Hz, 0.35A; Output: 5.0Vd.c., 1.0A), Lithium battery, 3.7V, 350mAh.
2. Sample size for test set & Data ProvenanceEfficacy: No clinical test set used; therefore, no sample size for efficacy testing. Usability: 45 initial subjects (3 withdrew), 42 subjects for usability and label comprehension. Provenance not specified beyond "simulated user's home environment".
3. Number/Qualifications of Experts for Ground TruthEfficacy: Not applicable (no direct clinical study).Usability: "Researchers" identified as making judgments/recordings. Qualifications not specified.
4. Adjudication Method for Test SetEfficacy: Not applicable.Usability: Direct observation and analysis by "researchers/investigators." No formal multi-reviewer adjudication process described.
5. MRMC Comparative Effectiveness Study (AI vs. human)Not applicable (device is a light therapy device, not an AI-based diagnostic/assistive tool).
6. Standalone Performance (Algorithm only)Not applicable (device is a physical light therapy device, not an algorithm).
7. Type of Ground Truth UsedEfficacy: Not applicable (efficacy based on substantial equivalence to predicate, not direct clinical ground truth).Usability: Subject performance (correct self-selection, device operation, label comprehension) as observed/evaluated by "researchers."
8. Sample size for training setNot applicable (no algorithm/AI training).
9. Ground truth establishment for training setNot applicable (no algorithm/AI training).

In conclusion: The device was cleared based on its substantial equivalence to a legally marketed predicate device, along with extensive non-clinical and usability testing, as explicitly stated in the "Discussion of Clinical Tests Performed" section: "So there is no clinical test on our device." This type of submission relies on proving the new device is as safe and effective as a previously cleared device, rather than conducting new clinical trials to establish de novo efficacy.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 7, 2022

Light Tree Ventures Europe B.V. Alain Dijkstra Manager Laan van Ypenburg 108, 2497 GC, The Hague, The Netherlands Hague. Netherlands

Re: K222205

Trade/Device Name: Cold Sore Device (Model: QPZ-01) Regulation Number: 21 CFR 878.4860 Regulation Name: Light Based Energy Source For Topical Application Regulatory Class: Class II Product Code: OKJ Dated: July 25, 2022 Received: July 25, 2022

Dear Alain Dijkstra:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K22205

Device Name Cold Sore Device (Model: QPZ-01)

Indications for Use (Describe)

The Cold Sore Device is indicated for shortening the time to healing of herpes simplex labialis lesions on or around the lips with time to healing defined as the time to patient described re-epithelialization.

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary of K222205

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 878.4810.

1. Submitter's Information

Sponsor Name: Light Tree Ventures Europe B.V. Address: Laan van Ypenburg 108, 2497 GC, The Haque, The Netherlands Contact Person (including title): Alain Dijkstra (Manager) Establishment Registration Number: 3017422691 Tel: +86-135-10378748 Fax: +86-755-25024651 E-mail: regulation@kaiyanmedical.com

Factory:

Company Name: Shenzhen Kaiyan Medical Equipment Co., Ltd Address: Building 3, No.40, Fuxin street, Huaide Community, Fuyong Town,Baoan District, Shenzhen, Guangdong, 518103, China

Application Correspondent:

Contact Person: Alain Dijkstra Sponsor Name: Light Tree Ventures Europe B.V. Address: Laan van Ypenburg 108, 2497 GC, The Hague, The Netherlands Tel: +86 755 82129361 Fax: +86 755 25024651 Email: regulation@kaiyanmedical.com

2. Subject Device Information

Classification Name: Light based energy source device for topical application Trade Name: Cold Sore Device (Model: QPZ-01) Model Name: QPZ-01 Common Name: Light Based Treatment For Cold Sores Herpes Simplex Virus-1 Review Panel: General & Plastic Surgery Product Code: OKJ Regulation Number: 21 CFR 878.4860 Regulatory Class: II

3. Predicate Device Information

Sponsor: VIRULITE LLC Trade Name: ViruLite Cold Sore Machine (ViruLite) Classification Name: Light based energy source device for topical application 510(K) Number: DEN090012/K083767 Review Panel: General & Plastic Surgery Product Code: OKJ Regulation Number: 21 CFR 878.4860 Regulation Class: II

{4}------------------------------------------------

4. Device Description

The Cold Sore Device is a solid state opto-electronic device that emits a controlled quantity of 1072mm +/-12nm peak wavelength near infrared light for a period of approximately 3 minutes. The maximum peak light intensity across the treatment surface is 20mW/cm². The light output and duration are monitored by a microprocessor. The tip of the device that contacts the patient is made out of Acrylonitrile Butadiene Styrene (ABS) + PC.

Treatment with the device is commenced at the first symptoms of a cold sore 3 times a day with 4 hours in between each treatment for 2 consecutive days. The treatment area is approximately 3 cm². There are 2 light emitting diodes (LEDs) in the treatment area. The light within the device is activated by opening the cover of the device and automatically shut down after the preprogrammed treatment time (3 minutes). The device is designed for external, limited duration skin contact in an environment free from fluids and is provided non-sterile. The LEDs do not come in direct contact with the patient based upon the design of the device. The device is for OTC use and single patient use as described in the patient and box labeling.

5. Intended Use / Indications for Use

The Cold Sore Device is indicated for shortening the time to healing of herpes simplex labialis lesions on or around the lips with time to healing defined as the time to patient described re-epithelialization.

6. Test Summary

6.1 Non-Clinical Tests Performed

Non-clinical tests were performed on the subject device in order to validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety, and electromagnetic compatibility:

  • AAMI/ANSI ES60601-1 Medical electrical equipment Part 1: General requirements for basic safety � and essential performance.
  • � IEC 60601-1-11 Medical Electrical Equipment --Part 1: General Requirements for Basic Safety and Essential Performance --Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment.
  • � IEC 60601-2-57 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use.
  • � IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
  • � IEC 62471 Photobiological safety of lamps and lamp systems.
  • � IEC 62133-2 Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems.
  • � IEC 60601-1-6 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
  • � IEC 62366-1 Medical devices - Part 1: Application of usability engineering to medical devices
  • � ISO 10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.
  • ISO 10993-10 Biological evaluation of medical devices Part 10: Tests for irritation and skin � sensitization.

6.2 Discussion of Clinical Tests Performed

The intended use and technical parameters of the subject device are highly consistent with the predicate device (DEN090012/K083767), and important parameters such as output wavelength, energy density, treatment time, etc. are all consistent with the predicate device (DEN090012/K083767). At the same time,

{5}------------------------------------------------

the safety of the equipment has also been tested by safety regulations, EMC and performance. So there is no clinical test on our device.

6.3 Software Validation and Verification Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

The software for this device was considered as a "Moderate" level of concern, since a failure or latent design flaw could directly result in minor injury to the patient or operator.

6.4 Human Factors/Usability

According to "Applying Human Factors and Usability Engineering to Medical Devices: Guidance for Industry and Food and Drug Administration Staff", the researchers conducted simulated usability tests, including self-selection, device usability, and label comprehension.

At first, the subjects will make self-selection according to the outer packaging to determine whether the subjects can make a correct self-selection, and the researchers will make judgments about the selection of the subjects. Subjects who are identified as inappropriate equipment in self-selection will be withdrawn from the study. The results showed that all 45 subjects could make self-selection correctly, and 3 subjects were not suitable for the device and withdrew from the study.

Next, the 42 subjects will undergo a usability study. The researchers will simulate the user's home environment and have subjects simulate the operation of the investigator will record the operation of the subject to determine whether the using instructions provided on the user manual are understood by the subject. The results showed that all 42 subjects could perform the device usability operation correctly.

Final, the subjects will answer a series of questions related to label comprehension. Investigators will record and analyze responses to these questions to determine whether the information and instructions provided on the label were understood by the subjects. The results showed that all 42 subjects could correctly understand the information and instructions provided on the label.

After analyzing, usability test results (participating subjects could correctly perform self-selection, usability manipulation, and understanding of label content) showed that the self-selection, device usability, and label understanding of Cold Sore Devices were acceptable for patients of.

Elements ofComparisonSubject DevicePredicate DeviceRemark
CompanyShenzhen Kaiyan MedicalEquipment Co., LtdVIRULITE LLC--
Trade NameCold Sore Device (Model: QPZ-01)ViruLite Cold Sore Machine(ViruLite)--
ClassificationNameLight based energy source devicefor topical applicationLight based energy source devicefor topical application--
510(k) NumberK222205DEN090012/K083767--
Product CodeOKJOKJSame
Intended Use /Indications for UseThe Cold Sore Device is indicatedfor shortening the time to healingThe ViruLite Cold Sore Machine isindicated for shortening the timeSame

7. Comparison to predicate device and conclusion

Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject

{6}------------------------------------------------

Elements ofComparisonSubject DevicePredicate DeviceRemark
of herpes simplex labialis lesionson or around the lips with time tohealing defined as the time topatient described re-epithelialization.to healing of herpes simplexlabialis lesions on or around thelips with time to healing definedas the time to patient describedre-epithelialization.
Wavelengths1072nm +/- 12nm1072nm +/- 12nmSame
Energy SourceLEDLEDSame
TreatmentSchedule3X/day, 4 hours betweentreatments, for 2 consecutive day3X/day, 4 hours betweentreatments, for 2 consecutive daySame
Auto-off feature(after 3 mintreatment)YesYesSame
Treatment time3-minute treatment cycle3-minute treatment cycleSame
Treatment area3 cm27 cm2Different,note 1
Energy density(mw/cm2)20 mw/cm220 mw/cm2Same
Power supplyAdapter:Input: 100-240Va.c., 50/60Hz,0.35AOutput: 5.0Vd.c., 1.0ALithium battery, 3.7V, 350mAhalkaline 9V batteryDifferent,note 2
Location for UseOTCOTCSame
Safety and EMCIEC 60601-1IEC 60601-1-2IEC 60601-1-11IEC 62471IEC 60601-2-57IEC 62133-2IEC 60601-1IEC 60601-1-2IEC 60601-1-11IEC 62471Same
BiocompatibilityISO 10993-1ISO 10993-5ISO 10993-10ISO 10993-1ISO 10993-5ISO 10993-10Same

Note 1: Although the "Treatment area" is different from the predicate devices, the treatment area is determined by the size of the treatment window, the energy density across the treatment area is the same as that of the predicate device, so the it does not affect the effectiveness of the treatment. And they all complied with the IEC 62471 and IEC 60601-2-57 safety standards' requirements. So, these slight differences will not raise any safety or effectiveness issues.

Note 2: Although the "Power supply" is different from the predicate devices, they all complied with the IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57, IEC 62471 and IEC 62133-2 safety standards' requirements. So, these slight differences will not raise any safety or effectiveness issues.

Final Conclusion:

The subject device is the same or similar to the legally marketed predicate device DEN090012/K083767.

{7}------------------------------------------------

8. Date of the summary prepared: September 23, 2022

§ 878.4860 Light based energy source device for topical application.

(a)
Identification. The device emits light energy at near infrared spectrum and is applied externally to the surface of herpes simplex labialis lesions on or around the lips.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The technical parameters of the device, including wavelength, treatment time, treatment area, energy density, spot size, and power, must be characterized.
(2) The cleaning and disinfection instructions for the device must be validated.
(3) The device must be demonstrated to be biocompatible.
(4) Performance testing must validate electromagnetic compatibility (EMC), ocular safety, and electrical safety of the device.
(5) Labeling must direct end-users to contact the device manufacturer and MedWatch if they experience any adverse events when using this device.
(6) Labeling must include specific information pertinent to use of the device by the intended patient population and the treatment regimen.
(7) Simulated use testing must include information from a usability, label comprehension and self-selection study to demonstrate that the device can be used by the intended patient population without any assistance.
(8) Clinical data must show adequate reduction in time to healing and assess risks of redness, discomfort, burns, and blisters.