K Number
DEN090012

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-10-18

(1206 days)

Product Code
Regulation Number
878.4860
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ViruLite Cold Sore Machine is indicated for shortening the time to healing of herpes simplex labialis lesions on or around the lips with time to healing defined as the time to patient described re-epithelialization.

Device Description

The ViruLite is a solid state opto-electronic device that emits a controlled quantity of 1072nm +/- 12nm peak wavelength near infrared light for a period of approximately 3 minutes. The maximum peak light intensity across the treatment surface is 20mW/cm². The light output and duration are monitored by a microprocessor. The power source is a standard alkaline 9V battery. which is replaceable. The tip of the device that contacts the patient is made out of Acrylonitiile Butadiene Stvrene (ABS).

AI/ML Overview

Here's an analysis of the provided text to extract the acceptance criteria and study details for the ViruLite Cold Sore Machine:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the ViruLite Cold Sore Machine are primarily focused on demonstrating clinical benefit (shortening time to healing for herpes simplex labialis lesions) and safety. The clinical performance data provided supports these criteria.

Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance (as demonstrated in studies)
Clinical EffectivenessAdequate reduction in time to healing for herpes simplex labialis lesions on or around the lips, with time to healing defined as patient-described re-epithelialization. This is the primary endpoint for all studies.The ViruLite demonstrated a reduction in healing time compared to placebo in all four supporting datasets.
Clinical SafetyAssess risks of redness, discomfort, burns, and blisters.Low risk profile demonstrated. The requester's randomized clinical trial reported no major long-lasting complications, only "mild blistering, burning or redness in some patients." This was consistent with what was observed in the four additional submitted datasets and with other light-based technologies. The overall risk was determined to be low.
Usability/Label Comprehension
  • Device can be used by the intended patient population without assistance.
  • Labeling comprehension for OTC use.
  • Appropriate for ages 16 and older. | Demonstrated through two usability studies.
  • Combined, 47 out of 49 subjects correctly answered questions about appropriate actions in hypothetical scenarios, indicating good labeling comprehension and usability.
  • The studies were conducted on subjects 16 years of age or older, and it was determined that 16-year-olds could comprehend and follow instructions, and there was little physiological difference between 16-19 and 20+ year-olds. |
    | Biocompatibility | Device must be demonstrated to be biocompatible. | Acceptable. The ABS material used for patient contact was evaluated per ISO 10993-1:2003 for cytotoxicity, irritation, and sensitization and found acceptable, as it's a widely used medical material. |
    | Cleaning & Disinfection | Cleaning and disinfection instructions for the device must be validated. | Validation Required pre-market. The text states that "Cleaning and disinfection validation data to verify the effectiveness of the cleaning and disinfection instructions are required prior to marketing the ViruLite, otherwise the device is considered to be misbranded and adulterated." This implies it was an outstanding requirement for full market acceptance, not necessarily a completed study at the time of this document, but critical for final acceptance. Labeling includes instructions. |
    | Ocular Safety | Performance testing must validate ocular safety. | Acceptable. ViruLite was tested according to IEC 64721:2008 (despite not being an FDA-recognized standard, accepted due to similar prior devices) operating at worst-case continuous LED operation for 8 hours. The measured exposure limit was below the standard's acceptable limit. |
    | Electrical Safety | Performance testing must validate electrical safety. | Accepted. The device was subjected to electrical safety testing performed in accordance with EN60601-1:1993. |
    | Electromagnetic Compatibility (EMC) | Performance testing must validate EMC. | Accepted. The device was subjected to EMC testing performed in accordance with EN60601-1-2:1993. |
    | Software | Software performs as intended and all software-related risks have been adequately mitigated. | Acceptable. The software was deemed to have a minor level of concern. The requester provided comprehensive documentation (requirements, design, traceability, V&V tests, risk analysis) that provided reasonable assurance the software performs as intended. |
    | Technical Parameters | The technical parameters of the device, including wavelength, treatment time, treatment area, energy density, spot size, and power, must be characterized. | Characterized. The document describes: 1072nm +/- 12nm peak wavelength, ~3-minute treatment duration, 20mW/cm² maximum peak light intensity, ~7cm² treatment area, 2 LEDs. |
    | Labeling |
  • Complies with 21 CFR 801 and FDA's Guidance on Medical Device Patient Labeling (2001).
  • Direct users to contact manufacturer and MedWatch for adverse events.
  • Includes specific information pertinent to use by intended patient population and treatment regimen. | Complies. The document explicitly states compliance with 21 CFR 801 and the guidance. The MedWatch adverse event reporting statement is directly quoted in the text. The various warnings, precautions, contraindications, and instructions for use (e.g., "Do not use except for cold sores...", "not to prevent cold sores...", "not studied in immunocompromised...") demonstrate specific patient information. |

2. Sample Sizes Used for the Test Set and Data Provenance

The "test set" for clinical effectiveness and safety primarily refers to the patient data used in the clinical studies.

  • Requester's Randomized Clinical Trial:

    • Sample Size: Total of 95 subjects (ages 20 years and above).
      • 48 subjects in the active (treatment) arm.
      • 47 subjects in the control (placebo) arm.
    • Data Provenance: Not explicitly stated but implied to be a prospective, double-blind, placebo-controlled, randomized clinical trial conducted by the requester. Location is not given, but likely the country where the company is based or a region allowing such trials.
  • Four Additional Data Sets (Supporting Effectiveness):

    • Dougal and Kelly 2001: 56 subjects
    • Dougal and Haslam 2004: 54 subjects
    • Hargate 2006: 32 subjects
    • Lee, Hargate and Dougal 2004: 87 subjects
    • Data Provenance: These are listed as studies from 2001, 2004, and 2006. The exact country of origin is not explicitly mentioned but the authors' names (Dougal, Kelly, Haslam, Hargate, Lee) suggest a Western context (e.g., UK, USA, Australia). They are retrospective in the sense that they were submitted to support the De Novo request but were conducted prior to this specific submission. They are described as double-blind, placebo studies.
  • Usability Studies:

    • Sample Size:
      • First study: 9 subjects.
      • Second study: 40 additional subjects (totaling 49, with 2 withdrawals).
      • Combined: 47 subjects (passed) out of 49.
    • Data Provenance: Not specified, but likely conducted by the requester in a relevant country for OTC use within the target population. Prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The ground truth for the clinical effectiveness (time to healing) was primarily established by patient self-assessment. The primary endpoint was "time to healing" defined as "patient described re-epithelialization". Therefore, no external experts were explicitly mentioned for establishing this specific clinical ground truth.

4. Adjudication Method for the Test Set

Given that the primary endpoint was "patient described re-epithelialization," there was no formal external expert adjudication method (like 2+1 or 3+1 consensus) explicitly described for the clinical effectiveness data. The patients themselves provided the "ground truth" for healing time.

For internal study consistency and data integrity, there would have been study coordinators or investigators involved, but the ultimate "truth" for the primary endpoint came directly from the patient's report.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No obvious MRMC comparative effectiveness study was described. The studies compared the device to a placebo (or acyclovir in one instance for a specific arm), and the endpoint was patient-reported healing time, not an assessment by human readers. The concept of "human readers improving with/without AI assistance" does not apply here as it's a device for direct physical treatment, not an imaging or diagnostic AI tool.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, the primary clinical effectiveness studies are essentially "standalone" evaluations of the device. The ViruLite device is a physical light therapy device, not an algorithm. Its performance is measured directly by its effect on the patient (reduced healing time) without human intervention in its outcome assessment process beyond the patient's own reporting. The intervention itself (device use) is by the human (patient).

7. The Type of Ground Truth Used

  • Clinical Effectiveness (Healing Time): Patient-described re-epithelialization. (Outcomes data - specifically patient-reported outcome).
  • Clinical Safety: Patient self-reports of adverse events (redness, discomfort, blistering, burning), and overall absence of "major long-lasting complications."
  • Usability: Subject responses to questions regarding labeling comprehension and appropriate actions in hypothetical scenarios.

8. The Sample Size for the Training Set

The provided document does not discuss a "training set" in the context of an AI/machine learning model. The ViruLite is a hardware device with "software" to monitor button presses, control LEDs, and manage the 3-minute treatment cycle. This software is not a machine learning algorithm that learns from data. Therefore, the concept of a training set as typically understood for AI models does not apply here. The software review was about demonstrating it performs as intended, not about training an AI.

9. How the Ground Truth for the Training Set Was Established

As explained in point 8, there is no "training set" for an AI model. The software's "ground truth" in this context would be defined by the functional requirements in its software requirements specification (e.g., "ON button press correctly initiates treatment cycle," "LEDs activate at specified wavelength," "treatment duration is 3 minutes +/- tolerance"). These were established by the device's design specifications and verified through standard software and hardware testing (verification/validation tests), not through empirical data training.

{0}------------------------------------------------

DE NOVO CLASSIFICATION REQUEST FOR VIRULITE COLD SORE MACHINE

REGULATORY INFORMATION

FDA identifies this generic type of device as:

Light based energy source device for topical application. The device emits light energy at near infrared spectrum and is applied externally to the surface of herpes simplex labialis lesions on or around the lips.

NEW REGULATION NUMBER: 21 CFR 878.4860

CLASSIFICATION: II

PRODUCT CODE: OKJ

BACKGROUND

DEVICE NAME: ViruLite Cold Sore Machine (ViruLite)

SUBMISSION NUMBER: DEN090012 (K083767)

DATE OF DE NOVO: June 25, 2009

Virulite LLC CONTACT: 27631 Vista De Dons Capistrano Beach, CA 92624

INDICATIONS FOR USE

The ViruLite Cold Sore Machine is indicated for shortening the time to healing of herpes simplex labialis lesions on or around the lips with time to healing defined as the time to patient described re-epithelialization.

{1}------------------------------------------------

LIMITATIONS

The ViruLite device is available as an over-the-counter (OTC) device and is intended for use by a single individual. As an additional measure to ensure that the end users of this device report all adverse events that may occur with its usage, the labeling states the following:

"If you experience any of the following adverse events since using the device lack of effectiveness, spread or worsening of herpes infection, blistering and skin pigmentation - then please contact MedWatch at 1-800-332-1088 or the internet at: https://www.accessdata.fda.gov/scripts/medwatch/medwatch-online.htm. "

Limitations on the ViruLite device are also achieved through the following statements included in the User Guide:

Do not use the ViruLite except for the treatment of cold sores on or around the lips. This device is not to be used for other skin conditions especially genital herpes. The effect of the device for cold sores in other areas has not been tested and the risk is unknown.

Do not use the ViruLite to prevent cold sores on or around the lips area. The device is not intended to prevent cold sores. The safety of the device for the prevention of cold sores has not been tested and the risk is unknown.

Virulite has not been studied in patients who have been diagnosed by their physician as having difficulty in fighting infections (immunocompromised) and should therefore not be used by these patients.

ViruLite has not been evaluated on or around the lips of people with darker pigmented skin. If you have darker pigmented skin you may use the device, however stop using the device immediately if you encounter any problems or changes to your skin.

Do not move the ViruLite from one lesion to another without cleaning and disinfecting the device between uses. The effect of moving the device between lesions without cleaning and disinfection has not been tested and there is a risk that you may spread the virus.

PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS. PRECAUTIONS AND CONTRAINDICATIONS.

DEVICE DESCRIPTION

The ViruLite is a solid state opto-electronic device that emits a controlled quantity of 1072nm +/- 12nm peak wavelength near infrared light for a period of approximately 3 minutes. The maximum peak light intensity across the treatment surface is 20mW/cm². The light output and

{2}------------------------------------------------

duration are monitored by a microprocessor. The power source is a standard alkaline 9V battery. which is replaceable. The tip of the device that contacts the patient is made out of Acrylonitiile Butadiene Stvrene (ABS).

Image /page/2/Figure/1 description: The image shows a white device with a blue power button on the top. The device has a treatment cavity at the bottom with two light-emitting diodes inside. The text labels the power button, light-emitting diodes, and treatment cavity.

Figure 1: ViruLite Cold Sore Machine

Treatment with the ViruLite is commenced at the first symptoms of a cold sore 3 times a day with 4 hours in between each treatment for 2 consecutive days. The treatment area is approximately 7cm2. There are 2 light emitting diodes (LEDs) in the treatment area. The light within the device is activated by the ON button and automatically powers down after the preprogrammed treatment time (3 minutes). The device is designed for external, limited duration skin contact in an environment free from fluids and is provided non-sterile. The LEDs do not come in direct contact with the patient based upon the design of the device.

The software for the ViruLite device is designed to monitor ON button press by users, monitor ON button and near infrared LED control during treatment cycle and perform a 3-minute treatment cycle.

The device is for OTC use and single patient use as described in the patient and box labeling.

SUMMARY OF NONCLINICAL/BENCH STUDIES

BIOCOMPATIBILITY/MATERIALS

The only patient-contacting component of the ViruLite is comprised of Acrylonitrile Butadiene Styrene (ABS).

The raw material contained in the patient-contacting component of the ViruLite was evaluated with respect to its intended use per ISO 10993-1:2003. The raw material was evaluated for cvtotoxicity (ISO 10993-5), irritation and sensitization (ISO 10993-10). This was found acceptable because the ViruLite is used externally and ABS is a widely used material for products in medical use.

{3}------------------------------------------------

STERILIZATION/SHELF LIFE/REUSE

The ViruLite is not provided sterile, and sterilization for this use is not necessary because the device is not being used in a sterile environment. The ViruLite does not have a stated shelf life. Based on the nature of the components, the absence of a shelf life is acceptable.

Because the casing of the ViruLite does come in contact with the patient's lips and is intended for reuse by a single patient with herpes simplex labialis lesions, the Instructions for Use include sufficient cleaning and disinfection instructions to mitigate the risks of infection and transmissibility between uses.

Cleaning and disinfection validation data to verify the effectiveness of the cleaning and disinfection instructions are required prior to marketing the ViruLite, otherwise the device is considered to be misbranded and adulterated.

ELECTROMAGNETIC COMPATIBILITY AND ELECTRICAL SAFETY

The ViruLite was subjected to electromagnetic compatibility (EMC) and electrical safety testing performed in accordance to EN60601-1-2: 1993 and EN60601-1:1993.

SOFTWARE

The software for the ViruLite device is designed to monitor ON button press by users, monitor ON button and near infrared LED control during treatment cycle and perform a 3-minute treatment cycle.

The software for the ViruLite device presents a minor level of concern based on answers to the questions listed in FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005).

In the software requirements specification, the requester provides the general requirements and the functional requirements for traceability. The software design description provides the high level design of the ViruLite software application and firmware. The requester provided the traceability matrix and showed the traceability by testing the functional requirements through the software requirements specifications. verification/validation tests and risk analysis. The requester provided the verification and validation documentation. The documentation provides sufficient evidence that the specified requirements have been fulfilled.

All expected elements of software documentation per FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff are included in the request and in sufficient detail to provide reasonable assurance that the software performs as intended and all software-related risks have been adequately mitigated.

{4}------------------------------------------------

PERFORMANCE TESTING - BENCH TESTING FOR OCULAR SAFETY

The ViruLite was subjected to testing performed in accordance with IEC 64721:2008 Photobiological Safety of Lamps and Lamp Systems to establish ocular safety. ViruLite was tested while operating under the worst case failure mode, which is continuous operation of the LEDs for 8 hours.

The requester then took the measured exposure duration limit of the ViruLite and compared it to the exposure duration limit described in the IEC 64271:2008 standard to establish ocular safety. The exposure limit in IEC 64271:2008 is 2556097 W/m³/sr. Although this is not an FDA recognized consensus standard, the exposure duration limit described in this standard is an acceptable limit because experiences with low level light devices within the near infrared spectrum with similar technical parameters and exposure limits have lead us to conclude that this exposure limit and test method is acceptable to determine ocular safety.

The requester stated that based on testing in accordance with IEC 64271:2008 at the worse case failure mode the exposure limit of the ViruLite is (b) (4) W/m3/sr. This is below the exposure limit in IEC 64271:2008 and therefore deem acceptable in establishing ocular safety.

SUMMARY OF CLINICAL INFORMATION

To support the De Novo request, the requester provided usability testing as well as clinical data set information comparing the device to a placebo or the standard of care

SIMULATED USE (USABILITY) TESTING

The ViruLite device is intended for over-the-counter (OTC use. Consequently, to ensure that the ViruLite device could be used according to the proposed labeling and instructions for use, the requester conducted two usability studies.

At first, the requester completed their usability analysis on® subjects, which was determined to not be a sufficient sample size to demonstrate generalizability to the intended patient population. The requester, therefore, conducted a second study on (b) (4) additional subjects to determine the usability and labeling comprehension of the ViruLite device. The subjects in both studies were given the contents of the proposed labeling. This was followed by a series of questions that were asked to determine the subjects' understanding of the appropriate action to take with the ViruLite device in several hypothetical scenarios. The responses to the questions were recorded and analyzed to determine whether the information presented on the box label and instructions for use were comprehensible to the target population. In the first study out of " and in the second study (0) (0) out of (0) (0) (0) (0) subjects did not complete due to withdrawing from the study) correctly answered the questions presented. Therefore, the combined data (47 out of 49 subjects) in the two usability studies were acceptable to

{5}------------------------------------------------

address the labeling comprehension and usability of the ViruLite device for the OTC patient population.

The labeling allows for use of the ViruLite device for ages 16 years or older. In the clinical data sets, the ViruLite device was tested in patients 20 years or older to demonstrate a reasonable assurance of effectiveness and safety. Because the usability study was conducted on subjects 16 years of age or older and it was determined to include 16 year old patients for the following reasons: (i) patients 16 and older demonstrated an ability to comprehend and follow instructions written for OTC use within the usability study; and (ii) from a clinical perspective, it was determined that there is little difference in expected physiological response in response to patients between the ages of 16-19 and patients older than 20 years.

CLINICAL PERFORMANCE DATA

To support the De Novo request, the clinical performance data was provided from two main sources that included: 1) individual patient data (n = 95, 48 active arm) based on the requester conducting a double-blind, placebo-controlled, randomized clinical trial; and, 2) the requester's submission of four data sets. These data sets provided additional clinical evidence for using the ViruLite device and/or another related device using a similar wavelength and treatment parameters to the final device. Collectively these 5 data sets were used to support both device effectiveness (the four data sets) and safety (the four data sets and the requester's clinical data set).

Through utilization of data from these 5 studies it was determined that the clinical benefit outweighed any known risks. Therefore, FDA's granting of the De Novo request for the ViruLite for its intended use is based on an overall assessment of the clinical experience as demonstrated in the five studies.

The primary endpoint used in the evaluation of all studies provided was "time to healing" (subject described re-epithelialization) from the time of onset of the cold sore (as described by the subject).

The individual patient data from the requester's randomized data set in the De Novo request included a total of 95 subjects, ages 20 years and above, randomized either to control (placebo) or active (treatment with the ViruLite) arms. Of these 95 subjects, forty-eight (48) subjects were originally randomized to the active (treatment) arm and used to evaluate safety of the VireLite device. In the active (treatment) arm, 8 subjects were lost after the first follow-up visit and an additional 4 subjects were lost during the 3 month follow-up period. Of the 47 subjects randomized to control, 1 was lost after the first follow-up visit, and an additional 5 subjects were lost to follow-up during the 3 month follow up period.

It should be noted that in an evaluation of this data set, there were some concerns over data collection, missing data, data inconsistencies, and the interpretability of this dataset. Consequently, the results of the randomized data set did not by itself demonstrate a reasonable assurance of effectiveness

{6}------------------------------------------------

However, in a safety assessment of the requester's data set, there were no major longlasting complications (there was some mild blistering, burning or redness in some patients) associated with the ViruLite device in this clinical data set. Further, similar low risk complications have been observed with other light-based technologies and in the four submitted data sets by the requester. Therefore, FDA has determined that these low risks can be appropriately mitigated through the final labeling and other measures identified as special controls.

Due to the limitations of the initial randomized dataset provided above to determine device efficacy, the requester as noted above, submitted a total of four data sets for Agency review describing the use of ViruLite and similar devices on subjects who presented with herpes simplex labilis.

These studies were therefore used to support the effectiveness of the ViruLite device as well as confirming a low risk profile as seen in the requester's clinical trial. The four data sets (Dougal/Kelly data set, Dougal/Haslam data set, Hargate, and Lee/Hargate/Dougal data set) each identified a clinical benefit (a difference in healing - subject described reepithelialization) when comparing a reduction in healing time of the placebo and the active device.

The data from each of these trials were summarized in Table 1 below

TrialDesignProtocolDatasetSizeHealingTime inDays(Device)HealingTime inDays(Placebo)Endpoint
Dougaland Kelly20011Double Blind,Placebo: 3Arms withAcyclovirSingle Application vsAcyclovir or Placebo creamor Placebo light. 36 hourinitiation(within 36 hours ofthe onset of symptoms)564.3 +/- 1.88.5 +/- 1.8Healing Timewasdetermined byself assessment
DougalandHaslam20042Double Blindvs. PlaceboDeviceUp to 2 days, up to 3 TX. 36hour initiation (within 36hours of the onset ofsymptoms)545.1 +/- 3.16.8 +/- 2.8Same
Hargate20063DoubleBlindedPlaceboSame (as Dougal Haslam2004)326.3 +/- 3.09.4 +/- 4.6Same
Lee,HargateandDougal20044DoubleBlindedPlaceboSame (as Dougal Haslam2004)875.9 +/- 2.67.9 +/- 0.3Same

Table 1: Additional Data Sets to Support Effectiveness

  1. G Dougal and P Kelly. "A pilot study of treatment of herpes labialis with 1072 nm narrow waveband light". Clinical and Experimental Dermatology. 2006. 2. G Dougal and J Haslam. 2004 data set.

{7}------------------------------------------------

  1. G Hargate. "A randomized double-blind study comparing the effect of 1072-nm light against placebo for the treatment of herpes labialis" Clinical and Experimental Dermatology. 2006. 4. G Dougal, G Hargate, SY Lee. 2004 data set.

LABELING

The ViruLite device complies with the labeling requirements under 21 CFR 801 and the recommendations within FDA's Guidance on Medical Device Patient Labeling (2001).

The ViruLite device is available as an OTC device and is intended for use by a single individual.

As an additional measure to ensure that the end users of this device report all adverse events that may occur with its usage, the labeling states the following:

"If you experience any of the following adverse events since using the device- lack of effectiveness, spread or worsening of herpes infection, blistering and skin pigmentation then please contact MedWatch at 1-800-332-1088 or the internet at: https://www.accessdata.fda.gov/scripts/medwatch/medwatch-online.htm. "

RISKS TO HEALTH

Table 2 below identifies the risks to health that may be associated with the use of a light based energy source device for topical application and the measures necessary to mitigate these risks.

Identified RiskMitigation Measures
Redness and discomfortClinical performance testingUsability testingLabeling
Burns and blistersClinical performance testingUsability testingLabeling
Adverse tissue reactionBiocompatibility evaluation
Infection/transmissibilityLabelingCleaning and disinfection validationUsability testing
Electrical shockElectrical safety testingLabeling
Electromagnetic incompatibilityElectromagnetic compatibility testingLabeling
User errorUsability testingLabeling
Ocular injuryLabelingNon-clinical performance testing for ocular safety

Table 2: Identified Risks to Health and Mitigation Measures

{8}------------------------------------------------

SPECIAL CONTROLS

In combination with the general controls of the FD&C Act, the light based energy source device for topical application is subject to the following special controls:

    1. The technical parameters of the device, including wavelength, treatment time, treatment area, energy density, spot size, and power, must be characterized.
    1. The cleaning and disinfection instructions for the device must be validated.
    1. The device must be demonstrated to be biocompatible.
    1. Performance testing must validate electromagnetic compatibility (EMC), ocular safety, and electrical safety of the device.
    1. Labeling must direct end-users to contact the device manufacturer and MedWatch if they experience any adverse events when using this device.
    1. Labeling must include specific information pertinent to use of the device by the intended patient population and the treatment regimen.
    1. Simulated use testing must include information from a usability, label comprehension and self-selection study to demonstrate that the device can be used by the intended patient population without any assistance.
    1. Clinical data must show adequate reduction in time to healing and assess risks of redness, discomfort, burns, and blisters.

BENEFIT/RISK DETERMINATION

To determine the benefit FDA used the non clinical information submitted by the requester as well as the four data sets in Table 1 that are in the clinical section above. To determine the risk. FDA used the data from the five clinical data sets (the requester's clinical data set and the four data sets in Table 1). The primary risks (adverse events) in the data included: 1) potential redness; and, 2) some discomfort with application. Incorrect use or user error also led to a small number of minor (transient) burns and blistering. No serious or long lasting complications were reported. Therefore, FDA has evaluated the risk as low.

In summary, based on the submitted data (pre-clinical and clinical) the device provides a reasonable assurance of a clinical benefit, and the low device risks can be mitigated by the use of general and special controls, including the submitted final labeling for the indicated population.

{9}------------------------------------------------

CONCLUSION

The De Novo request for the ViruLite is granted and the device is classified under the following:

Product Code: OKJ Device Type: Light based energy source device for topical application Class: II Regulation: 21 CFR 878.4860

§ 878.4860 Light based energy source device for topical application.

(a)
Identification. The device emits light energy at near infrared spectrum and is applied externally to the surface of herpes simplex labialis lesions on or around the lips.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The technical parameters of the device, including wavelength, treatment time, treatment area, energy density, spot size, and power, must be characterized.
(2) The cleaning and disinfection instructions for the device must be validated.
(3) The device must be demonstrated to be biocompatible.
(4) Performance testing must validate electromagnetic compatibility (EMC), ocular safety, and electrical safety of the device.
(5) Labeling must direct end-users to contact the device manufacturer and MedWatch if they experience any adverse events when using this device.
(6) Labeling must include specific information pertinent to use of the device by the intended patient population and the treatment regimen.
(7) Simulated use testing must include information from a usability, label comprehension and self-selection study to demonstrate that the device can be used by the intended patient population without any assistance.
(8) Clinical data must show adequate reduction in time to healing and assess risks of redness, discomfort, burns, and blisters.