Intended Use

An examination glove is a disposable glove intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Description

Nitrile Examination Powder Free Glove, Blue Tested For Use With 32 Chemotherapy Drugs and Fentanyl Permeation Resistance Claim, Nitrile Examination Powder Free Glove, Black Tested For Use With 32 Chemotherapy Drugs and Fentanyl Permeation Resistance Claim

AI/ML Overview

This document is a 510(k) premarket notification for nitrile examination gloves and describes their performance against chemotherapy drugs and Fentanyl. It does not describe an AI/ML powered device, therefore, the requested information regarding AI/ML device acceptance criteria and study details cannot be extracted.

The document provides the following information:

  • Device: Nitrile Examination Powder Free Glove, Blue and Black
  • Intended Use: Disposable medical glove worn on the examiner's hand to prevent contamination between patient and examiner. Tested for use with specific chemotherapy drugs and Fentanyl.
  • Testing Standard: ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
  • Performance Data: Tables showing the minimum breakthrough detection time in minutes for 32 chemotherapy drugs and Fentanyl for both blue and black gloves. Most drugs show a breakthrough time of >240 minutes, but Carmustine and Thiotepa have significantly shorter breakthrough times.
  • Warnings/Cautions: Specific warnings against using Carmustine and cautions regarding the short breakthrough time for Thiotepa.

It does not contain information about:

  • Acceptance criteria or studies for an AI/ML powered device.
  • Sample sizes for test sets or data provenance in the context of AI/ML.
  • Number or qualifications of experts for ground truth establishment.
  • Adjudication methods.
  • Multi Reader Multi Case (MRMC) comparative effectiveness studies.
  • Standalone algorithm performance.
  • Types of ground truth for AI/ML.
  • Training set sample sizes or ground truth establishment for AI/ML.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 10, 2023

Top Glove Medical (Thailand) Co., Ltd. % David Lim Executive Director TG Medical (USA) Inc 165 North Aspan Avenue Azusa, California 91702

Re: K222129

Trade/Device Name: Nitrile Examination Powder Free Glove, Blue Tested For Use With 32 Chemotherapy Drugs and Fentanyl Permeation Resistance Claim, Nitrile Examination Powder Free Glove, Black Tested For Use With 32 Chemotherapy Drugs and Fentanyl Permeation Resistance Claim Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, OPJ, QDO Dated: July 4, 2022 Received: November 7, 2022

Dear David Lim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Allan Guan -S

For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222129

Device Name

Nitrile Examination Powder Free Glove, Blue Tested for Use with 32 Chemotherapy Drugs and Fentance Claim

Indications for Use (Describe)

An examination glove is a disposable glove intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

NoTest Chemotherapy DrugConcentration (mg/ml)Minimum breakthrough Detection Time (min)
1Blenoxane15>240
2Busulfan6>240
3Carboplatin10>240
4Carmustine (BCNU)3.316.7
5Cisplatin1>240
6Cyclophosphamide20>240
7Cytarabine100>240
8Dacarbazine10>240
9Daunorubicin HCl5>240
10Docetaxel10>240
11Doxorubicin HCl2.0>240
12Etoposide20>240
13Ellence2>240
14Fludarabine25>240
15Fluorouracil50>240
16Ganciclovir10>240
17Gemcitabine38>240
18Idarubicin HCl1>240
19Ifosfamide50>240
20Irinotecan20>240
21Mechlorethamine HCl1>240
22Melphalan5>240
23Methotrexate25>240
24Mitomycin C0.5>240
25Mitoxantrone HCl2>240
26Oxaliplatin5>240
27Paclitaxel6.0>240
28Rituximab10>240
29Thiotepa10.035.0
30Trisenox1>240
31Vincristine Sulfate1>240
32Vinorelbine10>240
Test DrugConcentrationMinimum breakthrough Detection Time (min)
Fentanyl Citrate100 mcg/2ml>240

Please note that Carmustine and Thiotepa have extremely low permeation times of 16.7 minutes respectively.

{3}------------------------------------------------

WARNING: Do not use Carmustine

CAUTION:

Testing showed an average breakthrough time of 35 minutes with Thiotepa.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K222129

Device Name

Nitrile Examination Powder Free Glove, Black Tested for Use with 32 Chemotherapy Drugs and Fentance Claim

Indications for Use (Describe)

An examination glove is a disposable glove intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

NoTest Chemotherapy DrugConcentration (mg/ml)Minimum breakthrough Detection Time (min)
1Blenoxane15>240
2Busulfan6>240
3Carboplatin10>240
4Carmustine (BCNU)3.314.2
5Cisplatin1>240
6Cyclophosphamide20>240
7Cytarabine100>240
8Dacarbazine10>240
9Daunorubicin HCl5>240
10Docetaxel10>240
11Doxorubicin HCl2.0>240
12Etoposide20.0>240
13Ellence2>240
14Fludarabine2>240
15Fluorouracil50>240
16Ganciclovir10>240
17Gemcitabine38>240
18Idarubicin HCl1>240
19Ifosfamide50>240
20Irinotecan20>240
21Mechlorethamine HCl1>240
22Melphalan5>240
23Methotrexate25>240
24Mitomycin C0.5>240
25Mitoxantrone HCl2>240
26Oxaliplatin5>240
27Paclitaxel6.0>240
28Rituximab10>240
29Thiotepa10.024.1
30Trisenox1>240
31Vincristine Sulfate1>240
32Vinorelbine10>240
Test DrugConcentrationMinimum breakthrough Detection Time (min)
Fentanyl Citrate100 mcg/2ml>240

Please note that Carmustine and Thiotepa have extremely low permeation times of 14.2 minutes respectively.

{5}------------------------------------------------

WARNING: Do not use Carmustine and Thiotepa

CAUTION:

Testing showed an average breakthrough time of 24.1 minutes with Thiotepa.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.