K Number
K221900
Date Cleared
2023-09-29

(456 days)

Regulation Number
862.1120
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stat Profile Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings and for point-of-care usage for quantitative determination of pH, Partial Pressure of Carbon Dioxide (pCO2), Partial Pressure of Oxygen (pO2), Hematocrit, Sodium, Chloride, Ionized Calcium, Ionized Magnesium, Gucose, and Lactate in heparinized capillary whole blood.

Indication for Use: pH, pCO2, pO2 measurements are used in the diagnosis and treatment of life-threatening acid base disturbances.

Hematocrit (Hct) measurements of the packed red blood cell volume are used to distinguish normal states, such as anemia and erythrocytosis.

Glucose (Glu) measurement is used in the diagnosis and treatment of carbohydrate metabolism distuding diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Lactate (lactic acid) measurement is used to evaluate the acid-base status of patients suspected of having lactic acidosis.

Sodium (Na) measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, or diseases involving electrolyte imbalance.

Potassium (K) measurements are used in the diagnosis and treatment of disease conditions characterized by low or high potassium levels.

Chloride (Cl) measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

Ionized Calcium (iCa) measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

Ionized Magnesium (iMg) measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia (abnormally high levels of magnesium).

Device Description

The Stat Profile Prime Plus Analyzer System is an analyzer for use in hospital laboratory and point-of-care settings. It consists of the analyzer, sensor cartridges, and thermal paper for an onboard printer. Optionally, it provides for reading of barcode labels (such as operator badges and data sheets).

The Stat Profile Prime Plus Analyzer has slots to accommodate two sensor cartridges (Primary and Auxiliary). The analyzer will determine the configuration of the system by detecting which sensor cards are installed.

Primary Sensor Card Port:
There are two options for the primary sensor card:

  • Primary Sensor Card 1 shall enable and report the following listed analytes: .
    • PO2, PCO2, pH, Hct, tHb, SO2, O2Hb, COHb, MetHb, HHb, Glu, Lactate, Sodium, o Potassium, Chloride, Calcium, Ionized Magnesium
  • Primary Sensor Card 2 shall enable and report the following listed analytes: .
    • PO2, PCO2, pH, Hct, tHb, SO2, Glu, Lactate, Sodium, Chloride, Calcium, Ionized o Magnesium

Auxiliarv Sensor Card Port:
The reporting of Creatinine and BUN parameters (or not reporting them) shall be determined by the selection of the Auxiliary Sensor Card

  • . Auxiliary Sensor Card 1 shall enable the Creatinine and BUN parameters
  • Auxiliary Sensor Card 2 shall be a "dummy" sensor card and will not report any parameters. .

As with the predicate, the Stat Profile Prime Plus Analyzer is a blood gas, co-oximetry, electrolyte, chemistry, and hematology analyzer with an enhanced test menu and multiple quality control options. Both traditional internal and external quality control is available, as well as an on-board Quality Management System (QMS), and an electronic monitoring approach that ensures the analyzer is working properly.

The Stat Profile Prime Plus Analyzer accepts samples from syringes, open tubes, and capillary tubes. The sample size for analysis is 135 µL for the complete test panel or 90 µL for the capillary panel.

Sample collection, preparation and application to the analyzer are the same as for the previously cleared predicate. The end user can select which analytes are to be tested in the panel.

Stat Profile Prime Plus Analyzer System Components:
The Stat Profile Prime Plus Analyzer System is comprised of the following components.

  • . Stat Profile Prime Plus Analyzer System
  • Primary Sensor Cartridge .
  • Auxiliary Sensor Cartridge .
  • Stat Profile Prime Plus Auto-Cartridge Quality Control Pack
  • Stat Profile Prime Plus Calibrator Cartridge
  • Stat Profile Prime Plus External Ampule Control
  • . IFU/Labeling

Sample Types:
The Stat Profile Prime Plus Analyzer System accepts lithium heparinized arterial, venous, and capillary whole blood.

Measured Parameters:
The Stat Profile Prime Plus Analyzer measures:

  • . pH
  • . Partial Pressure of Carbon Dioxide (pCO2)
  • Partial Pressure of Oxygen (pO2) ●
  • Hematocrit (Hct) ●
  • . Glucose (Glu)
  • . Lactate (Lac)
  • Sodium (Na) ●
  • Potassium (K)
  • Chloride (CI)
  • . Ionized Calcium (iCa)
  • . lonized Magnesium (iMg)
AI/ML Overview

The Nova Biomedical Stat Profile Prime Plus Analyzer System is undergoing a 510(k) premarket notification to expand its indications for use to include capillary whole blood specimen testing for pH, pCO2, pO2, Sodium (Na+), Potassium (K+), Chloride (Cl-), Ionized Calcium (Ca2+), Ionized Magnesium (Mg2+), Glucose, Lactate, and Hematocrit. The study described focuses on demonstrating the substantial equivalence of the Stat Profile Prime Plus Analyzer system to its predicate device, the Nova Biomedical Stat Profile pHOx Ultra Analyzer, specifically for capillary whole blood samples.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for substantial equivalence are primarily demonstrated through method comparison and precision studies. While explicit numerical acceptance criteria for each parameter (e.g., specific ranges for slope, intercept, r-value in method comparison, or max SD/CV% for precision) are not directly stated in the provided text as a standalone table, the conclusion sections for each study indicate that the device "met the clinical accuracy acceptance criteria" or "met the performance criteria for precision." The reported performance is shown in the tables below, which are the primary evidence for meeting the implicit acceptance criteria.

Method Comparison (Clinical Accuracy - Comparison to Predicate Device)

ParameterN (Combined)Altered Samples (Combined)Whole Blood Range (Combined)SlopeInterceptr
pH249186.790-7.7290.98940.07360.9942
pO2, (mmHg)251207.5-567.11.00060.83200.9976
pCO2, (mmHg)245147.4-183.11.0075-0.59690.9968
Hct, (%)2411018-550.99000.80110.9876
Na, (mM)2431283.0-195.61.0129-2.22440.9885
K, (mM)245141.34-18.530.99400.04160.9987
Cl, (mM)2431264.5-191.60.99440.34940.9856
Ca, (mM)247160.37-2.460.99000.01550.9932
Mg, (mM)249180.13-1.220.96590.02140.9811
Glu, (mg/dL)2451428-4520.99500.90410.9969
Lac, (mM)243120.4-17.61.00010.01190.9989

Precision (Laboratory and Point-of-Care Settings)

The precision data is presented across multiple tables (Tables 4, 5, 6, 7, 8, 9, 10). Rather than reiterating all data here, the text explicitly states:

  • "The precision data for all samples in capillary mode met the within run and between analyzer imprecision specifications for the Prime Plus analyzers." (Summary of Capillary Mode Within Sample Precision)
  • "This study demonstrates the Stat Profile Prime Plus analyzer exhibits clinically acceptable imprecision specifications for pH, pCO2, pO2, sodium (Na+), chloride (C1-), potassium (K+), ionized calcium (Ca2+), ionized magnesium (Mg2+), glucose, lactate, and hematocrit measured by the Stat Profile Prime Plus Analyzer System in Capillary mode." (Conclusion of Within-Run Imprecision - Capillary Mode Fingerstick (External POC))
  • "The analyzer used for this evaluation met the performance criteria for within sample precision on capillary fingerstick specimens run by POC operators." (Conclusion of Within-Sample Imprecision - Capillary Mode Fingerstick (Internal POC))
  • "The Stat Profile Prime Plus analyzers provided consistently reliable performance throughout the evaluation study. The analyzers used for this evaluation met the acceptance criteria for precision." (Conclusion of Within-Run Imprecision - Capillary Mode)

The acceptance criteria are therefore implicitly met by the reported r-values nearing 1.0 and slopes nearing 1.0 with intercepts near 0 for method comparison, and the CV% and SD values falling within acceptable limits (though the limits themselves are not numerically specified in the provided text).

2. Sample Sizes Used for the Test Set and Data Provenance

  • Method Comparison Test Set (Capillary Mode):

    • For each measured parameter, the sample size (N) ranged from 118 to 123 at the ER site and 123 to 128 at the Hemodialysis site. The combined sample size (N) for each parameter ranged from 241 to 251.
    • Provenance: This was a prospective clinical study conducted at two external Point-of-Care (POC) sites within the United States (an Emergency Room and a Hemodialysis Unit). Some samples (less than 10%, indicating "Altered Samples" ranging from 5 to 10 for each site) were altered to cover the full dynamic range. These were "de-identified and discarded arterial blood specimens" for the external precision study (implicitly reflecting human samples, though the exact origin beyond "external POC site" is not specified beyond being collected from patients).
  • Precision Test Set (Capillary Mode):

    • Within Run Precision (Internal Lab): 20 replicates for each parameter, tested on two Prime Plus analyzers from venous blood transferred to capillary tubes. This appears to be lab-based, controlled samples.
    • Within Sample Precision (Internal Lab): 2 replicates from 30 different donors (Total N=60 for each analyte) of capillary whole blood. This implies human subjects.
    • Within-Run Imprecision (External POC): Sample analysis involved transferring discarded arterial blood specimens from a lithium heparin syringe to three balanced heparin capillary tubes. The number of unique discarded specimens is not explicitly stated but "each whole blood specimen" suggests multiple, distinct specimens were used.
    • Within-Sample Imprecision (Internal POC - Fingerstick): Capillary whole blood was collected via fingerstick puncture from individuals, with 2 replicates for each. N=60 for all sample pairs. This explicitly involves human subjects/donors.
    • Within-Run Imprecision (Internal Study - Lab): 5 different concentrations of deidentified venous whole blood specimens per analyte. Each concentration was run on 3 Prime Plus analyzers, 5 days, 1 run/day, 8 replicates/run/level. This totals 120 (5 concentrations * 3 analyzers * 5 days * 8 replicates) data points per analyte for the "N" value in Table 10. These are likely controlled lab samples simulating human blood.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The provided text does not explicitly state the number of experts used or their specific qualifications for establishing ground truth.

  • For the method comparison study, the predicate device (Nova Stat Profile pHOx Ultra Analyzer) serves as the "ground truth" or reference method for comparison. The performance of this predicate device itself is assumed to be established and accepted.
  • For the precision studies, the intrinsic analytical performance of the device is assessed, rather than against a human expert's interpretation.

4. Adjudication Method for the Test Set

This information is not applicable as the device measures objective chemical and physical parameters rather than interpreting images or clinical signs that would require human adjudication. The "ground truth" is the measurement from the predicate device or the inherent value in the sample for precision studies.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, Effect Size

This information is not applicable as the device is an in-vitro diagnostic (IVD) analyzer for quantitative measurements, not an AI imaging or diagnostic algorithm requiring human reader performance studies. The study focuses on instrument performance and equivalence rather than human reader improvement with AI assistance.

6. If a Standalone (Algorithm Only Without Human-in-the Loop Performance) Was Done

Yes, the studies conducted (method comparison and precision) are standalone performance evaluations of the device's accuracy and precision in measuring the analytes. There is no "human-in-the-loop" aspect to the analytical performance being evaluated; the device provides direct quantitative measurements.

7. The Type of Ground Truth Used

  • Method Comparison: The "ground truth" or reference standard for comparison was the predicate device, the Nova Stat Profile pHOx Ultra Analyzer. This is a comparative method where the new device's performance is assessed against an already legally marketed and accepted device.
  • Precision Studies: The "ground truth" for precision is the measured value itself and its statistical variation across multiple runs or samples. It's an assessment of the device's inherent reproducibility and repeatability, not against an external truth source like pathology or outcomes data. Human samples (venous and capillary whole blood) were used to test performance under realistic conditions.

8. The Sample Size for the Training Set

The provided text does not mention a training set as this is not a machine learning or AI-driven device in the sense of requiring an explicit training phase with labeled data in the way an imaging algorithm would. This is an analytical instrument based on established sensor technology and algorithms. Therefore, discussions of training sets and their sample sizes are typically not relevant for this type of device submission. The device uses "the same sensor technology, measurement algorithms, formulations of the internal and external controls, and calibrator cartridge" as its predicate, implying a well-established design.

9. How the Ground Truth for the Training Set Was Established

As no training set is discussed or implied to be applicable for this type of analytical device in the provided context, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

September 29, 2023

Nova Biomedical Corporation Bobby Zinck Sr. Regulatory Affairs Manager 200 Prospect St. Waltham, Massachusetts 02454

Re: K221900

Trade/Device Name: Stat Profile Prime Plus Analyzer System Regulation Number: 21 CFR 862.1120 Regulation Name: Blood Gases (pCO2, pO2) and Blood pH Test System Regulatory Class: Class II Product Code: CHL, JPI, CGA, KHP, JGS, CEM, CGZ, JFP, CFA Dated: February 17, 2023 Received: February 21, 2023

Dear Bobby Zinck:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Paula V. Caposino -2

Paula Caposino, Ph.D. Acting Division Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221900

Device Name Stat Profile Prime Plus Analyzer System

Indications for Use (Describe)

Intended Use:

The Stat Profile Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings and for point-of-care usage for quantitative determination of pH, Partial Pressure of Carbon Dioxide (pCO2), Partial Pressure of Oxygen (pO2), Hematocrit, Sodium, Chloride, Ionized Calcium, Ionized Magnesium, Gucose, and Lactate in heparinized capillary whole blood.

Indication for Use: pH, pCO2, pO2 measurements are used in the diagnosis and treatment of life-threatening acid base disturbances.

Hematocrit (Hct) measurements of the packed red blood cell volume are used to distinguish normal states, such as anemia and erythrocytosis.

Glucose (Glu) measurement is used in the diagnosis and treatment of carbohydrate metabolism distuding diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Lactate (lactic acid) measurement is used to evaluate the acid-base status of patients suspected of having lactic acidosis.

Sodium (Na) measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, or diseases involving electrolyte imbalance.

Potassium (K) measurements are used in the diagnosis and treatment of disease conditions characterized by low or high potassium levels.

Chloride (Cl) measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

Ionized Calcium (iCa) measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

Ionized Magnesium (iMg) measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia (abnormally high levels of magnesium).

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K221900 510(k) Summary

510(K) Owner: Nova Biomedical Corporation

Registration Number: 1219029

Address: 200 Prospect St. Waltham, MA 02454

Phone: 781-894-0800

784-891-4806 Fax Number:

Contact Person: Bobby Zinck

Date Prepared: September 28, 2023

Proprietary Name:

Stat Profile Prime Plus Analyzer System

Common or Usual Name:

Blood and Blood Gas Analyzer

Classification Name: Multiple

Classification NameRegulation #ClassProduct CodePanel
Blood Gases and Blood pH System862.1120IICHLChemistry (75)
Hematocrit measuring device864.6400IIJPIHematology (81)
Glucose Test System864.1345IICGAHematology (81)
Lactic Acid Test System*862.1450IKHPChemistry (75)
Sodium Test System862.1665IIJGSChemistry (75)
Potassium Test System862.1600IICEMChemistry (75)
Chloride Test System862.1170IICGZChemistry (75)
Calcium Test System862.1145IIJFPChemistry (75)
Magnesium Test System862.1495Class I, reservedCFAChemistry (75)
  • Meets Limitations to Exemptions 21 CFR § 862.9(c)(9)

Predicate Device: The predicate device for this submission is the Nova Biomedical Stat Profile pHOx Ultra Analyzer which received clearance through 510(k) K110648.

Purpose for Submission:

The purpose of this submission is to expand the indication for use to include capillary whole blood specimen testing for the following analytes tested on the Stat Profile Prime Plus Analyzer System, pH, pCO2, pO2, sodium (Nat), potassium (Kt), chloride (Cl-), ionized calcium (Ca2+), ionized magnesium (Mg2*), glucose, lactate, and hematocrit. The full panel of analytes measured on the Stat Profile Prime Plus including the ones listed in this submission previously received FDA clearance for point-of-care testing with arterial and venous whole blood specimens as a part of a bundled submission under the 510(k)s: K193246, K200204, K200349, K200403.

To support the expanded claims, a point-of-care study at 2 clinical sites was conducted using capillary specimens tested on the Stat Profile Prime Plus Analyzer and compared to the predicate device, Stat Profile pHOx Ultra Analyzer.

{5}------------------------------------------------

Device Description:

Stat Profile Prime Plus Analyzer

The Stat Profile Prime Plus Analyzer System is an analyzer for use in hospital laboratory and point-of-care settings. It consists of the analyzer, sensor cartridges, and thermal paper for an onboard printer. Optionally, it provides for reading of barcode labels (such as operator badges and data sheets).

The Stat Profile Prime Plus Analyzer has slots to accommodate two sensor cartridges (Primary and Auxiliary). The analyzer will determine the configuration of the system by detecting which sensor cards are installed.

Primary Sensor Card Port:

There are two options for the primary sensor card:

  • Primary Sensor Card 1 shall enable and report the following listed analytes: .
    • PO2, PCO2, pH, Hct, tHb, SO2, O2Hb, COHb, MetHb, HHb, Glu, Lactate, Sodium, o Potassium, Chloride, Calcium, Ionized Magnesium
  • Primary Sensor Card 2 shall enable and report the following listed analytes: .
    • PO2, PCO2, pH, Hct, tHb, SO2, Glu, Lactate, Sodium, Chloride, Calcium, Ionized o Magnesium

Auxiliarv Sensor Card Port:

The reporting of Creatinine and BUN parameters (or not reporting them) shall be determined by the selection of the Auxiliary Sensor Card

  • . Auxiliary Sensor Card 1 shall enable the Creatinine and BUN parameters
  • Auxiliary Sensor Card 2 shall be a "dummy" sensor card and will not report any parameters. .

As with the predicate, the Stat Profile Prime Plus Analyzer is a blood gas, co-oximetry, electrolyte, chemistry, and hematology analyzer with an enhanced test menu and multiple quality control options. Both traditional internal and external quality control is available, as well as an on-board Quality Management System (QMS), and an electronic monitoring approach that ensures the analyzer is working properly.

The Stat Profile Prime Plus Analyzer accepts samples from syringes, open tubes, and capillary tubes. The sample size for analysis is 135 µL for the complete test panel or 90 µL for the capillary panel.

Sample collection, preparation and application to the analyzer are the same as for the previously cleared predicate. The end user can select which analytes are to be tested in the panel.

Stat Profile Prime Plus Analyzer System Components:

The Stat Profile Prime Plus Analyzer System is comprised of the following components.

  • . Stat Profile Prime Plus Analyzer System
  • Primary Sensor Cartridge .
  • Auxiliary Sensor Cartridge .
  • Stat Profile Prime Plus Auto-Cartridge Quality Control Pack
  • Stat Profile Prime Plus Calibrator Cartridge
  • Stat Profile Prime Plus External Ampule Control
  • . IFU/Labeling

Sample Types:

The Stat Profile Prime Plus Analyzer System accepts lithium heparinized arterial, venous, and capillary whole blood.

Measured Parameters:

The Stat Profile Prime Plus Analyzer measures:

  • . pH
  • . Partial Pressure of Carbon Dioxide (pCO2)

{6}------------------------------------------------

  • Partial Pressure of Oxygen (pO2) ●
  • Hematocrit (Hct) ●
  • . Glucose (Glu)
  • . Lactate (Lac)
  • Sodium (Na) ●
  • Potassium (K)
  • Chloride (CI)
  • . Ionized Calcium (iCa)
  • . lonized Magnesium (iMg)

Intended Use:

The Stat Profile Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings and for point-of-care usage for quantitative determination of pH, Partial Pressure of Carbon Dioxide (pCO2), Partial Pressure of Oxygen (pO2), Hematocrit, Sodium, Potassium, Chloride, lonized Calcium, lonized Magnesium, Glucose, and Lactate in heparinized capillary whole blood.

Indication for Use:

pH, pC02, pO2Measurements are used in the diagnosis and treatment of life-threatening acid basedisturbances.
HctHematocrit (Hct) measurements of the packed red blood cell volume are used todistinguish normal from abnormal states, such as anemia and erythrocytosis.
Glucose (Glu)Glucose measurement is used in the diagnosis and treatment of carbohydratemetabolism disturbances including diabetes mellitus, neonatal hypoglycemia, andidiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Lactate (Lac)Lactate (lactic acid) measurement is used to evaluate the acid-base status of patientssuspected of having lactic acidosis.
Sodium (Na)Sodium measurements are used in the diagnosis and treatment of aldosteronism,diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, ordiseases involving electrolyte imbalance.
Potassium (K)Potassium measurements are used in the diagnosis and treatment of diseaseconditions characterized by low or high potassium levels.
Chloride (CI)Chloride measurements are used in the diagnosis and treatment of electrolyte andmetabolic disorders such as cystic fibrosis and diabetic acidosis.
lonized Calcium(iCa)lonized calcium measurements are used in the diagnosis and treatment ofparathyroid disease, a variety of bone diseases, chronic renal disease and tetany(intermittent muscular contractions or spasms).
lonizedMagnesium(iMg)lonized magnesium measurements are used in the diagnosis and treatment ofhypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia(abnormally high levels of magnesium).

{7}------------------------------------------------

Principle of Measurement:

pH:

pH is measured using a hydrogen ion selective glass membrane. One side of the glass is in contact with a solution of constant pH. The other side is in contact with a solution of unknown pH. A change in potential develops which is proportional to the pH difference of these solutions. This change in potential is measured against a reference electrode of constant potential. The magnitude of the potential difference is a measure, then, of the pH of the unknown solution.

pCO2:

PCO2 is measured with a modified pH sensor. Carbon dioxide in the unknown solution makes contact with a gas permeable membrane mounted on a combination measuring/ reference electrode. CO2 diffuses across the membrane into a thin layer of electrolyte solution in response to partial pressure difference. This solution then becomes equilibrated with the external gas pressure. CO2 in the solution becomes hydrated producing carbonic acid. which results in a change in hydrogen ion activity.

pO2:

PO2 is measured amperometrically by the generation of a current at the sensor surface. As oxygen diffuses through a gas permeable membrane, the oxygen molecules are reduced at the cathode, consuming 4 electrons for every molecule of oxygen reduced. This flow of electrons is then measured by the sensor and is directly proportional to the partial pressure of oxygen.

Hematocrit:

Hematocrit is defined as the percentage of red blood cells to the total blood volume and can be obtained by measuring electrical resistance of the blood sample. Two standard solutions are used to calibrate the hematocrit sensor and to obtain the slope. The analyzer the electrical resistance of the blood sample to obtain the hematocrit value. The hematocrit value obtained is corrected for the concentration of the sodium ion.

Glucose:

Glucose measurement is based on the level of H₂O₂ produced during the enzymatic reaction between glucose and oxygen molecules in the presence of the glucose oxidase enzyme. At a constant potential of 0.70 volts, electroactive H2O2 is oxidized at the surface of the platinum anode. The current generated by the flow of electrons at the surface of the platinum electrode is proportional to the glucose concentration of the sample.

Lactate:

Lactate measurement is based on the level of HzOz produced during the enzymatic reaction between lactate and oxygen molecules in the presence of the lactate oxidase enzyme. At a constant potential of 0.70 volts, electroactive H2O2 is oxidized at the surface of the platinum anode. The current generated by the flow of electrons at the surface of the platinum electrode is proportional to the lactate concentration of the sample.

Sodium, Potassium, Chloride, Ionized Calcium, Ionized Magnesium:

The parameters are measured by an Ion-Selective Electrode (ISE) that selectively measures the activity of ionic species. When the ISE is contacted with a sample, potential is developed. The potential is proportional to the logarithm of the ionic activity and is measured versus a reference electrode.

Summary of Technological Characteristics:

The Stat Profile Prime Plus Analyzer System is substantially equivalent to the previously cleared for market Stat Profile pHOx Ultra Analyzer System (K110648). It uses the same sensor technology, measurement algorithms, formulations of the internal and external controls, and callbrator cartridge for the tested parameters.

Summary of Performance Testing:

Bench testing was previously completed and summarized in the following Prime Plus Analyzer System bundled 510(k)s K173797, K180186, K180340, K180428, K200349 to demonstrate that the Stat Profile Prime Plus Analyzer is substantially equivalent in performance, safety, and efficacy in the predicate submission.

The results of that testing confirmed that the performance of the Stat Profile Prime Plus Analyzer System Page 4 of 15

{8}------------------------------------------------

is substantially equivalent to the Nova Stat Profile pHOx Ultra Analyzer System (predicate device),

Summary of Point-of-Care Testing:

Point-of-Care testing for arterial and venous whole blood samples was previously completed and summarized in the following Prime Plus Analyzer System bundled 510(k)s K193246, K200349, K200403 to demonstrate that the system is safe and effective for use in the POC setting.

The testing included method comparison studies in 3 POC sites, including a Cardiothoracic Intensive Care Unit (CTICU), an Emergency Department (ED) and a Respiratory Therapy Lab (RT).

The results of the POC clinical performance verification testing confirmed that the Stat Profile Prime Plus Analyzer is safe and effective for its intended purpose in POC settings and that the Stat Profile Prime Plus Analyzer System is substantially equivalent to that of the predicate Stat Profile Prime Plus Analyzer System.

Summary of Capillary Mode Testing: Method Comparison:

A separate clinical study was conducted in Point-of-Care (POC) settings to demonstrate the clinical performance of capillary whole blood specimens on the Stat Profile Prime Plus analyzers. The study was conducted to demonstrate the clinical performance of capillary whole blood specimens on the Stat Profile Prime Plus analyzers. The study compared the Stat Profile Prime Plus analyzers to the Nova Stat Profile pHOx Ultra analyzers (predicate device) to assess the equivalence of the analyzers in the measurement of pH, blood gases, hematocrit, electrolytes, and metabolites in capillary whole blood specimens. The method comparison study was performed at two (2) external POC sites. The study was conducted in the Emergency Room (ER) with five (5) POC nurses and within a Hemodialysis Unit with four (4) POC nurses over a minimum of 20 days. Each site used a single Prime Plus analyzer and a single pHOx Ultra analyzer. Some of the specimens were altered (less than 10%) to cover the full dynamic range.

Capillary Mode Method Comparison
Stat Profile Prime Plus vs pHOx Ultra (ER)
ParameterNAlteredSamplesWhole BloodRangeSlopeInterceptr
pH12296.790-7.7290.99490.03210.9914
pO2, (mmHg)123107.5-562.21.0109-1.32810.9965
pCO2, (mmHg)12077.4-183.01.0003-0.31350.9958
Hct, (%)118518-540.99440.78310.9864
Na, (mM)119683.0-194.81.0158-2.74730.9907
K, (mM)12071.34-18.470.99240.04610.9984
Cl, (mM)119664.5-191.10.95614.79100.9810
Ca, (mM)12180.37-2.440.98710.02220.9932
Mg, (mM)12290.13-1.170.96900.01870.9808
Glu, (mg/dL)120728.0-427.50.99550.44360.9959
Lac, (mM)11960.5-16.91.0148-0.01080.9988

Table 1: Capillary Mode Method Comparison: ER

{9}------------------------------------------------

Capillary Mode Method Comparison
Stat Profile Prime Plus vs pHOx Ultra (Hemodialysis)
ParameterNAlteredSamplesWhole BloodRangeSlopeInterceptr
pH12796.790-7.6550.98390.11480.9971
pO2, (mmHg)128108.1-567.10.99362.39800.9987
pCO2, (mmHg)12577.4-183.11.0125-0.81970.9975
Hct, (%)123518-550.99440.78310.9864
Na, (mM)124685.0-195.61.0097-1.65830.9866
K, (mM)12571.43-18.530.99560.03690.9990
Cl, (mM)124667.2-191.61.0235-3.07250.9897
Ca, (mM)12680.38-2.460.99350.00820.9935
Mg, (mM)12790.16-1.220.96270.02410.9813
Glu, (mg/dL)125728.5-452.00.99481.30660.9979
Lac, (mM)12460.4-17.60.98670.03070.9991

Table 2: Capillary Mode Method Comparison: Hemodialysis

Table 3: Capillary Mode Method Comparison: Combined Results

Capillary Mode Method Comparison
Stat Profile Prime Plus vs pHOx Ultra (Combined)
ParameterNAlteredSamplesWhole BloodRangeSlopeInterceptr
pH249186.790-7.7290.98940.07360.9942
p02, (mmHg)251207.5-567.11.00060.83200.9976
pCO2, (mmHg)245147.4-183.11.0075-0.59690.9968
Hct, (%)2411018-550.99000.80110.9876
Na, (mM)2431283.0-195.61.0129-2.22440.9885
K, (mM)245141.34-18.530.99400.04160.9987
Cl, (mM)2431264.5-191.60.99440.34940.9856
Ca, (mM)247160.37-2.460.99000.01550.9932
Mg, (mM)249180.13-1.220.96590.02140.9811
Glu, (mg/dL)2451428-4520.99500.90410.9969
Lac, (mM)243120.4-17.61.00010.01190.9989

Conclusion: The Nova Biomedical Stat Profile Prime Plus analyzer provided consistently reliable performance throughout the POC clinical evaluation study. The capillary blood comparison data for all test parameters for the Stat Profile Prime Plus analyzers met the clinical accuracy acceptance criteria for correlation coefficient (r) and bias at the medical decision levels.

Total Imprecision Performance:

Within Run Precision:

Estimates for total imprecision for within run precision were obtained from running 20 replicates of venous blood transferred to capillary tubes at targeted sample concentrations on two Prime Plus analyzers in a Customer Simulation Laboratory at Nova Biomedical. The average SD and CV% for each analyzer for each sample type and level was calculated and compared to the defined within run imprecision specifications. The Between Analyzer mean, SD, and CV% from two (2) analyzers for each level was calculated and compared to the defined

{10}------------------------------------------------

between analyzer imprecision specifications. Within run data are summarized in Tables 3 to 5.

Prime Plus Capillary Mode
Whole Blood Within Run Precision
Parametern = 20Analyzer 1Analyzer 2Between Analyzer
pHMean7.3587.3427.350
SD0.0060.0060.009
pHMean7.2527.246
SD0.0040.0040.005
pHMean7.1427.144
SD0.0040.0050.005
pCO2, mmHgMean25.024.7
SD0.60.50.6
CV%2.41.92.2
pCO2, mmHgMean54.859.657.2
SD1.10.82.6
CV%2.01.34.5
pCO2, mmHgMean70.273.772.0
SD1.81.32.4
CV%2.61.83.3
pO2, mmHgMean54.356.055.2
SD0.70.61.1
CV%1.31.02.0
pO2, mmHgMean178.4174.3176.4
SD1.81.62.7
CV%1.00.91.5
pO2, mmHgMean95.0100.797.9
SD1.71.53.3
CV%1.71.53.4

Table 4: Capillary Mode Within Run Precision: Blood Gas Summary

Table 5: Capillary Mode Within Run Precision: Electrolytes Summary

Prime Plus Capillary Mode
Whole Blood Within Run Precision
Parametern = 20Analyzer 1Analyzer 2Between Analyzer
Na+, mmol/LMean87.286.286.7
SD0.50.80.8
CV%0.61.01.0
Na+, mmol/LMean155.0155.6155.3
SD0.41.00.8
Prime Plus Capillary Mode
Whole Blood Within Run Precision
Parametern = 20Analyzer 1Analyzer 2Between Analyzer
CV%0.30.60.5
Na+, mmol/LMean134.5135.9135.2
SD0.40.60.9
CV%0.30.40.6
K+, mmol/LMean1.921.931.92
SD0.020.020.02
CV%1.11.11.1
K+, mmol/LMean6.176.116.14
SD0.110.110.11
CV%1.81.81.8
K+, mmol/LMean4.574.594.58
SD0.040.030.04
CV%0.90.70.8
Cl-, mmol/LMean72.171.771.9
SD0.50.90.8
CV%0.71.21.0
Cl-, mmol/LMean130.8131.5131.2
SD1.21.01.2
CV%0.90.70.9
Cl-, mmol/LMean102.7102.0102.3
SD0.40.50.6
CV%0.40.50.6
iCa, mmol/LMean0.310.290.30
SD0.010.010.01
CV%2.23.03.9
iCa, mmol/LMean2.592.622.61
SD0.030.060.05
CV%1.12.21.8
iCa, mmol/LMean1.331.351.34
SD0.010.010.02
CV%0.60.71.3
iMg, mmol/LMean0.130.130.13
SD0.010.000.00
CV%4.31.73.4
iMg, mmol/LMean0.911.000.95
SD0.020.010.05
CV%1.71.35.2
iMg, mmol/LMean0.580.600.59
Prime Plus Capillary Mode
Whole Blood Within Run Precision
Parametern = 20Analyzer 1Analyzer 2Between Analyzer
SD0.010.010.01
CV%1.31.22.4

{11}------------------------------------------------

{12}------------------------------------------------

Table 6: Capillary Mode Within Run Precision: Hematology and Chemistry Summary

Prime Plus Capillary Mode
Whole Blood Within Run Precision
Parametern = 20Analyzer 1Analyzer 2Between Analyzer
Hct, %Mean282828
Hct, %SD0.70.60.6
Hct, %Mean606361
Hct, %SD0.31.31.6
Hct, %Mean424342
Hct, %SD0.90.70.9
Glu, mg/dLMean919392.4
Glu, mg/dLSD2.32.22.4
Glu, mg/dLCV%2.52.42.5
Glu, mg/dLMean282279280.5
Glu, mg/dLSD4.22.63.8
Glu, mg/dLCV%1.50.91.3
Glu, mg/dLMean363636.1
Glu, mg/dLSD0.40.40.4
Glu, mg/dLCV%1.01.11.1
Lac, mmol/LMean1.81.81.79
Lac, mmol/LSD0.10.10.07
Lac, mmol/LCV%3.84.34.0
Lac, mmol/LMean4.24.44.3
Lac, mmol/LSD0.10.10.1
Lac, mmol/LCV%3.42.03.2
Lac, mmol/LMean9.69.79.6
Lac, mmol/LSD0.30.30.3
Lac, mmol/LCV%2.72.72.7

Within Sample Precision:

Sample within run precision in capillary mode was assessed for whole blood by measuring two replicates of capillary whole blood from thirty (30) different donors in a Customer Simulation Laboratory at Nova Biomedical. The mean, SD, and CV% for all samples were calculated and compared to the defined withinrun imprecision specifications. Sample within run data is summarized in Table 7.

{13}------------------------------------------------

Prime Plus Capillary ModeWhole Blood Within Sample Precision
AnalyteSD%CVnMeanSD%CV
pH0.010607.3220.003--------
pCO2 (mmHg)1.03.06031.00.72.1
pO2 (mmHg)2.03.060127.90.80.6
Hct (%)1.060320.4--------
Na+ (mM)1.01.060137.90.10.1
K+ (mM)0.21.5604.030.010.21
Cl- (mM)1.52.060109.50.50.5
iCa (mM)0.072.5601.110.000.44
iMg (mM)0.032.0600.630.011.58
Glu (mg/dL)2.03.0601080.50.5
Lac (mM)0.33.0604.10.00.6

Table 7: Capillary Mode Within Sample Precision: Summary

Conclusion: The precision data for all samples in capillary mode met the within run and between analyzer imprecision specifications for the Prime Plus analyzers. Each Prime Plus analyzer provided a consistently reliable performance throughout the evaluation study. The analyzers used for this evaluation met all performance criteria for precision.

Within-Run Imprecision - Capillary Mode Fingerstick (External POC):

A precision study was performed at an external site using discarded specimens transferred from a lithium heparin syringe to a capillary tube by POC (Point-of-Care) operators. The study used one Stat Profile Prime Plus Analyzer, and de-identified and discarded arterial blood specimens.

Sample Analysis

Each whole blood specimen was transferred from a syringe to three balanced heparin capillary tubes and analyzed in Capillary Micro mode on the Stat Profile Prime Plus Analyzer by one POC operator.

{14}------------------------------------------------

Data Analysis

The mean, SD, and CV% of the triplicate results of each sample were calculated.

AnalyteUnitGrand MeanGrand SDGrand CV%
pHpH units7.3570.008N/A
pCO2mmHg37.20.71.9
pO2mmHg145.62.11.5
Hct%31.51.2N/A
Na+mmol/L135.10.80.6
K+mmol/L3.940.040.9
Cl-mmol/L110.10.80.8
Ca2+mmol/L1.140.011.2
Mg2+mmol/L0.510.012.5
Glumg/dL1281.81.4
Lacmmol/L3.10.14.6

Table 8: Within-Run Imprecision - Capillary Mode Fingerstick Data Summary

Conclusion

This study demonstrates the Stat Profile Prime Plus analyzer exhibits clinically acceptable imprecision specifications for pH, pCO2, pO2, sodium (Na+), chloride (C1-), potassium (K+), ionized calcium (Ca2+), ionized magnesium (Mg2+), glucose, lactate, and hematocrit measured by the Stat Profile Prime Plus Analyzer System in Capillary mode.

Within-Sample Imprecision - Capillary Mode Fingerstick (Internal POC):

An internal precision study was performed in the Nova Customer Simulation Laboratory using a single Stat Profile Prime Plus analyzer. Capillary whole blood was collected via fingerstick puncture and run on the Stat Profile Prime Plus Analyzer in capillary mode by two (2) point-of-care (POC) operators (Nurse and Respiratory Therapist) in duplicate.

The mean, SD, and CV% for all sample pairs was calculated and compared to the defined imprecision specifications.

Table 9: Within-Sample Imprecision - Capillary Mode Fingerstick Data Summary

AnalyteMeanNWithin Sample (SD)Within Sample (%CV)
pH7.403600.008N/A
p02 (mmHg)81.8602.22.7
pCO2 (mmHg)32.0601.03.0
Hct (%)41601.0N/A
Na (mM)139.5600.80.6
K (mM)4.75600.173.5
Cl (mM)109.9600.60.6
Ca (mM)1.20600.010.9
Mg (mM)0.54600.011.1
Glu (mg/dl)109601.61.5
Lac (mM)1.7600.212.4

{15}------------------------------------------------

Conclusion

The Stat Profile Prime Plus analyzer provided a consistently reliable performance throughout the evaluation study. The analyzer used for this evaluation met the performance criteria for within sample precision on capillary fingerstick specimens run by POC operators.

Within-Run Imprecision - Capillary Samples (Internal Study)

An internal precision study was performed in the Nova Customer Simulation Laboratory using five (5) different concentrations of deidentified venous whole blood specimens per analyte, each run on three (3) Stat Profile Prime Plus analyzers, for five (5) days, with one (1) run performed each day and eight (8) replicates measured per run per level.

The mean, SD, and CV% for all samples were calculated and compared to the defined within-run imprecision specifications.

Table 10: Within-Run Imprecision - Capillary Mode

AnalyteLevelMeanNWithin Run (SD)Within Run (%CV)
pH17.1331200.003N/A
27.3411200.003N/A
37.4651200.004N/A
46.9331200.004N/A
57.6521200.009N/A
pO2(mmHg)147.91200.40.9%
2205.41200.80.4%
3415.41202.50.6%
4146.61200.70.5%
5521.61203.30.6%
pCO2(mmHg)184.01201.31.5%
239.51200.71.7%
321.91200.20.9%
4161.01202.31.4%
5117.91201.61.3%
Hct (%)1401200.8N/A
2641200.7N/A
3201200.6N/A
4301200.7N/A
5571200.8N/A
Na(mM)1107.71200.80.7%
2167.31201.10.7%
3181.21201.50.8%
4152.21200.60.4%
5128.91200.60.5%

{16}------------------------------------------------

AnalyteLevelMeanNWithin Run (SD)Within Run (%CV)
K (mM)13.341200.041.2%
28.801200.131.4%
31.791200.052.7%
414.951200.171.1%
56.101200.050.8%
Cl (mM)195.91200.70.8%
2161.21201.81.1%
3186.31202.01.1%
481.21200.91.1%
5131.71201.20.0%
Ca (mM)10.951200.011.4%
21.501200.021.3%
31.821200.021.4%
40.781200.011.1%
52.221200.031.3%
Mg (mM)10.801200.011.7%
21.131200.021.4%
30.301200.012.1%
40.561200.011.9%
51.381200.021.6%
Glu (mg/dl)1871200.91.0%21981203.51.8%34291203.90.9%41171201.61.4%5341200.72.0%Lac (mM)110.81200.22.3%22.11200.15.4%35.91200.12.4%413.61200.32.5%516.11200.31.9%
Glu (mg/dl)1871200.91.0%
21981203.51.8%
34291203.90.9%
41171201.61.4%
5341200.72.0%
Lac (mM)110.81200.22.3%22.11200.15.4%35.91200.12.4%413.61200.32.5%516.11200.31.9%
Lac (mM)110.81200.22.3%
22.11200.15.4%
35.91200.12.4%
413.61200.32.5%
516.11200.31.9%

Table 10: Within-Run Imprecision - Capillary Mode (continued)

Conclusion

The Stat Profile Prime Plus analyzers provided consistently reliable performance throughout the evaluation study. The analyzers used for this evaluation met the acceptance criteria for precision.

{17}------------------------------------------------

Table 11: Comparison of Predicate and Proposed Devices

CharacteristicPredicate:Proposed:
Indication For UseThe Stat Profile pHOx Ultra Analyzer withoutCO-Oximeter is intended for in vitro diagnosticuse by health care professionals and/or point-of-care usage in the quantitativedetermination of pH, PCO2, PO2, SO2%,Hematocrit (Hct), Hemoglobin (Hb) inheparinized whole blood; Na+, K'-, Cl-, Ca++,Mg++, Glucose (Glu), Lactate (Lac), BUN(Urea), and Creatinine (Great) in heparinizedwhole blood, serum, or plasma.The Stat Profile pHOx Ultra Analyzer with CO-Oximeter is intended for in vitro diagnostic useby health care professionals and for point-of-care usage in the quantitative determination ofpH, PCO2, PO2, SO2%, Hematocrit (Hct),total Hemoglobin (tHb), Oxyhemoglobin(O2Hb), Carboxyhemoglobin (COHb),Methemoglobin (MetHb), Deoxyhemoglobin(HHb), and total bilirubin (tBil) in heparinizedwhole blood; Nat, K'-, Cl-, Ca++, Mg++,Glucose (Glu), Lactate (Lac), BUN (Urea),and Creatinine (Creat) in Heparinized wholeThe Stat Profile Prime Plus Analyzer Systemis indicated for use by healthcareprofessionals in clinical laboratory settingsand for point-of-care usage for quantitativedetermination of pH, Partial Pressure ofCarbon Dioxide (pCO2), Partial Pressure ofOxygen (pO2), Hematocrit, Sodium,Potassium, Chloride, Ionized Calcium,Ionized Magnesium, Glucose, and Lactate inheparinized capillary whole blood.
blood, serum, or plasma. Total Bilirubin (tBil)
was not evaluated on neonatal samples.
Acceptable Samples
Sample TypesSodium or lithium heparinized whole blood,serum, or plasma samples from syringes,open tubes, small cups, and capillary tubes.Lithium heparin whole blood from syringes,open tubes, and capillary tubes.
Sample Volumes60-200µL (dependent on panel selected)135µL (Syringes and open tubes)90µL (Capillary tubes)
Measurement Range
pH6.500 - 8.000Same
PCO23.0 – 200 mmHgSame
PO20 - 800 mmHg5.0 - 765.0 mmHg
Hct12% - 70%Same
Glu15 - 500 mg/dLSame
Lac0.3 - 20.0 mmol/LSame
Sodium80 - 200 mmol/LSame
Potassium1.0 - 20.0 mmol/LSame
Chloride50 - 200 mmol/LSame
lonized Calcium0.1 - 2.7 mmol/LSame
lonized Magnesium0.1 - 1.5 mmol/LSame
Principles ofMeasurement
pHHydrogen ion-selective glass sensorSame
PCO2Severinghaus-type sensorSame
PO2Polarographic Clark-type sensorSame
HctImpedance sensorSame
GluEnzymatic sensorSame
LacEnzymatic sensorSame
SodiumIon-Selective ElectrodeSame
PotassiumIon-Selective ElectrodeSame
ChlorideIon-Selective ElectrodeSame
Ionized CalciumIon-Selective ElectrodeSame
Ionized MagnesiumIon-Selective ElectrodeSame
Touch Screen12.1" LCD, 1024x768 pixel, Resistive Touch10.1" WXGA 1280 x 800 color touch screen
MenuFully configurable test menu based onavailable sensorsSame
Bar Code ScannerInternal Integrated 1D/2DSame
Printer2" Roll, Thermal TransferSame
PumpPeristaltic Pump w/ Pressure Plate, TPETubing (Pharmed BPT)Same
Analog BoardPrecision low level analog front end w/amperometric and potentiometric amplifiers,air detector circuitry and temperature controlcircuitrySame

{18}------------------------------------------------

Conclusion

The performance of pH, Partial Pressure of Carbon Dioxide (pCO2), Partial Pressure of Oxygen (pO2), Hematocrit, Sodium, Potassium, Chloride, lonized Calcium, lonized Magnesium, Glucose, and Lactate in heparinized capillary whole blood on the Stat Profile Prime Plus Analyzer System is substantially equivalent to the Stat Profile pHOx Ultra Analyzer.

§ 862.1120 Blood gases (P

CO2 , PO2 ) and blood pH test system.(a)
Identification. A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.(b)
Classification. Class II.