K Number
K221668
Date Cleared
2022-08-08

(61 days)

Product Code
Regulation Number
872.3660
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dental Impression Material is intended for use with all crowns, bridges, and orthodontic impression techniques to reproduce the structure of a patient's teeth and gums.

Device Description

The Dental Impression Material is a kind of addition-cure rubber impression material composed of vinyl polysiloxane and various fillers, with neutral smell and applicable to impression in dentistry.

The Dental Impression Material is very easy to mix and has good dimensional stability which helps to make precise impression taking.

AI/ML Overview

The information provided describes the acceptance criteria and performance of the Dental Impression Material by Chemoment Materials Co.,ltd, as detailed in the 510(k) summary K221668.

Here's a breakdown of the requested information:

1. Table of acceptance criteria and the reported device performance

ItemProposed deviceAcceptance criteria (Type 0)Result
Consistency31.8mm≤35mmPass
Compatibility with gypsumComplied75μmPass
Detail reproductionComplied75μmPass
Linear dimensional change0.798%≤1.5%Pass
Elastic recovery97.718%≥96.5%Pass
Strain in compression5.44%0.8~20%Pass

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample size for each performance test. It states that "Non-clinical tests were conducted to verify that the proposed device met all design specifications as same or similar to the predicate device." These tests were conducted according to ISO standards (ISO 10993-05, ISO 10993-10, and ISO 4823:2021). The data provenance is not explicitly mentioned but is implied to be from laboratory testing performed by or for Chemoment Materials Co.,ltd as part of their regulatory submission. The tests are non-clinical, meaning they don't involve human subjects, and are therefore not classified as retrospective or prospective human data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable. The device's performance was evaluated against established engineering and material science standards (ISO standards), not against a ground truth established by medical experts for diagnostic purposes.

4. Adjudication method for the test set

This information is not applicable as the tests are objective measurements against established standards, not subjective assessments requiring expert adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No such study was conducted. This device is a material (dental impression material), not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a material, not an algorithm.

7. The type of ground truth used

The "ground truth" for the device's performance is defined by the objective pass/fail criteria set forth in the international standards ISO 4823:2021, and the biocompatibility standards ISO 10993-05:2009 and ISO 10993-10:2010. These standards provide specific measurable parameters for acceptable performance for dental impression materials.

8. The sample size for the training set

This is not applicable. The product is a material, not a machine learning algorithm that requires a training set.

9. How the ground truth for the training set was established

This is not applicable as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

August 8, 2022

Chemoment Materials Co.,ltd % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Century Avenue Shanghai. 200122 China

Re: K221668

Trade/Device Name: Dental Impression Material Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: Class II Product Code: ELW Dated: June 8, 2022 Received: June 8, 2022

Dear Ivy Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K221668

Device Name

Dental Impression Material (Type 0)

Indications for Use (Describe)

The Dental Impression Material is intended for use with all crowns, bridges, and orthodontic impression techniques to reproduce the structure of a patient's teeth and qums.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K221668

510(K) Summary

(As requirement by 21 CFR 807.92)

Date prepared: 26th, May, 2022

A. Applicant:

Name: Chemoment Materials Co.,ltd Address: No. 546-549, Renhe Industry town, Shuangfeng Economic Development Zone, Hefei , Anhui. 230081 Contact: Mabel. Lu Title: Sales Tel: +86-551-62329017 Email: mabel.lu(@foxmail.com

Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@@gmail.com

B. Device:

Trade Name: Dental Impression Material Common Name: Dental Impression Material

Regulatory Information Classification Name: Material, Impression Classification: Class II Product code: ELW Regulation Number: 21 CFR 872.3660 Review Panel: Dental

C. Predicate device: K211489 Nanchang Dental Bright Technology Co., Ltd. Trade Name: Dental Impression Material Common Name: Dental Impression Material

{4}------------------------------------------------

Regulatory Information Classification Name: Material, Impression Classification: Class II Product code: ELW Regulation Number: 21 CFR 872.3660 Review Panel: Dental

D. Indications for use of the device:

The Dental Impression Material is intended for use with all crowns, bridges, and orthodontic impression techniques to reproduce the structure of a patient's teeth and gums.

E. Device Description:

The Dental Impression Material is a kind of addition-cure rubber impression material composed of vinyl polysiloxane and various fillers, with neutral smell and applicable to impression in dentistry.

The Dental Impression Material is very easy to mix and has good dimensional stability which helps to make precise impression taking.

F. Non-clinical Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications as same or similar to the predicate device. The test results demonstrated that the proposed device complied with the following standards:

  • ISO 10993-05:2009 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity

  • A ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation And Skin Sensitization
  • ISO 4823:2021 Dentistry Elastomeric impression and bite registration materials A

Performance parameters are summarized in below Table 1.

ItemProposed deviceAcceptance criteria (Type 0)Result
Consistency31.8mm≤35mmPass
Compatibility with gypsumComplied75μmPass
Detail reproductionComplied75μmPass
Linear dimensional change0.798%≤1.5%Pass
Elastic recovery97.718%≥96.5%Pass
Strain in compression5.44%0.8~20%Pass

Table 1 Performance Parameter

G. Clinical Test Conclusion

No clinical study is included in this submission.

H. Comparison with predicate device

Table 2 Comparison of Proposed and Predicate Devices

DeviceProposed DevicePredicate DeviceResult
---------------------------------------------------

{5}------------------------------------------------

510K #K221668K211489-
ManufacturerChemoment Materials Co.,ltdNanchang Dental Bright Technology Co., Ltd.-
Product NameDental Impression MaterialDental Impression MaterialSame
ClassificationClass II Device, ELW (21 CFR 872.3660)Class II Device, ELW (21 CFR 872.3660)Same
Indications for useThe Dental Impression Material is intended for use with all crowns, bridges, and orthodontic impression techniques to reproduce the structure of a patient's teeth and gums.The Dental Impression Material is intended for use with all crowns, bridges, and orthodontic impression techniques to reproduce the structure of a patient's teeth and gums.Same
Standard conformedISO 4823ISO 4823Same
Physical properties-Classification according to ISO 4823:Type 0-Dimensional accuracy: Max. 1.5%-Consistency: Max.35mm-Compatibility with the die and cast materials:75µm reproduction-Curve of the shrinkage (strain in compression): Min.0.8-Max.20%-Classification according to ISO 4823:Type 0-Dimensional accuracy: Max. 1.5%-Consistency: Max.35mm-Compatibility with the die and cast materials:75µm reproduction-Curve of the shrinkage (strain in compression): Min.0.8-Max.20%Same
Raw materialVinyl polysiloxaneVinyl polysiloxaneSame
Mixing ratio1:11:1Same
SterilityNon-sterileNon-sterileSame
Method of manipulationHand-kneaded mixesHand-kneaded mixesSame
BiocompatibilityISO 10993ISO 10993Same

I. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K211489.

§ 872.3660 Impression material.

(a)
Identification. Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.(b)
Classification. Class II (Special Controls).