K Number
K221616
Date Cleared
2022-08-01

(59 days)

Regulation Number
870.1025
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The monitors are intended to be used for monitoring, storing, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments.

The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), cardiac output (C.O.), anesthetic gas (AG), bispectral index (BIS), respiration mechanice cardiography (ICG).

BIS is intended for use on adult and pediatric patients.

ICG monitoring is intended for use on adults only.

The arrhythmia detection and ST Segment analysis are intended for adult patients.

The monitors are additionally intended for use during patient transport inside hospitals.

The monitors are not intended for MRI environments.

Device Description

LMPLUS series Patient Monitor including LMPLUS-12, LMPLUS-15 and LMPLUS-17 which can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormity so that doctors and nurses can deal with them in time.

The LMPLUS series Patient Monitor realize the monitoring of physiological parameters by configuration with different parameter modules which include SpO2 (pulse oxygen saturation, pulse rate and SpO2 plethysmogram) with EDAN SpO2 module or Nellcor SPO2 module, NIBP (systolic pressure, diastolic pressure and pulse rate), TEMP, ECG, RESP (respiration), CO2, IBP, C.O. and AG (anesthetic gas), RM (respiratory mechanics), BIS (bispectral index) and ICG (impedance cardiography).

The above is the maximum configuration for LMPLUS series Patient Monitor, the user may select different monitoring parameters in according with their requirements.

LMPLUS-12 configures with 12.1-inch color TFT touch screen, LMPLUS-15 and LMPLUS-17 with same screen except different sizes 15-inch and 17-inch separately. Three models are all build-in Lithium-ion battery, support software upgrade online and networking.

AI/ML Overview

The provided document focuses on the 510(k) summary for the CAF Medical Solutions Inc. Patient Monitor (models LMPLUS-12, LMPLUS-15, and LMPLUS-17), demonstrating its substantial equivalence to a predicate device (Edan Instruments, Inc. Patient Monitor, models elite V5, elite V6, and elite V8). The document primarily presents a feature-by-feature comparison and non-clinical performance data, with a brief mention of clinical tests.

Therefore, the information regarding acceptance criteria and the study proving the device meets them will be limited to what is explicitly stated in the document or can be inferred from the provided test types and standards. A full, detailed study proving acceptance criteria for specific performance metrics (like sensitivity, specificity, or inter-reader variability for an AI model) is not present in this type of regulatory submission document, which focuses on substantial equivalence to a predicate rather than a novel AI algorithm.

Based on the provided document, here's what can be extracted and inferred regarding performance and validation:

The document indicates that the device's performance was evaluated against various recognized standards for patient monitors, which inherently define acceptable performance ranges for each physiological parameter. The study primarily aims to show that the new device meets these established standards and performs comparably to the predicate device.


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a "table of acceptance criteria" in the format of a pre-defined threshold that the device must meet for a specific study's outcome (e.g., "sensitivity > X%"). Instead, it shows a feature-by-feature comparison to a predicate device, including various performance specifications (e.g., accuracy, measurement range) that are in line with industry standards for patient monitors. The "Acceptance Criteria" are implicitly defined by the parameters and accuracy/range specifications of the predicate device and the relevant IEC/ISO standards the device claims compliance with. The "Reported Device Performance" for the subject device (LMPLUS models) is stated to be "Same" as the predicate device across all listed specifications.

Here's an illustrative table based on the provided comparison, highlighting key physiological parameters:

Feature/ParameterAcceptance Criteria (Implied by Predicate/Standards)Reported Device Performance (LMPLUS Models)Comparison to Acceptance
ECG Monitor
Measurement Range (Adult)15 to 300 bpm15 to 300 bpmMeets
Accuracy±1 bpm or ±1%, whichever is greater±1 bpm or ±1%, whichever is greaterMeets
ST Value Accuracy-0.8 to +0.8 mV: ±0.02 mV or 10%-0.8 to +0.8 mV: ±0.02 mV or 10%Meets
RESP Monitor
Measurement Range (Adult)0 to 120 rpm0 to 120 rpmMeets
Accuracy (Adult)6 to 120 rpm: ±2 rpm6 to 120 rpm: ±2 rpmMeets
Temperature Monitor
Measurement Range0 to 50°C0 to 50°CMeets
Accuracy±0.1°C (±0.2°F)±0.1°C (±0.2°F)Meets
SpO2 Monitor
Measurement Range0-100%0-100%Meets
Accuracy (Adult/Pediatric, no motion)70 to 100%: ±2%70 to 100%: ±2%Meets
NIBP Monitor
Max Mean Error±5 mmHg±5 mmHgMeets
Max Standard Deviation8 mmHg8 mmHgMeets
CO2 Monitor (EDAN)
Accuracy (≤60rpm)±2mmHg, 0-40mmHg; ±5%, 41-70mmHg; etc.±2mmHg, 0-40mmHg; ±5%, 41-70mmHg; etc.Meets
OtherCompliance with specific IEC/ISO standardsCompliance with specific IEC/ISO standardsMeets

Note: The table above is a summary of just a few representative parameters from the much larger comparison table (Table 1) in the document. The general "Acceptance Criteria" for all listed parameters are the identical specifications of the predicate device.


2. Sample Size Used for the Test Set and Data Provenance

The document states: "Clinical tests were performed on the LMPLUS 12, LMPLUS 15 and LMPLUS 17 monitors to validate their performance in terms of noninvasive blood pressure (NIBP) and SpO2 accuracy."

However, the specific sample sizes for these clinical tests (number of patients, number of measurements) and the data provenance (e.g., country of origin, retrospective or prospective nature) are not detailed in this 510(k) summary. This level of detail would typically be found in the full test report, which is referenced but not included.


3. Number of Experts Used to Establish Ground Truth and Qualifications

The document mentions "clinical tests" for NIBP and SpO2 accuracy. For these types of physiological measurements, the ground truth is typically established by:

  • Reference Devices: Using highly accurate, calibrated reference measurement devices.
  • Clinical Protocols: Adhering to established clinical protocols for data collection (e.g., for NIBP, a protocol like ISO 81060-2 which often involves comparisons to invasive arterial measurements or calibrated sphygmomanometers by trained healthcare professionals).

There is no mention of human experts being used to establish "ground truth" in the context of interpretation (e.g., radiologists for imaging, unlike an AI algorithm for image analysis). The device measures physiological parameters, and accuracy is validated against established, objective measurement techniques, not expert consensus on qualitative data. Therefore, the concept of "experts establishing ground truth" as it applies to subjective judgments or interpretations (which is common for AI/ML in imaging) is not directly applicable here.


4. Adjudication Method for the Test Set

Given that the clinical tests mentioned are for quantitative physiological parameter accuracy (NIBP and SpO2), adjudication methods like 2+1 or 3+1 (common in studies involving multiple readers for subjective assessments) are not applicable. Accuracy is determined by comparing device readings to a reference standard, not by expert consensus on interpretations.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC study was mentioned or implied.
This device is a patient monitor, not an AI-assisted diagnostic tool that would involve human readers interpreting cases. Therefore, a study to measure how much human readers improve with AI assistance is not relevant to this type of device and was not performed.


6. Standalone Performance (Algorithm Only without Human-in-the-Loop)

The document does not describe the device as having a distinct "algorithm" component for analysis that would be evaluated in isolation. It's a physiological monitoring device. Its accuracy in measuring parameters like NIBP and SpO2 is its "standalone performance." The clinical tests mentioned (for NIBP and SpO2 accuracy) would indeed be an assessment of the device's ability to accurately measure these parameters independently, which is effectively its standalone performance. The results are implied by the statement "the subject devices perform comparably to the predicate device."


7. Type of Ground Truth Used

For the clinical tests (NIBP and SpO2 accuracy), the ground truth would be established through:

  • Reference Standard Measurements: Using a highly accurate and validated reference device (e.g., an invasive arterial line for NIBP, or a co-oximeter for SpO2) or an established standardized method as per relevant ISO standards (e.g., ISO 81060-2 for NIBP).
  • Physiological Data: Direct physiological measurements, not pathology, outcomes data, or expert consensus on subjective interpretations.

8. Sample Size for the Training Set

This document describes a conventional patient monitor, not a medical device that utilizes AI/ML requiring a distinct "training set" of data to learn from. Therefore, there is no mention of a training set or its sample size. The device's algorithms for processing physiological signals are based on established engineering principles and signal processing, not machine learning from a large training dataset.


9. How the Ground Truth for the Training Set Was Established

As there is no training set for this type of device, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

August 1, 2022

CAF Medical Solutions Inc. % Juan Tezak Consultant Compliance 4 Devices 118 W Prive Cr. Delray Beach, Florida 33445

Re: K221616

Trade/Device Name: Patient Monitor, models LMPLUS-12, LMPLUS-15 and LMPLUS-17 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, DST, MLD, DRT, DXN, DSK, FLL, DQA, CCK, CBQ, NHQ, NHP, CBS, CBR, CCL, DRG, DPS Dated: June 3, 2022 Received: June 3, 2022

Dear Juan Tezak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221616

Device Name

Patient Monitor, models LMPLUS 12, LMPLUS 15 and LMPLUS 17.

Indications for Use (Describe)

The monitors are intended to be used for monitoring, storing, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments.

The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), cardiac output (C.O.), anesthetic gas (AG), bispectral index (BIS), respiration mechanice cardiography (ICG).

BIS is intended for use on adult and pediatric patients.

ICG monitoring is intended for use on adults only.

The arrhythmia detection and ST Segment analysis are intended for adult patients.

The monitors are additionally intended for use during patient transport inside hospitals.

The monitors are not intended for MRI environments.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Lifee Medical Devices. The logo consists of a red heart rate line on the left, followed by the word "lifee" in purple. Below the word "lifee" is the phrase "medical devices" in a smaller, red font.

510(k) Summary

Patient Monitor, models LMPLUS-12, LMPLUS-15 and LMPLUS-17 June, 2022

ADMINISTRATIVE INFORMATION

Applicant

CAF Medical Solutions Inc. 17539 Roberts Road, Hockley TX 77477 USA E-mail: c.ortiz@cafmedical.com

Establishment Registration Number

Official ContactOscar Arriaga Yamin
Legal Representative
17539 Roberts Road, Hockley
TX 77477 USA
Phone: +1 713 614 7049
E-mail: c.ortiz@cafmedical.com
Representative/ConsultantJuan Tozak
---------------------------------------
Representative/ConsultantJuan Tezak
Carlos Marín
Compliance4Devices
118 W Prive Cr. Delray Beach Fl, 33445
Phone: +1 561-789-2411
E-mail: compliance4devices@gmail.com

DEVICE AND CLASSIFICATION NAME

Device Trade Name:Patient Monitor, models LMPLUS-12, LMPLUS-15 and LMPLUS-17
Regulatory Class:Class II
Prior Submission:K160981
Device name, classification andproduct code:· 870.1025 monitor, physiological, patient (witharrhythmia detection or alarms)/ MHX· 870.1025 Detector and Alarm, Arrhythmia/ DST· 870.1025 Monitor, ST Segment with Alarm/ MLD· 870.2300 Cardiac monitor (includingcardiotachometer and rate alarm)/ DRT· 870.1130 Non-Invasive blood pressure/ DXN· 870.1110 Blood pressure computer/ DSK

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Lifee Medical Devices. The logo consists of a red heart rate line to the left of the word "lifee" in purple. Below the word "lifee" is the phrase "medical devices" in red.

  • · 880.2910 Clinical Electronic Thermometers-Temperature Monitor with Probe/ FLL
  • · 870.2700 Oximeter, Pulse/ DQA
  • · 868.1400 Carbon Dioxide Gas Analyzer/ CCK
  • · 868.1500 Analyzer, Gas, Enflurane, Gaseous-Phase (Anesthetic Concentration)/ CBQ
  • · 868.1500 Analyzer, Gas, Desflurane, Gaseous-Phase (Anesthetic Concentration)/NHO
  • · 868.1500 Analyzer, Gas, Isoflurane, Gaseous-Phase (Anesthetic Concentration)/NHQ
  • · 868.1500 Analyzer, Gas, Sevoflurane, Gaseous-Phase (Anesthetic Concentration)/NHP
  • · 868.1620 Halothane gas analyzer/ CBS
  • · 868.1700 Nitrous Oxide gas analyzer/ CBR
  • · 868.1720 Oxygen gas analyzer/ CCL
  • · 870.2770 Impedance plethysmograph/ DSB
  • · 868.1850 Monitoring spirometer/ BZK
  • · 868.2375 Monitor, Breathing Frequency/BZQ
  • · 870.2340 Electrocardiograph/DPS
  • · 870.2910 Radiofrequency physiological signal transmitter and receiver/ DRG

Predicate Device Information

Predicate Device:

K160981. Patient Monitor, models elite V5, elite V6 and elite V8. Edan Instruments, Inc.

Intended Use

The monitors are intended to be used for monitoring, storing, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments.

The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), cardiac output (C.O.), anesthetic gas (AG), bispectral index (BIS), respiration mechanics (RM) and impedance cardiography (ICG).

BIS is intended for use on adult and pediatric patients.

ICG monitoring is intended for use on adults only.

The arrhythmia detection and ST Segment analysis are intended for adult patients.

The monitors are additionally intended for use during patient transport inside hospitals.

The monitors are not intended for MRI environments.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Lifee Medical Devices. The logo consists of a red heart-shaped graphic with an EKG waveform on the left, followed by the word "lifee" in a dark purple, rounded sans-serif font. Below "lifee" are the words "medical devices" in a smaller, light red sans-serif font.

Device Description

LMPLUS series Patient Monitor including LMPLUS-12, LMPLUS-15 and LMPLUS-17 which can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormity so that doctors and nurses can deal with them in time.

The LMPLUS series Patient Monitor realize the monitoring of physiological parameters by configuration with different parameter modules which include SpO2 (pulse oxygen saturation, pulse rate and SpO2 plethysmogram) with EDAN SpO2 module or Nellcor SPO2 module, NIBP (systolic pressure, diastolic pressure and pulse rate), TEMP, ECG, RESP (respiration), CO2, IBP, C.O. and AG (anesthetic gas), RM (respiratory mechanics), BIS (bispectral index) and ICG (impedance cardiography).

The above is the maximum configuration for LMPLUS series Patient Monitor, the user may select different monitoring parameters in according with their requirements.

LMPLUS-12 configures with 12.1-inch color TFT touch screen, LMPLUS-15 and LMPLUS-17 with same screen except different sizes 15-inch and 17-inch separately. Three models are all build-in Lithium-ion battery, support software upgrade online and networking.

Contraindications:

There are no known contraindications for use.

Equivalence to Marketed Device

The LMPLUS series Patient Monitor is substantially equivalent to the predicate. In further support of a substantial equivalence determination, here-under is a comparison chart with the submitted device and predicate device.

FeatureSubject DevicePredicate DeviceComparison
K #Current submissionK160981
Intended useThe monitors are intended to be used for monitoring, storing, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments. The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial bloodThe monitors are intended to be used for monitoring, storing, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments. The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterialSame
FeatureSubject DevicePredicate DeviceComparison
(SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), cardiac output (C.O.), anesthetic gas (AG), bispectral index (BIS), respiration mechanics (RM) and impedance cardiography (ICG).BIS is intended for use on adult and pediatric patients.ICG monitoring is intended for use on adults only.The arrhythmia detection and ST Segment analysis are intended for adult patients.The monitors are additionally intended for use during patient transport inside hospitals.The monitors are not intended for MRI environments.blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), cardiac output (C.O.), anesthetic gas (AG), bispectral index (BIS), respiration mechanics (RM) and impedance cardiography (ICG).BIS is intended for use on adult and pediatric patients.ICG monitoring is intended for use on adults only.The arrhythmia detection and ST Segment analysis are intended for adult patients.The monitors are additionally intended for use during patient transport inside hospitals.The monitors are not intended for MRI environments.
IntendedpatientpopulationAdult, pediatric and neonatal patientsAdult, pediatric and neonatal patientsSame
IntendedapplicationenvironmentHospital environment.Hospital environment.Same
1) ECG monitor
Lead Mode3-Lead: I, II, III5-Leads: I, II, III, aVR, aVL, aVF, V12-leads: I, II, III, aVR, aVL, aVF, V1 to V63-Lead: I, II, III5-Leads: I, II, III, aVR, aVL, aVF, V12-leads: I, II, III, aVR, aVL, aVF, V1 to V6Same
Lead NamingStyleAHA, IECAHA, IECSame
DisplaySensitivity1.25mm/mV (x0.125),2.5mm/mV (x0.25), 5mm/mV (x0.5), 10mm/mV (x1),20mm/mV (x2), 40mm/mV (x4),AUTO gain1.25mm/mV (x0.125),2.5mm/mV (x0.25),5mm/mV (x0.5), 10mm/mV (x1), 20mm/mV (x2),40mm/mV (x4),AUTO gainSame
Sweep6.25mm/s, 12.5mm/s,25mm/s, 50mm/s6.25mm/s, 12.5mm/s,25mm/s, 50mm/sSame
MeasurementRangeNeonate: 15 to 350 bpmPediatric: 15 to 350 bpmNeonate: 15 to 350 bpmPediatric: 15 to 350 bpmSame
FeatureSubject DevicePredicate DeviceComparison
CMRR(Common ModeRejection Ratio)Diagnosis: >95dBMonitor: >105dBSurgery: >105dBDiagnosis: >95dBMonitor: >105dBSurgery: >105dBSame
Accuracy±1 bpm or ±1%, whichever isgreater±1 bpm or ±1%, whichever isgreaterSame
Resolution1 bpm1 bpmSame
Sensibility200μV200μVSame
DifferentialInputImpedance>5ΜΩ>5ΜΩSame
Leakage Currentof Patient<10μΑ<10μΑSame
ST valueMeasurementRange-2.0 mV ~ +2.0 mV-2.0 mV ~ +2.0 mVSame
Accuracy-0.8 mV to +0.8 mV: ±0.02 mVor 10%,whichever is greater.Beyond this range: notspecified.-0.8 mV to +0.8 mV: ±0.02mV or 10%,whichever is greater.Beyond this range: notspecified.Same
Pace
Pulse IndicatorAmplitude: ±2 mV to ±700 mVWidth: 0.1 ms to 2.0 msAscending time: 10 µs to 100µsAmplitude: ±2 mV to ±700mVWidth: 0.1 ms to 2.0 msAscending time: 10 µs to100 µsSame
Pulse RejectionAmplitude: ±2 mV to ±700 mVWidth: 0.1 ms to 2.0 msAscending time: 10 µs to 100µsAmplitude: ±2 mV to +700mVWidth: 0.1 ms to 2.0 msAscending time: 10 µs to100 µsSame
HR
MeasurementrangeNeonate: 15 to 350 bpmPediatric: 15 to 350 bpmAdult: 15 to 300 bpmNeonate: 15 to 350 bpmPediatric: 15 to 350 bpmAdult: 15 to 300 bpmSame
Accuracy±1% or 1 bpm, whichever isgreater±1% or 1 bpm, whichever isgreaterSame
Resolution1 bpm1 bpmSame
2) RESP monitor
Principle ofOperationThoracic impedanceThoracic impedanceSame
MethodImpedance between RA-LL,RA-LAImpedance between RA-LL,RA-LASame
MeasurementleadOptions are lead I and II.The default is lead II.Options are lead I and II.The default is lead II.Same
MeasurementRangeAdult:0 to 120 rpmAdult:0 to 120 rpmSame
FeatureSubject DevicePredicate DeviceComparison
AccuracyAdult:6 to 120 rpm: ±2 rpm,0 to 5 rpm: not specifiedPediatric/neonate:6 to 150 rpm: ±2 rpm,0 to 5 rpm: not specifiedAdult:6 to 120 rpm: ±2 rpm,0 to 5 rpm: not specifiedPediatric/neonate:6 to 150 rpm: ±2 rpm,0 to 5 rpm: not specifiedSame
CalculationTypeManual, AutomaticManual, AutomaticSame
BaselineImpedanceRange2000Ω to 2500Ω (with ECGcables of 1 kΩ resistance)2000Ω to 2500Ω (with ECGcables of 1 kΩ resistance)Same
MeasuringSensitivityWithin the baselineimpedance range: 0.3ΩWithin the baselineimpedance range: 0.3ΩSame
WaveformBandwidth0.2 Hz to 2.5 Hz (-3 dB)0.2 Hz to 2.5 Hz (-3 dB)Same
RespirationExcitationWaveformSinusoid, 62.8 kHz (±10%),<500 μΑSinusoid, 62.8 kHz (±10%),<500 μΑSame
Resolution1 rpm1 rpmSame
Apnea Alarm10s, 15s, 20s, 25s, 30s, 35s,40s10s, 15s, 20s, 25s, 30s, 35s,40sSame
3) Temperature monitor
TechniqueThermal resistanceThermal resistanceSame
Number ofchannels22Same
MeasurementRange0 to 50°C0 to 50°CSame
Accuracy±0.1°C (±0.2°F)±0.1°C (±0.2°F)Same
Resolution0.1°C0.1°CSame
4) SpO2 monitor
MeasurementRange0-100%0-100%Same
Accuracy70 to 100%: ±2% (measuredwithout motion inadult/pediatric mode).70 to 100%: ±3% (measuredwithout motion in neonatemode).0% to 69%: Not specified.70 to 100%: ±2% (measuredwithout motion inadult/pediatric mode).70 to 100%: ±3% (measuredwithout motion in neonatemode).0% to 69%: Not specified.Same
Resolution1 %1 %Same
5) PR
MeasurementRange25 bpm to 300 bpm25 bpm to 300 bpmSame
Accuracy±2bpm±2bpmSame
Resolution1 bpm1 bpmSame
6) NIBP monitor
Principle ofOperationOscillationOscillationSame
Adult:Adult:Same
FeatureSubject DevicePredicate DeviceComparison
MeasurementRange (mmHg)Systolic 40 to 270 mmHgSystolic 40 to 270 mmHgSame
Diastolic 10 to 215 mmHgDiastolic 10 to 215 mmHgSame
Mean 20 to 235 mmHgMean 20 to 235 mmHgSame
Pediatric:Pediatric:Same
Systolic 40 to 230 mmHgSystolic 40 to 230 mmHgSame
Diastolic 10 to 180 mmHgDiastolic 10 to 180 mmHgSame
Mean 20 to 195 mmHgMean 20 to 195 mmHgSame
Neonate:Neonate:Same
Systolic 40 to 135 mmHgSystolic 40 to 135 mmHgSame
Diastolic 10 to 100 mmHgDiastolic 10 to 100 mmHgSame
Mean 20 to 110 mmHgMean 20 to 110 mmHgSame
AccuracyMax mean error: ±5 mmHg,Max standard deviation: 8mmHgMax mean error: ±5 mmHg,Max standard deviation: 8mmHgSame
Resolution1mmHg1mmHgSame
OverpressureprotectionAdult:297±3mmHgAdult:297±3mmHgSame
Pediatric:245±3mmHgPediatric:245±3mmHg
Pediatric:147±3mmHgPediatric:147±3mmHg
ModeManual, Auto, ContinuousManual, Auto, ContinuousSame
Measuringinterval inAUTO Mode1/2/3/4/5/10/15/30/60/90/120/240/ 480min1/2/3/4/5/10/15/30/60/90/120/240/ 480minSame
PR from NIBP
MeasurementRange40 bpm to 240bpm40 bpm to 240bpmSame
Accuracy±3bpm or 3.5%, whichever isgreater±3bpm or 3.5%, whicheveris greaterSame
7) IBP monitor
TechniqueDirect invasive measurementDirect invasivemeasurementSame
MeasurementRange-50 ~ 300mmHg-50 ~ 300mmHgSame
Accuracy±2% or ±1 mmHg, whichever isgreater (without sensor)±2% or ±1 mmHg, whicheveris greater (without sensor)Same
Resolution1mmHg1mmHgSame
Transducer5 (μV/V/mmHg)200 to 3000 Ω5 (μV/V/mmHg)200 to 3000 ΩSame
8) CO2 Monitor
CO2 Module
Respironics CAPNOSTAT 5 LoFlo CO2 (Sidestream)
TechniqueInfra-red AbsorptionTechniqueInfra-red AbsorptionTechniqueSame
MeasureParametersEtCO2, FiCO2, AwRREtCO2, FiCO2, AwRRSame
FeatureSubject DevicePredicate DeviceComparison
CO2Measurementrange0 mmHg to 150 mmHg0 mmHg to 150 mmHgSame
AwRRMeasurementrange2 rpm to 150 rpm2 rpm to 150 rpmSame
AccuracyEDAN:Respiratory rate ≤60rpm:±2mmHg, 0mmHg to40mmHg, ±5% of reading,41mmHg to 70mmHg, ±8% ofreading, 71mmHg to100mmHg, ±10% of reading,101mmHg to 150mmHg,Respiratory rate >60rpm:±12% or ±4mmHg of reading,whichever is greater.Respironics:±2 mmHg, 0 to 40 mmHg, ±5%ofreading, 41 to 70 mmHg, ±8%of reading, 71 to 100 mmHg,±10% of reading, 101 to 150mmHg, ±12% of reading, RR isover 80 rpm.EDAN:Respiratory rate ≤60rpm:±2mmHg, 0mmHg to40mmHg, ±5% of reading,41mmHg to 70mmHg, ±8%of reading, 71mmHg to100mmHg, ±10% of reading,101mmHg to 150mmHg,Respiratory rate >60rpm:±12% or ±4mmHg ofreading, whichever isgreater.Respironics:±2 mmHg, 0 to 40 mmHg,±5% ofreading, 41 to 70 mmHg,±8% of reading, 71 to 100mmHg, ±10% of reading,101 to 150 mmHg, ±12% ofreading, RR is over 80 rpm.Same
Resolution1 mmHg1 mmHgSame
Sample flowrateEDAN: 70ml/min or100ml/min, optional(±15ml/min)Respironics:50 ±10 ml/minEDAN: 70ml/min or100ml/min, optional(±15ml/min)Respironics:50 ±10 ml/minSame
Apnea AlarmDelay10s, 15s, 20s, 25s, 30s, 35s,40s.10s, 15s, 20s, 25s, 30s, 35s,40s.Same
Respironics CAPNOSTAT 5 CO2 (Mainstream)
MeasureParametersEtCO2, FiCO2, AwRREtCO2, FiCO2, AwRRSame
CO2Measurementrange0 mmHg to 150 mmHg0 mmHg to 150 mmHgSame
AwRRmeasurementrange0 rpm to 150 rpm0 rpm to 150 rpmSame
Accuracy±2 mmHg, 0 to 40 mmHg, ±5%of reading, 41 to 70 mmHg,±8% of reading, 71 to 100mmHg, ±10% of reading, 101to 150 mmHg±2 mmHg, 0 to 40 mmHg,±5% of reading, 41 to 70mmHg, ±8% of reading, 71to 100 mmHg, ±10% ofreading, 101 to 150 mmHgSame
FeatureSubject DevicePredicate DeviceComparison
AwRRmeasurementaccuracy$\pm$ 1 rpm$\pm$ 1 rpmSame
9) C.O. Temperature
MeasurementmethodThermodilution methodThermodilution methodSame
MeasurementrangeC.O.: 0.1 to 20L/minTB: 23 to 43°CTI: -1 to 27°CC.O.: 0.1 to 20L/minTB: 23 to 43°CTI: -1 to 27°CSame
AccuracyC.O.: $\pm$ 5% or $\pm$ 0.2L/min, whichis greaterTB, TI: $\pm$ 0.1°C (without sensor)C.O.: $\pm$ 5% or $\pm$ 0.2L/min,which is greaterTB, TI: $\pm$ 0.1°C (withoutsensor)Same
ResolutionC.O.: 0.1L/minTB, TI:0.1°CC.O.: 0.1L/minTB, TI:0.1°CSame
10) AG Monitor (Phasein and Drager)
TechniqueInfrared absorptionInfrared absorptionSame
Warm-uptimeFull accuracy mode:<20s(Phasein)Full accuracy mode:<450s(Drager)Full accuracy mode:<20s(Phasein)Full accuracy mode:<450s(Drager)Same
Sample flowrate50 $\pm$ 10ml/min(Phasein)200 mL/min $\pm$ 20mL/min(Drager)50 $\pm$ 10ml/min(Phasein)200 mL/min $\pm$ 20mL/min(Drager)Same
MeasurementRange andAccuracyRange (%REL)Accuracy(%ABS)Same
CO2
Phasein
Sidestream:0 to 15 vol%15 to 25vol%Sidestream:$\pm$ (0.2 vol% +2% ofreading)UnspecifiedSame
Mainstream:0 to 10 vol%10 to 15vol%15 to 25vol%Mainstream:$\pm$ (0.2 vol% +2% ofreading)$\pm$ (0.3 vol% +2% ofreading)Unspecified

Table 1. Comparison with predicate device for Summary

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Lifee Medical Devices. The logo consists of a red heart-shaped electrocardiogram symbol on the left, followed by the word "lifee." in purple. Below the word "lifee." are the words "medical devices" in red.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Lifee Medical Devices. On the left is a red heart rate monitor symbol. To the right of the symbol is the word "lifee" in purple, and below that is the phrase "medical devices" in red.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Lifee Medical Devices. To the left of the logo is a red heart rate graphic. To the right of the graphic is the word "lifee" in purple, with the words "medical devices" in red underneath.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Lifee Medical Devices. The logo consists of a red heart-shaped line drawing on the left, followed by the word "lifee." in purple. Below the word "lifee." are the words "medical devices" in a smaller font, also in purple.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Lifee Medical Devices. The logo consists of a red heart rate waveform on the left, followed by the word "lifee" in purple. Below the word "lifee" are the words "medical devices" in a smaller, red font.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Lifee Medical Devices. To the left of the logo is a red heart rate symbol. To the right of the symbol is the word "lifee" in purple, with the words "medical devices" in red underneath.

{12}------------------------------------------------

Image /page/12/Picture/1 description: The image shows the logo for Lifee Medical Devices. On the left is a red heart rate monitor symbol. To the right of the heart rate monitor is the word "lifee" in a dark purple color. Underneath the word "lifee" are the words "medical devices" in a light red color.

FeatureSubject DevicePredicate DeviceComparison
DragerDrager
0 to 13.6Vol%$\pm$ (0.43 Vol%+ 8% rel.)0 to 13.6Vol%$\pm$ (0.43Vol% + 8% rel.)Same
O2O2
Phasein (Sidestream &Mainstream)Phasein (Sidestream &Mainstream)
0 to 100 vol%$\pm$ (1 vol% +2% ofreading)0 to 100 vol%$\pm$ (1 vol% +2% ofreading)Same
DragerDrager
0 to 100Vol%$\pm$ (2.5 Vol% +2.5% rel.)0 to 100Vol%$\pm$ (2.5 Vol%+ 2.5% rel.)Same
N2ON2O
Phasein (Sidestream &Mainstream)Phasein (Sidestream &Mainstream)Same
0 to 100vol%$\pm$ (2 vol% +2% ofreading)0 to 100vol%$\pm$ (2 vol% +2% ofreading)Same
DragerDrager
0 to 100Vol%$\pm$ (2 Vol% + 8% rel.)0 to 100Vol%$\pm$ (2 Vol% +8% rel.)Same
DesDes
Phasein (Sidestream &Mainstream)
0 to 22 vol %22 to 25 vol%$\pm$ (0.15 vol%+ 5% ofreading)Unspecified0 to 22 vol%22 to 25 vol%$\pm$ (0.15 vol%+ 5% ofreading)UnspecifiedSame
DragerDrager
0 to 20 Vol%$\pm$ (0.2 Vol% +15% rel.)0 to 20Vol%$\pm$ (0.2 Vol%+ 15% rel.)Same
SevSev
Phasein (Sidestream &Mainstream)
0 to 10 vol %10 to 25 vol%$\pm$ (0.15 vol%+ 5% ofreading)Unspecified0 to 10 vol%10 to 25 vol%$\pm$ (0.15 vol%+ 5% ofreading)UnspecifiedSame
DragerDrager
0 to 10 Vol%$\pm$ (0.2 Vol% +15% rel.)0 to 10Vol%$\pm$ (0.2 Vol%+ 15% rel.)Same
EnfEnf
Phasein (Sidestream &Mainstream)
0 to 8 vol %8 to 25 vol %$\pm$ (0.15 vol%+ 5% ofreading)Unspecified0 to 8 vol %8 to 25 vol%$\pm$ (0.15 vol%+ 5% ofreading)UnspecifiedSame
DragerDrager

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Lifee Medical Devices. To the left of the logo is a red heart rate monitor symbol. To the right of the symbol is the word "lifee" in purple, with the words "medical devices" in red underneath.

FeatureSubject DevicePredicate DeviceComparison
0 to 10 Vol%IsoPhasein (Sidestream & Mainstream)$\pm$ (0.2 Vol% + 15% rel.)Same
0 to 10 Vol%$\pm$ (0.2 Vol% + 15% rel.)
0 to 8 vol %8 to 25 vol %$\pm$ (0.15 vol% + 5% of reading) UnspecifiedSame
0 to 8 vol %8 to 25 vol %$\pm$ (0.15 vol% + 5% of reading) Unspecified
Drager0 to 8.5 Vol%Hal$\pm$ (0.2 Vol% + 15% rel.)Same
Drager0 to 8.5 Vol%Hal$\pm$ (0.2 Vol% + 15% rel.)
Phasein (Sidestream & Mainstream)
0 to 8 vol %8 to 25 vol %$\pm$ (0.15 vol% + 5% of reading) UnspecifiedSame
0 to 8 vol %8 to 25 vol %$\pm$ (0.15 vol% + 5% of reading) Unspecified
Drager0 to 8.5 Vol%$\pm$ (0.2 Vol% +15% rel.)Same
Drager0 to 8.5 Vol%$\pm$ (0.2 Vol% +15% rel.)
11) BIS
TechniqueBispectral indexBispectral indexSame
MeasuredparametersEEGBIS: 0 to 100EEGBIS: 0 to 100Same
Impedancerange0 to 999 kΩ0 to 999 kΩSame
Sweep speed6.25mm/s, 12.5mm/s,25mm/s, 50mm/s6.25mm/s, 12.5mm/s,25mm/s, 50mm/sSame
Noise (EEGWaveform)<0.3μV (0.25Hz~50Hz)<0.3μV (0.25Hz~50Hz)Same
EEG bandwidth0.25Hz~50Hz0.25Hz~50HzSame
12) RM
Frequencyresponse>10Hz>10HzSame
Flow
MeasurementrangeAdult:2.0 L/min to 180 L/minPediatric:0.75 L/min to 100 L/minNeonatal:0.25 L/min to 30 L/minAdult:2.0 L/min to 180 L/minPediatric:0.75 L/min to 100 L/minNeonatal:0.25 L/min to 30 L/minSame
AccuracyAdult:0.5 L/min or $\pm$ 3% of reading,whichever is greaterPediatric:Adult:0.5 L/min or $\pm$ 3% of reading, whichever is greaterPediatric:Same
FeatureSubject DevicePredicate DeviceComparison
0.25 L/min or ± 3% of reading, whichever is greater0.25 L/min or ± 3% of reading, whichever is greater
Neonatal:0.125 L/min or ± 3% of reading, whichever is greaterNeonatal:0.125 L/min or ± 3% of reading, whichever is greater
Resolution1.0 L/min1.0 L/minSame
Paw (or Airway Pressure)
Measurement range-120 cmH2O to 120 cmH2O-120 cmH2O to 120 cmH2OSame
Accuracy0.5 cmH2O or ± 2% of reading, whichever is greater0.5 cmH2O or ± 2% of reading, whichever is greaterSame
Resolution1 cmH2O1 cmH2OSame
MVe/MVi
Measurement rangeAdult:1 L/min to 30.00 L/minPediatric:0.3 L/min to 20 L/minNeonatal:0.1 L/min to 3 L/minAdult:1 L/min to 30.00 L/minPediatric:0.3 L/min to 20 L/minNeonatal:0.1 L/min to 3 L/minSame
AccuracyAdult: 0.1 L/minPediatric: 0.1 L/minNeonatal: 0.1 L/minAdult: 0.1 L/minPediatric: 0.1 L/minNeonatal: 0.1 L/minSame
TVe/TVi
Measurement rangeAdult: 40 mL to 2500 mLPediatric: 6 mL to 750 mLNeonatal: 2 mL to 100 mLAdult: 40 mL to 2500 mLPediatric: 6 mL to 750 mLNeonatal: 2 mL to 100 mLSame
Resolution1.0 mL1.0 mLSame
AccuracyAdult: ± 10.0 mL or ± 5% of reading, whichever is greaterPediatric: ± 3.0 mL or ± 5% of reading, whichever is greaterNeonatal: ± 1.0 mL or ± 5% of reading, whichever is greaterAdult: ± 10.0 mL or ± 5% of reading, whichever is greaterPediatric: ± 3.0 mL or ± 5% of reading, whichever is greaterNeonatal: ± 1.0 mL or ± 5% of reading, whichever is greaterSame
12) RR (RM)
Measurement rangeAdult:0 rpm to 120 rpmPediatric/Neonate:0 rpm to 150 rpmAdult:0 rpm to 120 rpmPediatric/Neonate:0 rpm to 150 rpmSame
AccuracyAdult:6 to 120 rpm: ±2 rpmAdult:6 to 120 rpm: ±2 rpmSame
0 to 5 rpm: not specified0 to 5 rpm: not specified
FeatureSubject DevicePredicate DeviceComparison
Pediatric/Neonate:Pediatric/Neonate:
6 to 150 rpm: ±2 rpm6 to 150 rpm: ±2 rpm
0 to 5 rpm: not specified0 to 5 rpm: not specified
Resolution1 rpm1 rpmSame
Gain Selectionx0.25, x0.5, x1, x2, x3, x4, 5x5x0.25, x0.5, x1, x2, x3, x4,Same
5x5
Sweep6.25 mm/s, 12.5 mm/s,25mm/s, 50 mm/s6.25 mm/s, 12.5 mm/s,25mm/s, 50 mm/sSame
Apnea Alarm10 s, 15 s, 20 s, 25 s, 30 s,35 s, 40 s; default value is 20 s.10 s, 15 s, 20 s, 25 s, 30 s,35 s, 40 s; default value is 20 s.Same
13) ICG
TechniqueThoracic electricalbioimpedanceThoracic electricalbioimpedanceSame
MeasurementSV: 0 ml/beat250 ml/beatHR: 40 bpm250bpmC.O.: 0 L/min~30 L/minSV: 0 ml/beat250 ml/beatHR: 40 bpm250bpmC.O.: 0 L/min~30 L/minSame
AccuracySV: UndefinedHR: ±2bpmC.O.: UndefinedSV: UndefinedHR: ±2bpmC.O.: UndefinedSame
WI-FI
IEEE802.11b/g/n802.11b/g/nSame
Frequency Band2.4GHz ISM band2.4GHz ISM bandSame
ModulationOFDM with BPSK, QPSK,16-QAM, and 64-QAM802.11b with CCK and DSSSOFDM with BPSK, QPSK,16-QAM, and 64-QAM802.11b with CCK and DSSSSame
Typical TransmitPower (±2 dBm)17 dBm for 802.11b DSSS17 dBm for 802.11b CCK15 dBm for 802.11g/n OFDM17 dBm for 802.11b DSSS17 dBm for 802.11b CCK15 dBm for 802.11g/nOFDMSame
Care and Cleaning
RecommendedcleaningagentsMild neutral detergentEthanol (75%)Isopropanol (70%)Mild neutral detergentEthanol (75%)Isopropanol (70%)Same
Recommendedtypes ofdisinfectingagentsEthanol (75%)Isopropanol (70%)Cidex OPA (High leveldisinfection of intracavitarytemperature probe only)Ethanol (75%)Isopropanol (70%)Cidex OPA (High leveldisinfection of intracavitarytemperature probe only)Same
CleaningSurface-clean the monitorwith a soft cloth dampenedwith the cleaning solutionSurface-clean the monitorwith a soft cloth dampenedwith the cleaning solutionSame
DisinfectingFollowing hospital's policyFollowing hospital's policySame
Safety Classifications
Type ofprotectionagainst electricshockClass IClass ISame
IngressProtectionIPX1IPX1Same
FeatureSubject DevicePredicate DeviceComparison
The degree ofRFGroup 1, Class AGroup 1, Class ASame
The degree of protection against electric shock
ECG, RESP,TEMP, IBP, C.O.CFCFSame
ICG, NIBP, SpO2,CO2, AG, BIS,RM, ICGBFBFSame
Electrical & Mechanical safety & Thermal safety Standards
GeneralStandardsIEC 60601-1:2005IEC 60601-1:2005Same
EMC StandardsIEC 60601-1-2:2007IEC 60601-1-2:2007Same
AlarmStandardsIEC 60601-1-8:2006IEC 60601-1-8:2006Same
BiocompatibilityStandardsISO 10993-1:2009ISO 10993-1:2009Same
SoftwareStandardsIEC 62304:2006IEC 62304:2006Same
Special Standards
Basic safety andessentialperformance forpatient monitorIEC 60601-2-49: 2011IEC 60601-2-49: 2011Same
ECGIEC 60601-2-27: 2011IEC 60601-2-25: 2011IEC 60601-2-27: 2011IEC 60601-2-25: 2011Same
NIBPIEC 80601-2-30: 2009ISO 81060-2IEC 80601-2-30: 2009ISO 81060-2Same
IBPIEC 60601-2-34: 2011IEC 60601-2-34: 2011Same
AG, CO2ISO 80601-2-55ISO 80601-2-55Same
TEMPISO 80601-2-56EN12470-4:2000ISO 80601-2-56EN12470-4:2000Same
Power supply
AC powerRequirement100-240V, 50/60Hz100-240V, 50/60HzSame
Battery
RechargeableBatteryYesYesSame
Operation characteristic
Installation andusePortable EquipmentFix Equipment (when thesystem is installed on WallMounting Bracket)Portable EquipmentFix Equipment (when thesystem is installed on WallMounting Bracket)Same
Working SystemContinuous operationContinuous operationSame
Physical Characteristics
WeightLMPLUS-12: <6.2kgLMPLUS-15: <7.5kgLMPLUS-17: <14kgLMPLUS-12: <6.2kgLMPLUS-15: <7.5kgLMPLUS-17: <14kgSame
DimensionsLMPLUS-12: 333±2 mm (L) x289±2 mm (H)x 211±2mm (W)LMPLUS-12: 333±2 mm (L) x289±2 mm (H)x 211±2mm (W)Same
FeatureSubject DevicePredicate DeviceComparison
LMPLUS-15: 384±2 mm (L) x320±2 mm (H)x 213±2mm (W)LMPLUS-17: 425 mm (L) x 245 mm(W)x 382 mm (H)LMPLUS-15: 384±2 mm (L) x320±2 mm (H)x 213±2mm (W)LMPLUS-17: 425 mm (L) x 245 mm(W)x 382 mm (H)
LCDLMPLUS-12: 12.1 inches LCDLMPLUS-15: 15 inches LCDLMPLUS-17: 17 inches LCDLMPLUS-12: 12.1 inches LCDLMPLUS-15: 15 inches LCDLMPLUS-17: 17 inches LCDSame
LCD ResolutionLMPLUS-12: 800 x 600LMPLUS-15: 1024 x 768LMPLUS-17: 1280 x 1024LMPLUS-12: 800 x 600LMPLUS-15: 1024 x 768LMPLUS-17: 1280 x 1024Same
Environmental Specifications
Temperature
Working+0°C to +40°C+0°C to +40°CSame
Transport andStorage-20°C to +55°C-20°C to +55°CSame
Humidity
Working15% to 95% (non-condensing)15% to 95% (non-condensing)Same
Transport andStorage15% to 95% (non-condensing)15% to 95% (non-condensing)Same
Altitude
Working860hPa to 1060hPa860hPa to 1060hPaSame
Transport andStorage700hPa to 1060hPa700hPa to 1060hPaSame
Other function
Indicators
Alarm indicator3(red/yellow/blue)3(red/yellow/blue)Same
AC powerindicator1(green)1(green)Same
Batteryindicator1(green)1(green)Same
SpeakerYesYesSame
RecorderYesYesSame
Data StorageTrend, NIBP MeasurementReview, Alarm Review,Arrhythmia Review, 12-LeadDiagnosis Review, Full-disclosure waveformsTrend, NIBP MeasurementReview, Alarm Review,Arrhythmia Review, 12-LeadDiagnosis Review, Full-disclosure waveformsSame
InterfaceUSB/VGA/Network/Nursecall/DefibrillatorSynchronization/AnalogOutput/SD/PAM/DVI/RS232USB/VGA/Network/Nursecall/DefibrillatorSynchronization/AnalogOutput/SD/PAM/DVI/RS232Same

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for Lifee Medical Devices. To the left of the logo is a red heart rate monitor symbol. To the right of the symbol is the word "lifee" in purple, with the words "medical devices" in a smaller font below the word "lifee".

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for Lifee Medical Devices. The logo consists of a red heart rate symbol on the left, followed by the word "lifee" in purple. Below the word "lifee" is the phrase "medical devices" in a smaller, lighter font.

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the logo for Lifee Medical Devices. To the left of the company name is a red heart rate monitor symbol. The company name is in purple, with the words "medical devices" in a smaller font below the name.

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the logo for Lifee Medical Devices. On the left is a red heart-shaped design made of jagged lines. To the right of the heart is the word "lifee" in a dark purple color. Below the word "lifee" are the words "medical devices" in a smaller, light red font.

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image shows the logo for Lifee Medical Devices. The logo consists of a red heart rate line on the left, followed by the word "lifee" in purple. Below the word "lifee" is the phrase "medical devices" in a smaller, lighter font.

Performance data:

Non-clinical data:

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

The biocompatibility evaluation for the LMPLUS-12, LMPLUS-15 and LMPLUS-17 Patient Monitor were conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for a duration of less than 24 hours. The tests performed are as follows:

Cytotoxicity Skin Sensitization Skin Irritation

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the LMPLUS-12, LMPLUS-15 and LMPLUS-17 Patient Monitor device, consisting of all the modules and accessories in the system. The system complies with the IEC 60601-1:2012, IEC 60601-1-8: 2006, IEC 60601-2-25: 2011, IEC 60601-2-27: 2011, ANSI/AAMI EC57: 2012, IEC 80601-2-30: 2009, IEC 60601-2-34: 2011, IEC 60601-2-49: 2011, ISO 80601-2-55: 2011, ISO 80601-2-56: 2009, ISO 80601-2-61: 2011 and ISO 81060-2: 2013 standards for safety and the IEC 60601-1-2:2007 standard for EMC.

Table 2 Standards compliance

StandardsConclusion
IEC 60601-1-8 - Medical electrical equipment - Part 1-8: Generalrequirements for basic safety and essential performance - CollateralStandard: General requirements, tests and guidance for alarmsystems in medical electrical equipment and medical electricalsystems- Test report of elite LMPLUS series (2006).Pass
IEC 60601-2-25 Medical electrical equipment - Part 2-25: Particularrequirements for the basic safety and essential performance ofelectrocardiographs - Test Report of elite LMPLUS series (2011)Pass
IEC 60601-2-27 Medical electrical equipment - Part 2-27: Particularrequirements for the basic safety and essential performance ofelectrocardiographic monitoring equipment - Test Report of eliteLMPLUS series 2011Pass
IEC 60601-2-34 - Medical electrical equipment - Part 2-34: Particularrequirements for the basic safety, including essential performance, ofinvasive blood pressure monitoring equipment - Test Report of eliteLMPLUS series 2011Pass
IEC60601-2-49 - Medical electrical equipment Part 2-49: Particularrequirements for the basic safety and essential performance ofmultifunction patient monitoring equipment - Test Report ofelite LMPLUS series (2011) - (This standard is not recognized byFDA)Pass
IEC 80601-2-30 - Medical electrical equipment - Part 2-30: Particularrequirements for the basic safety and essential performance ofautomated non-invasive sphygmomanometers - Test Report of eliteLMPLUS seriesPass
ISO 80601-2-55 - Medical electrical equipment - Part 2-55: Particularrequirements for the basic safety and essential performance ofrespiratory gas monitors - Test Report of elite LMPLUS seriesPass
ISO 80601-2-56 - Medical electrical equipment - Part 2-56: Particularrequirements for basic safety and essential performance of clinicalthermometers for body temperature measurement - Test Report ofelite LMPLUS series.Pass
ISO 80601-2-61 - Medical electrical equipment - Part 2-61: Particularrequirements for basic safety and essential performance of pulseoximeter equipment - Test Report of elite LMPLUS seriesPass
IEC62366 - Medical devices - Application of usability engineering toPass

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image shows the logo for Lifee Medical Devices. To the left of the logo is a red heart rate symbol. To the right of the heart rate symbol is the word "lifee" in purple. Below the word "lifee" is the phrase "medical devices" in red.

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.

{20}------------------------------------------------

Image /page/20/Picture/0 description: The image shows the logo for Lifee Medical Devices. The logo consists of a red heart rate waveform on the left, followed by the word "lifee" in purple. Below the word "lifee" are the words "medical devices" in red.

Clinical data:

Clinical tests were performed on the LMPLUS 12, LMPLUS 15 and LMPLUS 17 monitors to validate their performance in terms of noninvasive blood pressure (NIBP) and SpO2 accuracy.

Summary

Based on the non-clinical and clinical performance as documented in the system development, the subject devices were found to have a safety and effectiveness profile that is similar to the predicate device.

Conclusion

The non-clinical data support the safety of the device and the hardware and software verification and validation demonstrate that the LMPLUS-12, LMPLUS-15 and LMPLUS-17 Patient Monitor device should perform as intended in the specified use conditions. The clinical data demonstrate that the subject devices perform comparably to the predicate device that is currently marketed for the same intended use. In other words, the subject the LMPLUS-12, LMPLUS-15 and LMPLUS-17 Patient Monitor devices are substantially equivalent to the predicate devices.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.