(88 days)
The OneTouch® Delica® Plus lancets are sterile, single use devices that have been designed for single patient use by lay users in a home or general environment. Their intended use is for performing skin punctures on patients for the purpose of obtaining capillary blood samples from the fingertips. The OneTouch Delica Plus lancets are compatible with the single patient use OneTouch® Delica® Plus lancing device.
The OneTouch® Delica® Plus Lancing System consists of a non-sterile, reusable base blood lancing device that is intended for use on a single patient and is used in conjunction with a sterile, single-use lancet. The lancing device has 13 depth settings. The lancets are made available in size 30G and are packaged in boxes of 30 and 100 count.
This device is a OneTouch® Delica® Plus Lancing System, which is a blood lancing system for obtaining capillary blood samples from the fingertips. This submission is for a 510(k) premarket notification, which means it evaluates substantial equivalence to a legally marketed predicate device. Therefore, the "acceptance criteria" are demonstrating substantial equivalence through non-clinical performance testing.
Here's a breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
Since this is a 510(k) submission based on substantial equivalence, the "acceptance criteria" are the successful completion of various non-clinical performance tests to demonstrate that the device is as safe and effective as the predicate. The reported device performance is that it met the specifications for each test.
Acceptance Criteria (Test) | Reported Device Performance |
---|---|
Materials of Construction Specifications met | Specifications met |
Visual, Physical and Dimensional Verification | Specifications met |
Functional Testing | Specifications met |
Performance Testing | Specifications met |
Cleaning and Disinfection | Specification met |
Cytotoxicity Testing per ISO 10993-5 | Passed |
Sensitization Testing per ISO 10993-10 | Passed |
Irritation per ISO 10993-10 | Passed |
Sterilization Validation per ISO 11137-1/-2 | Demonstrates SAL 10⁻⁶ |
Product Sterility per ISO 11737-1 | Demonstrates product is free of microorganisms |
Shelf-life Testing per ASTM F1980 (Lancing device: 4 years, Lancet: 5 years) | Supports a shelf-life of 4 and 5 years |
Transportation Testing per ASTM 4169 | Demonstrates package integrity is maintained |
Risk Analysis per ISO 14971 | Hazards identified, ranked and risk mitigation measures implemented |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify the exact sample sizes used for each non-clinical test. The data provenance is also not explicitly stated in terms of country of origin or whether it was retrospective or prospective, though it would be prospective for these types of verification and validation tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is typically not applicable to non-clinical performance testing of a lancing device. The "ground truth" here is objective measurements and adherence to established standards and specifications, not expert interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable to non-clinical performance testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a blood lancing system, not an AI software. No MRMC study was performed.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This device is hardware, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the non-clinical performance tests, the "ground truth" refers to the established technical specifications, quality standards (e.g., ISO, ASTM), and regulatory requirements that the device must meet. The tests themselves provide objective data to verify compliance with these standards.
8. The sample size for the training set
Not applicable. This is a medical device, not an AI model requiring a training set.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for this device.
§ 878.4850 Blood lancets.
(a)
Single use only blood lancet with an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base (including an integral sharps injury prevention feature) that is used to puncture the skin to obtain a drop of blood for diagnostic purposes. The integral sharps injury prevention feature allows the device to be used once and then renders it inoperable and incapable of further use.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
Single use only blood lancet without an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
Multiple use blood lancet for single patient use only —(1)Identification. A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
Multiple use blood lancet for multiple patient use —(1)Identification. A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class III (premarket approval).(3)
Date PMA or notice of completion of a PDP is required: A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.