K Number
K221546

Validate with FDA (Live)

Date Cleared
2022-08-23

(88 days)

Product Code
Regulation Number
878.4850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OneTouch® Delica® Plus lancets are sterile, single use devices that have been designed for single patient use by lay users in a home or general environment. Their intended use is for performing skin punctures on patients for the purpose of obtaining capillary blood samples from the fingertips. The OneTouch Delica Plus lancets are compatible with the single patient use OneTouch® Delica® Plus lancing device.

Device Description

The OneTouch® Delica® Plus Lancing System consists of a non-sterile, reusable base blood lancing device that is intended for use on a single patient and is used in conjunction with a sterile, single-use lancet. The lancing device has 13 depth settings. The lancets are made available in size 30G and are packaged in boxes of 30 and 100 count.

AI/ML Overview

This device is a OneTouch® Delica® Plus Lancing System, which is a blood lancing system for obtaining capillary blood samples from the fingertips. This submission is for a 510(k) premarket notification, which means it evaluates substantial equivalence to a legally marketed predicate device. Therefore, the "acceptance criteria" are demonstrating substantial equivalence through non-clinical performance testing.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

Since this is a 510(k) submission based on substantial equivalence, the "acceptance criteria" are the successful completion of various non-clinical performance tests to demonstrate that the device is as safe and effective as the predicate. The reported device performance is that it met the specifications for each test.

Acceptance Criteria (Test)Reported Device Performance
Materials of Construction Specifications metSpecifications met
Visual, Physical and Dimensional VerificationSpecifications met
Functional TestingSpecifications met
Performance TestingSpecifications met
Cleaning and DisinfectionSpecification met
Cytotoxicity Testing per ISO 10993-5Passed
Sensitization Testing per ISO 10993-10Passed
Irritation per ISO 10993-10Passed
Sterilization Validation per ISO 11137-1/-2Demonstrates SAL 10⁻⁶
Product Sterility per ISO 11737-1Demonstrates product is free of microorganisms
Shelf-life Testing per ASTM F1980 (Lancing device: 4 years, Lancet: 5 years)Supports a shelf-life of 4 and 5 years
Transportation Testing per ASTM 4169Demonstrates package integrity is maintained
Risk Analysis per ISO 14971Hazards identified, ranked and risk mitigation measures implemented

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the exact sample sizes used for each non-clinical test. The data provenance is also not explicitly stated in terms of country of origin or whether it was retrospective or prospective, though it would be prospective for these types of verification and validation tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is typically not applicable to non-clinical performance testing of a lancing device. The "ground truth" here is objective measurements and adherence to established standards and specifications, not expert interpretation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable to non-clinical performance testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a blood lancing system, not an AI software. No MRMC study was performed.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This device is hardware, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical performance tests, the "ground truth" refers to the established technical specifications, quality standards (e.g., ISO, ASTM), and regulatory requirements that the device must meet. The tests themselves provide objective data to verify compliance with these standards.

8. The sample size for the training set

Not applicable. This is a medical device, not an AI model requiring a training set.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 23, 2022

Asahi Polyslider Co., Ltd. % Stuart Goldman Senior Consultant, RA/QA Emergo by UL 2500 Bee Cave Road, Bldg.1 Suite 300 Austin, Texas 78746

Re: K221546

Trade/Device Name: OneTouch Delica Plus Lancing System Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: QRL, QRK Dated: Mav 25, 2022 Received: May 27, 2022

Dear Stuart Goldman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name

OneTouch® Delica® Plus Lancing System

Indications for Use (Describe)

The OneTouch® Delica® Plus lancets are sterile, single use designed for single patient use by lay users in a home or general environment. Their intended use is for performing skin punctures on patients for the purpose of obtaining capillary blood samples from the fingertips. The OneTouch Delica Plus lancets are compatible with the single patient use OneTouch® Delica® Plus lancing device.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary OneTouch® Delica® Plus Lancing System

1. Submission Sponsor

Asahi Polyslider Company, Ltd. 860-2 Misaki, Maniwa Okayama 719-3226 Japan Yoshitaka Akagi Sr. Manager, Quality Assurance 0867-42-1171

2. Submission Correspondent

Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Stuart R. Goldman Sr. Consultant (512) 327-9997

3. Date Prepared

July 21, 2022

4. Device Identification

Trade/Proprietary Name:OneTouch® Delica® Plus Lancing System
Common/Usual Name:Blood lancets
Classification Name:Blood lancets
Regulation Number:878.4850
Product Code:QRL, QRK
Class:II
Classification Panel:General & Plastic Surgery

5. Legally Marketed Predicate Device

Device name:Accu-Chek Softclix Blood Lancing System
510(k) number:K214022
Manufacturer:Roche Diabetes Care, Inc.

6. Indication for Use Statement

The OneTouch® Delica® Plus lancets are sterile, single use devices that have been designed for single patient use by lay users in a home or general environment. Their intended use is for performing skin punctures on patients for the purpose of obtaining capillary blood samples from the fingertips. The

{4}------------------------------------------------

OneTouch Delica Plus lancets are compatible with the single patient use OneTouch® Delica® Plus lancing device.

7. Device Description

The OneTouch® Delica® Plus Lancing System consists of a non-sterile, reusable base blood lancing device that is intended for use on a single patient and is used in conjunction with a sterile, single-use lancet. The lancing device has 13 depth settings. The lancets are made available in size 30G and are packaged in boxes of 30 and 100 count.

8. Substantial Equivalence Discussion

Table 5-1 compares the OneTouch® Delica® Plus Lancing System to the predicate device with respect to its intended use, indications for use, and technological characteristics and performance, forming the basis for the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device.

AttributeSubject DevicePredicate Device
ManufacturerAsahi PolysliderRoche
510(k) #PendingK214022
Product CodesQRL/QRKQRL/QRK
Device NameOneTouch® Delica® Plus Lancing SystemAccu-Chek Softclix BloodLancing System
DeviceDescriptionThe OneTouch® Delica® Plus Lancing System consistsof the lancing device and compatible lancets to obtaina drop of blood from a fingertip.The Accu-Chek Softclix LancingDevice uses compatible Accu-ChekSoftclix Lancets to obtain a drop ofblood from a fingertip or alternativesites using the Accu-Chek SoftclixAlternative Site Testing (AST) Cap.
Intended UseFor the hygienic collection of capillary blood fortesting purposes.For the hygienic collection ofcapillary blood for testing purposes.
Indications forUseThe OneTouch® Delica® Plus lancets are sterile, singleuse devices that have been designed for single patientuse by lay users in a home or general environment.Their intended use is for performing skin punctures onpatients for the purpose of obtaining capillary bloodsamples from the fingertips. The OneTouch Delica Pluslancets are compatible with the single patient useOneTouch® Delica® Plus lancing device.The Accu-Chek Softclix BloodLancing System is intended for thehygienic collection of capillary bloodfor testing purposes from the side ofa fingertip and from alternativesites, such as the palm, the upperarm, and the forearm.The sterile, single-use lancets are tobe used with the reusable lancingdevice that is to be cleaned anddisinfected between each use, andthen the lancets are to be disposedof.
AttributeSubject DevicePredicate Device
This system is for use only on asingle patient in a home setting.This system is not suitable for use byhealthcare professionals with
multiple patients in a healthcaresetting.
Type of UseOTCOTC
Mechanism ofActionSpring-loadedSpring-loaded
Lancet Sizes(needlediameter)30G33G28G
DepthAdjustments13 levels11 levels
AnatomicalSite(s)FingertipFingertipBall of the handUpper armLower arm
MaterialsLancing device housing:polymerLancing device housing:polymer
Lancet: cap and body (polymer),needle (stainless steel)Lancet: cap and body (polymer),needle (stainless steel)
Sharps InjuryPreventionYes (lancing device)Yes (lancing device)
SterileLancing device: noLancing device: no
Lancet: yes, gamma radiation(SAL = $10^{-6}$ )Lancet: yes, gamma radiation(SAL = $10^{-6}$ )
Single-UseLancing device: no, multiple useLancing device: no, multiple use
Lancet: yesLancet: yes
Shelf-LifeLancing device: (4 years)Lancet: (5 years)

Table 5-1 – Substantial Equivalence Comparison of OneTouch® Delica® Plus Lancing System to the Predicate Device

{5}------------------------------------------------

9. Non-Clinical Performance Data

To demonstrate safety and effectiveness of the OneTouch® Delica® Plus Lancing System and to show substantial equivalence to the predicate device, Asahi Polyslider completed the following verification and validation activities, including non-clinical tests. Results confirm that the design inputs and performance specifications for the device are met. The OneTouch® Delica® Plus Lancing System passed the testing in accordance with internal requirements, national standards, and international standards shown below, supporting its safety and effectiveness, and its substantial equivalence to the predicate device:

{6}------------------------------------------------

  • Materials of Construction Specifications met ●
  • . Visual, Physical and Dimensional Verification – Specifications met
  • Functional Testing – Specifications met
  • Performance Testing – Specifications met
  • Cleaning and Disinfection – Specification met
  • Cytotoxicity Testing per ISO 10993-5 – Passed
  • . Sensitization Testing per ISO 10993-10 – Passed
  • . Irritation per ISO 10993-10 – Passed
  • Sterilization Validation per ISO 11137-1/-2 – Demonstrates SAL 10°
  • . Product Sterility per ISO 11737-1 – Demonstrates product is free of microorganisms
  • Shelf-life Testing per ASTM F1980 – Supports a shelf-life of 4 and 5 years
  • . Transportation Testing per ASTM 4169 – Demonstrates package integrity is maintained
  • . Risk Analysis per ISO 14971 – Hazards identified, ranked and risk mitigation measures implemented

10. Clinical Performance Data

No clinical tests were performed.

11. Substantial Equivalence Conclusion

The OneTouch® Delica® Plus Lancing System has the same intended use as the Accu-Chek Softclix Lancing System and the same or similar technological characteristics. The differences in technological characteristics do not raise new or different questions of safety and effectiveness. Performance testing has demonstrated the subject device is as safe and effective as the predicate device. Therefore, the OneTouch® Delica® Plus Lancing System is substantially equivalent to the predicate device.

§ 878.4850 Blood lancets.

(a)
Single use only blood lancet with an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base (including an integral sharps injury prevention feature) that is used to puncture the skin to obtain a drop of blood for diagnostic purposes. The integral sharps injury prevention feature allows the device to be used once and then renders it inoperable and incapable of further use.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
Single use only blood lancet without an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
Multiple use blood lancet for single patient use only —(1)Identification. A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
Multiple use blood lancet for multiple patient use —(1)Identification. A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class III (premarket approval).(3)
Date PMA or notice of completion of a PDP is required: A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.