K Number
K221522
Date Cleared
2023-01-19

(238 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

HF cables for electrosurgical use in laparoscopic, and open surgery in combination with compatible active accessories and compatible electrosurgical generators

Device Description

The HF-cables (see Table 2.1) subject to this submission are reusable, non-sterile devices, that connect electrosurgical generators as sender to compatible active accessories as receiver for electrosurgical use in laparoscopic, endoscopic, and open surgery. The devices are used as part of a system.

The HF-cables are class II medical device accessories under the regulation number 878.4400 and the product code GEI - "Electrosurgical cutting and coagulation device and accessories". Regulation Medical Specialty: General & Plastic Surgery.

There are monopolar as well as bipolar HF-cable models subject of this submission. All of the subject HF-cables are delivered non-sterile. They are reusable and fully autoclavable. Before first and each subsequent use the cables must be inspected and reprocessed according to defined reprocessing methods in the Instructions for Use.

The HF-cables are available with different cable lengths and with different plugs for the working element/instrument as well as for the electrosurgical generator side. The different plugs allow different compatibilities in accordance with their respective labeling.

AI/ML Overview

Based on the provided text, the device in question is a set of HF-cables (reusable) for electrosurgical use. The document outlines the performance data and testing conducted to demonstrate the safety and effectiveness of these cables for a 510(k) premarket notification.

It's crucial to understand that this document describes the data provided for a 510(k) submission, which aims to demonstrate "substantial equivalence" to a legally marketed predicate device, rather than proving efficacy from scratch like a PMA might. As such, the types of studies and acceptance criteria are focused on demonstrating that the new device performs similarly and safely compared to the predicate, and does not raise new questions of safety or effectiveness.

Here's an analysis of the acceptance criteria and study detailed in the document, keeping in mind the context of a 510(k) submission for reusable HF-cables:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't provide a direct "table of acceptance criteria and reported device performance" in the typical AI/diagnostic software study format (e.g., sensitivity, specificity thresholds). Instead, the acceptance criteria are implicitly met by successful completion of various engineering and safety tests according to recognized standards. The "performance" is the successful completion of these tests.

Acceptance Criteria CategorySpecific Tests/Standards MetReported Device Performance (Implicitly "Met")
BiocompatibilityISO 10993 (Note: Not required as direct/indirect patient contact components are absent)Not applicable / Deemed compliant by design.
Electrical SafetyAAMI/ANSI ES 60601-1:2005/(R)2012, IEC 60601-1-2 Ed. 4.0, IEC 60601-2-2 Ed. 6.0Successfully tested for electrical safety in accordance with listed standards.
Thermal SafetyAAMI/ANSI ES 60601-1:2005/(R)2012, IEC 60601-1:2005/(R)2012Successfully tested for thermal safety in accordance with listed standards.
Functional Performance (Bench)Tests regarding design, transport and storage, repeated stress, mechanical performance testing.Performed as intended; meets design specifications.
Risk ManagementISO 14971 (Risk analysis)Risk analysis carried out in accordance with established internal acceptance criteria.
ReprocessingValidation test reports for cleaning and sterilization procedures.All described methods supported by respective validation test reports.
Usability/Human FactorsIEC 60601-1-6 Edition 3.1, IEC 62366-1 Edition 1.0Compliance with usability engineering standards.
General Device StandardsISO 15223-1 (Symbols), ISO 17664 (Processing info)Compliance with general medical device standards.
Transport SimulationASTM D4169-16Standard Practice for Performance Testing of Shipping Containers and Systems followed.

2. Sample Size Used for the Test Set and Data Provenance:

The document describes engineering bench testing and compliance with standards, not performance on a clinical test set with patient data. Therefore, the concept of "sample size used for the test set" (referring to patient data) and "data provenance" (country of origin, retrospective/prospective) is not applicable to this type of device and submission. The "samples" would be physical units of the HF-cables. The document does not specify the number of cable units tested, which is common for this type of submission.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

This information is not applicable as the ground truth for this device is not established by expert clinical review of data (e.g., medical images). Instead, the "ground truth" is established by the successful passing of engineering and safety tests against predefined, recognized standards and internal specifications.

4. Adjudication Method for the Test Set:

This is not applicable as it's not a study involving human readers or interpretation of clinical data that would require adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

This is not applicable. An MRMC study is relevant for AI or diagnostic imaging devices where human reader performance is being evaluated, with or without AI assistance. This submission is for electrosurgical cables.

6. Standalone (Algorithm Only) Performance:

This is not applicable. This concept applies to AI algorithms. The HF-cables are a physical accessory, not an algorithm. Their "performance" is based on their physical and electrical characteristics when used in a system.

7. The Type of Ground Truth Used:

The "ground truth" for this device is based on:

  • Engineering Specifications: The design and functional requirements for the cables.
  • Recognized Consensus Standards: International and national standards (e.g., IEC 60601 series, ISO 14971) that define safety, electrical performance, thermal limits, and other critical aspects for electrosurgical accessories.
  • Validation of Reprocessing Methods: Verification that cleaning and sterilization procedures are effective.

It is not based on expert consensus, pathology, or outcomes data in a clinical sense as would be the case for a diagnostic device.

8. The Sample Size for the Training Set:

This is not applicable. The concept of a "training set" refers to data used to train machine learning models. This submission is for a physical medical device (HF-cables), not an AI algorithm.

9. How the Ground Truth for the Training Set Was Established:

This is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 19, 2023

Olympus Winter & Ibe GmbH % Christina Flores, RAC Manager, Regulatory Affairs Olympus Surgical Technologies of the Americas 800 West Park Drive Westborough, MA 01581

Re: K221522

Trade/Device Name: HF-cables (reusable) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: II Product Code: GEI Dated: December 20, 2022 Received: December 20, 2022

Dear Christina Flores:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark J. Antonino -S

Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221522

Device Name HF-cables (reusable)

Indications for Use (Describe)

HF cables for electrosurgical use in laparoscopic, and open surgery in combination with compatible active accessories and compatible electrosurgical generators

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary of Safety and Effectiveness

General Information 2.1

Manufacturer/HolderOlympus Winter and Ibe GmbH
Kuehnstr. 61
22045 Hamburg
Germany
Establishment Registration No.: 9610773
Official CorrespondentMs Christina Flores, RAC
Manager, Regulatory Affairs
Olympus Surgical Technologies of the Americas
800 West Park Drive
Westborough, MA 01581
Phone: (508) 808-3341
Email: Christina.Flores@olympus.com
Establishment Registration No.: 3003790304

2.2 Device Identification

Common Name:HF-cables (reusable)
Regulation Number:21 CFR 878.4400
Classification Name:Electrosurgical cutting and coagulation device andaccessories
Device Class:II
Product Code:GEI
Review Panel:General & Plastic Surgery
Proprietary/Trade Name:HF-cables (reusable)

Table 2.1: List of devices subject to this submission

Article No.Article name
A00010AHF-cable, monopolar
A00011AHF-cable, monopolar
A00012AHF-cable, monopolar
A0335.1HF-cable, monopolar
A0355HF-cable, monopolar

{4}------------------------------------------------

510(k) Premarket Notification
---------------------------------
Article No.Article name
A0357HF-cable, monopolar
A0358HF-cable, monopolar
A0391HF-cable, monopolar
A0392HF-cable, monopolar
A0393HF-cable, monopolar
A60000CHF-cable, bipolar
A60001CHF-cable, bipolar
A60002CHF-cable, bipolar
A60003CHF-cable, bipolar
WA00013AHF-cable
WA00014AHF-cable, bipolar

2.3 Predicate Device

The HF-cables (reusable) subject to this submission Table 2.1 are considered substantially equivalent to the following legally marketed predicate device identified in Table 2.2.

Table 2.2: Predicate Device

Predicate DeviceDevice Name510(k) No.
WA00014AHF-cable, bipolarK120418

2.4 Device Description

The HF-cables (see Table 2.1) subject to this submission are reusable, non-sterile devices, that connect electrosurgical generators as sender to compatible active accessories as receiver for electrosurgical use in laparoscopic, endoscopic, and open surgery. The devices are used as part of a system.

The HF-cables are class II medical device accessories under the regulation number 878.4400 and the product code GEI - "Electrosurgical cutting and coagulation device and accessories". Regulation Medical Specialty: General & Plastic Surgery.

There are monopolar as well as bipolar HF-cable models subject of this submission. All of the subject HF-cables are delivered non-sterile. They are reusable and fully autoclavable. Before first and each subsequent use the cables must be inspected and reprocessed according to defined reprocessing methods in the Instructions for Use.

The HF-cables are available with different cable lengths and with different plugs for the working element/instrument as well as for the electrosurgical generator side. The different plugs allow

{5}------------------------------------------------

different compatibilities in accordance with their respective labeling.

2.5 Intended use and indications

HF cables for electrosurqical use in laparoscopic, and open surgery in combination with compatible active accessories and compatible electrosurgical generators.

The indications for use statement for the subject HF-cables is comparable to that of the predicate device. A slightly different wording is chosen and specific indications are added. Furthermore, the sentence "Do not use for any other purposes." was removed.

The differences do not alter the intended use of the devices nor do they raise different questions of safety and effectiveness of the device relative to the predicate.

2.6 Summary of Technological Characteristics

The technological characteristics of the subject devices are considered equivalent to the predicate device.

The subject devices and predicate device share the following characteristics:

  • . general technology (HF-cables connecting electrosurgical generators as sender to compatible active accessories as receiver for electrosurgical use),
  • . validated reprocessing methods, and
  • o expected service life.

The following differences to the predicate device exist depending on the subject device model:

  • design of plugs (different configurations of plugs for connectivity to the compatible . instrument and the electrosurgical generator of each HF-cable),
  • max. rated voltage (only different for monopolar subject devices),
  • . electrical resistance,
  • . cable weight,
  • comparable cable length,
  • wire cross section,
  • . number of single wires, and
  • cable diameter/cable cross section.

As stated above, the subject and predicate devices have similar design characteristics and show comparable performance. As demonstrated in the non-clinical testing the different technological characteristics do not negatively alter the safety and effectiveness of the subject device.

{6}------------------------------------------------

2.7 Performance Data

Performance tests were carried out to ensure that the subject devices function as intended and meet design specifications. The following performance data were provided in support of the substantial equivalence determination.

■ Biocompatibility Testing

The HF-cables do not contain components that come directly or indirectly in patient contact. Biocompatibility testing according to ISO 10993 is not required.

Electrical Safety

The subject HF-cables have been successfully tested for electrical safety in accordance with AAMI/ANSI and IEC 60601 standards (please see list of applied FDA recognized standards below).

Thermal Safety

The subject HF-cables have been successfully tested for thermal safety in accordance with AAMI/ANSI and IEC 60601-1:2005/(R)2012 standards (please see list of applied standards below).

■ Performance Testing Bench

Conducted tests include tests with regard to design, transport and storage, repeated stress and mechanical performance testing.

Risk analysis was carried out in accordance with established internal acceptance criteria based on ISO 14971.

Reprocessing

Required cleaning and sterilization procedures for the HF-cables are described in the respective instructions for use. All described methods are supported by respective validation test reports.

{7}------------------------------------------------

■ Applied Standards

Table 2.3: Applied FDA recognized standards

Standards NumberStandard TitleFDA-RecognitionNumber + date
AAMI/ANSI ES60601-1:2005/(R)2012 andC1:2009/(R)2012 and,A2:2010/(R)2012Medical electrical equipment - Part 1: Generalrequirements for basic safety and essentialperformance (IEC 60601-1:2005, MOD)19-407/09/2014
IEC60601-1-2Ed. 4.0:2014-02Medical electrical equipment - Part 1-2: Generalrequirements for basic safety and essentialperformance - Collateral Standard: Electromagneticdisturbances - Requirements and tests19-809/17/2018
IEC60601-2-2Ed. 6.0:2017-03Medical electrical equipment - Part 2-2: Particularrequirements for the basic safety and essentialperformance of high frequency surgical equipment andhigh frequency surgical accessories6-38906/07/2021
IEC 60601-1-6Edition 3.12013-10Medical electrical equipment - Part 1-6: Generalrequirements for basic safety and essentialperformance - Collateral standard: Usability5-8906/27/2016
IEC 62366-1Edition 1.02015-02Medical devices - Part 1: Application of usabilityengineering to medical devices [IncludingCORRIGENDUM 1 (2016)]5-11412/23/2016
ISO 14971Second edition2007-03-01Medical devices - Application of risk management tomedical devices5-4006/27/2016
ISO 10993-1Fifth edition2018-08Biological evaluation of medical devices - Part 1:Evaluation and testing within a risk managementprocess2-25801/14/2019
ISO 15223-1Third edition2016-11-01Medical devices -- Symbols to be used with medicaldevice labels, labelling and information to be supplied --Part 1: General requirements5-11708/21/2017
ISO 17664Second edition2017-10Processing of health care products - Information to beprovided by the medical device manufacturer for theprocessing of medical devices14-51507/06/2020
ASTM D4169-16Standard Practice for Performance Testing of ShippingContainers and Systems14-49912/23/2016

2.8 Conclusion

The performance data demonstrates the safety of the device as well as compliance with listed recognized consensus standards.

In summary, Olympus believes the subject HF-cables are substantially equivalent to the predicate device with respect to the general design approach, function, and the intended use. The subject HF-cables raise no new concerns of safety or effectiveness compared to the predicate device.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.