(245 days)
Precision Blender (K053232)
Not Found
No
The device description and performance studies focus on pneumatic and mechanical aspects of gas mixing and flow regulation, with no mention of AI or ML technologies.
No
The device is described as an air-oxygen mixer used to administer a precise mix of medical air and oxygen. While it is a medical device used in patient care, its function is to blend gases, not directly provide therapy. It is explicitly stated that "This product is not intended for use as a life supporting device."
No
Explanation: The device is an air-oxygen mixer used to administer a continuous, precise mix of medical air and medical oxygen. It is a therapy device, not a diagnostic one, as it provides a treatment rather than identifying a condition or disease.
No
The device description clearly states it is a "compact air/oxygen mixing device" and a "pneumatic powered device with flowmeters," indicating it is a physical hardware device that mixes gases.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is to "administer a continuous, precise mix of medical air and medical oxygen through its outlet ports" to patients. This is a therapeutic or life-support function, not a diagnostic one.
- Device Description: The device is described as an "air/oxygen mixing device" that provides a "precise mixing of medical grade air and oxygen gases." This aligns with a respiratory support device, not a device used to examine specimens from the human body.
- Lack of IVD Characteristics: There is no mention of the device being used to test or analyze biological samples (blood, urine, tissue, etc.) or to provide information about a patient's health status based on such analysis.
IVD devices are specifically designed to perform tests on specimens derived from the human body to provide information for diagnostic, monitoring, or screening purposes. This device's function is to deliver a specific gas mixture to a patient, which is a different category of medical device.
N/A
Intended Use / Indications for Use
The air-oxygen mixer Quality Mix is used to administer a continuous, precise mix of medical air and medical oxygen through its outlet ports - to infants, children and adults. The exact fractional inspiratory oxygen concentration (FiO2) corresponds to the selected FiO2 setting on the control knob (dial).
It is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician in professional healthcare settings.
This product is not intended for use as a life supporting device.
Warning: By using flows above 70 LPM, the risk of high flow such as ar trapping, barotrauma, and gastric insufflation (leading to possible aspiration) can occur.
Product codes
BZR
Device Description
The air-oxygen blender Quality Mix is a compact air/oxygen mixing device, which provides a precise mixing of medical grade air and oxygen gases. The blender is a pneumatic powered device with flowmeters which can be connected to outlet ports to regulate the gas flow.
The Quality Mix series medical air / oxygen blender includes following models: Quality Mix High Flow; Quality Mix High Flow XL and Quality Mix Low Flow; Quality Mix Low Flow XL.
The Quality Mix series air / oxygen blender is designed, tested and manufactured in accordance with the standard ISO 11195:2018 Gas mixers for medical use - Stand - alone qas mixers.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
infants, children and adults.
Intended User / Care Setting
qualified, trained personnel under the direction of a physician in professional healthcare settings.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Summary of non-clinical tests:
The subject device underwent non-clinical tests according to equivalent standards to demonstrate that it performs as intended and is substantially equivalent to the predicate device Precision Blender (K053232).
-
ISO 11195:2018 (second edition) Gas mixers for medical use - Standalone gas mixers:
- Standard used for: General Performance Test Report related to Standalone Blenders. This testing included applicable elements of the standard for blenders according clause 4 essential performance requirements:
- Clause 9: Reverse gas flow
- Clause 11: Alarm systems
- Clause 12: Accuracy of indicated oxygen concentration
- Clause 13: Gas supply failure
- Result: Pass. The Quality Mix Blender High Flow and Low Flow performs as intended within the specified functions and performance and fulfills the essential performance requirements clause 4 according to the standard ISO 11195:2018-01 for Gas mixers for medical use - Stand - alone gas mixers.
- Standard used for: General Performance Test Report related to Standalone Blenders. This testing included applicable elements of the standard for blenders according clause 4 essential performance requirements:
-
ISO 14971 Second edition 2007-03-01 Medical devices - Application of risk management to medical devices:
- Standard used for: Identification and assessment of hazards and associated risks, the control of these risks and the monitoring of the effectiveness of risk control measures.
- Result: Pass.
-
IEC 62366 – 1 Edition 1.0 2015-02 Medical devices Part 1: Application of usability engineering to medical devices (including CORRIGENDUM 1 2016):
- Standard used for: Verifying and validating the usability of the Quality Mix blender series.
- Result: Pass. Usability report Quality Mix blender series performed in cooperation with an authorized, certified competence center for medical oxygen blenders.
-
German version EN 1789:2007+A2:2014 Medical vehicles and their equipment - Road ambulances; German version EN 60068-2-6:2008 Environmental testing - Part 2-6: Tests - Test Fc: Vibration (sinusoidal); German version EN 60068-2-27:2009 Environmental testing - Part 2-27: Tests - Test Ea and guidance: Shock; German version EN 60068-2-64:2008 Environmental testing – Part 2-64: Tests – Test Fh: Vibration, broadband random and guidance:
- Standard used for: Laboratory Test with Report environmental testing regarding strength, vibration, shock of the device.
- Result: Pass. Vibration (sinusoidal) Test, Shock Test, Vibration, broadband random Test.
-
DIN EN ISO 15002:2020-05 Flow-metering devices for connection to terminal units of medical gas pipeline systems:
- Standard used for: General performance test report for flow-metering devices for connection to blenders of Quality Mix Series (Compensated Thorpe tube flowmeter).
- Result: Pass. This testing included Flowmeter accuracy, Leak test, environmental and mechanical condition.
-
ISO 15001:2010 Anesthetic and respiratory equipment - Compatibility with oxygen:
- Standard used for: To ensure the oxygen compatibility of the used materials.
- Result: Pass. Aspects of compatibility that are addressed include cleanliness, resistance to ignition and the toxicity of products of combustion and/or decomposition.
-
ISO 18562-1 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management process; ISO 18562-2 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 2: Tests for emissions of particulate matter; ISO 18562-3 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compounds:
- Standard used for: Biocompatibility evaluation of breathing gas pathways in healthcare applications.
- Result: Pass. Performed Biocompatibility testing: Gas emission VOC testing with a risk assessment of any identified chemical. Contact classification of blender: Indirect patient contact provided via the mixture of gases brought together by the blender and supplied to the patient (gas pathway contact), for medical devices with gas pathway contact to the patient, compliance to ISO 18562-1 is required. Duration: The duration of use is classified as permanent (more than 30 days) within the scope of the standard ISO 18562-1. Testing: Tests for particulate matter emissions according to ISO 18562-2, and Tests for volatile organic compound (VOC) emissions according to ISO 18562-3 were carried out.
Conclusion: In all cases the subject device passed or meets the acceptance criteria of general requirements for performance and basic safety. The subject device is found substantially equivalent to the predicate device Precision Blender (K053232).
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 868.5330 Breathing gas mixer.
(a)
Identification. A breathing gas mixer is a device intended for use in conjunction with a respiratory support apparatus to control the mixing of gases that are to be breathed by a patient.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services - USA logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 23, 2023
DEHAS Medical Systems GmbH % Tom Hirte Senior Director, Regulatory Affairs Draeger Medical Systems GmBH 3135 Ouarry Road Telford, Pennsylvania 18969
Re: K221494
Trade/Device Name: Quality Mix Blender, Oxymixer Regulation Number: 21 CFR 868.5330 Regulation Name: Breathing Gas Mixer Regulatory Class: Class II Product Code: BZR Dated: December 19, 2022 Received: December 22, 2022
Dear Tom Hirte:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
James J. Lee -S
James J. Lee, Ph.D. Division Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K221494
Device Name Quality Mix Blender, Oxymixer
Indications for Use (Describe)
The air-oxygen mixer Quality Mix is used to administer a continuous, precise mix of medical air and medical oxygen through its outlet ports - to infants, children and adults. The exact fractional inspiratory oxygen concentration (FiO2) corresponds to the selected FiO2 setting on the control knob (dial).
It is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician in professional healthcare settings.
This product is not intended for use as a life supporting device.
Warning: By using flows above 70 LPM, the risk of high flow such as ar trapping, barotrauma, and gastric insufflation (leading to possible aspiration) can occur.
Type of Use (Select one or both, as applicable) |
---|
------------------------------------------------- |
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
8 510(K) SUMMARY
In accordance with 21 CFR 807.92 the following summary of information is provided:
Date: January 12, 2023
Type: Traditional 510(k)
510(K) Number: K221494/S001
8.1. Submitter
DEHAS Medical Systems GmbH
Wesloer Straße 107-109
23568 Luebeck, Germany
Primary contact person
Jens Mittendorf
Managing Director / Head of Engineering & Development
Phone: +49-451-80904-200
+49-451-80904-111 Fax:
E-Mail: J.Mittendorf@dehas.de
Statement
Information supporting claims of substantial equivalence, as defined under the Federal Food, Drug and Cosmetic Act, respecting safety and effectiveness is summarized below. For the convenience of the Reviewer, this summary is formatted in accordance with the Agency's final rule" .... 510(k) Summaries and 510(k) Statements...." (21 CFR 807) and can be used to provide a substantial equivalence summary to anyone requesting it from the Agency.
8.2. Device
Device trade name: Quality Mix blender, Oxymixer
Common name: Breathing gas mixer
Classification name: Anesthesia Inhalation, Breathing Gas Mixer (21 CFR 868.5330)
Regulatory Class: II
Product Code : BZR
Models:
Quality Mix Low Flow, range 3-30 liters per minute
4
Quality Mix Low Flow XL, range 3-30 liters per minute
Quality Mix High Flow, range 15-120 liters per minute
Quality Mix High Flow XL, range 15-120 liters per minute
8.3. Predicated device equivalence
DEHAS Medical Systems GmbH is claiming substantial equivalence to the Precision Blender (K053232) from Precision Medical Inc.
8.4. Indications for use
The air-oxygen mixer Quality Mix is used to administer a continuous, precise mix of medical air and medical oxygen – through its outlet ports – to infants, children and adults. The exact fractional inspiratory oxygen concentration (FiO2) corresponds to the selected FiO2 setting on the control knob (dial).
It is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician in professional healthcare settings.
This product is not intended for use as a life supporting device.
Warning: By using flows above 70 LPM, the risk of high flow such as air trapping, barotrauma, and gastric insufflation (leading to possible aspiration) can occur.
8.5. Patient Group
The Quality Mix blender may be used on equipment where patient population is infant, pediatric and adult patients.
8.6. Environments of use
Professional healthcare settings, where the delivery and monitoring of air/ oxygen mixtures is required.
8.7. Device description
The air-oxygen blender Quality Mix is a compact air/oxygen mixing device, which provides a precise mixing of medical grade air and oxygen gases. The blender is a pneumatic powered device with flowmeters which can be connected to outlet ports to regulate the gas flow.
The Quality Mix series medical air / oxygen blender includes following models: Quality Mix High Flow; Quality Mix High Flow XL and Quality Mix Low Flow; Quality Mix Low Flow XL.
The Quality Mix series air / oxygen blender is designed, tested and manufactured in accordance with the standard ISO 11195:2018 Gas mixers for medical use - Stand - alone qas mixers.
8.8. Technology
The Quality Mix blender is designed to use two gas sources of medical air and oxygen (46,41- 94,27 psi). The two gas sources enter through the diameter - indexed air and oxygen inlet connectors located on the bottom of the Quality Mix blender. Once through the filters, each gas passes through a duckbill check valve which prevents possible reverse gas flow from either the oxygen or the air supply systems. The two gases then pass
5
through a two-stage balance regulator module. The purpose of this regulator module is to equalize the operating pressures of the air and oxygen gas sources.
Once these pressures have been balanced, the gases are proportioned according to the oxygen concentration selected on the oxygen concentration selection dial. The oxygen concentration dial allows the clinician to select a desired oxygen concentration from 21% to 100% O2.
When the pressure difference between the two inlet gases exceeds the prescribed value, or if there is a gas supply failure of one of the gas sources because of disconnection, the bypass poppet will move to one site and create a path for the high-pressure gas to flow into the alarm. This triggers the audible alarm. From this point, the mixed gas flows to the outlet port. The bleed valve on the auxiliary outlet port needs to be activated to ensure the oxygen concentration accuracy of ±3% when the intended output flow in low flow applications is below the specified range of ≤15 lpm for the High Flow blender and ≤3 lpm for the Low Flow blender.
8.9. Determination of substantial equivalence
The subject device has the same indication for use, the same patient group, and the same environment of use as the predicate device Precision Blender (K053232).
The subject device also has the same technological characteristics for example design components, the same function and uses the same energy source as the predicate device. Both devices accept room air and 100% oxygen gas and adjust the molar fraction of oxygen gas in the room air by mechanical means, and blended gas is supplied to the patient. They have identical specifications for the delivery of blended gas and they use the same technoloqy for the audible alarm if the pressure differential between the two inlet gases exceeds the prescribed value or if one gas source is disconnected. Both devices are pneumatically powered only and they produce an audible alarm. The subject device and the predicate device are not intended to be used sterile, nor do either device contain electrical components. The subject device and the predicate device are provided in versions of High Flow and Low Flow blenders.
Please see the table below for more details.
| Item /
Manufacturer | Quality Mix
Blender | Quality Mix
Blender | PRECISION
BLENDER | PRECISION
BLENDER |
|--------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | DEHAS Medical
Systems GmbH
Wesloer Str.
107-109
23568 Luebeck,
Germany | DEHAS Medical
Systems GmbH
Wesloer Str.
107-109
23568 Luebeck,
Germany | Precision
Medical, Inc.®
300 Held Drive,
Northampton,
PA 18067, USA | Precision
Medical, Inc.®
300 Held Drive,
Northampton,
PA 18067, USA |
| Model | Quality Mix
High Flow; XL | Quality Mix Low
Flow; XL | PM5200 High
Flow | PM5300 Low
Flow |
| Item /
Manufacturer | Quality Mix
Blender
DEHAS Medical
Systems GmbH
Wesloer Str.
107-109
23568 Luebeck,
Germany | Quality Mix
Blender
DEHAS Medical
Systems GmbH
Wesloer Str.
107-109
23568 Luebeck,
Germany | PRECISION
BLENDER
Precision
Medical, Inc.®
300 Held Drive,
Northampton,
PA 18067, USA | PRECISION
BLENDER
Precision
Medical, Inc.®
300 Held Drive,
Northampton,
PA 18067, USA |
| 510 (K)
Number | K221494/S001 | K221494/S001 | (K053232) | (K053232) |
| Product Code | BZR | BZR | BZR | BZR |
| Regulation
Number | 21 CFR
868.5330 | 21 CFR
868.5330 | 21 CFR
868.5330 | 21 CFR
868.5330 |
| Patient Group | Infant, pediatric,
and adult
patients | Infant, pediatric,
and adult
patients | Infant, pediatric
and adult
patients | Infant, pediatric,
and adult
patients |
| Primary Outlet
Flow Range | 15 – 120 LPM
With both supply pressures at 50 psi
(3,4 bar) with BLEED closed | 3-30 LPM
With both supply pressures at 50 psi
(3,4 bar) with BLEED closed | 15 – 120 LPM
With both supply pressures at 50 psi
(3,4 bar) with BLEED closed | 3-30 LPM
With both supply pressures at 50 psi
(3,4 bar) with BLEED closed |
| Auxiliary
Outlet Flow
Range | 15 – 120 LPM
With both supply pressures at 50 psi
(3,4 bar) with BLEED closed
0 – 105 LPM
With both supply pressures at 50 psi
(3,4 bar) with BLEED open | 3-30 LPM
With both supply pressures at 50 psi
(3,4 bar) with BLEED closed
0 – 27 LPM
With both supply pressures at 50 psi
(3,4 bar) with BLEED open | 15 – 120 LPM
With both supply pressures at 50 psi
(3,4 bar) with BLEED closed
2-100 LPM
With both supply pressures at 50 psi
(3,4 bar) with BLEED open | 3-30 LPM
With both supply pressures at 50 psi
(3,4 bar) with BLEED closed
0-30 LPM
With both supply pressures at 50 psi
(3,4 bar) with BLEED open |
| Bleed Flow | ≤ 13 LPM at 50
psi (3,4 bar) | ≤ 3 LPM at 50 psi
(3,4 bar) | 13 LPM or less
at 50 psi (3,4
bar) | 3 LPM or less at
50 psi (3,4 bar) |
| Maximum
Combined
Flow (all
outlets) | ≥ 120 LPM | ≥ 30 LPM | ≥ 120 LPM | ≥ 30 LPM |
| Bypass Flow
(loss of air or
oxygen
supply) | > 85 LPM | > 15 LPM | > 85 LPM | > 45 LPM |
| Bypass Alarm
Activation
(range of
activation if | At inlet Pressure
46,41 psi (3,2
bar) | At inlet Pressure
46,41 psi (3,2
bar) | 50 psi (3,45 bar)
13,25 - 25 psi
(0,9 - 1.7 bar) | 50 psi (3,45 bar)
18 - 22 psi
(1.2 - 1.5 bar) |
| Item /
Manufacturer | Quality Mix
Blender | Quality Mix
Blender | PRECISION
BLENDER | PRECISION
BLENDER |
| | DEHAS Medical
Systems GmbH | DEHAS Medical
Systems GmbH | Precision
Medical, Inc.® | Precision
Medical, Inc.® |
| | Wesloer Str.
107-109 | Wesloer Str.
107-109 | 300 Held Drive,
Northampton, | 300 Held Drive,
Northampton, |
| | 23568 Luebeck,
Germany | 23568 Luebeck,
Germany | PA 18067, USA | PA 18067, USA |
| there is a
pressure
difference
between both
gas sources
or loss of gas
supply) | When pressure
difference of one
inlet gas is
detected of ≥
13,05 psi (≥ 0,9
bar). | When pressure
difference of one
inlet gas is
detected of ≥
13,05 psi (≥ 0,9
bar). | | |
| | At inlet Pressure
60,92 psi (4,2
bar) | At inlet Pressure
60,92 psi (4,2
bar) | 60 psi (4.14 bar) | 60 psi (3,.5 bar) |
| | When pressure
difference of one
inlet gas is
detected of ≥
13,05 psi (≥ 0,9
bar). | When pressure
difference of one
inlet gas is
detected of ≥
13,05 psi (≥ 0,9
bar). | 16 – 24 psi
(1.1 - 1.65 bar) | 16 – 24 psi
(1.1 - 1.65 bar) |
| | At inlet Pressure
94,27 psi (6,5
bar) | At inlet Pressure
94,27 psi (6,5
bar) | | |
| | When pressure
difference of one
inlet gas is
detected of ≥
13,05 psi (≥ 0,9
bar). | When pressure
difference of one
inlet gas is
detected of ≥
13,05 psi (≥ 0,9
bar). | | |
| Alarm Reset | When the pressure difference
between the two inlet gases is
restored to ≤ 4,35 psi (0,3 bar) to
each other, the alarm will be reset. | | When pressure differential is 6 psi
(0.4 bar) or less | |
| Alarm Sound
Level | ≥ to 80 dB at a distance of 1 ft. (0,3 m) | | ≥ to 80 dB at 1 ft. (0,3 m) | |
| Oxygen
Concentration
Adjustment
Range | 21 – 100 % | | 21 – 100 % | |
| Gas Supply
pressure | 46,41 – 94,27 psi (3,2 – 6,5 bar) | | 30 - 75 psi (2.1 – 5.2 bar) | |
| Item /
Manufacturer | Quality Mix
Blender | Quality Mix
Blender | PRECISION
BLENDER | PRECISION
BLENDER |
| | DEHAS Medical
Systems GmbH | DEHAS Medical
Systems GmbH | Precision
Medical, Inc.® | Precision
Medical, Inc.® |
| | Wesloer Str.
107-109 | Wesloer Str.
107-109 | 300 Held Drive,
Northampton, | 300 Held Drive,
Northampton, |
| | 23568 Luebeck,
Germany | 23568 Luebeck,
Germany | PA 18067, USA | PA 18067, USA |
| | Air and oxygen pressure differential
should be within max 0.7 bar | | Air and oxygen within 10 psi (0.69
bar) of each other | |
| Mixed Gas
Stability | ± 1% Oxygen | | ± 1% Oxygen | |
| Connection
Types | DISS Type - air & oxygen inlets &
outlets and / or NIST Type - air
&oxygen inlets | | DISS Type - air & oxygen inlets
&outlets and / or NIST Type - air &
oxygen inlets | |
| Dimensions | 12.2" High x 16,5" Wide x 13" Deep
[cm] | | 10.4" High x 5.7" Wide x 12.5" Deep
[cm] | |
| Weight | 1.6 kg | | 2.29 lbs. (1.04 kg) | |
| Shipping
Weight | 4.0 kg | | 2.95 lbs. (1.34 kg) | |
| Operating
Temperature
Range | 41° F to 122°F (5°C to 50°C) | | 59° F to 104°F (15°C to 40°C) | |
| % Oxygen
Control | 21-100% | 21-100% | 21 - 100% | 21-100% |
| Accuracy
(FiO2) | ± 3% | ± 3% | ± 3% | ± 3% |
| Normal
operating
pressure | air/ oxygen | air/ oxygen | air and oxygen | air and oxygen |
| | 46,41 - 94,27
psi | 46,41 - 94,27 psi | 30 - 70 psi | 30 - 70 psi |
| Pressure Drop | High Flow:
≤ 0.21 bar at
inlet pressures
from 3,2 to 6,5
bar, with a flow
rate of 30 LPM
at 60% FiO2 | Low Flow:
≤ 0.14 bar at inlet
pressures from
3,2 to 6,5 bar,
with a flow rate of
10 LPM at 60%
FiO2 | High Flow:
≤ 3 psi (0.21 bar)
at 30-90 psi (2.1
- 6.2 bar) and at
30 LPM flow rate
at 60% FiO2 | Low Flow:
≤ 2 psi (0.14 bar)
at 30-90 psi (2.1 - 6.2 bar) and at
10 LPM flow rate
at 60% FiO2 |
6
7
8
Table 1: Table of comparison
The following quality assurance measures were applied to the development of the device:
9
- Risk analysis .
- Requirements evaluation reports .
- Raw materials verification .
- . Final acceptance testing requirements
- Performance testing verification .
8.10. Summary of non - clinical tests
For comparison, we performed a number of non-clinical tests according to the equivalent standards to demonstrate that the subject device performs as intended and is substantial equivalent to the predicate device Precision Blender (K053232).
We used the following international standards for testing to demonstrate substantial equivalence:
| Used Standard | Standard used for | Result:
Pass /fail |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 11195:2018
(second edition) Gas
mixers for medical use -
Standalone gas mixers | ISO 11195:2018-01 Gas
mixers for medical use -
Stand-alone gas mixers
(second edition)
General Performance Test
Report related to Standalone
Blenders. This testing included
applicable elements of the
standard for blenders
according clause 4 essential
performance requirements:
Clause 9:
Reverse gas flow
Clause 11:
Alarm systems
Clause 12:
Accuracy of indicated oxygen
concentration
Clause 13:
Gas supply failure | The Quality Mix Blender
High Flow and Low Flow
performs as intended
within the specified
functions and
performance and fulfils
the essential performance
requirements clause 4
according to the standard
ISO 11195:2018-01 for
Gas mixers for medical
use - Stand - alone gas
mixers.
Result: pass |
| ISO 14971 Second
edition 2007-03-01
Medical devices - | This international standard has
been used as a process for
our medical devices for the | Result: pass |
| Application of risk
management to medical
devices
FR Recognition List
Number: 043
FR Recognition Number:
5-40 | identification and assessment
of hazards and associated
risks, the control of these risks
and the monitoring of the
effectiveness of risk control
measures. | |
| IEC 62366 – 1 Edition
1.0 2015-02 Medical
devices Part 1:
Application of usability
engineering to medical
devices (including
CORRIGENDUM 1
2016) | This standard has been used
to verify and validate the
usability of the Quality Mix
blender series. | Usability report Quality
Mix blender series
performed in cooperation
with an authorized,
certified competence
center for medical oxygen
blenders.
Result: pass |
| FR Recognition List
Number: 046
FR Recognition Number:
5-114 | | |
| German version EN
1789:2007+A2:2014
Medical vehicles and
their equipment - Road
ambulances | We performed Laboratory Test
with Report environmental
testing regarding strength,
vibration, shock of the device. | Vibration (sinusoidal) Test
Shock Test |
| German version EN
60068-2-6:2008
Environmental testing -
Part 2-6: Tests - Test
Fc: Vibration (sinusoidal) | | Vibration, broadband
random Test
Result: pass |
| German version EN
60068-2-27:2009
Environmental testing -
Part 2-27: Tests - Test
Ea and guidance: Shock | | |
| German version EN
60068-2-64:2008 | | |
| Environmental testing –
Part 2-64: Tests – Test
Fh: Vibration, broadband
random and guidance | | |
| DIN EN ISO
15002:2020-05 Flow-
metering devices for
connection to terminal
units of medical gas
pipeline systems | General performance test
report for flow-metering
devices for connection to
blenders of Quality Mix Series
(Compensated Thorpe tube
flowmeter. | This testing included
Flowmeter accuracy,
Leak test, environmental
and mechanical condition.
Result: pass |
| ISO 15001:2010
Anesthetic and
respiratory equipment -
Compatibility with
oxygen | The standard has been used
to ensure the oxygen
compatibility of the used
materials. | Aspects of compatibility
that are addressed
include cleanliness,
resistance to ignition and
the toxicity of products of
combustion and/or
decomposition.
Result: pass |
| ISO 18562-1
Biocompatibility
evaluation of breathing
gas pathways in
healthcare applications -
Part 1: Evaluation and
testing within a risk
management process
ISO 18562-2
Biocompatibility
evaluation of breathing
gas pathways in
healthcare applications -
Part 2: Tests for
emissions of particulate
matter
ISO 18562-3
Biocompatibility
evaluation of breathing
gas pathways in
healthcare applications -
Part 3: Tests for
emissions of volatile
organic compounds | Biocompatibility evaluation of
breathing gas pathways in
healthcare applications. | We did perform
Biocompatibility testing:
Gas emission VOC
testing with a risk
assessment of any
identified chemical.
Contact classification of
blender:
Indirect patient contact
provided via the mixture
of gases brought together
by the blender and
supplied to the patient
(gas pathway contact), for
medical devices with gas
pathway contact to the
patient, compliance to
ISO 18562-1 is required
Duration:
The duration of use is
classified as permanent
(more than 30 days)
within the scope of the
standard ISO 18562-1 |
| | | Testing:
The following
biocompatibility tests
were carried out: |
| | | Tests for particulate
matter emissions
according to ISO 18562-2 |
| | | Tests for volatile organic
compound (VOC)
emissions according to
ISO 18562-3 |
| | | Result: pass |
10
11
12
Table 2: Summary of non - clinical tests
Conclusion:
In all cases the subject device passed or meets the acceptance criteria of general requirements for performance and basic safety. We can find the subject device substantially equivalent to the predicate device Precision Blender (K053232).
There are no differences which raise any different questions of safety and effectiveness.
8.11. Animal testing
No animal testing was performed.
8.12. Clinical tests
The subject of this premarket submission, Quality Mix blender series, do not require clinical studies to support substantial equivalence.
8.13. Discussion of Differences
There are a few differences between the subject device and the predicate device Precision Blender (K053232). The differences between the predicate and the subject device are addressed when we compare the subject device to the reference devices
They are:
Environments MRI
The environments of use of the Quality Mix series are same as the predicate device Precision Blender (K053232). But the Subjected Device is not for use in MRI Environments, due to the use of ferrous materials / metallic components. Therefore the Device as stated in the safety information's and labeling is marked as MR unsafe. The predicate Device Precision Blender (K053232) can be used in MRI environments of only 3 Tesla MR Systems. Both devices are not intended to be used sterile.
Physical characteristics:
13
510 (K) SUMMARY
Regarding the dimensions and weight, the physical characteristics are similar. All devices have been physically designed and evaluated to be used in clinical settings for their intended use.
Auxiliary outlet flow range (with both supply pressures at 50 psi (3,4 bar) with Bleed Flow open):
The specification of the auxiliary outlet flow range of the air-oxygen blender QualityMix series (High Flow, Low Flow) is similar to the predicate blender devices. The Low Flow and the High Flow blender require a minimum flow rate depending on the model if the auxiliary outlet is in use at low flow applications. This bleed flow is necessary to ensure the concentration accuracy ±3 % when using the blender on this outlet port for low flows, typically below ≤3 lpm (Low Flow blender) and below ≤15 lpm (High Flow blender). Therefore the bleed flow in low flow applications must be activated on this outlet port. Because of this necessary bleed flow, the flow rate in this application on the auxiliary outlet port ensures with an accuracy of ±3 % 0-27 lpm for Quality Mix Low Flow and 0 -105 lpm for Quality Mix High Flow blenders.
The predicate device ensures under same conditions an auxiliary outlet flow range of 0-30 lpm (Low Flow) and 2-100 lpm (High Flow) with an accuracy of ± 3%.
Bypass flow (loss of air or oxygen supply):
Regarding specification of the bypass flow of the Air Oxygen blender Quality Mix Series (High Flow, Low Flow) the performance is similar to the predicate blender devices.
The Quality Mix blender series has an alarm / bypass function, which quarantees a flow of > 85 LPM for the High Flow blenders and > 15 LPM for the Low Flow blenders in emergency situations, if one of the gas supply sources has a malfunction or loss.
The predicate device insures under same conditions a bypass flow of > 85 LPM for the High Flow blender and >45 LPM for the Low Flow blender.
Technical function of bypass flow of the subject device and the predicate device is the same. The bypass flow function ensures that the flow of the remaining gas is maintained. if other one of the two gas supply sources is malfunctioning and the remaining gas supply is at the indicated supply pressure value of normal operating condition.
At both devices, the gas bypass function operates in unison with the alarm. Once the pressure alarm is activated, the bypass function is actuated and the qas with the higher pressure flows directly to the outlet port, bypassing the mixing function of the blender. The oxygen concentration flowing out of the blender will be that of gas with the higher pressure.
Therefore, both devices meet the requirements of the ISO 11195 standard in the case of gas supply failure conditions.
Bypass alarm activation:
The difference is, that the alarm function of the subject device is more sensitive than the predicate device to inform the user of pressure differentials between medical air and medical oxygen. This safety mechanism was taken into account during the development and risk analysis and tested during the usability assessment. In this case, the alarm of the subject device arises at any inlet pressure range of tested normal operating conditions in
14
510 (K) SUMMARY
Page 39
a case of a difference in pressure (O2 and Air) of≥13,05psi (0,9 bar) of the used gases and cease again before the difference in pressure drops below the value of ≤4,35 psi (≤0,3 bar).
The alarm of the predicate device Precision Blender (K053232) arises under tested normal operating conditions at a pressure difference between both gases between ≥16 - 24 psi (1.1 - 1.65 bar).Furthermore, the alarm function of the predicate device is infected and reacts by any different inlet pressure range.
Gas supply pressure / normal operating pressure of the blender:
Regarding normal operating pressure, the subject device was tested and specified within 46.41 - 94.27 psi (3.2 - 6.5 bar) inlet pressure of both gases. The predicate device Precision Blender (K053232) within 30 – 75 psi (2.1 – 5.2 bar). This difference does not raise any questions of safety and effectiveness. Both blenders have been tested by verification and meet the requirements of ISO 11195: 2018 that the gas blenders must cover and work in the entire nominal range of medical gas pipeline systems for compressed medical gases and vacuum according to ISO 7396-1 in hospital environments.
Operating temperature:
The operating temperature of the subject device is similar to the predicate device.
The operating temperature of the subject device is +5C° - +50C°.
The operating temperature of the predicate Precision Blender (K053232) device is 15C° – 40C°.
Pressure drop:
In comparison and according to the technical data and performance test, the subject device has similar performance of pressure drop in the range of ≤ 3 psi with a flow rate of 30 I/ min at 60% FiO2 (High Flow) and ≤ 2 psi with a flow rate of 10 I/ min at 60% FiO2 (Low Flow) at a set inlet pressure of 46,41 - 94,27 psi (3,2 - 6,5 bar).
The predicate device has a performance of pressure drop in the range of ≤ 3 psi with a flow rate of 30 l/ min at 60% FiO2 (High Flow) and ≤ 2 psi with a flow rate of 10 l/ min at 60% FiO2 (Low Flow) at a set inlet pressure of 2.1 - 6.2 bar (normal operating conditions).
The Quality Mix blenders has nearly the same performance specifications as the predicate device. The only difference is the specified range of the optimal and used inlet pressure under normal operating conditions.
8.14. Substantial equivalence conclusion
The subject device has the same intended use as the predicate device, and the technological differences do not raise different questions of safety and effectiveness.
15