K Number
K221249
Device Name
L1P-1F (TRIOS 5)
Manufacturer
Date Cleared
2022-09-16

(137 days)

Product Code
Regulation Number
872.1745
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The L1P-1F intraoral scanner (IOS) system is intended for aid in diagnosis of caries.

Device Description

The L1P-1F intraoral scanner (IOS) system is intended for aid in diagnosis of caries. The L1P-1F system consists of a scanner and a detachable scanner tip which enables scanning using white and blue (fluorescence) Light Emitting Diodes (LEDs). A single-use scanner body sleeve is mounted on the scanner before being introduced into the mouth of a patient for scanning of the teeth.

The interface to the computer is handled by ScanSuite TRIOS which functions as a Hardware Abstraction Layer (HAL). The scan control of, and information from the scanner is handled by the TRIOS module software, which holds the interface to the user.

The L1P-1F system also consists of the TRIOS Patient Monitoring software which aids in diagnosis of occlusal and surface caries.

The functional principle of L1P-1F (TRIOS 5) is to collect images of the object scanned by means of focus scanning, where light from the LEDs travels through the optical system to the object being scanned and back to an image sensor. 3D true colored images of the teeth combined with 2D green and red image as an additional texture are generated by emitting blue and white light in turn.

AI/ML Overview

The provided text describes the 3Shape TRIOS 5 (L1P-1F) intraoral scanner system, which is intended to aid in the diagnosis of caries. The documentation focuses on demonstrating substantial equivalence to a predicate device (DÜRR DENTAL AG – VistaCam iX "Proof" – K150672). However, it does not explicitly provide a table of acceptance criteria and reported device performance in the format requested, nor does it detail a specific study proving the device meets acceptance criteria with all the requested information.

The document discusses key performance attributes tested and compared, which serve as implicit acceptance criteria for establishing substantial equivalence. These include:

  1. Fluorescence scanning
  2. Color Separation
  3. In vitro caries detection study
  4. Spatial resolution

Based on the available information, here's an attempt to structure the response, acknowledging where specific details for the requested categories are not explicitly provided in the document:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied)Reported Device Performance
Fluorescence Scanning PerformanceThe L1P-1F uses two blue LEDs (405 nm) for fluorescence detection, yielding "sufficient performance." The device detects green to red visible light (>435 nm), similar to the predicate device. The blue-violet light causes tooth structures and bacterial metabolites to fluoresce, similar to the predicate.
Color Separation/Categorization for Caries AidThe L1P-1F presents two severity categories: "Initial" (covering early-stage enamel caries, enamel caries up to enamel/dentine junction, and dentin junction already exceeded from the predicate's classification) and "Moderate/Extensive" (for deep dentin caries). This aligns with ICDAS merged codes.
In vitro Caries Detection PerformancePerformance testing demonstrates that L1P-1F is "as safe and effective as a legally marketed device" and "does not raise different questions of safety and effectiveness than the predicate device."
Spatial Resolution PerformanceA qualitative comparison shows "the resolution of L1P-1F must be at least as good or better than the predicate device."
Electrical Safety / Electromagnetic Compatibility (EMC)Complies with IEC 60601-1:2005/AMD1:2012, IEC 60601-1-2:2014, and IEC 62366-1:2015+A1:2020.
BiocompatibilityEvaluated according to relevant ISO 10993-series standards. No biocompatibility concerns were found for materials.
Reprocessing (Scanner tip sterilization, microbial barrier)Scanner tip is autoclavable. Body Sleeve evaluated for tear resistance, tensile strength, puncture resistance, viral penetration, and synthetic blood penetration. Microbial barrier tested against viral ingress (ASTM F1671/F1671M-22). Reprocessing methods align with AAMI TIR12:2010 and AAMI TIR30:2011+R2016.
Software Verification and ValidationVerification and validation conducted as per FDA guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and EN 62304:2006+A1:2015.

2. Sample size used for the test set and the data provenance

The document mentions an "In vitro caries detection study" and performance testing, but does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective nature).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not provide information regarding the number or qualifications of experts used to establish ground truth for any test sets.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

The document does not specify any adjudication method used for establishing ground truth in a test set.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance

The document does not indicate that an MRMC comparative effectiveness study was done involving human readers with and without AI assistance for the L1P-1F device. The focus is on the device's standalone performance compared to a predicate device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

The document describes "Performance Testing" including fluorescence scanning, color separation, in vitro caries detection, and spatial resolution. These tests would inherently involve the device's standalone performance to characterize its technical capabilities, which is then concluded to be "as safe and effective as a legally marketed device." While not explicitly termed "standalone performance study," the non-clinical data and performance testing sections describe the device's intrinsic capabilities.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The document mentions an "in vitro caries detection study," which suggests that the ground truth would likely be established by a recognized method for in vitro caries assessment, such as histology, micro-CT, or a standardized artificial caries model assessed by experts, but this is not explicitly stated.

8. The sample size for the training set

The document does not specify the sample size for any training set. Given the nature of the device (an intraoral scanner with integrated fluorescence technology), it's possible that machine learning or AI components might be involved in the "TRIOS Patient Monitoring software," but details on specific training sets are absent.

9. How the ground truth for the training set was established

Since no training set details are provided, how its ground truth was established is not mentioned.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 16, 2022

3Shape TRIOS A/S Klaus Hoj Senior Regulatory Affairs Specialist Holmens Kanal 7 Copenhagen K, Dk-1060 DENMARK

Re: K221249

Trade/Device Name: L1P-1F (TRIOS 5) Regulation Number: 21 CFR 872.1745 Regulation Name: Laser Fluorescence Caries Detection Device Regulatory Class: Class II Product Code: NBL Dated: August 12, 2022 Received: August 18, 2022

Dear Klaus Hoj:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221249

Device Name TRIOS 5 (L1P-1F)

Indications for Use (Describe) The L1P-1F intraoral scanner (IOS) system is intended for aid in diagnosis of caries.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart G)

Prescription Use (Part 21 CFR 801 Subpart D)

|__ | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) PREMARKET NOTIFICATION 3SHAPE L1P-1F SYSTEM

510(K) SUMMARY - K221249

Date Summary Prepared

15 September 2022

Submitter's Identification

3Shape TRIOS A/S Holmens Kanal 7 DK-1060 Copenhagen K Denmark Phone: +45 7027 2620 E-mail: info@3Shape.com Web: www.3Shape.com

Contact: Klaus Rudbæk Høj Senior Regulatory Affairs Specialist Direct phone: +45 3148 8032 E-mail: Klaus r@3Shape.com

US Agent: F. David Rothkopf MEDIcept Inc. 200 Homer Ave Ashland, MA 01721 Phone: 617-899-3449 E-mail: Drothkopf@MEDIcept.com

Device Identification

Trade / ProprietaryName:TRIOS 5
Models:L1P-1F
Common Name:Intraoral optical scanning system including aid in diagnosis of caries,pod with wireless scanner
Requlation number:Regulation Name:21 CFR 872.1745Laser fluorescence caries detection device.
Product Code:Classification:Panel:NBLClass IIDental (76)

{4}------------------------------------------------

Predicate Device

DÜRR DENTAL AG – VistaCam iX "Proof" – K150672

Device Description

The L1P-1F intraoral scanner (IOS) system is intended for aid in diagnosis of caries. The L1P-1F system consists of a scanner and a detachable scanner tip which enables scanning using white and blue (fluorescence) Light Emitting Diodes (LEDs). A single-use scanner body sleeve is mounted on the scanner before being introduced into the mouth of a patient for scanning of the teeth.

The interface to the computer is handled by ScanSuite TRIOS which functions as a Hardware Abstraction Layer (HAL). The scan control of, and information from the scanner is handled by the TRIOS module software, which holds the interface to the user.

The L1P-1F system also consists of the TRIOS Patient Monitoring software which aids in diagnosis of occlusal and surface caries.

The functional principle of L1P-1F (TRIOS 5) is to collect images of the object scanned by means of focus scanning, where light from the LEDs travels through the optical system to the object being scanned and back to an image sensor. 3D true colored images of the teeth combined with 2D green and red image as an additional texture are generated by emitting blue and white light in turn.

Indications for Use

The L1P-1F intraoral scanner (IOS) system is intended for aid in diagnosis of caries.

Summary of the technological characteristics of the new device compared to predicate device

The L1P-1F system is substantially equivalent in terms of indications for use and technology to the VistaCam iX "Proof" manufactured by Dürr Dental AG, which currently is in commercial distribution and FDA cleared under 510(k) #K150672.

Table 1 below, summarizes the technological characteristics of the L1P-1F system vs. the predicate device.

{5}------------------------------------------------

Table 1 Comparison to the predicate device

ComparisoncriteriaNew DeviceL1P-1F (TRIOS 5)#K221249Predicate DeviceVistaCam iX "Proof" 510(k)#K150672
Owner / Operator3Shape TRIOS A/SDürr Dental AG
Manufacturer3Shape TRIOS A/SDürr Dental AG
Comparison
Intended UseThe L1P-1F intraoral scanner(IOS) system is intended to aidin the detection of tooth decayby measuring increased laserinduced fluorescence.Caries detection aid
Indications for UseThe L1P-1F intraoral scanner(IOS) system is intended for aidin diagnosis of caries.The VistaCam iX "Proof" isintended to be used as an aid inthe detection and diagnosis ofdental caries.
Intended UsersThe L1P-1F System is to beoperated by legally qualifiedhealth care professionals indental clinics.Dentists and dental personnel.
Main partsScannerDetachable scanner tipCommunication softwarebetween external operatingsystem and scannerHandpieceInterchangeable head (Proof)Handpiece holderUSB connection cableCommunication softwarebetween PC and handpiece
Accessories andComponentsScanner holder (pod)Wireless dongleProtection tipBatteries and chargerSoftwareBody SleeveSpacersHygienic protective cover
VariantsDifferent packagingconfigurationsThere are different packagingconfigurations includingdifferent heads.
TechnologyTechnology to obtain 3D digitalimages based on white lightillumination and fluorescencetechnology to aid in diagnosis ofcaries.Fluorescence technology to aidin the detection of cariouslesions.
Mode of OperationThe blue-violet light causestooth structures and bacterialmetabolites to fluoresce.The blue-violet light causestooth structures and bacterialmetabolites to fluoresce
Light Source1 White LED (broad emissionspectrum in the visible) + 2 Blue4 LEDs at 405 nm (blue-violetlight)
ComparisoncriteriaNew DeviceL1P-1F (TRIOS 5)Predicate DeviceVistaCam iX "Proof" 510(k)#K150672
LEDs at 405 nm (blue-violetlight)
DetectionWavelengthGreen to red visible light,> 435 nmGreen to red visible light,495 nm peak wavelength
Power SupplyPower from batteriesUses power from PC USB port
ConnectivityWirelessWired
SoftwareTRIOS Module (imagingsoftware)ScanSuite TRIOS (advanceddriver software)TRIOS Patient Monitoring (cariesdetection aid software)VistaCam iX "Proof" driversoftware and DSBWIN.VistaCam iX "Proof" can only beused with DBSWIN imagingsoftware
Image Evaluation3D image color indicating cariesstageRed: Moderate/ExtensiveYellow: InitialWhite: Insufficient scanThe colors of the diagnosticoverlay are derived from thefluorescence texture recordedfrom the tooth substanceImage ColorRelative levelGreen: 0 - 1.0Blue: 1.0 - 1.5Red 1.5 - 2.0Orange 2.0 - 2.5Yellow 2.5 - 3.0Lighter colors and highernumbers indicate an increasingmagnitude of the red / greenfluorescence ratio and thelikelihood of poorer tooth health.
Spatial ResolutionObtained from many 2D imagesat different focus positionsstitched togetherN/A. No publicly availableinformation on spatial resolutionperformance.
Returned LightReflection for white LEDilluminationFluorescence for blue LEDilluminationFluorescence
OperatingEnvironmentAmbient temperature range: 15-26°CRelative humidity: 10-85%(non-condensing)Atmospheric pressure: 800-1100hPaThe scanner should be used inan environment maintaining aconsistent room temperatureAmbient conditions duringoperation:Temperature range: 10-40°CRelative humidity: 20-75%(non-condensing)Air pressure: 800-1060 hPa
MaterialsScanner:Shell: PolycarbonateBattery inlet: PC/ABSScanner tip:Handpiece housing:ASA + PC (Acrylonitrile StyreneAcrylate + Polycarbonate)Handpiece control buttons:
ComparisoncriteriaNew DeviceL1P-1F (TRIOS 5)Predicate DeviceVistaCam iX "Proof" 510(k)#K150672
PSUScanner tip window:Sapphire with FDTS and Al2O3Protection tip:PSUPolyester Umbilical – SiliconeCamera cover: Ethylene methylLens window: Glass
ComponentsimpactingreprocessingScanner tipBody SleeveCamera coverSpacer
BiocompatibilityISO 10993-1ISO 10993-5ISO 10993-10ISO 10993-11ISO 10993-12ISO 10993-17ISO 10993-18ISO 10993-23ISO 10993-5ISO 10993-10
Method ofsterilizationAutoclave (scanner tip)Autoclave (only spacer)
Electrical Safety &EMCIEC 60601-1IEC 60601-1-2IEC 60601-1IEC 60601-1-2IEC 80601-2-60

{6}------------------------------------------------

{7}------------------------------------------------

Discussion of Differences

Power Source / Connectivity

The VistaCam iX "Proof" only comes in a wired version whereas the L1P-1F is made in a wireless variant where the power source is batteries.

The batteries are certified and tested according to IEC 62133-2:2017 ('Secondary cells and batteries containing alkaline or other non-acid electrolytes. Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications').

The wireless part complies with 47 CFR, Part 15, and has been tested to comply with IEC 60601-1-2:2014 ('Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances. Requirements and tests').

Light Source

The L1P-1F includes two types of LEDs as the light source: one white LED (for 3D image generation) and two blue LEDs (405 nm) as contrary to the predicate device which has four LEDs with blue-violet light (405 nm). The arrangement with two blue LEDs yields sufficient performance for fluorescence detection.

{8}------------------------------------------------

The LEDs have been tested and assessed to be in Risk Group 1 (Low-Risk) according to IEC 62471:2008 ('Photobiological safety of lamps and lamp systems').

Detection wavelength

The L1P-1F and predicate device are detecting similar wavelengths and the images are equivalent.

Detection wavelength is >435 nm for L1P-1F and 495 nm for the predicate device. Both devices detect green to red visible light (in particular, L1P-1P for wavelengths above 435 nm).

It is assumed that the predicate device detection wavelength of 495 nm refers to the wavelength of highest sensitivity.

Software

The software package for the L1P-1F consists of the following parts: TRIOS Module (imaging software), ScanSuite TRIOS (advanced driver software), and TRIOS Patient Monitoring (caries detection aid software).

The software package for the VistaCam iX "Proof" consists of the VistaCam iX "Proof" driver software and DSBWIN. VistaCam iX "Proof" can only be used with DBSWIN imaging software.

Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

The software has been validated according to EN 62304:2006+A1:2015, Medical Device Software Life-Cycle Processes.

Image Evaluation

The color categories used by the two devices to aid image interpretation are named differently.

The predicate device presents four caries severity categories, while the L1P-1F presents two severity caries categories. Table 2 shows how they correspond to each other.

Table 2 Color categories

VistaCam iX ProofL1P-1F
Early-stage caries,incipient enamel cariesInitial
Enamel caries up to enamel/dentine junctionInitial
Dentin junction already exceededInitial
Deep dentin cariesModerate/Extensive

L1P-1F is following the merged codes of the International Caries Detection and Assessment System (ICDAS) summarizing the first three categories of the VistaCam stages into one.

510(K) SUMMARY

{9}------------------------------------------------

With the intended use to aid caries diagnosis, the important clinical performance characteristic is to distinguish between a healthy tooth surface and a surface with initial decay or early-stage lesion. The additional caries severity categories that are provided in the predicate device are not expected to trigger any difference in the equivalent clinical performance and safety of caries assessments. However, these early caries stages would usually receive the same non-operative treatment approach by the dental practitioners and the distinction between them would therefore be of less clinical significance as well as reliability.

Spatial Resolution

The spatial (3D) model is obtained from many 2D images at different focus positions stitched together by the imaging software.

There is no publicly available data on spatial resolution for the predicate device. But a qualitative comparison of obtained shows that the resolution of L1P-1F must be at least as good or better than the predicate device.

Components impacting reprocessing

Both L1P-1F and the predicate device have exchangeable tips for scanning and single-use covers.

The predicate device has a protective cover which covers the entire scanner and thus acts as a microbial barrier. The microbial barrier for L1P-1F consists of a protective cover (Body Sleeve) and an autoclavable tip. The Body Sleeve is designed to cover the body of the scanner as well as the base of the detachable tip, which is closed with a window in the front end.

The Body Sleeve has been evaluated to ensure that requirements for clearance of dental barriers and sleeves requlated under 21 C.F.R §878.4370, FDA product code PEM, are met. This included evaluation of tear resistance, tensile strength, puncture resistance, viral penetration, and synthetic blood penetration.

The microbial barrier functionality of the scanner tip has been tested to show that no viral ingress through the tip window will reach the scanner body as tested using the test organism described in ASTM F1671/F1671M-22.

In conclusion, the microbial barrier properties implemented for the L1P-1F system are equivalent to the predicate device and ensures that cleaning and intermediate-level disinfection can be recommended for the scanner body between patients.

Non-Clinical Data

Key performance attributes tested and compared include:

    1. Fluorescence scanning
    1. Color Separation
    1. In vitro caries detection study
    1. Spatial resolution

{10}------------------------------------------------

The results of the performance testing demonstrates that L1P-1F is as safe and effective as a legally marketed device and does not raise different questions of safety and effectiveness than the predicate device.

Performance Testing

Electrical Safety / Electromagnetic Compatibility (EMC)

Testing was performed for the L1P-1F and complies with:

  • . IEC 60601-1:2005/AMD1:2012 (Medical electrical equipment - Part 1: General requirements for basic safety and essential performance)
  • . IEC 60601-1-2:2014 (Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and Tests)
  • . IEC 62366-1:2015+ A1:2020 (Medical Devices - Part 1: Application of Usability Engineering to Medical Devices).

Biocompatibility Testing

Patient contact components were evaluated according to relevant recognized standards in the ISO 10993-series. It was concluded that there are no biocompatibility concerns for any of the materials used to manufacture the L1P-1F.

Sterilization

The L1P-1F is not sold as a sterile device.

The Scanner tip is required to be processed prior to first use and between patients. Testing to support recommended methods for reprocessing was performed in accordance with AAMI TIR12:2010 ('Designing, testing, and labeling reusable medical devices for reprocessing in health care facilities: A guide for medical manufacturers') and AAMI TIR30:2011+R2016 ('A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices').

Conclusion

The proposed device, the L1P-1F, is substantially equivalent to the predicate device Vistacam iX "Proof".

The differences between the proposed and predicate device do not impact the safety and effectiveness of the proposed device. Performance testing supports that L1P-1F is substantially equivalent to the legally marketed predicate device.

§ 872.1745 Laser fluorescence caries detection device.

(a)
Identification. A laser fluorescence caries detection device is a laser, a fluorescence detector housed in a dental handpiece, and a control console that performs device calibration, as well as variable tone emitting and fluorescence measurement functions. The intended use of the device is to aid in the detection of tooth decay by measuring increased laser induced fluorescence.(b)
Classification. Class II, subject to the following special controls:(1) Sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter;
(2) Premarket notifications must include clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
(3) The labeling must include detailed use instructions with precautions that urge users to:
(i) Read and understand all directions before using the device,
(ii) Store probe tips under proper conditions,
(iii) Properly sterilize the emitter-detector handpick before each use, and
(iv) Properly maintain and handle the instrument in the specified manner and condition.