K Number
K220946
Manufacturer
Date Cleared
2022-04-15

(14 days)

Product Code
Regulation Number
892.1650
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SURGIVISIO device is intended to be used during surgical procedures in which the physician would benefit from the visualization of 2D medical imaging and/or intraoperatively generated 3D medical structures or objects with high x-ray attenuation such as bony anatomy or metallic objects. Such procedures during which the spine, pelvis or articulation structures are visualized.

Device Description

The SURGIVISIO Device is a mobile x-ray system which provides 2D imaging and allows the generation of intraoperative 3D information of high contrast objects and anatomical structures. The system consists of two mobile interconnected units: a mobile C-arm and a mobile viewing Workstation. These units are moved manually and are interconnected by a single cable that provides power and transfer of data and controls. The mobile C-Arm comprises a high voltage generator, foot switches for radiation release, laser target devices, electronics cabinet, collision avoidance system, and a C-shaped structure mounting the X-ray tube assembly and the flat X-ray detector on distal end of the 'C'. The mobile viewing workstation comprises a computer, an image detector process unit, the main power supply, radiation indicator, dual viewing monitors and a user interface for patient management and image handling.

The system integrates a Computer Aided Surgery (CAS) feature that supports instruments positioning during surgical procedures.

The SPX1 Instruments are a set of dedicated accessories intended to be used in conjunction with the SURGIVISIO Device in order to enable stereotaxic guidance. More specifically, and when used with the 3D imaging, they support the registration of patient anatomy with the 3D reconstructed image. the Submissions K202547/ K220627 consider SPX1 instruments set with the standard 3D Imaging Set SPX1 accessory.

This submission aims at clearing a new imaging set accessory possibility: the 3D Offset Imaging Set SPX1 accessory, containing same patient reference and wedges as 3D imaging Set SPX1 accessory already 510(k) cleared, and an offset phantom (new instrument). The principle of use of the Offset Phantom SPX1 is identical to the already 510(k) cleared Phantom SPX1 (belonging to 3D Imaging set). However, the Offset Phantom SPX1 additional value is to offset the imaging area that will be navigated during surgery by 10 cm from the standard configuration. The 3D Imaging Set SPX1 accessory will be considered as the predicate for the 3D Offset Imaging Set SPX1 accessory.

AI/ML Overview

I am sorry, but I cannot fulfill your request to describe the acceptance criteria and study proving device performance based on the provided text. The document is primarily a 510(k) summary for a medical device (SURGIVISIO Device) and focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed acceptance criteria and the comprehensive study results that prove the device meets those criteria.

Here's why I cannot extract the requested information:

  • Lack of Specific Acceptance Criteria: The text avers to "standards, acceptance criteria and results" in the "Performance Data" section but explicitly states that those details can be found in a "Design Control Activities Summary," which is not included in the provided document. Therefore, the specific numerical or qualitative acceptance criteria for the device's performance are missing.
  • Focus on Substantial Equivalence, Not De Novo Performance Study: The document's core purpose is to show that a new accessory (3D Offset Imaging Set SPX1) for an existing device (SURGIVISIO Device) is substantially equivalent to a previously cleared version. It states, "No clinical tests were conducted to demonstrate substantial equivalence." The non-clinical tests mentioned are primarily design verification, accuracy verification, software verification, and packaging validation, all aimed at confirming the new accessory functions similarly and safely to the predicate, not necessarily to establish de novo performance against a set of independent acceptance criteria for the entire system from scratch.
  • Missing Study Details: While some non-clinical tests are listed (design verification, accuracy verification, software verification, packaging validation), the text does not provide the methodology, sample sizes, data provenance, ground truth establishment, expert qualifications, adjudication methods, or direct performance metrics in the way you've requested for a detailed study. It merely states that these tests "have been performed as intended" or "have been conducted as intended."

In summary, the provided FDA document is a regulatory submission for substantial equivalence, not a detailed scientific study report containing the specifics of acceptance criteria and the comprehensive data proving the device meets them. The crucial information you've asked for (e.g., actual performance values against acceptance criteria, detailed methodologies, sample sizes for test sets, ground truth establishment details, MRMC study results) is explicitly stated to be elsewhere ("Design Control Activities Summary") or is simply not part of this type of regulatory summary.

{0}------------------------------------------------

April 15, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is the Department of Health & Human Services logo. The right side of the logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Ecential Robotics % Mathilde Saulpic Quality Assurance & Regulatory Affairs Engineer Zone Mayencin II, Parc Equation - Bâtiment 1, 2 avenue de Vignate Gieres, 38610 FRANCE

Re: K220946

Trade/Device Name: SURGIVISIO Device Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, OXO, JAA Dated: March 29, 2022 Received: April 1, 2022

Dear Mathilde Saulpic:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220946

Device Name SURGIVISIO Device

Indications for Use (Describe)

The SURGIVISIO device is intended to be used during surgical procedures in which the physician would benefit from the visualization of 2D medical imaging and/or intraoperatively generated 3D medical structures or objects with high x-ray attenuation such as bony anatomy or metallic objects. Such procedures during which the spine, pelvis or articulation structures are visualized.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitter Information

Submitter:ECENTIAL ROBOTICS2 avenue de VignateZone Mayencin II, Parc Equation – Bâtiment 138610 GièresFrance
Contact Person:Mathilde SAULPICQuality Assurance & Regulatory Affairs EngineerPhone : 0033 4 58 00 35 86Email: mathilde.saulpic@ecential-robotics.com
Date Summary Prepared:08 April 2022

{4}------------------------------------------------

Summary of the technological characteristics of the device compared to the predicate device

There have been no modifications on indication for use on SURGIVISIO Device since the previously approved submission K202547 & K220627.

This submission aims at adding a new Imaging set: 3D Offset Imaging Set SPX1 accessory, which enables to deport imaging zone regarding the already cleared 3D Imaging Set SPX1 accessory, giving user more flexibility by freeing space around the Patient reference SPX1 and the zone of interest. The new 3D Offset Imaging Set SPX1 accessory will not replace the already cleared 3D Imaging Set SPX1 accessory.

Device Information

Trade or proprietary name:SURGIVISIO Device
Common or usual name:Mobile Interventional Fluoroscopic X-ray System
Classification Name:Interventional Fluoroscopic X-Ray System
Regulation Number:21 CFR 892.1650 - Image-intensified fluoroscopic x-ray system
Regulatory class:II
Primary product code:OWB
Legally marketed device to which equivalence is claimed:K202547 & K220627 - SURGIVISIO device - Manufacturer: ECENTIAL ROBOTICS (previously SURGIVISIO)
Device Description:The SURGIVISIO Device is a mobile x-ray system which provides 2D imaging and allows the generation of intraoperative 3D information of high contrast objects and anatomical structures. The system consists of two mobile interconnected units: a mobile C-arm and a mobile viewing Workstation. These units are moved manually and are interconnected by a single cable that provides power and transfer of data and controls. The mobile C-Arm comprises a high voltage generator, foot switches for radiation release, laser target devices, electronics cabinet, collision avoidance system, and a C-shaped structure mounting the X-ray tube assembly and the flat X-ray detector on distal end of the 'C'. The mobile viewing workstation comprises a computer, an image detector process unit, the main power supply, radiation indicator, dual viewing monitors and a user interface for patient management and image handling.

{5}------------------------------------------------

The system integrates a Computer Aided Surgery (CAS) feature that supports instruments positioning during surgical procedures.

The SPX1 Instruments are a set of dedicated accessories intended to be used in conjunction with the SURGIVISIO Device in order to enable stereotaxic guidance. More specifically, and when used with the 3D imaging, they support the registration of patient anatomy with the 3D reconstructed image. the Submissions K202547/ K220627 consider SPX1 instruments set with the standard 3D Imaging Set SPX1 accessory.

This submission aims at clearing a new imaging set accessory possibility: the 3D Offset Imaging Set SPX1 accessory, containing same patient reference and wedges as 3D imaging Set SPX1 accessory already 510(k) cleared, and an offset phantom (new instrument). The principle of use of the Offset Phantom SPX1 is identical to the already 510(k) cleared Phantom SPX1 (belonging to 3D Imaging set). However, the Offset Phantom SPX1 additional value is to offset the imaging area that will be navigated during surgery by 10 cm from the standard configuration. The 3D Imaging Set SPX1 accessory will be considered as the predicate for the 3D Offset Imaging Set SPX1 accessory.

  • Indication for use: The SURGIVISIO device is intended to be used during surgical procedures in which the physician would benefit from the visualization of 2D medical imaging and/or intraoperatively generated 3D medical imaging of anatomical structures or objects with high x-ray attenuation such as bony anatomy or metallic objects. Such procedures include procedures during which the spine, pelvis or articulation structures are visualized. Indication for use is the same as the predicate.
  • Technology comparison Same as the predicate. The technology is not impacted by the discussed device change, object of this Special 510(K) submission. This is an x-ray system that generates 2D and 3D images of anatomical structures. The 3D Offset Imaging Set accessory is used in combination with the imaging system in the same way than the 3D Imaging Set SPX1 accessory.
  • Biocompatibility Same as the predicate. The discussed device change, object of this Special 510(K) submission does not require to re-perform Biocompatibility testing as the raw material, manufacturing processes are the same for 3D Imaging Set SPX1 accessory than for 3D Offset Imaging Set SPX1 accessory.

{6}------------------------------------------------

Electrical Safety Testing Same as the predicate. The discussed device change, object of this Special 510(K) submission does not impact Electrical Safety.

Performance Testing Same as the predicate.

Software Testing Same as the predicate.

Software change to enable the identification of the Offset Phantom SPX1 were already taken into account in submission K220627. The clearance of the present Special 510(k) would enable to release the associated functionalities by clearing the associated required instrument.

SURGIVISIO software was verified and validated according to IEC 62304 Medical Device Software. Results of these validations and verifications confirm that the SURGIVISIO software complies with its specifications.

  • Clinical images Same as the predicate. The discussed device change, object of this Special 510(K) submission does not impact clinical images.
  • Conclusion Based upon comparison of devices and performance testing results, SURGIVISIO Device using 3D Offset Imaging Set SPX1 accessory is substantially equivalent to the predicate device.

{7}------------------------------------------------

DevicesSURGIVISIO Device usedwith 3D Offset ImagingSet SPX1 accessorySURGIVISIO Device(K202547 & K220627)used with 3D ImagingSet SPX1 accessory
Intended use / Indication for useThe SURGIVISIO device isintended to be used duringsurgical procedures inwhich the physician wouldbenefit from thevisualization of 2D medicalimaging and/orintraoperatively generated3D medical imaging ofanatomical structures orobjects with high x-rayattenuation such as bonyanatomy or metallicobjects. Such proceduresinclude procedures duringwhich the spine, pelvis orarticulation structures arevisualized.The SURGIVISIO device isintended to be used duringsurgical procedures inwhich the physician wouldbenefit from thevisualization of 2D medicalimaging and/orintraoperatively generated3D medical imaging ofanatomical structures orobjects with high x-rayattenuation such as bonyanatomy or metallicobjects. Such proceduresinclude procedures duringwhich the spine, pelvis orarticulation structures arevisualized.
Primary product codeOWBOWB
Mechanical configurationMobile C-ArmMobile C-Arm
Movement rangeVertical: Up to 43.5cmHorizontal: Up to 28cmOrbital: 200°Angulation: 180°Vertical: Up to 43.5cmHorizontal: Up to 28cmOrbital: 200°Angulation: 180°
Movement controlMotor-drivenMotor-driven
kV Range40 - 120 kV40 - 120 kV
mA Range1 – 120 mA1 – 120 mA
Pulse frequency1 - 12.5 fps1 - 12.5 fps
Detector technologyFlat panel, 287mm x 265mmFlat panel, 287mm x 265mm
Image Matrix Size1560x1440 pixels780x720 pixels1560x1440 pixels780x720 pixels
X-ray tube technologyRotating anode0.3 - 0.8 focal spotRotating anode0.3 - 0.8 focal spot
2D Imaging2D Fluoroscopic2D Fluoroscopic
Pulsed FluoroscopyYesYes
AERC systemYes, kv/mA curve typesYes, kv/mA curve types
3D ImagingYesYes
Rotating movement for 3Dimaging180°180°
DevicesSURGIVISIO Device usedwith 3D Offset ImagingSet SPX1 accessorySURGIVISIO Device(K202547 & K220627)used with 3D ImagingSet SPX1 accessory
3D Imaging characteristicsCylindrical volume: 15 xØ13 cmElliptic cylindrical volume:15 x Ø116 x Ø218 cmResolution 400 x 400 x 400voxelsCylindrical volume: 15 xØ13 cmElliptic cylindrical volume:15 x Ø116 x Ø218 cmResolution 400 x 400 x 400voxels
Monitor cart/WorkstationYesYes
Screen displays2 Monitors 22" tactileScreen Displays2 Monitors 22" tactileScreen Displays
Image output formatDICOMDICOM
LAN network connectionYesYes
Computer AidedSurgery(CAS) interfaceYes, integratedYes, integrated
3D Imaging set accessorycomponentsOffset Phantom SPX1 withits ID cardPatient reference SPX1with its ID cardAssociated wedgesPhantom SPX1 with its IDcardPatient reference SPX1with its ID cardAssociated wedges
3D Imaging set accessorycomponents packagingOffset Phantom SPX1 andpatient reference withassociated wedgespackaged in one uniqueblisterBlister Material andreference:PETG & Tyvek 1073BPhantom SPX1 andpatient referencepackaged in its own BlisterPatient reference SPX1with associated wedgespackaged in its own BlisterBlister Material andreference:PETG & Tyvek 1073B
Software workflow3D Spine Universalworkflow3D Spine Universalworkflow

{8}------------------------------------------------

{9}------------------------------------------------

Performance Data

For further information about testing, standards, acceptance criteria and results, please refer to section Design Control Activities Summary.

Nonclinical tests: The following nonclinical tests were performed on the SURGIVISIO Device using 3D Offset Imaging Set SPX1 accessory to demonstrate substantial equivalence of safety and efficacy with the predicate device:

  • -Design verification activities have been performed for the Offset Phantom SPX1
  • -Accuracy verification has been conducted at the system level with the 3D Offset Imaging Set SPX1 accessory as intended
  • -Software verification testing verifying the software requirements has been performed as intended and are already taken into account by the previous 510(k) cleared version (Submission K220627).
  • Packaging process validation and verification activities, including Shelf-life validation, have been conducted as intended

Clinical tests: No clinical tests were conducted to demonstrate substantial equivalence.

Conclusions drawn from Performance Data

The SURGIVISIO Device using 3D Offset Imaging Set SPX1 accessory is identical in indications for use and technological characteristics as the proposed predicate device. These aspects, along with the performance testing conducted, demonstrate the substantial equivalence to the SURGIVISIO Device (K202547/ K220627) and that the SURGIVISIO Device using 3D Offset Imaging Set SPX1 accessory does not raise different questions of safety and effectiveness when compared to this predicate.

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.