(884 days)
The Ensol EnTM Collection and Transport System is intended for the collection of clinical samples containing upper respiratory viruses including Influenza A, Human Coronavirus 229E, and Respiratory Syncytial Virus (RSV) from the collection site to the testing laboratory to be used with standard diagnostic identification techniques that utilize stable recoverable infectious viral particles.
The EnTM Collection and Transport System medium consists of a polypropylene conical tube filled with 2 ml of the transport medium (pale brown to red color medium solution), affixed with a polyethylene screw cap. The bottom part of the tube has a self-standing shape. Tubes are provided in a paper rack. The media tubes can be provided with one or two kinds of sterile specimen collection swabs, one for oropharyngeal (OP) oral use and the other nasopharyngeal (NP) for nasal use. The swab shaft is polystyrene with a breaking point, and the swab tip is flocked nylon fibers. The media is provided in two different configurations with and without the sterile peel pouch containing swabs.
The provided text describes the Ensol EnTM Specimen Collection and Transport System, a viral transport medium, and its non-clinical performance evaluation. It does not contain information about an AI/ML powered device, therefore, many of the requested categories are not applicable.
Here's an analysis of the provided information:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Physical Stability | |
| Appearance | Acceptable (red and clear solution) |
| UV Spectrum Analysis (250-800 nm) | Similar results for all lots at all storage times |
| Absorbance at 290 nm | No changes, similar results for all lots at all storage times |
| pH Measurement | Within 7.4 ± 1 for all lots at all storage times |
| Sterility | No bacterial or fungal proliferation detected |
| Culture-Based Viral Recovery | |
| Average viral titer percent change within ±90% (1 log change) for Influenza A, Human Coronavirus 229E, and Respiratory Syncytial Virus (RSV) at 4°C storage for 0, 24, and 48 hours. | All tested viral strains (Influenza A, Human Coronavirus 229E, and RSV) maintained viral viability at 4°C storage for up to 48 hours, with percent changes within the ±90% (1 log change) acceptance criteria. Refer to Table 1 in the original text for specific values. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Physical Stability and Sterility: The study used three lots of the Ensol EnTM Collection and Transport System. The provenance of the data (country of origin, retrospective/prospective) is not explicitly stated.
- Culture-Based Viral Recovery Study:
- Sample Size: Three lots of VTM (ENTM-2001, ENTM-22003, and ENTM-23001) were used. For each lot and each virus (Influenza A, Human Coronavirus 229E, and Respiratory Syncytial Virus), aliquots were stored at 4°C for 0, 24, and 48 hours. The number of replicates for each viral sample is not specified beyond "aliquots of each replicate".
- Data Provenance: The study was conducted in a laboratory setting using specific viral strains and host cells. The country of origin of the data and whether it was retrospective or prospective is not explicitly stated, but it implicitly describes a prospective study designed to evaluate the device's performance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This information is not applicable as the device is a viral transport medium, not an AI/ML powered diagnostic device that requires expert ground truth establishment for image or data interpretation. The "ground truth" in this context refers to the controlled viral titers and their expected viability, which are established through standard laboratory techniques and biological assays.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This information is not applicable. Adjudication methods are typically used in clinical studies involving multiple readers or assessors to resolve discrepancies in subjective interpretations, which is not relevant for the objective measurements performed on a viral transport medium.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. The device is a viral transport medium, not an AI-assisted diagnostic tool. Therefore, no MRMC study or AI assistance evaluation was performed or is relevant.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
This information is not applicable. The device is a viral transport medium and does not involve any algorithm or AI for standalone performance evaluation.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth for the viral recovery study was based on laboratory-established viral titers (TCID50). This involves:
- Using established, characterized viral strains (Influenza A, Human Coronavirus 229E, and RSV).
- Determining the initial viral concentration at 0 hours using standard cell culture infectivity assays (Reed-Muench method calculation of TCID50).
- Comparing the viral titers at 24 and 48 hours to the initial titers to assess changes in viability.
8. The sample size for the training set
This information is not applicable. The device is a physical medical device (viral transport medium), not an AI/ML model that requires a training set.
9. How the ground truth for the training set was established
This information is not applicable as there is no training set for this type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a human figure, and the text on the right is in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.
August 29, 2024
Ensol Biosciences Inc. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 18881 Von Karman Ave. STE 160 Irvine, California 92612
Re: K220908
Trade/Device Name: Ensol EnTM Specimen Collection and Transport System Regulation Number: 21 CFR 866.2390 Regulation Name: Transport Culture Medium Regulatory Class: Class I, reserved Product Code: JSM Dated: October 30, 2023 Received: October 30, 2023
Dear Priscilla Chung:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
Please note that if you modify your IVD in the future to exceed any of the limitations to the exemption found in 21 CFR 866.9(c), your device will require a new 510(k) prior to marketing this device in the United States and will not be exempt from the premarket notification requirements so long as it exceeds the limitations to the exemption found in 21 CFR 866.9.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Ribhi Shawar -S
Ribhi Shawar, Ph.D. (ABMM) Chief General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K220908
Device Name Ensol EnTM Collection and Transport System
Indications for Use (Describe)
The Ensol EnTM Collection and Transport System is intended for the collection of clinical samples containing upper respiratory viruses including Influenza A, Human Coronavirus 229E, and Respiratory Syncytial Virus (RSV) from the collection site to the testing laboratory to be used with standard diagnostic identification techniques that utilize stable recoverable infectious viral particles.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) Summary: K220908
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: 8/29/2024
Applicant / Submitter
Ensol Biosciences Inc. 51, Techno 10-ro, Yuseong-gu, Daejeon 34036 Republic of Korea Tel +82-42-939-4500
U.S. Designated Agent
Priscilla Chung LK Consulting Group USA, Inc. 18881 Von Karman Ave. STE 160 Irvine, CA 92612 Fax: 714.409.3357 Tel: 714.202.5789 Email: juhee.c@LKconsultingGroup.com
Trade/Proprietary Name:
Ensol EnTM Specimen Collection and Transport System
Common Name:
Viral Transport Medium (VTM)
Classification:
Non-propagating transport culture media (21CFR 866.2390, Product code JSM, Class 1, Microbiology)
Device Description:
The EnTM Collection and Transport System medium consists of a polypropylene conical tube filled with 2 ml of the transport medium (pale brown to red color medium solution), affixed with a polyethylene screw cap. The bottom part of the tube has a self-standing shape. Tubes are provided in a paper rack. The media tubes can be provided with one or two kinds of sterile specimen collection swabs, one for oropharyngeal (OP) oral use and the other nasopharyngeal (NP) for nasal use. The swab shaft is polystyrene with a breaking point, and the swab tip is flocked nylon fibers. The media is provided in two different configurations with and without the sterile peel pouch containing swabs as shown below:
| Cat No. | Tube with VTM | Swab | Packaging |
|---|---|---|---|
| ES-TM-01 | 2 mL of VTM solution in16 x 105 mm conicalshape bottom screw-captube | One applicator NP swab withflocked nylon fiber tip withbreaking point | 50 tubes with culture mediumper box50ea. NP swabs50ea. OP swabs |
{5}------------------------------------------------
| One applicator throat swab withflocked nylon fiber tip withbreaking pointBoth swabs packaged together ina sterilized peel pouch | |||
|---|---|---|---|
| ES-TM-02 | 2 mL of VTM solution in16 x 105 mm conicalshape bottom screw-captube | No swabs | 50 tubes with culture mediumper box |
Principle of Operation:
The EnTM Collection and Transport System is an osmotically balanced and buffered medium that contains Hank's balanced salt solution. Also, the medium contains a pH indicator, a protein stabilizer, sucrose as a preservative and antibiotics to prevent contamination from bacteria or fungi. The EnTM Collection and Transport System is used to safely collect and transport upper respiratory viruses from collection sites to the testing laboratories. It is intended to be used by health care professionals.
After collecting a specimen using the swab, the swab is placed into the uncapped polypropylene tube filled with 2 mL of the transport medium, the swab shaft is broken off at the breakpoint by hand and the tube is capped closed. The swab specimen can be stored in the collection medium for up to 48 hours at 4℃. The swab specimen is ready for use in diagnostic to determine the presence of a target pathogen. Sample collection is performed by health care professionals.
Indication for use:
The EnTM Collection and Transport System is intended for the collection and transportation of clinical samples containing upper respiratory viruses including Influenza A, Human Coronavirus 229E, and Respiratory Syncytial Virus (RSV) from the collection site to the testing laboratory to be used with standard diagnostic/identification techniques that utilize stable recoverable infectious viral particles.
Predicate Device:
Copan Universal Transport Medium (UTM-RT) System (K042970) by COPAN DIAGNOSTICS, INC.
Substantial Equivalence:
The subject and predicate devices comparison is described in the table below:
| Device and Predicate Device | Device: K220908 | Predicate: K042970 |
|---|---|---|
| Device Trade Name | EnTM Collection and Transport System | Copan Universal Transport Medium(UTM-RT) System |
| Device Product Code andClassification | JSM | JSM |
| General DeviceCharacteristic Similarities |
{6}------------------------------------------------
| Intended Use/Indications ForUse | The EnTM Collection and TransportSystem is intended for the collectionand transportation of clinical samplescontaining upper respiratory virusesincluding Influenza A, HumanCoronavirus 229E, and RespiratorySyncytial Virus (RSV) from thecollection site to the testinglaboratory to be used with standarddiagnostic/identification techniquesthat utilize stable recoverableinfectious viral particles. | Copan Universal Transport Medium(UTM-RT) System is intended forthe collection and transport ofclinical specimens containingviruses, chlamydiae, mycoplasma orureaplasma from the collection siteto the testing laboratory. UTM-RTcan be processed using standardclinical laboratory operatingprocedures for viral, chlamydial,mycoplasma and ureaplasmaculture. |
|---|---|---|
| Tube Material | Plastic | Same |
| Single Use Device | Yes | Same |
| Sterile Device | Yes | Same |
| Shelf Life | 12 months | Same |
| General Device | ||
| Characteristic Differences | ||
| List of Ingredients | Modified Hank's Balanced Salts withcalcium and magnesium ionsSucroseBovine serum albuminHEPESGelatinL-CysteineL-Glutamic acidGentamicin sulfateAmphotericin BVancomycinColistin sulfate saltPhenol Red sodium salt | Hank's Balanced Salts,Sucrose,Bovine serum albumin,HEPES BufferGelatinL-CysteineL-Glutamic acidAmphotericin BVancomycinColistinPhenol Red |
| Samples Transported toPerform | Assays to detect the following upperrespiratory viruses: Influenza A,Human Coronavirus 229E, andRespiratory Syncytial Virus (RSV) | |
| Specimen transportconditions and storage time | Specimen stored at 4°C andprocessed within 48 hours | Specimen stored at 4 -25 °C andprocessed within 48 hours |
Shelf-life stability:
The shelf life for the EnTM Collection and Transport System was determined to be 12 months from the date of manufacture when stored at a temperature of 25℃. The shelf life of the EnTM Collection and Transport Medium was evaluated using real-time aging performance test.
Physical Stability: Physical stability was evaluated by Appearance, UV spectrum analysis (Scan: absorbance at 250-800 nm), and pH measurement at time points T = 0, 6, and 12 months using three lots stored at 25℃ and higher. The appearance was evaluated by inspecting visually to examine for a red and clear solution. Appearance results were acceptable. To detect any physical changes in the media, UV Spectrum analysis was conducted by setting the wavelength range to 250-800 nm (scan). The UV spectrum scanning analysis for 250-800 nm produced similar results for all lots at all storage times. Absorbance at 290 nm was also measured for any physical changes and results showed no
{7}------------------------------------------------
changes and produced similar results for all lots at all storage times. For pH measurement study, all the transport media demonstrated pH level within 7.4±1 for all lots at all storage times.
Sterility test: The EnTM Collection and Transport System is not claimed to be sterile nor is it intended to be sterilized by the end user. To reduce the contamination, the screw-cap tube is sterilized by irradiation with a maximum dose of 40 kGv using the E-beam irradiation process. A check for possible contamination was conducted by transferring transport media to growth medium Thioglycolate and incubated at 35 ℃ for 3 days, and in Tryptic Soy Broth at 25℃ for 5 days and examined for growth turbidity. No bacterial or fungal proliferation was detected at either condition tested.
Culture-Based Viral Recovery Study:
Performance of the EnTM Collection and Transport System media was evaluated for virus viability at refrigerated temperatures (4 °C). Three lots of VTM with newly manufactured, middle-aged lot and close to expiry lots were used to evaluate viral recovery (ENTM-2001, ENTM-22003. and ENTM-23001). Influenza A (California/07/2009 H1N1, NCCP 42467). Human coronavirus 229E (Zeptometrix 0810229CF), and Human respiratory syncytial virus type A (type A strain Long, ATCC VR-26) were used for the viral recovery study. Each virus was diluted in pooled negative nasopharyngeal matrix transferred to the collection swab and then added to the Viral Transport Media. Aliquots of each replicate were stored at 4°C for 0, 24, and 48 hours. Host cells (MDCK for Influenza A, MRC-5 for Human coronavirus, and HEp-2 for Human respiratory syncytial virus type A) were plated for 1 day. At each timepoint the viral samples from EnTM were serially diluted and inoculated onto the host cell-line and cultured in a 96 well plate. The plates were incubated at 37°C and viral CPE was observed after 4-5 days. Viral titer was determined by the Reed-Muench method calculation of TCID50 (50% tissue culture infective dose).
Results were considered acceptable if the average viral titer demonstrates any percent changes within ±90% (i.e.,1 log change). The results are presented in the Table 1. Any changes in the viral titer in the timepoints was shown as percent changes and negative percent change indicates reduction.
| Test Virus | Lot | Average Virus Titer (TCID50/mL) | Percent Changes(0 to 24 hrs.)* | Percent Changes(0 to 48 hrs.)* | ||
|---|---|---|---|---|---|---|
| 0 hr | 24 hrs. | 48 hrs. | ||||
| Influenza A | ENTM-22001 | 3.58E+07 | 6.49E+06 | 9.07E+06 | -81.84% | -74.64% |
| Influenza A | ENTM-22003 | 2.80E+07 | 1.07E+07 | 1.33E+07 | -61.70% | -52.53% |
| Influenza A | ENTM-23001 | 3.34E+07 | 1.35E+07 | 6.04E+06 | -59.53% | -81.89% |
| Human Coronavirus 229E | ENTM-22001 | 1.11E+04 | 1.16E+04 | 1.18E+04 | 3.73% | 5.58% |
| Human Coronavirus 229E | ENTM-22003 | 1.93E+04 | 2.36E+04 | 8.04E+03 | 22.25% | -58.41% |
| Human Coronavirus 229E | ENTM-23001 | 1.14E+04 | 1.08E+04 | 8.87E+03 | -5.69% | -22.36% |
| Respiratory Syncytial Virus | ENTM-22001 | 7.20E+05 | 5.20E+05 | 6.65E+05 | -27.73% | -7.66% |
| Respiratory Syncytial Virus | ENTM-22003 | 8.71E+05 | 4.05E+05 | 7.02E+05 | -53.47% | -19.43% |
| Respiratory Syncytial Virus | ENTM-23001 | 1.06E+06 | 3.72E+05 | 5.82E+05 | -64.99% | -45.26% |
Table 1: Recovery of Viruses at 4°C Storage.
{8}------------------------------------------------
*-ve indicates decline.
EnTM Collection and Transport System media demonstrated the recovery of viruses at 4°C storage when evaluated for 48 hrs. in all replicates of tested viral strains (Influenza A, Human Coronavirus 229E, and Respiratory Syncytial Virus). All the results support specimen transport for up to 48 hours when stored at 4°C.
Conclusion:
The documentation submitted in this premarket notification demonstrates that the subject device K220908 has comparable features and performance and is substantially equivalent to the predicate device.
§ 866.2390 Transport culture medium.
(a)
Identification. A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.(b)
Classification. Class I (general controls).