K Number
K220852
Device Name
iCare IC200
Manufacturer
Date Cleared
2022-10-06

(197 days)

Product Code
Regulation Number
886.1930
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The iCare IC200 tonometer is intended to be used for the measurement of intraocular pressure of the human eye.
Device Description
The iCare IC200 tonometer (model TA031) is a hand-held, battery-operated device which measures intraocular pressure (IOP) without the need for topical anaesthesia by rebound tonometry. The tonometer is to be used by a healthcare professional. The tonometer uses the rebound method. A small and light, sterile, single-use probe makes brief contact with the eye. The device measures the deceleration of the probe and the rebound time and calculates the IOP from these parameters. Deceleration of the probe is slower at low IOP compared to high IOP. The measurement method, the IOP measurement algorithm and rebound technology (including disposable probe) are identical with the predicate device. iCare IC200 tonometer can obtain IOP measurements in all angles between horizontal and supine directions. External design has not been modified. The user interface menu has been updated to contain a Quick Measure mode. The new Quick Measure feature is used to measure patient's IOP with fewer rebound measurements and faster measurement cycle. Quick Measure takes two or three rebound measurements; two if both results are within 2 mmHg and third if the difference between the first two measurements is greater than 2 mmHg.
More Information

Not Found

No
The description focuses on a simple algorithm based on comparing two or three measurements and does not mention AI or ML.

No.
The device is used for measurement (diagnosis), not for treatment or therapy.

Yes

The device measures intraocular pressure, which is a physiological parameter used to assess the health of the eye and diagnose conditions like glaucoma. The "Intended Use" explicitly states it's for "measurement of intraocular pressure of the human eye," indicating its role in providing objective data for diagnostic purposes.

No

The device description explicitly states it is a "hand-held, battery-operated device" and describes physical components like a "probe" and the "rebound method," indicating it is a hardware device with integrated software.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is to measure intraocular pressure of the human eye. This is a direct measurement of a physical parameter within the body, not an analysis of a sample taken from the body (like blood, urine, or tissue).
  • Device Description: The device measures the deceleration and rebound time of a probe making contact with the eye. This is a physical measurement technique, not a chemical, biological, or immunological test performed on a sample.
  • Lack of IVD Characteristics: The description does not mention any analysis of biological samples, reagents, or laboratory procedures typically associated with IVD devices.

IVD devices are specifically designed to examine specimens derived from the human body to provide information for diagnostic, monitoring, or compatibility purposes. This tonometer measures a physical property of the eye itself.

N/A

Intended Use / Indications for Use

The iCare IC200 tonometer is intended to be used for the measurement of intraocular pressure of the human eye.

Product codes

HKY

Device Description

The iCare IC200 tonometer (model TA031) is a hand-held, battery-operated device which measures intraocular pressure (IOP) without the need for topical anaesthesia by rebound tonometry. The tonometer is to be used by a healthcare professional.

The tonometer uses the rebound method. A small and light, sterile, single-use probe makes brief contact with the eye. The device measures the deceleration of the probe and the rebound time and calculates the IOP from these parameters. Deceleration of the probe is slower at low IOP compared to high IOP. The measurement method, the IOP measurement algorithm and rebound technology (including disposable probe) are identical with the predicate device.

iCare IC200 tonometer can obtain IOP measurements in all angles between horizontal and supine directions. External design has not been modified. The user interface menu has been updated to contain a Quick Measure mode. The new Quick Measure feature is used to measure patient's IOP with fewer rebound measurements and faster measurement cycle. Quick Measure takes two or three rebound measurements; two if both results are within 2 mmHg and third if the difference between the first two measurements is greater than 2 mmHg.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

human eye

Indicated Patient Age Range

Not Found

Intended User / Care Setting

healthcare professional

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

Quick Measure performance data was obtained from a retrospective clinical data analysis in which Quick Measure function was retrospectively compared to the clinical data of iCare IC200. In the analysis Quick Measure met ANSI Z80.10:2014 (in accordance with FDA's extent of recognition) performance goals per GAT reference tonometer.
Accuracy of the iCare IC200 quick measure mode was assessed in a bench test using a manometrically controlled artificial cornea. Testing compared the new Quick Measure mode against the existing default measurement mode (same as in predicate device IC200), clinically validated subnormal reference tonometer (iCare IC200), and manometric pressure.
Repeatability and accuracy testing completed with iCare IC200 quick measure mode demonstrated high agreement with manometric pressure and reference tonometer used in default measurement mode (iCare IC200).
Reproducibility test completed with iCare IC200 quick measure mode demonstrated high agreement with manometric pressure regardless of the device, operator, or the measurement angle.

Key Metrics

Outside ±5 mmHg vs. GAT reference (Sitting)
GAT group = low ≤ 16 mmHg: 0/44 (0.0 %)
GAT group = medium > 16 to 16 to

§ 886.1930 Tonometer and accessories.

(a)
Identification. A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.(b)
Classification. Class II.

0

October 6, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Icare Finland Oy Hannes Hyvönen Regulatory Affairs Manager Ayritie 22 Vantaa. 01510 Finland

Re: K220852

Trade/Device Name: iCare IC200 Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer And Accessories Regulatory Class: Class II Product Code: HKY Dated: September 5, 2022 Received: September 8, 2022

Dear Hannes Hyvönen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices

1

or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia. Respiratory, ENT & Dental Devices Office of Product Evaluation and Ouality (OPEO) Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K220852

Device Name iCare IC200

Indications for Use (Describe)

The iCare IC200 tonometer is intended to be used for the measurement of intraocular pressure of the human eye.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

ical

iCare IC200 Traditional 510(k)

510(k) Summary

1.1 Submitter Name and Address

Icare Finland Oy Äyritie 22 Vantaa, Finland FI-01510

Contact: Hannes Hyvönen Phone: 358 9 8875 1150 Email: regulatory(at)icare-world.com Date prepared: October 5th, 2022

1.2 Device Name

Trade Name:iCare IC200
Type/model:TA031
Common/Usual Name:Tonometer
Classification Name:Tonometer and Accessories
Regulation No:21 CFR 886.1930
Device Regulatory Class:II
Review Panel:Ophthalmic
Product Code:HKY

Premarket Notification 510(k) Number: -

This premarket notification type is Traditional 510(k).

1.3 Predicate Device

Predicate device: iCare IC200 tonometer (type TA031), K190316.

The predicate device has not been subject to a design-related recall.

4

icare

1.4 Indications for Use / Intended Use

"The iCare IC200 tonometer is intended to be used for the measurement of intraocular pressure of the human eye."

The intended use of the iCare IC200 tonometer remains unchanged by this design modification.

1.5 Device Description

The iCare IC200 tonometer (model TA031) is a hand-held, battery-operated device which measures intraocular pressure (IOP) without the need for topical anaesthesia by rebound tonometry. The tonometer is to be used by a healthcare professional.

The tonometer uses the rebound method. A small and light, sterile, single-use probe makes brief contact with the eye. The device measures the deceleration of the probe and the rebound time and calculates the IOP from these parameters. Deceleration of the probe is slower at low IOP compared to high IOP. The measurement method, the IOP measurement algorithm and rebound technology (including disposable probe) are identical with the predicate device.

iCare IC200 tonometer can obtain IOP measurements in all angles between horizontal and supine directions. External design has not been modified. The user interface menu has been updated to contain a Quick Measure mode. The new Quick Measure feature is used to measure patient's IOP with fewer rebound measurements and faster measurement cycle. Quick Measure takes two or three rebound measurements; two if both results are within 2 mmHg and third if the difference between the first two measurements is greater than 2 mmHg.

510(k) Summary

5

Image /page/5/Picture/1 description: The image shows the word "icare" in a bold, sans-serif font. The letters are black, and there is a thin blue line underneath the word. The word appears to be a logo or brand name.

1.6 Comparison of Technological Characteristics with the Predicate Device

Similarities of the subject device and the predicate device is presented in the Table 1 below:

Table 1 Similarities of the subject device and the predicate device
-------------------------------------------------------------------------------
CharacteristicModified device (subject of this 510k)Cleared Device (K190316)
Product
Appearance and
materialsImage: Modified deviceImage: Cleared device
SameSame
Product/Device
IdentificationSameiCare IC200 Tonometer
(Type: TA031)
Intended Use /
Indications for
Use StatementSameIOP Measurement
“The iCare IC200 tonometer is intended
to be used for the measurement of
intraocular pressure of the human eye.”
Intended usersSameHealthcare professionals
Measurement
methodSameRebound tonometry
Measurement
rangeSame7-50 mmHg
Default
measurement
sequenceSameDefault measurement mode: 6
measurements, calculation of median IOP
and standard deviation based on 4
measurements (highest and lowest
excluded)
Versatility of
Measurement
positionSameTonometer can be used in any angle
between 0° (sitting, standing) and 90°
(patient in supine position)
Device DisplaySame1.50" OLED display
Automatic eye
recognition
systemSameNot applicable
DesignSameHandheld microprocessor based
CalibrationSameNo maintenance calibration required
Contact tip
(probe)SameLightweight, disposable, single use,
plastic probe
(TP01s)
Contact tip
sterilizationSameGamma-sterilized
Anaesthesia
requiredSameNo
Power supplySame4 x 1,5V AA Alkaline LR6 batteries
Device
dimensions and
weightSameDimensions:
43mm x 104mm x 214mm
Weight: 165g (without batteries), 267g
(with batteries)
Connectivity
interfaceSameBluetooth (Microchip RN4678 Module)
ElectronicsSameInclination sensor, External Flash
memory, Real time clock
MechanicsSameAs described in K190316
User interfaceAudio indications:
SameAudio indications:
Beeps for device too near or too far
situation, and for other errors.
Graphical User Interface:
SameGraphical User Interface:
OLED display
Probe base light:
SameProbe base light:
Green: angle correct, device readiness
Red: angle incorrect, measurement not possible
Blinking red: measurement error messages

510(k) Summary

6

Image /page/6/Picture/1 description: The image shows the word "icare" in a bold, sans-serif font. The word is black and is positioned on a white background. A thin blue line is located directly underneath the word.

7

ical

iCare IC200 Traditional 510(k)

Changes listed in Table 2 below.

CharacteristicModified device (subject of this 510k)Cleared Device (K190316)
Quick Measure
FeatureDifference: a new Quick Measure feature
added to measure patient's IOP with fewer
rebound measurements and faster
measurement cycle.
-Quick Measure takes two or three rebound
measurements; two if both results are
within 2mmHg and third if the difference
between the first two measurements is
greater than 2mmHg.Not applicable
User interface
colorGraphical User Interface Colors:
UI in Default measurement mode remains
blue
Difference: measurement sequence
information and results in Quick Measure
mode are shown in magenta color to
differentiate from default measurement
modeGraphical User Interface Colors:
Blue color
Alignment error
indicationAlignment error remains enabled in default
measurement mode.
Difference: alignment error is disabled in
Quick Measure mode.Enabled in default measure mode
Device Software
AccessoriesSame during default measurement
Difference: iCare CLINIC/EXPORT is not
usable for Quick Measure resultsiCare CLINIC, iCare EXPORT

Table 2 Differences of the subject device to the predicate device

510(k) Summary

8

| Labelling | Difference: User manual is updated to
include Quick Measure instructions and
applicable warnings.
Minor edits to quick guide and type label. | User manual, quick guide, type label |

--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

1.7 Performance and Safety Data

The device has been tested according to relevant FDA recognized consensus standards. The following performance and safety data are provided in support of the substantial equivalence:

  • ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012 C1:2009/(R)2012 and . A2:2010/(R)2012 Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance
  • ISO 15004-1:2006 Ophthalmic instruments Fundamental requirements and test . methods - Part 1: General requirements
  • IEC 60601-1-6:2010 + A1:2013 Medical electrical equipment Part 1-6: General ● requirements for basic safety and essential performance - Collateral standard: Usability
  • IEC 62304:2006 + A1:2015 Medical device software Software life-cycle . processes (Device firmware and software level of concern: Class B)
  • . ANSI Z80.10:2014 Ophthalmic Instruments - Tonometers

There were no changes related to sterilization, biocompatibility, EMC, or light hazard protection. Device and software risk analysis has been performed in accordance with ISO 14971 risk management standard.

Quick Measure performance data was obtained from a retrospective clinical data analysis in which Quick Measure function was retrospectively compared to the clinical data of iCare IC200. In the analysis Quick Measure met ANSI Z80.10:2014 (in accordance with

510(k) Summary

9

FDA's extent of recognition) performance goals per GAT reference tonometer (see Table 3).

| Group | N (eyes) | Outside ±5 mmHg
vs. GAT reference
(Sitting) | Outside ±5 mmHg
vs. Perkins reference
(Supine) |
|-----------------------------------------|----------|---------------------------------------------------|------------------------------------------------------|
| GAT group = low
≤ 16 mmHg | 44 | 0/44
(0.0 %) | 0/44
(0.0 %) |
| GAT group = medium

16 to 510(k) Summary