K Number
K220852
Device Name
iCare IC200
Manufacturer
Date Cleared
2022-10-06

(197 days)

Product Code
Regulation Number
886.1930
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The iCare IC200 tonometer is intended to be used for the measurement of intraocular pressure of the human eye.

Device Description

The iCare IC200 tonometer (model TA031) is a hand-held, battery-operated device which measures intraocular pressure (IOP) without the need for topical anaesthesia by rebound tonometry. The tonometer is to be used by a healthcare professional. The tonometer uses the rebound method. A small and light, sterile, single-use probe makes brief contact with the eye. The device measures the deceleration of the probe and the rebound time and calculates the IOP from these parameters. Deceleration of the probe is slower at low IOP compared to high IOP. The measurement method, the IOP measurement algorithm and rebound technology (including disposable probe) are identical with the predicate device. iCare IC200 tonometer can obtain IOP measurements in all angles between horizontal and supine directions. External design has not been modified. The user interface menu has been updated to contain a Quick Measure mode. The new Quick Measure feature is used to measure patient's IOP with fewer rebound measurements and faster measurement cycle. Quick Measure takes two or three rebound measurements; two if both results are within 2 mmHg and third if the difference between the first two measurements is greater than 2 mmHg.

AI/ML Overview

The provided document, a 510(k) Premarket Notification review for the iCare IC200 tonometer, outlines the acceptance criteria and the study performed to demonstrate substantial equivalence to a predicate device. This submission focuses on a new feature of the device, the "Quick Measure" mode, rather than the entire device itself, as the underlying technology remains the same as the predicate.

Here's a breakdown of the requested information based on the provided text:

Acceptance Criteria and Reported Device Performance

The acceptance criterion for the Quick Measure mode's performance is its ability to meet the performance goals of the ANSI Z80.10:2014 standard for ophthalmic tonometers, specifically regarding agreement with reference tonometers (GAT and Perkins) within ±5 mmHg. This is presented across different Intraocular Pressure (IOP) subgroups.

Table 1: Acceptance Criteria and Reported Device Performance for iCare IC200 Quick Measure Mode

Group (IOP Range)Acceptance Criteria (Max % Outside ±5 mmHg vs. Reference)Reported Performance (N Eyes: % Outside ±5 mmHg vs. GAT Reference - Sitting)Reported Performance (N Eyes: % Outside ±5 mmHg vs. Perkins Reference - Supine)
Low (≤ 16 mmHg)Not explicitly stated as a percentage, but implied to be low, as 0% achieved.44: 0.0 %44: 0.0 %
Medium (> 16 to < 23 mmHg)Not explicitly stated as a percentage, but implied to be low, as 1.5% and 6.2% achieved.65: 1.5 %65: 6.2 %
High (≥ 23 mmHg)Not explicitly stated as a percentage, but implied to be low, as 2.5% and 0% achieved.40: 2.5 %40: 0.0 %

Note: The document only provides the reported performance in terms of percentage of eyes outside ±5 mmHg. It does not explicitly state a numerical acceptance criterion (e.g., "must be less than X%"). However, meeting "performance goals per GAT reference tonometer (see Table 3)" indicates that these achieved percentages were deemed acceptable.

Study Details

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size (Eyes):
    • GAT Group (Low IOP): 44 eyes
    • GAT Group (Medium IOP): 65 eyes
    • GAT Group (High IOP): 40 eyes
    • Total (from table 3): 149 eyes
  • Data Provenance: The Quick Measure performance data was obtained from a retrospective clinical data analysis. The country of origin is not specified in the document.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • The document does not specify the number or qualifications of experts used to establish the ground truth for the clinical data analysis. The ground truth appears to be established by comparison to recognized reference tonometers (GAT and Perkins).

4. Adjudication Method for the Test Set:

  • The document does not mention any specific adjudication method (e.g., 2+1, 3+1). The study retrospectively compared the Quick Measure function to existing clinical data, which presumably had established reference measurements.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC comparative effectiveness study was not done. This submission is for a tonometer, a device that directly measures a physiological parameter (IOP), not an AI-assisted diagnostic imaging tool that requires human interpretation. The "Quick Measure" is a new mode of the device affecting its measurement sequence, not an AI assisting human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Yes, in effect. The "Quick Measure" mode, like the existing default mode, is an algorithm within the tonometer that calculates IOP. The performance data presented in Table 3 directly assesses the accuracy of this algorithm (the Quick Measure mode) against established reference tonometers. The document also states:
    • "Accuracy of the iCare IC200 quick measure mode was assessed in a bench test using a manometrically controlled artificial cornea."
    • "Repeatability and accuracy testing completed with iCare IC200 quick measure mode demonstrated high agreement with manometric pressure and reference tonometer used in default measurement mode (iCare IC200)."
    • "Reproducibility test completed with iCare IC200 quick measure mode demonstrated high agreement with manometric pressure regardless of the device, operator, or the measurement angle."
      These statements indicate testing of the algorithm's performance independent of human variability beyond simply operating the device.

7. The Type of Ground Truth Used:

  • The ground truth used for the clinical data analysis was comparison to recognized reference tonometers: the Goldmann Applanation Tonometer (GAT) for sitting measurements and the Perkins tonometer for supine measurements. It also mentions comparison to manometric pressure in bench testing for accuracy and reproducibility.

8. The Sample Size for the Training Set:

  • The document does not provide information on a specific training set size. The "Quick Measure" feature is described as a software modification to an existing device, and the method, algorithm, and technology are stated to be "identical with the predicate device." It's possible that the "training" was implicit in the existing device's development or that the "retrospective clinical data analysis" served as the primary validation, rather than a separate, large-scale training phase distinct from the existing device's development.

9. How the Ground Truth for the Training Set was Established:

  • As the document does not specify a distinct training set for the Quick Measure mode, it also does not detail how ground truth for such a set was established. The ground truth for the validation (test set) was established via GAT and Perkins reference tonometers and manometric pressure. The fundamental measurement mechanism and algorithm are inherited from the predicate device, which would have undergone its own rigorous validation (likely with similar ground truth methods) during its initial clearance (K190316).

{0}------------------------------------------------

October 6, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Icare Finland Oy Hannes Hyvönen Regulatory Affairs Manager Ayritie 22 Vantaa. 01510 Finland

Re: K220852

Trade/Device Name: iCare IC200 Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer And Accessories Regulatory Class: Class II Product Code: HKY Dated: September 5, 2022 Received: September 8, 2022

Dear Hannes Hyvönen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices

{1}------------------------------------------------

or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia. Respiratory, ENT & Dental Devices Office of Product Evaluation and Ouality (OPEO) Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220852

Device Name iCare IC200

Indications for Use (Describe)

The iCare IC200 tonometer is intended to be used for the measurement of intraocular pressure of the human eye.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

ical

iCare IC200 Traditional 510(k)

510(k) Summary

1.1 Submitter Name and Address

Icare Finland Oy Äyritie 22 Vantaa, Finland FI-01510

Contact: Hannes Hyvönen Phone: 358 9 8875 1150 Email: regulatory(at)icare-world.com Date prepared: October 5th, 2022

1.2 Device Name

Trade Name:iCare IC200
Type/model:TA031
Common/Usual Name:Tonometer
Classification Name:Tonometer and Accessories
Regulation No:21 CFR 886.1930
Device Regulatory Class:II
Review Panel:Ophthalmic
Product Code:HKY

Premarket Notification 510(k) Number: -

This premarket notification type is Traditional 510(k).

1.3 Predicate Device

Predicate device: iCare IC200 tonometer (type TA031), K190316.

The predicate device has not been subject to a design-related recall.

{4}------------------------------------------------

icare

1.4 Indications for Use / Intended Use

"The iCare IC200 tonometer is intended to be used for the measurement of intraocular pressure of the human eye."

The intended use of the iCare IC200 tonometer remains unchanged by this design modification.

1.5 Device Description

The iCare IC200 tonometer (model TA031) is a hand-held, battery-operated device which measures intraocular pressure (IOP) without the need for topical anaesthesia by rebound tonometry. The tonometer is to be used by a healthcare professional.

The tonometer uses the rebound method. A small and light, sterile, single-use probe makes brief contact with the eye. The device measures the deceleration of the probe and the rebound time and calculates the IOP from these parameters. Deceleration of the probe is slower at low IOP compared to high IOP. The measurement method, the IOP measurement algorithm and rebound technology (including disposable probe) are identical with the predicate device.

iCare IC200 tonometer can obtain IOP measurements in all angles between horizontal and supine directions. External design has not been modified. The user interface menu has been updated to contain a Quick Measure mode. The new Quick Measure feature is used to measure patient's IOP with fewer rebound measurements and faster measurement cycle. Quick Measure takes two or three rebound measurements; two if both results are within 2 mmHg and third if the difference between the first two measurements is greater than 2 mmHg.

510(k) Summary

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "icare" in a bold, sans-serif font. The letters are black, and there is a thin blue line underneath the word. The word appears to be a logo or brand name.

1.6 Comparison of Technological Characteristics with the Predicate Device

Similarities of the subject device and the predicate device is presented in the Table 1 below:

Table 1 Similarities of the subject device and the predicate device
-------------------------------------------------------------------------------
CharacteristicModified device (subject of this 510k)Cleared Device (K190316)
ProductAppearance andmaterialsImage: Modified deviceImage: Cleared device
SameSame
Product/DeviceIdentificationSameiCare IC200 Tonometer(Type: TA031)
Intended Use /Indications forUse StatementSameIOP Measurement“The iCare IC200 tonometer is intendedto be used for the measurement ofintraocular pressure of the human eye.”
Intended usersSameHealthcare professionals
MeasurementmethodSameRebound tonometry
MeasurementrangeSame7-50 mmHg
DefaultmeasurementsequenceSameDefault measurement mode: 6measurements, calculation of median IOPand standard deviation based on 4measurements (highest and lowestexcluded)
Versatility ofMeasurementpositionSameTonometer can be used in any anglebetween 0° (sitting, standing) and 90°(patient in supine position)
Device DisplaySame1.50" OLED display
Automatic eyerecognitionsystemSameNot applicable
DesignSameHandheld microprocessor based
CalibrationSameNo maintenance calibration required
Contact tip(probe)SameLightweight, disposable, single use,plastic probe(TP01s)
Contact tipsterilizationSameGamma-sterilized
AnaesthesiarequiredSameNo
Power supplySame4 x 1,5V AA Alkaline LR6 batteries
Devicedimensions andweightSameDimensions:43mm x 104mm x 214mmWeight: 165g (without batteries), 267g(with batteries)
ConnectivityinterfaceSameBluetooth (Microchip RN4678 Module)
ElectronicsSameInclination sensor, External Flashmemory, Real time clock
MechanicsSameAs described in K190316
User interfaceAudio indications:SameAudio indications:Beeps for device too near or too farsituation, and for other errors.
Graphical User Interface:SameGraphical User Interface:OLED display
Probe base light:SameProbe base light:Green: angle correct, device readiness
Red: angle incorrect, measurement not possibleBlinking red: measurement error messages

510(k) Summary

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the word "icare" in a bold, sans-serif font. The word is black and is positioned on a white background. A thin blue line is located directly underneath the word.

{7}------------------------------------------------

ical

iCare IC200 Traditional 510(k)

Changes listed in Table 2 below.

CharacteristicModified device (subject of this 510k)Cleared Device (K190316)
Quick MeasureFeatureDifference: a new Quick Measure featureadded to measure patient's IOP with fewerrebound measurements and fastermeasurement cycle.-Quick Measure takes two or three reboundmeasurements; two if both results arewithin 2mmHg and third if the differencebetween the first two measurements isgreater than 2mmHg.Not applicable
User interfacecolorGraphical User Interface Colors:UI in Default measurement mode remainsblueDifference: measurement sequenceinformation and results in Quick Measuremode are shown in magenta color todifferentiate from default measurementmodeGraphical User Interface Colors:Blue color
Alignment errorindicationAlignment error remains enabled in defaultmeasurement mode.Difference: alignment error is disabled inQuick Measure mode.Enabled in default measure mode
Device SoftwareAccessoriesSame during default measurementDifference: iCare CLINIC/EXPORT is notusable for Quick Measure resultsiCare CLINIC, iCare EXPORT

Table 2 Differences of the subject device to the predicate device

510(k) Summary

{8}------------------------------------------------

LabellingDifference: User manual is updated toinclude Quick Measure instructions andapplicable warnings.Minor edits to quick guide and type label.User manual, quick guide, type label
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

1.7 Performance and Safety Data

The device has been tested according to relevant FDA recognized consensus standards. The following performance and safety data are provided in support of the substantial equivalence:

  • ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012 C1:2009/(R)2012 and . A2:2010/(R)2012 Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance
  • ISO 15004-1:2006 Ophthalmic instruments Fundamental requirements and test . methods - Part 1: General requirements
  • IEC 60601-1-6:2010 + A1:2013 Medical electrical equipment Part 1-6: General ● requirements for basic safety and essential performance - Collateral standard: Usability
  • IEC 62304:2006 + A1:2015 Medical device software Software life-cycle . processes (Device firmware and software level of concern: Class B)
  • . ANSI Z80.10:2014 Ophthalmic Instruments - Tonometers

There were no changes related to sterilization, biocompatibility, EMC, or light hazard protection. Device and software risk analysis has been performed in accordance with ISO 14971 risk management standard.

Quick Measure performance data was obtained from a retrospective clinical data analysis in which Quick Measure function was retrospectively compared to the clinical data of iCare IC200. In the analysis Quick Measure met ANSI Z80.10:2014 (in accordance with

510(k) Summary

{9}------------------------------------------------

FDA's extent of recognition) performance goals per GAT reference tonometer (see Table 3).

GroupN (eyes)Outside ±5 mmHgvs. GAT reference(Sitting)Outside ±5 mmHgvs. Perkins reference(Supine)
GAT group = low≤ 16 mmHg440/44(0.0 %)0/44(0.0 %)
GAT group = medium> 16 to < 23 mmHg651/65(1.5 %)4/65(6.2 %)
GAT group = high≤ 23 mmHg401/40(2.5 %)0/40(0.0 %)

Table 3 Comparability Chart between IC200 Quick Measure and GAT, and Perkins, within IOP subgroups

Accuracy of the iCare IC200 quick measure mode was assessed in a bench test using a manometrically controlled artificial cornea. Testing compared the new Quick Measure mode against the existing default measurement mode (same as in predicate device IC200), clinically validated subnormal reference tonometer (iCare IC200), and manometric pressure.

Repeatability and accuracy testing completed with iCare IC200 quick measure mode demonstrated high agreement with manometric pressure and reference tonometer used in default measurement mode (iCare IC200).

Reproducibility test completed with iCare IC200 quick measure mode demonstrated high agreement with manometric pressure regardless of the device, operator, or the measurement angle.

{10}------------------------------------------------

Software Verification and Validation

Testing Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern, since a failure or latent flaw in the software could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider.

1.8 Substantial Equivalence

The modified iCare IC200 including the Quick measure mode is substantially equivalent to the predicate device. The devices have the same or substantially equivalent technological characteristics and the iCare IC200 raises no new issues of safety and effectiveness. Performance and safety data demonstrate that the modified iCare IC200 is as safe and effective as the predicate device.

510(k) Summary

§ 886.1930 Tonometer and accessories.

(a)
Identification. A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.(b)
Classification. Class II.