K Number
K220739

Validate with FDA (Live)

Date Cleared
2022-11-17

(248 days)

Product Code
Regulation Number
866.2390
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Disposable Urine Collection Tube is intended for the collection, transport and preservation of urine specimens from the collection site to the testing laboratory, urine specimens are processed using standard clinical laboratory operating procedures for the cultivation of uropathogenic bacteria and yeasts.

Device Description

The Disposable Urine Collection Tubes for the collection, transport and preservation system consist of a safety rubber plug, a plastic cap and boric acid as an additive inside the tube. The tube is made of Polyethylene terephthalate (PET) and the safety plug is made of butvl-rubber while the disposable cap is made of Polyethylene (PE). The disposable tube device sustains the viability of organisms during transport for up to 48 hrs. at 2-4°C and 22-25°C.There are two main tube types (round and conical bottom), and four configurations by volume of urine for the round bottom tubes (4 ml, 7 ml, 8 ml, 10 ml). The conical bottom tubes only have one configuration by volume (8 ml). Each volume is available in yellow and white caps.

AI/ML Overview

This document describes the premarket notification (510(k)) for the Disposable Urine Collection Tube, which is a Class I medical device. The information provided outlines the device's intended use, comparison to a predicate device, and performance data from analytical studies.

Here's an analysis of the provided text in relation to the requested information:

1. A table of acceptance criteria and the reported device performance

The document details performance studies, primarily for microbial recovery and fill volume impact.

Acceptance Criteria (from CLSI document M40-A2 and internal study criteria)Reported Device Performance (from Tables 2 & 3)
Microbial Recovery: Within 1 log10 of the initial microorganism count.All strains tested (Candida albicans, Escherichia coli, Enterococcus Faecalis, Pseudomonas aeruginosa, Proteus mirabilis, Staphylococcus saprophyticus, Citrobacter freundii, Candida glabrata, Enterobacter cloacae, Morganella morganii, Streptococcus agalactiae) showed a change in Log10 (0 - 48 hrs) within the range of -0.17 to -0.51 at both 2-4°C and 22-25°C. All reported values are within the 1 log10 acceptance criterion.
Fill Volume Impact Recovery: Less than 1 log10 change when the fill volume is above or below the nominal line.For E. coli, P. aeruginosa, and P. mirabilis at 90% and 110% fill volumes, the change in Log10 (0 - 48 hrs) ranged from -0.42 to -0.55. All reported values are within the 1 log10 acceptance criterion.
Shelf Life:**
Appearance, Vacuum, Leak testing, Strength test, 50Kpa Pressure Test, PET tube's mechanical stress testing, Sterility test, Standards of boric acid content, Package integrity must pass over 18 months.All samples passed the real-time aging test for 18 months at 22-25°C for all listed criteria.

2. Sample size used for the test set and the data provenance

  • Microbial Recovery Studies:
    • Sample Size: "Three lots of the Disposable Urine Collection Tubes with three samples per lot were evaluated" (3 lots * 3 samples/lot = 9 total samples per condition/time point). The table indicates N=9 for average CFU/ml recovered.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The studies used "pooled human negative clinical urine matrix." The study is presented as part of a 510(k) submission to the US FDA by a Chinese manufacturer. It is a prospective analytical study comparing the device's performance over time.
  • Fill Volume Impact Recovery Study:
    • Sample Size: "Three production lots were tested in three replicates with each urine samples spiked with E. coli, P. aeruginosa or P. mirabilis" (3 lots * 3 replicates/lot = 9 total samples per microorganism/fill volume/time point). The table indicates N=9 for average CFU/ml recovered.
    • Data Provenance: Similar to microbial recovery, not explicitly stated beyond being an analytical study supporting the device's performance from a Chinese manufacturer.
  • Shelf Life Study:
    • Sample Size: "Six lots of varying production were used for the studies." Specific number of samples per lot or per test is not detailed beyond that.
    • Data Provenance: Not explicitly stated. This is a prospective real-time aging study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This device is a urine collection, transport, and preservation tube, not an AI or diagnostic imaging device that requires expert interpretation of output. The "ground truth" for its performance studies is established by quantitative microbiological methods (e.g., counting colony-forming units (CFU/ml) and comparing them to initial counts), and physical/chemical tests (e.g., vacuum, leak, strength, borid acid content). Therefore, the concept of "experts" in the context of radiologists or similar interpretive professionals establishing ground truth is not applicable here. The ground truth is objective, laboratory-derived data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is an analytical device where performance is measured objectively through laboratory tests (microbial culture, physical properties) rather than through subjective human interpretation or consensus. No adjudication method for "ground truth" is needed in this context.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device for specimen collection and transport, not an AI-assisted diagnostic tool. No MRMC study or AI assistance evaluation was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth for this device's performance is:

  • Microbiological Culture Counts (CFU/ml): For microbial recovery and fill volume impact studies, the "ground truth" is the quantitative measurement of viable microorganisms at different time points and conditions. This is an objective, laboratory-based measurement.
  • Physical and Chemical Test Results: For shelf-life studies, the "ground truth" is determined by objective measurements of vacuum, leak integrity, mechanical strength, sterility, and boric acid content.

8. The sample size for the training set

Not applicable. This is a physical medical device, not a machine learning model, so there is no "training set."

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

November 17, 2022

Zhejiang Gongdong Medical Technology Co., Ltd. % Evan Hu Marketing & Technical Manager Shanghai Mind-link Consulting Co., Ltd. 1399 Jiangyue Road, Minhang Shanghai, Shanghai 201114 China

Re: K220739

Trade/Device Name: Disposable Urine Collection Tube Regulation Number: 21 CFR 866.2390 Regulation Name: Transport Culture Medium Regulatory Class: Class I, reserved Product Code: JSM, LIO Dated: March 7, 2022 Received: March 14, 2022

Dear Evan Hu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

Please note that if you modify your IVD in the future to exceed any of the limitations to the exemption found in 21 CFR 866.9(c), your device will require a new 510(k) prior to marketing this device in the United States and will not be exempt from the premarket notification requirements so long as it exceeds the limitations to the exemption found in 21 CFR 866.9.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Ribhi Shawar -S

Ribhi Shawar, Ph.D. (ABMM) Branch Chief Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220739

Device Name Disposable Urine Collection Tube

Indications for Use (Describe)

Disposable Urine Collection Tube is intended for the collection, transport and preservation of urine specimens from the collection site to the testing laboratory, urine specimens are processed using standard clinical laboratory operating procedures for the cultivation of uropathogenic bacteria and yeasts.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

I. SUBMITTER:

Zhejiang Gongdong Medical Technology Co., Ltd.

No.10 Beiyuan Ave., Huangyan, Taizhou, Zhejiang, 318020, China

Contact Person: Handson Wei

Title: Sales Director

Tel: 86-0576-84115678

Email: handson(@chinagongdong.com

Submission Correspondent: Evan Hu

Email: Evan.hu@mind-link.net Tel:86-18616124827

Date Prepared: November 9, 2022

II. DEVICE

Name of Device: Disposable Urine Collection Tube Classification Name: Culture Media, Non-Propagating Transport FDA Panel: Microbiology Classification Number: 21 CFR 866.2390, Transport Culture Medium Regulatory Class: Class I Product Code: JSM, LIO

III. PREDICATE DEVICE

Primary predicate device: UriSwab™ - Urine Collection, Transport and Preservation System (K180052) Classification Name: Culture Media, Non-Propagating Transport FDA Panel: Microbiology Classification Number: 21 CFR 866.2390, Transport Culture Medium Regulatory Class: Class I Product Code: JSM, LIO

{4}------------------------------------------------

IV. DEVICE DESCRIPTION

The Disposable Urine Collection Tubes for the collection, transport and preservation system consist of a safety rubber plug, a plastic cap and boric acid as an additive inside the tube. The tube is made of Polyethylene terephthalate (PET) and the safety plug is made of butvl-rubber while the disposable cap is made of Polyethylene (PE). The disposable tube device sustains the viability of organisms during transport for up to 48 hrs. at 2-4°C and 22-25°C.There are two main tube types (round and conical bottom), and four configurations by volume of urine for the round bottom tubes (4 ml, 7 ml, 8 ml, 10 ml). The conical bottom tubes only have one configuration by volume (8 ml). Each volume is available in yellow and white caps. The different device configurations and information on the amount of preservative contained in each tube is provided in table 1.

Table 1. Description and Packaging configurations of Disposable Urine Collection Tube

REF NO.Draw volume (mL)Tube size (Diameter × Length) in mmAdditive Boric acid in mgMaterialType of tubeCap colour
GD040UB413×7540PETRound bottomyellow
GD040UBW413×7540PETRound bottomwhite
GD070UB713×10070PETRound bottomyellow
GD070UBW713×10070PETRound bottomwhite
GD080UB816×10080PETRound bottomyellow
GD080UBW816×10080PETRound bottomwhite
GD100UB1016×100100PETRound bottomyellow
GD100UBW1016×100100PETRound bottomwhite
GD080UAC816×10080PETConical bottomyellow

{5}------------------------------------------------

GD080UACW816×10080PETConicalwhite
bottom

V. INDICATIONS FOR USE/INTENDED USE

Disposable Urine Collection Tube is intended for the collection, transport and preservation of urine specimens from the collection site to the testing laboratory. In the laboratory, urine specimens are processed using clinical laboratory operating procedures for the cultivation of uropathogenic bacteria and yeasts.

PRINCIPLE OF OPERATION VI.

The Disposable Urine Collection Tube system is used to safely collect and transport urine from collection sites to the testing laboratories. It is intended to be used by health care professionals. The vacuum in the collection tube allows autofill the urine sample through the puncture needle and the boric acid in the tubes prevents overgrowth of uropathogenic bacteria and yeasts.

COMPARISON WITH THE PREDICATE DEVICE VII.

The device, Disposable Urine Collection Tube is compared with the predicate device UriSwab - Urine Collection, Transport and Preservation System (K180052). The results are shown below in the comparison table:

Device & PredicateDevice(s):Device: K220739Predicate: K180052
Device Trade NameDisposable UrineCollection TubeUriSwab-UrineCollection, Transportand PreservationSystem
Device Product Code andClassificationJSM, LIO, Class ISame
General DeviceCharacteristic Similarities
Indications For UseDisposable UrineCollection Tube isintended for thecollection, transport,and preservation ofCopan UriSwab™ -Urine Collection,Transport andPreservation System isintended for the
urine specimens fromthe collection site to thetesting laboratory. In thelaboratory, urinespecimens areprocessed usingstandard clinicallaboratory operatingprocedures for thecultivation ofuropathogenic bacteriaand yeasts.collection, transport,and preservation ofurine specimens fromthe collection site to thetesting laboratory. In thelaboratory, UriSwab™specimens areprocessed usingstandard clinicallaboratory operatingprocedures for thecultivation ofuropathogenic bacteriaand yeasts.
Specimen Stability2-4°C and 22-25°C;48 hrs.Same
Device Storage Temperature22-25°CSame
SterilizationRadiationSame
SterileYes; 10-6 SALSame
General DeviceCharacteristic Differences
Tube materialPolyethyleneTerephthalate (PET)Plastic
Tube typeRound and ConicalbottomConical bottom
Tube size13 mm × 75 mm;13 mm x 100 mm;16 mm × 100 mm12 mm × 80 mm;16 mm × 100 mm
Tube volume4 ml, 7 ml, 8 ml, 10 ml1.5 ml, 3.2 ml
CapSafety cap, yellow andwhite.Screw cap, yellow

{6}------------------------------------------------

{7}------------------------------------------------

AdditiveBoric acidBoric acid, Sodiumformate
Collection devicePET tube with additivePlastic tube, stick withsponge
Tube fillingPredefined vacuumSponge applicator

VIII. PERFORMANCE DATA

Analytical Performance

Microbial Recovery Studies:

Microbial recovery (viability) studies were conducted to support the claim of the Disposable Urine Collection Tube for the collection, transport and preservation of urine specimens from the collection site to the testing laboratory. Sample panels were prepared from the listed bacteria and yeast (below) that were commonly associated with urinary tract infections. Cultures from 1.5 × 10° (equivalent 0.5 McFarland standard) were diluted into the pooled human negative clinical urine matrix to give a final concentration of 1.5 x 102 CFU/ml. The appropriate volume of urine samples spiked with each organism was added to the Disposable Urine Collection Tubes. Three lots of the Disposable Urine Collection Tubes with three samples per lot were evaluated for the microbial recovery (viability) assay. The inoculated sample tubes were stored at 2-4ºC and 22-25ºC for 24, and 48 hrs. After each storage time and temperature, the samples were plated to grow colonies which were counted to determine CFU/ml for each time points. According to CLSI document M40-A2, the acceptance criteria for microbial recovery method was set to be within 1 log10 of the initial microorganism count. Results of recovery studies are presented in table 2.

Strain testedStoragetemperatureAverage CFU/ml recovered(N=9) at time pointsChange inLog10(0 - 48 hrs)
0 hr24 hrs48 hrs
Candida albicans(ATCC 24433)2-4°C14010785-0.22
22-25°C1407469-0.31
Escherichia coli(ATCC 25922)2-4°C13710492-0.17
22-25°C1378542-0.51
Enterococcus Faecalis(ATCC 29212)2-4°C14510291-0.20
22-25°C1458054-0.43
Pseudomonas aeruginosa(ATCC 27853)2-4°C14310369-0.32
22-25°C14310672-0.30

|--|

{8}------------------------------------------------

Proteus mirabilis2-4°C14011172-0.29
(ATCC 7002)22-25°C14010571-0.29
Staphylococcus saprophyticus2-4°C13910969-0.30
(ATCC 15305)22-25°C13910866-0.32
Citrobacter freundii2-4°C14010473-0.28
(ATCC 8090)22-25°C14010573-0.28
Candida glabrata2-4°C13711177-0.25
(ATCC MYA-2950)22-25°C13710277-0.25
Enterobacter cloacae2-4°C13810767-0.31
(ATCC 13047)22-25°C1388169-0.30
Morganella morganii2-4°C13911169-0.30
(ATCC 25829)22-25°C13911053-0.42
Streptococcus agalactiae2-4°C14311269-0.32
(ATCC 12386)22-25°C14311271-0.30

Fill Volume Impact Recovery Study:

A fill volume impact recovery study was performed to demonstrate that underfilling of Disposable Urine Collection Tube containing preservative does not negatively impact the viability of urine microorganism. Sample panels were prepared by listed microorganisms (below) diluted from 1.5 x 108 CFU/mL (equivalent to a 0.5 McFarland standard) into the pooled human negative clinical urine matrix to give a final concentration of 1.5 x 102 CFU/ml. Three production lots were tested in three replicates with each urine samples spiked with E. coli, P. aeruginosa or P. mirabilis at the sample fill volume of 90% and 110%. The inoculated sample tubes were stored at 22-25°C for 0 and 48 hrs. After each storage time, the samples were plated to grow colonies which were counted to determine CFU/ml for each time points. The acceptance criterium for the study is less than 1 log10 change when the fill volume is above or below the line. The results were shown in table 3.

Table 3. Performance of Disposable Urine Collection Tube in fill volume impact
recovery study.
Strain testedFill volumeAverage CFU/ml recovered(N=9) at time pointsChange in Log10(0 - 48 hrs)
0 hr48 hrs
E. coli(ATCC 25922)90%12548-0.42
E. coli(ATCC 25922)110%12746-0.44
P. aeruginosa(ATCC 27853)90%14349-0.47
P. aeruginosa(ATCC 27853)110%14250-0.45

{9}------------------------------------------------

P. mirabilis(ATCC 7002)90%14040-0.54
110%13939-0.55

IX. STERILIZATION

The Disposable Urine Collection Tubes were sterilized by electron beam radiation with an average dose of 20 kGy to achieve a sterility assurance level (SAL) of 10 ° with a dosimeter attached to monitor dosage through the sterilization process. Sterilization was validated following ISO 11137-1: 2006 and ISO 11137-2: 2013 Sterilization of Health Care Products Radiation guidelines. Sterility test was carried out according to ISO11137-2:2009 Medical Devices Radiation Sterilization-Microbiological Methods-Part 2. In this sterility test, 100 samples after irradiated at the verification dose were inoculated with 200 ml soybean casein liquid medium and incubated at 30±2℃ for 14 days. The results were observed with no growth.

X. SHELF LIFE

The shelf life for the Disposable Urine Collection Tube was determined to be 18 months from the date of manufacture when stored in well-ventilated clean room with the environment of relative humidity not more than 80% and at temperature 22 -25°C. The shelf life of the Disposable Urine Collection Tube was evaluated using real-time aging performance test at time points T = 0, 6, 12 and 18 months. At each time point, following performance test were conducted:

Appearance, Vacuum, Leak testing, Strength test, 50Kpa Pressure Test, PET tube's mechanical stress testing, Sterility test, Standards of boric acid content, Package integrity

Appearance test was carried out by inspecting visually for clear label, transparent PET tube and any sign of deformation or breakage. Vacuum test was performed by measuring volume drawn into the Disposable Urine Collection Tube between 90 to 110% of the nominal liquid capacity allowed. Physical and mechanical performance tests included leak test, strength test, and a 50 KPa pressure test. Mechanical stress test was conducted by checking the tubes could withstand centrifugation. The strength test, and urine collection volume capacity test were carried out as part of the evaluation. Sterility test and the standard of boric acid content (measuring the boric acid content in the tube) used were also assessed. The package integrity was evaluated by visually checking for any changes or damage to the package. Six lots of varying production were used for the studies and results showed that all samples passed the real time aging test for 18 months at 22-25°C.

{10}------------------------------------------------

The results for shelf-life study support that the Disposable Urine Collection Tube is physically or visually stable for 18 months.

§ 866.2390 Transport culture medium.

(a)
Identification. A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.(b)
Classification. Class I (general controls).