K Number
K220654
Manufacturer
Date Cleared
2022-04-06

(30 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ANGIOGUARD XP and ANGIOGUARD RX Emboli Capture Guidewire devices are indicated for use as a guidewire and embolic protection system to contain and remove embolic material (thrombus/debris) while performing carotid artery angioplasty and stenting procedures in carotid arteries. The diameter of the site of filter basket placement should be from 3mm to 7.5mm.

Device Description

Both the subject and predicate ANGIOGUARD XP Emboli Capture Guidewire and ANGIOGUARD RX Emboli Capture Guidewire devices consists of a guidewire with integrated emboli filter basket at the distal end. The devices function as an interventional guidewire and distal protection device during delivery and placement of stents and interventional devices in carotid procedures. The guidewire is delivered via an OTW (overthe-wire) or RX (rapid-exchange) deployment sheath and is captured via an OTW or RX capture sheath. The ANGIOGUARD XP Emboli Capture Guidewire and ANGIOGUARD RX Emboli Capture Guidewire devices have a filter basket at the distal end that is deployed prior to the stenting procedure. When deployed, the filter basket opens in an umbrella-like fashion, allowing passive hemo-filtration with subsequent emboli capture. At the end of the procedure, the filter is collapsed and retrieved.

AI/ML Overview

The provided document is a 510(k) premarket notification decision letter from the FDA regarding the ANGIOGUARD XP Emboli Capture Guidewire and ANGIOGUARD RX Emboli Capture Guidewire. This document determines substantial equivalence to a predicate device, but it does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of device performance related to a diagnostic or therapeutic algorithm.

Instead, it focuses on the substantial equivalence of an updated version of a medical device (the ANGIOGUARD XP/RX Emboli Capture Guidewire) to a previously cleared predicate device. The changes are limited to a material change for the Deployment and Capture Sheath components.

The "Performance Data" section (Page 4-5) lists the types of tests performed to demonstrate that the new material meets the same requirements as the previous material and that the device's performance and functionality remain substantially equivalent. However, these are not "acceptance criteria" in the sense of performance metrics for an AI algorithm or a diagnostic test with corresponding study results. They are engineering and biocompatibility tests to ensure the safety and effectiveness of the physical device after a material change.

Therefore, for the requested information:

  1. A table of acceptance criteria and the reported device performance: Not available in the provided document in the context of a diagnostic or therapeutic performance study. The document focuses on showing that a material change to an existing device maintains equivalence to its predicate. The "performance data" listed are physical and biological tests, not performance metrics like sensitivity, specificity, etc., with associated target levels and results.

  2. Sample size used for the test set and the data provenance: Not applicable. There is no "test set" in the context of an algorithm or diagnostic study. The tests mentioned (e.g., biocompatibility) involve laboratory samples or animal models, not human patient data in the way a diagnostic study would.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No ground truth establishment for a diagnostic test is described.

  4. Adjudication method: Not applicable.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable. This is a physical device, not an AI or diagnostic algorithm improving human reader performance.

  6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable.

  7. The type of ground truth used: Not applicable.

  8. The sample size for the training set: Not applicable. There is no training set mentioned, as this is a physical device submission, not an AI algorithm submission.

  9. How the ground truth for the training set was established: Not applicable.

In summary, the provided FDA letter confirms the substantial equivalence of a physical medical device following a material change. It does not describe a study involving acceptance criteria for diagnostic/therapeutic performance, nor does it involve AI or algorithmic performance evaluation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 6, 2022

Dunia Bram Principal Specialist, Regulatory Affairs 14201 N.W. 60th Avenue Miami Lakes, Florida 33014

Re: K220654

Trade/Device Name: ANGIOGUARD XP Emboli Capture Guidewire, ANGIOGUARD RX Emboli Capture Guidewire Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NTE Dated: March 4, 2022 Received: March 7, 2022

Dear Dunia Bram:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Finn Donaldson Acting Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220654

Device Name ANGIOGUARD XP and ANGIOGUARD RX Emboli Capture Guidewire

Indications for Use (Describe)

ANGIOGUARD XP and ANGIOGUARD RX Emboli Capture Guidewire devices are indicated for use as a guidewire and embolic protection system to contain and remove embolic material (thrombus/debris) while performing carotid artery angioplasty and stenting procedures in carotid arteries. The diameter of the site of filter basket placement should be from 3mm to 7.5mm.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SUBMITTER I.

Cordis US Corporation 14201 N.W. 60th AVE. Miami Lakes, FL 33014, USA Phone: (408) 273-3423 Fax: (408) 955-0704 Establishment registration number: 1016427

Contact: Dunia Bram Cordis US Corp. Date Prepared: March 4, 2022

II. DEVICE

Trade Name: ANGIOGUARD® XP Emboli Capture Guidewire ANGIOGUARD® RX Emboli Capture Guidewire

Common Name: Embolic Protection Guidewire

Classification Name: Cardiovascular Percutaneous Catheter and Catheter Guidewire

Device Classification: 21 CFR §870.1250 and 21 CFR §870.1330

Regulatory Class: Class II

Product Code: NTE - temporary carotid catheter for embolic capture

III. PREDICATE DEVICE

ANGIOGUARD XP Emboli Capture Guidewire and ANGIOGUARD RX Emboli Capture Guidewire, previously cleared under K101651 on July 9, 2010, and under K062531 on September 22, 2006.

IV. DEVICE DESCRIPTION

Both the subject and predicate ANGIOGUARD XP Emboli Capture Guidewire and ANGIOGUARD RX Emboli Capture Guidewire devices consists of a guidewire with integrated emboli filter basket at the distal end. The devices function as an interventional guidewire and distal protection device during delivery and placement of stents and interventional devices in carotid procedures. The guidewire is delivered via an OTW (overthe-wire) or RX (rapid-exchange) deployment sheath and is captured via an OTW or RX capture sheath. The ANGIOGUARD XP Emboli Capture Guidewire and ANGIOGUARD RX Emboli Capture Guidewire devices have a filter basket at the distal end that is deployed prior to the stenting procedure. When deployed, the filter basket opens in an umbrella-like

{4}------------------------------------------------

fashion, allowing passive hemo-filtration with subsequent emboli capture. At the end of the procedure, the filter is collapsed and retrieved.

V. INDICATIONS FOR USE

ANGIOGUARD XP Emboli Capture Guidewire and ANGIOGUARD RX Emboli Capture Guidewire devices are indicated for use as a guidewire and embolic protection system to contain and remove embolic material (thrombus/debris) while performing carotid artery angioplasty and stenting procedures in carotid arteries. The diameter of the artery at the site of filter basket placement should be from 3mm to 7.5mm.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Both the subject and predicate ANGIOGUARD XP Emboli Capture Guidewire and ANGIOGUARD RX Emboli Capture Guidewire are identical in its basic design, intended use, indications for use statement, contraindications, mechanism of action, operating principle, sterilization method and Sterility Assurance Level (SAL). The changes to the subject device relative to the predicate are limited to a material change for the Deployment and Capture Sheath components.

With the exception of the Deployment Sheath and Capture Sheath material change, the subject and predicate devices share the same materials, design, dimensions, size range, intended use, principle of operation, mechanism of action, method of sterilization and sterility assurance level, biocompatibility classification, labeling, packaging materials and configuration, and shelf life. The new material for the Deployment Sheath and Capture Sheath meets the same requirements as the current material.

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination:

Biocompatibility Testing

Biocompatibility testing was performed on finished and sterilized ANGIOGUARD XP and ANGIOGUARD RX Emboli Capture Guidewire in compliance with U.S. Food and Drug Administration Good Laboratory Practice (GLP) regulations set forth in 21 CFR Part 58 and per International Standard ISO 10993-1:2018 and FDA guidance: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", September 2020.

Biocompatibility testing performed included the following:

  • Chemical Characterization
  • Cytotoxicity MEM Elution ●
  • Sensitization Guinea Pig Maximization ●
  • Intracutaneous Irritation Reactivity ●
  • Acute Systemic Toxicity

{5}------------------------------------------------

  • Material Mediated Pyrogenicity
  • Hemocompatibility
    • Hemolysis Extract & Direct Contact O
    • O Platelet and Leukocyte Count
    • Partial Thromboplastin Time O
    • Complete Activation Assay O
    • Thrombin Antithrombin Assay (TAT) O
    • In Vitro Blood Loop Assay O

Bench Testing

  • Visual inspection ●
  • Dimensional verification
  • Deployment and Capture Sheath Tensile Test
  • Material evaluation ●
    • o Fourier Transform Infrared Spectroscopy (FTIR)
    • Differential Scanning Calorimetry (DSC) o
    • Scanning Electron Microscopy (SEM) / Elemental Analysis (EDAX) O

VIII. CONCLUSIONS

The subject ANGIOGUARD XP Emboli Capture Guidewire and ANGIOGUARD RX Emboli Capture Guidewire are the same in basic design and have the same intended use as the legally marketed predicate, ANGIOGUARD XP Emboli Capture Guidewire and ANGIOGUARD RX Emboli Capture Guidewire. The modifications made to the Deployment and Capture Sheath material do not alter the fundamental scientific technology of the device, the device's operating principles, mechanism of action, intended use, or the indication for use of the device. The material modification made to the ANGIOGUARD XP Emboli Capture Guidewire and ANGIOGUARD RX Emboli Capture Guidewire were verified and validated through a series of tests ensuring that the subject guidewire meets all specifications and that the performance and functionality are substantially equivalent to the predicate device. The ANGIOGUARD XP Emboli Capture Guidewire and ANGIOGUARD RX Emboli Capture Guidewire continues to meet all previous performance specifications and none of the critical clinical performance parameters have changed. The modifications do not raise new questions of safety and effectiveness. ANGIOGUARD XP Emboli Capture Guidewire and ANGIOGUARD RX Emboli Capture Guidewire can be used according to its intended use and in an equivalent manner to the predicate device. The subject ANGIOGUARD XP Emboli Capture Guidewire and ANGIOGUARD RX Emboli Capture Guidewire devices are substantially equivalent to the predicate ANGIOGUARD XP Emboli Capture Guidewire and ANGIOGUARD RX Emboli Capture Guidewire devices.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).