K Number
K220601
Device Name
CoolTone
Date Cleared
2022-04-01

(30 days)

Product Code
Regulation Number
890.5850
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CoolTone is indicated to be used for:

  • Improvement of abdominal tone, strengthening of the abdominal muscles, development for firmer abdomen.
  • Strengthening, toning and firming of buttocks and thighs.
Device Description

The CoolTone system is a non-invasive, therapeutic device. The device produces an electromagnetic field that interacts with the tissues of the human body. By muscle stimulation, the CoolTone helps to strengthen, tone and firm the abdomen, buttocks and thighs.
The device is a mobile, standalone unit with four wheels and the control unit housing protects the patient from electrical shock and mechanical injury.
Two large applicators are connected to the control unit and can be used simultaneously depending on the treatment. CoolTone is a medical device that generates a magnetic field by applying a strong current to an applicator. It is equipped with the securement system, designed to maintain the position of the applicator throughout treatment.
A large color touch screen facilitates the use of the device. The on-screen information guides the user, step-by-step, through the entire treatment process. The treatment is operated through variable parameters such as frequency, time and intensity. Three, pre-set treatment options are available for users to choose from: Abdomen, buttocks and thighs.

AI/ML Overview

This document is a 510(k) Premarket Notification for the CoolTone device. It states that the device is substantially equivalent to a previously cleared predicate device (K192940). As such, the purpose of this submission is not to prove new efficacy or safety but rather to demonstrate that modifications made to the device do not change its substantial equivalence to the predicate. Therefore, this document does not contain a new clinical study from which to extract acceptance criteria and performance data in the typical sense of a de novo device or a device with new indications.

The relevant sections to address your request will derive from the comparison to the predicate device and the performance testing conducted to show the current device continues to meet the predicate's performance.

Based on the provided text, here's an attempt to answer your questions, but many points will indicate that a new study to prove the device meets acceptance criteria was not required for this 510(k) submission, as it's establishing substantial equivalence to an existing cleared device with minor modifications.


Here's the analysis of the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of "acceptance criteria" for a new clinical study. Instead, it demonstrates that the modified device maintains the same performance characteristics as the predicate device (CoolTone K192940). The performance data section primarily focuses on compliance with voluntary standards and the effects of the engineering changes.

Acceptance Criterion (Implicitly referring to predicate device performance)Reported Device Performance (Subject CoolTone)
Magnetic Field Intensity: 0.5-1.35T ± 20%0.5 – 1.35T +/- 20% The System Level Testing confirmed that the two applicators, either operated separately or simultaneously, performed within this range.
Tissue Temperature Rise: No appreciable rise in temperature at maximum intensity to cause patient risk.The System Level Testing confirmed that "the tissue being treated by the device is not subjected to an appreciable rise in temperature at maximum intensity to cause a risk to the patient."
Electrical Safety: Compliance with ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012, C1:2009/(R) 2012 and A2:2010/(R)2012Complies The system (control unit and applicator) underwent electrical and mechanical safety performance testing and complies with this standard.
Electromagnetic Compatibility (EMC): Compliance with IEC 60601-1-2:2014 (Edition 4.0)Complies The control unit underwent electromagnetic compatibility testing and complies with this standard. The applicator did not require additional EMC testing because the changes did not affect its electromagnetic compatibility.
Software Verification & Validation: Meeting requirements as per FDA guidanceMet all acceptance criteria. Software verification was performed in three steps: verification of the software requirements, verification of the design specifications and verification of the software architecture. All tests, including Integration Tests and System Level Test, were performed successfully and met their acceptance criteria.
Biocompatibility: No changes to patient-contacting materials, thus maintaining previous biocompatibility.No changes There were no changes to patient-contacting materials in comparison with the previously cleared predicate device, implying continued biocompatibility.
Cybersecurity: No impact on cybersecurity from software changes.No impact. Cybersecurity risk management was performed. The changes made to the predicate's system software have no impact on cybersecurity.
Therapy Time: Up to 30 minSame as predicate.
Pulse Repetition Rate: 1 - 150 HzSame as predicate.
Pulse Duration: 370 µs ± 20%Same as predicate.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample size: Not applicable for a clinical test set from this document. The "test set" here refers to engineering and performance verification tests of the device itself (e.g., electrical safety, EMC, software). Specific numbers are not provided for electrical and mechanical safety performance testing or EMC testing, but it states these were "undergone." The system level testing mentioned refers to device performance parameters (magnetic field, temperature) and not a patient-based clinical study.
  • Data Provenance: Not applicable for a clinical test set. The testing is for the device's technical specifications.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This 510(k) submission is for modifications to a physical therapeutic device, not an AI/imaging device requiring expert interpretation of outputs. Therefore, there are no "experts" establishing ground truth in the sense of medical image annotation or diagnosis. The "ground truth" for the device's performance is its technical specifications and compliance with established engineering and safety standards (e.g., IEC, ANSI AAMI).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is not an image-based or diagnostic study requiring adjudication of expert interpretations.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a powered muscle stimulator, not an AI or imaging device, and no clinical MRMC study was conducted or required for this 510(k) submission (as stated in Section VII: "No clinical testing was required for these changes.").

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This refers to the device's functional performance tests (e.g., magnetic field output, temperature control) as described under "Performance Testing" and compliance with engineering standards. It is a standalone device in that its primary function is to directly stimulate muscles, not provide algorithmic analysis to a human. The document confirms that the device's physical performance characteristics (e.g., magnetic field intensity, temperature rise) were tested.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" here is based on:

  • Engineering Specifications and Standards: The device's technical parameters (e.g., magnetic field intensity, pulse duration, frequency) were verified against its design specifications and relevant international standards (e.g., ANSI AAMI ES 60601-1, IEC 60601-1-2, ISO 62304).
  • Predicate Device Performance: The primary "ground truth" for substantial equivalence is that the modified device performs identically or equivalently to the predicate device (CoolTone K192940) for its intended use.
  • Safety and Biocompatibility Standards: Compliance with standards like ISO 10993 for biocompatibility and IEC 60601 series for electrical safety.

8. The sample size for the training set

Not applicable. This is not an AI/machine learning device with a training set.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this type of device and submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

April 1, 2022

Zimmer MedizinSysteme GmbH % Scott Blood Principal Consultant Quality and Regulatory Services 151 Glensondale Road Stow, Massachusetts 01775

Re: K220601

Trade/Device Name: CoolTone Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: March 1, 2022 Received: March 2, 2022

Dear Scott Blood:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Heather Dean, PhD Assistant Director THT5B3: Acute Injury Devices DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220601

Device Name CoolTone

Indications for Use (Describe)

CoolTone is indicated to be used for:

  • Improvement of abdominal tone, strengthening of the abdominal muscles, development for firmer abdomen.
    · Strengthening, toning and firming of buttocks and thighs.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a horizontal line above the "z" and "i". Below the word "zimmer" is the phrase "MedizinSysteme" in a smaller, sans-serif font. The logo is simple and modern, and the use of a bold font makes it easily recognizable.

6. 510(K) SUMMARY

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

I. SUBMITTER:Zimmer MedizinSysteme GmbH
Junkersstrasse 9
89231 Neu-Ulm, Germany
Establishment Registration: 8010720
Ms. Julia Kredler
Manager, Regulatory Affairs
Phone: +49-731-9761-625
Fax: +49-731-9761-118
E-mail: j.kredler@zimmer.de
CONTACT:Scott Blood
Principal Consultant
Phone: 978-729-5978
Fax: +49-731-9761-118
Email: scottqara@gmail.com
DATE PREPARED:March 31, 2022
II. DEVICE:
TRADE NAME:CoolTone
COMMON NAME:Powered Muscle Stimulator
CLASSIFICATION NAME:Stimulator, Muscle, Powered, For Muscle Conditioning
DEVICE CLASSIFICATION:Class II, 21 CFR §890.5850
PRODUCT CODE:NGX
III. PREDICATE DEVICE:CoolTone (K192940)

IV. DEVICE DESCRIPTION:

The CoolTone system is a non-invasive, therapeutic device. The device produces an electromagnetic field that interacts with the tissues of the human body. By muscle stimulation, the CoolTone helps to strengthen, tone and firm the abdomen, buttocks and thighs.

{4}------------------------------------------------

The device is a mobile, standalone unit with four wheels and the control unit housing protects the patient from electrical shock and mechanical injury.. These features are unchanged.

Two large applicators are connected to the control unit and can be used simultaneously depending on the treatment. CoolTone is a medical device that generates a magnetic field by applying a strong current to an applicator. It is equipped with the securement system, designed to maintain the position of the applicator throughout treatment.

A large color touch screen facilitates the use of the device. The on-screen information guides the user, step-by-step, through the entire treatment process. The treatment is operated through variable parameters such as frequency, time and intensity. Three, pre-set treatment options are available for users to choose from: Abdomen, buttocks and thighs. These features are unchanged.

V. INDICATION FOR USE:

CoolTone is indicated to be used for:

  • Improvement of abdominal tone, strengthening of the abdominal muscles, development ● for firmer abdomen.
  • . Strengthening, toning and firming of buttocks and thighs.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:

The technological characteristics and operating principles associated with the treatment remain unchanged from the predicate device. The subject device is identical to the primary predicate device, the CoolTone system, as cleared under K192940.

In comparison with the predicate device, the following changes have been made on the subject device:

Hardware: Control unit:

  • A compressor cooler was added to the cooling circuit to enhance cooling ● performance;
  • . The single diode was replaced by a double diode on the capacitor stage to reduce the potential for treatment interruption;
  • . Foam tape was added to the compressor module to reduce noise;

Applicator:

  • . Design and manufacture process improvement was made to the applicator to prevent cracking and the leaking of coolant fluid;

Software:

  • . Modified to allow customers limited access to a device check and an advanced customer screen to facilitate troubleshooting;

{5}------------------------------------------------

  • . Implemented a treatment protocol that was within the previously cleared range of software specification;
  • Adjusted the coolant flow, implemented relax-sequences and extended cooldown to reduce the potential interruption of treatment incurred by applicator warmth;
  • Optimized pulsing sequence to reduce noise;

Labeling:

  • The user's manual was updated to capture information obtained in post-market . surveillance activity.
    The technological characteristics of the subject device, as outlined in the comparison table below, remain the same as those of the predicate device.
Device NamePredicate DeviceCoolTone K192940SubjectDeviceCoolTone
Company NameZimmer MedizinSysteme GmbHSame
Product Code andRegulationPhysical Medicine 21 CFR 890.5850NGX-Stimulator, Muscle, Powered, MuscleConditioningSame
Intended UseImprovement of abdominal tone,•strengthening of the abdominal muscles,development of firmer abdomen.Strengthening, Toning and Firming of•buttocks and thighs.Same
Principle of ActionInitiating action potential of nerves results inmuscle contractionSame
Clinical UsePrescription useSame
Electrical ProtectionClass I, BFSame
User InterfaceTouch screen, 12"Same
Firmware ControlledYesSame
Type of EnergyMagnetic fieldSame
Number of outputs2Same
Number of ApplicatorsUp to two applicators can be operational atsame timeSame
Applicator ConnectionDetachable from the control unitSame
Total Induced Current inTissue (mA)327Same
Type of OperationContinuousSame
Number of Magnetic Coilsin the Applicator1Same
Positioning of ApplicatorSecurement systemSame
Magnetic Field Intensity0.5-1.35T ± 20%Same
Pulse Repetition Rate1 - 150 HzSame
Pulse Duration370 µs ± 20%Same
Pulse Amplitude0 - 100%Same

Table 6.1: Technological Characteristics between Subject and Predicate Device

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo is in black and white. The word "zimmer" is in a bold, sans-serif font. Below the word "zimmer" is the phrase "MedizinSysteme" in a smaller, sans-serif font.

Device NamePredicate DeviceCoolTone K192940SubjectDeviceCoolTone
Selection of parameters(Intensity, Time)YesSame
Therapy TimeUp to 30 minSame
Shape of StimulationPulseSymmetrical Biphasic SineWaveSame
Energy Source220-240V, 50/60 HzSame
System Dimensions(WxHxD)600x1100x600 mmSame
Operating AmbientTemperature10°C – 28 °CSame
EnvironmentalSpecificationsFor indoor use onlySame

Table 6.2: Technological Similarities between Proposed and Predicate Device per FDA Guidance for Industry for Powered Muscle Stimulators for 510(k)s (June 9, 1999)

TechnologicalCharacteristicsPredicate DeviceCoolTone K192940Subject DeviceCoolToneComments
Power Source(s)- Method of LineCurrent Isolation- Patient LeakageCurrentNormal conditionSingle fault condition60601 compliant< 1 μΑ3.9 μΑSame60601 Compliant
Number of OutputModes1Same
Number of OutputChannels- Synchronous orAlternating?- Method of ChannelIsolationSynchronousN/ASameNo electrodes -applicators arenot connected topatient
Regulated Current orRegulated Voltage?VoltageSameControlledvoltage to the coil
Software/Firmware/Microprocessor Control?YesSame
Automatic OverloadTrip?N/AN/ANo electrodes -applicators arenot connected topatient
Automatic No-LoadTrip?N/AN/ANo electrodes -applicators arenot connected topatient
TechnologicalCharacteristicsPredicate DeviceCoolTone K192940Subject DeviceCoolToneComments
Automatic Shut Off?YesSameUnit shuts offwith specifiedtimer
Patient OverrideControl?NoSameTreatment isdelivered byhealthcareprovider
Indicator Display:- On/Off Status?- Low Battery?- Voltage/Current Level?YesN/ANoSame
Timer Range(minutes)Up to 30 minSame
Compliance withVoluntary Standards?YesSameRefer toperformance datafor details
Compliance* with21 CFR 898? (*Becomesmandatory beginningMay 9, 2000)YesSame
Weight80 KgSame
Housing Materialsand ConstructionSteel and InjectionMolded PlasticsSame
Waveform (e.g., pulsedmonophasic, biphasic)Symmetrical BiphasicSineWaveSame
Shape (e.g., rectangular,spike, rectifiedsinusoidal)SinusoidalSame
Maximum OutputVoltage (specify units)N/AN/ANo electrodes -applicators arenot connected topatient
Maximum OutputCurrent (specify units)N/AN/ANo electrodes -applicators arenot connected topatient
Frequency (Hz)1-150 HzSame
For interferentialmodes only: - BeatFrequency (Hz)N/AN/A
For multiphasicwaveforms only: -Symmetrical phases?YesSameBiphasic
Phase Duration(include units) (staterange, if applicable)370 $\mu$ s +/- 20%Same
TechnologicalCharacteristicsPredicate DeviceCoolTone K192940Subject DeviceCoolToneComments
(both phases, ifasymmetrical)N/AN/A
Net Charge (mC perpulse) (If zero, statemethod of achievingzero net charge.)N/AN/ANo electrodes -applicators arenot connected topatient
Maximum PhaseCharge, (mC)N/AN/ANo electrodes -applicators arenot connected topatient
Maximum CurrentDensity, (mA/cm²)N/AN/ANo electrodes -applicators arenot connected topatient
Maximum PowerDensity, (W/cm²)(using smallestelectrode conductivesurface area)N/AN/ANo electrodes -applicators arenot connected topatient
Burst Mode7 (i.e., pulsetrains) a. Pulses perburst b. Bursts persecond c. Burstduration (seconds) d.Duty Cycle[Line (b) x Line (c)]N/AN/A
ON Time (seconds)N/AN/A
OFF Time (seconds)N/AN/A
Additional Features (ifapplicable)N/AN/A

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a line above the "z" and "i". Below the word "zimmer" is the text "MedizinSysteme" in a smaller, sans-serif font. The logo is in grayscale.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The word "zimmer" is in bold, lowercase letters on the top line. Below that, the words "MedizinSysteme" are in a smaller font size. The logo is in grayscale.

VII. PERFORMANCE DATA:

The CoolTone System is the same as the previously cleared predicate device (K192940). It has been tested/assessed against and fully complies with the following voluntary standards:

Table 6.3: Compliance with Voluntary Standards
--------------------------------------------------
StandardsStandardsOrganizationStandards Title
ANSI AAMI ES60601-1:2005/(R)2012and A1:2012,C1:2009/(R) 2012 andA2:2010/(R)2012ANSIAAMIMedical electrical equipment – Part 1: Generalrequirements for basic safety and essentialperformance
StandardsStandardsOrganizationStandards Title
60601-1-2:2014(Edition 4.0)IECMedical electrical equipment – Part 1-2: Generalrequirements for basic safety and essentialperformance – Collateral standard:Electromagnetic disturbances - Requirementsand tests
60601-1-6:2013(Edition 3.1)IECMedical electrical equipment – Part 1-6: Generalrequirements for basic safety and essentialperformance - Collateral standard: Usability
60601-2-10:2016(Edition 2.1)IECMedical electrical equipment - Part 2-10:Particular requirements for the Basic Safety andEssential Performance of Nerve and MuscleStimulators
62366-1:2015(Edition 1.0)IECMedical devices - Application of usabilityengineering to medical devices
62304:2015(Edition 1.1)ISOMedical devices software –software life cycleprocesses
14971:2012ISOMedical devices - Application of risk managementto medical devices
10993-1: 2018ISOBiological evaluation of medical devices - Part 1:Evaluation and testing within a risk managementprocess
10993-5: 2009ISOBiological evaluation of medical devices - Part 5:Tests for in vitro cytotoxicity
10993-10: 2010ISOBiological evaluation of medical devices - Part 10:Tests for irritation and skin sensitization

{9}------------------------------------------------

Software Verification and Validation Testing

Software verification and validation testing were conducted and documented as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

Software verification was performed in three steps: verification of the software requirements, verification of the design specifications and verification of the software architecture. All tests, including Integration Tests (Tests of the System Architecture and the

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The word "zimmer" is in bold, sans-serif font. Below the word "zimmer" is the word "MedizinSysteme" in a smaller, sans-serif font. The logo is in grayscale.

Design Specification) and the System Level Test (Test of functional requirements that were evaluated in the software requirements verification) were performed successfully and met their acceptance criteria.

Cybersecurity

Cybersecurity risk management for the device was performed as part of the overall risk management process for the medical device and follows the guidance in the FDA document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices". The changes made to the predicate's (K192940) system software have no impact on cybersecurity.

Biocompatibility Testing

There were no changes to patient-contacting materials in comparison with the previously cleared predicate device.

Electrical safety and electromagnetic compatibility (EMC)

The CoolTone system (control unit and applicator) has undergone electrical and mechanical safety performance testing as a result of the changes referenced. Additionally, the control unit underwent electromagnetic compatibility testing. The applicator did not require additional electromagnetic compatibility testing because the changes did not affect its electromagnetic compatibility.

The system complies with ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012, C1:2009/(R) 2012 and A2:2010/(R)2012), and IEC 60601-1-2:2014 (Fourth Edition).

Performance Testing

The proposed CoolTone system performance is the same as the previously cleared predicate device.

The System Level Testing confirmed that the two applicators, either operated separately or simultaneously, performed within the magnetic field intensity of 0.5 – 1.35T +/- 20% and that the tissue being treated by the device is not subjected to an appreciable rise in temperature at maximum intensity to cause a risk to the patient.

Clinical Study

No clinical testing was required for these changes.

VIII. CONCLUSION:

The Indication for Use for the subject CoolTone system is the same as the predicate device cleared in K192940. The changes that have been made to the system's hardware and software do not affect the intended use, performance or risk profile of the device. The subject CoolTone system is therefore substantially equivalent to the predicate device.

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).