K Number
K220583
Device Name
ClearCheck Model RADCC V2
Manufacturer
Date Cleared
2022-08-23

(175 days)

Product Code
Regulation Number
892.5050
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
ClearCheck is intended to assist radiation therapy professionals in generating and assessing the quality of radiotherapy treatment plans. ClearCheck is also intended to assist radiation treatment planners in predicting when a treatment plan might result in a collision between the treatment machine and the patient or support structures.
Device Description
The ClearCheck Model RADCC V2 device is software that uses treatment data, image data, and structure set data obtained from supported Treatment Planning System and Application Programming Interfaces to present radiotherapy treatment plans in a user-friendly way for user approval of the treatment plan. The ClearCheck device (Model RADCC V2) is also intended to assist users to identify where collisions between the treatment machine and the patient or support structures may occur in a treatment plan. It is designed to run on Windows Operating Systems. ClearCheck Model RADCC V2 performs calculations on the incoming supported treatment data. Supported Treatment Planning Systems are used by trained medical professionals to simulate radiation therapy treatments for malignant or benign diseases.
More Information

No
The summary describes software that performs calculations and presents data, but there is no mention of AI or ML techniques being used for prediction or analysis. The "predicting when a treatment plan might result in a collision" is described as a function of the software performing calculations, not through a learned model.

No
The device is intended to assist radiation therapy professionals in planning and assessing the quality of radiotherapy treatment plans, and in predicting collisions. It does not directly treat or diagnose a disease or condition, which is characteristic of a therapeutic device.

No

This device is intended to assist in generating and assessing the quality of radiotherapy treatment plans and predicting collisions. It does not diagnose a disease or condition in a patient.

Yes

The device description explicitly states that ClearCheck Model RADCC V2 is "software that uses treatment data, image data, and structure set data obtained from supported Treatment Planning System and Application Programming Interfaces". It is designed to run on Windows Operating Systems and performs calculations. There is no mention of accompanying hardware components or hardware-specific functions beyond running on a standard operating system.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is to assist radiation therapy professionals in generating and assessing treatment plans and predicting collisions. This is related to the planning and delivery of medical treatment, not the examination of specimens derived from the human body to provide information for diagnosis, monitoring, or screening.
  • Device Description: The device is software that processes treatment data, image data, and structure set data from treatment planning systems. It performs calculations related to radiation therapy planning and collision prediction. This is not consistent with the typical function of an IVD, which analyzes biological samples.
  • Lack of IVD Characteristics: There is no mention of analyzing biological specimens (blood, urine, tissue, etc.), providing diagnostic information, or being used for screening or monitoring disease states through the analysis of such specimens.

In summary, the ClearCheck device is a software tool for radiation therapy planning and quality assurance, which falls outside the scope of In Vitro Diagnostics.

N/A

Intended Use / Indications for Use

ClearCheck is intended to assist radiation therapy professionals in generating and assessing the quality of radiotherapy treatment plans. ClearCheck is also intended to assist radiation treatment planners in predicting when a treatment plan might result in a collision between the treatment machine and the patient or support structures.

Product codes

IYE

Device Description

The ClearCheck Model RADCC V2 device is software that uses treatment data, image data, and structure set data obtained from supported Treatment Planning System and Application Programming Interfaces to present radiotherapy treatment plans in a user-friendly way for user approval of the treatment plan. The ClearCheck device (Model RADCC V2) is also intended to assist users to identify where collisions between the treatment machine and the patient or support structures may occur in a treatment plan.

It is designed to run on Windows Operating Systems. ClearCheck Model RADCC V2 performs calculations on the incoming supported treatment data. Supported Treatment Planning Systems are used by trained medical professionals to simulate radiation therapy treatments for malignant or benign diseases.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Files/Treatment Planning System API-provided data containing CT, Structure Set, and Treatment Plan (including treatment field parameters) data

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Trained clinically qualified radiation oncology personnel

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

As with the Primary and Secondary Predicates, no clinical trials were performed for ClearCheck Model RADCC V2. Verification tests were performed to ensure that the software works as intended and pass/fail criteria were used to verify requirements.

Validation testing for the added BED / EQD2 functionality compared ClearCheck's results on a plan and plan sum against values calculated by hand using the well-known BED / EQD2 formulas. A passing criteria of 0.5% difference for dose type constraints and 3% for Volume type constraints was used. Validation testing for the added Deformed Dose functionality compared known dose deformations to the deformed dose results from ClearCheck. Qualitative dose-volume histogram (DVH) analysis showed good agreement for all cases and evaluation of Dman and Dmax differences used a quantitative +/-3% difference to achieve a passing result. The verification and validation testing passed in all test cases.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K162468, K171350

Reference Device(s)

K220598

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

0

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The FDA logo is in blue.

Radformation, Inc. % Kurt Sysock Co-founder/CEO 335 Madison Avenue, 4th Floor NEW YORK NY 10017

Re: K220583

Trade/Device Name: ClearCheck Model RADCC V2 Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: IYE Dated: July 20, 2022 Received: July 22, 2022

Dear Kurt Sysock:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act, or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

1

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Julie Sullivan. Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K220583

R220963

Device Name ClearCheck

Indications for Use (Describe)

ClearCheck is intended to assist radiation therapy professionals in generating and assessing the quality of radiotherapy treatment plans. ClearCheck is also intended to assist radiation treatment planners in predicting when a treatment plan might result in a collision between the treatment machine and the patient or support structures.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

This 510(k) Summary has been created per the requirements of the Safe Medical Device Act (SMDA) of 1990, and the content is provided in conformance with 21 CFR Part 807.92.

5.1. Submitter's Information

Table 1 : Submitter's Information
Submitter's Name:Kurt Sysock
Company:Radformation, Inc.
Address:335 Madison Avenue, 16th Floor
New York, NY 10017
Contact Person:Alan Nelson
Chief Science Officer, Radformation
Phone:518-888-5727
Fax:----------
Email:anelson@radformation.com
Date of Summary Preparation08/23/2022

5.2. Device Information

Table 2 : Device Information
Trade Name:ClearCheck
Common Name:Oncology Information System
Classification Name:Class II
Classification:Medical charged-particle radiation therapy system
Regulation Number:892.5050
Product Code:IYE
Classification Panel:Radiology

4

5.3. Predicate Device Information

Primary Predicate Device

ClearCheck Model RADCC V2 (Subject Device) makes use of its prior submission -ClearCheck Model RADCC (K162468) - as the Primary Predicate Device.

Secondary Predicate Device

With the addition of the Collision Check module included in the submission, ClearCheck Model RADCC V2 also makes use of CollisionCheck Model RADCO (K171350) as a Secondary Predicate Device for the additional functionality.

5.4. Device Description

The ClearCheck Model RADCC V2 device is software that uses treatment data, image data, and structure set data obtained from supported Treatment Planning System and Application Programming Interfaces to present radiotherapy treatment plans in a user-friendly way for user approval of the treatment plan. The ClearCheck device (Model RADCC V2) is also intended to assist users to identify where collisions between the treatment machine and the patient or support structures may occur in a treatment plan.

It is designed to run on Windows Operating Systems. ClearCheck Model RADCC V2 performs calculations on the incoming supported treatment data. Supported Treatment Planning Systems are used by trained medical professionals to simulate radiation therapy treatments for malignant or benign diseases.

5.5. Indications for Use

ClearCheck is intended to assist radiation therapy professionals in generating and assessing the quality of radiotherapy treatment plans. ClearCheck is also intended to assist radiation treatment planners in predicting when a treatment plan might result in a collision between the treatment machine and the patient or support structures.

5.6. Technological Characteristics

Primary Predicate Device

ClearCheck Model RADCC V2 (Subject Device) makes use of its prior submission -ClearCheck Model RADCC (K162468) - as the Primary Predicate Device.

The functionality and technical components of this prior submission remain unchanged in ClearCheck Model RADCC V2. This submission is intended to build on the functionality and technological components of the 510(k) cleared ClearCheck Model RADCC.

5

The main technological characteristics of the subject device and the Primary Predicate device remain the same. Both are computer-based software devices designed to run on Windows Operating Systems. Both use treatment data, image data, and structure set data obtained from supported Treatment Planning System and Application Programming Interfaces to present radiotherapy treatment plans in a user-friendly way for user approval of the treatment plan. Both devices share major functionality such as Dose Constraints, Plan Checks, Plan Reporting and Plan Comparison. Both are used by trained medical professionals to assist in the treatment planning process.

The submission for ClearCheck Model RADCC V2 contains additional functionality to assist in the plan generation process. The significant changes to the previously cleared devices (Primary and Secondary Predicates) are:

    1. Prescriptions may be entered and managed within the software.
    1. Radiation dose may be evaluated on different image sets through the use of a deformable registration object provided by Radformation's AutoContour Model RADAC V2 (K220598) software.
    1. Biologically Equivalent Dose (BED) and Equivalent Dose in 2 Gy fractions (EQD2) may be evaluated for plans and plan sums.
    1. Additional Dose Constraints and Plan Checks added.
    1. Addition of a Collision Check module that alerts the user to a potential collision between the treatment machine and the patient or support structures. The Secondary Predicate Device is now included in the Subject device with additional features as stated below.
    1. Additional reporting features.

Secondary Predicate Device

ClearCheck Model RADCC V2 (Subject Device) also makes use of CollisionCheck Model RADCO (K171350) as a Secondary Predicate Device for the additional functionality of the Collision Check module.

The Collision Check module within ClearCheck Model RADCC V2 is an updated version of the previously cleared CollisionCheck Model RADCO device. The core functionality of the Collision Check module is the same as in the Secondary Predicate Device, however additional features have been added. Treatment device components such as gantry head, SRS cones and Electron cones are modeled in the same way as they are in the Secondary Predicate Device. Similarly, SUPPORT, BOLUS and patient EXTERNAL structures are modeled in the same way as in the Secondary Predicate device. Collision calculations between the two classes of structures remain the same as in the Secondary Predicate Device, where a collision occurs if any part of a Treatment device component is found within a SUPPORT, BOLUS, or patient EXTERNAL structure for the particular

6

plan geometry. HU collisions are calculated in the same way as in the Secondary Predicate Device as well.

The added features to the Collision Check module in the Subject Device are as follows:

    1. The ability to assess collisions with kV On Board Imaging arms (OBI)
    1. The ability to detect collision warnings, i.e. situations where there is no collision detected but two objects are within a user defined warning distance of each other.
    1. Sticky settings
    1. Printing capabilities

| Table 3: Substantial Equivalence ClearCheck Model RADCC V2 vs.
Primary Predicate: ClearCheck Model RADCC (K162468)

and Secondary Predicate: CollisionCheck Model RADCO (K171350)
ParametersSubject Device:
ClearCheck Model
RADCC V2Primary
Predicate:
ClearCheck
Model RADCC
(K162468)Secondary
Predicate:
CollisionCheck
Model RADCO
(K171350)Equivalence
Indications for
UseClearCheck is
intended to assist
radiation therapy
professionals in
generating and
assessing the quality
of radiotherapy
treatment plans.
ClearCheck is also
intended to assist
radiation treatment
planners in predicting
when a treatment
plan might result in a
collision between the
treatment machine
and the patient or
support structures.ClearCheck is
intended for
quality
assessment of
radiotherapy
treatment plansCollisionCheck is
intended to assist
radiation treatment
planners in predicting
when a treatment
plan might result in a
collision between the
treatment machine
and the patient or
support structures.Equivalent to
Primary and
Secondary
Predicates
Energy Used
and/or DeliveredNone - software-only
application. The
software application
does not deliver or
depend on energy
delivered to or from
patients.None -
software-only
application. The
software
application does
not deliver or
depend on
energy delivered
to or fromNone - software-only
application. The
software application
does not deliver or
depend on energy
delivered to or from
patients.Equivalent to
both Primary
and Secondary
Predicates
patients.
Intended usersTrained clinically
qualified radiation
oncology personnelTrained clinically
qualified
radiation
oncology
personnelTrained clinically
qualified radiation
oncology personnelEquivalent to
both Primary
and Secondary
Predicates
OTC/RxRxRxRxEquivalent to
both Primary
and Secondary
Predicates
FunctionalityPerforms dosimetric
and plan evaluation
for Radiation
Treatment Plans. Also
simulates the plan
and predicts whether
that no gantry
collisions occur with
patient or support
structuresPerforms
dosimetric and
plan evaluation
for Radiation
Treatment Plans.Simulates the plan
and predicts
whether any gantry
collisions occur with
patient or support
structures.Equivalent to
both Primary
and Secondary
Predicates
Design:
Graphical User
InterfaceContains a Data
Visualization /
Graphical User
InterfaceContains a Data
Visualization /
Graphical User
InterfaceContains a Data
Visualization /
Graphical User
InterfaceEquivalent to
both Primary
and Secondary
Predicates
Design:
Supported FilesFiles/Treatment
Planning System
API-provided data
containing CT,
Structure Set, and
Treatment Plan
(including treatment
field parameters) dataFiles/Treatment
Planning System
API-provided
data containing
CT, Structure
Set, and
Treatment Plan
(including
treatment field
parameters) dataFiles/Treatment
Planning System
API-provided data
containing CT,
Structure Set, and
Treatment Plan
(including treatment
field parameters) dataEquivalent to
both Primary
and Secondary
Predicates
Design:
Calculation
RequirementsUses local hardwareUses local
hardwareUses local hardwareEquivalent to
both Primary
and Secondary
Predicates
Design:
ReportingReporting built-in and
user has ability to
customizeReporting built-in
and user has
ability to
customizeNo reporting
capabilitiesEquivalent to
Primary
Predicate
Pure softwareYesYesYesEquivalent to
both Primary
and Secondary
Predicates
Operating
SystemWindows Operating
SystemWindows
Operating
SystemWindows Operating
SystemEquivalent to
both Primary
and Secondary
Predicates
InputTreatment data,
image data, and
structure set data
obtained from
supported Treatment
Planning System and
Application
Programming
interfacesTreatment data,
image data, and
structure set data
obtained from
supported
Treatment
Planning System
and Application
Programming
interfacesTreatment data,
image data, and
structure set data
obtained from
supported Treatment
Planning System and
Application
Programming
interfacesEquivalent to
both Primary
and Secondary
Predicates
ClearCheck Model RADCC V2 vs. ClearCheck Model RADCC (K162468) Functionality Comparison
Dose
ConstraintsAdditional dose
constraints addedInitial set of dose
constraintsN/AMinor
differences
with Primary
Predicate
Plan ChecksAdditional plan
checks addedInitial set of plan
checksN/AMinor
differences
with Primary
Predicate
Create Users
and Edit User
RightsAddition of option to
authenticate with
Active Directory
credentialsUser rights
managed by
Administration
ApplicationN/AMinor
differences
with Primary
Predicate
Global and
Patient Specific
TemplatesAddition of
Prescription
templates and overall
improvements to
template usabilityInitial set of dose
constraint and
plan check
templatesN/ANew Feature /
Minor
differences
with Primary
Predicate
ReportingImprovements to
report usabilityInitial reporting
functionalityN/AMinor
differences
with Primary
Predicate
Compare
Treatment PlansImprovements to plan
comparison usabilityInitial plan
comparison
functionalityN/AMinor
differences
with Primary
Predicate
PrescriptionPhysician prescriptions may be entered and managed in the ClearCheck software.N/AN/ANew Feature
Deformed Dose EvaluationRadiation dose may be evaluated on different image sets through the use of a deformable registration object.N/AN/ANew Feature
BED/EQD2 doseAbility to evaluate BED/EQD2 for plans and plan sumsN/AN/ANew Feature
ClearCheck Model RADCC V2 vs. CollisionCheck Model RADCO (K171350) Functionality Comparison
ParametersSubject Device:
ClearCheck Model
RADCC V2N/ASecondary Predicate:
CollisionCheck
(K171350)Equivalence
Simulation DetailsSimulates the plan and predicts whether any gantry collisions occur with patient or support structures. Calculates gantry clearance by modeling the linac as a cylinder with a user-configured value for distance between isocenter and the face of the gantry. Collision Check also supports additional applicators: Stereotactic radiosurgery cones (also modeled as a cylinder) and Electron Applicators (modeled as a rectangular prism). ClearCheck also simulates the on-board imaging (OBI) arms as a rectangular prism. User can define aN/ASimulates the plan and predicts whether any gantry collisions occur with patient or support structures. Calculates gantry clearance by modeling the linac as a cylinder with a user-configured value for distance between isocenter and the face of the gantry. CollisionCheck also supports additional applicators: Stereotactic radiosurgery cones (also modeled as a cylinder) and Electron Applicators (modeled as a rectangular prism).Minor differences with Secondary Predicate
warning distance that
adds a margin to the
shapes in the
simulation.
Collision Check
OutputCollision Check tests
thousands of sample
points against CT
data and patient and
couch structures and
reports the number of
sample points that
resulted in a collision.
Collision Check also
displays these
sample point test
results with a 3D
display and an axial
2D image plane
viewer for the user to
inspect the results.
If the user has
defined warning
distances, then
collisions within the
warning margin are
presented to the user
with a different color
and label.N/ACollision Check tests
thousands of sample
points against CT
data and patient and
couch structures and
reports the number of
sample points that
resulted in a collision.
Collision Check also
displays these
sample point test
results with a 3D
display and an axial
2D image plane
viewer for the user to
inspect the results.Minor
differences
with Secondary
Predicate

7

8

9

10

5.7. Performance Data

As with the Primary and Secondary Predicates, no clinical trials were performed for ClearCheck Model RADCC V2. Verification tests were performed to ensure that the software works as intended and pass/fail criteria were used to verify requirements.

Validation testing for the added BED / EQD2 functionality compared ClearCheck's results on a plan and plan sum against values calculated by hand using the well-known BED / EQD2 formulas. A passing criteria of 0.5% difference for dose type constraints and 3% for Volume type constraints was used. Validation testing for the added Deformed

11

Dose functionality compared known dose deformations to the deformed dose results from ClearCheck. Qualitative dose-volume histogram (DVH) analysis showed good agreement for all cases and evaluation of Dman and Dmax differences used a quantitative +/-3% difference to achieve a passing result. The verification and validation testing passed in all test cases.

5.8. Conclusion

ClearCheck Model RADCC V2 is substantially equivalent to the Primary Predicate Device, ClearCheck Model RADCC (K162468). Additionally, the Collision Check module found within ClearCheck Model RADCC V2 is substantially equivalent to the Secondary Predicate Device, CollisionCheck Model RADCO (K171350). The minor technological differences between ClearCheck and the Primary and Secondary Predicate Devices do not raise any questions on the safety and effectiveness of the Subject Device. Verification and Validation testing and the Risk Management Report demonstrate that the ClearCheck is as safe and effective as the Primary and Secondary Predicates.