K Number
K220411
Date Cleared
2022-05-13

(88 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Reprocessed LF1212A LigaSure Sealer/Divider is a bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/ Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery and in such surgical specialties as urologic, thoracic, plastic, and reconstructive. Procedure may include, but are not limited to, bowel resections, gall bladder procedures, Nissen fundoplication, and adhesiolysis.

The instrument is also indicated for open ENT procedures in adults (thyroidectomy, radical neck dissection, parotidectorny, and tonsillectorny) for ligation of vessels, lymphatics and tissue bundles 2-3 mm away from unintended thermally-sensitive structures such as nerves and parathyroid glands.

The LigaSure system has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

Device Description

The Reprocessed LigaSure Curved, Small Jaw, Open Sealer/Divider (LF1212A) is a bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The instrument is intended to be used with Covidien electrosurgical generators that include vessel sealing capability. The instrument can be used on vessels (arteries and veins) up to and including 7 mm. The instrument creates a seal by application of radiofrequency (RF) electrosurgical energy to vascular structures (vessels and lymphatics) or tissue bundles interposed between the jaws of the instrument. A blade within the instrument is surgeon actuated to divide tissue. The following controls are located on the device:

  • Scissor like design for opening and closing the instrument jaws and activating RF energy. The mechanism must be held in the closed position during vessel sealing and cutting.
  • An activation button at the bottom of the handle for generator power to initiate vessel sealing. The button has two stages that provide tactile feedback to the user. The first stage button click does not apply energy, whereas the second stage click begins energy application.
  • A trigger for actuating the blade.
    All controls can be operated with either the right or left hand. Vessel sealing can be initiated using the activation button or utilizing a footswitch connected to the generator. The instrument attached to the generator via a cable with a connector that identifies the instrument type to the generator.

The instrument is compatible with the Covidien ForceTriad Enerqy Platform and Valleylab FT10 Energy Platform.

The scope of the submission only includes the Reprocessed LigaSure Curved, Small Jaw, Open Sealer/Divider and not the Covidien ForceTriad Energy Platform or Valleylab FT10 Energy Platform that are used to power the device, or the footswitch that connects to the generator. Stryker Sustainability Solutions does not reprocess or market the generators or footswitch.

AI/ML Overview

This document is a 510(k) premarket notification for a reprocessed medical device, the Reprocessed LigaSure Curved, Small Jaw, Open Sealer/Divider (LF1212A). It demonstrates substantial equivalence to a predicate device.

Important Note: The provided document is for a reprocessed surgical instrument, not an AI-based diagnostic or therapeutic device. Therefore, the questions related to AI-specific criteria (such as multi-reader multi-case studies, expert consensus for ground truth, training set information, effects of AI assistance on human readers) are not applicable to this submission. The "acceptance criteria" discussed in this document refer to the performance of the reprocessed physical device, not an AI algorithm's performance against clinical endpoints.

Here's an analysis based on the provided text, addressing the relevant points and indicating where information is not applicable:


1. A table of acceptance criteria and the reported device performance

The document does not present acceptance criteria and reported device performance in a formal table with specific numerical thresholds. Instead, it lists the types of tests conducted to demonstrate substantial equivalence to the predicate device. The general acceptance criterion for this 510(k) submission is that the reprocessed device performs "at least as safe and effective as the predicate" and "as well as the identified legally marketed predicate device."

Here's a summary of the performance tests conducted, implying the "reported device performance" met the equivalence standard for each:

Acceptance Criteria (Implied: Performance comparable to or equivalent to the predicate device for safe and effective use)

Test CategorySpecific Tests MentionedReported Device Performance (Implied)
BiocompatibilityNot detailed, but generally involves testing for biological safety (e.g., cytotoxicity, sensitization, irritation).Demonstrated biocompatibility, indicating the reprocessed device does not pose new or unacceptable biological risks.
Validation of ReprocessingNot detailed, but typically includes cleaning, disinfection/sterilization, and functional restoration processes. This includes "removal of adherent visible soil and decontamination."Reprocessing methods were validated to ensure the device is safely and effectively prepared for reuse. Each device is "tested for appropriate function of its components prior to packaging and labeling operations."
SterilizationNot detailed, but involves validation of the sterilization method.Sterilization was validated to ensure the device is sterile for patient use.
Electrical Safety and EMCIEC 60601-1 (General requirements for basic safety and essential performance), IEC 60601-1-2 (Electromagnetic compatibility), IEC 60601-2-2 (Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories).The device met the requirements of the specified IEC standards, indicating safe electrical operation and electromagnetic compatibility.
Functional Performance TestsBlade Trigger Actuation Force, Clamp Arm Closing and Seal Button Force, Clamp Arm Opening Force, Clamp Arm Deflection, Jaw Clamp Force, Blade Excursion, Jaw Opening Angle, Burst Pressure, Maximum Jaw Temperature, Device Reliability, Functional Attribute Testing. Evaluated ability to "seal and divide vessels from 1mm to 7mm."All functional tests demonstrated performance "at least as safe and effective as the predicate" and "as well as the identified legally marketed predicate device." Specifics include successful sealing and division of vessels 1mm to 7mm, with satisfactory burst pressure, maximum jaw temperature, reliability, and functional attributes.
Preclinical Laboratory EvaluationsAcute and chronic survival studies in an animal model to evaluate "thermal spread and the ability to achieve hemostasis of vessels."Animal studies confirmed the reprocessed device performs similarly to the predicate in terms of thermal spread control and hemostasis effectiveness. The conclusion states these studies "demonstrate that the Reprocessed LigaSure Curved, Small Jaw, Open Sealer/Divider is at least as safe and effective as the predicate and perform as well as the identified legally marketed predicate device."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: The document does not specify exact numerical sample sizes for each test conducted (e.g., number of devices tested for each functional performance metric, number of animals in preclinical studies). It generally refers to "bench and laboratory testing" and "preclinical laboratory evaluations in an animal model."
  • Data Provenance: The tests were conducted by Stryker Sustainability Solutions, a U.S. company. The data would primarily originate from their internal testing facilities. The nature of the testing (bench, in-vitro, and animal studies) is prospective for the purpose of this 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Applicable: This pertains to an AI/imaging device. For this reprocessed surgical instrument, "ground truth" is established through engineering and performance testing against predefined mechanical, electrical, and functional specifications, as well as comparative performance in an animal model. There is no concept of expert human image interpretation or "ground truth" in the AI sense for this device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable: This pertains to an AI/imaging device where human readers might disagree. For a physical device, testing involves quantitative measurements and observations against specifications. Any discrepancies would be handled through standard engineering quality control and retesting protocols, not through expert adjudication in a clinical reading context.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable: This is specific to AI devices and their impact on human reading performance. This document is for a reprocessed surgical instrument.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable: This is specific to AI algorithms. This is a physical surgical device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

For this physical device, the "ground truth" is based on:

  • Engineering Specifications and Benchmarking: Mechanical properties (e.g., force, angles, excursion), electrical properties, and material integrity tested against established standards and the performance of the predicate device.
  • Physiological Performance in Animal Models: "Thermal spread and the ability to achieve hemostasis of vessels" in an animal model serve as a form of "ground truth" for the device's functional performance in a biological system, compared to the expected performance of new devices or the predicate.

There is no "expert consensus" or "pathology" as "ground truth" in the context of clinical endpoints for an AI system.


8. The sample size for the training set

  • Not Applicable: This question relates to AI model training. This document is about a reprocessed physical device, which does not have a "training set" in the machine learning sense. The device is re-manufactured and tested, not "trained."

9. How the ground truth for the training set was established

  • Not Applicable: As above, this pertains to AI model training.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The text is in blue, with "FDA" in a larger font size and enclosed in a blue square.

May 13, 2020

Stryker Sustainability Solutions Mr.Scott English, BS Principal Regulatory Affairs Specialist 1810 W Drake Drive Tempe, Arizona 85283

Re: K220411

Trade/Device Name: Reprocessed LigaSure Curved, Small Jaw, Open Sealer/Divider (LF1212A) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: NUJ Dated: April 15, 2022 Received: April 18, 2022

Dear Scott English:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent ffor the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (0S) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220411

Device Name

Reprocessed LigaSure Curved, Small Jaw, Open Sealer/Divider (LF1212A)

Indications for Use (Describe)

The Reprocessed LF1212A LigaSure Sealer/Divider is a bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/ Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery and in such surgical specialties as urologic, thoracic, plastic, and reconstructive. Procedure may include, but are not limited to, bowel resections, gall bladder procedures, Nissen fundoplication, and adhesiolysis.

The instrument is also indicated for open ENT procedures in adults (thyroidectomy, radical neck dissection, parotidectorny, and tonsillectorny) for ligation of vessels, lymphatics and tissue bundles 2-3 mm away from unintended thermally-sensitive structures such as nerves and parathyroid glands.

The LigaSure system has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Reprocessed Single-Use Device Model Included in Clearance

Device ModelDevice NameOriginal Manufacturer
LF1212ALigaSure Curved, Small Jaw, Open Sealer/DividerCovidien

{4}------------------------------------------------

510(k) SUMMARY

Submitter:

Stryker Sustainability Solutions 1810 W. Drake Drive Tempe, Arizona 85283

Contact:

Scott English, BS Principal Regulatory Affairs Specialist 901-451-1456 (o) 480-763-5310 (f) scott.english@stryker.com

Date of Preparation: May 13, 2022

Name of Device:

Trade/Proprietary Name:Reprocessed LigaSure Curved, Small Jaw, Open Sealer/Divider(LF1212A)
Common Name:Electrosurgical cutting and coagulation device and accessories
Classification Information:Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic &Endoscopic, Reprocessed(21 CFR§878.4400, Product Code NUJ, Class II)

Predicate Devices:

ModelNumber510(k)Number510(k) TitleOriginalManufacturer
LF1212AK152286LigaSure Curved, Small Jaw, Open Sealer/DividerCovidien

Device Description:

The Reprocessed LigaSure Curved, Small Jaw, Open Sealer/Divider (LF1212A) is a bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The instrument is intended to be used with Covidien electrosurgical generators that include vessel sealing capability. The instrument can be used on vessels (arteries and veins) up to and including 7 mm. The instrument creates a seal by application of radiofrequency (RF) electrosurgical energy to vascular structures (vessels and lymphatics) or tissue bundles interposed between the jaws of the instrument. A blade within the instrument is surgeon actuated to divide tissue. The following controls are located on the device:

  • Scissor like design for opening and closing the instrument jaws and activating RF energy. The mechanism must be held in the closed position during vessel sealing and cutting.
  • An activation button at the bottom of the handle for generator power to initiate vessel . sealing. The button has two stages that provide tactile feedback to the user. The first stage

{5}------------------------------------------------

button click does not apply energy, whereas the second stage click begins energy application.

  • . A trigger for actuating the blade.
    All controls can be operated with either the right or left hand. Vessel sealing can be initiated using the activation button or utilizing a footswitch connected to the generator. The instrument attached to the generator via a cable with a connector that identifies the instrument type to the generator.

The instrument is compatible with the Covidien ForceTriad Enerqy Platform and Valleylab FT10 Energy Platform.

The scope of the submission only includes the Reprocessed LigaSure Curved, Small Jaw, Open Sealer/Divider and not the Covidien ForceTriad Energy Platform or Valleylab FT10 Energy Platform that are used to power the device, or the footswitch that connects to the generator. Stryker Sustainability Solutions does not reprocess or market the generators or footswitch.

Intended Use:

The Reprocessed LF1212A LigaSure Sealer/Divider is a bipolar electrosurgical instrument intended for use in open surgical procedures where ligation of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery and in such surgical specialties as urologic, thoracic, plastic, and reconstructive. Procedure may include, but are not limited to, bowel resections, gall bladder procedures, Nissen fundoplication, and adhesiolysis.

The instrument is also indicated for open ENT procedures in adults (thyroidectorny, radical neck dissection, parotidectorny, and tonsillectomy) for ligation of vessels, lymphatics and tissue bundles 2-3 mm away from unintended thermally-sensitive structures such as nerves and parathyroid glands.

The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

Indications for Use Comparison:

The indications for use for the proposed device are the same in comparison to the predicate device.

Technological Comparison:

The design, materials, and intended use of the Reprocessed LigaSure Curved, Small Jaw, Open Sealer/Divider are equivalent to the predicate device. The mechanism of action of the reprocessed device is identical to the predicate device in that the same standard mechanical design, materials, and size is utilized. There are no changes to the claims, intended use, clinical application, patient population, performance specifications, or method of operation. In addition, Stryker Sustainability Solutions' reprocessing of the device includes removal of adherent visible soil and decontamination. Each individual device is tested for appropriate function of its components prior to packaging and labeling operations. The only differences between the Reprocessed LigaSure Curved, Small Jaw, Open Sealer/Divider and the predicate LigaSure

{6}------------------------------------------------

Curved. Small Jaw. Open Sealer/Divider are that the device is reprocessed, and some device components are replaced with equivalent components during the reprocessing operation.

Predicate DeviceSubject Device
UsesSingle Patient UseSingle Patient Use
Device Length19cm19cm
Jaw Length16.5mm16.5mm
Jaw ShapeCurvedCurved
Mechanism for CuttingMechanical - User actuated bladeMechanical - User actuated blade

Performance Data:

Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed LigaSure Curved, Small Jaw, Open Sealer/Divider. This included the following tests:

  • Biocompatibility ●
  • Validation of Reprocessing
  • Sterilization
  • Electrical Safety and Electromagnetic Compatibility
  • Functional Performance Tests
    • Blade Trigger Actuation Force O
    • O Clamp Arm Closing and Seal Button Force
    • Clamp Arm Opening Force o
    • Clamp Arm Deflection O
    • O Jaw Clamp Force
    • Blade Excursion O
    • Jaw Opening Angle O
    • Burst Pressure O
    • Maximum Jaw Temperature O
    • Device Reliability O
    • Functional Attribute Testing O

The functional performance testing involved electrical safety and electromagnetic compatibility testing in accordance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2, and verification/comparative testing (to the predicate device). The bench testing involved evaluation of the device's performance and ability to seal and divide vessels from 1mm to 7mm, including burst pressure, maximum jaw temperature, device functionality, device reliability, and functional attribute tests.

Additionally, preclinical laboratory evaluations in an animal model were performed, which included acute and chronic survival studies were done to evaluate thermal spread and the ability to achieve hemostasis of vessels of the reprocessed device.

Conclusion:

The results of bench testing and preclinical laboratory evaluations demonstrate that the Reprocessed LigaSure Curved, Small Jaw, Open Sealer/Divider is at least as safe and effective as the predicate and perform as well as the identified legally marketed predicate device as described herein.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.