K Number
K220369
Date Cleared
2022-09-23

(227 days)

Product Code
Regulation Number
872.6660
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Esthetic Colorant is used for coloring pre-sintered zirconia structures for the fabrication of all-ceramic restorations.

Device Description

Esthetic Colorant, a colorant with a zirconia structure, is a dilute aqueous solution of transition metal salts and lanthanoid metal salts and enables appropriate color adjustment when combined with a pre-sintered zirconia structure such as KATANA Zirconia (K143439).

Esthetic Colorant colors the pre-sintered zirconia structure by applying it with a metal-free brush. The structure is then dried and then sintered at temperatures above 1400°C. Sintering welds the coloring components in Esthetic Colorant, making it possible to adjust the color shade of the zirconia structure.

Esthetic Colorant has 12 shades in bottle size (12 ml). In addition, Esthetic Colorant can only be used by professional technicians in the field of dentistry.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the "Esthetic Colorant" device, based on the provided FDA 510(k) summary:

This document is a 510(k) summary for a medical device that colors pre-sintered zirconia structures for dental restorations. It is not an AI/ML device, and therefore several of the requested sections (sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, sample size for training set, ground truth for training set) are not applicable.

1. Table of acceptance criteria and the reported device performance

Acceptance Criteria (from ISO 6872:2015)Reported Device Performance (Esthetic Colorant with KATANA Zirconia HT)
Flexural Strength:
Equal to or greater than 900 MPaCOMPLIES (for all tested colorants)
Coefficient of Thermal Expansion:
Shall not deviate by more than 0.5 x 10^-6 K^-1 from manufacturer's stated valueCOMPLIES (for all tested colorants)
Chemical Solubility:
Less than to 100 µg/cm²COMPLIES (for all tested colorants)

2. Sample sized used for the test set and the data provenance

  • Sample Size: Not explicitly stated how many samples were tested for each physical and chemical property, other than that the properties were evaluated according to ISO 6872:2015.
  • Data Provenance: Not explicitly stated, but the company is in Japan (Kuraray Noritake Dental Inc.) so it can be inferred that the testing was conducted there. The testing is prospective as it's for a new product submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. This is a material science device, not an AI/ML device requiring expert ground truth for image interpretation or diagnosis. The ground truth refers to adherence to physical and chemical property standards.

4. Adjudication method for the test set

  • Not Applicable. See point 3. Testing involves objective measurements against established ISO standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is a material science device, not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is a material science device; there is no algorithm or standalone performance in the context of AI.

7. The type of ground truth used

  • International Standard Compliance: The ground truth is established by the requirements of the international standard ISO 6872:2015 for dental ceramics. This standard defines the acceptable ranges for physical (flexural strength, coefficient of thermal expansion) and chemical (chemical solubility) properties.

8. The sample size for the training set

  • Not Applicable. This is not an AI/ML device with a training set.

9. How the ground truth for the training set was established

  • Not Applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

September 23, 2022

Kuraray Noritake Dental Inc. Yasujiro Ohara Manager Tokiwabashi Tower, 2-6-4, Otemachi Chiyoda-ku, Tokyo 100-0004 JAPAN

Re: K220369

Trade/Device Name: Esthetic Colorant Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH, Dated: August 24, 2022 Received: August 26, 2022

Dear Yasujiro Ohara:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220369

Device Name Esthetic Colorant

Indications for Use (Describe)

Esthetic Colorant is used for coloring pre-sintered zirconia structures for the fabrication of all-ceramic restorations.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Date: March 8, 2022

510(k) Summary

K220369

1. 510(k) submitters information
1) NameKuraray Noritake Dental Inc.
2) Address1621 Sakazu, Kurashiki, Okayama 710-0801, Japan
3) Contact personYasujiro OharaManagerQuality Assurance Department
4) Contact person in USAManabu SuzukiDirectorKURARAY AMERICA, INC.32 Old Slip, 7th Floor, New York, NY 10005Tel: (212)-986-2230 (Ext. 115) or (800)-879-1676Fax: (212)-867-3543

2. Subject device identification

1) Trade / Proprietary nameEsthetic Colorant
2) Common NameColoring liquid for zirconia structures
3) Classification Number / Name21 CFR 872.6660 / Porcelain Powder for Clinical Use
4) ClassII
5) Product codeEIH
6) PanelDental

3. Legally marketed predicate device

1) CompanyZirkonzahn srl
2) Device nameZirkonzahn COLOUR LIQUID
3) Common NameLiquid for Dental Zirconia Prosthesis
4) Classification Number / Name21 CFR 872.6660 / Porcelain Powder for Clinical Use
5) ClassII
6) Product codeEIH
7) PanelDental
8) 510(k) NumberK190518

4. Device description

Esthetic Colorant, a colorant with a zirconia structure, is a dilute aqueous solution of transition metal salts and lanthanoid metal salts and enables appropriate color adjustment when combined with a pre-sintered zirconia structure such as KATANA Zirconia (K143439).

Esthetic Colorant colors the pre-sintered zirconia structure by applying it with a metal-free brush. The structure is then dried and then sintered at temperatures above 1400°C. Sintering welds the coloring components in Esthetic Colorant, making it possible to adjust the color shade of the zirconia structure.

Esthetic Colorant has 12 shades in bottle size (12 ml). In addition, Esthetic Colorant can only be used by professional technicians in the field of dentistry.

5. Indications for Use / Intended Use

Esthetic Colorant is used for coloring pre-sintered zirconia structures for the fabrication of all-ceramic restorations.

6. Performance data

Non-clinical testing was performed to evaluate the physical and chemical properties of the colored pre-sintered zirconia structures to demonstrate that the subject device of interest did not adversely affect the function of the zirconia structures. Since the product used for the evaluation is a dental zirconia structure, the physical and chemical properties were evaluated according to ISO 6872: 2015. As a result of the subject device does not have an adverse effect, so it is suitable for the purpose of use.

{4}------------------------------------------------

7. Performance data summary

7-1. Summary of physical properties to the product standard (ISO 6872:2015)

ProductFlexural strengthTest ResultProductCoefficient of thermal expansionTest Result
KATANA Zirconia (HT)Equal to or greater than 900MPa.COMPLIESKATANA Zirconia (HT)The coefficient of thermal expansion ofthe ceramics shall not deviate by morethan $0.5 ×10^{-6}K^{-1}$ from the value statedby the manufacturer.COMPLIES
KATANA Zirconia (HT)Shaded with Esthetic Colorant (B plus)COMPLIESKATANA Zirconia (HT)Shaded with Esthetic Colorant (B plus)COMPLIES
KATANA Zirconia (HT)Shaded with Esthetic Colorant (BROWN)COMPLIESKATANA Zirconia (HT)Shaded with Esthetic Colorant (BROWN)COMPLIES
KATANA Zirconia (HT)Shaded with Esthetic Colorant (ORANGE)COMPLIESKATANA Zirconia (HT)Shaded with Esthetic Colorant (ORANGE)COMPLIES
KATANA Zirconia (HT)Shaded with Esthetic Colorant (BLUE)COMPLIESKATANA Zirconia (HT)Shaded with Esthetic Colorant (BLUE)COMPLIES
KATANA Zirconia (HT)Shaded with Esthetic Colorant (GRAY)COMPLIESKATANA Zirconia (HT)Shaded with Esthetic Colorant (GRAY)COMPLIES
KATANA Zirconia (HT)Shaded with Esthetic Colorant (VIOLET)COMPLIESKATANA Zirconia (HT)Shaded with Esthetic Colorant (VIOLET)COMPLIES
KATANA Zirconia (HT)Shaded with Esthetic Colorant (OPAQUE)COMPLIESKATANA Zirconia (HT)Shaded with Esthetic Colorant (OPAQUE)COMPLIES
KATANA Zirconia (HT)Shaded with Esthetic Colorant (PINK)COMPLIESKATANA Zirconia (HT)Shaded with Esthetic Colorant (PINK)COMPLIES

7-2. Summary of chemical properties to the product standard (ISO 6872:2015)

ProductChemical solubilityTest Result
KATANA Zirconia (HT)Less than to 100µg/cm².COMPLIES
KATANA Zirconia (HT)Shaded with Esthetic Colorant (B plus)COMPLIES
KATANA Zirconia (HT)Shaded with Esthetic Colorant (BROWN)COMPLIES
KATANA Zirconia (HT)Shaded with Esthetic Colorant (ORANGE)COMPLIES
KATANA Zirconia (HT)Shaded with Esthetic Colorant (BLUE)COMPLIES
KATANA Zirconia (HT)Shaded with Esthetic Colorant (GRAY)COMPLIES
KATANA Zirconia (HT)Shaded with Esthetic Colorant (VIOLET)COMPLIES
KATANA Zirconia (HT)Shaded with Esthetic Colorant (OPAQUE)COMPLIES
KATANA Zirconia (HT)Shaded with Esthetic Colorant (PINK)COMPLIES

{5}------------------------------------------------

8. Substantial Equivalence Discussion
Propertiesand informationSubject device:Esthetic ColorantPredicate device: (K190518)Zirkonzahn COLOUR LIQUIDComparisoncomments
CompanyKuraray Noritake Dental Inc.Zirkonzahn srl
Product codeEIHEIHSame
Regulation name(Regulation No.)Porcelain powder for clinical use(872.6660)Porcelain powder for clinical use(872.6660)Same
Indications for use/Intended useEsthetic Colorant is used forcoloring pre-sintered zirconiastructures for the fabrication ofall-ceramic restorations.Zirkonzahn COLOUR LIQUID isused for coloring pre-sinteredzirconia structures.Similar
TechnologyAqueous solutions of transitionand lanthanoid metal salts.Aqueous solutions of transitionand lanthanoid metal salts.Similar
ChemicalcompositionA solvent consisting mostly ofwater.A solute consisting of transitionand lanthanoid metal salts.Other chemical ingredients.A solvent consisting mostly ofwater.A solute consisting of transitionand lanthanoid metal salts.Minordifference(Note 5-8-1.)
OperatingprincipleBrush zirconia ceramic structureswith coloring liquid beforesinteringBrush or immerse zirconiaceramic structures with coloringliquid before sinteringSimilar
Type of packagingand volumeBottle; 12 mlBottle; 20 ml, 50 ml, 100 mlSimilar
ShadeVariousVariousSimilar
Storage conditionsMax: 30°C, 2 yearsMax: 25°C, 5 yearsMinordifference(Note 5-8-2.)
SterileNon-sterileNon-sterileSame
BiocompatibilityEstablished(the worst case biocompatibilitywith ISO 10993-1:2018)Established(the worst case biocompatibilitywith ISO 10993-1:2010)Similar(Note 5-8-3.)

0 C--latoration Lia---------

8-1. Discussion of chemical composition

The subject device and the predicate device are the same in that they are aqueous solutions of transition metal salts and lanthanoid metal salts. As the subject device contains other chemical ingredients, we evaluated the biocompatibility of the subject device in combination with 510(k) cleared dental zirconia and confirmed that these differences of chemical composition did not affect the biological safety of the subject device.

8-2. Discussion of storage conditions

In order to establish the storage conditions of the subject device, the test sample was accelerated and the requirements of ISO 6872:2015 was evaluated, and it was confirmed that there was no problem.

8-3. Established of biocompatibility

Biocompatibility was evaluated using the same ISO 10993-1 as the predicate device. Biocompatibility test was evaluated on the worst case model of the subject device as in the predicate device, and it was confirmed that the subject device satisfied biocompatibility.

9. Conclusion

The subject device and the predicate device are composed of similar chemical composition, and the difference does not matter. It was also confirmed that there was no problem with storage conditions and biocompatibility. From the above. it was judged that the two were substantially equivalent. In addition, from performance tests, the subject device serves its intended use.

§ 872.6660 Porcelain powder for clinical use.

(a)
Identification. Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.(b)
Classification. Class II.