(51 days)
This device is intended for medical purpose that is worn on the examination between patient and examiner. In addition, these gloves were tested for use with the following chemotherapy drug concentrations per ASTM D6978-05(2019): The following drugs had NO breakthrough detected up to 240 minutes:
Cisplatin, 1.0 mg/ml (1,000 ppm) >240 min.
Cyclophosphamide (Cytoxan),20.0 mg/ml (20,000 ppm) >240 min.
Dacarbazine,10.0 mg/ml (10,000 ppm) >240 min.
Doxorubicin HCI, 2.0 mg/ml (2,000 ppm) >240 min.
Etoposide, 20.0 mg/ml (20,000 ppm) >240 min.
Fluorouracil,50.0 mg/ml (50,000 ppm) >240 min.
Paclitaxel,6.0 mg/ml (6,000 ppm) >240 min.
Carmustine (BCNU) 3.3 mg/ml (3,300 ppm), 38.7 min.
ThioTepa10.0 mg/ml (10,000 ppm), 56.9 min
Please note that the following drug has low permeation times of less than 60 minutes:
Carmustine (BCNU) 3.3 mg/ml (3,300 ppm), 38.7minutes
ThioTepa10.0 mg/ml (10.000 ppm),56.9 minutes
WARNING: Do not use Carmustine and Thiotepa
Not Found
This is an FDA 510(k) clearance letter for Disposable Medical Nitrile Examination Gloves -Tested for Use with Chemotherapy Drugs. It's a Class I reserved device and, as such, the provided information does not describe an AI/ML device but rather a physical device (gloves). Therefore, many of the requested criteria regarding AI/ML study design are not applicable.
Here's the relevant information that can be extracted from the provided text, focusing on the device's performance regarding chemotherapy drug resistance:
Acceptance Criteria and Device Performance for Chemotherapy Drug Permeation
The acceptance criteria are implicitly defined by the test results indicating "NO breakthrough detected up to 240 minutes" for most listed chemotherapy drugs, in accordance with ASTM D6978-05(2019). For drugs with breakthrough times below 240 minutes, those times are reported, along with a warning for those under 60 minutes.
| Acceptance Criteria (Target Breakthrough Time per ASTM D6978-05(2019)) | Reported Device Performance (Minimum Breakthrough Detection Time) |
|---|---|
| No breakthrough for at least 240 minutes | Cisplatin, 1.0 mg/ml (1,000 ppm): >240 min. |
| No breakthrough for at least 240 minutes | Cyclophosphamide (Cytoxan), 20.0 mg/ml (20,000 ppm): >240 min. |
| No breakthrough for at least 240 minutes | Dacarbazine, 10.0 mg/ml (10,000 ppm): >240 min. |
| No breakthrough for at least 240 minutes | Doxorubicin HCI, 2.0 mg/ml (2,000 ppm): >240 min. |
| No breakthrough for at least 240 minutes | Etoposide, 20.0 mg/ml (20,000 ppm): >240 min. |
| No breakthrough for at least 240 minutes | Fluorouracil, 50.0 mg/ml (50,000 ppm): >240 min. |
| No breakthrough for at least 240 minutes | Paclitaxel, 6.0 mg/ml (6,000 ppm): >240 min. |
| Breakthrough time reported | Carmustine (BCNU) 3.3 mg/ml (3,300 ppm): 38.7 min. |
| Breakthrough time reported | ThioTepa 10.0 mg/ml (10,000 ppm): 56.9 min. |
Important Note: The document explicitly warns against using the gloves with Carmustine and ThioTepa due to their low permeation times (less than 60 minutes).
The following points are not applicable to this device as it is a physical medical device (examination gloves) and not an AI/ML powered diagnostic or therapeutic device. The requested information pertains to studies typically conducted for AI/ML performance evaluation.
- Sample size used for the test set and the data provenance: Not applicable. This refers to physical material testing, not data analysis.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Material testing to a standard like ASTM D6978-05(2019) involves laboratory procedures, not expert consensus on medical images or diagnoses.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" used here is the physical barrier performance of the glove against specific chemical permeation, determined according to the standardized laboratory methods described in ASTM D6978-05(2019). This is a scientific, objective measurement, not based on expert consensus or clinical outcomes for diagnosis.
- The sample size for the training set: Not applicable. This concept is for AI/ML models.
- How the ground truth for the training set was established: Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 31, 2022
Niujian Technology Co., Ltd % Johnson Liu Consultant Cnmed Consulting 31 Archer St Upper Mount Gravatt, QLD 4122 Australia
Re: K220353
Trade/Device Name: Disposal Medical Nitrile Examination Gloves-Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: January 24, 2022 Received: February 8, 2022
Dear Johnson Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K220353
Device Name
Disposable Medical Nitrile Examination Gloves -Tested for Use with Chemotherapy Drugs
Indications for Use (Describe)
This device is intended for medical purpose that is worn on the examination between patient and examiner. In addition, these gloves were tested for use with the following chemotherapy drug concentrations per ASTM D6978-05(2019): The following drugs had NO breakthrough detected up to 240 minutes:
| Test Chemotherapy Drugs | Minimum Breakthrough |
|---|---|
| Detection Time (Minutes) | |
| Cisplatin, 1.0 mg/ml (1,000 ppm) | >240 min. |
| Cyclophosphamide (Cytoxan),20.0 mg/ml (20,000 ppm) | >240 min. |
| Dacarbazine,10.0 mg/ml (10,000 ppm) | >240 min. |
| Doxorubicin HCI, 2.0 mg/ml (2,000 ppm) | >240 min. |
| Etoposide, 20.0 mg/ml (20,000 ppm) | >240 min. |
| Fluorouracil,50.0 mg/ml (50,000 ppm) | >240 min. |
| Paclitaxel,6.0 mg/ml (6,000 ppm) | >240 min. |
| Carmustine (BCNU) 3.3 mg/ml (3,300 ppm), | 38.7 min. |
| ThioTepa10.0 mg/ml (10,000 ppm), | 56.9 min |
Please note that the following drug has low permeation times of less than 60 minutes:
Carmustine (BCNU) 3.3 mg/ml (3,300 ppm), 38.7minutes
ThioTepa10.0 mg/ml (10.000 ppm),56.9 minutes
WARNING: Do not use Carmustine and Thiotepa
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.