(129 days)
The ALLY™ Adaptive Cataract Treatment System is an ophthalmic surgical laser indicated for use:
- · in the creation of an anterior capsulotomy.
- · in patients undergoing surgery requiring laser-assisted fragmentation of the cataractous lens.
- · in the creation of full and partial thickness single-plane and multi-plane arc cuts/incisions in the cornea.
The ALLY" Adaptive Cataract Treatment System (ALLY" System) is a medical device intended for use in ophthalmic surgery. The ALL Y™ System brings the precision of femtosecond laser to the cataract procedure. The ALL Y™ System allows for an initial femtosecond laser procedure using a dual-pulse-width laser used to cut a precision capsulotomy in the anterior lens capsule; laser-assisted fragmentation of the cataractous lens for removal during cataract surgery; and full and partial thickness single-plane and multi-plane arc cuts/incisions in the cornea. Each of which may be performed either individually or consecutively during the same procedure.
The provided text describes the LENSAR ALLY™ Adaptive Cataract Treatment System and its substantial equivalence to a predicate device, the LENSAR Laser System – fs 3D (LLS-fs 3D). However, it does not contain a specific study demonstrating performance against acceptance criteria in the way typically required for AI/ML device evaluations (e.g., diagnostic performance metrics like sensitivity, specificity, or AUC).
Instead, the document focuses on demonstrating that the ALLY™ System is substantially equivalent to its predicate device through non-clinical performance data, ensuring it meets safety and effectiveness requirements.
Here's an attempt to answer your questions based on the available information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly list acceptance criteria in the form of quantitative performance metrics (e.g., sensitivity, specificity, accuracy) for a specific task. Instead, it describes various non-clinical tests conducted to ensure the device's safety, effectiveness, and equivalence to the predicate. The "performance" here refers to meeting engineering and safety standards.
| Acceptance Criteria Category | Specific Criteria (Inferred from testing) | Reported Device Performance |
|---|---|---|
| Overall Performance | Device complies with specifications and requirements | All criteria met, device meets all performance specifications. |
| Eye Safety | Laser is safe regarding retinal and corneal illumination under worst-case normal and single failure conditions. | Conclusion: ALLY™ System laser is safe with regards to retinal and corneal illumination for defined indications, per ANSI Z-136.1:2014. |
| Biocompatibility | PID materials and PVC alignment card meet biocompatibility requirements. | All biocompatibility requirements met per ISO 10993-1. |
| Sterilization Validation | No changes to PID/PID Ring Arm sterilization methodology or sterility assurance level from predicate. Drapes kit updated and validated. | Drapes kit (same packaging as predicate's) updated and adopted into validated family of products. Implies continued compliance of sterilization. |
| Sterile Device Packaging | Packaging meets acceptance criteria. | Test results satisfied acceptance criteria per ISTA 3A, ASTM F2096, ASTM F88, and found compliant. |
| EMC & Electrical Safety | Device meets electromagnetic compatibility and electrical safety standards. | Test results satisfied acceptance criteria per ANSI/AAMI/IEC 60601-1-2:2014, ANSV/AAMI/ES 60601-1:2005 A1:2012, and found compliant. |
| Hazard Analysis | Identified hazards evaluated, mitigation measures defined and tested, residual risks acceptable. | Benefit of ALLY™ System prevails over residual risks; all potential hazards have acceptable probability/severity characteristics. |
| Software V&V | Software verifies and validates according to standards. | Compliance with IEC 62304. Documentation provided per FDA guidance. |
2. Sample Sizes and Data Provenance
- Test Set Sample Size: Not applicable in the context of diagnostic performance studies. The "test set" here refers to various samples used for non-clinical engineering and safety testing (e.g., material samples for biocompatibility, packaging samples for integrity). Specific sample sizes for each non-clinical test are not detailed in this summary.
- Data Provenance: The nature of the non-clinical tests implies that the data was generated specifically for the ALLY™ device during its development and testing phases. There is no mention of external, retrospective, or prospective patient data sets for performance evaluation in the diagnostic sense. The testing was conducted by LENSAR, Inc., likely in the US, as suggested by the FDA submission.
3. Number of Experts and Qualifications for Ground Truth
Not applicable. This submission focuses on engineering and safety validation, not on diagnostic performance or interpretation by human experts.
4. Adjudication Method
Not applicable. There is no mention of expert adjudication for ground truth for diagnostic or interpretative tasks.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted and is not mentioned in the document. The device is a surgical laser system, and the evaluation focuses on its technical performance and safety, not on improving human reader performance in interpreting medical images with AI assistance.
6. Standalone Performance Study (Algorithm Only)
No, a standalone algorithm-only performance study (in the context of AI/ML diagnostic or interpretative algorithms) was not explicitly conducted or described. The document describes the performance of the integrated ALLY™ system itself through various engineering and safety tests. While the system uses "automated precision" and "treatment planning tools for precision-guided laser treatments," the evaluation presented is not a standalone AI algorithm performance study in the typical sense of measuring diagnostic accuracy.
7. Type of Ground Truth Used
The "ground truth" for the various non-clinical tests would be the established engineering standards, safety regulations, and predetermined specifications that the device had to meet. For example:
- Eye Safety: ANSI Z-136.1:2014 standards.
- Biocompatibility: ISO 10993-1 standards.
- Sterile Device Packaging: ISTA 3A, ASTM F2096, ASTM F88 standards.
- EMC & Electrical Safety: ANSI/AAMI/IEC 60601-1-2:2014, ANSV/AAMI/ES 60601-1:2005 A1:2012 standards.
- Software: IEC 62304 standard and FDA guidance.
8. Sample Size for the Training Set
Not applicable. The document does not describe an AI/ML model that would require a distinct "training set" of data for diagnostic or interpretative learning. The "automated precision" and "treatment planning tools" are likely based on deterministic algorithms and calibrated parameters from engineering design, not deep learning from a large training dataset.
9. How the Ground Truth for the Training Set was Established
Not applicable, as no AI/ML training set is described.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 9, 2022
LENSAR, Inc. Mr. Keith Peck Director, Quality Assurance 2800 Discovery Drive, Suite 100 Orlando, Florida 32826
Re: K220259
Trade/Device Name: ALLY Adaptive Cataract Treatment System Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: OOE Dated: April 29, 2022 Received: May 2, 2022
Dear Mr. Keith Peck:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anjana Jain, PhD Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K220259
Device Name
ALLYTM Adaptive Cataract Treatment System
Indications for Use (Describe)
- The ALL Y™ Adaptive Cataract Treatment System is an ophthalmic surgical laser indicated for use:
- · in the creation of an anterior capsulotomy.
- · in patients undergoing surgery requiring laser-assisted fragmentation of the cataractous lens.
- · in the creation of full and partial thickness single-plane and multi-plane arc cuts/incisions in the cornea.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for LENSAR. The logo consists of a spherical, web-like design on the left, followed by the word "LENSAR" in a simple, sans-serif font. The color scheme is primarily light blue or teal, giving the logo a clean and modern appearance.
510(k) Summary
This 510(k) Summary has been prepared in accordance with the requirements of 21 CFR 807.92.
1.0 SUBMITTER INFORMATION
| Company/Owner: | LENSAR, Inc.2800 Discovery Drive, Suite 10Orlando, FL 32826Phone: (407) 641-4889Fax: (407) 386-7228 |
|---|---|
| Contact Person: | Keith PeckPhone: (888) 575-6412Fax: (407) 386-7228 |
Date Summary Prepared: June 8, 2022
2.0 DEVICE INFORMATION
| Trade/Proprietary Name: | ALLY™ Adaptive Cataract Treatment System |
|---|---|
| Regulation Number: | 21 CFR 886.4390 |
| Regulation Name: | Ophthalmic Laser |
| Regulatory Classification: | Class II |
| Product Code: | OOE |
| Review Panel: | Ophthalmic |
3.0 LEGALLY MARKETED DEVICES
The legally marketed devices to which LENSAR is claiming equivalence to are:
| 510(k) Number | Device Name | Manufacturer |
|---|---|---|
| K182795 | LENSAR Laser System – fs 3D (LLS-fs 3D) | LENSAR, Inc. |
DEVICE DESCRIPTION 4.0
The ALLY" Adaptive Cataract Treatment System (ALLY" System) is a medical device intended for use in ophthalmic surgery. The ALL Y™ System brings the precision of femtosecond laser to the cataract procedure. The ALL Y™ System allows for an initial femtosecond laser procedure using a dual-pulse-width laser used to cut a precision capsulotomy in the anterior lens capsule; laser-assisted fragmentation of the cataractous lens for removal during cataract surgery; and full and partial thickness single-plane and multi-plane arc cuts/incisions in the cornea. Each of which may be performed either individually or consecutively during the same procedure.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for LENSAR. The logo consists of a blue, spherical graphic on the left, followed by the word "LENSAR" in a light gray, sans-serif font. The graphic appears to be a stylized representation of a lens or a globe, with intersecting lines forming a mesh-like pattern.
Use of the ALLY"" System provides automated precision and control of the size of the capsular opening; the type and parameters of laser fragmentation treatment within the lens; the size, architecture, and location of full-thickness incisions within the cornea; and the size, architecture, location, depth, and quantity of partial thickness incisions within the cornea.
The ALLY"" System uses Streamline® to integrate pre-op analysis devices, automated iris registration with automatic cyclorotation adjustment, IntelliAxis-C® (corneal), and IntelliAxis Refractive Capsulorhexis® (lens) markers for alignment of toric IOLs, as well as treatment planning tools for precision-guided laser treatments.
INDICATIONS FOR USE 5.0
The ALLY™ Adaptive Cataract Treatment System is an ophthalmic surgical laser indicated for use:
- · in the creation of an anterior capsulotomy.
- · in patients undergoing surgery requiring laser-assisted fragmentation of the cataractous lens.
- · in the creation of full and partial thickness single-plane and multi-plane arc cuts/ incisions in the cornea.
TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES 6.0
The following table provides a comparative between the ALLY™ System and the predicate device.
| Summary of Technological Characteristics | ||
|---|---|---|
| Subject DeviceALLY™ Adaptive Cataract TreatmentSystem, 510(k) K220259 | Predicate DeviceLLS-fs 3D Laser System510(k) K182795 | |
| Characteristic | ||
| Class / Product Code /Regulation No. | Class II / OOE / 886.4390 | Class II / OOE / 886.4390 |
| Intended Type of Use | Prescription Use | Prescription Use |
| Indications for Use | The ALLY™ Adaptive Cataract TreatmentSystem is an ophthalmic surgical laserindicated for use:in the creation of an anterior capsulotomy. in patients undergoing surgery requiringlaser-assisted fragmentation of thecataractous lens. in the creation of full and partial thicknesssingle-plane and multi-plane arc cuts/incisions in the cornea. | The LENSAR Laser System – fs 3D(LLS-fs 3D) with Streamline™ is anophthalmic surgical laser indicated for use:in the creation of an anterior capsulotomy. in patients undergoing surgery requiringlaser-assisted fragmentation of thecataractous lens. in the creation of full and partial thicknesssingle-plane and multi-plane arc cuts/incisions in the cornea. in patients undergoing ophthalmic surgeryor other treatments requiring pocketcuts/incisions in the cornea. in the creation of a corneal flap in patientsundergoing treatment requiring initiallamellar resection of the cornea. in patients undergoing surgery or othertreatment requiring initial lamellar |
| Summary of Technological Characteristics | ||
| Subject Device | Predicate Device | |
| Characteristic | ALLY™ Adaptive Cataract TreatmentSystem, 510(k) K220259 | LLS-fs 3D Laser System510(k) K182795 |
| resection of the cornea to create tunnelsfor placement of corneal ring segments.• in the creation of partial thickness single-plane radial cuts/incisions in the cornea. | ||
| Operating Principle | Photodisruption and plasma mediatedablation | Photodisruption and plasma mediatedablation |
| Patient ContactPortion of Device (PID) | Disposable polycarbonate/silicone patientinterface accessory; suction ring assembly | Disposable polycarbonate/silicone patientinterface accessory; suction ring assembly |
| Sterilization Method(Accessories) | Ethylene Oxide (EO) | Gamma irradiation (not including CurvedContact PID) or Ethylene Oxide (EO) |
| Laser Type | Yb:YAG | Yb:YAG |
| Wavelength | 1030 nm | 1030 nm |
| Pulsing | Mode-locked | Mode-locked |
| Mode | Fundamental (TEM00) | Fundamental (TEM00) |
| Pulse Duration | 320 - 1500 fs | $1.5 x10^{-12}$ sec |
| Repetition Rate | Variable (320 kHz max) | 20-80 kHz, adjustable |
| Maximum PulseEnergy | 15 μJ | 15 μJ |
| Biometrics | Scheimpflug Camera-based Imaging Systemwith 3D Confocal Structured Illumination(3D-CSI) | Scheimpflug Camera-based Imaging Systemwith 3D Confocal Structured Illumination(3D-CSI) |
| Beam Targeting andDelivery | Computer-controlled scanner with F-thetalens | Computer-controlled scanner with F-thetalens |
| PhotodisruptiveFluence Threshold inLens Tissue | 4 J/cm² | 4 J/cm² |
| Intended Use for IrisRegistration Feature | Same as predicate device (K182795). | • System provides a means to assist thesurgeon during anterior segmentophthalmic surgical procedures, to makean incision.• Using standard pre-operative clinical data,together with surgeon-driven, onscreentemplates and guides, the systemprovides graphical assistance to thesurgeon as desired during the surgery.• The system utilizes surgeon confirmationat each step for planning and positioningof guidance for incision placement.• The ocular rotation difference betweenthe upright position (as observed duringpreoperative measurements) and thesupine position (as observed during thesurgical event) is determined either bymanually marking the eye in the uprightposition and allowing the surgeon to lineup the pre-marked eye in the uprightposition to that of the supine position by |
| Summary of Technological Characteristics | ||
| Characteristic | Subject Device | Predicate Device |
| ALLY™ Adaptive Cataract TreatmentSystem, 510(k) K220259 | LLS-fs 3D Laser System510(k) K182795 | |
| Intended Use for theCataract DensityImaging Feature | Same as predicate device (K182795). | System's ability to analyze and match iris features between images acquired in each position. |
| The imaging system finds lens surfaces in order to correctly place pulses into the lens. Upon finding the lens surfaces, the imaging system also determines the location and density/degree of the nuclear cataract within the lens. The surgeon then has the ability to pick from a variety of fragmentation patterns as the surgeon sees fit. Additionally, the pattern itself can be altered in size and thereby the number of shots as the surgeon sees fit. | ||
| The system provides a mechanism for the doctor to change the size and type of pattern as the surgeon sees fit. For convenience, the System is also provided with an optional mechanism for intra-operatively matching the nuclear cataract condition to one of five categories for surgeon pre-selected fragmentation (CustomFrag). | ||
| The system utilizes surgeon confirmation for planning and positioning of guidance for laser shot placement. | ||
| Intended Use for theIntelliAxis RefractiveCapsulorhexis®Feature | Same as predicate device (K182795). | Using standard pre-operative data, the system places laser marks onto the lens capsule as part of the capsulotomy to identify the astigmatic axis as prescribed by the surgeon. |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for LENSAR. The logo consists of a blue, spherical, grid-like design on the left, followed by the word "LENSAR" in a light gray, sans-serif font. The text is aligned horizontally with the spherical design.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for LENSAR. The logo consists of a blue, spherical, geometric design on the left, resembling a stylized globe or network. To the right of the design is the word "LENSAR" in a light gray, sans-serif font. The logo appears clean and modern.
7.0 DETERMINATION OF SUBSTANTIAL EQUIVALENCE
7.1 Non-Clinical Performance Data
Design verification and validation testing was completed to demonstrate that the proposed device performance complies with specifications and requirements identified for the ALLY" Adaptive Cataract Treatment System. Each function and/or feature was verified through unit testing and system testing by means of the appropriate test case or test specification. The unit/system verification test reports provide the test cases, expected results for each test case, and the actual results obtained. All criteria for this testing were met and the results demonstrate that the ALLY™ Adaptive Cataract Treatment System meets all performance specifications and requirements. The objectives defined in the validation plan were achieved according to the validation results.
The comparison to the predicate device shows that the ALLY™ Adaptive Cataract Treatment System is substantially equivalent. The minor differences between the additional ALLY" Adaptive Cataract Treatment System device features and the predicate device do not raise any new questions of safety or effectiveness.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for LENSAR. The logo consists of a spherical graphic to the left of the company name. The spherical graphic is made up of many small squares. The company name, LENSAR, is written in a sans-serif font.
7.2 Eye Safety Analysis Testing
A detailed analysis of the retinal and corneal burn hazard analysis for capsulotomy, fragmentation, clear corneal and arcuate incisions under worst-case normal and single failure conditions was performed in accordance with ANSI Z-136.1:2014, with the conclusion that the ALLY™ Adaptive Cataract Treatment System laser is safe with regards to retinal and corneal illumination for the defined indications.
7.3 Biocompatibility Testing
Minor changes to materials used in the ALLY™ System Patient Interface Device (PID) and the addition of the PVC alignment card that is temporarily placed on the patient's head to facilitate system alignment underwent biocompatibility testing per ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process (3 ed.). Test results indicated that the PVC alignment card and the minor materials changes met all biocompatibility requirements.
7.4 Sterilization Validation
There have been no changes to the ALLY™ System PID or PID Ring Arm as detailed and previously cleared with the LLS-fs 3D (K182795) to the sterilization provider, sterilization methodology, profile/chamber, or sterility assurance level.
The ALL Y™ System includes a sterile drapes kit (used for monitors, handles, joystick, tray; and an alignment card). This drapes kit has the same packaging as the drapes kit previously cleared for the predicate device in K182795. The drapes for the ALLY 100 System have been updated and were adopted into the family of products already validated.
7.5 Sterile Device Packaging
Packaging for the sterile Patient Interface Device (PID) and Drapes Kit has been performed. Test results satisfied the acceptance criteria as defined by the recognized standards (ISTA 3A, ASTM F2096, ASTM F88) and found to be in compliance with the recognized standards.
7.6 Electromagnetic Compatibility and Electrical Safety
The ALLY™ Adaptive Cataract Treatment System has undergone electromagnetic compatibility and electrical safety testing to support this premarket notification. Test results satisfied the acceptance criteria as defined by the recognized standards (ANSI/AAMI/IEC 60601-1-2:2014, ANSV/AAMI/ES 60601-1:2005 A1:2012) and found to be in compliance with the applicable recognized safety standards.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for LENSAR. The logo consists of a blue, spherical, geometric design on the left, resembling a stylized globe or network. To the right of the design is the word "LENSAR" in a light blue, sans-serif font. The overall impression is modern and corporate.
7.7 Hazard Analysis
A hazard analysis was conducted to identify all potential hazards to the patient, surgeon, and other system operators resulting from differences between the ALL Y" Adaptive Cataract Treatment System and the predicate device, LLS-fs 3D (K182795). The identified hazards were duly evaluated in the risk analysis, and mitigation measures were defined and tested. After the implementation and testing of mitigation measures, the benefit of the ALLY™ Adaptive Cataract Treatment System prevails over the residual risks and all potential hazards have acceptable levels of probability/severity characteristics.
7.8 Software Verification and Validation Testing
A complete software verification and validation testing of the ALLY™ Adaptive Cataract Treatment System was conducted. The results demonstrate that the ALLY™ Adaptive Cataract Treatment System is in compliance with the following recognized standard: IEC 62304. Documentation was provided as recommended in Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices: Guidance for Industry and FDA Staff (May 11, 2005). The software for this device was considered as a "major" level of concern since a failure or latent flaw in the software could directly result in serious injury to the patient or operator.
7.9 Animal Studies Performance Data
No animal studies were performed to support substantial equivalence of this premarket notification.
7.10 Clinical Performance Data
No clinical studies were deemed necessary to demonstrate the safety and effectiveness or substantial equivalence of the ALLY™ Adaptive Cataract Treatment System to the predicate device. The non-clinical performance data demonstrated that the device performs as intended.
8.0 CONCLUSIONS FROM NON-CLINICAL AND CLINICAL DATA
The ALLY" Adaptive Cataract Treatment System has the same intended use, indications for use, and similar design and fundamental and scientific technology as the predicate device. The non-clinical performance data demonstrate that the ALLY™ Adaptive Cataract Treatment System is substantially equivalent to the predicate device and does not raise any new issues of safety or effectiveness. Thus, the ALLY " Adaptive Cataract Treatment System is substantially equivalent to the predicate device.
§ 886.4390 Ophthalmic laser.
(a)
Identification. An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.(b)
Classification. Class II.