K Number
K220239
Manufacturer
Date Cleared
2022-09-13

(228 days)

Product Code
Regulation Number
892.1680
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VIVIX-S 4386W is indicated for digital imaging solution designed as a general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purposes of diagnostic procedures. It is not to be used for mammography. It is intended for both adult and pediatric populations.

Device Description

VIVIX-S 4386W, a flat panel detector model named; FXRD-4386WB with imaging areas of 43cm x 86cm. The device intercepts x-ray photons and the scintillator emits visible spectrum photons that illuminate an array of photo (a-SI)-detectors that create electrical signals. After the electrical signals are generated, it is converted to digital value, and the Software which acquires and processes the data values from the detector. The resulting digital images will be displayed on monitors. These devices should be integrated with an operating PC and an X-Ray generator. It can be utilized to digitalize x-ray images and transfer for radiography diagnostic.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the VIVIX-S 4386W device, based on the provided text:

Acceptance Criteria and Reported Device Performance

The provided text focuses on demonstrating substantial equivalence to a predicate device (VIVIX-S 1717V, K181003) rather than defining specific acceptance criteria for de novo performance. The acceptance criteria implicitly are that the device performs as well as or better than the predicate device for relevant metrics, and that any differences do not raise new safety or effectiveness concerns.

The comparison table below highlights the performance metrics reported and their relationship to the predicate device.

ParameterPredicate Device (VIVIX-S 1717V)Subject Device (VIVIX-S 4386W)Equivalence Assessment (from text)
MTF (1lp/mm)FXRD-1717VA: Horizontal 66, Vertical 67; FXRD-1717VB: Horizontal 58, Vertical 5862Substantially Equivalent (Stated "higher than the one of the predicate device models" for FXRD-1717VB, which is consistent with 62 vs 58)
DQE (1lp/mm)FXRD-1717VA: Horizontal 53, Vertical 54; FXRD-1717VB: Horizontal 29, Vertical 2928Substantially Equivalent (Stated "little lower" than the predicate device FXRD-1717VB, explicitly noting "not significant (just one 1)p/mm value difference)")
Resolution3.5 lp/mm3.5 lp/mmEquivalent
Spatial Resolution3.5 lp/mmMin. 3.5 lp/mmEquivalent
Field of View17 x 17 inch17 x 34 inchSimilar (Larger)
Pixel Pitch0.14mm0.14mmEquivalent
Weight4.5kg8kg (With two battery packs)Equivalent (Though numerically different, likely considered within acceptable range for a larger form factor)
Materials ScintillatorCsl: TI, Gd2O2S:TbGd2O2S:TbSubstantially Equivalent
Communication MethodWiredWired / WirelessSubstantially Equivalent (Added Wireless capability)
Image Acquisition Time (Wired)1.5 secMax. 5 sec. (Exposure time is set to 500ms, Excluding exposure time)Similar (Longer, but likely deemed acceptable given overall performance)
Active Area430.08 x 430.08 (mm)430.08 mm × 860.16 mm; 430.08 mm × 430.08 mmSubstantially Equivalent (Larger option available)
Active Array3072 x 3072 pixels3072 x 6144 pixels; 3072 x 3072 pixelsSubstantially Equivalent (Larger option available)
Method of Generator InterfaceDR Trigger (External Line Trigger); AED (Auto Exposure Detection); Passive Trigger (External Line Trigger)DR Trigger (External Line Trigger); AED (Auto Exposure Detection); Software TriggerSubstantially Equivalent (Added Software Trigger)

Study Details

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: The document does not explicitly state the number of images or cases used in the clinical study. It only mentions that "a single-blinded concurrence study was conducted."
    • Data Provenance: Not specified (e.g., country of origin). The study is described as a "concurrence clinical study," implying prospective data collection for the purpose of the study, but this is not definitively stated.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified (e.g., radiologist with 10 years of experience).
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Adjudication Method: Not specified. The study is described as "single-blinded concurrence study," which suggests readers were evaluating images without knowing if they were from the subject or predicate device, but the method for reconciling disagreements or establishing a consensus ground truth is not detailed.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: Yes, a "single-blinded concurrence study" was conducted comparing the subject device with the predicate. This is a form of comparative effectiveness study.
    • Effect Size (AI Assistance): Not applicable. This study evaluates a digital X-ray detector (hardware plus basic image processing software), not an AI-assisted diagnostic tool. The study aims to show equivalence between the subject device and the predicate device in terms of diagnostic capability, not to demonstrate improvement with AI assistance. The document states, "the study confirmed that the new VIVIX-S 4386W provides images of equivalent diagnostic capability to the predicate devices."
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, non-clinical performance evaluations (MTF, DQE, spatial resolution) were conducted in a standalone manner to assess the physical performance characteristics of the device. However, the "clinical study" aspect would inherently involve human readers to assess "diagnostic capability." The term "standalone" in the context of diagnostic algorithms typically refers to the algorithm making a direct diagnostic output without human review, which is not what this device does. This device is a diagnostic imaging system component, and its images are intended for human interpretation.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The ground truth for the clinical study is based on expert assessment of diagnostic capability via comparison of images from the subject device to those from the predicate device. The document states "provides images of equivalent diagnostic capability." It does not specify if an independent "ground truth" (e.g., pathology reports, follow-up outcomes) was established for the cases themselves, but rather the relative diagnostic utility of the images produced by the two devices.
  7. The sample size for the training set:

    • Training Set Sample Size: Not applicable. This document describes a medical device (digital X-ray detector) and its associated image acquisition and processing software. It is not an AI/ML algorithm that typically requires a large, annotated training set in the conventional sense for learning specific diagnostic tasks. The "training" for such a system involves engineering and calibration to produce high-quality images.
  8. How the ground truth for the training set was established:

    • Ground Truth for Training Set: Not applicable for the reasons mentioned above. The development of the VIVIX-S 4386W involves engineering design, physical measurements, and quality control, rather than machine learning training with labeled data.

{0}------------------------------------------------

September 13, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Vieworks Co., Ltd. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 18881 Von Karman Ave., STE 160 IRVINE CA 92612

Re: K220239

Trade/Device Name: VIVIX-S 4386W Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: MQB Dated: August 10, 2022 Received: August 12, 2022

Dear Priscilla Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220239

Device Name VIVIX-S 4386W

Indications for Use (Describe)

VIVIX-S 4386W is indicated for digital imaging solution designed as a general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purposes of diagnostic procedures. It is not to be used for mammography. It is intended for both adult and pediatric populations.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary («220239)

[As required by 21 CFR 807.92]

This 510(k)-summary information is prepared in accordance with 21 CFR807.92

1. Date Prepared [21 CFR 807.92(a) (1)]

09/06/2022

2. Submitter's Information [21 CFR 807.92(a) (1)]

Name of Sponsor: Vieworks Co., Ltd. Address: 41-3, Burim-ro 170beon-gil, Dongan-gu, Anyang-si, Gyeonggido, 14055 Republic of Korea Contact Name: Oh, Kevin / Regulatory Affairs Associate Registration Number: 3006013411 Name of Manufacturer: Same as Sponsor

3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]

Trade Name:VIVIX-S 4386W
Model Name:FXRD-4386WB
Common Name:Digital Flat Panel X-ray Detector
Classification Name:Solid State X-Ray Imager(Flat Panel/Digital Imager)
Classification Panel:Radiology
Classification Regulation:21 CFR 892.1680
Product Code:MQB
Device Class:2

4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]

510(k) Number: K181003 Product Code: MQB Applicant: Vieworks Co., Ltd. Trade Name: VIVIX-S 1717V Decision Date: 03/15/2018 Type: Traditional

5. Description of the Device [21 CFR 807.92(a) (4)]

General Description

VIVIX-S 4386W, a flat panel detector model named; FXRD-4386WB with imaging areas of 43cm x 86cm. The device intercepts x-ray photons and the scintillator emits visible spectrum photons that illuminate an array of photo (a-SI)-detectors that create electrical signals. After the electrical signals

{4}------------------------------------------------

are generated, it is converted to digital value, and the Software which acquires and processes the data values from the detector. The resulting digital images will be displayed on monitors. These devices should be integrated with an operating PC and an X-Ray generator. It can be utilized to digitalize x-ray images and transfer for radiography diagnostic.

6. Indications for Use [21 CFR 807.92(a)(5)]

VIVIX-S 4386W is indicated for digital imaging solution designed as a general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purposes of diagnostic procedures. It is not to be used for mammography. It is intended for both adult and pediatric populations.

7. Technological Characteristics [21 CFR 807.92(a) (6)]

Comparisons with the predicate, the system shows the technological characteristics of the proposed device to be substantially equivalent to the predicate devices. The proposed devices are functionally identical to the predicate devices.

8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]

When compared to the predicate devices (K181003), the VIVIX-S 4386W presented in this submission has the subsequently equivalent:

  • Intended Use
  • Technological characteristics
  • Operating principle
  • · Design features
  • Communication Method

The MTF value of the subject device is higher than the one of the predicate device models (FXRD-1717VB) but DQE value of is little lower than the predicate device. However, the difference is not significant (just one 1)p/mm value difference), and the test result of the concurrence clinical study shows that this difference does not raise an issue in substantial equivalence.

Based on the information and the test result we submitted, we conclude that the subject device is substantially equivalent to these devices in design, function, materials, operational principles and intended use.

ParameterPredicate DeviceSubject DeviceEquivalence
510(k) NumberK181003K220239-
ManufacturerVieworks Co., Ltd.-
Device NameVIVIX-S 1717VVIVIX-S 4386W-
Compo-nentDetectorFXRD-1717VA,FXRD-1717VBFXRD-4386WB-
SCU-FXRS-04APower Supply
SoftwareVXvueVXvueEquivalent
Common NameDigital Flat Panel X-ray DetectorEquivalent
Classification NameSolid State X-Ray Imager (Flat Panel/Digital Imager)Equivalent
Classification PanelRadiologyEquivalent
Classification Regulation21 CFR 892.1680Equivalent
Product CodeMQBEquivalent
Device Class2Equivalent
Indications for UseVIVIX-S 1717V series isindicated for digitalimaging solution designedas a general radiographicsystem for humananatomy. It is intended toreplace film or screenbased radiographicsystems in all generalpurposes of diagnosticprocedures. It is not to beused for mammography.VIVIX-S 4386W isindicated for digitalimaging solutiondesigned as a generalradiographic system forhuman anatomy. It isintended to replace filmor screen basedradiographic systems inall general purposes ofdiagnostic procedures. Itis not to be used formammography. It isintended for both adultand pediatricpopulations.Equivalent
Panel ShapeSquare PanelSquare PanelEquivalent
Field of View17 x 17inch17 x 34inchSimilar
DesignDimensions(H x W x D)460.0 x 460.0 x15.5mm900 mm × 465 mm ×17.0 mmEquivalent
Pixel Pitch0.14mm0.14mmEquivalent
Weight4.5kg8kg (With two batterypacks)Equivalent
Materials ScintillatorCsl: TI,Gd2O2S:TbGd2O2S:TbSubstantiallyEquivalent
Image sensorTFT: a-Si (AmorphousSilicon)TFT: a-Si (AmorphousSilicon)Equivalent
DataTransmis-sionWiredMax. 1GbpsMax. 1GbpsEquivalent
Communication MethodWiredWired / WirelessSubstantiallyEquivalent
Perform-anceMTF(1lp/mm)FXRD-1717VA:Horizontal 66Vertical 67FXRD-1717VB:Horizontal 58Vertical 58FXRD-4386W:62SubstantiallyEquivalent
DQE(1lp/mm)FXRD-1717VA:Horizontal 53Vertical 54FXRD-4386W:28SubstantiallyEquivalent
FXRD-1717VB:Horizontal 29Vertical 29
Resolution3.5 lp/mm3.5 lp/mmEquivalent
Active Area430.08 x 430.08 (mm)430.08 mm × 860.16 mm430.08 mm × 430.08 mmSubstantiallyEquivalent
Pixel Size140 μm140 μmEquivalent
Grayscale16 bit16 bitEquivalent
ImageAcquisitionTimeWired1.5 secMax. 5 sec.(Exposure time is set to500ms, Excludingexposure time)Similar
Wireless• Max. 5 sec.(IEEE802.11ac, MiMO3x3, 5GHz, 80MHz)(Exposure time is set to500ms, Excludingexposure time)
Active Array3072 x 3072 pixels3072 x 6144 pixels3072 x 3072 pixelsSubstantiallyEquivalent
Method of GeneratorInterface• DR Trigger (ExternalLine Trigger)• AED (Auto ExposureDetection)• Passive Trigger(External Line Trigger)• DR Trigger (External LineTrigger)• AED (Auto ExposureDetection)• Software TriggerSubstantiallyEquivalent
Spatial Resolution3.5lp/mmMin. 3.5lp/mmEquivalent
TechnologyStructured scintillatoris attached in thedetector to increasethe X-ray photoninteraction, X-rayabsorption, and theamount of availablevisible light. Thisenhances potentialdetective quantumefficiency and partialresolution.Structured scintillatoris attached in thedetector to increasethe X-ray photoninteraction, X-rayabsorption, and theamount of availablevisible light. Thisenhances potentialdetective quantumefficiency and partialresolution.Equivalent
Principle of OperationGenerated X-rayphotons strike thescintillator causing thevisible light emissionthat is to be convertedinto an electric chargeby the photodiodearray. The electricGenerated X-rayphotons strike thescintillator causing thevisible light emissionthat is to be convertedinto an electric chargeby the photodiodearray. The electricEquivalent
charge is thencollected at eachphotodiode and turnsinto a digital value byusing the underlyingreadout electronics.charge is then collectedat each photodiodeand turns into a digitalvalue by using theunderlying readoutelectronics.

Table 1: Comparison of Subject Device VIVIVX-S 4386W to the Predicate

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

9. Summary of Non-Clinical Data

A comparison test was conducted between the subject devices (VIVIX-S 4386W) and the predicate device (K181003) on the items such as DQE, MTF and spatial resolution.

These detectors comply with the following international and FDA-recognized consensus standards:

  • 21CFR1020.30, Diagnostic X-ray Systems and their major components
  • . 21CFR1020.31, Radiographic equipment
    • IEC 60601-1 Medical Electrical Equipment -- Part 1: General Requirements for Basic Safety and Essential Performance.
  • . CAN/CSA-C22.2 No. 60601-1 (2008) (Medical Equipment --Part 1 : General Requirements for Basic Safety and Essential Performance) (includes National Differences for Canada)
  • ANSI/AAMI ES60601-1 (2005+ C1:09+A2:10) (Medical Electrical Equipment – Part 1
  • IEC 60601-1-2 Medical Electrical Equipment - Part 1-2 : General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests

10. Summary of Clinical Data

A single-blinded concurrence study was conducted, and the study confirmed that the new VIVIX-S 4386W provides images of equivalent diagnostic capability to the predicate devices, the VIVIX-S 1717V and its results demonstrate substantial equivalence.

11. Conclusion [21 CFR 807.92(b) (3)]

The VIVIX-S 4386W is substantially equivalent to the currently marketed and predicate devices (K181003) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.

Additionally, Substantial equivalence was demonstrated through the non-clinical performance, which complied with the requirements specified in the international and FDA-recognized consensus standards, IEC60601-1, IEC 60601-1-2, 3 and the clinical test.

The results of these tests demonstrate that VIVIX-S 4386W meets the acceptance criteria and is adequate for this intended use. The comparison of technological characteristics, non-clinical performance data, safety testing, and clinical image concurrence data demonstrates that the device is as safe, as effective, and performs as well the predicate devices.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.